Application of ultrasound artificial intelligence in the differential diagnosis between benign and malignant breast lesions of BI-RADS 4A

verfasst von: Sihua Niu, Jianhua Huang, Jia Li, Xueling Liu, Dan Wang, Ruifang Zhang, Yingyan Wang, Huiming Shen, Min Qi, Yi Xiao, Mengyao Guan, Haiyan Liu, Diancheng Li, Feifei Liu, Xiuming Wang, Yu Xiong, Siqi Gao, Xue Wang, Jiaan Zhu

Erschienen in: BMC Cancer | Ausgabe 1/2020

Abstract

Background

The classification of Breast Imaging Reporting and Data System 4A (BI-RADS 4A) lesions is mostly based on the personal experience of doctors and lacks specific and clear classification standards. The development of artificial intelligence (AI) provides a new method for BI-RADS categorisation. We analysed the ultrasonic morphological and texture characteristics of BI-RADS 4A benign and malignant lesions using AI, and these ultrasonic characteristics of BI-RADS 4A benign and malignant lesions were compared to examine the value of AI in the differential diagnosis of BI-RADS 4A benign and malignant lesions.

Methods

A total of 206 lesions of BI-RADS 4A examined using ultrasonography were analysed retrospectively, including 174 benign lesions and 32 malignant lesions. All of the lesions were contoured manually, and the ultrasonic morphological and texture features of the lesions, such as circularity, height-to-width ratio, margin spicules, margin coarseness, margin indistinctness, margin lobulation, energy, entropy, grey mean, internal calcification and angle between the long axis of the lesion and skin, were calculated using grey level gradient co-occurrence matrix analysis. Differences between benign and malignant lesions of BI-RADS 4A were analysed.

Results

Significant differences in margin lobulation, entropy, internal calcification and ALS were noted between the benign group and malignant group (P = 0.013, 0.045, 0.045, and 0.002, respectively). The malignant group had more margin lobulations and lower entropy compared with the benign group, and the benign group had more internal calcifications and a greater angle between the long axis of the lesion and skin compared with the malignant group. No significant differences in circularity, height-to-width ratio, margin spicules, margin coarseness, margin indistinctness, energy, and grey mean were noted between benign and malignant lesions.

Conclusions

Compared with the naked eye, AI can reveal more subtle differences between benign and malignant BI-RADS 4A lesions. These results remind us carefully observation of the margin and the internal echo is of great significance. With the help of morphological and texture information provided by AI, doctors can make a more accurate judgment on such atypical benign and malignant lesions.
Hinweise

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
BI-RADS
Breast Imaging Reporting and Data system
AI
Artificial intelligence
ROI
Region of interest
ALS
Angle between the long axis of the lesion and skin
Cir
Circularity
HWR
Height-to-width ratio
MS
Margin spicules
MC
Margin coarseness
MI
Margin indistinctness
ML
Margin lobulation
E
Energy
ENT
Entropy
GM
Grey mean

Background

The Breast Imaging Reporting and Data System (BI-RADS) facilitates communications among radiologists, clinicians and patients via the use of standardised descriptions of lesions and reports, which greatly promotes the application of breast imaging in clinical practice. BI-RADS 4A lesions exhibit a low suspicion for malignancy of 2–10% and primarily include some atypical benign and malignant lesions [1]. The 2013 BI-RADS does not provide specific guidance for the sub-category of BI-RADS 4 lesions. The classification of these lesions is mostly based on the personal experience of doctors and lacks specific and clear classification standards. The large ultrasonic feature span of atypical benign and malignant lesions creates the possibility of misclassification in the BI-RADS 4A category.
The development of artificial intelligence (AI) provides a new method for BI-RADS classification [2]. AI can calculate the morphological and texture features of breast lesions in ultrasonic images and overcome the shortcomings of human visual observation [35]. At present, the application of AI in BI-RADS classification mainly focuses on the feasibility and accuracy of different AI procedures [610]. AI can achieve a classification level similar to that of radiologists [6, 9]. Through the quantitative study of BI-RADS classification features, some studies have reported morphological and textural features that are different between benign and malignant lesions. The shape, margin, internal echo and posterior echo of tumour can be used as the differential diagnosis points of benign and malignant lesions [6, 8, 11]. Some other studies focus on the differences in morphological and textural features among different BI-RADS categories or specific diseases, for example, triple-negative breast cancer and fibroadenoma [6, 1214]. Studies investigating the application of AI between BI-RADS 4A benign and malignant lesions are limited. The present study analysed the ultrasonic morphological and texture characteristics of BI-RADS 4A benign and malignant lesions using AI and aimed at examining the value of AI in the differential diagnosis of BI-RADS 4A benign and malignant lesions.

