Erschienen in: Skeletal Radiology 8/2023

21.03.2023 | Scientific Article

Comparison of deep learning-based reconstruction of PROPELLER Shoulder MRI with conventional reconstruction

verfasst von: Seok Hahn, Jisook Yi, Ho-Joon Lee, Yedaun Lee, Joonsung Lee, Xinzeng Wang, Maggie Fung

Erschienen in: Skeletal Radiology | Ausgabe 8/2023

Einloggen, um Zugang zu erhalten

Abstract

Objective

To compare the image quality and agreement among conventional and accelerated periodically rotated overlapping parallel lines with enhanced reconstruction (PROPELLER) MRI with both conventional reconstruction (CR) and deep learning–based reconstruction (DLR) methods for evaluation of shoulder.

Materials and methods

We included patients who underwent conventional (acquisition time, 8 min) and accelerated (acquisition time, 4 min and 24 s; 45% reduction) PROPELLER shoulder MRI using both CR and DLR methods between February 2021 and February 2022 on a 3 T MRI system. Quantitative evaluation was performed by calculating the signal-to-noise ratio (SNR). Two musculoskeletal radiologists compared the image quality using conventional sequence with CR as the reference standard. Interobserver agreement between image sets for evaluating shoulder was analyzed using weighted/unweighted kappa statistics.

Results

Ninety-two patients with 100 shoulder MRI scans were included. Conventional sequence with DLR had the highest SNR (P < .001), followed by accelerated sequence with DLR, conventional sequence with CR, and accelerated sequence with CR. Comparison of image quality by both readers revealed that conventional sequence with DLR (P = .003 and P < .001) and accelerated sequence with DLR (P = .016 and P < .001) had better image quality than the conventional sequence with CR. Interobserver agreement was substantial to almost perfect for detecting shoulder abnormalities (κ = 0.600–0.884). Agreement between the image sets was substantial to almost perfect (κ = 0.691–1).

