Open Access 08.04.2017 | Original Article

Reliability of quadriceps surface electromyography measurements is improved by two vs. single site recordings

verfasst von: T. G. Balshaw, A. Fry, T. M. Maden-Wilkinson, P. W. Kong, J. P. Folland

Erschienen in: European Journal of Applied Physiology | Ausgabe 6/2017

Abstract

Purpose

The reliability of surface electromyography (sEMG) is typically modest even with rigorous methods, and therefore further improvements in sEMG reliability are desirable. This study compared the between-session reliability (both within participant absolute reliability and between-participant relative reliability) of sEMG amplitude from single vs. average of two distinct recording sites, for individual muscle (IM) and whole quadriceps (WQ) measures during voluntary and evoked contractions.

Methods

Healthy males (n = 20) performed unilateral isometric knee extension contractions: voluntary maximum and submaximum (60%), as well as evoked twitch contractions on two separate days. sEMG was recorded from two distinct sites on each superficial quadriceps muscle.

Results

Averaging two recording sites vs. using single site measures improved reliability for IM and WQ measurements during voluntary (16–26% reduction in within-participant coefficient of variation, CVW) and evoked contractions (40–56% reduction in CVW).

Conclusions

For sEMG measurements from large muscles, averaging the recording of two distinct sites is recommended as it improves within-participant reliability. This improved sensitivity has application to clinical and research measurement of sEMG amplitude.
Hinweise
Communicated by Toshio Moritani.
Abkürzungen
60%MVF
Sixty percent of maximum voluntary force
CVW
Within-participant coefficient of variability
DIST
Distal recording site
sEMG
Surface electromyography
ICC
Intraclass correlation coefficient
IM
Individual muscle
MMAX
Supra-maximal M-wave
MVF
Maximum voluntary force
PROX
Proximal recording site
RF
Rectus femoris
VL
Vastus lateralis
VM
Vastus medialis
WQ
Whole quadriceps

Introduction

Surface electromyography (sEMG) is used extensively to measure the electrical activity within skeletal muscles in clinical and research applications, including: the investigation of neurological diseases (Martin et al. 2006; Perrin et al. 2011; Rissanen et al. 2007); the assessment of motor control and muscle dysfunction (Birch et al. 2000; Nederhand et al. 2002; MacDonald et al. 2009); and the evaluation of rehabilitation/exercise interventions (Aagaard et al. 2002; Buckthorpe et al. 2015; Fimland et al. 2010). Despite the relative ease with which sEMG measurements can be performed, there are numerous technical and methodological issues that are recommended to maximise signal fidelity and measurement reliability (De Luca 1997). These considerations include skin preparation (Cram and Rommen 1989), sensor placement (Hermens et al. 2000; Rainoldi et al. 2004) and the use and selection of normalisation methods (Burden 2010; Balshaw and Hunter 2012; Buckthorpe et al. 2012). Nonetheless, despite careful attention to these issues the reliability of absolute sEMG amplitude recording during both voluntary and evoked (involuntary) contractions remains modest (Ball and Scurr 2010; Buckthorpe et al. 2012; Rota et al. 2013). Therefore, methods to further improve the reliability of sEMG measurements of neuromuscular activity are desirable.
The between-session reliability of sEMG measurements are sensitive to any variations in volume conduction (Rutkove 2007), skin impedance (Hermens et al. 2000), and the skin-electrode interface at that particular recording site (Huigen et al. 2002), even if sensor location is precisely replicated. Recordings from a single site may be particularly susceptible to these sources of noise. Furthermore, recordings from a single sEMG sensor measure electrical activity from a relatively minor fraction of large muscles (e.g. the constituent members of the quadriceps femoris muscle group). In contrast, recording sEMG from more than one sensor/site, when averaged, may provide a more robust and reliable measurement of neuromuscular activity that is less susceptible to the noise present at a single site and provide a better representation of electrical activity within the whole muscle (Rash and Quesada 2006). Therefore, we hypothesised that two sensors placed at distinct locations on the same muscle, to derive an average, may improve the between-day reliability of sEMG measurements and facilitate a more stable measure of neuromuscular activity across a large muscle. Measuring a larger proportion of the motor unit pool and the statistical effect of increasing the number of measurements performed might be expected to increase reliability. If this were the case the use of two site sEMG recording might offer greater reliability, and thus also sensitivity, of sEMG measurements for clinical and research applications. Indeed, recent studies have adopted the approach of averaging sEMG from two distinct sensor locations on the same muscle (Fry and Folland 2014; Haider and Folland 2014; Balshaw et al. 2016). However, the effect of using the average of two distinct sEMG recording sites, opposed to one, on the reliability of sEMG amplitude measurements has not been investigated.
The purpose of this study was to compare the between-session reliability of sEMG amplitude measurements from single vs. mean of two sEMG recording sites. The primary measure of reliability was within-participant absolute reliability (coefficient of variation, CVW) and the secondary measure was between-participant relative reliability (intraclass correlation coefficient, ICC). Comparisons were made for each of the individual superficial quadriceps muscles as well as for the quadriceps as a whole (averaged based on either 1 or 2 electrode recording sites per individual muscle) during voluntary (maximum and submaximum) contractions and electrically evoked maximal M-waves (MMAX).

