Open Access 10.02.2021 | Technical Innovation

Real-time magnetic resonance imaging in pediatric radiology — new approach to movement and moving children

verfasst von: Franz Wolfgang Hirsch, Jens Frahm, Ina Sorge, Christian Roth, Dirk Voit, Daniel Gräfe

Erschienen in: Pediatric Radiology | Ausgabe 5/2021

Abstract

The recent development of highly undersampled radial gradient echo sequences in combination with nonlinear inverse image reconstruction now allows for MRI examinations in real time. Image acquisition times as short as 20 ms yield MRI videos with rates of up to 50 frames per second with spin density, T1- and T2-type contrast. The addition of an initial 180° inversion pulse achieves accurate T1 mapping within only 4 s. These technical advances promise specific advantages for studies of infants and young children by eliminating the need for sedation or anesthesia. Our preliminary data demonstrate new diagnostic opportunities ranging from dynamic studies of speech and swallowing processes and body movements to a rapid volumetric assessment of brain cerebrospinal fluid spaces in only few seconds. Real-time MRI of the heart and blood flow can be performed without electrocardiogram gating and under free breathing. The present findings support the idea that real-time MRI will complement existing methods by providing long-awaited diagnostic options for patients in early childhood. Major advantages are the avoidance of sedation or anesthesia and the yet unexplored potential to gain insights into arbitrary body functions.
Hinweise
A scientific study on the clinical use of real-time MRI on infant head has been published in 2020: Gräfe D, et al., Outpacing movement - ultrafast volume coverage in neuropediatric magnetic resonance imaging, https://​doi.​org/​10.​1007/​s00247-020-04771-5.

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Introduction

Recent advances in dynamic MRI allow for acquisition times of serial images of 20–30 ms, which correspond to MRI videos with rates of 30–50 frames per second [1]. While preliminary clinical applications suggest new diagnostic opportunities as well as challenges for radiologic practices [2], no studies in the field of pediatric radiology have been reported. This is surprising in view of several specific advantages for MRI examinations in early childhood. In this first overview article we summarize our initial experience gathered since September 2019 in a specialized department for pediatric radiology. To allow the reader to fully appreciate the information of an MRI video even from the printed form of a scientific journal, we have chosen a new format for Pediatric Radiology that provides access to video sequences by scanning a QR Code with any personal computer, tablet or smartphone. A first example (Video 1) shows online navigation of the imaging plane by interactive real-time MRI at 30 frames per second, while planning a kidney examination of a 10-year-old boy with urine transport disturbance. Localizer scans serve to freely position the desired cross-section with use of the scanner mouse.
Video 1

Description

The study was approved by the local ethics committee. Informed consent was obtained from patients or legal guardians. Patients or legal guardians signed informed consent regarding publishing their data and images.

Technical prerequisites

According to its definition, real-time MRI combines serial high-speed acquisitions with immediate reconstruction and display of the images with minimum latency. In our implementation this was accomplished by adding a dedicated computer with eight graphical processing units to a standard 3.0-tesla (T) MRI system (Prismafit; Siemens, Erlangen, Germany). The real-time environment is currently only available for Siemens 3-T MR scanners. The measuring protocols appear on the user interface of the scanner like on any other vendor sequence and the resulting image series occur in the usual databank as regular Digital Imaging and Communications in Medicine (DICOM) images.
The real-time MRI technique used here is based on highly undersampled radial gradient echo sequences in conjunction with a joint reconstruction of images and coil sensitivity maps by nonlinear inversion with temporal regularization to the preceding frame; for details see Uecker et al. [1]. The high temporal fidelity of the method has been experimentally confirmed [3]. Depending on radiofrequency flip angles and gradient schemes, the typical contrast opportunities of gradient echo sequences comprise spin density and T1 contrast as well as steady-state free precession (SSFP) T2-type contrast also known as T2/T1 contrast. Figure 1 shows a selected frame from a T1-weighted real-time MRI study of the heart of a 5-month-old girl. In general, all real-time MRI sequences comply with standard regulations for radiofrequency power deposition and peripheral nerve stimulation. In fact, fast low-angle shot (FLASH)-type acquisitions result in very low specific absorption rate (SAR) values because of their use of small flip angles.

