Open Access 01.12.2007 | Research article

Phono-spectrographic analysis of heart murmur in children

verfasst von: Anna-Leena Noponen, Sakari Lukkarinen, Anna Angerla, Raimo Sepponen

Erschienen in: BMC Pediatrics | Ausgabe 1/2007

Abstract

Background

More than 90% of heart murmurs in children are innocent. Frequently the skills of the first examiner are not adequate to differentiate between innocent and pathological murmurs. Our goal was to evaluate the value of a simple and low-cost phonocardiographic recording and analysis system in determining the characteristic features of heart murmurs in children and in distinguishing innocent systolic murmurs from pathological.

Methods

The system consisting of an electronic stethoscope and a multimedia laptop computer was used for the recording, monitoring and analysis of auscultation findings. The recorded sounds were examined graphically and numerically using combined phono-spectrograms. The data consisted of heart sound recordings from 807 pediatric patients, including 88 normal cases without any murmur, 447 innocent murmurs and 272 pathological murmurs. The phono-spectrographic features of heart murmurs were examined visually and numerically. From this database, 50 innocent vibratory murmurs, 25 innocent ejection murmurs and 50 easily confusable, mildly pathological systolic murmurs were selected to test whether quantitative phono-spectrographic analysis could be used as an accurate screening tool for systolic heart murmurs in children.

Results

The phono-spectrograms of the most common innocent and pathological murmurs were presented as examples of the whole data set. Typically, innocent murmurs had lower frequencies (below 200 Hz) and a frequency spectrum with a more harmonic structure than pathological cases. Quantitative analysis revealed no significant differences in the duration of S1 and S2 or loudness of systolic murmurs between the pathological and physiological systolic murmurs. However, the pathological murmurs included both lower and higher frequencies than the physiological ones (p < 0.001 for both low and high frequency limits). If the systolic murmur contained intensive frequency components of over 200 Hz, or its length accounted for over 80 % of the whole systolic duration, it was considered pathological. Using these criteria, 90 % specificity and 91 % sensitivity in screening were achieved.

Conclusion

Phono-spectrographic analysis improves the accuracy of primary heart murmur evaluation and educates inexperienced listener. Using simple quantitative criterias a level of pediatric cardiologist is easily achieved in screening heart murmurs in children.
Hinweise

Electronic supplementary material

The online version of this article (doi:10.​1186/​1471-2431-7-23) contains supplementary material, which is available to authorized users.

Competing interests

The author(s) declare that they have no competing interests.

Authors' contributions

ALN carried out most of the clinical work, extracted the numerical data for the statistical analysis and wrote the clinical aspects of the manuscript. SL designed the recording system, performed the statistical analysis, wrote the technical description of the system, and reviewed the manuscript. AA participated in the system and study design and data acquisition. RS conceived of the study, and participated in its design and coordination and helped to draft the manuscript. All authors read and approved the final manuscript.

Background

Although Dr. Laennec's invention, the stethoscope, has been in clinical use for more than 180 years, and electronic stethoscopes with variable amplification gain have been available for over 80 years, it is still difficult to understand auscultation findings [13]. The phonocardiogram, first developed in 1894, visualizes auscultatory signals [2, 4, 5]. The spectral phonocardiogram has proven to be a reliable tool that gives information of whether or not the murmur is pathological. Based on earlier studies and clinical observations, it has been assumed that pathological murmurs involve sounds of higher frequency [2, 5]. Phonocardiography and electronic stethoscopy attempt to improve the diagnostic accuracy of cardiac auscultation. In the most recent studies, digital acoustic analysis has demonstrated the validity of these methods [611]. Since the 1980's, phonocardiographic research activity had decreased due to the improvements of echocardiography, which yields more visual information. During the past few years, however, the improvements of personal computers have made it possible to design new low-cost, high-quality phonocardiographic devices [1217]. Spectral phonocardiography emulates the ear and may be ideal for teaching clinical stethoscopy [4]. The phono-spectrogram combines traditional phonocardiogram with time-frequency distribution presentation of the signal. The spectrogram was introduced for heart sound analysis as early as 1955 by McKusik et al, but was afterwards almost forgotten [4, 11].
The first evaluation of children's heart murmur is one of the basic tasks of welfare clinic practitioners and school medical officers. However, based on their frequently limited auscultation experience, they might not be able to recognize the innocence of a heart murmur. Even the auscultation skills of pediatric residents have been found to be suboptimal [1821]. Several healthy children are referred to pediatric cardiologists or for echocardiography. Parental anxiety may also be a reason to refer the patient for unnecessary examinations, and it is important to recognize an innocent murmur as soon as it is found [2225]. Auscultation training will naturally improve practitioners' listening skills [2628].
Previous studies have shown that pediatric cardiologists, based on clinical examination, can differentiate innocent from pathological murmurs with high sensitivity (82...92 %) and specificity (76...99%) [2933]. Even better results have been attained by using advanced signal processing and pattern recognition tools. Artificial neural network-based screening is reported to have 100% sensitivity and specificity [13, 14].
The capability of a doctor or a computerized system to differentiate between pathological and innocent murmurs depends on the quality of the pathological murmurs. It is easy to recognize loud murmurs as pathological by means of clinical or digital analysis [9]. Serious defects are seldom missed, and the challenge is actually to detect mild defects. Small muscular VSD, mild PS or AS causes no hemodynamic harm, but they may require endocarditis prophylaxis. ASD secundum may also be hard to diagnose only by auscultation [3438].
The purpose of this article was to demonstrate the capabilities of a modern digital system for phonocardiographic recording and analysis and to evaluate its potential for differentiating the characteristic features of heart murmurs in children. Another goal was to investigate how the numerical features measured from a phono-spectrogram can be used to distinguish innocent systolic murmurs from pathological murmurs, which are generally hard to identify in a routine clinical examination. This article additionally concludes more than ten years' experience of studying heart sounds.