Methods

All of the patients were from Peking University People’s Hospital, Southeast University Zhongda Hospital, the First Affiliated Hospital of Guangxi University of Chinese Medicine and Zhengzhou University First Affiliated Hospital. The ethics committees of the four hospitals approved this study. Written informed consents were obtained from all participants. All the doctors participated in the ultrasonic examinations. All lesions diagnosed as BI-RADS 4A before surgery from January 2019 to December 2019 were collected and analysed retrospectively. According to the ACR BI-RADS® Atlas Fifth Edition, two doctors (SHN and XW) with more than 10 years’ experience in breast ultrasound diagnosis who were blind to the pathological results evaluated the suspicion for malignancy of all the lesions separately, and lesions with low suspicion for malignancy (2–10%) were classified as BI-RADS 4A.
The inclusion criteria were as follows: (1) lesions were classified as BI-RADS 4A by the two doctors finally; (2) the lesions were clear in grey-scale images without measurement labels and the sample window of colour Doppler; (3) lesions should be displayed within a high-frequency probe, and those less than 5 cm were included according to the width of high-frequency probes; (4) all lesions were surgically resected and pathologically diagnosed. The following exclusion criteria were employed: (1) lesions were displayed in colour Doppler ultrasound images; (2) measurement labels were present in grey scale images; (3) the transverse diameter of lesions exceeded the width of probes.
Among them, 194 lesions were both classified as BI-RADS 4A by the two doctors. Twelve cases with inconsistent classification were determined as BI-RADS 4A after discussion by the two doctors. Finally, 206 lesions were enrolled in our study.
The ultrasound instruments used in this study included Aixplorer (Supersonic Imagine, Aix-en-Provence, France), Aplio 500 (Toshiba Medical Systems, Otawara, Tochigi, Japan), and Logic E9 (GE Healthcare, Milwaukee, WI, USA) instruments.
The AI software used in this research was the breast ultrasound intelligent diagnosis system developed by the Harbin Institute of Technology. All lesions were manually contoured, and the region of interest (ROI) was calculated using grey gradient co-occurrence matrix analysis to obtain the morphological and texture features.
The morphological features included circularity, height-to-width ratio, margin spicules, margin coarseness, margin indistinctness, margin lobulation, internal calcification and angle between the long axis of the lesion and skin (ALS). The principles of these features were as follows:
(1)
Circularity
 
Circularity (Cir) described the similarity between tumours and circle, and it was calculated according to the following formula (1):
$$ Cir=\frac{C^2}{S} $$
(1)
C was the number of pixels in the tumour boundary, which was equivalent to the perimeter of the tumour, and S was the number of pixels contained in the tumour area, which could be regarded as the area of the tumour.
2
Height-to-width ratio
 
The height-to-width ratio (HWR) calculated the circumscribed rectangle of the tumour boundary first to obtain the height and width of the circumscribed rectangle and then calculated the ratio of the two using the formula (2):
$$ HWR=\frac{H}{W} $$
(2)
3
Margin spicules
 
The coordinates of the margin pixels (xi, yi) were set to coordinates in polar coordinates (ri, θi) according to centroid coordinates (x0, y0). Then, the coordinates were rearranged clockwise (or anticlockwise). Then, Fourier transformation was performed, and the frequency spectrum data were obtained. The number of margin spicules (MS) was calculated according to the following formula (3):
$$ MS=\frac{\sum \limits_{\omega =0}^{\frac{\pi }{4}}R\left(\omega \right)}{\sum \limits_{\omega =\frac{\pi }{4}}^{\pi }R\left(\omega \right)} $$
(3)
4
Margin coarseness
 
Margin coarseness (MC) reflectd the degree of coarseness of tumour margin, which was given by Eq. (4):
$$ MC=\frac{1}{N}\sum \limits_{i=1}^N\mid {d}_i-{d}_{i+1}\mid $$
(4)
Here, di reflected the distance (in pixel units) of the ith pixel on the boundary to the centroid coordinates of the tumour, and di was arranged and calculated according to the clockwise (or anticlockwise) order of the corresponding pixels on the boundary.
5
Margin indistinctness
 
The coarse boundary of tumour in the original grey-scale ultrasound image was calculated using a rough segmented ROI image, and tissue surrounding the tumour was regarded as the boundary area. The pixel gradient in horizontal and vertical directions of the boundary area was calculated using the Sobel operator, and the margin indistinctness (MI) was calculated according to the following formula (5):
$$ MI=\sum \limits_{i=1}^n\sum \limits_{j=1}^m\sqrt{d_x{\left(i,j\right)}^2+{d}_y{\left(i,j\right)}^2} $$
(5)
M and n represented the size of the image, and dx and dy represented the gradient in the horizontal and vertical directions of the pixel at the tumour boundary, respectively.
6
Margin lobulation
 
