Skip to main content
Erschienen in: Archives of Gynecology and Obstetrics 4/2024

Open Access 27.03.2023 | Maternal-Fetal Medicine

D-Transposition of the great arteries with restrictive foramen ovale in the fetus: the dilemma of predicting the need for postnatal urgent balloon atrial septostomy

verfasst von: I. Gottschalk, A. Walter, T. Menzel, E. C. Weber, S. Wendt, N. Sreeram, U. Gembruch, C. Berg, J. S. Abel

Erschienen in: Archives of Gynecology and Obstetrics | Ausgabe 4/2024

Abstract

Objective

Restrictive foramen ovale (FO) in dextro-transposition of the great arteries (d-TGA) with intact ventricular septum may lead to severe life-threatening hypoxia within the first hours of life, making urgent balloon atrial septostomy (BAS) inevitable. Reliable prenatal prediction of restrictive FO is crucial in these cases. However, current prenatal echocardiographic markers show low predictive value, and prenatal prediction often fails with fatal consequences for a subset of newborns. In this study, we described our experience and aimed to identify reliable predictive markers for BAS.

Methods

We included 45 fetuses with isolated d-TGA that were diagnosed and delivered between 2010 and 2022 in two large German tertiary referral centers. Inclusion criteria were the availability of former prenatal ultrasound reports, of stored echocardiographic videos and still images, which had to be obtained within the last 14 days prior to delivery and that were of sufficient quality for retrospective re-analysis. Cardiac parameters were retrospectively assessed and their predictive value was evaluated.

Results

Among the 45 included fetuses with d-TGA, 22 neonates had restrictive FO postnatally and required urgent BAS within the first 24 h of life. In contrast, 23 neonates had normal FO anatomy, but 4 of them unexpectedly showed inadequate interatrial mixing despite their normal FO anatomy, rapidly developed hypoxia and also required urgent BAS (‘bad mixer’). Overall, 26 (58%) neonates required urgent BAS, whereas 19 (42%) achieved good O2 saturation and did not undergo urgent BAS. In the former prenatal ultrasound reports, restrictive FO with subsequent urgent BAS was correctly predicted in 11 of 22 cases (50% sensitivity), whereas a normal FO anatomy was correctly predicted in 19 of 23 cases (83% specificity). After current re-analysis of the stored videos and images, we identified three highly significant markers for restrictive FO: a FO diameter < 7 mm (p < 0.01), a fixed (p = 0.035) and a hypermobile (p = 0.014) FO flap. The maximum systolic flow velocities in the pulmonary veins were also significantly increased in restrictive FO (p = 0.021), but no cut-off value to reliably predict restrictive FO could be identified. If the above markers are applied, all 22 cases with restrictive FO and all 23 cases with normal FO anatomy could correctly be predicted (100% positive predictive value). Correct prediction of urgent BAS also succeeded in all 22 cases with restrictive FO (100% PPV), but naturally failed in 4 of the 23 cases with correctly predicted normal FO (‘bad mixer’) (82.6% negative predictive value).

Conclusion

Precise assessment of FO size and FO flap motility allows a reliable prenatal prediction of both restrictive and normal FO anatomy postnatally. Prediction of likelihood of urgent BAS also succeeds reliably in all fetuses with restrictive FO, but identification of the small subset of fetuses that also requires urgent BAS despite their normal FO anatomy fails, because the ability of sufficient postnatal interatrial mixing cannot be predicted prenatally. Therefore, all fetuses with prenatally diagnosed d-TGA should always be delivered in a tertiary center with cardiac catheter stand-by, allowing BAS within the first 24 h after birth, regardless of their predicted FO anatomy.
Hinweise

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
What does this study add to the clinical work
Restrictive foramen ovale (FO) in dextro-transposition of the great arteries (d-TGA) with intact ventricular septum may lead to severe life-threatening hypoxia within the first hours of life, making urgent balloon atrial septostomy (BAS) inevitable. Precise assessment of FO size and FO flap motility is possible prenatally and allows a reliable prediction of both restrictive and normal FO anatomy postnatally. In addition, prediction of likelihood of urgent BAS also succeeds reliably in all fetuses with restrictive FO, but identification of the small subset of fetuses that also requires urgent BAS despite their normal FO anatomy fails. Therefore, all fetuses with prenatally diagnosed d-TGA should always be delivered in a tertiary center with cardiac catheter stand-by, allowing BAS within the first 24 hours after birth, regardless of their predicted FO anatomy.

Introduction

The simple dextro-Transposition of the Great Arteries with intact ventricular septum (d-TGA) is one of the most common cyanotic heart defects with a prevalence of 2–3 of 10.000 live births. It accounts for 5–7% of all congenital heart defects [1, 2]. Simple d-TGA is characterized by atrioventricular concordance and ventriculo-arterial discordance, leading to redistribution of blood flow with subsequent development of severe hypoxemia and rapid hemodynamic compromise postnatally. Both, fetal pulmonary circulation and ductus arteriosus become increasingly sensitive to the high oxygen content toward the end of gestation, which may predispose to a restriction of the foramen ovale (FO) and ductus arteriosus at term by increase of left atrial pressure leading to restrictive FO flow [37]. In addition, the increased right ventricular outflow and decreased antegrade flow through the arterial duct lead to a reduced right-to-left shunt across the FO and to a stagnation of FO growth during the 3rd trimester [8, 9]. Consequently, the FO may be significantly restrictive in a significant proportion of newborns, making an urgent balloon atrial septostomy (BAS) immediately after birth inevitable in about 28% of newborns [9, 10]. Indisputably, prenatal diagnosis of critical d-TGA with restrictive FO with planned delivery in a tertiary center with cardiac catheter stand-by, allowing immediate BAS after birth, improves neonatal survival, whereas a significant delay of diagnosis and intervention postnatally may result in severe life-threatening neonatal hypoxia [1113].
Several conventional prenatal echocardiographic parameters were previously evaluated with contradictory sensitivity and specificity in predicting restrictive FO and the likelihood for urgent BAS [6, 7, 9, 1421]. Recently, modern techniques on cardiac functional assessment like spectral pulsed wave tissue-Doppler imaging, speckle-tracking echocardiography, or fetal cardiac magnetic resonance imaging have demonstrated a potentially higher sensitivity for detection of subtle subclinical changes in myocardial dysfunction [7, 2224], but data on functional assessment in fetuses with d-TGA are limited to only one single small cohort of only 13 fetuses without validation in larger prospective trials [7]. Additionally, the calculation of functional cardiac indices is time-consuming and requires highly specified instrumental equipment. Therefore, the use of these technologies is so far limited to a very few specialized centers and research purposes only.
This study aimed to describe our experience with the prenatal prediction of normal or restrictive FO anatomy in fetuses with d-TGA. We first reviewed the reliability of our own former prenatal ultrasound diagnoses made at the time of the patients’ visits in our centers prior to birth. Second, we re-assessed all stored echocardiographic videos and still images and correlated them with the outcome. Thereby, we retrospectively identified the most reliable echocardiographic parameters for restrictive FO. Finally, we described the predictive value of these parameters by calculating their sensitivity and specificity.