Methods

Data

Heart sounds were collected at the outpatient pediatric cardiology clinics of the rural hospitals in the vicinity of Helsinki from 1995 to 1999. The Ethics Committees of Jorvi (4/95 number 17) and rural hospitals approved the study. The study series consisted of 807 children, whose ages ranged from newborn to 16 years. The recordings were done at random times, i.e. without pre-set dates. The patient roster was not available in advance and every patient with a positive attitude towards the study were included. From each patient, the auscultation finding at the point of maximal intensity was recorded. The patients were either children with murmurs referred for a cardiologic evaluation for the first time or patients with a known diagnosis. In all cases, the diagnosis was confirmed with echocardiography. The control series consisted of children without heart murmur, including children with paroxysmal supraventricular tachyarrhythmia, spontaneously closing ventricular septal defect or ductus arteriosus. Healthy child volunteers were also recorded and included in the control series. Table 1 summarizes the diagnostic findings.
Table 1
Recorded heart sounds
Diagnosis
Number
%
Vibratory murmur
310
38.4
Ejection murmur
107
13.3
Venous hum
26
3.2
Other innocent musical murmurs
4
0.5
Ventricular septal defect (VSD)
87
10.8
Pulmonary stenosis (PS)
47
5.8
Aortic stenosis (AS) or coarctation (CoA)
43
5.3
Mitral valve defect with (MI) or without leakage
29
3.6
Patent ductus arteriosus (PDA)
19
2.4
Atrial septal defect (ASD)
17
2.1
Other
30
3.7
No murmur (control)
88
10.9
Total
807
100

Equipment

A phonocardiographic system developed in Helsinki University of Technology was used to record heart sounds. The system consisted of an electronic stethoscope and a multimedia laptop computer. A parabolic-shaped cup combined with a high-quality electric microphone and variable-gain battery-powered amplifier in the electronic stethoscope gave a flat frequency response in the whole frequency range of usual heart sounds and murmurs (from 75 to 1500 Hz). The laptop computer had standard multimedia capabilities for audio input and output (16-bit resolution in amplitude, variable recording gain and sampling frequencies from 8 kHz up to 44.1 kHz). Sound signals were recorded digitally using special software for the monitoring and analysis of auscultation. The software was specifically written for this project, and it had several end user (general practitioner) friendly capabilities, such as a database for recordings, basic patient information input dialog, real-time graphical monitoring during recording, selectable and changeable digital filters, tunable settings for the graphical display and analysis, free zooming and replaying of the interesting parts and tools for measuring intensity, duration and frequency range. The software was compatible with standard Microsoft Windows (NT, 2000, XP) environments.
The examiner heard the sounds through earphones and monitored the signals on the computer screen. Thus, the recording was completed in almost the same time as traditional auscultation. Some additional effort was needed to enter the patient information. The new generation commercial electronic stethoscopes were also compatible with this recording program. During the recording it was possible to both to listen to the sound and to follow the phonospectrogram from the display screen, the process was completed often in less than ten minutes.

Methods

In the post-analysis phase, the recorded sounds were re-played, and the fingerprints of innocent and pathological heart murmurs were simultaneously examined visually by using a phono-spectrogram. First the recorded sounds were digitally filtered using pass-band filtering (75–1500 Hz) (the 3rd order Butterworth type high-pass and low-pass filters). These filter settings were selected by a long term subjective trials where the objective was to trim the display to show the details of the murmurs as an experienced practitioner would understand them. For the traditional waveform display the signal was scaled by the signal's absolute maximum. The resulted waveform showed the relative intensities of the heart sounds as the ear recognizes them and thus absolute intensity scaling is not needed. Similarly in the spectrogram the intensities were scaled by finding the maximum intensity on the time-frequency distribution and calculating relative intensities in the decibel scale. Thus the value of 0 dB corresponded to the maximum intensity and value of -60 dB, which is almost unbearable to listen to. Short time Fast Fourier Transform (STFT) with Hanning windowing, 512 data samples (46 ms time resolution) and total of 1024 FFT points (10.7 Hz frequency resolution) were used for calculating the spectrogram.
The three most common innocent murmurs, i.e. vibratory murmur, pulmonary ejection murmur and venous hum, and the five most common congenital heart defects, i.e. ventricular septal defect (VSD), aortic valve stenosis (AS), pulmonary valve stenosis (PS), patent ductus arteriosus (PDA) and atrial septal defect (ASD), are presented as examples of the whole data set.
To differentiate between innocent and pathological murmurs, 50 vibratory innocent heart murmurs, 25 innocent pulmonary or aortic ejection murmurs and 50 mild pathological systolic murmurs with intensity equal to or less than grade 3/6 were selected from a larger heart sound database for more detailed analysis. 14 of the selected pathological cases were small VSD, 5 hemodynamically loading ASD secundum, 8 PS with a pressure gradient less than 30 mmHg, 8 AS with a pressure gradient less than 30 mmHg, 3 bicuspid aortic valves with velocity less than 2.0 m/s, 1 mild CoA with a measured 15 mmHg RR difference, 1 hypertrophic cardiomyopathy (HOCM) with a septum thickness of 12 mm but with laminar aortic flow, 5 mitral leakage with or without prolapse (MI) and 3 tricuspid valve leakage (TI).

