Reliability and validity of the COSMED K5 portable metabolic system during walking

verfasst von: Jacob P. DeBlois, Lindsey E. White, Tiago V. Barreira

Erschienen in: European Journal of Applied Physiology | Ausgabe 1/2021

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Portable methods for assessing energy expenditure outside the laboratory and clinical environments are becoming more widely used. As such, it is important to understand the accuracy of such devices. Therefore, the purpose was to determine the reliability and validity of the COSMED K5 portable metabolic system.

Methods

Reliability and validity were assessed in 27 adults (age: 27 ± 5 years; n = 15 women) using a walking protocol. The protocol consisted of a 5-min walk/2-min rest cycle starting at 1.5 mph and increasing in 0.5-mph increments to 4.0 mph. During visit one, participants wore the K5 to assess oxygen consumption (\({\dot{\text{V}}}\)O2), carbon dioxide production (\({\dot{\text{V}}}\)CO2), and other metabolic variables. Two to seven days later, the protocol was repeated twice with the COSMED K5 and K4b2 systems in a randomized, counterbalanced order.

Results

Intraclass correlation coefficients (ICC) revealed that the K5 reliably measured \({\dot{\text{V}}}\)O2 (ICC 0.64–0.85) and \({\dot{\text{V}}}\)CO2 across all walking speeds (ICC 0.50–0.80), with stronger reliability at faster walking speeds compared with slower speeds. Moderate-to-strong relationships were observed for measured gases between the K5 and K4b2. Specifically, \({\dot{\text{V}}}\)O2 exhibited a moderately high-to-high relationship between devices (r = 0.72–0.82), and a similarly moderately high-to-high relationship was observed for \({\dot{\text{V}}}\)CO2 (r = 0.68–0.82). While there were no differences in \({\dot{\text{V}}}\)O2 measured between devices (p ≥ 0.10), the K5 provided lower \({\dot{\text{V}}}\)CO2 readings than the K4b2 during the 3.0, 3.5, and 4.0 mph walking speeds (p ≤ 0.02).

Conclusions

The K5 provided reliable and valid measures of metabolic variables, with greater reliability and validity at faster walking speeds.
Literatur
Zurück zum Zitat Allen MJ, Yen WM (1979) Introduction to measurement theory. Brooks/Cole Publishing Company, Monterey Allen MJ, Yen WM (1979) Introduction to measurement theory. Brooks/Cole Publishing Company, Monterey
Zurück zum Zitat Janse De Jonge XAK, Thompson MW, Chuter VH et al (2012) Exercise performance over the menstrual cycle in temperate and hot, humid conditions. Med Sci Sports Exerc 44(11):2190–2198CrossRef Janse De Jonge XAK, Thompson MW, Chuter VH et al (2012) Exercise performance over the menstrual cycle in temperate and hot, humid conditions. Med Sci Sports Exerc 44(11):2190–2198CrossRef
Zurück zum Zitat Min Lee K, Lee J, Youb Chung C et al (2012) Pitfalls and important issues in testing reliability using intraclass correlation coefficients in orthopaedic research. Clin Orthop Surg 4:149–155CrossRef Min Lee K, Lee J, Youb Chung C et al (2012) Pitfalls and important issues in testing reliability using intraclass correlation coefficients in orthopaedic research. Clin Orthop Surg 4:149–155CrossRef
Zurück zum Zitat Overstreet B, Bassett D Jr, Crouter S et al (2017) Portable open-circuit spirometry systems. J Sports Med Phys Fit 57:227–237 Overstreet B, Bassett D Jr, Crouter S et al (2017) Portable open-circuit spirometry systems. J Sports Med Phys Fit 57:227–237
Zurück zum Zitat Pinnington HC, Wong P, Tay J et al (2001) The level of accuracy and agreement in measures of FEO2, FECO2 and VE between the Cosmed K4b2 portable, respiratory gas analysis system and a metabolic cart. J Sci Med Sport 4:324–335CrossRef Pinnington HC, Wong P, Tay J et al (2001) The level of accuracy and agreement in measures of FEO2, FECO2 and VE between the Cosmed K4b2 portable, respiratory gas analysis system and a metabolic cart. J Sci Med Sport 4:324–335CrossRef
Zurück zum Zitat Safrit M, Wood T (1995) Introduction to Measurement in physical education and exercise science, 3rd edn. McGraw-Hill Education, St. Louis Safrit M, Wood T (1995) Introduction to Measurement in physical education and exercise science, 3rd edn. McGraw-Hill Education, St. Louis
Metadaten
Titel
Reliability and validity of the COSMED K5 portable metabolic system during walking
verfasst von
Jacob P. DeBlois
Lindsey E. White
Tiago V. Barreira
Publikationsdatum
04.10.2020
Verlag
Springer Berlin Heidelberg
Erschienen in
European Journal of Applied Physiology / Ausgabe 1/2021
Print ISSN: 1439-6319
Elektronische ISSN: 1439-6327
DOI
https://doi.org/10.1007/s00421-020-04514-2

Weitere Artikel der Ausgabe 1/2021

European Journal of Applied Physiology 1/2021 Zur Ausgabe

Neu im Fachgebiet Arbeitsmedizin

Das Geschlechterparadoxon in der gesundheitlich beeinträchtigten Lebenszeit – Ende eines Mythos?

Beginnend mit den 1920er-Jahren hat sich eine Vorstellung über die Geschlechterdifferenzen in Gesundheit und Mortalität etabliert, die von Lorber und Moore in dem einprägsamen Satz: „Women get sicker, but men die quicker“, zusammengefasst wurde [ 1 …, S. 13]. Tatsächlich erscheinen vor dem Hintergrund der höheren Lebenserwartung der Frauen die Studienergebnisse zu den Geschlechterdifferenzen in der Morbidität überraschend, wonach Frauen im Durchschnitt einen schlechteren Gesundheitszustand aufweisen als Männer [

Gesunde Lebenserwartung: Ein kritischer Blick auf Nutzen und Potenziale des demographischen Gesundheitsindikators

Open Access Leitthema

Die demographische Alterung hat vielfältige gesellschaftliche Konsequenzen, deren Ausmaß wesentlich vom Gesundheitszustand der Bevölkerung abhängt. Um diesen analysieren und bewerten zu können, wurden spezielle Kennziffern entwickelt, die in …

Wie hat sich die Lebenserwartung ohne funktionelle Einschränkungen in Deutschland entwickelt? Eine Analyse mit Daten des Deutschen Alterssurveys (DEAS)

Deutschland erfährt, wie andere Hocheinkommensstaaten, aufgrund kontinuierlich rückläufiger Mortalität und niedriger Geburtenraten tiefgreifende demografische Veränderungen. Der demografische Wandel führt in Deutschland zu einem zunehmend höheren …

Hitzeschutz im Fokus der hessischen Betreuungs- und Pflegeaufsicht

Open Access Klimawandel Übersichtsartikel

Im Sommer 2023 kündigte das Bundesministerium für Gesundheit (BMG) einen nationalen Hitzeschutzplan an und forderte die Länder auf, zu prüfen, „ob die Warnstufen des [Deutschen Wetterdienstes] DWD mit der Durchführung von Akutmaßnahmen …