The coordinates of the margin pixels (xi, yi) were converted to coordinates in polar coordinates (ri, θi) according to centroid coordinates (x0, y0). Here, θi was converted to the polar coordinate sequence (r1, θ1), (r2, θ2), (r3, θ3), (ri, θi), (rn, θn) according to the clockwise (or anticlockwise) order. The median filter of frame size 21 was used to reduce the influence of image noise, and the sequence was fitted with a polynomial of degree 20. The sum of the maximum and minimum points was obtained as the value of margin lobulation (ML) listed in the formula (6).
https://static-content.springer.com/image/art%3A10.1186%2Fs12885-020-07413-z/MediaObjects/12885_2020_7413_Figa_HTML.png
(6)
7
Internal calcification
 
First, the irrelevant region outside the tumour was set as zero pixels according to the coarse segmentation results, and the interior region of the tumour was binarized according to the mean grey value and the maximum value. Then, the binary image was processed by morphology expansion and corrosion to remove the interference pixels; finally, the number of connected regions of the white spots in the binary image was the number of internal calcifications in the image.
8
ALS
 
ALS θ described the angle between the tumour area and the horizontal direction. The ellipse fitting algorithm was used to fit the tumour boundary of ROI image, and the fitted ellipse centre, long axis, short axis, the positive angles of long axis and X axis were obtained. The following transformation was performed according to the formula (7):
$$ \Big\{{\displaystyle \begin{array}{c}\theta, 0\le \theta \le \frac{\pi }{2}\\ {}\pi -\theta, \frac{\pi }{2}\le \theta \le \pi \\ {}\theta -\pi, \pi \le \theta \le \frac{3\pi }{2}\\ {}2\pi -\theta, \frac{3\pi }{2}\le \theta \le 2\pi \end{array}} $$
(7)
Texture features included energy, entropy and grey mean. The number of pixels with a grey level of i and gradient of j in the gradient image simultaneously was the value of H(i, j). Here, H(i, j) was normalised to obtain P(i, j), and P(i, j) was used to calculate these texture features. The calculations of energy, entropy and grey mean were according to the formulas (8, 9 and 10), respectively:
(1)
Energy (E)
 
$$ E=\sum \limits_{i=1}^{L_s}\sum \limits_{j=1}^{L_g}P{\left(i,j\right)}^2 $$
(8)
2
Entropy (Ent)
 
$$ Ent=-\sum \limits_{i=1}^{L_s}\left(\sum \limits_{j=1}^{L_g}P\left(i,j\right)\ast \log \sum \limits_{j=1}^{L_g}P\left(i,j\right)\right) $$
(9)
3
Grey mean (GM)
 
$$ GM=\sum \limits_{i=1}^{L_s}i\ast \sum \limits_{j=1}^{L_g}P\left(i,j\right) $$
(10)

Statistical analysis

The SPSS version 17.0 software package for Windows (IBM Corporation, Armonk, NY, USA) was used for data analyses. Descriptive statistics and frequencies were provided for circularity, height-to-width ratio, margin spicules, margin coarseness, margin indistinctness, margin lobulation, energy, entropy, grey mean, internal calcification and ALS, which were all nomal distribution. Means ± standard deviation were used to describe these features. Two independent samples t-test was used to compare two means in the sample. P < 0.05 indicated a statistically significant difference.

Results

All of the 206 patients were female. All of the lesions were isolated. A total of 174 cases were benign. The median patient age was 39 years (range: 26–57 years), and the median lesion size was 1.6 cm (range: 0.6–4.2 cm). Thirty two cases were malignant. The median patient age was 43 years (range: 32–63 years), and the median lesion size was 1.3 cm (range: 0.8–2.5 cm). The pathological types of benign lesions and malignant lesions were presented in Table 1.
Table 1
Pathological types of benign lesions and malignant lesions
 
Pathological types
n
%
Benign lesions
Adenosis
67
38.5
Benign phyllodes tumour
6
3.4
Fibroadenoma
68
39.1
Inflammation
8
4.6
Intraductal papilloma
20
11.5
Complex cyst
5
2.9
Malignant lesions
Intraductal carcinoma in situ
12
37.5
Intraductal papillary carcinoma
2
6.3
Invasive ductal carcinoma
18
56.2
Data for the circularity, height-to-width ratio, margin spicules, margin coarseness, margin indistinctness, margin lobulation, energy, entropy, mean of grey level, internal calcification and ALS were presented in Table 2.
Table 2
Morphological and texture characteristics of benign and malignant groups
 