Suitable real-time magnetic resonance imaging methods for pediatric radiology

The availability of images with very short acquisition times and MRI videos with correspondingly high frame rates can be diagnostically exploited for several examinations and situations. Here we present three approaches to rapid MRI diagnostics in pediatric radiology.
A first method directly focuses on the functional anatomy, or, in other words, on the monitoring of arbitrary, i.e. also aperiodic, dynamic processes. For example, this holds true for studies of the oral cavity during speaking [4] or swallowing [5, 6] and also applies to joint movements [7]. Moreover, in contrast to conventional electrocardiogram (ECG)-gated cine techniques, which retrospectively merge data from multiple heartbeats into a single cardiac cycle, myocardial movements can now be recorded in real time. These applications have been extended to real-time phase-contrast flow MRI [8, 9]. Multi-slice real-time MRI videos can be obtained in a sequential manner.
A second method refers to quantitative tissue characterization, more precisely to rapid T1 mapping. This is accomplished by a brief FLASH-type real-time MRI acquisition of typically 4 s, preceded by a 180° inversion pulse. The resulting image series defines a single-shot inversion-recovery dataset that has been shown to yield highly accurate T1 maps by pixel-wise fitting [10, 11]. When using slice-selective inversion pulses, this method can be extended to multi-slice T1 mapping of whole organs (i.e. 4 s per slice).
A third method allows for rapid scanning of a large volume by advancing the slice position of each frame of a cross-sectional real-time MRI study by a certain percentage of the slice thickness [12]. The resulting spatial overlap generates sufficient similarity of successive, i.e. neighboring, sections and therefore allows for the same nonlinear inverse reconstruction as for dynamic real-time MRI. Because of the short acquisition times, individual frames are robust against motion, which is in contrast to conventional 3-D MRI techniques. Typically, volume coverage of a child’s brain with 200–300 overlapping images can be performed in a few seconds per orientation.

Body movements

Until now, it has not been possible to dynamically evaluate aperiodic body movements such as, for example, the mobility of the cervical spine. Video 2 depicts a sagittal T1-weighted real-time MRI recording of a 12-year-old boy with paresthesia in the right arm and suspected vertebral gliding; the video demonstrates a continuous motion without constriction of the craniocervical junction. The movie further reveals the pulsatile nature of the cerebrospinal fluid (CSF) dynamics, which is important in a variety of pathological conditions including Chiari malformation.
Similarly, conventional MRI techniques cannot visualize post-traumatic joint constellations, which for children means they eventually require X-ray fluoroscopy. In contrast, Video 3 shows a coronal T1-weighted real-time MRI recording of a 3-year-old girl without sedation. The examination served as a postoperative control after repositioning of the left hip and the video easily documents correct position and movement of the femoral head. Abduction and adduction of the leg is performed by a person lying in prone position on the patient table.

Speaking, swallowing and beyond

Monitoring the complex motility of tongue, velum and soft palate during speaking is critically important in logopedics as well as for pre-surgical planning and post-surgical evaluation of speech problems. The case shown in Video 4 refers to a child who learned to correctly vocalize words (e.g., “krokodil,” the German equivalent of “crocodile”) after surgical fixation of the velum to the posterior wall of the oral cavity.
Swallowing processes can be studied by voluntary swallows of pineapple juice because it contains an inherent concentration of paramagnetic manganese, which serves as an MRI contrast agent for T1-weighted images. This is demonstrated in Video 5 in a healthy subject. However, this diagnostic can play a role in all coordination disorders of the swallowing process for which fluoroscopy has been used, as well as in the diagnosis of gastroesophageal reflux [13]. Children after esophageal atresia should also benefit from MRI swallowing studies.
Evaluation of the filling of the stomach and passage of the fluid into the small bowel is another diagnostic option, analogous to X-ray fluoroscopy but without ionizing radiation and access to 3-D information. The case shown in Video 6 represents a coronal T1-weighted volume coverage scan of a 6-week-old boy with suspected anastomotic stricture after duodenojejunostomy in duodenal atresia, obtained within 11 s without sedation.

Cardiac magnetic resonance imaging

Cardiac MRI studies in real time appear as a most natural and obvious application in the field of pediatrics. Such measurements do not require ECG or self-gating techniques and can be performed during free breathing without the use of navigator scans. Moreover, when based on FLASH-type T1-weighted versions at 3.0 T, the short gradient-echo times do not require time-consuming shimming procedures, while completely avoiding SSFP-inherent banding artifacts. Preliminary applications in adults have focused on those with arrhythmias [14], where ECG synchronization is barely possible and commonly leads to poor image quality. Here, Video 7 illustrates the case of a 5-day-old boy with interrupted aortic arch where the slice position is automatically shifted every 3 s to cover a larger volume in sequential recordings. Despite pronounced movements of the child, the wall of the thorax and abdomen appear without blurring. The second example in Video 8 is an 8-month-old boy with right aortic arch and aberrant left subclavian artery leading to alterations of the distal trachea and right main bronchus, while the case in Video 8 refers to a 17-year-old with funnel chest. The application of a vacuum bell at increasingly negative pressure demonstrates the elevation of the sternum with decompression of the heart.