Methods

Study design

We conducted a retrospective cohort study in two large German tertiary referral centers (University Hospitals of Cologne and Bonn) over a period of 13 years (2010–2022) and included all cases of isolated fetal d-TGA with intact ventricular septum and known outcome. Requirements for inclusion in the study were the availability of the former prenatal ultrasound reports as well as sufficient prenatal echocardiographic videos and still images for re-evaluation and to establish prognostic markers. We only included cases where last echocardiographic examination was performed not more than 14 days prior to delivery. Exclusion criteria were d-TGA with hemodynamically relevant ventricular septal defects, presence of additional major anomalies, delivery before completed 34 weeks, lack of sufficient prenatal echocardiographic videos, and images for retrospective re-evaluation as well as unknown outcome.
The fetal and maternal variables assessed in every case are shown in Table 1. Prenatal testing for aneuploidies including 22q11 microdeletion was offered to all parents. The assessment of postnatal outcome included gestational age at delivery, mode of delivery, birth weight, Apgar score at 1 min, oxygen saturation, need for urgent BAS within the first 24 h after birth, and day of life at arterial switch operation. All neonatal data and clinical files were retrieved from our Departments of Pediatric Cardiology in Cologne and Bonn.
Table 1
Fetal and maternal characteristics in the urgent and no urgent BAS group
Characteristics
All
Urgent BAS
No urgent BAS
p value
(n = 45)
(n = 26)
(n = 19)
Maternal age (years)
32.6 ± 4.8
32.5 ± 4.7
32.6 ± 5.0
0.934
Gestational age at last echo (weeks)
38 + 6 (34 + 5–40 + 1)
38 + 4 (37 + 1–39 + 1)
38 + 1 (34 + 5–40 + 2)
0.368
Estimated fetal weight at last ultrasound (g)
3133 ± 665
3290 ± 650
3240 ± 677
0.505
Period between last echocardiography—birth
7.9 ± 4.5
7.2 ± 3,4
8.3 ± 4.4
0.544
Gestational age at delivery (weeks)
38 + 6 (35 + 0–40 + 6)
38 + 6 (37 + 6–40 + 1)
38 + 4 (35 + 0–40 + 6)
0.467
Cesarean section
26 (57.8%)
16 (61.5%)
10 (52.6%)
0.550
Birth weight (g)
3400 (range 2500–4100)
3390 (range 2500–4100)
3450 (range 2700–4000)
0.848
Age of life at arterial switch operation (days)
9 (1–30)
8 (1–16)
10 (4–30)
0.600
Retrospective studies were approved by the local ethical committee of human research (No 20-1517).

Predictive reliability of the former prenatal ultrasound reports

For the purpose of this study, ultrasound reports were retrieved from the ViewPoint Software system (GE Healthcare) and compared with the outcome to retrospectively evaluate the predictive reliability of the FO anatomy. At that time, prenatal sonographic assessment as well as prediction of the likelihood of urgent BAS was performed by varying operators and according to subjective criteria.

Establishing reliable diagnostic cardiac parameters

The cardiac parameters, that were retrospectively calculated to establish reliable prognostic parameters, were retrieved from the latest echocardiographic videos and still images that were recorded not more than 2 weeks prior to delivery. This retrospective calculation was always carried out by the same author (I.G.) and blinded to the outcome. Then, the author once again predicted the likelihood for urgent BAS and the specificity and sensitivity of this current assessment was described.
The cardiac parameters that were assessed, are shown in Table 2. We measured both the right and left atrial diameter in the 4-chamber view and calculated the right-to-left atrial diameter ratio as previously described by Firpo et al. [25]. We assessed the sonographic appearance of the septum primum and the FO flap swinging movement and measured the diameter of the FO as well as the maximal angle of the flap at its attachment point to the septum which usually varies between 30 and 50° as described by Wilson et al. [26]. In case of an abnormal mobility of the FO flap, we described the FO flap either as (a) redundant, if the aneurysmal septum primum flap bulged more than 50% across the left atrium wall (‘spinnaker-like’), (b) flat, if the angle between the septum primum flap and the rest of the atrial septum was < 30°, (c) fixed, if the septum primum did not demonstrate any swinging mode during cardiac cycle, or (d) hypermobile, if the flap oscillated bidirectionally through the small FO orifice between the left and right atrium (Fig. 1).
Table 2
Retrospectively calculated cardiac parameters
Cardiac parameters
All
Urgent BAS
No urgent BAS
p value
(n = 45)
(n = 26)
(n = 19)
RA diameter (mm)
22.61 ± 3.46
22.86 ± 3.02
22.11 ± 2.85
0.428
LA diameter (mm)
20.65 ± 2.89
20.86 ± 3.02
20.38 ± 2.78
0.595
RA-to-LA diameter ratio, mean ± SD
1.12 ± 0.12
1.12 ± 0.07
1.11 ± 0.19
0.317
FO diameter (mm)
7.05 ± 2.25
5.71 ± 2.06
8.75 ± 0.95
 < 0.01*
FO restrictive appearance (< 7 mm) (n)
20 (44.4%)
20 (76.9%)
0 (0%)
 < 0.01*
Redundant ‘spinnaker-like’ appearance of the FO flap (n)
4 (8.9%)
3 (11.5%)
1 (5.3%)
0.465
Flat appearance of the FO flap (mobility < 30°) (n)
4 (8.9%)
3 (11.5%)
1 (5.3%)
0.465
Fixed appearance of the FO flap (n)
9 (20.0%)
8 (30.8%)
1 (5.3%)
0.035*
Hypermobile FO flap (n)
7 (15.6%)
7 (26.9%)
0 (0%)
0.014*
‘Plateau-like’ PV flow pattern (n)
24/28 (85.7%)
16/18 (88.9%)
8/10 (80%)
0.520
‘pulsatile’ PV flow pattern (n)
2/28 (7.1%)
2/18 (11.1%)
2/10 (20%)
0.520
To-and-fro PV flow pattern (n)
2/28 (7.1%)
2/18 (11.1%)
0/10 (0%)
0.274
PV maximum systolic flow velocity (cm/s), mean ± SD
34 ± 8
35 + / 9
31 ± 6
0.021*
PV maximum systolic flow velocity ≥ 35 cm/s (n)
15/28 (53.6%)
11/18 (61.1%)
4/10 (40%)
0.062
PV maximum systolic flow velocity ≥ 41 cm/s (n)
4/28 (14.3%)
4/18 (22.2%)
0/10 (0%)
0.107
Restrictive DA (n)
2 (4.4%)
2 (7.7%)
0 (0%)
0.216
Reversed flow in DA (n)
2 (4.4%)
2 (7.7%)
0 (0%)
0.216
The Doppler flow patterns in the pulmonary veins were classified either as (a) continuous forward flow with a small a-wave reversal (‘plateau-like’ flow pattern), (b) continuous forward flow with an increased a-wave reversal (‘pulsatile’ flow pattern) or (c) to-and-fro flow pattern with absent diastolic forward flow. In accordance with Taketazu et al. [27] we classified the ‘plateau-like’ and the ‘pulsatile’ flow patterns as not suspicious and the to-and-fro flow pattern as highly suspicious for restrictive FO (Fig. 2).
The Doppler flow velocity was measured in at least one pulmonary vein at the entrance to the left atrium as described by Slodki et al. [19], and the maximum systolic flow velocity was assessed (Fig. 3). The angle of insonation was preferably 0°, but always less than 30°. As previously described, a maximum systolic flow velocity of ≥ 35 cm/s was classified as increased and ≥ 41 cm/s as highly suspicious for restrictive FO.
The size of the ductus arteriosus (DA) and the direction of flow in the DA were routinely assessed in all examinations in the 3-vessel view by B-mode and Color Doppler, respectively (Fig. 4). A reversed flow in the DA was classified as abnormal. Atrial diameters, FO diameter, FO flap appearance and flap mobility, ductal size, and flow direction in arterial duct could be assessed in all fetuses, whereas the pulmonary vein (PV) flow patterns and the PV systolic maximum flow velocity could retrospectively be calculated in only 28 of our 45 fetuses.