Benign group
Malignant group
P
Circularity
23.62 ± 6.85
27.53 ± 13.45
0.240
Height-to-width ratio
0.59 ± 0.27
0.69 ± 0.16
0.343
Margin spicules
14.60 ± 9.40
14.67 ± 11.01
0.960
Margin coarseness (pixel)
2.78 ± 2.52
3.43 ± 3.00
0.420
Margin indistinctness (E6)
2.19 ± 9.03
2.23 ± 7.36
0.610
Margin lobulation
7.68 ± 2.55
9.58 ± 3.40
0.013
Energy
0.03 ± 0.02
0.04 ± 0.03
0.271
Entropy
2.34 ± 0.13
2.23 ± 0.20
0.045
Grey mean
0.52 ± 0.17
0.49 ± 0.28
0.957
Internal calcification
3.78 ± 4.29
2.53 ± 2.05
0.045
ALS (degree)
13.58 ± 18.42
6.32 ± 5.65
0.002
Statistically significant differences in margin lobulation, entropy, internal calcification and ALS were noted between the benign and malignant groups. The malignant group exhibited increased margin lobulation (Fig. 1) and lower entropy compared with the benign group, and the benign group had more internal calcifications and increased ALS compared with the malignant group (Fig. 2). No significant differences in circularity, height-to-width ratio, margin spicules, margin coarseness, margin indistinctness, energy, and grey mean were noted between the benign and malignant groups.

Discussion

AI exhibits high accuracy in the diagnosis of breast lesions [15, 16]. AI significantly improves the diagnostic accuracy of doctors and improves the consistency among observers [7]. According to a study of BI-RADS 3 lesions, the computer-aided diagnosis system could correctly upgrade most malignant tumours misdiagnosed as Category 3 by doctors [12]. For Category 4A, AI also exhibited high diagnostic efficiency, and the classification accuracy of BI-RADS 4A can be greater than 0.9 [10, 14, 17].
Morphological and texture features are the main factors for AI diagnosis. According to the literatures, the use of morphological features and texture features is not limited to the diagnosis of benign and malignant diseases, and these features also help classify malignant tumour subtypes [13, 15, 1719]. Entropy reflects the complexity and heterogeneous character of lesion texture. Larger entropy indicates more information contained in an image and greater uniformity of the pixel matrix of the image [20]. Compared to benign tumours, the internal components of malignant tumours are more complex. The different proportions of fibrous components, haemorrhage, necrosis, and calcification, result in a heterogeneous echo of malignant tumour. The increase in scattering media causes variation in backscattering, which reduces entropy. Therefore, compared with benign tumours, the entropy of malignant tumours is often reduced [20, 21]. Category 4A benign and malignant lesions are atypical benign and malignant lesions. Our study demonstrated that these atypical benign and malignant lesions were still consistent with previous studies [20, 21]. These findings suggest that careful observation of the internal echo of the lesions will help doctors improve the accuracy of naked eye diagnosis of difficult differentiations between benign and malignant tumours.
Category 4A benign and malignant lesions exhibited a significant difference in the number of margin lobulations. The biological behavior of the tumour determines the ultrasonic characteristics. The growth of cancer cells is not uniform and results in an irregular tumour morphology, which is lobulated. On the other hand, the ultrasonic characteristics of the lesions reflect the essential characteristics of the tumour, which is the basis for differentiating between benign and malignant lesions. Therefore, the characteristics of tumour margin are significant in the differentiation of atypical benign and malignant lesions, which is consistent with the literature [11].
Calcification can occur in both benign and malignant breast lesions. Most of the calcifications are benign, but a small portion is malignant [22]. Some benign tumours may have mucinous degeneration or hyaline degeneration with dystrophic calcification, which is occasionally difficult to distinguish from breast cancer calcification [23]. More calcifications were found in benign lesions in our research, which is consistent with early literatures [22, 24]. These characteristics increase the pathological uncertainty of benign lesions and make these lesions more atypical.
Most of the benign lesions grow in parallel, but atypical benign and malignant lesions may also exhibit unconventional characteristics. In this study, the ALS of benign lesions was larger than that of malignant lesions. In a sense, category 4A benign lesions are more like malignant lesions based on some ultrasound features. Category 4A malignant lesions exhibit fewer typical malignant signs, and some of their ultrasound features are more similar to those of benign lesions. These differences reflect the characteristics of category 4A lesions. The boundaries of some characteristics between category 4A benign and malignant lesions are indistinct or even inverted and deviate from the signs of typical benign and malignant lesions [25]. Difficulty in the differential diagnosis of the two groups causes the classification of benign lesions to be upgraded, whereas the classification of malignant lesions is downgraded.
Our study had some limitations. First, the size of our sample was relatively small. Future studies will include a larger number of cases. Second, in the aspect of intralesional calcification, we only studied the value of the number of calcification in the differential diagnosis of BI-RADS 4A benign and malignant lesions, but the significance of the size and shape of calcification in the differential diagnosis was not clear. Finally, this study was based on manually contoured images for quantitative analyses of ROI, which was different from other studies that focused on lesions that are automatically contoured by AI [26]. The present study did not evaluate the automatic identification efficiency for BI-RADS 4A lesions of our AI diagnosis system, and these aspects will be studied in the future.

