Skip to main content
Erschienen in: Journal of Orthopaedic Surgery and Research 1/2019

Open Access 01.12.2019 | Research article

Percutaneous kyphoplasty assisted with/without mixed reality technology in treatment of OVCF with IVC: a prospective study

verfasst von: Peiran Wei, Qingqiang Yao, Yan Xu, Huikang Zhang, Yue Gu, Liming Wang

Erschienen in: Journal of Orthopaedic Surgery and Research | Ausgabe 1/2019

Abstract

Background

The purpose of this study was to assess the clinical outcome of percutaneous kyphoplasty (PKP) assisted with mixed reality (MR) technology in treatment of osteoporotic vertebral compression fracture (OVCF) with intravertebral vacuum cleft (IVC).

Method

Forty cases of OVCF with IVC undergoing PKP were randomized into a MR technology-assisted group (group A) and a traditional C-arm fluoroscopy group (group B). Both groups were performed PKP and evaluated by VAS scores, ODI scores, radiological evidence of vertebral body height, and kyphotic angle (KA) at pre-operation and post-operation. The volume of injected cement, fluoroscopy times, and operation time were recorded. And cases of non-PMMA-endplates-contact(NPEC) in radiological evidence was also recorded postoperatively. The clinical outcomes and complications were evaluated afterwards. All patients received 10 to 14 months follow-up, with an average of 12 months.

Result

This MR-assisted group (group A) acquired more about the amount of the polymethyl methacrylate (PMMA) injection and postoperative vertebral height and less about postoperative KA, fluoroscopy times, and operation time compared with the control group (group B) (P < 0.05). The VAS scores and ODI scores in both groups have improved, but more significantly in group A (P < 0.05). Also, more cases achieve both-endplates-touching of cement in group A (P < 0.05). And there are less of the loss of vertebral height, KA, and occurrence of re-collapse of the vertebra in group A during the follow-up (P < 0.05).

Conclusion

PKP assisted with MR technology can accurately orientate the position of IVC area, which can be augmented by the balloon leading to more satisfied vertebral height improvement, cement diffusion, and pain relief.

Trial registration

ClinicalTrials.gov Identifier: NCT03959059. Registered 25 September 2016.
Hinweise

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
BMD
Bone mineral density
CAD
Computer aided design
CT
Computed tomography
IVC
Intravertebral vacuum cleft
KA
Kyphotic angle
MR
Mixed reality
MRI
Magnetic resonance imaging
ODI
Oswestry Disability Index
OVCF
Osteoporotic vertebral compression fracture
PKP
Percutaneous kyphoplasty
PMMA
Poly methyl methacrylate
VAS
Visual analog scale

Key points

1.
Destruction of membrane at the periphery of the IVC is probably the key point for distribution of cement to be interdigitated with the surrounding cancellous bone for the sake of better clinical efficacy.
 
2.
The location of IVC in the vertebral body could be accurately acquired preoperatively from the CAD virtual anatomic images obtained by fusion of MRI and three-dimensional CT images, which was a great advantage for the guidance.
 
3.
PKP assisted with MR technology could acquire real-time and accurate guidance to the IVC area during operation.
 

Background

Percutaneous kyphoplasty (PKP) is an effective and currently widely used method for the treatment of osteoporotic vertebral compression fracture (OVCF), the procedure was done usually under local anesthesia, and the patient was well tolerated [1]. Early studies have shown that patients could obtain promising clinical outcome of immediate pain relief and functional improvement, especially among the elderly [26]. However, there is still a very number of patients who were unsatisfied with the result after operation. As for these patients, they complain about their unsatisfied or unchanged pain relief or even worse pain, which may indicate continued compression or a recurrent fracture in the treated vertebra. Previous studies have shown that intravertebral vacuum cleft (IVC) in acute OVCF was not a rare phenomenon [710] and it also has been considered as an important risk factor for persistent back pain and severe vertebral collapse and might be the main reason responsible for unsatisfied outcome after PKP [11, 12]. On magnetic resonance imaging (MRI), an IVC presents linear or cystic hypointensity similar to air on T1-weighted sequences and hyperintensity similar to cerebrospinal fluid on T2-weighted sequences [13, 14]. It also has been demonstrated [15, 16] that during the PKP procedure, instead of diffusing into the surrounding cancellous bone, the cement was formed into a solid lump that causes the fibrocartilaginous membrane at the periphery of IVC inhibited its dispersion. The fibrocartilaginous membrane can be broken by the balloon injected into the IVC area, leading to improvement of the cement diffusion [17, 18]. Therefore, in order to break the fibrocartilaginous membrane of IVC by the balloon during PKP procedure, so as to make the cement sufficiently diffuse into surrounding cancellous bone, we attempted to acquire accurate navigation to the IVC area.
Mixed reality (MR) is a combination of VR and AR in 3D applications [19], and it divorces digital reality and virtual digital images from the virtual world on the screen, so that three-dimensional virtual objects and the real world can be accurately combined [20, 21]. Previous studies have reported that MR technology has been used to show three-dimensional reconstructed model of all aspects, both preoperatively and intraoperatively; the muscles, blood vessels, and bones could be accurately seen; and the anatomical data could be measured [22]. Thus, it could help the surgeon to select the best operative program individually and precisely [21]. Under this technology, the real tissues and preoperative image data can be matched and fused to achieve real-time three-dimensional visual positioning and real-time display of the position of surgical tools, so as to carry out surgical navigation [23, 24]. The purpose of the present study was to evaluate PKP assisted with precise navigation by MR technology in the treatment of OVCF with IVC.

Methods

Patients

All patients participated in the trial and signed informed consent voluntarily. There was also no financial relationship between the investigators and study participants. Patients were required to meet the following inclusion criteria: (1) single-level OVCF in the thoracic and lumbar levels (T10-L4) [25], suffering with severe back pain, and (2) OVCF without damaged vertebral posterior wall and nerve lesion; (3) the patient age was more than 50 years old, and diagnosis of osteopenia or osteoporosis was confirmed by bone densitometry; and (4) all patients underwent plain radiography, computed tomography (CT), and MRI before surgery, with IVC presented in MRI. From June 2017 to September 2017, there were 40 patients who met the criteria above; who participated the trial, aging from 50 to 95 years old; and were randomized into MR technology-assisted group (group A) and traditional C-arm fluoroscopy group (group B), 20 cases for each group.
Approval was obtained from the Institutional Review Board of Nanjing First Hospital prior to performing the study.