Standardized postnatal management

Postnatal management in both of our participating institutes was standardized. A cardiological team with ultrasound system and full catheterization equipment was on stand-by as soon as the birth was about to begin, so that a BAS could be performed immediately after birth, if necessary. In both institutes, we recommended delivery by planned cesarean section in all cases with prenatally suspected restrictive FO, whereas in cases of a predicted normal FO anatomy, the mode of delivery was chosen by the obstetric condition or by maternal request. Immediately after birth, a focused echocardiography was performed to first ascertain the prenatal diagnosis of d-TGA with intact ventricular septum and second to evaluate the FO anatomy and the interatrial shunt. Initial treatment included administration of prostaglandin E1, intubation, and ventilation on the basis of the clinical status of the neonate. Urgent BAS was performed within the first 24 h of life in all neonates with a severely restrictive FO and persistent neonatal hypoxemia at first assessment with oxygen saturation of ≤ 65%. If oxygen saturation was normal despite reduced FO diameter, BAS was not considered as urgent and was not performed within the first 24 h.

Statistical analysis

Statistical analysis was performed using Student’s t Test for continuous variables and the χ2 or Fisher’s exact test for categorical variables. A p value of < 0.05 was considered statistically significant, and all p-values were based upon two-tailed tests. Statistical analysis was performed using SPSS for Mac (SPSS Inc., Chicago, Ill., USA).

Results

During the study period, 87 fetuses with isolated d-TGA with intact ventricular septum and known outcome were diagnosed and delivered in one of our two participating centers. 42 cases had to be excluded from the study. Reasons for exclusion were either preterm delivery before the completed 34th week or lack of suitable prenatal echocardiographic videos or still images for reliable re-analysis. Finally, 45 cases were included in this series. Fetal or neonatal genetic testing, including karyotype and 22q11 microdeletion, was normal in all cases. There was no termination of pregnancy and no intrauterine or neonatal death in our cohort.
Cesarean section was performed in 26 (57.7%) of 45 cases. Reasons for surgical delivery were either prenatal prediction of restrictive FO in 15 cases, obstetrical reasons in seven cases and maternal request in four cases. Median gestational age at delivery was 39 weeks (range, 36–41 weeks).
Among the 45 fetuses, 22 neonates had restricted FO at birth and underwent urgent BAS within the first 24 h of life, whereas 23 neonates had normal FO size at birth. Among the latter, four neonates unexpectedly showed inadequate interatrial mixing despite a normal FO anatomy. They did not achieve sufficient oxygen saturation, rapidly developed hypoxia, and also required urgent BAS (‘bad mixer’). Finally, 26 (57.8%) neonates required urgent BAS within the first 24 h of life (urgent BAS group), whereas 19 (42.2%) neonates achieved good oxygen saturation and did not underwent urgent BAS (non-urgent BAS group) (Fig. 5).
Among the 4 “bad mixers”, all had normal FO appearance with normal FO diameters between 8 and 9 mm, all had a normal FO flap mobility of 30–50° and none of them had a restrictive DA prenatally. Additionally, in 3 of those 4 cases, the pulmonary veins were also assessed prenatally and also showed a normal, plateau-like flow pattern with peak velocities between 20 and 30 cm/s (Table 3, patients no 23–26). All those prenatal findings were confirmed on first postnatal echocardiography that was performed immediately after birth.
Table 3
Prediction of FO anatomy and likelihood of urgent BAS in the former ultrasound reports and after retrospective re-analysis of 26 fetuses that required urgent BAS
Patient no
Former prenatal ultrasound report
Retrospective re-analysis of FO assessment
Retrospective re-analysis of FO diameter (mm)
Retrospective re-analysis of FO flap movement
Retrospective re-analysis of PV flow pattern
Retrospective re-analysis of PV peak velocity(cm/s)
Retrospective re-analysis of DA flow direction
Delivery Mode
Birth weight (kg)
Postnatal findings
Urgent BAS
1
Normal
Restrictive
10
Flat
Plateau
43
Antegr
Cesarean
2500
fixed flap
Yes
2
Normal
Restrictive
5
Fixed
To-and-fro
37
Antegr
Cesarean
3100
FO < 7 mm, fixed flap
Yes
3
Normal
Restrictive
5
Hypermobile
Plateau
38
Antegr
Cesarean
3400
FO < 7 mm, hypermobile flap
Yes
4
Restrictive
Restrictive
6
Fixed
Plateau
30
Antegr
Cesarean
3600
FO < 7 mm, fixed flap
Yes
5
Normal
Restrictive
3
Hypermobile
Plateau
20
Antegr
Vaginal
3800
FO < 7 mm, hypermobile flap
Yes
6
Restrictive
Restrictive
4
Fixed
To-and-fro
42
Antegr
Cesarean
3000
FO < 7 mm, fixed flap
Yes
7
Restrictive
Restrictive
10
Redundant
Plateau
43
Antegr
Cesarean
4000
Fixed flap
Yes
8
Normal
Restrictive
6
Hypermobile
Plateau
38
Antegr
Vaginal
3800
FO < 7 mm
Yes
9
Normal
Restrictive
5
Redundant
Plateau
29
Reversed
Vaginal
3600
FO < 7 mm, hypermobile flap
Yes
10
Restrictive
Restrictive
4
Flat
Plateau
35
Reversed
Cesarean
3100
FO < 7 mm, fixed flap
Yes
11
Restrictive
Restrictive
4
Fixed
Plateau
35
Antegr
Cesarean
3200
FO < 7 mm, fixed flap
Yes
12
Normal
Restrictive
4
Redundant
Plateau
58
Antegr
Cesarean
2700
FO < 7 mm
Yes
13
Restrictive
Restrictive
4
Hypermobile
Plateau
35
Antegr
Cesarean
3900
FO < 7 mm, hypermobile flap
Yes
14
Restrictive
Restrictive
3
Hypermobile
Plateau
40
Antegr
Cesarean
3700
FO < 7 mm, hypermobile flap
Yes
15
Restrictive
Restrictive
4
Flat
Plateau
28
Antegr
Cesarean
4000
FO < 7 mm, flat flap
Yes
16
Normal
Restrictive
5
Normal
  
Antegr
Vaginal
3400
FO < 7 mm
Yes
17
Normal
Restrictive
5
Fixed
  
Antegr
Vaginal
3500
FO < 7 mm, fixed flap
Yes
18
Normal
Restrictive
5
Fixed
  
Antegr
Cesarean
3000
FO < 7 mm, fixed flap
Yes
19
Restrictive
Restrictive
3
Hypermobile
  
Antegr
Cesarean
3700
FO < 7 mm, hypermobile flap
Yes
20
Normal
Restrictive
5
Hypermobile
  
Antegr
Vaginal
3300
FO < 7 mm
Yes
21
Restrictive
Restrictive
5
Fixed
  
Antegr
Cesarean
4100
FO < 7 mm, fixed flap
Yes
22
Restrictive
Restrictive
6
Fixed
  
Antegr
Cesarean
3100
FO < 7 mm, fixed flap
Yes
23
Normal
Normal
8
Normal
Plateau
30
Antegr
Vaginal
3100
FO > 7 mm, normal flap motility, but bad mixer
Yes
24
Normal
Normal
8
Normal
Plateau
30
Antegr
Vaginal
2900
FO > 7 mm, normal flap motility, but bad mixer
Yes
25
Normal
Normal
9
Normal
Plateau
20
Antegr
Vaginal
3600
FO > 7 mm, normal flap motility, but bad mixer
Yes
26
Normal
Normal
8
Normal
  
Antegr
Vaginal
3000
FO > 7 mm, normal flap motility, but bad mixer
Yes

Urgent BAS group

Despite initial prostaglandin infusion and mechanical ventilation, 26 (57.8%) neonates developed severe hypoxemia with oxygen saturation of less than 65% and received urgent BAS within the first 24 h. Among them, two neonates also needed subsequent arterial switch operation within the following 24 h after BAS, because of still inadequate oxygen saturation after BAS. The remaining 24 neonates underwent elective arterial switch operation at an average age of 8 days (range, 4–16 days).