Conclusion

Accelerated PROPELLER with DLR showed even better image quality than conventional PROPELLER with CR and interobserver agreement for shoulder pathologies comparable to that of conventional PROPELLER with CR, despite the shorter scan time.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Subhas N, Benedick A, Obuchowski NA, Polster JM, Beltran LS, Schils J, et al. Comparison of a fast 5-minute shoulder MRI protocol with a standard shoulder MRI protocol: a multiinstitutional multireader study. AJR Am J Roentgenol. 2017;208(4):W146–54.CrossRefPubMed Subhas N, Benedick A, Obuchowski NA, Polster JM, Beltran LS, Schils J, et al. Comparison of a fast 5-minute shoulder MRI protocol with a standard shoulder MRI protocol: a multiinstitutional multireader study. AJR Am J Roentgenol. 2017;208(4):W146–54.CrossRefPubMed
2.
Zurück zum Zitat Lee SH, Lee YH, Song H-T, Suh J-S. Rapid acquisition of magnetic resonance imaging of the shoulder using three-dimensional fast spin echo sequence with compressed sensing. Magn Reson Imaging. 2017;42:152–7.CrossRefPubMed Lee SH, Lee YH, Song H-T, Suh J-S. Rapid acquisition of magnetic resonance imaging of the shoulder using three-dimensional fast spin echo sequence with compressed sensing. Magn Reson Imaging. 2017;42:152–7.CrossRefPubMed
3.
Zurück zum Zitat Del Grande F, Rashidi A, Luna R, Delcogliano M, Stern SE, Dalili D, et al. Five-minute five-sequence knee MRI using combined simultaneous multislice and parallel imaging acceleration: comparison with 10-minute parallel imaging knee MRI. Radiology. 2021;299(3):635–46.CrossRefPubMed Del Grande F, Rashidi A, Luna R, Delcogliano M, Stern SE, Dalili D, et al. Five-minute five-sequence knee MRI using combined simultaneous multislice and parallel imaging acceleration: comparison with 10-minute parallel imaging knee MRI. Radiology. 2021;299(3):635–46.CrossRefPubMed
4.
Zurück zum Zitat Fritz J, Guggenberger R, Grande FD. Rapid musculoskeletal MRI in 2021: clinical application of advanced accelerated techniques. AJR Am J Roentgenol. 2021;216(3):718–33.CrossRefPubMed Fritz J, Guggenberger R, Grande FD. Rapid musculoskeletal MRI in 2021: clinical application of advanced accelerated techniques. AJR Am J Roentgenol. 2021;216(3):718–33.CrossRefPubMed
5.
Zurück zum Zitat Kijowski R, Rosas H, Samsonov A, King K, Peters R, Liu F. Knee imaging: Rapid three-dimensional fast spin-echo using compressed sensing. J Magn Reson Imaging. 2017;45(6):1712–22.CrossRefPubMed Kijowski R, Rosas H, Samsonov A, King K, Peters R, Liu F. Knee imaging: Rapid three-dimensional fast spin-echo using compressed sensing. J Magn Reson Imaging. 2017;45(6):1712–22.CrossRefPubMed
6.
Zurück zum Zitat Hou B, Li Y, Xiong Y, Morelli JN, Wang J, Liu C, et al. Comparison of CAIPIRINHA-accelerated 3D fat-saturated-SPACE MRI with 2D MRI sequences for the assessment of shoulder pathology. Eur Radiol. 2022;32(1):593–601.CrossRefPubMed Hou B, Li Y, Xiong Y, Morelli JN, Wang J, Liu C, et al. Comparison of CAIPIRINHA-accelerated 3D fat-saturated-SPACE MRI with 2D MRI sequences for the assessment of shoulder pathology. Eur Radiol. 2022;32(1):593–601.CrossRefPubMed
7.
Zurück zum Zitat Pipe JG. Motion correction with PROPELLER MRI: application to head motion and free breathing cardiac imaging. Magn Reson Med. 1999;42(5):963–9.CrossRefPubMed Pipe JG. Motion correction with PROPELLER MRI: application to head motion and free breathing cardiac imaging. Magn Reson Med. 1999;42(5):963–9.CrossRefPubMed
8.
Zurück zum Zitat Fellner C, Menzel C, Fellner F, Ginthoer C, Zorger N, Schreyer A, et al. BLADE in sagittal T2-weighted MR imaging of the cervical spine. AJNR Am J Neuroradiol. 2010;31(4):674–81.CrossRefPubMedPubMedCentral Fellner C, Menzel C, Fellner F, Ginthoer C, Zorger N, Schreyer A, et al. BLADE in sagittal T2-weighted MR imaging of the cervical spine. AJNR Am J Neuroradiol. 2010;31(4):674–81.CrossRefPubMedPubMedCentral