Parameters

The timings, the intensities and the frequency contents of the sounds were manually extracted from the phono-spectrogram using the software's graphical measurement tools (see any of the Figures from 1, 2, 3, 4, 5, 6, 7, 8). First, the whole recording is listened to carefully. The locations of S1 and S2 are recognized from the graphical presentation showing a moving marker over the phono-spectrogram during replaying. Then, a shorter interesting area containing three whole heart cycles is selected by zooming the signal. The timings are read from the time-scale (at the bottom) by graphically studying both the phonocardiogram and the spectrogram, deciding the duration of the sound events and dragging the marker over the selected time segment. The intensities of S1, S2 and the murmur are read from the scale (on the right) by moving the marker with the mouse over the phonocardiogram. The loudness of the murmur was estimated from the phonocardiogram by comparing the maximum amplitude of the murmur to the average value of the maximum first and second heart sounds Finally, the frequencies in Hertz are read by moving the marker over the spectrogram. The typical intensity for the high and low frequency limits of the systolic murmurs was around -45–50 dB.

Statistics

To find statistically significant parameters, two-tailed heteroscedastic t-test for independent samples was used to calculate the p-values. Three sequential beats were manually selected from each recording. The mean value and standard deviation of the parameters over three beats were calculated. Based on the statistical results, the relative duration of the systolic murmur (percentage of murmur duration out of the interval between the end of S1 and the beginning of S2) and the occurrence of intensive high-frequency components were used as criteria for testing the pathology of the murmur.

Results

Illustrative examples of most common heart murmurs

The system was able to display and reproduce cardiovascular sound events. The musical murmur caused by harmonic movements of the heart or the vasculature, of which vibratory innocent murmur is a good example, was usually visualized as a well-defined area or line in the spectrogram. Innocent systolic murmur appeared to have a lower peak frequency, below 200 Hz, and shorter duration than pathological murmurs, and it always faded before the second heart sound. The higher the velocity of flow in echogardiography, the more intensive the murmur and the wider the frequency scale. This phenomenon was clearly visible in the spectrogram.
The attached illustrations (Figures from 1, 2, 3, 4, 5, 6, 7, 8) show examples of the most typical murmurs in children. The upper part is a traditional phonocardiogram showing the relative amplitude of the sound. The lower part, i.e. the spectrogram, shows the sound intensity as colors on a frequency scale of 0–1000 Hz. The corresponding intensity scale in decibels is shown on the left. The duration, peak frequency and volume of the murmur were estimated based on a mean of three sequential beats. Volume was compared to the mean of the amplitudes of S1 and S2, obtained from the traditional phonocardiogram.

Quantitative comparison of innocent and confusing pathological systolic murmurs

The mean and standard deviation of the patients' ages for the three groups are presented in Table 2. The patients with innocent vibratory murmurs were referred for examination at a younger age than those with ejection murmurs, which seemed confusing to the examiners and were referred closer to school age. The pathological murmurs in this series included recordings of both new patients and repeated control recordings of outpatients.
Table 2
Mean and standard deviation of patients' ages in the three groups
Group
Age
Vibratory
4.8 ± 3.3
Ejection
7.3 ± 5.0
Pathological
6.2 ± 4.9
There were no significant differences in the duration of S1 and S2 between the pathological and physiological systolic murmurs (Table 3). Although S2 was constantly split in ASDs, the duration of S2 in these cases did not differ from the average in the series. By contrast, the pathological murmurs, measured in either absolute or relative terms, were longer than the innocent vibratory or ejection systolic murmurs (p < 0.001).
Table 3
Duration of S1, S2, systolic murmur and relative duration of systolic murmur
Group
Duration
 
S1 (ms)
S2 (ms)
Murmur (ms)
Murmur (%)
Vibratory
81 ± 15
69 ± 12
100 ± 21
53 ± 12
Ejection
92 ± 16
75 ± 14
84 ± 36
44 ± 17
Pathological
87 ± 19
77 ± 17
147 ± 49
78 ± 21
p-value
0.47
0.029
< 0.001
< 0.001
Although the pathological systolic murmurs were somewhat louder than the physiological ones, there were no significant differences between the groups (Relative Amplitude in Table 4). This confirmed that the series fulfilled our aim to select easily confusable cases. However, the pathological murmurs included both lower and higher frequencies than physiological ones (p < 0.001 for both low and high frequency limits).
Table 4
Relative amplitude and low and high frequency limits of systolic murmur
Group
Amplitude (%)
Low freq limit (Hz)
High freq limit (Hz)
Vibratory
23 ± 9
72 ± 15
161 ± 22
Ejection
20 ± 9
60 ± 9
142 ± 51
Pathological
30 ± 20
52 ± 19
299 ± 133
p-value
0.013
< 0.001
< 0.001
Sensitivity of 76 % and specificity of 84 % were achieved using a cut-off value of 65 % for the relative duration of systolic murmur to differentiate between pathological and physiological murmurs (Table 5 and Figure 9 – upper left). Using that cut-off, 12 (24%) pathological and 12 (16 %) physiological murmurs were misclassified.
Table 5
Capability of relative length to distinguish between murmurs: sensitivity is 76 % and specificity 84 %
 