Assessment of clinical parameters

Several parameters including age, sex, and bone mineral density (BMD) of the two groups have been evaluated preoperatively. Operation duration, fluoroscopy times, and volume of cement injected to both groups were recorded respectively. Symptoms of two groups were evaluated by visual analog scale (VAS) score of back pain and Oswestry Disability Index (ODI) of movement function. Immediate postoperative, 1-, 3-, 6 months, and 1 year VAS and ODI scores were compared.

Assessment of radiological parameters

We measured preoperative and postoperative (1 day, 1-, 3-, 6 months, and 1 year) radiological parameters including kyphotic angle (KA) from sagittal radiographs, and anterior, central, and posterior vertebral height; anterior/posterior height ratio was defined as relative anterior vertebral height, and central/posterior height ratio was defined as relative central vertebral height (Fig. 1a). Kyphotic angle (KA) was formed by the upper and lower endplates of the fractured vertebral body (Fig. 1b). Patients underwent CT scanning pre- and postoperatively (Fig. 1d, f). IVC was presented in MRI image (Fig. 1e). And cases of cement touching both endplates in each of the two groups were recorded postoperatively (Fig. 1c). In group A, patients also required accurate three-dimensional images to confirm the location of IVC in the vertebral body, which was obtained by computer-aided design (CAD) of preoperative MRI and three-dimensional CT images, for the purpose of navigation by MR technology (Fig. 2).

Surgery

All of the surgeries were performed by one group of surgeons and adopted bipedicular approach in a prone posture under local anesthesia (1% lidocaine) in all cases.
Group A: All the PKP operations were performed by one group of surgeons. Persons were in the prone position with local anesthesia (1% lidocaine), through bilateral transpedicle puncture approach. Cases in MR-PKP group (group A) underwent dual-source 64-slice CT volume scanning (Siemens Sensation, Germany) with a slice thickness of 1 mm and a scanning matrix of 512 × 512. Input the data obtained into the M3D digital medical software (Bahulu, Shanghai Front Computing Company, China); hence, the three-dimensional model was obtained by using its threshold segmentation and regional growth functions. The responsible vertebral body and adjacent vertebral body were labeled with different colors, the location and direction of pedicle insertion were pre-designed, and the three-dimensional virtual digital CAD image was constructed (Fig. 2). Before the operation, the CAD image was imported into the computer of MR system (Medivi, Changzhou, China). Doctors wore AR glasses HoloLens (Microsoft, USA). As the original data for CT reconstruction was taken in supine position, the patient is placed in prone position during operation. We need C-arm fluoroscopy during operation to adjust the CAD image to optimally match the patient in prone position. X-ray fluoroscopy was performed, and the spatial registration of virtual image and patient’s body was completed by manual positioning. With MR technology and C-arm fluoroscopy, these 3D virtual images can be accurately combined with the patient, so the surgeon could see the spinal 3D virtual images (CAD images) accurately in the patient’s body. Through this combination of virtual images and real body, also combined with C-arm fluoroscopy, needles were accurately inserted through bilateral pedicle of the injured vertebra until the tip was optimally positioned in the IVC. Then the needles were removed from the trochar, and a balloon was placed into the channel and inflated in order to augment the IVC area, elevate the endplate, and restore the height of the vertebral body. Cement (poly methyl methacrylate, PMMA) was injected into the IVC area for attempting to make the cement to be sufficiently interdigitated with the surrounding cancellous bone and complete filling of the cleft, maximizing the stabilization of the IVC and fracture (Fig. 3).
Group B: Insertion of the transpedicular access was performed bilaterally under C-arm fluoroscopic guidance. Combined with the X-ray, CT, MRI images preoperatively and the C-arm fluoroscopic images, the tip of needle was set to the most possible area of the IVC, the needle was removed and a balloon was inserted to the position, which was inflated and expanded to restore the height of the vertebra, and then the cement (PMMA) was injected. The filling and diffusion of bone cement were observed while injection was carried out. And the injection was stopped when the bone cement reached the posterior edge of the vertebral body.

Statistical analysis

SPSS V.11.0 statistical software was used for analysis. Comparisons were made before and at each postoperative follow-up appointment. Qualitative characteristics of groups were assessed using one-way or repeated measures ANOVA and t test. χ2 tests were performed for categorical variables. P < 0.05 was considered statistically significant.

Results

Forty patients (M/F 13:27) were randomly categorized into two groups, each group with 20 cases. The mean age of the patients was 73.48 ± 8.72 years in group A and 74.02 ± 8.85 years in group B, and the length of follow-up ranged from 10 to 14 months (median 12 months). There was no significant difference between the two groups in baseline parameters including age, sex, BMD, VAS, ODI, vertebral height, KA, level of injured vertebra, and the time from injury to surgery (P > 0.05, Table 1).
Table 1
Basic characteristics and preoperative data of the two groups
Patient characteristics
Group A
Group B
P
Age (years)
73.48 ± 8.72
74.02 ± 8.85
0.97
Women, n (%)
14 (70%)
13 (65%)
0.77
BMD score
− 4.35 ± 0.94
− 4.44 ± 0.76
0.89
VAS scores
8.38 ± 1.28
8.14 ± 1.35
0.31
ODI scores
83.95 ± 8.96
84.27 ± 7.48
0.13
Relative anterior vertebral height (%)
77.51 ± 5.91
75.72 ± 6.73
0.79
Relative central vertebral height (%)
56.74 ± 4.13
58.28 ± 3.62
0.96
KA (°)
14.02 ± 4.06
12.78 ± 4.75
0.39
Time from injury to surgery (number of patients)
 1 day
0
1
0.97
 2 days
0
1
 