9.
Zurück zum Zitat Forbes KP, Pipe JG, Karis JP, Farthing V, Heiserman JE. Brain imaging in the unsedated pediatric patient: comparison of periodically rotated overlapping parallel lines with enhanced reconstruction and single-shot fast spin-echo sequences. AJNR Am J Neuroradiol. 2003;24(5):794–8.PubMedPubMedCentral Forbes KP, Pipe JG, Karis JP, Farthing V, Heiserman JE. Brain imaging in the unsedated pediatric patient: comparison of periodically rotated overlapping parallel lines with enhanced reconstruction and single-shot fast spin-echo sequences. AJNR Am J Neuroradiol. 2003;24(5):794–8.PubMedPubMedCentral
10.
Zurück zum Zitat Ciet P, Serra G, Bertolo S, Spronk S, Ros M, Fraioli F, et al. Assessment of CF lung disease using motion corrected PROPELLER MRI: a comparison with CT. Eur Radiol. 2016;26(3):780–7.CrossRefPubMed Ciet P, Serra G, Bertolo S, Spronk S, Ros M, Fraioli F, et al. Assessment of CF lung disease using motion corrected PROPELLER MRI: a comparison with CT. Eur Radiol. 2016;26(3):780–7.CrossRefPubMed
11.
Zurück zum Zitat Lane BF, Vandermeer FQ, Oz RC, Irwin EW, McMillan AB, Wong-You-Cheong JJ. Comparison of sagittal T2-weighted BLADE and fast spin-echo MRI of the female pelvis for motion artifact and lesion detection. AJR Am J Roentgenol. 2011;197(2):W307–13.CrossRefPubMed Lane BF, Vandermeer FQ, Oz RC, Irwin EW, McMillan AB, Wong-You-Cheong JJ. Comparison of sagittal T2-weighted BLADE and fast spin-echo MRI of the female pelvis for motion artifact and lesion detection. AJR Am J Roentgenol. 2011;197(2):W307–13.CrossRefPubMed
12.
Zurück zum Zitat Dietrich TJ, Ulbrich EJ, Zanetti M, Fucentese SF, Pfirrmann CWA. PROPELLER technique to improve image quality of MRI of the shoulder. AJR Am J Roentgenol. 2011;197(6):W1093–100.CrossRefPubMed Dietrich TJ, Ulbrich EJ, Zanetti M, Fucentese SF, Pfirrmann CWA. PROPELLER technique to improve image quality of MRI of the shoulder. AJR Am J Roentgenol. 2011;197(6):W1093–100.CrossRefPubMed
13.
Zurück zum Zitat Nagatomo K, Yabuuchi H, Yamasaki Y, Narita H, Kumazawa S, Kojima T, et al. Efficacy of periodically rotated overlapping parallel lines with enhanced reconstruction (PROPELLER) for shoulder magnetic resonance (MR) imaging. Eur J Radiol. 2016;85(10):1735–43.CrossRefPubMed Nagatomo K, Yabuuchi H, Yamasaki Y, Narita H, Kumazawa S, Kojima T, et al. Efficacy of periodically rotated overlapping parallel lines with enhanced reconstruction (PROPELLER) for shoulder magnetic resonance (MR) imaging. Eur J Radiol. 2016;85(10):1735–43.CrossRefPubMed
14.
Zurück zum Zitat Lavdas E, Vlychou M, Zaloni E, Vassiou K, Tsagkalis A, Dailiana Z, et al. Elimination of motion and pulsation artifacts using BLADE sequences in shoulder MR imaging. Skeletal Radiol. 2015;44(11):1619–26.CrossRefPubMed Lavdas E, Vlychou M, Zaloni E, Vassiou K, Tsagkalis A, Dailiana Z, et al. Elimination of motion and pulsation artifacts using BLADE sequences in shoulder MR imaging. Skeletal Radiol. 2015;44(11):1619–26.CrossRefPubMed
15.
Zurück zum Zitat Kohli A, Pilkinton DT, Xi Y, Cho G, Moore D, Mohammadi D, et al. Image quality improvement and motion degradation reduction in shoulder MR imaging: comparison of BLADE and rectilinear techniques at 3-Tesla scanning. Skeletal Radiol. 2022;51(12):2291–7. Kohli A, Pilkinton DT, Xi Y, Cho G, Moore D, Mohammadi D, et al. Image quality improvement and motion degradation reduction in shoulder MR imaging: comparison of BLADE and rectilinear techniques at 3-Tesla scanning. Skeletal Radiol. 2022;51(12):2291–7.
17.