Pathological
Physiological
Total
> = 65 %
38
12
24
< 65 %
12
63
101
Total
50
75
125
When the frequency of 190 Hz for the high frequency limit was used as a classification criterion, both sensitivity and specificity increased to 88% (Table 6 and Figure 9 – lower left). The number of misclassified cases decreased to 6 (12%) in the group of pathological murmurs and to 9 (12%) in the group of physiological murmurs.
Table 6
Capability of high frequency limit to distinguish between murmurs: sensitivity is 88 % and specificity 88 %
 
Pathological
Physiological
Total
> = 190 Hz
44
9
53
< 190 Hz
6
66
72
Total
50
75
125
By combining these two criteria, optimal results were achieved at 80% of the relative duration and with a frequency of 200 Hz as the high frequency limit. Sensitivity approached 90 % and specificity 91 % (Table 7 and Figure 10). There were 5 (10 %) false negative findings and 7 (9 %) false positives.
Table 7
Combinated criteria: sensitivity 90 %, specificity 91 %
 
Pathological
Physiological
Total
> = 80 % OR > = 200 Hz
45
7
52
All other
5
68
73
Total
50
75
125

Discussion

Over 90 % of heart murmurs in children are physiological. Moreover, 75 % of them are innocent vibratory murmurs (39, 40, 41, 42). The physician needs to quickly confirm the benign nature of the heart murmur and thus to avoid misdiagnosis [25, 43]. The combination of a spectrogram and a traditional phonocardiogram can be an adequate method for distinguishing innocent murmurs from pathological. When quantitative analysis is used systematically, the clinician should know what signals the system is processing and how. Visual analysis is the first step towards understanding automatic analysis, and it is a more reliable way to classify the findings as pathological or non-pathological than auscultation alone. Interpretation of the spectrogram helps to understand the hemodynamic events and the origin of heart sounds. Laminar flow will cause a harmonic wave movement in surrounding tissues, as exemplified by vibratory murmur with a peak frequency of approximately 150 Hz, which has also been reported in earlier studies [39, 44, 45]. Rapid flow will cause turbulence. The faster the flow velocity, the higher the sound frequencies [4, 5]. These findings are also illustrated by the analysis of aortic stenosis or ventricular septal defect [610].
It is not simple to estimate absolute volume of heart sound. The point of maximal intensity of the murmur and the thickness of the chest vary and affects on the intensities of the sound components. In our study, the sound volume of murmur was calculated as a percentage of the volume of murmur compared to the mean volumes of S1 and S2. This may be misleading, as we can see in case 4. The volume of the aortic closure sound is decreased because of valvular stenosis, and the volume of murmur is over estimated compared to auscultation findings.
By analyzing the combined phono-spectrogram, it is possible to achieve the level of an experienced pediatric cardiologist in screening heart murmurs. In this study, by using special criteria, sensitivity of 90 % and specificity of 91% were attained. Both sensitivity and specificity were higher than using phonocradiographic or spectral analysis alone. The results were less good than those obtained in previous studies using the methods of advanced signal processing, pattern recognition and artificial networks [10, 1315]. In these previous studies, however, the series consisted of pathological cases with distinct and easily recognizable features of heart sounds and murmurs. We agree that it is easy to distinguish fast turbulent flow with a pressure gradient of over 25 mmHg, and in such cases phono-spectral analysis is reliable. We here examined the grey area, i.e. typical physiological cases compared to mildly pathological murmurs, and the most common pitfalls.
Typically, the screening method is allowed to include some false positive cases, but false negatives are unacceptable. In this study, ten (10) small muscular ventricular septal defects, three (3) cases of mild aortic stenosis, one (1) mild peripheral pulmonary stenosis, one (1) mitral valve prolapse with leakage and one tricuspid leakage fell below the 80 % duration criterion. However, each of these cases exceeded the 200 Hz frequency limit. Three (3) atrial septal defects and one (1) slight pulmonary valve stenosis included frequencies below 200 Hz, but each of them was sustained for over 80% of the systolic duration. Five (5) of the pathological murmurs were not caught with either of these two criteria. One (1) of them was a clinically insignificant tricuspid valve leakage, one (1) a very mild mitral valve leakage, one (1) a bicuspic aortic valve without stenosis and one (1) a bicuspic aortic valve with slight stenosis (pressure gradient 16 mmHg). The last and the worst missed pathological murmur was recorded from a three-year-old boy with hypertrophic cardiomyopathy (thickness of ventricular septum 12 mm). Seven out of 75 innocent murmurs exceeded one of the defined screening criteria. None of the ejection murmurs exceeded the 80 % duration limit, and 3 cases exceeded the frequency limit. All of these cases were aortic ejection murmurs with aortic velocity of 1.5–1.8 m/s. Of the vibratory murmurs, one (1) continued for over 80 % of the systolic duration, three (3) contained frequency components of over 200 Hz, but none of them exceeded both limits.
Slight tricuspid leakage is a harmless finding. Mitral valve leakage and bicuspic aortic valve should be recognized because they require endocarditis prophylaxis. The patient with mitral leakage was a fearful retarded boy, and the recorded signal was weak in amplitude, which is why the highest frequencies were missed in the analysis, but the murmur was sustained until the beginning of the second heart sound, implying a pathological finding. The murmurs due to a bicuspic aortic valve without stenosis or with slight stenosis did not differ from the hyperkinetic aortic ejection murmur. The murmur of the patient with hypertrophic cardiomyopathy resembled vibratory innocent murmur [39]. The difference between these murmurs was hardly seen in the spectrogram. Typically, the frequency of vibratory innocent murmur decreases within the systolic interval, while in this case of hypertropic cardiomyopathy it remained constant. The ECG finding was pathological, and the diagnosis was known beforehand. This particular murmur was rather loud, 31% in spectral analysis and 3/6 estimated clinically. However, 10 (20%) of vibratory murmurs were equal to or louder than 30% measured from phonogram.
As the examples show, the method did not recognize hypertrophic cardiomyopathy or mild aortic valve malformation and failed with the mitral leakage, because the recording was suboptimal. We recommend an echocardiography, when the vibratory or aortic ejection murmur is repeatedly found in adolescent age or when the recording is suboptimal. Naturally, young babies need always an echocardiography, because their pulmonary pressure may still be high and the results of heart sound analysis can be misleading.
Modern electronic stethoscopes are suitable for heart sound recording. Visual presentation of the auscultation finding provides an opportunity to study more objectively and quantitatively the timing, quality (frequency contents) and intensity of different heart sound and murmur components and helps us to understand and learn about the hemodynamics of the heart and the origin of cardiac murmurs.
Computer analysis is a powerful tool in teaching an unexperienced listener the art of auscultation [4, 11, 46]. Furthermore, we believe that computer analysis can be a promising new diagnostic method that has a potential to reduce medical costs by eliminating unnecessary referrals, visits to specialists and echocardiograms, especially when the systolic murmur is weak. Recording of auscultation findings also makes teleconsultations possible [4749] and gives an opportunity to save cardiac or respiratory sounds, phonograms and spectrograms in electronic patient documents.