Phase-contrast flow magnetic resonance imaging

Access to blood flow in real time is a logical extension of anatomical real-time MRI and requires the adoption of phase-contrast flow MRI principles to determine velocity maps for through-plane flow [8]. Extending conventional phase-contrast flow MRI, the use of a model-based reconstruction eliminates the “salt-and-pepper” noise in regions such as the lung that are without MRI signal support, and it further improves the spatio-temporal resolution [9]. Figure 2 presents the results of a real-time flow MRI study of the ascending aorta in the same child as in Video 9. The true time courses of flow volume over multiple heartbeats reveal respiration-dependent variations. Clinically relevant evaluations in pediatric cardiology are the determination of flow volumes (or rates) of the large thoracic vessels in people with arrhythmias as well as in young anxious children.
In addition to blood flow measurements at high velocities, real-time flow MRI techniques provide access to the low-velocity dynamics of the CSF. Such studies in healthy adults recently demonstrated respiration to be a major regulator of spinal CSF flow mildly modulated by cardiac pulsatility [15].

Rapid T1 mapping

The ability to obtain quantitative maps of absolute T1 relaxation times in arbitrary tissues within only 4 s per map emerges as an attractive tool that might serve as a biomarker. Such studies are possible on all parenchymatous organs and can contribute to tumor characterization in the context of multiparametric measurements. For example, T1 maps of age-dependent differences in white matter during infancy and childhood can be used to assess delayed myelination in an objective manner. Figure 3 presents a color-coded T1 map of a 6-year-old boy with pilocytic astrocytoma, where high T1 values (red=2,800 ms) and low T1 values (blue=480 ms) distinguish cystic tumor compartments from solid tissues. The use of a slice-selective inversion pulse extends the measurement to an automatic multi-slice acquisition covering the entire brain. Video 10 depicts such a series of 35 contiguous T1 maps (4 s each) of the brain of a 5-year-old girl with Neurofibromatosis Type 1, revealing multiple hamartomas, especially within the left globus pallidus. The total scan time of 2 min 20 s was much shorter than achievable with conventional methods.

Rapid volume coverage

In pediatric radiology the need for volumetric assessments, for example by 3-D MRI with measuring times of minutes, is often compromised by involuntary movements or requires the use of sedation or anesthesia. The new volume coverage scan sequences by real-time MRI with automatic slice advancement overcome this problem: the method is not only motion-robust because individual frames are acquired within tens of milliseconds, but it is also fast in scanning large fields-of-view with overlapping images [12]. Figure 4 demonstrates the reduced sensitivity to motion in a 2.3-year-old boy after intracranial bleeding and ventriculo-peritoneal shunt by comparing a selected frame from a T2/T1-weighted volume coverage scan with a conventional T2-weighted fast spin-echo image. A full evaluation of the brain in all three orientations is shown in Video 11 for the case of a 4-year-old girl with Blake pouch cyst. Each of the transverse, coronal and sagittal T2/T1-weighted volume coverage scans with up to 300 overlapping images (3-mm section thickness, 0.5-mm slice shift) were obtained within a total scan time of 7–15 s. Accordingly, the overall measurement time of the patient including planning could be reduced to less than a minute.
The rapid volume coverage scans have demonstrated good to very good image quality in the evaluation of inner and outer CSF spaces in the brain (Gräfe et al. 2020, article in press in Pediatric Radiology). In all cases the diagnostic question could be answered. Moreover, in comparison to CT and US, the method offers more comprehensive information, so that it is now part of our routine portfolio. It should especially be noted that for all children up to 6 years of age, the MRI examination could be performed without sedation when a parent or another person mildly fixated the child with the hands. Finally, a correspondingly adjusted version might in the future be applied to fetal imaging and complement or even replace conventional T2-weighted sequences for a fretful and moving fetus, as already supported by preliminary trials.