Materials and methods

Participants

Twenty healthy males (mean ± SD, age 22 ± 4 years, height 1.80 ± 0.06 m, body mass 75 ± 9 kg) who were not involved in any systematic physical training provided written informed consent prior to participation in this study, which was approved by Loughborough University Ethical Advisory Committee. Participants had low to moderate physical activity levels [2106 ± 2248 METmin wk−1; international physical activity questionnaire (IPAQ): short format (Craig et al. 2003)] and no history of systematic strength and/or power training.

Overview

Participants attended three test sessions (one familiarisation and two identical test sessions), each at a consistent time of day (12:00–18:00) and separated by 7 days. Participants were instructed to abstain from caffeine, alcohol, and strenuous exercise for 36 h prior to each visit. Throughout each session, participants were seated on a rigid custom-made isometric knee extension dynamometer (Fig. 1a) with knee and hip joint angles of 120° and 100° (180° representing full extension), respectively. Knee extension force and quadriceps sEMG were recorded throughout the two test sessions, whilst participants performed knee extensor contractions of the dominant leg: submaximum and maximum voluntary contractions (MVCs), and electrically evoked maximal twitch contractions with MMAX responses. During the familiarisation session participants completed the same contractions but no data were recorded.

Knee extension force

The configuration of the isometric dynamometer was determined during the familiarisation session and replicated for subsequent test sessions. Adjustable waist and shoulder straps were used to ensure participants were secured firmly in the chair, and prevented extraneous movements. An ankle strap (35 mm width reinforced canvas webbing) was placed proximal to the ankle (15% of tibial length above the medial malleolus), positioned perpendicular to the tibia and in series with a calibrated S-beam strain gauge (Force Logic, Swallowfield, UK). The analogue force signal from the strain gauge was amplified (×370) and sampled at 2000 Hz using an external A/D converter (Micro 1401, CED Ltd., Cambridge, UK) interfaced with Spike 2 computer software (CED Ltd., Cambridge, UK). In offline analysis, force data were low-pass filtered at 500 Hz using a fourth-order zero-lag Butterworth filter. As a custom built dynamometer with a low noise strain gauge (noise range: <0.1 Nm) was used in this study rather than a commercial dynamometer (noise range ~5 Nm) minimal filtering was applied to the force signal (Maffiuletti et al. 2016). Force data were gravity corrected by subtracting baseline force from active force measures.

Electromyography

sEMG was recorded using two Delsys Bagnoli-4 acquisition systems (Delsys Inc., Boston, MA). Two separate double differential wired sEMG sensors (Bagnoli DE-3.1, Delsys, Boston, MA) were placed over the belly of each superficial quadriceps muscle at specific percentages of thigh length (greater trochanter to lateral knee joint space) from the superior border of the patella as follows: vastus medialis (VM), 35% (VMPROX) and 25% (VMDIST); vastus lateralis (VL), 55% (VLPROX) and 45% (VLDIST); and rectus femoris (RF), 65% (RFPROX) and 55% (RFDIST) (Fig. 1b). The proximal and distal sensors on each muscle were offset medio-laterally from the longitudinal mid-line of the muscle belly by 10 mm, and aligned parallel to the presumed orientation of the muscle fibres. The separation of the two electrodes placed over each muscle (proximal–distal and medio-lateral) was used to avoid/minimise recording from the same muscle fibres/motor units. The proportions of thigh length chosen to position sensors at were selected to avoid the innervation zones on each muscle (Rainoldi et al. 2004). A reference electrode was situated over the patella of the same leg. The sEMG signals were amplified (×1000; double differential amplifier, EMG signal bandwidth: 20–450 Hz), sampled at 2000 Hz and synchronised with the force data using the same data acquisition equipment.

Protocol

Following a series of submaximum unilateral isometric knee extension warm-up contractions [50% (×3), 75% (×2), and 90% (×1) of perceived maximum effort] experimental measurements were completed in the following order.

Maximum voluntary contractions

Participants performed four MVCs and were instructed to “push as hard as possible” for 3–5 s during MVCs and rest for ≥30 s between each effort. A force–time curve with a horizontal cursor indicating the greatest force obtained within the session was displayed for biofeedback, and verbal encouragement was provided during all MVCs. Knee extension maximum voluntary force (MVF) during each measurement session was taken from the individual MVC that produced the greatest instantaneous force and was calculated as the mean force over a 500 ms window (250 ms either side of the greatest instantaneous force). Root mean square (RMS) EMG for the same 500 ms epoch at MVF (EMGMVF) was calculated for each quadriceps sEMG sensor before determining single and two site measurements (see Data analysis and statistics).

Submaximum voluntary contractions

Once MVF had been established a horizontal cursor indicating a target force level of 60% MVF was placed on the force–time curve to ensure the desired force level was achieved during a single submaximum contraction where participants were required to match and hold this target force level for ~5 s. Mean force was calculated for a 500 ms time period when there was a steady plateau in force at ~60% MVF. RMS EMG was measured for this same epoch (EMG60%MVF) for each quadriceps sEMG sensor before calculating single and two site measurements (see Data analysis and statistics section).