Conclusions

AI gives us a lot of inspiration. First of all, AI can find out the difference between benign and malignant lesions of BI-RADS 4A, which exceeds the recognition ability of human eyes. Secondly, AI reminds us we should carefully observe whether the lesions are more lobulated and whether the internal echo is more heterogeneous. Especially, the combination of the two features has higher diagnostic value. However, it need a large quantity of cases to determine the threshold of margin lobulation, entropy and internal calcification to diagnose malignant lesions of BI-RADS 4A, our cases are far from enough, especially for the malignant lesions. In the future, we will collect more lesions of BI-RADS 4A and summarize their characteristics so as to obtain a more accurate differential diagnosis threshold.

Acknowledgments

Not applicable.
This study was approved by the ethics committees of Peking University People’s Hospital, Southeast University Zhongda Hospital, the First Affiliated Hospital of Guangxi University of Chinese Medicine and Zhengzhou University First Affiliated Hospital. The four hospitals all granted permission to access and use the medical records described in our study. Written informed consents were obtained from all participants.
Not applicable.

Competing interests

The authors declare that they have no competing interests.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
Zurück zum Zitat Mendelson EB, Böhm-Vélez M, Berg WA, et al. ACR BI-RADS® ultrasound. ACR BI-RADS® atlas, breast imaging reporting and data system. Reston: American College of Radiology; 2013. Mendelson EB, Böhm-Vélez M, Berg WA, et al. ACR BI-RADS® ultrasound. ACR BI-RADS® atlas, breast imaging reporting and data system. Reston: American College of Radiology; 2013.
2.
Zurück zum Zitat Sadoughi F, Kazemy Z, Hamedan F, Owji L, Rahmanikatigari M, Azadboni TT. Artificial intelligence methods for the diagnosis of breast cancer by image processing: a review. Breast Cancer (Dove Med Press). 2018;10:219–30. Sadoughi F, Kazemy Z, Hamedan F, Owji L, Rahmanikatigari M, Azadboni TT. Artificial intelligence methods for the diagnosis of breast cancer by image processing: a review. Breast Cancer (Dove Med Press). 2018;10:219–30.
3.
Zurück zum Zitat Bi WL, Hosny A, Schabath MB, Giger ML, Birkbak NJ, Mehrtash A. Artificial intelligence in cancer imaging: clinical challenges and applications. CA Cancer J Clin. 2019;69(2):127–57.PubMedPubMedCentral Bi WL, Hosny A, Schabath MB, Giger ML, Birkbak NJ, Mehrtash A. Artificial intelligence in cancer imaging: clinical challenges and applications. CA Cancer J Clin. 2019;69(2):127–57.PubMedPubMedCentral
4.
Zurück zum Zitat Moon WK, Chen IL, Chang JM, Shin SU, Lo CM, Chang RF. The adaptive computer-aided diagnosis system based on tumor sizes for the classification of breast tumors detected at screening ultrasound. Ultrasonics. 2017;76:70–7.PubMedCrossRef Moon WK, Chen IL, Chang JM, Shin SU, Lo CM, Chang RF. The adaptive computer-aided diagnosis system based on tumor sizes for the classification of breast tumors detected at screening ultrasound. Ultrasonics. 2017;76:70–7.PubMedCrossRef
5.
Zurück zum Zitat Becker AS, Mueller M, Stoffel E, Marcon M, Ghafoor S, Boss A. Classification of breast cancer in ultrasound imaging using a generic deep learning analysis software: a pilot study. Br J Radiol. 2018;91(1083):20170576.PubMedPubMedCentral Becker AS, Mueller M, Stoffel E, Marcon M, Ghafoor S, Boss A. Classification of breast cancer in ultrasound imaging using a generic deep learning analysis software: a pilot study. Br J Radiol. 2018;91(1083):20170576.PubMedPubMedCentral
6.
Zurück zum Zitat Shen WC, Chang RF, Moon WK. Computer aided classification system for breast ultrasound based on breast imaging reporting and data system (BI-RADS). Ultrasound Med Biol. 2007;33(11):1688–98.PubMedCrossRef Shen WC, Chang RF, Moon WK. Computer aided classification system for breast ultrasound based on breast imaging reporting and data system (BI-RADS). Ultrasound Med Biol. 2007;33(11):1688–98.PubMedCrossRef
7.
Zurück zum Zitat Rodríguez-Cristerna A, Gómez-Flores W, de Albuquerque Pereira WC. A computer-aided diagnosis system for breast ultrasound based on weighted BI-RADS classes. Comput Methods Prog Biomed. 2018;153:33–40.CrossRef Rodríguez-Cristerna A, Gómez-Flores W, de Albuquerque Pereira WC. A computer-aided diagnosis system for breast ultrasound based on weighted BI-RADS classes. Comput Methods Prog Biomed. 2018;153:33–40.CrossRef
8.