Discussion

To the best of our knowledge this is the first report about the use of real-time MRI in pediatric radiology. This is surprising in view of its several advantages such as extremely short examinations, insensitivity to motion, very low energy deposition for FLASH-type contrast, and avoidance of sedation in children up to 6 years of age. As demonstrated in this initial overview, the technology offers a variety of applications, all of which were met by clinical indication. Our preliminary experience in these cases clearly supports the idea that real-time MRI without sedation or anesthesia could gain significant importance for MRI studies in early childhood. However, because of some compromises in spatial resolution and contrast, real-time MRI applications are likely to complement rather than replace conventional state-of-the-art MRI techniques in older children. Nevertheless, this first report identifies several clinical indications where real-time MRI offers more rapid and better information, in particular when compared to US and its well-known limitations. Similar arguments hold true for CT because of the exposure to ionizing radiation and the reduced soft-tissue contrast. As a consequence, there might be a change of paradigm in pediatric radiology: if a brief MRI examination yields sufficient information to allow for a conclusive diagnosis, then the role of MRI in general and the access to suitable MRI techniques need to be reconsidered, at least for children up to the age of 6 years. Of course, this process requires extensive scientific study to fully understand all of the technical and clinical pros and cons of the novel real-time MRI technology and its dependent applications. Such study, we hope, would substantiate and confirm our current expectations as a prerequisite for routine clinical use. The present findings support the idea that real-time MRI will complement existing methods by long-awaited diagnostic options for patients in early childhood. Major advantages are the avoidance of sedation or anesthesia and the unexplored potential to gain insights into arbitrary body functions.

Acknowledgments

Open Access funding provided by Projekt DEAL. The study was funded by the European Society of Paediatric Radiology (Guy Sebag Grant 2019 for Real-Time MRI).

Compliance with ethical standards

Conflicts of interest

Drs. Frahm and Voit are co-inventors of a patent and software describing the real-time MRI technique used here. The other authors declare no conflicts of interest.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Pädiatrie

Kombi-Abonnement

Mit e.Med Pädiatrie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Pädiatrie, den Premium-Inhalten der pädiatrischen Fachzeitschriften, inklusive einer gedruckten Pädiatrie-Zeitschrift Ihrer Wahl.