 3 days
15
14
 
 4 days
2
2
 
 5 days
3
2
 
Level of injured vertebra (number of patients)
 T10
1
1
0.66
 T11
1
1
 
 T12
5
6
 
 L1
6
7
 
 L2
4
3
 
 L3
2
2
 
 L4
1
0
 
The operation duration and fluoroscopy times in group A were less than those in group B (P < 0.05). The amount of PMMA injection during MR guidance (group A) was more than that during C-arm fluoroscopic guidance (Table 2, P < 0.05).
Table 2
Postoperative data of the two groups
 
Group A
Group B
P
Operative time (min)
25.12 ± 5.36
45.14 ± 3.86
< 0.05
Fluoroscopy times
31.87 ± 9.77
76.94 ± 8.65
< 0.05
Amount of cement injection (ml)
6.41 ± 0.54
4.22 ± 0.35
< 0.05
Cases of cement of both-endplates-contact
13
6
< 0.05
Relative anterior vertebral height (%)
92.42 ± 5.14
71.21 ± 4.83
< 0.05
Relative central vertebral height (%)
79.74 ± 3.92
60.53 ± 3.31
< 0.05
Improvement of relative anterior vertebral height (%)
12.81 ± 4.76
6.94 ± 3.42
< 0.05
Improvement of relative central vertebral height (%)
19.71 ± 2.82
3.42 ± 3.57
< 0.05
KA (°)
7.11 ± 3.01
10.89 ± 2.91
< 0.05
Change of KA (°)
7.11 ± 1.21
3.72 ± 1.53
< 0.05
The postoperative VAS and ODI scores were followed about 1 year. Immediately after surgery, the mean VAS and ODI scores of all patients in both of the two groups decreased significantly (P < 0.05), but the scores decreased more in group A (P < 0.05). However, we observed a gradual significant increase in VAS and ODI scores during the 1-year follow-up after surgery in group B. The VAS and ODI scores were significantly higher in group B than those in group A at the 1-year postoperative follow-up (P < 0.05). The VAS and ODI scores in group B immediately after surgery were 2.58 ± 1.09 and 29.11 ± 8.07, respectively, and increased to 3.42 ± 1.16 and 35.07 ± 9.1 at 1 year, respectively (Tables 2 and 3).
Table 3
VAS and ODI scores and imaging parameters during 1 year follow-up
 
Group A
Group B
VAS
 Immediate postoperative
1.37 ± 1.01
2.58 ± 1.09
 1 month
1.33 ± 1.03
2.96 ± 1.15
 3 month
1.39 ± 1.04
3.17 ± 1.20
 6 month
1.36 ± 1.08
3.21 ± 1.08
 12 month
1.31 ± 1.12
3.42 ± 1.16
ODI
 Immediate postoperative
18.97 ± 7.79
29.11 ± 8.07
 1 month
19.05 ± 8.03
31.34 ± 8.73
 3 month
19.23 ± 8.77
33.68 ± 7.85
 6 month
19.51 ± 8.02
33.94 ± 8.67
 12 month
19.87 ± 8.34
35.07 ± 9.10
Relative anterior vertebral height (%)
 Immediate postoperative
92.42 ± 5.14
71.21 ± 4.83
 1 month
92.21 ± 4.49
68.88 ± 5.01
 3 month
91.75 ± 6.44
66.35 ± 5.94
 6 month
91.69 ± 5.02
63.81 ± 6.23
 12 month
91.58 ± 5.34
61.72 ± 6.96
Relative central vertebral height (%)
 Immediate postoperative
79.74 ± 3.92
60.53 ± 3.31
 1 month
79.19 ± 3.88
58.91 ± 3.56
 3 month
78.77 ± 4.07
58.21 ± 3.87
 6 month
78.65 ± 3.98
57.37 ± 3.61
 12 month
78.31 ± 4.11
56.96 ± 3.71
KA (°)
 Immediate postoperative
7.11 ± 3.01
10.89 ± 2.91
 1 month
7.73 ± 3.11
11.56 ± 3.34
 3 month
7.78 ± 2.89
11.88 ± 2.85
 6 month
7.85 ± 3.08
11.96 ± 3.41
 12 month
8.01 ± 3.67
12.47 ± 3.28
Vertebral height was defined as relative anterior vertebral height (anterior/posterior height ratio) and relative central vertebral height (central/posterior height ratio). KA and preoperative reduction in height of the vertebra were not significantly different in the two groups (P > 0.05). However, postoperative height of the vertebra (relative anterior height, relative central height) was significantly taller in group A than in groups B (Table 2, P < 0.05). Vertebral kyphotic angle (KA) in group A was also significantly lower than that in group B (Table 2, P < 0.05). Cases of cement-both-endplates-contact were 13 in group A and 5 in group B, which was significantly different (P < 0.05) (Tables 2 and 3).

Complications

No cement leakage, neurologic deficits, re-fracture of the treated vertebra, and other complications were observed in these two groups. There was one case of adjacent vertebral fracture in group B, while none in group A. The perioperative mortality was 0%.

