NEO E-vita—NEO era!

verfasst von: Heinz Jakob, Fanar Mourad

Erschienen in: Indian Journal of Thoracic and Cardiovascular Surgery | Sonderheft 1/2022

download
DOWNLOAD
print
DRUCKEN
insite
SUCHEN

Abstract

In this review article, the history of the first commercially available thoracic aortic hybrid graft, the E-vita Open, later modified to the blood-impermeable E-vita Open Plus, is reported from its beginning in 2005 until its newest variation, the E-vita Open NEO, European conformity (CE) marked in 2020. Besides the background of its design and clinical experience in Essen, concomitant evolutionary steps in surgery as well as in strategic approaches like the hybrid operating room concept are displayed, finally leading to a well-rounded surgical package with a device that can be applied in all elective as well as emergency situations with complex arch involving aortic pathologies. With the E-vita Open NEO, now, surgery has been facilitated to convenient anastomosing in any of the arch zones, with the opportunity to use the island technique with a straight graft variation as well as individual head vessel anastomosing with either a trifurcated graft for zone 0 or the branched graft for zone 2 or 3 implantation. With its proven long-term stability, the surgical armamentarium to cope with complex multisegmental thoracic aortic pathologies has been significantly improved.
Hinweise

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Based on early experience with many redo cases after proximal repair of type I aortic dissection, the operative Essen program using covered stent grafts (Talent, Medtronic, Meerbusch, Germany) started in June 2001 to treat arch and descending aorta aneurysm formation or distal malperfusion, by inserting the reversed mounted stent graft onto the introducer, through the resected aortic arch into the true lumen (TL) of the descending aorta [1].
Due to a dismal experience with the stiff introducer and the creation of proximal endoleaks, caused by penetrating bare springs, despite stay sutures for fixation within the new arch prosthesis, the program was stopped after 14 cases. Based on our conception, a one-piece hybrid graft, without bare springs, was designed and a clinical prototype was produced by JOTEC (Hechingen, Germany) in 2004 (Fig. 1). In 2005, this hybrid graft became the first commercially available hybrid vascular prosthesis in Europe, later called frozen elephant trunk (FET), thus initiating the era of intended one-stage repair of complex thoracic aortic disease involving the arch [2, 3]. Due to the fact that “mad cow disease” in those days prohibited coating of the new device with bovine collagen or gelatine, as well as the intended invagination of the arch part into the stented portion within the minimized introducer, the woven prosthesis had to be preclotted with fibrin sealant (Immuno AG, Vienna, Austria, later Baxter, Höchstadt, Germany) (Fig. 2). Though efficient, primary blood impermeability became the next step forward and was achieved by a tighter weaving process. This could be demonstrated in animal experiments and first clinical application, leading to the E-vita Open Plus product, which was then unanimously accepted by the surgical community in 2008 [4] (Figs. 3 and 4).
In parallel, several additional developments in surgical strategies took place (Table 1). These include important concomitant innovations like the introduction of our hybrid room concept in 2004, enabling simultaneous diagnostics and therapy on one table [5, 6]; the anesthesiological improvements in coagulation management by rotational thromboelastometry in 2005 [7]; the direct true lumen cannulation in disastrous acute type A aortic dissections in 2006 [8]; and the institution of angioscopy for guidance of the FET into the true lumen of the invisible descending aorta in aortic dissection and control of its positioning [9].
Table 1
Surgical strategy and device development in Essen
 