Zurück zum Zitat Jamshidii N, Chang J, Mock K, Nguyen B, Dauphine C, Kuo MD. Evaluation of the predictive ability of ultrasound-based assessment of breast cancer using BI-RADS natural language reporting against commercial transcriptome-based tests. PLoS One. 2020;15(1):e0226634.PubMedPubMedCentralCrossRef Jamshidii N, Chang J, Mock K, Nguyen B, Dauphine C, Kuo MD. Evaluation of the predictive ability of ultrasound-based assessment of breast cancer using BI-RADS natural language reporting against commercial transcriptome-based tests. PLoS One. 2020;15(1):e0226634.PubMedPubMedCentralCrossRef
9.
Zurück zum Zitat Ciritsis A, Rossi C, Eberhard M, Marcon M, Becker AS, Boss A. Automatic classification of ultrasound breast lesions using a deep convolutional neural network mimicking human decision-making. Eur Radiol. 2019;29(10):5458–68.PubMedCrossRef Ciritsis A, Rossi C, Eberhard M, Marcon M, Becker AS, Boss A. Automatic classification of ultrasound breast lesions using a deep convolutional neural network mimicking human decision-making. Eur Radiol. 2019;29(10):5458–68.PubMedCrossRef
10.
Zurück zum Zitat Huang Y, Han L, Dou H, Luo H, Yuan Z, Liu Q, et al. Two-stage CNNs for computerized BI-RADS categorization in breast ultrasound images. Biomed Eng Online. 2019;18(1):8.PubMedPubMedCentralCrossRef Huang Y, Han L, Dou H, Luo H, Yuan Z, Liu Q, et al. Two-stage CNNs for computerized BI-RADS categorization in breast ultrasound images. Biomed Eng Online. 2019;18(1):8.PubMedPubMedCentralCrossRef
11.
Zurück zum Zitat Shan J, Alam SK, Garra B, Zhang Y, Ahmed T. Computer-aided diagnosis for breast ultrasound using computerized BI-RADS features and machine learning methods. Ultrasound Med Biol. 2016;42(4):980–8.PubMedCrossRef Shan J, Alam SK, Garra B, Zhang Y, Ahmed T. Computer-aided diagnosis for breast ultrasound using computerized BI-RADS features and machine learning methods. Ultrasound Med Biol. 2016;42(4):980–8.PubMedCrossRef
12.
Zurück zum Zitat Moon WK, Lo CM, Chang JM, Huang CS, Chen JH, Chang RF. Quantitative ultrasound analysis for classification of BI-RADS category 3 breast masses. J Digit Imaging. 2013;26(6):1091–8.PubMedPubMedCentralCrossRef Moon WK, Lo CM, Chang JM, Huang CS, Chen JH, Chang RF. Quantitative ultrasound analysis for classification of BI-RADS category 3 breast masses. J Digit Imaging. 2013;26(6):1091–8.PubMedPubMedCentralCrossRef
13.
Zurück zum Zitat Wu T, Sultan LR, Tian J, Cary TW, Sehgal CM. Machine learning for diagnostic ultrasound of triple-negative breast cancer. Breast Cancer Res Treat. 2019;173(2):365–73.PubMedCrossRef Wu T, Sultan LR, Tian J, Cary TW, Sehgal CM. Machine learning for diagnostic ultrasound of triple-negative breast cancer. Breast Cancer Res Treat. 2019;173(2):365–73.PubMedCrossRef
14.
Zurück zum Zitat Lee SE, Han K, Kwak JY, Lee E, Kim EK. Radiomics of US texture features in differential diagnosis between triple-negative breast cancer and fibroadenoma. Sci Rep. 2018;8(1):13546.PubMedPubMedCentralCrossRef Lee SE, Han K, Kwak JY, Lee E, Kim EK. Radiomics of US texture features in differential diagnosis between triple-negative breast cancer and fibroadenoma. Sci Rep. 2018;8(1):13546.PubMedPubMedCentralCrossRef
15.
Zurück zum Zitat Venkatesh SS, Levenback BJ, Sultan LR, Bouzghar G, Sehgal CM. Going beyond a first reader: a machine learning methodology for optimizing cost and performance in breast ultrasound diagnosis. Ultrasound Med Biol. 2015;41(12):3148–62.PubMedCrossRef Venkatesh SS, Levenback BJ, Sultan LR, Bouzghar G, Sehgal CM. Going beyond a first reader: a machine learning methodology for optimizing cost and performance in breast ultrasound diagnosis. Ultrasound Med Biol. 2015;41(12):3148–62.PubMedCrossRef
16.
Zurück zum Zitat Le EPV, Wang Y, Huang Y, Hickman S, Gilbert FJ. Artificial intelligence in breast imaging. Clin Radiol. 2019;74(5):357–66.PubMedCrossRef Le EPV, Wang Y, Huang Y, Hickman S, Gilbert FJ. Artificial intelligence in breast imaging. Clin Radiol. 2019;74(5):357–66.PubMedCrossRef
17.
Zurück zum Zitat Kim K, Song MK, Kim EK, Yoon JH. Clinical application of S-detect to breast masses on ultrasonography: a study evaluating the diagnostic performance and agreement with a dedicated breast radiologist. Ultrasonography. 2017;36:3–9.PubMedCrossRef Kim K, Song MK, Kim EK, Yoon JH. Clinical application of S-detect to breast masses on ultrasonography: a study evaluating the diagnostic performance and agreement with a dedicated breast radiologist. Ultrasonography. 2017;36:3–9.PubMedCrossRef
18.