Discussion

IVC in the fractured vertebra has been reported to be the main reason to intractable pain of OVCF and unsatisfied pain relief after PKP [26]. And IVC may cause pseudarthrosis of the vertebra leading to spinal instability, which may give rise to the severe chronic pain. Therefore, it was thought to be a good indication for PKP to acquire pain relief and spinal stabilization through filling IVC with cement. However, Heo et al. [15] found that the cement cannot be diffused to be interdigitated with the surrounding cancellous bone because of the inhibition by the fibrocartilaginous membrane of IVC. It formed into a solid lump which may induce greater stress to the surrounding cancellous bone, leading to recollapse or refracture [15, 18, 27]. It has been reported that distribution of the cement is more important than the volume injected, specifically acquiring the cement touching both endplates may be more crucial for stabilization of the compression fracture [2831]. Both the superior and inferior endplates should be touched with cement to provide complete support of the vertebra to prevent continuing collapse and recurrent fractures [9, 11, 3234]. Biomechanical studies demonstrated that better diffusion of cement distributes the load of the fractured vertebrae [35, 36]. Breaking the peripheral membrane, filling the cleft, and making the cement be interdigitated with the surrounding cancellous bone are the key point to pain relief and preventing further collapse [37]. However, the needle tip in previous study was still traditionally placed in the IVC area and lacks accurate navigation and guidance.
Therefore, we combined MR technology with the real patient in order to achieve accurate navigation to IVC area of OVCF. Mixed reality (MR) is a technology that divorces digital reality and virtual digital images from the virtual world on the screen, so that three-dimensional virtual objects and the real world can be accurately combined, which enables accurate replication of patient anatomy. Using the presented methodology, patient anatomy and surgical instruments can be aligned correctly in spatial relation to each other [38, 39]. Combining 3D reconstruction and MR technology is a new mode of production, which is applied in many fields because of the advantages of visualization, plasticity, and rapid printing [4042]. Moreover, it can also help surgeons to achieve accurate navigation and position in spinal surgery [43, 44].
In our study, the location of IVC in the vertebral body could be accurately acquired in CAD anatomic images preoperatively as a result of fusion of MRI and three-dimensional CT images, and the CAD virtual digital images could be accurately combined with the patient’s body during operation, which was a great advantage for the guidance. Combined with C-arm fluoroscopic images during surgery, MR technology could assist surgeons to acquire real-time and accurate guidance to the IVC area, reducing operation time and numbers of C-arm fluoroscopy and increasing the accuracy and efficacy of surgery.
Our study also demonstrates that mixed reality (MR) technology can change the bottleneck of unarmed puncture in IVC-PKP surgery. The vertebra can be reconstructed and restored by three-dimensional CT reconstruction before operation, so that the location of IVC lesions can be displayed more intuitively in front of doctors. During the operation, the surgeons wore HoloLens glasses and obtained holographic image information, so that the patient’s virtual 3D digital model overlapped completely with the patient’s lesion site; thus, the doctor has a new way of operation [45].
With the assistance of MR in our study, the accuracy of puncture was significantly higher than that of traditional methods during the PKP procedure, especially in some patients with lesions of the vertebrae. The vertebral body may be accompanied by mild lateral bending, rotation, the formation of a bridge, and so on, which increase the difficulty of the operation, and the entry point and direction are difficult to control, especially in patients with IVC. Combined with MR, it can assist virtual positioning, give doctors the correct positioning point and direction when punctured, make the procedure more accurate and safer, is easy to operate, and obviously reduce the number of fluoroscopy and operation time. Elmi-terander et al. [46] confirmed that combined with intraoperative 3D image, navigation technology is expected to replace intraoperative X-ray or fluoroscopy in the process of thoracolumbar pedicle puncture and has high accuracy and safety.
Our study also indicates that the number of X-ray irradiation of C-arm machine in MR-PKP group was significantly lower than that in the traditional group. In the past, C-arm machine was needed to judge the position of the needle many times during operation, which was harmful to doctors. However, with MR navigation technology, the number of C-arm machine used in operation could be significantly reduced, and the injury to doctors and patients could also be obviously reduced.
In our study, we also found that the amount of cement injected and relative vertebral height in the MR-assisted group increased more compared with the control group. The presence of the peripheral membrane may prevent the diffusion of cement; thus, some part of the vertebra cannot be filled with cement [47], leading to the smaller volume injected in group B (Fig. 1f). The vertebral height was reconstructed by balloon and prone postural reduction, without sufficient cement reinforcing, especially touching both endplates; the vertebral height in group B may not be maintained and might be lost after operation [4749]. Daisuke et al. [50] have reported that increased cement injection may be due to the increased balloon expansions, and this increased amount of cement injection may affect postoperative vertebral height improvement.
However, delayed reaction, small field of vision, dim brightness, and insufficient stability of binocular shared vision of MR technology inevitably lead to the need of C-arm fluoroscopy assistance. In the specialized soft- and hardware, the cost was relatively higher.

Conclusion

In conclusion, MR technology was adopted as an efficient method for navigation and position. Compared with the traditional procedure of PKP, it acquired an accurate position of IVC in OVCF and preferable clinical outcomes. We consider this technology as a perfect approach for guidance in minimally invasive surgery.

Limitations

1.
This research lacks of biomechanical study, and the sample size is too small; this also needs long-term follow-up and research.
 
2.
Although with fused CAD anatomic images preoperatively, MR technology cannot acquire real-time MRI images of IVC during operation.
 
3.
For the disadvantages of delayed reaction, small field of vision, dim brightness, and insufficient stability of binocular shared vision, MR technology still needs C-arm fluoroscopy assistance.
 

Acknowledgements

We would like to thank all the participants who took part in this research for their time and help.

Statement

Our study adheres to CONSORT guidelines, and a completed CONSORT checklist was submitted as an additional file.
This study was approved by the ethics committee of Nanjing First Hospital. Written informed consent was obtained from all patients.
Not applicable.