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

Literatur
1.
Zurück zum Zitat Uecker M, Zhang S, Voit D et al (2010) Real-time MRI at a resolution of 20 ms. NMR Biomed 23:986–994CrossRef Uecker M, Zhang S, Voit D et al (2010) Real-time MRI at a resolution of 20 ms. NMR Biomed 23:986–994CrossRef
2.
Zurück zum Zitat Frahm J, Voit D, Uecker M (2019) Real-time magnetic resonance imaging — radial gradient-echo sequences with nonlinear inverse reconstruction. Investig Radiol 54:757–766CrossRef Frahm J, Voit D, Uecker M (2019) Real-time magnetic resonance imaging — radial gradient-echo sequences with nonlinear inverse reconstruction. Investig Radiol 54:757–766CrossRef
3.
Zurück zum Zitat Frahm J, Schätz S, Untenberger M et al (2014) On the temporal fidelity of nonlinear inverse reconstructions for real-time MRI — the motion challenge. Open Med Imaging J 8:1–7CrossRef Frahm J, Schätz S, Untenberger M et al (2014) On the temporal fidelity of nonlinear inverse reconstructions for real-time MRI — the motion challenge. Open Med Imaging J 8:1–7CrossRef
4.
Zurück zum Zitat Niebergall A, Zhang S, Kunay E et al (2013) Real-time MRI of speaking at a resolution of 33 ms: undersampled radial FLASH with nonlinear inverse reconstruction. Magn Reson Med 69:477–485CrossRef Niebergall A, Zhang S, Kunay E et al (2013) Real-time MRI of speaking at a resolution of 33 ms: undersampled radial FLASH with nonlinear inverse reconstruction. Magn Reson Med 69:477–485CrossRef
6.
Zurück zum Zitat Olthoff A, Carstens P-O, Zhang S et al (2016) Evaluation of dysphagia by novel real-time MRI. Neurology 87:2132–2138CrossRef Olthoff A, Carstens P-O, Zhang S et al (2016) Evaluation of dysphagia by novel real-time MRI. Neurology 87:2132–2138CrossRef
7.
Zurück zum Zitat Krohn S, Joseph AA, Voit D et al (2019) Multi-slice real-time MRI of temporomandibular joint dynamics. Dentomaxillofacial Radiol 48:20180162CrossRef Krohn S, Joseph AA, Voit D et al (2019) Multi-slice real-time MRI of temporomandibular joint dynamics. Dentomaxillofacial Radiol 48:20180162CrossRef
8.
Zurück zum Zitat Untenberger M, Tan Z, Voit D et al (2016) Advances in real-time phase-contrast flow MRI using asymmetric radial gradient echoes. Magn Reson Med 75:1901–1908CrossRef Untenberger M, Tan Z, Voit D et al (2016) Advances in real-time phase-contrast flow MRI using asymmetric radial gradient echoes. Magn Reson Med 75:1901–1908CrossRef
9.
Zurück zum Zitat Tan Z, Roeloffs V, Voit D et al (2017) Model-based reconstruction for real-time phase-contrast flow MRI: improved spatiotemporal accuracy. Magn Reson Med 77:1082–1093CrossRef Tan Z, Roeloffs V, Voit D et al (2017) Model-based reconstruction for real-time phase-contrast flow MRI: improved spatiotemporal accuracy. Magn Reson Med 77:1082–1093CrossRef
10.
Zurück zum Zitat Wang X, Roeloffs V, Merboldt KD et al (2015) Single-shot multi-slice T1 mapping at high spatial resolution — inversion-recovery FLASH with radial undersampling and iterative reconstruction. Open Med Imaging J 9:1–8CrossRef Wang X, Roeloffs V, Merboldt KD et al (2015) Single-shot multi-slice T1 mapping at high spatial resolution — inversion-recovery FLASH with radial undersampling and iterative reconstruction. Open Med Imaging J 9:1–8CrossRef
11.
Zurück zum Zitat Wang X, Joseph AA, Kalentev O et al (2016) High-resolution myocardial T1 mapping using single-shot inversion recovery fast low-angle shot MRI with radial undersampling and iterative reconstruction. Br J Radiol 89:20160255CrossRef Wang X, Joseph AA, Kalentev O et al (2016) High-resolution myocardial T1 mapping using single-shot inversion recovery fast low-angle shot MRI with radial undersampling and iterative reconstruction. Br J Radiol 89:20160255CrossRef
12.
Zurück zum Zitat Voit D, Kalentev O, Zalk M et al (2020) Rapid and motion-robust volume coverage using cross-sectional real-time MRI. Magn Reson Med 83:1652–1658CrossRef Voit D, Kalentev O, Zalk M et al (2020) Rapid and motion-robust volume coverage using cross-sectional real-time MRI. Magn Reson Med 83:1652–1658CrossRef
13.
Zurück zum Zitat Seif Amir Hosseini A, Beham A, Uhlig J et al (2018) Intra- and interobserver variability in the diagnosis of GERD by real-time MRI. Eur J Radiol 104:14–19CrossRef Seif Amir Hosseini A, Beham A, Uhlig J et al (2018) Intra- and interobserver variability in the diagnosis of GERD by real-time MRI. Eur J Radiol 104:14–19CrossRef
14.
Zurück zum Zitat Zhang S, Joseph AA, Voit D et al (2014) Real-time magnetic resonance imaging of cardiac function and flow — recent progress. Quant Imaging Med Surg 4:313–329PubMedPubMedCentral Zhang S, Joseph AA, Voit D et al (2014) Real-time magnetic resonance imaging of cardiac function and flow — recent progress. Quant Imaging Med Surg 4:313–329PubMedPubMedCentral
15.
Zurück zum Zitat Dreha-Kulaczewski S, Konopka M, Joseph AA et al (2018) Respiration and the watershed of spinal CSF flow in humans. Sci Rep 8:5594CrossRef Dreha-Kulaczewski S, Konopka M, Joseph AA et al (2018) Respiration and the watershed of spinal CSF flow in humans. Sci Rep 8:5594CrossRef
Metadaten
Titel
Real-time magnetic resonance imaging in pediatric radiology — new approach to movement and moving children
verfasst von
Franz Wolfgang Hirsch
Jens Frahm
Ina Sorge
Christian Roth
Dirk Voit
Daniel Gräfe
Publikationsdatum
10.02.2021
Verlag
Springer Berlin Heidelberg
Erschienen in
Pediatric Radiology / Ausgabe 5/2021
Print ISSN: 0301-0449
Elektronische ISSN: 1432-1998
DOI
https://doi.org/10.1007/s00247-020-04828-5

Weitere Artikel der Ausgabe 5/2021

Pediatric Radiology 5/2021 Zur Ausgabe

Minisymposium: Pediatric MRI quality and safety

Introduction

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

„Nur wer sich gut aufgehoben fühlt, kann auch für Patientensicherheit sorgen“

13.04.2024 Klinik aktuell Kongressbericht

Die Teilnehmer eines Forums beim DGIM-Kongress waren sich einig: Fehler in der Medizin sind häufig in ungeeigneten Prozessen und mangelnder Kommunikation begründet. Gespräche mit Patienten und im Team können helfen.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.