Zurück zum Zitat Zhang L, Li J, Xiao Y, Cui H, Du G, Wang Y, et al. Identifying ultrasound and clinical features of breast cancer molecular subtypes by ensemble decision. Sci Rep. 2015;5:11085.PubMedPubMedCentralCrossRef Zhang L, Li J, Xiao Y, Cui H, Du G, Wang Y, et al. Identifying ultrasound and clinical features of breast cancer molecular subtypes by ensemble decision. Sci Rep. 2015;5:11085.PubMedPubMedCentralCrossRef
19.
Zurück zum Zitat Klimonda Z, Karwat P, Dobruch-Sobczak K, Piotrzkowska-Wróblewska H, Litniewski J. Breast-lesions characterization using quantitative ultrasound features of peritumoral tissue. Sci Rep. 2019;9(1):7963.PubMedPubMedCentralCrossRef Klimonda Z, Karwat P, Dobruch-Sobczak K, Piotrzkowska-Wróblewska H, Litniewski J. Breast-lesions characterization using quantitative ultrasound features of peritumoral tissue. Sci Rep. 2019;9(1):7963.PubMedPubMedCentralCrossRef
20.
Zurück zum Zitat Tsui PH, Chen CK, Kuo WH, Chang KJ, Fang J, Ma HY, et al. Small-window parametric imaging based on information entropy for ultrasound tissue characterization. Sci Rep. 2017;7:41004.PubMedPubMedCentralCrossRef Tsui PH, Chen CK, Kuo WH, Chang KJ, Fang J, Ma HY, et al. Small-window parametric imaging based on information entropy for ultrasound tissue characterization. Sci Rep. 2017;7:41004.PubMedPubMedCentralCrossRef
21.
Zurück zum Zitat Marcon M, Ciritsis A, Rossi C, Becker AS, Berger N, Wurnig MC, et al. Diagnostic performance of machine learning applied to texture analysis-derived features for breast lesion characterisation at automated breast ultrasound: a pilot study. Eur Radiol Exp. 2019;3(1):44.PubMedPubMedCentralCrossRef Marcon M, Ciritsis A, Rossi C, Becker AS, Berger N, Wurnig MC, et al. Diagnostic performance of machine learning applied to texture analysis-derived features for breast lesion characterisation at automated breast ultrasound: a pilot study. Eur Radiol Exp. 2019;3(1):44.PubMedPubMedCentralCrossRef
22.
Zurück zum Zitat Baldwin P. Breast calcification imaging. Radiol Technol. 2013;84(4):383M–404M. quiz 405M–408M.PubMed Baldwin P. Breast calcification imaging. Radiol Technol. 2013;84(4):383M–404M. quiz 405M–408M.PubMed
23.
Zurück zum Zitat Stöblen F, Landt S, Ishaq R, Stelkens-Gebhardt R, Rezai M, Skaane P, et al. High-frequency breast ultrasound for the detection of microcalcifications and associated masses in BI-RADS 4a patients. Anticancer Res. 2011;31(8):2575–81.PubMed Stöblen F, Landt S, Ishaq R, Stelkens-Gebhardt R, Rezai M, Skaane P, et al. High-frequency breast ultrasound for the detection of microcalcifications and associated masses in BI-RADS 4a patients. Anticancer Res. 2011;31(8):2575–81.PubMed
24.
Zurück zum Zitat Hsu W, Zhou X, Petruse A, Chau N, Lee-Felker S, Hoyt A, et al. Role of clinical and imaging risk factors in predicting breast Cancer diagnosis among BI-RADS 4 cases. Clin Breast Cancer. 2019;19(1):e142–51.PubMedCrossRef Hsu W, Zhou X, Petruse A, Chau N, Lee-Felker S, Hoyt A, et al. Role of clinical and imaging risk factors in predicting breast Cancer diagnosis among BI-RADS 4 cases. Clin Breast Cancer. 2019;19(1):e142–51.PubMedCrossRef
25.
Zurück zum Zitat Elverici E, Barça AN, Aktaş H, Özsoy A, Zengin B, Çavuşoğlu M, et al. Nonpalpable BI-RADS 4 breast lesions: sonographic findings and pathology correlation. Diagn Interv Radiol. 2015;21(3):189–94.PubMedPubMedCentralCrossRef Elverici E, Barça AN, Aktaş H, Özsoy A, Zengin B, Çavuşoğlu M, et al. Nonpalpable BI-RADS 4 breast lesions: sonographic findings and pathology correlation. Diagn Interv Radiol. 2015;21(3):189–94.PubMedPubMedCentralCrossRef
26.
Zurück zum Zitat Zhao C, Xiao M, Jiang Y, Liu H, Wang M, Wang H, et al. Feasibility of computer-assisted diagnosis for breast ultrasound: the results of the diagnostic performance of S-detect from a single center in China. Cancer Manag Res. 2019;11:921–30.PubMedPubMedCentralCrossRef Zhao C, Xiao M, Jiang Y, Liu H, Wang M, Wang H, et al. Feasibility of computer-assisted diagnosis for breast ultrasound: the results of the diagnostic performance of S-detect from a single center in China. Cancer Manag Res. 2019;11:921–30.PubMedPubMedCentralCrossRef
Metadaten
Titel
Application of ultrasound artificial intelligence in the differential diagnosis between benign and malignant breast lesions of BI-RADS 4A
verfasst von
Sihua Niu
Jianhua Huang
Jia Li
Xueling Liu
Dan Wang
Ruifang Zhang
Yingyan Wang
Huiming Shen
Min Qi
Yi Xiao
Mengyao Guan
Haiyan Liu
Diancheng Li
Feifei Liu
Xiuming Wang
Yu Xiong
Siqi Gao
Xue Wang
Jiaan Zhu
Publikationsdatum
01.12.2020
Verlag
BioMed Central
Erschienen in
BMC Cancer / Ausgabe 1/2020
Elektronische ISSN: 1471-2407
DOI
https://doi.org/10.1186/s12885-020-07413-z