1st application
Reference
Arch repair + TEVAR antegrade
2001
EJCTS 2002 [1]
Hybrid room concept
2004
Herz 2011, EJCTS 2013 [5, 6]
E-vita Open implantation
Combining classic surgery with descending stent grafting
2005
Herz 2005 [2]
Ann Thorac Surg 2008 [3]
Thromboelastometry-directed coagulation management
2005
Transfus Med Hemother 2012 [7]
Direct true lumen asc. aorta cannulation
2006
JTCVS 2007 [8]
Angioscopy—aortic disease and landing zone evaluation
2007
EJCTS 2010 [9]
E-vita Open Plus
2008
J Endovasc Ther 2010 [4]
Stent graft length reduction to 130 mm
Proximalization of distal anastomosis to zone 2
2008/2009
EJCTS 2012 [14, 16]
International E-vita Open registry
2008
J Cardiovasc Surg 2011 [8, 11]
Left subclavian artery bypass first (on/off pump)
2010
Thorac Cardiovasc Surg 2012
LiquoGuard CSF drainage in CAD and TAA
2011
No publication
Immediate visceral reperfusion after FET anastomosis
2012
MITAT 2015 [17]
Enovia (3 Zone Graft—AAD)
2013
Charing Cross 2015
JTCVS 2020
E-vita Neo (Zone-0 FET)
2017
2020
Aortic Live 2017
CTS Net 2020
TEVAR thoracic endovascular aortic repair, CSF cerebrospinal fluid, CAD chronic aortic dissection, TAA thoraco-abdominal aortic aneurysm, FET frozen elephant trunk, AAAD acute type A aortic dissection
Over more than one decade, data was sampled and validated at Essen and got affirmation by the International E-vita Open Registry, which was initiated in 2008 by us. It was expanding from 8 to 19 participating centers over a 12-year period, with up to 1200 patient data sets demonstrating the effectiveness and durability of the device in the treatment of acute and chronic dissection as well as aneurysmal disease [1013]
This accumulation of knowledge identified unacceptable high rates of paraplegia in chronic dissection cases [14, 15] and led to modifications of our surgical approach: reduction of the stent graft length from 15 to 13 cm, distal positioning not exceeding T-level 8 by the stent graft end, and the introduction of the LiquoGuard system (Moeller Medical GmbH, Fulda, Germany) to regulate intraspinal pressure to a maximum of 10 mmHg, by continuous automated liquor drainage throughout surgery. To prevent delayed paraplegia, the device was kept in place for 72 h postoperatively. In addition, mean arterial pressure was kept beyond 80 mmHg and the central venous pressure was reduced to below 12 mmHg. To make sure that all cerebral regions were homogeneously perfused beyond the near-infrared spectroscopy (NIRS)–controlled coverage of the anterior parts of the brain, simultaneous perfusion of all head vessels during hypothermic circulatory arrest (HCA) was started in 2009. An additional positive side effect was the fact that spinal cord nutrition by collateral blood flow from the left subclavian arterial system was augmented, and ever since then, persistent spinal cord ischemic damage became very rare (1%). This was made possible due to rerouting of the sometimes difficult-to-reach left subclavian artery (LSA), by creating an extra-anatomic left axillary artery bypass with an 8-mm graft, which was then transferred into the mediastinum via the first intercostal space and connected to a second pump [1618] (Fig. 5A). Thus, the following strategy with extracorporeal circulation (ECC) was elaborated. In detail, routine arterial cannulation of the right axillary artery and venous cannulation of the right atrium using a dual-stage cannula were done, as well as left ventricular venting, while cooling the patient to a core temperature of 28–30 °C (Fig. 6A).
The second arterial pump is connected to the left axillary artery via the extra-anatomic bypass, which is started immediately after finishing the proximal aortic repair, start of hypothermic circulatory arrest, and transection of the distal ascending aorta. Deairing and ligature of the proximal left subclavian artery precede this maneuver. Via a Y-connector, the second arm of the first arterial pump is cannulated into the left common carotid artery (LCCA); thus, full perfusion of all 3 head vessels is achieved, while cooling the perfusate to 22–24 °C (Fig. 6B). After resection of the aortic arch to zone 2, the E-vita Open Plus hybrid graft is inserted along a stiff guide wire into the TL of the descending aorta in dissection cases, or beyond the aneurysmal pathology in the distal arch guided by angioscopy. After the aortic anastomosis, a Foley 30 Ch balloon-tipped catheter, connected via a Y-connector to the second arterial pump, is inserted into the stented portion of the hybrid graft [17] (Figs. 5B and 6C). After meticulous deairing for removal of potential air bubbles obstructing the offspring of the spinal arteries, full body perfusion and slow rewarming of the patient are started. After retraction of the intussuscepted non-stented graft portion into arch position, the reconnection of the head vessels into the arch graft is done either by island- or isolated reimplantation technique. After completion of this anastomosis, the Foley catheter is removed, deairing takes place, and the arch graft is clamped for finishing the procedure by anastomosing the previously implanted ascending graft with the arch graft, while finally rewarming the patient. After cross-clamp release, the extra-anatomic bypass via the second arterial pump is transected and reimplanted to either the ascending or arch graft, where appropriate. Discontinuation of ECC with reversal of heparin by protamine application and meticulous hemostasis including thromboelastometry-guided differentiated replenishment of fibrinogen, platelets, and prothrombin complex concentrate, if required, complete surgery [19].