Evoked twitch contractions with MMAX responses

Femoral nerve stimulation was conducted with a constant current variable voltage stimulator (DS7AH; Digitimer Ltd., Welwyn Garden City, UK), cathode probe (1 cm diameter, Electro-Medical Supplies Ltd., Wantage, UK), and anode electrode (7 × 10 cm carbon rubber electrode; Electro-Medical Supplies Ltd., Wantage, UK). The cathode and anode were coated with electrode gel and securely taped to the skin over the femoral nerve in the femoral triangle and over the greater trochanter, respectively. Cathode location was determined by delivering single electrical impulses (square wave-pulses of 0.2 ms duration, ≥12 s apart) to identify the position that elicited the greatest submaximum twitch response. Thereafter, current intensity was increased until plateaus in peak twitch force and M-wave peak-to-peak (P–P) amplitude were observed. The current intensity was then increased to a supra-maximal level (+50%) and a further three single impulses (15 s apart) were delivered to elicit three MMAX responses. MMAX P–P amplitude and MMAX area were averaged across the three supra-maximal twitch contractions for each of the individual sEMG recording sites. MMAX area was calculated as the cumulative area from EMG onset (after stimulation artefact) to the point where the signal returned to baseline. Peak force from the three supra-maximal twitches was also averaged (twitch peak force).

Data analysis and statistics

All sEMG measurements during the voluntary (EMGMVF, EMG60%MVF) and evoked (MMAX area, MMAX P–P) contractions were first determined for each individual test session. Measurements from each of the six recording sites, two on each of the VL, VM and RF, were considered individually as single site measurements. Two site measurements for each individual muscles were averaged across the two individual sites (e.g. VMTWO = [VMPROX + VMDIST] / 2). To calculate whole quadriceps (WQ) values using only single site recordings from each individual muscle, averages were determined from the three proximal and the three distal recording sites of the individual muscles (e.g. WQSINGLE−PROX = [VMPROX + VLPROX + RFPROX]/3). Whole quadriceps sEMG measurements based on two recording sites per muscle were averaged across the two site measurements from each individual muscle (e.g. WQTWO = [VMTWO + VLTWO + RFTWO]/3). Data are reported as mean ± SD. SPSS Version 22.0 (IBM Corp., Armonk, NY) was used to conduct all statistical analysis and statistical significance was set at P ≤ 0.05.
The primary outcome measure was within-participant coefficient of variation (CVW, [SD/mean] × 100) of sEMG amplitude values from each test session as a measure of absolute reliability. CVW values were calculated for single and two site measurements from each individual muscle and were averaged to calculate representative single site [CVW of IMSINGLE mean = (CVW of VMDIST + CVW of VMPROX + CVW of VLDIST + CVW of VLPROX + CVW of RFDIST + CVW of RFPROX)/6] and two site measurements that were not site/muscle specific [CVW of IMTWO mean = (CVW of VMTWO + CVW of VLTWO + CVW of RFTWO)/3]. Similarly, single site whole quadriceps CVw values were averaged to provide a representative CVW value [CVW of WQSINGLE mean= (CVW of WQSINGLE−DIST + CVW of WQSINGLE−PROX)/2] independent of proximal/distal sites. CVW values were interpreted as “acceptable” <12%, “intermediate” 12–20%, or “unacceptable” >20% (Albertus-Kajee et al. 2011). Shapiro–Wilk tests were used to assess the normality of the sEMG data and CVW values for each single site and two site variable. Several of the variables were not normally distributed and consequently non-parametric statistical tests were used. Wilcoxon signed-rank tests were conducted to compare CVW values between: single and two site measurements; and individual muscle vs. whole quadriceps measurements.
The secondary outcome measure was the ICC. As several of the sEMG variables were not normally distributed they were log-transformed to meet the assumptions of the parametric ICC prior to this statistical test being conducted. ICC values were interpreted as ‘‘very high’’ 0.9–1.0, ‘‘high’’ 0.7–0.9, “moderate” 0.5–0.7, “low” 0.3–0.5, “negligible” 0.0–0.3 (Hinkle et al. 2002). The interpretation of ICC values was done broadly by comparing mean ICC values across several variables (e.g. ICC of IMSINGLE mean vs. ICC of IMTWO mean) and when these were consistently higher (e.g. across most of the voluntary and evoked measures), were considered qualitatively different.

Results

Reliability of voluntary and evoked force

MVF displayed CVW values of 2.6 ± 2.0% (acceptable) and an ICC value of 0.977 (very high) with no difference between test days (session 1: 807 ± 126 N, session 2: 818 ± 129 N; Wilcoxon P = 0.159). Force production during submaximum contractions displayed CVW values of 2.3 ± 1.5% (acceptable) and an ICC value of 0.984 (very high) with no difference between tests days (session 1: 478 ± 75 N, session 2: 482 ± 74 N; Wilcoxon P = 0.247). The submaximum contractions were a consistent proportion of MVF on each test day (session 1: 59.3 ± 2.3%MVF, session 2: 59.0 ± 2.4%MVF). Twitch peak force displayed CVW values of 6.3 ± 5.0% (acceptable) and an ICC value of 0.955 (very high) with no difference between test days (session 1: 144 ± 35 N, session 2: 152 ± 41 N; Wilcoxon P = 0.052).