Competing interests

The authors declare that they have no competing interests.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
Zurück zum Zitat Bonnard E, Foti P, Kastler A, Amoretti N. Percutaneous vertebroplasty under local anaesthesia: feasibility regarding patients’ experience. Eur Radiol. 2016;27:1512–6.CrossRef Bonnard E, Foti P, Kastler A, Amoretti N. Percutaneous vertebroplasty under local anaesthesia: feasibility regarding patients’ experience. Eur Radiol. 2016;27:1512–6.CrossRef
2.
Zurück zum Zitat Mathis JM, Barr JD, Belkoff SM, et al. Percutaneous vertebroplasty: a developing standard of care for vertebral compression fractures. Am J Neuroradiol. 2001;22:373–81.PubMed Mathis JM, Barr JD, Belkoff SM, et al. Percutaneous vertebroplasty: a developing standard of care for vertebral compression fractures. Am J Neuroradiol. 2001;22:373–81.PubMed
3.
Zurück zum Zitat Diamond TH, Champion B, Clark WA. Management of acute osteoporotic vertebral fractures: a nonrandomized trial comparing percutaneous vertebroplasty with conservative therapy. Am J Med. 2003;114:257–65.CrossRef Diamond TH, Champion B, Clark WA. Management of acute osteoporotic vertebral fractures: a nonrandomized trial comparing percutaneous vertebroplasty with conservative therapy. Am J Med. 2003;114:257–65.CrossRef
4.
Zurück zum Zitat Do HM, Kim BS, Marcellus ML, et al. Prospective analysis of clinical outcomes after percutaneous vertebroplasty for painful osteoporotic vertebral body fractures. AJNR Am J Neuroradiol. 2005;26:1623–8.PubMed Do HM, Kim BS, Marcellus ML, et al. Prospective analysis of clinical outcomes after percutaneous vertebroplasty for painful osteoporotic vertebral body fractures. AJNR Am J Neuroradiol. 2005;26:1623–8.PubMed
5.
Zurück zum Zitat Voormolen MHJ, Mali WPTM, Lohle PNM, et al. Percutaneous vertebroplasty compared with optimal pain medication treatment: short-term clinical outcome of patients with subacute or chronic painful osteoporotic vertebral compression fractures. The VERTOS study. AJNR Am J Neuroradiol. 2007;28:555–60.PubMed Voormolen MHJ, Mali WPTM, Lohle PNM, et al. Percutaneous vertebroplasty compared with optimal pain medication treatment: short-term clinical outcome of patients with subacute or chronic painful osteoporotic vertebral compression fractures. The VERTOS study. AJNR Am J Neuroradiol. 2007;28:555–60.PubMed
6.
Zurück zum Zitat Clarençon F, Fahed R, Gabrieli J, et al. Safety and clinical effectiveness of percutaneous vertebroplasty in the elderly (≥80 years). Eur Radiol. 2016;26:2352–8.CrossRef Clarençon F, Fahed R, Gabrieli J, et al. Safety and clinical effectiveness of percutaneous vertebroplasty in the elderly (≥80 years). Eur Radiol. 2016;26:2352–8.CrossRef
7.
Zurück zum Zitat Kim YJ, Lee JW, Kim KJ, Chung SK, Kim HJ, Park JM, Kang HS. Percutaneous vertebroplasty for intravertebral cleft: analysis of therapeutic effects and outcome predictors. Skeletal Radiol. 2010;39:757–66.CrossRef Kim YJ, Lee JW, Kim KJ, Chung SK, Kim HJ, Park JM, Kang HS. Percutaneous vertebroplasty for intravertebral cleft: analysis of therapeutic effects and outcome predictors. Skeletal Radiol. 2010;39:757–66.CrossRef
8.
Zurück zum Zitat Kim YC, Kim YH, Ha KY. Pathomechanism of intravertebral clefts in osteoporotic compression fractures of the spine. Spine J. 2014;14:659–66.CrossRef Kim YC, Kim YH, Ha KY. Pathomechanism of intravertebral clefts in osteoporotic compression fractures of the spine. Spine J. 2014;14:659–66.CrossRef
9.
Zurück zum Zitat Linn J, Birkenmaier C, Hoffmann RT, Reiser M, Baur-Melnyk A. The intravertebral cleft in acute osteoporotic fractures: fluid in magnetic resonance imaging-vacuum in computed tomography? Spine. 2009;34:E88–93.CrossRef Linn J, Birkenmaier C, Hoffmann RT, Reiser M, Baur-Melnyk A. The intravertebral cleft in acute osteoporotic fractures: fluid in magnetic resonance imaging-vacuum in computed tomography? Spine. 2009;34:E88–93.CrossRef
10.
Zurück zum Zitat Zhang J, He X, Fan Y, Du J, Hao D. Risk factors for conservative treatment failure in acute osteoporotic vertebral compression fractures (OVCFs). Arch Osteoporos. 2019;14:24.CrossRef Zhang J, He X, Fan Y, Du J, Hao D. Risk factors for conservative treatment failure in acute osteoporotic vertebral compression fractures (OVCFs). Arch Osteoporos. 2019;14:24.CrossRef
11.
Zurück zum Zitat Sasaki Y, Aoki Y, Nakajima A, Shibata Y, Sonobe M, Takahashi K, et al. Delayed neurologic deficit due to foraminal stenosis following osteoporotic late collapse of a lumbar spine vertebral body. Case Reports in Orthopedics. 2013;2013:1–5.CrossRef Sasaki Y, Aoki Y, Nakajima A, Shibata Y, Sonobe M, Takahashi K, et al. Delayed neurologic deficit due to foraminal stenosis following osteoporotic late collapse of a lumbar spine vertebral body. Case Reports in Orthopedics. 2013;2013:1–5.CrossRef
12.
Zurück zum Zitat Ito Y, Hasegawa Y, Toda K, Nakahara S. Pathogenesis and diagnosis of delayed vertebral collapse resulting from osteoporotic spinal fracture. Spine J. 2002;2:101–6.CrossRef Ito Y, Hasegawa Y, Toda K, Nakahara S. Pathogenesis and diagnosis of delayed vertebral collapse resulting from osteoporotic spinal fracture. Spine J. 2002;2:101–6.CrossRef
13.