Weitere Artikel der Ausgabe 1/2020

BMC Cancer 1/2020 Zur Ausgabe

„Überwältigende“ Evidenz für Tripeltherapie beim metastasierten Prostata-Ca.

22.05.2024 Prostatakarzinom Nachrichten

Patienten mit metastasiertem hormonsensitivem Prostatakarzinom sollten nicht mehr mit einer alleinigen Androgendeprivationstherapie (ADT) behandelt werden, mahnt ein US-Team nach Sichtung der aktuellen Datenlage. Mit einer Tripeltherapie haben die Betroffenen offenbar die besten Überlebenschancen.

So sicher sind Tattoos: Neue Daten zur Risikobewertung

22.05.2024 Melanom Nachrichten

Das größte medizinische Problem bei Tattoos bleiben allergische Reaktionen. Melanome werden dadurch offensichtlich nicht gefördert, die Farbpigmente könnten aber andere Tumoren begünstigen.

CAR-M-Zellen: Warten auf das große Fressen

22.05.2024 Onkologische Immuntherapie Nachrichten

Auch myeloide Immunzellen lassen sich mit chimären Antigenrezeptoren gegen Tumoren ausstatten. Solche CAR-Fresszell-Therapien werden jetzt für solide Tumoren entwickelt. Künftig soll dieser Prozess nicht mehr ex vivo, sondern per mRNA im Körper der Betroffenen erfolgen.

Blutdrucksenkung könnte Uterusmyome verhindern

Frauen mit unbehandelter oder neu auftretender Hypertonie haben ein deutlich erhöhtes Risiko für Uterusmyome. Eine Therapie mit Antihypertensiva geht hingegen mit einer verringerten Inzidenz der gutartigen Tumoren einher.

Update Onkologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.