Though variable in our hands for either island technique or selective head vessel reimplantation into the arch portion of the E-vita Open Plus, many surgeons prefer the individual reconnection of the head vessels with a branched FET like the Thoraflex device (Vascutek, Renfrewshire, Scotland). This prompted the company to embark on the development of 3 variations of the E-vita Open Plus for additional variability and ease of implantation. With European conformity (CE) mark approval in March 2020, now a straight or triple-branched as well as a trifurcated version, all with an additional perfusion port for immediate lower-body perfusion after anastomosis either in zone 3, 2, 1, or even 0, became available (Fig. 7). With a short and surgeon-friendly application kit for fast and easy implantation, this device means a true progress in adopting the surgical approach to all kinds of pathologies and situations, minimizing the need for hypothermic circulatory arrest time. First in man implantation was performed by us during the Aortic Live Symposium in Hamburg 2017 (Table 1).
Stable results and positive remodeling in the stented area of more than 90% during follow-up (Fig. 8A, B) [20, 21] could be achieved. More distally, about two thirds of the patients face positive remodeling depending on the size and number of re-entry sites downstream [22]. In addition, we could prove that acute type I dissections with distal arch/proximal descending aorta re-entry sites could also be a robust indication for FET application, reducing the need for downstream re-intervention or open re-operation within 10 years in our center to 26%, in contrast to 52% after proximal repair, at a low peri-operative mortality risk [23].
It is a well-known fact that the implantation of a stent graft into the descending aorta carries the inherent risk of provoking a distal stent–induced new entry (DSINE). Learning from the past with too much oversizing of the stent graft in aortic dissection cases by taking the total aortic diameter as a guideline, in some cases, subsequent rupture, penetration, and fistula formation were experienced. This prompted our more conservative approach with taking the length of the separating lamella between the TL and false lumen (FL) as a measure for the graft size to be chosen, regardless of whether dissection was acute or chronic. In aneurysmal cases, a maximum of 10% oversizing was accepted taking care to fully cover the pathology with a 2–3-cm overlap into the non-aneurysmal distal part of the descending aorta. This resulted in a DSINE incidence of 2.9% in acute type A aortic dissection (ATAAD) cases [24]. In chronic aortic dissection (CAD) cases treated either by FET or thoracic endovascular aortic repair (TEVAR), the reported DSINE incidence ranges from 6.5 to 28% within 3 years [2427].
To probably further reduce the incidence of DSINE, the z-stent orientation within the E-vita Open NEO device was switched from tip-to-tip spring orientation to a tip-to-valley constellation, thus creating better alignment to curved structures, which probably will reduce the radial recoil force of the stented portion. In addition, the distal 2 springs were positioned inside the fabric. Together with the non-oversizing concept, both components are key to the reduction of local wall stress to the separating membrane in dissection cases (Fig. 9).
More recently, oozing through the pores of the E-vita Open NEO was made an issue, although the weaving process of the new graft was identical to the previous one, which had not caused this phenomenon for over a decade [28]. However, modern coagulation management using rotational thromboelastometry provides important information about hemostasis, particularly in regard to fibrin polymerization and platelet function, and is available in most tertiary-care institutions. Accordingly, cryoprecipitate can be prepared timely at cross-clamp release, if fibrinogen concentrate is not available. Adequate fibrin and platelet contributions to the clot firmness are essential for the formation of an internal fibrin layer to seal the prosthesis. Thus, volume overload using fresh frozen plasma infusions, with its very low fibrinogen contents, can be specifically avoided [19, 29].
Other groups also report comparable results with different FET devices like the branched Thoraflex graft and the Japanese version (Japan LifeLine, Tokyo, Japan) [30, 31]. The Chinese Cronus device (MicroPort Medical, Shanghai, China), though available since 2004, is mainly implanted in China with reported excellent results [32]. Cronus, as well as Frozenix, requires the addition of a second arch graft to be anastomosed to the cut FET, either in zone 2 or 3.
Overall, the 15-year experience with the E-vita Open and Open Plus graft, with all the demonstrated influential strategic innovations, can serve as a reference for comparison with other approaches with different FET devices. The recently reported results, especially with the last 177 patient cohort, are quite satisfying with their 11% 30-day mortality, 1% persisting spinal cord injury, and 7% new stroke occurrence [33].
With this prerequisite in mind, the E-vita Open NEO represents the next generation of perfected hybrid prostheses for the treatment of complex thoracic aortic diseases encompassing the arch and adjacent structures. Thus, a new era starts and the upcoming less invasive endovascular treatment options, for this kind of pathologies, have to be measured with this gold standard and its proven long-term stability.