Conclusion

Phono-spectrographic analysis of heart murmur in children is a useful additional method for the clinical approach and an interesting educational tool as well. The 90 % sensitivity level of the tool may well be at the level of a trained paediatrician. However, it should be considered a single, suboptimal screening tool. As the case of HOCM indicates, ECG and echocardiography are advisable if the intensity of the murmur is fairly high and frequently audible. Patient history and careful physical examination remain the cornerstones in clinical practice.
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://​creativecommons.​org/​licenses/​by/​2.​0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Competing interests

The author(s) declare that they have no competing interests.

Authors' contributions

ALN carried out most of the clinical work, extracted the numerical data for the statistical analysis and wrote the clinical aspects of the manuscript. SL designed the recording system, performed the statistical analysis, wrote the technical description of the system, and reviewed the manuscript. AA participated in the system and study design and data acquisition. RS conceived of the study, and participated in its design and coordination and helped to draft the manuscript. All authors read and approved the final manuscript.
Literatur
1.
Zurück zum Zitat Rappaport MB, Sprague HB: Physiologic and physical laws that govern auscultation, and their clinical application; the acoustic stethoscope and the electrical amplifying stethoscope and stethograph. Am Heart J. 1941, 21: 257-318. 10.1016/S0002-8703(41)90904-3.CrossRef Rappaport MB, Sprague HB: Physiologic and physical laws that govern auscultation, and their clinical application; the acoustic stethoscope and the electrical amplifying stethoscope and stethograph. Am Heart J. 1941, 21: 257-318. 10.1016/S0002-8703(41)90904-3.CrossRef
2.
Zurück zum Zitat Selig MB: Stethoscope and phonoaudio devices: Historical and future perspectives. Am Heart J. 1993, 126: 262-268.CrossRefPubMed Selig MB: Stethoscope and phonoaudio devices: Historical and future perspectives. Am Heart J. 1993, 126: 262-268.CrossRefPubMed
3.
Zurück zum Zitat Tavel ME: Cardiac auscultation. A glorious past–but does it have a future?. Circulation. 1996, 93: 1250-1253.CrossRefPubMed Tavel ME: Cardiac auscultation. A glorious past–but does it have a future?. Circulation. 1996, 93: 1250-1253.CrossRefPubMed
4.
Zurück zum Zitat McKusick VA, Webb GN, O'Neal Humphries J, Reid JA: On cardiovascular sound: further observations by means of spectral phonocardiography. Circulation. 1955, 11: 849-870.CrossRefPubMed McKusick VA, Webb GN, O'Neal Humphries J, Reid JA: On cardiovascular sound: further observations by means of spectral phonocardiography. Circulation. 1955, 11: 849-870.CrossRefPubMed
5.
Zurück zum Zitat Leatham A: Auscultation of the Heart and Phonocardiogrphy. J & A Churchill London. 1970 Leatham A: Auscultation of the Heart and Phonocardiogrphy. J & A Churchill London. 1970
6.
Zurück zum Zitat Tavel ME, Brown DD, Shander D: Enhanced auscultation with a new graphic display system. Arch Intern Med. 1994, 154: 893-898. 10.1001/archinte.154.8.893.CrossRefPubMed Tavel ME, Brown DD, Shander D: Enhanced auscultation with a new graphic display system. Arch Intern Med. 1994, 154: 893-898. 10.1001/archinte.154.8.893.CrossRefPubMed
7.
Zurück zum Zitat Balster DA, Chan DP, Rowland DG, Hugh DA: Digital acoustic analysis of precordial innocent versus ventricular septal defect murmurs in children. Am J Cardiol. 1997, 79: 1552-1555. 10.1016/S0002-9149(97)00194-X.CrossRefPubMed Balster DA, Chan DP, Rowland DG, Hugh DA: Digital acoustic analysis of precordial innocent versus ventricular septal defect murmurs in children. Am J Cardiol. 1997, 79: 1552-1555. 10.1016/S0002-9149(97)00194-X.CrossRefPubMed
8.