Zurück zum Zitat Ueda T, Ohno Y, Yamamoto K, Murayama K, Ikedo M, Yui M, et al. Deep Learning Reconstruction of Diffusion-weighted MRI Improves Image Quality for Prostatic Imaging. Radiology. 2022;303(2):373–81.CrossRefPubMed Ueda T, Ohno Y, Yamamoto K, Murayama K, Ikedo M, Yui M, et al. Deep Learning Reconstruction of Diffusion-weighted MRI Improves Image Quality for Prostatic Imaging. Radiology. 2022;303(2):373–81.CrossRefPubMed
18.
Zurück zum Zitat Recht MP, Zbontar J, Sodickson DK, Knoll F, Yakubova N, Sriam A, et al. Using deep learning to accelerate knee MRI at 3 T: results of an interchangeability study. AJR Am J Roentgenol. 2020;215(6):1421.CrossRefPubMedPubMedCentral Recht MP, Zbontar J, Sodickson DK, Knoll F, Yakubova N, Sriam A, et al. Using deep learning to accelerate knee MRI at 3 T: results of an interchangeability study. AJR Am J Roentgenol. 2020;215(6):1421.CrossRefPubMedPubMedCentral
19.
Zurück zum Zitat Kim M, Kim HS, Kim HJ, Park JE, Park SY, Kim Y, et al. Thin-slice pituitary MRI with deep learning-based reconstruction: diagnostic performance in a postoperative setting. Radiology. 2021;298(1):114–22.CrossRefPubMed Kim M, Kim HS, Kim HJ, Park JE, Park SY, Kim Y, et al. Thin-slice pituitary MRI with deep learning-based reconstruction: diagnostic performance in a postoperative setting. Radiology. 2021;298(1):114–22.CrossRefPubMed
20.
Zurück zum Zitat Sun S, Tan ET, Mintz DN, Sahr M, Endo Y, Nguyen J, et al. Evaluation of deep learning reconstructed high-resolution 3D lumbar spine MRI. Eur Radiol. 2022;32(9):6167–77.CrossRefPubMed Sun S, Tan ET, Mintz DN, Sahr M, Endo Y, Nguyen J, et al. Evaluation of deep learning reconstructed high-resolution 3D lumbar spine MRI. Eur Radiol. 2022;32(9):6167–77.CrossRefPubMed
21.
Zurück zum Zitat Yasaka K, Tanishima T, Ohtake Y, Tajima T, Akai H, Ohtomo K, et al. Deep learning reconstruction for 1.5 T cervical spine MRI: effect on interobserver agreement in the evaluation of degenerative changes. Eur Radiol 2022;32(9):6118–6125. Yasaka K, Tanishima T, Ohtake Y, Tajima T, Akai H, Ohtomo K, et al. Deep learning reconstruction for 1.5 T cervical spine MRI: effect on interobserver agreement in the evaluation of degenerative changes. Eur Radiol 2022;32(9):6118–6125.
22.
Zurück zum Zitat Hahn S, Yi J, Lee H-J, Lee Y, Lim Y-J, Bang J-Y, et al. Image quality and diagnostic performance of accelerated shoulder MRI with deep learning–based reconstruction. AJR Am J Roentgenol 2022;218(3):506–16. Hahn S, Yi J, Lee H-J, Lee Y, Lim Y-J, Bang J-Y, et al. Image quality and diagnostic performance of accelerated shoulder MRI with deep learning–based reconstruction. AJR Am J Roentgenol 2022;218(3):506–16.
23.
Zurück zum Zitat Johnson PM, Tong A, Donthireddy A, Melamud K, Petrocelli R, Smereka P, et al. Deep learning reconstruction enables highly accelerated biparametric mr imaging of the prostate. J Magn Reson Imaging. 2022;56(1):184–95.CrossRefPubMed Johnson PM, Tong A, Donthireddy A, Melamud K, Petrocelli R, Smereka P, et al. Deep learning reconstruction enables highly accelerated biparametric mr imaging of the prostate. J Magn Reson Imaging. 2022;56(1):184–95.CrossRefPubMed
24.
Zurück zum Zitat van der Velde N, Hassing HC, Bakker BJ, Wielopolski PA, Lebel RM, Janich MA, et al. Improvement of late gadolinium enhancement image quality using a deep learning-based reconstruction algorithm and its influence on myocardial scar quantification. Eur Radiol. 2021;31(6):3846–55.CrossRefPubMed van der Velde N, Hassing HC, Bakker BJ, Wielopolski PA, Lebel RM, Janich MA, et al. Improvement of late gadolinium enhancement image quality using a deep learning-based reconstruction algorithm and its influence on myocardial scar quantification. Eur Radiol. 2021;31(6):3846–55.CrossRefPubMed
25.