Non-urgent BAS group

19 (42.2%) neonates had good oxygen saturation initially and did not require urgent BAS, but four of them received elective BAS at the 2nd and 3rd day of life, respectively. Elective arterial switch operation in this group was performed at an average age of 10 days (range, 5–30 days).
The postoperative course in the cardiac intensive care unit was uncomplicated for all but three children. One child developed laryngomalacia, a second one had pulmonary valve stenosis that had to be widened at the age of 2 months, and a third one developed persistent pulmonary hypertension. Median follow-up period in our cohort was 41 months (range, 6 weeks–7 years).
There were no statistically significant differences between the urgent and non-urgent BAS group with regard to maternal age, gestational age at delivery, prevalence of cesarean section, birth weight, Apgar score at the 1st minute of life, and age of life at arterial switch operation (Table 1).

Predictive reliability of the former prenatal ultrasound reports

In the former prenatal ultrasound reports, 15 fetuses were predicted to have a restrictive FO postnatally and to require urgent BAS. Subsequently, they were delivered by cesarean section. Postnatally, only 11 of them had restrictive FO (73.3% true-positive diagnoses), whereas four had normal FO anatomy and achieved good oxygen saturation contrary to their prenatal prediction and did not require urgent BAS (26.7% false-positive diagnoses). In contrast, 30 fetuses were predicted to have a normal FO anatomy which was confirmed postnatally in only 19 neonates (63.3% true-negative diagnoses), whereas 11 neonates had a restrictive FO contrary to the prenatal prediction and also required urgent BAS (36.7% false-negative diagnoses) (Tables 3 and 4).
Table 4
Prediction of FO anatomy and of BAS in former ultrasound reports and after retrospective re-analysis of 19 fetuses that did not underwent urgent BAS
Patient no
Former prenatal ultrasound report
Retrospective re-analysis of FO assessment
Retrospective re-analysis of FO diameter (mm)
Retrospective re-analysis of FO flap movement
Retrospective re-analysis of PV flow pattern
Retrospective re-analysis of PV peak velocity(cm/s)
Retrospective re-analysis of DA flow direction
Delivery Mode
Birth weight (kg)
Postnatal findings
Urgent BAS
1
Restrictive
Normal
8
Normal
Pulsatile
25
Antegr
CESAREAN
2700
FO > 7 mm, normal flap motility
No
2
Restrictive
Normal
10
Normal
Plateau
38
Antegr
Cesarean
4000
FO > 7 mm, normal flap motility
No
3
Normal
Normal
8
Normal
Pulsatile
39
Antegr
Vaginal
3300
FO > 7 mm, normal flap motility
No
4
Normal
Normal
10
Normal
Plateau
25
Antegr
Vaginal
3500
FO > 7 mm, normal flap motility
No
5
Normal
Normal
8
Normal
Plateau
25
Antegr
Vaginal
3900
FO > 7 mm, normal flap motility
No
6
Normal
Normal
9
Normal
Plateau
35
Antegr
Vaginal
3800
FO > 7 mm, normal flap motility
No
7
Normal
Normal
8
Normal
Plateau
30
Antegr
Vaginal
3700
FO > 7 mm, normal flap motility
No
8
Normal
Normal
10
Normal
Plateau
40
Antegr
Cesarean
3500
FO > 7 mm, normal flap motility
No
9
Normal
Normal
10
Normal
Plateau
25
Antegr
Cesarean
3400
FO > 7 mm, normal flap motility
No
10
Normal
Normal
8
Normal
Plateau
30
Antegr
Cesarean
3600
FO > 7 mm, normal flap motility
No
11
Restrictive
Normal
10
Fixed
Antegr
Cesarean
3000
FO > 7 mm, normal flap motility
No
12
Restrictive
Normal
8
Flat
Antegr
Cesarean
3600
FO > 7 mm, normal flap motility
No
13
Normal
Normal
8
Normal
Antegr
Vaginal
3200
FO > 7 mm, normal flap motility
No
14
Normal
Normal
8
Normal
Antegr
Cesarean
3400
FO > 7 mm, normal flap motility
No
15
Normal
Normal
8
Normal
Antegr
Vaginal
3300
FO > 7 mm, normal flap motility
No
16
Normal
Normal
8
Normal
Antegr
CESAREAN
3000
FO > 7 mm, normal flap motility
No
17
Normal
Normal
9
Normal
Antegr
CESAREAN
3700
FO > 7 mm, normal flap motility
No
18
Normal
Normal
9
Normal
Antegr
Vaginal
3400
FO > 7 mm, normal flap motility
No
19
Normal
Normal
10
Normal
Antegr
Vaginal
3500
FO > 7 mm, normal flap motility
No
Overall, restrictive FO was correctly predicted in only 11 of 22 cases in the former prenatal ultrasound reports (sensitivity of 50%) whereas a normal FO size was correctly predicted in 19 of 23 cases (specificity of 83%).

Establishing reliable diagnostic cardiac parameters

The cardiac parameters, that were retrospectively re-calculated, are shown in Table 2. Overall, the right and left atrial diameters, the prevalence of a redundant (‘spinnaker-like’) appearance of the FO flap, the flow patterns in the PV, the diameter of the DA, and the flow direction through the DA did not differ significantly between both groups.
Significant differences were observed in the FO diameter, the FO flap mobility, and the PV maximum systolic flow velocity: The FO diameter was significantly smaller in the urgent BAS group with a median of 5.7 mm compared with 8.7 mm in the non-urgent BAS group, and a FO diameter of less than 7 mm was highly predictive for restrictive FO (p < 0.01). The FO flap mobility was significantly altered with either a fixed appearance in 30.8% of cases of the urgent BAS group compared with 5.3% in the non-urgent BAS group (p = 0.035), or a hypermobile flap movement in 26.9% of cases in the urgent BAS group. Because a hypermobile flap was exclusively seen in the urgent BAS group, it was highly predictive for a restrictive FO (p = 0.014).
The maximum systolic flow velocities in the PV were also significantly increased in the urgent BAS group with 35 cm/s compared with 31 cm/s in the non-urgent BAS group (p = 0.021), but no cut-off value to reliably distinguish between a restrictive and normal FO could be identified.
Thus, the three most significant markers to predict a restrictive FO were the FO diameter of less than 7 mm (p < 0.01) and a fixed (p = 0.035) or hypermobile FO flap (p = 0.014).

Reliability to correctly predict FO anatomy

After retrospective re-analysis of the FO size and the FO flap mobility, all 22 cases with restrictive FO as well as all 23 cases with normal FO anatomy could correctly be predicted. Both the positive predictive value (PPV) to predict a restrictive FO and the negative predictive value (NPV) to predict normal FO anatomy were 100% each (with a sensitivity and specificity of 100%).