Zurück zum Zitat Kim D, Tavel ME: Assessment of severity of aortic stenosis through timefrequency analysis of murmur. Chest. 2003, 124: 1638-1644. 10.1378/chest.124.5.1638.CrossRefPubMed Kim D, Tavel ME: Assessment of severity of aortic stenosis through timefrequency analysis of murmur. Chest. 2003, 124: 1638-1644. 10.1378/chest.124.5.1638.CrossRefPubMed
9.
Zurück zum Zitat Tavel ME, Katz H: Usefulness of a new sound spectral avaraging technique to distinguish an innocent systolic murmur from that of aortic stenosis. Am J Cardiol. 2005, 95: 902-904. 10.1016/j.amjcard.2004.12.018.CrossRefPubMed Tavel ME, Katz H: Usefulness of a new sound spectral avaraging technique to distinguish an innocent systolic murmur from that of aortic stenosis. Am J Cardiol. 2005, 95: 902-904. 10.1016/j.amjcard.2004.12.018.CrossRefPubMed
10.
Zurück zum Zitat Voss A, Mix A, Hubner T: Diagnosing aortic valve stenosis by parameter extraction of heart sound signals. Ann Biomed Eng. 2005, 33: 1167-1174. 10.1007/s10439-005-5347-x.CrossRefPubMed Voss A, Mix A, Hubner T: Diagnosing aortic valve stenosis by parameter extraction of heart sound signals. Ann Biomed Eng. 2005, 33: 1167-1174. 10.1007/s10439-005-5347-x.CrossRefPubMed
11.
Zurück zum Zitat Tavel ME: Cardiac auscultation: a glorious past–and it does have a future!. Circulation. 2006, 113: 1255-1259. 10.1161/CIRCULATIONAHA.105.591149.CrossRefPubMed Tavel ME: Cardiac auscultation: a glorious past–and it does have a future!. Circulation. 2006, 113: 1255-1259. 10.1161/CIRCULATIONAHA.105.591149.CrossRefPubMed
12.
Zurück zum Zitat Lukkarinen S, Noponen AL, Sikiö K, Angerla A: A new phonocardiographic recording system. Comput Cardiol. 1997, 24: 117-120. Lukkarinen S, Noponen AL, Sikiö K, Angerla A: A new phonocardiographic recording system. Comput Cardiol. 1997, 24: 117-120.
13.
Zurück zum Zitat Thompson WR, Hayek CS, Tuchinda C, Telford JK, Lombardo JS: Automated cardiac auscultation for detection of pathologic heart murmurs. Pediatr Cardiol. 2001, 22: 373-379.CrossRefPubMed Thompson WR, Hayek CS, Tuchinda C, Telford JK, Lombardo JS: Automated cardiac auscultation for detection of pathologic heart murmurs. Pediatr Cardiol. 2001, 22: 373-379.CrossRefPubMed
14.
Zurück zum Zitat DeGroff CG, Bhatikar S, Hertzberg J, Shandas R, Valdes-Cruz L, Mahajan RL: Artificial neural network-based method of screening heart murmurs in children. Circulation. 2001, 103: 2711-2716.CrossRefPubMed DeGroff CG, Bhatikar S, Hertzberg J, Shandas R, Valdes-Cruz L, Mahajan RL: Artificial neural network-based method of screening heart murmurs in children. Circulation. 2001, 103: 2711-2716.CrossRefPubMed
15.
Zurück zum Zitat Bhatikar SR, Degroff C, Mahajan RL: A classifier based on the artificial network approach for cardiologic auscultation in pediatrics. Artif Intell Med. 2005, 33: 251-260. 10.1016/j.artmed.2004.07.008.CrossRefPubMed Bhatikar SR, Degroff C, Mahajan RL: A classifier based on the artificial network approach for cardiologic auscultation in pediatrics. Artif Intell Med. 2005, 33: 251-260. 10.1016/j.artmed.2004.07.008.CrossRefPubMed
16.
Zurück zum Zitat El-Segaier M, Lilja O, Lukkarinen S, Sornmo L, Sepponen R, Pesonen E: Computer-based detection and analysis of heart sound and murmur. Ann Biomed Eng. 2005, 33 (7): 937-942. 10.1007/s10439-005-4053-3.CrossRefPubMed El-Segaier M, Lilja O, Lukkarinen S, Sornmo L, Sepponen R, Pesonen E: Computer-based detection and analysis of heart sound and murmur. Ann Biomed Eng. 2005, 33 (7): 937-942. 10.1007/s10439-005-4053-3.CrossRefPubMed
17.
Zurück zum Zitat Andresevic N, Ejaz K, Rios-Gutierrez F, Alba-Flores R, Nordehn G, Burns S: Detection of heart murmurs using wavelet analysis and artificial neural networks. J Biomech Eng. 2005, 127: 899-904. 10.1115/1.2049327.CrossRef Andresevic N, Ejaz K, Rios-Gutierrez F, Alba-Flores R, Nordehn G, Burns S: Detection of heart murmurs using wavelet analysis and artificial neural networks. J Biomech Eng. 2005, 127: 899-904. 10.1115/1.2049327.CrossRef
18.