Zurück zum Zitat Pimpalkhute VA, Page R, Kothari A, Bhurchandi KM, Kamble VM. Digital image noise estimation using DWT coefficients. IEEE Trans Image Process. 2021;30:1962–72.CrossRefPubMed Pimpalkhute VA, Page R, Kothari A, Bhurchandi KM, Kamble VM. Digital image noise estimation using DWT coefficients. IEEE Trans Image Process. 2021;30:1962–72.CrossRefPubMed
26.
Zurück zum Zitat Liu P, Wang Q, Peng C, Luo B, Zhang J. Combined application of isotropic three-dimensional fast spin echo (3D-FSE-Cube) with 2-point Dixon fat/water separation (FLEX) and 3D-FSE-cube in MR dacryocystography. Br J Radiol. 2019;92(1094):20180157. 1 Liu P, Wang Q, Peng C, Luo B, Zhang J. Combined application of isotropic three-dimensional fast spin echo (3D-FSE-Cube) with 2-point Dixon fat/water separation (FLEX) and 3D-FSE-cube in MR dacryocystography. Br J Radiol. 2019;92(1094):20180157. 1
27.
Zurück zum Zitat Siegel S. Nonparametric statistics. Am Stat. 1957;11(3):13–9. Siegel S. Nonparametric statistics. Am Stat. 1957;11(3):13–9.
28.
Zurück zum Zitat Wilcoxon F. Individual comparisons by ranking methods. Biom Bull. 1945;1:80–3.CrossRef Wilcoxon F. Individual comparisons by ranking methods. Biom Bull. 1945;1:80–3.CrossRef
29.
Zurück zum Zitat Rayat CS. Statistical Methods in Medical Research. Singapore: Springer; 2018. p. 69–79. Rayat CS. Statistical Methods in Medical Research. Singapore: Springer; 2018. p. 69–79.
30.
Zurück zum Zitat Landis JR, Koch GG. The measurement of observer agreement for categorical data. Biometrics. 1977;33:159–74.CrossRefPubMed Landis JR, Koch GG. The measurement of observer agreement for categorical data. Biometrics. 1977;33:159–74.CrossRefPubMed
31.
Zurück zum Zitat Holm S. A simple sequentially rejective multiple test procedure. Scand J Stat. 1979;65–70. Holm S. A simple sequentially rejective multiple test procedure. Scand J Stat. 1979;65–70.
32.
Zurück zum Zitat Glockner JF, Hu HH, Stanley DW, Angelos L, King K. Parallel MR imaging: a user’s guide. Radiographics. 2005;25(5):1279–97.CrossRefPubMed Glockner JF, Hu HH, Stanley DW, Angelos L, King K. Parallel MR imaging: a user’s guide. Radiographics. 2005;25(5):1279–97.CrossRefPubMed
33.
Zurück zum Zitat Reeder SB. Measurement of signal-to-noise ratio and parallel imaging. Parallel imaging in clinical MR applications. Springer; 2007, p. 49–61. Reeder SB. Measurement of signal-to-noise ratio and parallel imaging. Parallel imaging in clinical MR applications. Springer; 2007, p. 49–61.
34.
35.
Zurück zum Zitat Chang Y, Pipe JG, Karis JP, Gibbs WN, Zwart NR, Schar M. The effects of SENSE on PROPELLER imaging. Magn Reson Med. 2015;74(6):1598–608.CrossRefPubMed Chang Y, Pipe JG, Karis JP, Gibbs WN, Zwart NR, Schar M. The effects of SENSE on PROPELLER imaging. Magn Reson Med. 2015;74(6):1598–608.CrossRefPubMed
Metadaten
Titel
Comparison of deep learning-based reconstruction of PROPELLER Shoulder MRI with conventional reconstruction
verfasst von
Seok Hahn
Jisook Yi
Ho-Joon Lee
Yedaun Lee
Joonsung Lee
Xinzeng Wang
Maggie Fung
Publikationsdatum
21.03.2023
Verlag
Springer Berlin Heidelberg
Erschienen in
Skeletal Radiology / Ausgabe 8/2023
Print ISSN: 0364-2348
Elektronische ISSN: 1432-2161
DOI
https://doi.org/10.1007/s00256-023-04321-8

Weitere Artikel der Ausgabe 8/2023

Skeletal Radiology 8/2023 Zur Ausgabe

Test Yourself: Answer

Neck pain in a teenager

Test Yourself: Question

Neck pain in a teenager

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

„Nur wer sich gut aufgehoben fühlt, kann auch für Patientensicherheit sorgen“

13.04.2024 Klinik aktuell Kongressbericht

Die Teilnehmer eines Forums beim DGIM-Kongress waren sich einig: Fehler in der Medizin sind häufig in ungeeigneten Prozessen und mangelnder Kommunikation begründet. Gespräche mit Patienten und im Team können helfen.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.