Reliability to predict the likelihood of urgent BAS in restrictive FO

After retrospective re-assessment, the author correctly predicted the need for urgent BAS in all 22 fetuses with restrictive FO (100% true-positive diagnoses) (Table 3). Among these 22 fetuses, 20 (90.9%) had a FO diameter of less than 7 mm, 8 (36.4%) had a fixed and 7 (31.8%) had a hypermobile FO flap. The maximum systolic flow velocity in the PV was also significantly increased (p = 0.021) with a median of 35 cm/s, including four fetuses with a maximum flow velocity of more than 41 cm/s. Overall, the PPV to predict the likelihood of urgent BAS in fetuses with a correctly predicted restrictive FO was 100%.

Reliability to predict the likelihood of urgent BAS in normal FO anatomy

In addition, the author also correctly predicted the remaining 23 fetuses to have an adequate FO size. As presumed, 19 of the 23 neonates achieved good oxygen saturation and did not require urgent BAS (Table 4). All of them had an adequate FO diameter of more than 7 mm and normal FO flap swinging movements of 30–50°. But the remaining four neonates unexpectedly developed early cyanosis (‘bad mixers’) and also required urgent BAS, although all of them had postnatally confirmed adequate FO diameters of more than 7 mm and normal FO flap swinging movements. In three of these four neonates, the pulmonary veins were also assessed prenatally and also showed a normal plateau-like flow pattern and normal systolic maximum flow velocities of not more than 30 cm/s (Table 3). As this situation cannot be predicted prenatally, the negative predictive value (NPV) for the likelihood of urgent BAS in fetuses with normal FO anatomy was only 82.6% with a false-negative rate (FNR) of 17.8%.

Discussion

Indisputably, prenatal prediction of a restrictive FO with planned delivery in a tertiary cardiac center and consecutively cardiac intervention within 24 h after birth improves survival, whereas a significant delay of diagnosis and intervention may result in severe and life-threatening neonatal hypoxia. Several prenatal cardiac parameters and classification systems have previously been evaluated to predict premature restriction of the FO and the likelihood of urgent BAS [6, 7, 9, 1421]. Table 5 shows an overview of the previously published series and the included cardiac parameters.
Table 5
Significance of cardiac parameters in fetuses with TGA in previously published studies
 
Maeno et al. [14]
Jouannic et al. [15]
Punn et Silverman [16]
Vigneswaran et al. [6]
Tuo et al. [18]
Slodki et al. [19]
Gatta et al. [20]
Patey et al. [7]
Lachaut et al. [9]
Bucca et al. [21]
Gottschalk et al
1999
2004
2011
2017
2017
2018
2021
2021
2021
2022
2023
n = 16
n = 130
n = 26
n = 40
n = 40
n = 51
n = 60
n = 13
n = 59
n = 292
n = 45
Restrictive
s
s
s
s
n.s
s
s
s
s
s
FO appearance
Redundant
n.s
n.s
n.s
s
s
s
s
n.s
n.s
FO septum
Flat
n.s
n.s
n.s
s
s
n.s
n.s
n.s
n.s
FO septum
Fixed
n.s
n.s
n.s
s
n.s
n.s
n.s
s
FO septum
Hypermobile FO septum
n.s
s
n.s
n.s
-
n.s
n.s
-
s
s
Increased
s
s
PV maximum systolic flow
Abnormal PV flow pattern
n.s
n.s
Restrictive
s
s
n.s
n.s
n.s
n.s
n.s
s
n.s
n.s
DA appearance
Reversed
s
n.s
n.s
n.s
s
n.s
n.s
DA flow
Reduced
s
n.s
s
n.s
RA diameter
Mean ratio FO/AV diameter
s
s
Mean ratio FO/septal length
s
s
Prevalence of urgent BAS
23% BAS
54% BAS
30% BAS
50% BAS
57% BAS
45% BAS
54% BAS
58% BAS
s significant, n.s. not significant, – not evaluated
These series showed contradictory reliabilities of the different cardiac parameters in predicting the postnatal FO anatomy and the likelihood of urgent BAS. These difficulties in predicting postnatal hypoxia may be explained by several factors. First, fetal hemodynamic parameters in late pregnancy are influenced by several factors other than those related to FO diameter, such as breathing movements, greater sensitivity to oxygen, and physiological right ventricular preponderance [15]. In addition, all published data relied on retrospective studies and included different cardiac parameters that were obtained at different gestational ages with highly variable time intervals between last prenatal echocardiography and birth. Moreover, postnatal oxygen saturation cut-off values differed, and the indication to perform an urgent BAS was not standardized in previously published reports, making those data only to a limited extend comparable.
At first, Maeno et al. described 16 fetuses with isolated d-TGA in 1999 [14]. They assessed the appearance of the FO and the FO flap as well as the DA diameter, but last sonographic assessment prior to delivery was performed at a mean gestational age of 31 weeks without performing an additional late echocardiography. They concluded that the sensitivity to predict a restrictive FO is low. But in a significant subset of fetuses, FO restriction develops not until the late third trimester and last echocardiography at 31 weeks of gestation may be too early to detect significant restriction of the FO in these cases. As the critical period is near term, we strongly recommend to perform an additional late echocardiography after 35 weeks of gestation in all fetuses with d-TGA to once again re-assess the FO anatomy and to re-evaluate the likelihood of postnatal urgent BAS.
Since then, most recent series including a large meta-analysis by Buca et al. described a significantly smaller FO diameter in fetuses that required urgent BAS [6, 9, 15, 18, 20, 21], but a FO diameter cut-off value to reliably predict a restrictive FO was only described in one single series [20]. In this series, 60 fetuses with isolated d-TGA were retrospectively assessed and the authors concluded that a FO diameter of more than 6.5 mm predicts a normal FO with 100% sensitivity and a false-positive rate of 45%. We could confirm, that a FO diameter of less than 7 mm is highly predictive for a restrictive FO and need for urgent BAS (p < 0.01).
Inconsistent findings were published with regard to the septum primum appearance and the FO flap movement. Some authors stated that a redundant, ‘spinnaker-like’ [9, 18, 20, 21] or flat [6, 19] FO flap significantly predicts a restrictive FO, whereas other authors could not confirm this finding [1416]. Vigneswaran et al. postulated that a redundant FO may be explained as a variant of the normal third trimester atrial septum deviation rather than as a pathological marker [6]. In contrast, a redundant FO flap may cause an occlusion of the FO orifice in some cases as it may mimic primary restriction of FO in cases of an aneurysmal septum primum. Some authors described a hypermobile FO flap as a significant predictor for restrictive FO [16, 21], whereas others could not support this finding and assumed that this may be related to simulated breathing movements or hiccoughs rather than to a restrictive FO [6]. In our series, both a fixed and hypermobile flap mobility were the most significant predictors for a restrictive FO and the need for an urgent BAS (p = 0.035 and p = 0.014, respectively).
Only one previously published study assessed the flow velocity in the pulmonary veins (PV) and described a significantly increased maximum systolic flow in the PV in restrictive FO. They stated, that a cut-off value of 41 cm/s maximum flow velocity provided maximum specificity of 100% with a positive predictive value of 100% [19]. In accordance, we could confirm that the maximum flow velocity in the PV is increased in restricted FO, but this difference reached no statistical significance (p = 0.216) in our data and no cut-off value to reliably predict the restrictive FO could be identified. In addition, the flow pattern in the PV was not significantly altered in fetuses with restrictive FO. Therefore, assessment of PV flow did not improve the reliability of prediction.
Although we were able to correctly predict the postnatal FO anatomy with a sensitivity and specificity of 100% in our re-analysis, the neonates’ ability to achieve a sufficient O2 saturation remains unpredictable. Among our 23 neonates with a normal FO size and normal FO flap mobility, 17% were ‘bad mixer’, who postnatally did not achieve a sufficient O2 saturation and unexpectedly required urgent BAS also. As this situation escapes prenatal prediction, a normal FO anatomy consequently does not ensure an adequate interatrial shunting in all cases and the negative predictive value to reliably exclude an urgent BAS can naturally never reach 100%. This finding is in accordance with the study by Tuo et al. who described 10 of 40 fetuses that unexpectedly required urgent BAS (false-negative diagnoses in 25%) and who stated that “we should never trust fetuses with d-TGA, even when they seem to have a foramen ovale of normal appearance” [18]. Therefore, all fetuses with prenatally diagnosed d-TGA should always be delivered in a tertiary center with cardiac catheter stand-by allowing BAS within the first 24 h after birth, regardless of their predicted FO anatomy.
The prevalence of urgent BAS within the first 24 h in our cohort was 58%. In previously published prenatal series, the prevalence varied between 23% [15] and 57% [19]. These data are also comparable only to a certain extent because the indication to perform a BAS within the first 24 h was not standardized in different centers and depended on multiple factors including subjective assessment of first neonatal clinical presentation, variable institutional thresholds of oxygen saturation for urgent BAS, variable dosages of applied prostaglandin to prevent ductal closure, or the availability of a permanent cardiac catheter stand-by that allows urgent BAS at any time.
Besides the conventional echocardiographic assessment, very few recent studies described modern techniques on cardiac functional assessment to improve diagnostic accuracy and to optimize perinatal management [7, 24]. Patey and colleagues proposed a novel fetal index with cardiac parameters obtained by spectral pulsed wave tissue-Doppler imaging (PW-TDI) and speckle-tracking echocardiography (STE) and demonstrated significant cardiac geometrical and functional alterations in fetuses with d-TGA that required urgent BAS compared with fetuses that did not. They described the left ventricle rotation-to-shortening ratio, a ratio of apical systolic rotation to basal circumferential strain, as a marker of sub-endocardial dysfunction with the highest sensitivity for prediction of urgent BAS. However, this marker was only evaluated in a very small number of 13 pregnancies and has to be validated in larger prospective trials yet. So far it remains unclear, whether the use of PW-TDI or STE may also improve the prediction of the likelihood for urgent BAS in the subgroup of ‘bad mixer’ fetuses with normal FO size that escapes conventional echocardiography. In addition, calculation of the PW-TDI or STE parameters is more time-consuming, and the specificity of 83% seems to be no higher than the sensitivity of the previously published conventional echocardiographic parameters.
Other few series described the clinical utility of fetal cardiac magnetic resonance imaging (cardiac MRI) as a complementary tool to conventional echocardiography [23, 24, 28, 29]. Although MRI may improve fetal cardiovascular imaging, data on the use of cardiac MRI in prenatal diagnosis of cardiac defects including fetal d-TGA are still missing. Future studies are needed to provide more information on the reliability of fetal cardio MRI in the diagnosis of cardiac defects. Until then, conventional fetal echocardiography remains the ‘gold standard’ in fetal cardiology.