Zurück zum Zitat Ha KY, Lee JS, Kim KW, et al. Percutaneous vertebroplasty for vertebral compression fractures with and without intravertebral clefts. J Bone Joint Surg Br. 2006;88:629–33.CrossRef Ha KY, Lee JS, Kim KW, et al. Percutaneous vertebroplasty for vertebral compression fractures with and without intravertebral clefts. J Bone Joint Surg Br. 2006;88:629–33.CrossRef
14.
Zurück zum Zitat Koh YH, Han D, Cha JH, Seong CK, Kim J, Choi YH. Vertebroplasty: magnetic resonance findings related to cement leakage risk. Acta Radiologica. 2007;48(3):315–20.CrossRef Koh YH, Han D, Cha JH, Seong CK, Kim J, Choi YH. Vertebroplasty: magnetic resonance findings related to cement leakage risk. Acta Radiologica. 2007;48(3):315–20.CrossRef
15.
Zurück zum Zitat Heo DH, Chin DK, Yoon YS, Kuh SU. Recollapse of previous vertebral compression fracture after percutaneous vertebroplasty. Osteoporosis International. 2009;20(3):473–80.CrossRef Heo DH, Chin DK, Yoon YS, Kuh SU. Recollapse of previous vertebral compression fracture after percutaneous vertebroplasty. Osteoporosis International. 2009;20(3):473–80.CrossRef
16.
Zurück zum Zitat Li KC, Wong TU, Kung FC, Li A, Hsieh CH. Staging of Kümmell’s disease. J Musculoskelet Res. 2004;8(01):13.CrossRef Li KC, Wong TU, Kung FC, Li A, Hsieh CH. Staging of Kümmell’s disease. J Musculoskelet Res. 2004;8(01):13.CrossRef
17.
Zurück zum Zitat Wu AM, Ni WF, Weng W, Chi YL, Xu HZ, Wang XY. Outcomes of percutaneous kyphoplasty in patients with intravertebral vacuum cleft. Acta Orthop Belg. 2012;78:790–5.PubMed Wu AM, Ni WF, Weng W, Chi YL, Xu HZ, Wang XY. Outcomes of percutaneous kyphoplasty in patients with intravertebral vacuum cleft. Acta Orthop Belg. 2012;78:790–5.PubMed
18.
Zurück zum Zitat Kim YY, Rhyu KW. Recompression of vertebral body after balloon kyphoplasty for osteoporotic vertebral compression fracture. Eur Spine J. 2010;19(11):1907–12.CrossRef Kim YY, Rhyu KW. Recompression of vertebral body after balloon kyphoplasty for osteoporotic vertebral compression fracture. Eur Spine J. 2010;19(11):1907–12.CrossRef
19.
Zurück zum Zitat Nicola S, Stoicu-Tivadar L. Mixed reality supporting modern medical education. Stud Health Technol Inform. 2018;255:242–6.PubMed Nicola S, Stoicu-Tivadar L. Mixed reality supporting modern medical education. Stud Health Technol Inform. 2018;255:242–6.PubMed
20.
Zurück zum Zitat Eck U, Winkler A. Display technologies for augmented reality in medical applications. Unfallchirurg. 2018;121:278–85.CrossRef Eck U, Winkler A. Display technologies for augmented reality in medical applications. Unfallchirurg. 2018;121:278–85.CrossRef
21.
Zurück zum Zitat Bova FJ, Rajon DA, Friedman WA, Murad GJ, Hoh DJ, Jacob RP, Lampotang S, Lizdas DE, Lombard G, Lister JR. Mixed-reality simulation for neurosurgical procedures. Neurosurg. 2013:138–45. Bova FJ, Rajon DA, Friedman WA, Murad GJ, Hoh DJ, Jacob RP, Lampotang S, Lizdas DE, Lombard G, Lister JR. Mixed-reality simulation for neurosurgical procedures. Neurosurg. 2013:138–45.
22.
Zurück zum Zitat Sappenfield JW, Smith WB, Cooper LA, et al. Visualization improves supraclavicular access to the subclavian vein in a mixed reality simulator. Anesth Analg. 2018;127(1):83–9.CrossRef Sappenfield JW, Smith WB, Cooper LA, et al. Visualization improves supraclavicular access to the subclavian vein in a mixed reality simulator. Anesth Analg. 2018;127(1):83–9.CrossRef
23.
Zurück zum Zitat Lee SC, Fuerst B, Tateno K, et al. Multi-modal imaging, model-based tracking, and mixed reality visualisation for orthopaedic surgery. Healthc Technol Lett. 2017;4(5):168–73.CrossRef Lee SC, Fuerst B, Tateno K, et al. Multi-modal imaging, model-based tracking, and mixed reality visualisation for orthopaedic surgery. Healthc Technol Lett. 2017;4(5):168–73.CrossRef
24.
Zurück zum Zitat Fischer M, Fuerst B, Lee SC, et al. Preclinical usability study of multiple augmented reality concepts for K-wire placement. Int J Comput Assist Radiol Surg. 2016;11(6):1007–14.CrossRef Fischer M, Fuerst B, Lee SC, et al. Preclinical usability study of multiple augmented reality concepts for K-wire placement. Int J Comput Assist Radiol Surg. 2016;11(6):1007–14.CrossRef
25.
Zurück zum Zitat Limthongkul W, Karaikovic EE, Savage JW, Markovic A. Volumetric analysis of thoracic and lumbar vertebral bodies. Spine J. 2010;10:153–8.CrossRef Limthongkul W, Karaikovic EE, Savage JW, Markovic A. Volumetric analysis of thoracic and lumbar vertebral bodies. Spine J. 2010;10:153–8.CrossRef
26.
Zurück zum Zitat Kawaguchi S, Horigome K, Yajima H, et al. Symptomatic relevance of intravertebral cleft in patients with osteoporotic vertebral fracture. J Neurosurg Spine. 2010;13:267–75.CrossRef Kawaguchi S, Horigome K, Yajima H, et al. Symptomatic relevance of intravertebral cleft in patients with osteoporotic vertebral fracture. J Neurosurg Spine. 2010;13:267–75.CrossRef
27.
Zurück zum Zitat Fang X, Yu F, Fu S, Song H. Intravertebral clefts in osteoporotic compression fractures of the spine: incidence, characteristics, and therapeutic efficacy. Int J Clin Exp Med. 2015;8(9):16960.PubMedPubMedCentral Fang X, Yu F, Fu S, Song H. Intravertebral clefts in osteoporotic compression fractures of the spine: incidence, characteristics, and therapeutic efficacy. Int J Clin Exp Med. 2015;8(9):16960.PubMedPubMedCentral
28.