Reliability of sEMG measurements

EMG data from session 1 and 2 are displayed in Table 1. When averaged across the individual quadriceps muscles, the CVW of two site EMG measurements (IMTWO mean) was significantly lower than for single site measurements (IMSINGLE mean) for voluntary (EMGMVF: Wilcoxon P = 0.002; and EMG60%MVF P < 0.001; Fig. 2) and evoked (MMAX area: P < 0.001; and MMAX P–P: P < 0.001; Fig. 3) contractions. Representing 16–26% and 41–44% reductions in CVW values for voluntary and evoked contractions, respectively. Mean ICC values were higher for two vs. single site EMG measurements (e.g. IMTWO mean vs. IMSINGLE mean) during both voluntary and evoked contractions (Table 2). Representing a 2–9% and 10–15% improvement in ICC values for voluntary and evoked contractions, respectively. For each of the individual muscles CVW values were lower for two vs. single site measures (e.g. VMTWO vs. VMSINGLE mean, VLTWO vs. VLSINGLE mean, RFTWO vs. RFSINGLE mean) during evoked contractions (MMAX area and MMAX P–P: Wilcoxon 0.001 < P ≤ 0.009). CVW values were also lower or tended to be lower for VMTWO vs. VMSINGLE mean (EMGMVF and EMG60%MVF: Wilcoxon 0.002 ≤ P ≤ 0.013) and VLTWO vs. VLSINGLE mean (EMGMVF: P = 0.062; and EMG60%MVF: P = 0.031) during voluntary contractions. CVW values were 15–30% lower for RFTWO vs. RFSINGLE mean during voluntary contractions but this did not reach significance (EMGMVF: Wilcoxon P = 0.218; and EMG60%MVF: P = 0.100).
Table 1
Voluntary maximum (EMGMVF), voluntary submaximum (EMG60%MVF) and evoked [MMAX area and MMAX peak-to-peak (P–P)] surface EMG parameters measured on two separate test days
  
EMGMVF (mV)
EMG60%MVF (mV)
MMAX area (mV.s− 1)
MMAX P-P (mV)
 
Test day
1
2
1
2
1
2
1
2
Whole quadriceps
 Two sites
WQTWO
0.17 ± 0.06
0.18 ± 0.06
0.09 ± 0.03
0.08 ± 0.03
0.009 ± 0.004
0.009 ± 0.004
2.34 ± 1.04
2.32 ± 0.99
 Single site
WQSINGLE−DIST
0.18 ± 0.05
0.19 ± 0.07
0.09 ± 0.03
0.08 ± 0.03
0.011 ± 0.004
0.010 ± 0.004
2.66 ± 1.27
2.56 ± 0.91
WQSINGLE−PROX
0.16 ± 0.08
0.16 ± 0.07
0.08 ± 0.04
0.08 ± 0.03
0.008 ± 0.003
0.008 ± 0.004
2.01 ± 1.02
2.07 ± 1.19
Individual muscles
 Two sites
VMTWO
0.20 ± 0.07
0.19 ± 0.08
0.09 ± 0.03
0.09 ± 0.03
0.013 ± 0.005
0.012 ± 0.006
3.13 ± 1.66
2.85 ± 1.34
VLTWO
0.19 ± 0.09
0.20 ± 0.08
0.10 ± 0.05
0.10 ± 0.04
0.010 ± 0.005
0.010 ± 0.005
2.63 ± 1.51
2.87 ± 1.56
RFTWO
0.12 ± 0.05
0.13 ± 0.07
0.07 ± 0.03
0.06 ± 0.03
0.005 ± 0.003
0.005 ± 0.002
1.25 ± 0.57
1.22 ± 0.62
 Single site
VMDIST
0.24 ± 0.09
0.24 ± 0.10
0.11 ± 0.04
0.11 ± 0.05
0.017 ± 0.007
0.016 ± 0.007
4.31 ± 2.25
3.79 ± 1.59
VMPROX
0.15 ± 0.08
0.14 ± 0.08
0.07 ± 0.04
0.06 ± 0.03
0.009 ± 0.005
0.008 ± 0.006
1.95 ± 1.30
1.91 ± 1.74
VLDIST
0.18 ± 0.07
0.19 ± 0.09
0.09 ± 0.04
0.08 ± 0.04
0.010 ± 0.006
0.009 ± 0.005
2.45 ± 1.73
2.64 ± 1.74
VLPROX
0.19 ± 0.13
0.21 ± 0.09
0.10 ± 0.07
0.11 ± 0.05
0.010 ± 0.007
0.011 ± 0.006
2.81 ± 2.01
3.11 ± 1.61
RFDIST
0.13 ± 0.06
0.14 ± 0.09
0.07 ± 0.04
0.06 ± 0.03
0.005 ± 0.002
0.005 ± 0.003
1.23 ± 0.53
1.24 ± 0.67
RFPROX
0.12 ± 0.05
0.13 ± 0.07
0.07 ± 0.04
0.06 ± 0.03
0.005 ± 0.003
0.005 ± 0.003
1.27 ± 0.78
1.20 ± 0.74
Data are mean ± SD (n = 20)
VM vastus medialis, VL vastus lateralis, RF rectus femoris, DIST distal sEMG recording site, PROX proximal sEMG recording site
Table 2
Between-session intraclass correlation coefficient (ICC) and classifications for single (SINGLE) and average of two sites (TWO) sEMG measurements from individual muscles (IM) and the whole quadriceps (WQ) during maximum (EMGMVF) and submaximum (EMG60%MVF) voluntary contractions (500 ms epochs), as well as during evoked twitch contractions [MMAX area and MMAX peak-to-peak (P–P amplitude)]
 