Limitations of our study

The retrospective nature of our study is the main limitation, and the retrospective offline re-analysis of cardiac parameters may be operator-dependent. Larger prospective series with longitudinal follow-up and standardized postnatal treatment and standardized indications for urgent BAS are needed. The institutional threshold of 65% for pre-ductal oxygen saturation that was selected in our cohort to perform urgent BAS may be questionable, but it is commonly used in clinical practice and in most of the previously published prenatal series.

Conclusion

In fetuses with d-TGA, conventional fetal echocardiography with assessment of the FO size and FO flap mobility allows a reliable prenatal prediction of both, the restrictive and the normal FO, if fetal echocardiography is performed late in the third trimester within the last 2 weeks before birth. Prediction of the likelihood of urgent BAS also succeeds reliably in all fetuses with restrictive FO, but identification of the small subset of fetuses with normal FO anatomy that also requires urgent BAS (‘bad mixer’) fails, because the ability of sufficient postnatal interatrial mixing cannot be predicted prenatally. Therefore, all fetuses with prenatally diagnosed d-TGA should always be delivered in a tertiary center with cardiac catheter stand-by allowing BAS within the first 24 h after birth, regardless of their predicted FO anatomy.

Declarations

Conflict of interest

The authors have not disclosed any competing interests.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Jetzt e.Med zum Sonderpreis bestellen!

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

Jetzt bestellen und 100 € sparen!

e.Med Gynäkologie

Kombi-Abonnement

Mit e.Med Gynäkologie erhalten Sie Zugang zu CME-Fortbildungen der beiden Fachgebiete, den Premium-Inhalten der Fachzeitschriften, inklusive einer gedruckten gynäkologischen oder urologischen Zeitschrift Ihrer Wahl.