Declarations

Conflict of interest

Heinz Jakob is a consultant to JOTEC/CryoLife, former inventor, and recipient of royalties until 2019. Fanar Mourad has no conflict of interest.

Statement of human and animal rights

Not applicable.
N/A.

Ethical statement

N/A.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

Die Chirurgie

Print-Titel

Das Abo mit mehr Tiefe

Mit der Zeitschrift Die Chirurgie erhalten Sie zusätzlich Online-Zugriff auf weitere 43 chirurgische Fachzeitschriften, CME-Fortbildungen, Webinare, Vorbereitungskursen zur Facharztprüfung und die digitale Enzyklopädie e.Medpedia.

Bis 30. April 2024 bestellen und im ersten Jahr nur 199 € zahlen!

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

Literatur
11.
Zurück zum Zitat Jakob H, Tsagakis K, Pacini D, et al. The International E-vita Open Registry: data sets of 274 patients. J Cardiovasc Surg (Torino). 2011;52:717–23. Jakob H, Tsagakis K, Pacini D, et al. The International E-vita Open Registry: data sets of 274 patients. J Cardiovasc Surg (Torino). 2011;52:717–23.
15.
Zurück zum Zitat Leontyev S, Tsagakis K, Pacini D, et al. Impact of clinical factors and surgical techniques on early outcome of patients treated with frozen elephant trunk technique by using EVITA open stent-graft: results of a multicentre study. Eur J Cardiothorac Surg. 2016;49:660–6. https://doi.org/10.1093/ejcts/ezv150.CrossRefPubMed Leontyev S, Tsagakis K, Pacini D, et al. Impact of clinical factors and surgical techniques on early outcome of patients treated with frozen elephant trunk technique by using EVITA open stent-graft: results of a multicentre study. Eur J Cardiothorac Surg. 2016;49:660–6. https://​doi.​org/​10.​1093/​ejcts/​ezv150.CrossRefPubMed
Metadaten
Titel
NEO E-vita—NEO era!
verfasst von
Heinz Jakob
Fanar Mourad
Publikationsdatum
19.01.2022
Verlag
Springer Singapore
Erschienen in
Indian Journal of Thoracic and Cardiovascular Surgery / Ausgabe Sonderheft 1/2022
Print ISSN: 0970-9134
Elektronische ISSN: 0973-7723
DOI
https://doi.org/10.1007/s12055-021-01303-0

Weitere Artikel der Sonderheft 1/2022

Indian Journal of Thoracic and Cardiovascular Surgery 1/2022 Zur Ausgabe

Wie erfolgreich ist eine Re-Ablation nach Rezidiv?