EMGMVF
EMG60%MVF
MMAX area
MMAX P-P
Whole quadriceps
 Two sites
  WQTWO
0.851
0.943
0.960
0.984
“High”
“Very high”
“Very high”
“Very high”
Single site
  WQSINGLE mean
0.797
0.927
0.885
0.919
“High”
“Very high”
“High”
“Very high”
  WQSINGLE−DIST
0.750
0.929
0.836
0.887
“High”
“Very high”
“High”
“High”
  WQSINGLE−PROX
0.843
0.924
0.934
0.951
“High”
“Very high”
“Very high”
“Very high”
Individual muscles
 Two sites
  IMTWO mean
0.875
0.925
0.923
0.953
“High”
“Very high”
“Very high”
“Very high”
  VMTWO
0.893
0.893
0.910
0.935
“High”
“High”
“Very high”
“Very high”
  VLTWO
0.879
0.937
0.916
0.963
“High”
“Very high”
“Very high”
“Very high”
  RFTWO
0.853
0.946
0.944
0.962
“High”
“Very high”
“Very high”
“Very high”
 Single site
  IMSINGLE mean
0.801
0.904
0.806
0.870
“High”
“Very high”
“High”
“High”
  VMDIST
0.868
0.870
0.865
0.814
“High”
“High”
“High”
“High”
  VMPROX
0.923
0.920
0.822
0.91
“Very high”
“Very high”
“High”
“Very high”
  VLDIST
0.637
0.931
0.684
0.828
“Moderate”
“Very high”
“Moderate”
“High”
  VLPROX
0.832
0.895
0.802
0.887
“High”
“High”
“High”
“High”
  RFDIST
0.665
0.915
0.769
0.882
“Moderate”
“Very high”
“High”
“High”
  RFPROX
0.880
0.895
0.896
0.900
“High”
“High”
“High”
“Very high”
VM vastus medialis, VL vastus lateralis, RF rectus femoris, DIST distal sEMG recording site, PROX proximal sEMG recording site
Whole quadriceps measurements from two sites displayed lower CVW values than for single sites (e.g. WQTWO vs. WQSINGLE mean) for maximum voluntary (EMGMVF: Wilcoxon P = 0.002; Fig. 2) and evoked (MMAX area: P < 0.001; and MMAX P–P: P < 0.001; Fig. 3) contractions. Thereby, representing 21% and 40–56% reductions in CVW values for maximum voluntary and evoked contractions, respectively. CVW values were 15% lower for WQTWO vs. WQSINGLE mean during submaximum contractions but this did not reach statistical significance (EMG60%MVF: Wilcoxon P = 0.121). Whole quadriceps ICC values were greater for two site vs. single site EMG measurements (e.g. WQTWO vs. WQSINGLE mean) during voluntary and evoked contractions (Table 2). Representing 2–7% and 7–8% improvements in ICC values for voluntary and evoked contractions, respectively.
Additionally, the CVW for the whole quadriceps were lower than those of the individual muscle for both single and two site measures (e.g. WQSINGLE mean vs. IMSINGLE mean and WQTWO vs. IMTWO mean) for voluntary (EMGMVF: Wilcoxon 0.001 < P ≤ 0.007; and EMG60%MVF: 0.001 < P ≤ 0.006; Fig. 2) and evoked (MMAX area: P ≤ 0.001; and MMAX P–P: [both] P = 0.002; Fig. 2) contractions. ICC values were 3–10% greater for whole quadriceps vs. single quadriceps EMG measurements for single (WQSINGLE mean vs. IMSINGLE mean) and two (WQTWO vs. IMTWO mean) site measures for evoked contraction parameters (MMAX area and MMAX P-P). During voluntary contractions ICC values were similar (1–3% difference) for single (WQSINGLE mean vs. IMSINGLE mean) and two (WQTWO vs. IMTWO mean) site measures (Table 2).