Zurück zum Zitat Liang Z, Qiang W, Lin W, Jian S, Qiwei Z, Changtai S. Bone cement distribution in the vertebral body affects chances of recompression after percutaneous vertebroplasty treatment in elderly patients with osteoporotic vertebral compression fractures. Clin Interv Aging. 2017;12:431–6.CrossRef Liang Z, Qiang W, Lin W, Jian S, Qiwei Z, Changtai S. Bone cement distribution in the vertebral body affects chances of recompression after percutaneous vertebroplasty treatment in elderly patients with osteoporotic vertebral compression fractures. Clin Interv Aging. 2017;12:431–6.CrossRef
29.
Zurück zum Zitat Gaughen JR, Jensen ME, Schweickert PA, Marx WF, Kallmes DF. The therapeutic benefit of repeat percutaneous vertebroplasty at previously treated vertebral levels. AJNR Am J Neuroradiol. 2002;23(10):1657–61.PubMed Gaughen JR, Jensen ME, Schweickert PA, Marx WF, Kallmes DF. The therapeutic benefit of repeat percutaneous vertebroplasty at previously treated vertebral levels. AJNR Am J Neuroradiol. 2002;23(10):1657–61.PubMed
30.
Zurück zum Zitat Kim YJ, Chae SU, Kim GD, Park KH, Lee YS, Lee HY. Radiographic detection of osteoporotic vertebral fracture without collapse. J Bone Metab. 2013;20(2).CrossRef Kim YJ, Chae SU, Kim GD, Park KH, Lee YS, Lee HY. Radiographic detection of osteoporotic vertebral fracture without collapse. J Bone Metab. 2013;20(2).CrossRef
31.
Zurück zum Zitat Knavel EM, Rad AE, Thielen KR, Kallmes DF. Clinical outcomes with hemivertebral filling during percutaneous vertebroplasty. AJNR Am J Neuroradiol. 2009;30(3):496.CrossRef Knavel EM, Rad AE, Thielen KR, Kallmes DF. Clinical outcomes with hemivertebral filling during percutaneous vertebroplasty. AJNR Am J Neuroradiol. 2009;30(3):496.CrossRef
32.
Zurück zum Zitat Wu AM, Chi YL, Ni WF. Vertebral compression fracture with intravertebral vacuum cleft sign: pathogenesis, image, and surgical intervention. Asian Spine J. 2013;7(2):148–55.CrossRef Wu AM, Chi YL, Ni WF. Vertebral compression fracture with intravertebral vacuum cleft sign: pathogenesis, image, and surgical intervention. Asian Spine J. 2013;7(2):148–55.CrossRef
33.
Zurück zum Zitat Becker S, Tuschel A, Chavanne A, Meissner J, Ogon M. Balloon kyphoplasty for vertebra plana with or without osteonecrosis. J Orthop Surg. 2008;16(1):14–9.CrossRef Becker S, Tuschel A, Chavanne A, Meissner J, Ogon M. Balloon kyphoplasty for vertebra plana with or without osteonecrosis. J Orthop Surg. 2008;16(1):14–9.CrossRef
34.
Zurück zum Zitat Ishiyama M, Numaguchi Y, Makidono A, Kobayashi N, Fuwa S, Ohde S, et al. Contrast-enhanced MRI for detecting intravertebral cleft formation: relation to the time since onset of vertebral fracture. Am J Roentgenol. 2013;201(1):W117–23.CrossRef Ishiyama M, Numaguchi Y, Makidono A, Kobayashi N, Fuwa S, Ohde S, et al. Contrast-enhanced MRI for detecting intravertebral cleft formation: relation to the time since onset of vertebral fracture. Am J Roentgenol. 2013;201(1):W117–23.CrossRef
35.
Zurück zum Zitat Erbe EM, Clineff TD, Gualtieri G. Comparison of a new bisphenol-a-glycidyl dimethacrylate-based cortical bone void filler with polymethyl methacrylate. Eur Spine J. 2001;10:S147–52.CrossRef Erbe EM, Clineff TD, Gualtieri G. Comparison of a new bisphenol-a-glycidyl dimethacrylate-based cortical bone void filler with polymethyl methacrylate. Eur Spine J. 2001;10:S147–52.CrossRef
36.
Zurück zum Zitat Bae H, Hatten HP, Linovitz R, Tahernia AD, Schaufele MK, Mccollom V, et al. A prospective randomized FDA-IDE trial comparing Cortoss with PMMA for vertebroplasty: a comparative effectiveness research study with 24-month follow-up. Spine. 2012;37(7):544–50.CrossRef Bae H, Hatten HP, Linovitz R, Tahernia AD, Schaufele MK, Mccollom V, et al. A prospective randomized FDA-IDE trial comparing Cortoss with PMMA for vertebroplasty: a comparative effectiveness research study with 24-month follow-up. Spine. 2012;37(7):544–50.CrossRef
38.
Zurück zum Zitat Torres I, De Luccia N. A simulator for training in endovascular aneurysm repair: the use of three dimensional printers. Eur J Vasc Endovasc Surg. 2017;54(2):247–53.CrossRef Torres I, De Luccia N. A simulator for training in endovascular aneurysm repair: the use of three dimensional printers. Eur J Vasc Endovasc Surg. 2017;54(2):247–53.CrossRef
39.
Zurück zum Zitat Stefan P, Habert S, Winkler A, Lazarovici M, Fürmetz J, Eck U, Navab N. A radiation-free mixed-reality training environment and assessment concept for C-arm-based surgery. Int J Comput Assist Radiol Surg. 2018;13:1335–44.CrossRef Stefan P, Habert S, Winkler A, Lazarovici M, Fürmetz J, Eck U, Navab N. A radiation-free mixed-reality training environment and assessment concept for C-arm-based surgery. Int J Comput Assist Radiol Surg. 2018;13:1335–44.CrossRef
40.
Zurück zum Zitat Wu AM, Wang S, Weng WQ, Shao ZX, Yang XD, Wang JS, et al. The radiological feature of anterior occiput-to-axis screw fixation as it guides the screw trajectory on 3D printed models: a feasibility study on 3D images and 3D printed models. Medicine. 2014;93(28):e242.CrossRef Wu AM, Wang S, Weng WQ, Shao ZX, Yang XD, Wang JS, et al. The radiological feature of anterior occiput-to-axis screw fixation as it guides the screw trajectory on 3D printed models: a feasibility study on 3D images and 3D printed models. Medicine. 2014;93(28):e242.CrossRef
41.