Literatur
1.
Zurück zum Zitat Hoffman JI, Kaplan S (2002) The incidence of congenital heart disease. J Am Coll Cardiol 39:1890–1900CrossRefPubMed Hoffman JI, Kaplan S (2002) The incidence of congenital heart disease. J Am Coll Cardiol 39:1890–1900CrossRefPubMed
2.
Zurück zum Zitat Garne E, Loane M, Dolk H, De Vigan C, Scarano G, Tucker D, Soll C, Gener B, Pierini A, Nelen V, Rösch C, Gillerot Y, Feijoo M, Tincheva R, Queisser-Luft A, Addor MC, Mosquera C, Gall M, Barisic I (2005) Prenatal diagnosis of severe structural congenital malformations in Europe. Ultrasound Obstet Gynecol 25:6–11CrossRefPubMed Garne E, Loane M, Dolk H, De Vigan C, Scarano G, Tucker D, Soll C, Gener B, Pierini A, Nelen V, Rösch C, Gillerot Y, Feijoo M, Tincheva R, Queisser-Luft A, Addor MC, Mosquera C, Gall M, Barisic I (2005) Prenatal diagnosis of severe structural congenital malformations in Europe. Ultrasound Obstet Gynecol 25:6–11CrossRefPubMed
3.
Zurück zum Zitat Rudolph AM (2007) Aortopulmonary transposition in the fetus: speculation on pathophysiology and therapy. Pediatr Res 61:375–380CrossRefPubMed Rudolph AM (2007) Aortopulmonary transposition in the fetus: speculation on pathophysiology and therapy. Pediatr Res 61:375–380CrossRefPubMed
4.
Zurück zum Zitat Rudolph AM (2010) Congenital cardiovascular malformations and the fetal circulation. Arch Dis Child Fetal Neonatal Ed 95:F132–F136CrossRefPubMed Rudolph AM (2010) Congenital cardiovascular malformations and the fetal circulation. Arch Dis Child Fetal Neonatal Ed 95:F132–F136CrossRefPubMed
5.
Zurück zum Zitat Porayette P, van Amerom JFP, Yoo SJ, Jaeggi E, MacGowan CK, Seed M (2015) MRI shows limited mixing between systemic and pulmonary circulation in foetal transposition of the great arteries: a potential cause of in utero pulmonary vascular disease. Cardio Young 25:737–744CrossRef Porayette P, van Amerom JFP, Yoo SJ, Jaeggi E, MacGowan CK, Seed M (2015) MRI shows limited mixing between systemic and pulmonary circulation in foetal transposition of the great arteries: a potential cause of in utero pulmonary vascular disease. Cardio Young 25:737–744CrossRef
6.
Zurück zum Zitat Vigneswaran TV, Zidere V, Miller OI, Simpson JM, Sharland GK (2017) Usefulness of the prenatal echocardiogram in fetuses with isolated transposition of the great arteries to predict the need for balloon atrial septostomy. Am J Cardiol 119:1463–1467CrossRefPubMed Vigneswaran TV, Zidere V, Miller OI, Simpson JM, Sharland GK (2017) Usefulness of the prenatal echocardiogram in fetuses with isolated transposition of the great arteries to predict the need for balloon atrial septostomy. Am J Cardiol 119:1463–1467CrossRefPubMed
7.
Zurück zum Zitat Patey O, Carvalho JS, Thilaganathan B (2021) Urgent neonatal balloon atrial septostomy in simple transposition of the great arteries: predictive value of fetal cardiac parameters. Ultrasound Obstet Gynecol 57:756–768CrossRefPubMed Patey O, Carvalho JS, Thilaganathan B (2021) Urgent neonatal balloon atrial septostomy in simple transposition of the great arteries: predictive value of fetal cardiac parameters. Ultrasound Obstet Gynecol 57:756–768CrossRefPubMed
8.
Zurück zum Zitat Godfrey ME, Friedman KG, Drogosz M, Rudolph AM, Tworetzky W (2017) Cardiac output and blood flow redistribution in fetuses with D-loop transposition of the great arteries and intact ventricular septum: insights into pathophysiology. Ultrasound Obstet Gynecol 50:612–617CrossRefPubMed Godfrey ME, Friedman KG, Drogosz M, Rudolph AM, Tworetzky W (2017) Cardiac output and blood flow redistribution in fetuses with D-loop transposition of the great arteries and intact ventricular septum: insights into pathophysiology. Ultrasound Obstet Gynecol 50:612–617CrossRefPubMed
9.
Zurück zum Zitat Lachaud M, Dionne A, Brassard M, Charron MA, Birca A, Dehaes M, Raboisson MJ (2021) Cardiac hemodynamics in fetuses with transposition of the great arteries and intact ventricular septum from diagnosis to end of pregnancy: longitudinal follow-up. Ultrasound Obstet Gynecol 57:273–281CrossRefPubMed Lachaud M, Dionne A, Brassard M, Charron MA, Birca A, Dehaes M, Raboisson MJ (2021) Cardiac hemodynamics in fetuses with transposition of the great arteries and intact ventricular septum from diagnosis to end of pregnancy: longitudinal follow-up. Ultrasound Obstet Gynecol 57:273–281CrossRefPubMed
10.
Zurück zum Zitat Hamzah M, Othman HF, Peluso AM, Sammour I, Aly H (2020) Prevalence and outcomes of balloon atrial septostomy in neonates with transposition of great arteries. Pediatr Crit Care Med 21:324–331CrossRefPubMed Hamzah M, Othman HF, Peluso AM, Sammour I, Aly H (2020) Prevalence and outcomes of balloon atrial septostomy in neonates with transposition of great arteries. Pediatr Crit Care Med 21:324–331CrossRefPubMed
11.
Zurück zum Zitat Bartlett JM, Wypij D, Bellinger DC, Rappaport LA, Heffner LJ, Jonas RA, Newburger JW (2004) Effect of prenatal diagnosis on outcome in D-transposition of the great arteries. Pediatrics 113:e335–e340CrossRefPubMed Bartlett JM, Wypij D, Bellinger DC, Rappaport LA, Heffner LJ, Jonas RA, Newburger JW (2004) Effect of prenatal diagnosis on outcome in D-transposition of the great arteries. Pediatrics 113:e335–e340CrossRefPubMed
12.
Zurück zum Zitat Fuchs IB, Muller H, Abdul-Khaliq H, Harder T, Dudenhausen JW, Henrich W (2007) Immediate and long-term outcomes in children with prenatal diagnosis of selected isolated congenital heart defects. Ultrasound Obstet Gynecol 29:38–43CrossRefPubMed Fuchs IB, Muller H, Abdul-Khaliq H, Harder T, Dudenhausen JW, Henrich W (2007) Immediate and long-term outcomes in children with prenatal diagnosis of selected isolated congenital heart defects. Ultrasound Obstet Gynecol 29:38–43CrossRefPubMed
13.
Zurück zum Zitat Calderon J, Angeard N, Moutier S, Plumet MH, Jambaque I, Bonnet D (2012) Impact of prenatal diagnosis on neurocognitive outcomes in children with transposition of the great arteries. J Pediatr 161:94–98CrossRefPubMed Calderon J, Angeard N, Moutier S, Plumet MH, Jambaque I, Bonnet D (2012) Impact of prenatal diagnosis on neurocognitive outcomes in children with transposition of the great arteries. J Pediatr 161:94–98CrossRefPubMed
14.
Zurück zum Zitat Maeno YV, Kamenir SA, Sinclair B, van der Velde ME, Smallhorn JF, Hornberger LK (1999) Prenatal features of ductus arteriosus constriction and restrictive foramen ovale in d-transposition of the great arteries. Circulation 99:1209–1214CrossRefPubMed Maeno YV, Kamenir SA, Sinclair B, van der Velde ME, Smallhorn JF, Hornberger LK (1999) Prenatal features of ductus arteriosus constriction and restrictive foramen ovale in d-transposition of the great arteries. Circulation 99:1209–1214CrossRefPubMed
15.
Zurück zum Zitat Jouannic JM, Gavard L, Fermont L, Le Bidois J, Parat S, Vouhe PR, Dumez Y, Sidi D, Bonnet D (2004) Sensitivity and specificity of prenatal features of physiological shunts to predict neonatal clinical status in transposition of the great arteries. Circulation 110:1743–1746CrossRefPubMed Jouannic JM, Gavard L, Fermont L, Le Bidois J, Parat S, Vouhe PR, Dumez Y, Sidi D, Bonnet D (2004) Sensitivity and specificity of prenatal features of physiological shunts to predict neonatal clinical status in transposition of the great arteries. Circulation 110:1743–1746CrossRefPubMed
16.
Zurück zum Zitat Punn R, Silverman NH (2011) Fetal predictors of urgent balloon atrial septostomy in neonates with complete transposition. J Am Soc Echocardiogr 24:425–430CrossRefPubMed Punn R, Silverman NH (2011) Fetal predictors of urgent balloon atrial septostomy in neonates with complete transposition. J Am Soc Echocardiogr 24:425–430CrossRefPubMed
17.
Zurück zum Zitat Donofrio MT, Skurow-Todd K, Berger JT, McCarter R, Fulgium A, Krishnan A, Sable CA (2015) Risk-stratified postnatal care of newborns with congenital heart disease determined by fetal echocardiography. J Am Soc Echocardiogr 28:1339–1349CrossRefPubMed Donofrio MT, Skurow-Todd K, Berger JT, McCarter R, Fulgium A, Krishnan A, Sable CA (2015) Risk-stratified postnatal care of newborns with congenital heart disease determined by fetal echocardiography. J Am Soc Echocardiogr 28:1339–1349CrossRefPubMed
18.
Zurück zum Zitat Tuo G, Paladini D, Montobbio G, Volpe P, Cheli M, Calevo MG, Marasini M (2017) Prenatal echocardiographic assessment of foramen ovale appearance in fetuses with d-transposition of the great arteries and impact on neonatal outcome. Fetal Diagn Ther 2:48–56CrossRef Tuo G, Paladini D, Montobbio G, Volpe P, Cheli M, Calevo MG, Marasini M (2017) Prenatal echocardiographic assessment of foramen ovale appearance in fetuses with d-transposition of the great arteries and impact on neonatal outcome. Fetal Diagn Ther 2:48–56CrossRef
19.
Zurück zum Zitat Slodki M, Axt-Fliedner R, Zych-Krekora K, Wolter A, Kawecki A, Enzensberger C, Gulczynska E, Respondek-Liberska M (2018) New method to predict need for Rashkind procedure in fetuses with dextro-transposition of the great arteries. Ultrasound Obstet Gynecol 15:531–536CrossRef Slodki M, Axt-Fliedner R, Zych-Krekora K, Wolter A, Kawecki A, Enzensberger C, Gulczynska E, Respondek-Liberska M (2018) New method to predict need for Rashkind procedure in fetuses with dextro-transposition of the great arteries. Ultrasound Obstet Gynecol 15:531–536CrossRef
20.
21.
Zurück zum Zitat Buca D, Winberg P, Rizzo G, Khalil A, Liberati M, Makatsariya A, Greco F, Nappi L, Acharya G, D’Antonio (2022) Prenatal risk factors for urgent atrial septostomy at birth in fetuses with transposition of the great arteries: a systematic review and meta-analysis. J Matern Fetal Neonatal Med 3:598–606CrossRef Buca D, Winberg P, Rizzo G, Khalil A, Liberati M, Makatsariya A, Greco F, Nappi L, Acharya G, D’Antonio (2022) Prenatal risk factors for urgent atrial septostomy at birth in fetuses with transposition of the great arteries: a systematic review and meta-analysis. J Matern Fetal Neonatal Med 3:598–606CrossRef
22.
Zurück zum Zitat Crispi F, Bijnens B, Sepulveda-Swatson E, Cruz-Lemini M, Rojas-Benavente J, Gonzalez-Tendero A, Garcia-Posada R, Rodriguez-Lopez M, Demicheva E, Sitges M, Gratacos E (2014) Postsystolic shortening by myocardial deformation imaging as a sign of cardiac adaptation to pressure overload in fetal growth restriction. Circ Cardiovasc Imaging 7:781–787CrossRefPubMed Crispi F, Bijnens B, Sepulveda-Swatson E, Cruz-Lemini M, Rojas-Benavente J, Gonzalez-Tendero A, Garcia-Posada R, Rodriguez-Lopez M, Demicheva E, Sitges M, Gratacos E (2014) Postsystolic shortening by myocardial deformation imaging as a sign of cardiac adaptation to pressure overload in fetal growth restriction. Circ Cardiovasc Imaging 7:781–787CrossRefPubMed
25.
Zurück zum Zitat Firpo C, Hoffman JIE, Silverman NH (2001) Evaluation of fetal heart dimensions from 12 weeks to term. Am J Cardiol 87:594–600CrossRefPubMed Firpo C, Hoffman JIE, Silverman NH (2001) Evaluation of fetal heart dimensions from 12 weeks to term. Am J Cardiol 87:594–600CrossRefPubMed
26.
Zurück zum Zitat Wilson AD, Rao S, Aeschlimann S (1990) Normal fetal foramen ovale flap and transatrial Doppler velocity pattern. J Am Soc Echocardiogr 3:491–494CrossRefPubMed Wilson AD, Rao S, Aeschlimann S (1990) Normal fetal foramen ovale flap and transatrial Doppler velocity pattern. J Am Soc Echocardiogr 3:491–494CrossRefPubMed
27.
Zurück zum Zitat Taketazu M, Barrea C, Smallhorn JF, Wilson GJ, Hornberger LK (2004) Intrauterine pulmonary venous flow and restrictive foramen ovale in fetal hypoplastic left heart syndrome. J Am Coll Cardiol 10:1902–1907CrossRef Taketazu M, Barrea C, Smallhorn JF, Wilson GJ, Hornberger LK (2004) Intrauterine pulmonary venous flow and restrictive foramen ovale in fetal hypoplastic left heart syndrome. J Am Coll Cardiol 10:1902–1907CrossRef
29.
Zurück zum Zitat Lloyd DFA, van Amerom JFP, Pushparajah K, Simpson JM, Zidere V, Miller O, Sharland G, Allsop J, Fox M, Lohezic M, Murgasova M, Malamateniou C, Hajnal JV, Rutherford M, Razavi R (2016) An exploration of the potential utility of fetal cardiovascular MRI as an adjunct to fetal echocardiography. Prenat Diagn 36:916–925CrossRefPubMedPubMedCentral Lloyd DFA, van Amerom JFP, Pushparajah K, Simpson JM, Zidere V, Miller O, Sharland G, Allsop J, Fox M, Lohezic M, Murgasova M, Malamateniou C, Hajnal JV, Rutherford M, Razavi R (2016) An exploration of the potential utility of fetal cardiovascular MRI as an adjunct to fetal echocardiography. Prenat Diagn 36:916–925CrossRefPubMedPubMedCentral
Metadaten
Titel
D-Transposition of the great arteries with restrictive foramen ovale in the fetus: the dilemma of predicting the need for postnatal urgent balloon atrial septostomy
verfasst von
I. Gottschalk
A. Walter
T. Menzel
E. C. Weber
S. Wendt
N. Sreeram
U. Gembruch
C. Berg
J. S. Abel
Publikationsdatum
27.03.2023
Verlag
Springer Berlin Heidelberg
Erschienen in
Archives of Gynecology and Obstetrics / Ausgabe 4/2024
Print ISSN: 0932-0067
Elektronische ISSN: 1432-0711
DOI
https://doi.org/10.1007/s00404-023-06997-8