Discussion

This study compared the between-session reliability (both within-participant absolute reliability and between-participant relative reliability) of sEMG amplitude measurements derived from single vs. average of two recording sites during maximum (EMGMVF) and submaximum (EMG60%MVF) voluntary, as well as electrically evoked (MMAX area and MMAX P–P) contractions for individual muscles and the whole quadriceps. The use of two vs. single recording sites improved within-participant absolute reliability (15–56% reduction in CVW) for individual muscles and whole quadriceps measurements during voluntary and evoked contractions. The results of this study indicate that quantifying voluntary and evoked sEMG measures from two, rather than single, recording sites substantially improved CVW values for these variables. Therefore, it is strongly recommended that quantitative clinical and research measurements of sEMG amplitude, particularly those focusing on within-participant changes, record and average across two sites when addressing large locomotory muscles.
It appears that the enhanced within-participant absolute reliability of the sEMG parameters derived from averaging across two recording sites vs. single site measures in the present investigation was likely due to: quantifying electrical activity from a greater proportion of the motor unit pool; and the statistical effect of increasing the number of measurements performed. Measuring a greater proportion of the motor unit pool might be expected to provide a better, and more stable reflection of the whole muscle or overall muscle group. Averaging across two sites, may also exert a statistical effect simply by reducing measurement variability compared to recording from only one location on the muscle. Single site EMG reliability values of the present investigation were comparable to those reported from several previous studies conducting voluntary (Yang and Winter 1983; Mathur et al. 2005; Ball and Scurr 2010; Fauth et al. 2010; Buckthorpe et al. 2012; Rota et al. 2013) and evoked (Gondin et al. 2005; Place et al. 2007; Buckthorpe et al. 2012) isometric contractions. The current study performed measures during isometric contractions but these findings of enhanced within-participant absolute reliability from two site recordings would be expected to translate to all types of contractions although this needs to be confirmed by future research.
A further consideration when recording sEMG from large individual muscles (such as the VM, VL, and RF) is that often measurements from only one or two muscles of an overall group are collected (Higbie et al. 1996; Häkkinen et al. 1998; Brandon et al. 2014; Trulsson et al. 2015). Individual muscle measures have sometimes been assumed to provide reliable representation of whole muscle group activation. However, the greater within-participant reliability of whole quadriceps vs. individual muscle measurements (for voluntary and evoked contractions, as well as single and two site measures) in the current study suggests calculating whole quadriceps sEMG measures is preferable to enhance absolute reliability vs. measuring one or two of the individual constituent muscles. Therefore, it is strongly suggested that when overall acute or chronic changes in quadriceps sEMG parameters are of interest that measures averaged across the VM, VL, and RF are used.
The placement of two EMG sensors on the same muscle may introduce cross-talk between the sensors, i.e. some commonality to the recorded signals. To reduce this possibility, we took the following approaches: used double differential EMG sensors that are known to have a smaller detection volume than single differential sensors (Stepp 2012); used sensors with small inter-electrode distance (10 mm) that are thought to minimize cross-talk (De Luca et al. 2012); performed measurements in the current study on some of the largest muscles in the human body of healthy young men; and spatially separated the two sensors in both proximo-distal and medio-lateral directions. Qualitatively, the signals from two sensors on the same muscle appeared to be independent, nevertheless it is possible that there could have been some cross-talk between sensors, but currently there is no accepted analytical procedure to assess the extent of cross-talk within an EMG signal (Farina et al. 2014). The observation that averaging the two signals improves the reliability of EMG amplitude measurements may indicate that the signals were substantially independent, although it is unknown if this finding of improved reliability was specific to the conditions (cohort, muscles and electrodes) of our study.
In conclusion, the use of two vs. single sEMG recording sites improved the within-participant reliability of sEMG parameters across a range of different contraction types (voluntary maximum, submaximum, and electrically evoked). The effects of using two recording sites to quantify sEMG measures had the greatest benefit for within-participant reliability (CVW), but also produced some small but consistent improvement in relative measures of reliability (ICC). This greater reliability would be expected to increase the sensitivity of sEMG measurements to detect changes within, and differences between individuals. In addition, whole quadriceps sEMG within-participant reliability was greater than that of the individual muscles for both single and two site measures. Given the importance of reliability for clinical and research applications of sEMG, it is recommended that when measuring large muscles, such as the quadriceps femoris, that sEMG parameters are quantified by taking mean measures across two distinct recording sites before reporting absolute EMG values or normalising data.

Acknowledgements

The authors thank the participants for their time in taking part in the study.

Funding

The study was financially supported by the Arthritis Research UK Centre for Sport, Exercise, and Osteoarthritis (Grant reference 20194).
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Allgemeinmedizin

Kombi-Abonnement

Mit e.Med Allgemeinmedizin erhalten Sie Zugang zu allen CME-Fortbildungen und Premium-Inhalten der allgemeinmedizinischen Zeitschriften, inklusive einer gedruckten Allgemeinmedizin-Zeitschrift Ihrer Wahl.