Zurück zum Zitat Haney I, Ipp M, Feldmen W, McCrindle BW: Accuracy of clinical assessment of heart murmurs by office-based (general practice) paediatricians. Arch Dis Child. 1999, 81: 409-412.CrossRefPubMedPubMedCentral Haney I, Ipp M, Feldmen W, McCrindle BW: Accuracy of clinical assessment of heart murmurs by office-based (general practice) paediatricians. Arch Dis Child. 1999, 81: 409-412.CrossRefPubMedPubMedCentral
19.
Zurück zum Zitat Gaskin PRA, Owens SE, Talner NS, Sanders SP, Li JS: Clinical auscultation skills in pediatric residents. Pediatrics. 2000, 105: 1184-1187. 10.1542/peds.105.6.1184.CrossRefPubMed Gaskin PRA, Owens SE, Talner NS, Sanders SP, Li JS: Clinical auscultation skills in pediatric residents. Pediatrics. 2000, 105: 1184-1187. 10.1542/peds.105.6.1184.CrossRefPubMed
20.
Zurück zum Zitat Mahnke CB, Nowalk A, Hofkosh D, Zuberbuhler JR, Law YM: Comparison of two educational interventions on pediatric resident auskultation skills. Pediatrics. 2004, 113: 1331-1335. 10.1542/peds.113.5.1331.CrossRefPubMed Mahnke CB, Nowalk A, Hofkosh D, Zuberbuhler JR, Law YM: Comparison of two educational interventions on pediatric resident auskultation skills. Pediatrics. 2004, 113: 1331-1335. 10.1542/peds.113.5.1331.CrossRefPubMed
21.
Zurück zum Zitat Vukanovic-Criley JM, Criley S, Warde CM, Boker JR, Guevara-Matheus L, Churchill WH, Nelson WP, Criley JM: Competency in cardiac examination skills in medical students, trainees, physicians, and faculty: a multicenter study. Arch Intern Med. 2006, 166: 610-616. 10.1001/archinte.166.6.610.CrossRefPubMed Vukanovic-Criley JM, Criley S, Warde CM, Boker JR, Guevara-Matheus L, Churchill WH, Nelson WP, Criley JM: Competency in cardiac examination skills in medical students, trainees, physicians, and faculty: a multicenter study. Arch Intern Med. 2006, 166: 610-616. 10.1001/archinte.166.6.610.CrossRefPubMed
22.
Zurück zum Zitat Danford DA, Nasir A, Gumbiner C: Cost assessment of the evaluation of heart murmurs in children. Pediatrics. 1993, 91: 365-368.PubMed Danford DA, Nasir A, Gumbiner C: Cost assessment of the evaluation of heart murmurs in children. Pediatrics. 1993, 91: 365-368.PubMed
23.
Zurück zum Zitat McCrindle BW, Shaffer KM, Kan JS, Zahka KG, Rowe SA, Kidd I: An evaluation of parental concerns and misperceptions about heart murmurs. Clin Pediatr. 1995, 34: 25-31.CrossRef McCrindle BW, Shaffer KM, Kan JS, Zahka KG, Rowe SA, Kidd I: An evaluation of parental concerns and misperceptions about heart murmurs. Clin Pediatr. 1995, 34: 25-31.CrossRef
24.
Zurück zum Zitat Danford DA, Martin AB, Fletcher SE, Cumbiner CH: Echocardiographic yield in children when innocent murmur seems likely but doupts linger. Pediatr Cardiol. 2002, 23: 410-414. 10.1007/s00246-002-1390-y.CrossRefPubMed Danford DA, Martin AB, Fletcher SE, Cumbiner CH: Echocardiographic yield in children when innocent murmur seems likely but doupts linger. Pediatr Cardiol. 2002, 23: 410-414. 10.1007/s00246-002-1390-y.CrossRefPubMed
25.
Zurück zum Zitat Geggel RL, Horowitz LM, Brown EA, Parsons M, Wang PS, Fulton DR: Parental anxiety associated with referral of a child to a pediatric cardiologist for evaluation of a Still's murmur. J Pediatr. 2002, 140: 747-752. 10.1067/mpd.2002.124379.CrossRefPubMed Geggel RL, Horowitz LM, Brown EA, Parsons M, Wang PS, Fulton DR: Parental anxiety associated with referral of a child to a pediatric cardiologist for evaluation of a Still's murmur. J Pediatr. 2002, 140: 747-752. 10.1067/mpd.2002.124379.CrossRefPubMed
26.
Zurück zum Zitat Norgaard G, Greve G, Rosland GA, Berg A: Referral practice and clinical assessment of heart murmurs in children [in Norwegian]. Tidsskr Nor Laegeforen. 2005, 125: 996-998. Norgaard G, Greve G, Rosland GA, Berg A: Referral practice and clinical assessment of heart murmurs in children [in Norwegian]. Tidsskr Nor Laegeforen. 2005, 125: 996-998.
27.
Zurück zum Zitat March SK, Bedynek JL, Chizner MA: Teatching cardiac Auscultation: effectiveness of a patient-centrered teatching conference on improving cardiac auscultatory skills. Mayo Clin Proc. 2005, 80: 1443-1448.CrossRefPubMed March SK, Bedynek JL, Chizner MA: Teatching cardiac Auscultation: effectiveness of a patient-centrered teatching conference on improving cardiac auscultatory skills. Mayo Clin Proc. 2005, 80: 1443-1448.CrossRefPubMed