Weitere Artikel der Ausgabe 4/2024

Archives of Gynecology and Obstetrics 4/2024 Zur Ausgabe

Alter der Mutter beeinflusst Risiko für kongenitale Anomalie

28.05.2024 Kinder- und Jugendgynäkologie Nachrichten

Welchen Einfluss das Alter ihrer Mutter auf das Risiko hat, dass Kinder mit nicht chromosomal bedingter Malformation zur Welt kommen, hat eine ungarische Studie untersucht. Sie zeigt: Nicht nur fortgeschrittenes Alter ist riskant.

Fehlerkultur in der Medizin – Offenheit zählt!

28.05.2024 Fehlerkultur Podcast

Darüber reden und aus Fehlern lernen, sollte das Motto in der Medizin lauten. Und zwar nicht nur im Sinne der Patientensicherheit. Eine negative Fehlerkultur kann auch die Behandelnden ernsthaft krank machen, warnt Prof. Dr. Reinhard Strametz. Ein Plädoyer und ein Leitfaden für den offenen Umgang mit kritischen Ereignissen in Medizin und Pflege.

Mammakarzinom: Brustdichte beeinflusst rezidivfreies Überleben

26.05.2024 Mammakarzinom Nachrichten

Frauen, die zum Zeitpunkt der Brustkrebsdiagnose eine hohe mammografische Brustdichte aufweisen, haben ein erhöhtes Risiko für ein baldiges Rezidiv, legen neue Daten nahe.

Mehr Lebenszeit mit Abemaciclib bei fortgeschrittenem Brustkrebs?

24.05.2024 Mammakarzinom Nachrichten

In der MONARCHE-3-Studie lebten Frauen mit fortgeschrittenem Hormonrezeptor-positivem, HER2-negativem Brustkrebs länger, wenn sie zusätzlich zu einem nicht steroidalen Aromatasehemmer mit Abemaciclib behandelt wurden; allerdings verfehlte der numerische Zugewinn die statistische Signifikanz.

Update Gynäkologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert – ganz bequem per eMail.