Literatur
Zurück zum Zitat Birch L, Graven-Nielsen T, Christensen H, Arendt-Nielsen L (2000) Experimental muscle pain modulates muscle activity and work performance differently during high and low precision use of a computer mouse. Eur J Appl Physiol 83:492–498. doi:10.1007/s004210000320 CrossRefPubMed Birch L, Graven-Nielsen T, Christensen H, Arendt-Nielsen L (2000) Experimental muscle pain modulates muscle activity and work performance differently during high and low precision use of a computer mouse. Eur J Appl Physiol 83:492–498. doi:10.​1007/​s004210000320 CrossRefPubMed
Zurück zum Zitat Brandon R, Howatson G, Strachan F, Hunter AM (2014) Neuromuscular response differences to power vs strength back squat exercise in elite athletes. Scand J Med Sci Sports 25:630–639. doi:10.1111/sms.12289 CrossRefPubMed Brandon R, Howatson G, Strachan F, Hunter AM (2014) Neuromuscular response differences to power vs strength back squat exercise in elite athletes. Scand J Med Sci Sports 25:630–639. doi:10.​1111/​sms.​12289 CrossRefPubMed
Zurück zum Zitat Buckthorpe MW, Hannah R, Pain TG, Folland JP (2012) Reliability of neuromuscular measurements during explosive isometric contractions, with special reference to electromyography normalization techniques. Muscle Nerve 46:566–576. doi:10.1002/mus.23322 CrossRefPubMed Buckthorpe MW, Hannah R, Pain TG, Folland JP (2012) Reliability of neuromuscular measurements during explosive isometric contractions, with special reference to electromyography normalization techniques. Muscle Nerve 46:566–576. doi:10.​1002/​mus.​23322 CrossRefPubMed
Zurück zum Zitat Buckthorpe M, Erskine RM, Fletcher G et al (2015) Task-specific neural adaptations to isoinertial resistance training. Scand J Med Sci Sports 25:640–649CrossRefPubMed Buckthorpe M, Erskine RM, Fletcher G et al (2015) Task-specific neural adaptations to isoinertial resistance training. Scand J Med Sci Sports 25:640–649CrossRefPubMed
Zurück zum Zitat Cram JR, Rommen D (1989) Effects of skin preparation on data collected using an EMG muscle-scanning procedure. Biofeedback Self Regul 14:75–82CrossRefPubMed Cram JR, Rommen D (1989) Effects of skin preparation on data collected using an EMG muscle-scanning procedure. Biofeedback Self Regul 14:75–82CrossRefPubMed
Zurück zum Zitat De Luca CJ (1997) The use of surface electromyography in biomechanics. J Appl Biomech 13:135–163CrossRef De Luca CJ (1997) The use of surface electromyography in biomechanics. J Appl Biomech 13:135–163CrossRef
Zurück zum Zitat Häkkinen K, Newton RU, Gordon SE et al (1998) Changes in muscle morphology, electromyographic activity, and force production characteristics during progressive strength training in young and older men. J Gerontol Ser A-Biol Sci. Med Sci 53:B415–B423 Häkkinen K, Newton RU, Gordon SE et al (1998) Changes in muscle morphology, electromyographic activity, and force production characteristics during progressive strength training in young and older men. J Gerontol Ser A-Biol Sci. Med Sci 53:B415–B423
Zurück zum Zitat Hermens HJ, Freriks B, Disselhorst-Klug C, Rau G (2000) Development of recommendations for SEMG sensors and sensor placement procedures. J Electromyogr Kinesiol 10:361–374CrossRefPubMed Hermens HJ, Freriks B, Disselhorst-Klug C, Rau G (2000) Development of recommendations for SEMG sensors and sensor placement procedures. J Electromyogr Kinesiol 10:361–374CrossRefPubMed
Zurück zum Zitat Higbie EJ, Cureton KJ, Warren GL, Prior BM (1996) Effects of concentric and eccentric training on muscle strength, cross-sectional area, and neural activation. J Appl Physiol 81:2173–2181PubMed Higbie EJ, Cureton KJ, Warren GL, Prior BM (1996) Effects of concentric and eccentric training on muscle strength, cross-sectional area, and neural activation. J Appl Physiol 81:2173–2181PubMed
Zurück zum Zitat Hinkle DE, Wiersma W, Jurs SG (2002) Applied Statistics for the Behavioral Sciences, 5 edn. Houghton Mifflin, Boston Hinkle DE, Wiersma W, Jurs SG (2002) Applied Statistics for the Behavioral Sciences, 5 edn. Houghton Mifflin, Boston
Zurück zum Zitat Huigen E, Peper A, Grimbergen CA (2002) Investigation into the origin of the noise of surface electrodes. Med Biol Eng Comput 40:332–338CrossRefPubMed Huigen E, Peper A, Grimbergen CA (2002) Investigation into the origin of the noise of surface electrodes. Med Biol Eng Comput 40:332–338CrossRefPubMed
Zurück zum Zitat Nederhand MJ, Hermens HJ, IJzerman MJ et al (2002) Cervical muscle dysfunction in chronic whiplash-associated disorder grade 2: the relevance of the trauma. Spine 27:1056–1061CrossRefPubMed Nederhand MJ, Hermens HJ, IJzerman MJ et al (2002) Cervical muscle dysfunction in chronic whiplash-associated disorder grade 2: the relevance of the trauma. Spine 27:1056–1061CrossRefPubMed
Zurück zum Zitat Perrin RN, Richards JD, Pentreath V, Percy DF (2011) Muscle fatigue in chronic fatigue syndrome/myalgic encephalomyelitis (CFS/ME) and its response to a manual therapeutic approach: a pilot study. Int J Osteopath Med 14:96–105. doi:10.1016/j.ijosm.2010.12.002 CrossRef Perrin RN, Richards JD, Pentreath V, Percy DF (2011) Muscle fatigue in chronic fatigue syndrome/myalgic encephalomyelitis (CFS/ME) and its response to a manual therapeutic approach: a pilot study. Int J Osteopath Med 14:96–105. doi:10.​1016/​j.​ijosm.​2010.​12.​002 CrossRef
Zurück zum Zitat Place N, Maffiuletti NA, Martin A, Lepers R (2007) Assessment of the reliability of central and peripheral fatigue after sustained maximal voluntary contraction of the quadriceps muscle. Muscle Nerve 35:486–495. doi:10.1002/mus.20714 CrossRefPubMed Place N, Maffiuletti NA, Martin A, Lepers R (2007) Assessment of the reliability of central and peripheral fatigue after sustained maximal voluntary contraction of the quadriceps muscle. Muscle Nerve 35:486–495. doi:10.​1002/​mus.​20714 CrossRefPubMed
Zurück zum Zitat Rash G, Quesada P (2006) Electromyography: fundamentals. In: Karwowski W (ed) International encyclopedia of ergonomics and human factors, 2nd edn. Boca Raton Rash G, Quesada P (2006) Electromyography: fundamentals. In: Karwowski W (ed) International encyclopedia of ergonomics and human factors, 2nd edn. Boca Raton
Zurück zum Zitat Rutkove S (2007) Introduction to volume conduction. In: Blum A, Rutkove S (eds) The clinical neurophysiology primer. Totowa, NJ, pp 43–53 Rutkove S (2007) Introduction to volume conduction. In: Blum A, Rutkove S (eds) The clinical neurophysiology primer. Totowa, NJ, pp 43–53
Zurück zum Zitat Yang JF, Winter DA (1983) Electromyography reliability in maximal and submaximal isometric contractions. Arch Phys Med Rehabil 64:417–420PubMed Yang JF, Winter DA (1983) Electromyography reliability in maximal and submaximal isometric contractions. Arch Phys Med Rehabil 64:417–420PubMed
Metadaten
Titel
Reliability of quadriceps surface electromyography measurements is improved by two vs. single site recordings
verfasst von
T. G. Balshaw
A. Fry
T. M. Maden-Wilkinson
P. W. Kong
J. P. Folland
Publikationsdatum
08.04.2017
Verlag
Springer Berlin Heidelberg
Erschienen in
European Journal of Applied Physiology / Ausgabe 6/2017
Print ISSN: 1439-6319
Elektronische ISSN: 1439-6327
DOI
https://doi.org/10.1007/s00421-017-3595-z