Zurück zum Zitat Wu AM, Shao ZX, Wang JS, Yang XD, Weng WQ, Wang XY, Xu HZ, Chi YL, Lin ZK. The accuracy of a method for printing three-dimensional spinal models. PLoS ONE. 2015;10:e0124291.CrossRef Wu AM, Shao ZX, Wang JS, Yang XD, Weng WQ, Wang XY, Xu HZ, Chi YL, Lin ZK. The accuracy of a method for printing three-dimensional spinal models. PLoS ONE. 2015;10:e0124291.CrossRef
42.
Zurück zum Zitat Maruthappu M, Keogh B. (2014) How might 3D printing affect clinical practice?. BMJ, 349(dec30 1), g7709-g7709.CrossRef Maruthappu M, Keogh B. (2014) How might 3D printing affect clinical practice?. BMJ, 349(dec30 1), g7709-g7709.CrossRef
43.
Zurück zum Zitat Sugawara T, Higashiyama N, Kaneyama S, Sumi M. Accurate and simple screw insertion procedure with patient-specific screw guide templates for posterior C1-C2 fixation. Spine. 2016;42(6):1. Sugawara T, Higashiyama N, Kaneyama S, Sumi M. Accurate and simple screw insertion procedure with patient-specific screw guide templates for posterior C1-C2 fixation. Spine. 2016;42(6):1.
44.
Zurück zum Zitat Liu K, Zhang Q, Li X, Zhao C, Quan X, Zhao R, et al. (2016) Preliminary application of a multi-level 3D printing drill guide template for pedicle screw placement in severe and rigid scoliosis. European Spine Journal. Liu K, Zhang Q, Li X, Zhao C, Quan X, Zhao R, et al. (2016) Preliminary application of a multi-level 3D printing drill guide template for pedicle screw placement in severe and rigid scoliosis. European Spine Journal.
45.
Zurück zum Zitat Ghanai S, Marmulla R, Wiechnik J, et al. Computer-assisted three -dimensional surgical planning: 3D virtual articulator: technical note. Int J Oral Maxillofac Surg. 2010;39(1):75–82.CrossRef Ghanai S, Marmulla R, Wiechnik J, et al. Computer-assisted three -dimensional surgical planning: 3D virtual articulator: technical note. Int J Oral Maxillofac Surg. 2010;39(1):75–82.CrossRef
46.
Zurück zum Zitat Terander AE, Nachabe R, Skulason H, Pedersen K, Edström E. Feasibility and accuracy of thoracolumbar minimally invasive pedicle screw placement with augmented reality navigation technology. Spine. 2017;43(14):1. Terander AE, Nachabe R, Skulason H, Pedersen K, Edström E. Feasibility and accuracy of thoracolumbar minimally invasive pedicle screw placement with augmented reality navigation technology. Spine. 2017;43(14):1.
47.
Zurück zum Zitat Clark W, Bird P, Gonski P, Diamond TH, Smerdely P, McNeil HP, Schlaphoff G, Bryant C, Barnes E, Gebski V. Safety and efficacy of vertebroplasty for acute painful osteoporotic fractures (VAPOUR): a multicentre, randomised, double-blind, placebo-controlled trial. Lancet. 2016;388:1408–16.CrossRef Clark W, Bird P, Gonski P, Diamond TH, Smerdely P, McNeil HP, Schlaphoff G, Bryant C, Barnes E, Gebski V. Safety and efficacy of vertebroplasty for acute painful osteoporotic fractures (VAPOUR): a multicentre, randomised, double-blind, placebo-controlled trial. Lancet. 2016;388:1408–16.CrossRef
48.
Zurück zum Zitat Chen YJ, Chen HY, Lo DF, Chen HT, Hsu HC. Kirschner wire-guided technique for inserting a second needle into inadequately filled vertebrae in vertebroplasty: a technical report. Spine J. 2014;14:3025–9.CrossRef Chen YJ, Chen HY, Lo DF, Chen HT, Hsu HC. Kirschner wire-guided technique for inserting a second needle into inadequately filled vertebrae in vertebroplasty: a technical report. Spine J. 2014;14:3025–9.CrossRef
49.
Zurück zum Zitat Chen YJ, Chen WH, Chen HT, Hsu HC. Repeat needle insertion in vertebroplasty to prevent re-collapse of the treated vertebrae. Eur J Radiol. 2012;81:558–61.CrossRef Chen YJ, Chen WH, Chen HT, Hsu HC. Repeat needle insertion in vertebroplasty to prevent re-collapse of the treated vertebrae. Eur J Radiol. 2012;81:558–61.CrossRef
50.
Zurück zum Zitat Daisuke U, Yu Y, Yasuhiro N, et al. Balloon kyphoplasty under three-dimensional radiography guidance. Neurologia medico-chirurgica. 2017;57(9):489–95.CrossRef Daisuke U, Yu Y, Yasuhiro N, et al. Balloon kyphoplasty under three-dimensional radiography guidance. Neurologia medico-chirurgica. 2017;57(9):489–95.CrossRef
Metadaten
Titel
Percutaneous kyphoplasty assisted with/without mixed reality technology in treatment of OVCF with IVC: a prospective study
verfasst von
Peiran Wei
Qingqiang Yao
Yan Xu
Huikang Zhang
Yue Gu
Liming Wang
Publikationsdatum
01.12.2019
Verlag
BioMed Central
Erschienen in
Journal of Orthopaedic Surgery and Research / Ausgabe 1/2019
Elektronische ISSN: 1749-799X
DOI
https://doi.org/10.1186/s13018-019-1303-x

Weitere Artikel der Ausgabe 1/2019

Journal of Orthopaedic Surgery and Research 1/2019 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Mehr Frauen im OP – weniger postoperative Komplikationen

21.05.2024 Allgemeine Chirurgie Nachrichten

Ein Frauenanteil von mindestens einem Drittel im ärztlichen Op.-Team war in einer großen retrospektiven Studie aus Kanada mit einer signifikanten Reduktion der postoperativen Morbidität assoziiert.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

TEP mit Roboterhilfe führt nicht zu größerer Zufriedenheit

15.05.2024 Knie-TEP Nachrichten

Der Einsatz von Operationsrobotern für den Einbau von Totalendoprothesen des Kniegelenks hat die Präzision der Eingriffe erhöht. Für die postoperative Zufriedenheit der Patienten scheint das aber unerheblich zu sein, wie eine Studie zeigt.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.