23.04.2024 Ablationstherapie Nachrichten

Nach der Katheterablation von Vorhofflimmern kommt es bei etwa einem Drittel der Patienten zu Rezidiven, meist binnen eines Jahres. Wie sich spätere Rückfälle auf die Erfolgschancen einer erneuten Ablation auswirken, haben Schweizer Kardiologen erforscht.

Hinter dieser Appendizitis steckte ein Erreger

23.04.2024 Appendizitis Nachrichten

Schmerzen im Unterbauch, aber sonst nicht viel, was auf eine Appendizitis hindeutete: Ein junger Mann hatte Glück, dass trotzdem eine Laparoskopie mit Appendektomie durchgeführt und der Wurmfortsatz histologisch untersucht wurde.

Mehr Schaden als Nutzen durch präoperatives Aussetzen von GLP-1-Agonisten?

23.04.2024 Operationsvorbereitung Nachrichten

Derzeit wird empfohlen, eine Therapie mit GLP-1-Rezeptoragonisten präoperativ zu unterbrechen. Eine neue Studie nährt jedoch Zweifel an der Notwendigkeit der Maßnahme.

Ureterstriktur: Innovative OP-Technik bewährt sich

19.04.2024 EAU 2024 Kongressbericht

Die Ureterstriktur ist eine relativ seltene Komplikation, trotzdem bedarf sie einer differenzierten Versorgung. In komplexen Fällen wird dies durch die roboterassistierte OP-Technik gewährleistet. Erste Resultate ermutigen.

Update Chirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.

S3-Leitlinie „Diagnostik und Therapie des Karpaltunnelsyndroms“

Karpaltunnelsyndrom BDC Leitlinien Webinare
CME: 2 Punkte

Das Karpaltunnelsyndrom ist die häufigste Kompressionsneuropathie peripherer Nerven. Obwohl die Anamnese mit dem nächtlichen Einschlafen der Hand (Brachialgia parästhetica nocturna) sehr typisch ist, ist eine klinisch-neurologische Untersuchung und Elektroneurografie in manchen Fällen auch eine Neurosonografie erforderlich. Im Anfangsstadium sind konservative Maßnahmen (Handgelenksschiene, Ergotherapie) empfehlenswert. Bei nicht Ansprechen der konservativen Therapie oder Auftreten von neurologischen Ausfällen ist eine Dekompression des N. medianus am Karpaltunnel indiziert.

Prof. Dr. med. Gregor Antoniadis
Berufsverband der Deutschen Chirurgie e.V.

S2e-Leitlinie „Distale Radiusfraktur“

Radiusfraktur BDC Leitlinien Webinare
CME: 2 Punkte

Das Webinar beschäftigt sich mit Fragen und Antworten zu Diagnostik und Klassifikation sowie Möglichkeiten des Ausschlusses von Zusatzverletzungen. Die Referenten erläutern, welche Frakturen konservativ behandelt werden können und wie. Das Webinar beantwortet die Frage nach aktuellen operativen Therapiekonzepten: Welcher Zugang, welches Osteosynthesematerial? Auf was muss bei der Nachbehandlung der distalen Radiusfraktur geachtet werden?

PD Dr. med. Oliver Pieske
Dr. med. Benjamin Meyknecht
Berufsverband der Deutschen Chirurgie e.V.

S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“

Appendizitis BDC Leitlinien Webinare
CME: 2 Punkte

Inhalte des Webinars zur S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“ sind die Darstellung des Projektes und des Erstellungswegs zur S1-Leitlinie, die Erläuterung der klinischen Relevanz der Klassifikation EAES 2015, die wissenschaftliche Begründung der wichtigsten Empfehlungen und die Darstellung stadiengerechter Therapieoptionen.

Dr. med. Mihailo Andric
Berufsverband der Deutschen Chirurgie e.V.