Weitere Artikel der Ausgabe 6/2017

European Journal of Applied Physiology 6/2017 Zur Ausgabe

Neu im Fachgebiet Arbeitsmedizin

Das Geschlechterparadoxon in der gesundheitlich beeinträchtigten Lebenszeit – Ende eines Mythos?

Beginnend mit den 1920er-Jahren hat sich eine Vorstellung über die Geschlechterdifferenzen in Gesundheit und Mortalität etabliert, die von Lorber und Moore in dem einprägsamen Satz: „Women get sicker, but men die quicker“, zusammengefasst wurde [ 1 …, S. 13]. Tatsächlich erscheinen vor dem Hintergrund der höheren Lebenserwartung der Frauen die Studienergebnisse zu den Geschlechterdifferenzen in der Morbidität überraschend, wonach Frauen im Durchschnitt einen schlechteren Gesundheitszustand aufweisen als Männer [

Gesunde Lebenserwartung: Ein kritischer Blick auf Nutzen und Potenziale des demographischen Gesundheitsindikators

Open Access Leitthema

Die demographische Alterung hat vielfältige gesellschaftliche Konsequenzen, deren Ausmaß wesentlich vom Gesundheitszustand der Bevölkerung abhängt. Um diesen analysieren und bewerten zu können, wurden spezielle Kennziffern entwickelt, die in …

Wie hat sich die Lebenserwartung ohne funktionelle Einschränkungen in Deutschland entwickelt? Eine Analyse mit Daten des Deutschen Alterssurveys (DEAS)

Deutschland erfährt, wie andere Hocheinkommensstaaten, aufgrund kontinuierlich rückläufiger Mortalität und niedriger Geburtenraten tiefgreifende demografische Veränderungen. Der demografische Wandel führt in Deutschland zu einem zunehmend höheren …

Hitzeschutz im Fokus der hessischen Betreuungs- und Pflegeaufsicht

Open Access Klimawandel Übersichtsartikel

Im Sommer 2023 kündigte das Bundesministerium für Gesundheit (BMG) einen nationalen Hitzeschutzplan an und forderte die Länder auf, zu prüfen, „ob die Warnstufen des [Deutschen Wetterdienstes] DWD mit der Durchführung von Akutmaßnahmen …