28.
Zurück zum Zitat Barrett MJ, Lacey CS, Sekara AE, Linden EA, Gracely EJ: Mastering cardiac murmurs: the power of repetition. Chest. 2004, 126: 470-475. 10.1378/chest.126.2.470.CrossRefPubMed Barrett MJ, Lacey CS, Sekara AE, Linden EA, Gracely EJ: Mastering cardiac murmurs: the power of repetition. Chest. 2004, 126: 470-475. 10.1378/chest.126.2.470.CrossRefPubMed
29.
Zurück zum Zitat Van Oort A, Le Blanc-Botden M, De Boo T, Van Der Werf T, Daniels O: The vibratory innocent heart murmur in schoolchildren: difference in auscultatory findings between school medical officers and a pediatric cardiologist. Pediatr Cardiol. 1994, 15: 282-287. 10.1007/BF00798121.CrossRefPubMed Van Oort A, Le Blanc-Botden M, De Boo T, Van Der Werf T, Daniels O: The vibratory innocent heart murmur in schoolchildren: difference in auscultatory findings between school medical officers and a pediatric cardiologist. Pediatr Cardiol. 1994, 15: 282-287. 10.1007/BF00798121.CrossRefPubMed
30.
Zurück zum Zitat McCrindle BW, Shaffer KM, Kan JS, Zahka KG, Rowe SA, Kidd L: Cardinal clinical signs in the differentiation of heart murmurs in children. Arch Pediatr Adolesc Med. 1996, 150: 169-174.CrossRefPubMed McCrindle BW, Shaffer KM, Kan JS, Zahka KG, Rowe SA, Kidd L: Cardinal clinical signs in the differentiation of heart murmurs in children. Arch Pediatr Adolesc Med. 1996, 150: 169-174.CrossRefPubMed
31.
Zurück zum Zitat Rajakumar K, Weisse M, Rosas A, Gunel E, Pyles L, Neal WA, Balian A, Einzig S: Comparative study of clinical evaluation of heart murmurs by general pediatricians and pediatric cardiologists. Clin Pediatr. 1999, 38: 511-518.CrossRef Rajakumar K, Weisse M, Rosas A, Gunel E, Pyles L, Neal WA, Balian A, Einzig S: Comparative study of clinical evaluation of heart murmurs by general pediatricians and pediatric cardiologists. Clin Pediatr. 1999, 38: 511-518.CrossRef
32.
Zurück zum Zitat Benun J, Fisher SD, Orav EJ, Schwartz ML, Exil V, Messere C, Lips SE: Cardiac management by pediatricians versus pediatric cardiologists in an inpatient acadmic center. Am Heart J. 2003, 145: 424-429. 10.1067/mhj.2003.173.CrossRefPubMed Benun J, Fisher SD, Orav EJ, Schwartz ML, Exil V, Messere C, Lips SE: Cardiac management by pediatricians versus pediatric cardiologists in an inpatient acadmic center. Am Heart J. 2003, 145: 424-429. 10.1067/mhj.2003.173.CrossRefPubMed
33.
Zurück zum Zitat Norgard G, Greve G, Rosland GA, Berg A: Referral practice and clinical assesment of heart murmurs in children [in Norwegian]. Tidsskr Nor Laegeforen. 2005, 125: 996-998.PubMed Norgard G, Greve G, Rosland GA, Berg A: Referral practice and clinical assesment of heart murmurs in children [in Norwegian]. Tidsskr Nor Laegeforen. 2005, 125: 996-998.PubMed
34.
Zurück zum Zitat Van Oort A, Le Blanc-Botden M, De Boo T, Van Der Werf T, Daniels O: The vibratory innocent heart murmur in schoolchildren: difference in auscultatory findings between school medical officers and a pediatric cardiologist. Pediatr Cardiol. 1994, 15: 282-287. 10.1007/BF00798121.CrossRefPubMed Van Oort A, Le Blanc-Botden M, De Boo T, Van Der Werf T, Daniels O: The vibratory innocent heart murmur in schoolchildren: difference in auscultatory findings between school medical officers and a pediatric cardiologist. Pediatr Cardiol. 1994, 15: 282-287. 10.1007/BF00798121.CrossRefPubMed
35.
Zurück zum Zitat McCrindle BW, Shaffer KM, Kan JS, Zahka KG, Rowe SA, Kidd L: Cardinal clinical signs in the differentiation of heart murmurs in children. Arch Pediatr Adolesc Med. 1996, 150: 169-174.CrossRefPubMed McCrindle BW, Shaffer KM, Kan JS, Zahka KG, Rowe SA, Kidd L: Cardinal clinical signs in the differentiation of heart murmurs in children. Arch Pediatr Adolesc Med. 1996, 150: 169-174.CrossRefPubMed
36.
Zurück zum Zitat Danfor DA, Martin AB, Fletcher SE, Gumbiner CH, Cheatham JP, Hofschire PJ, Kugler JD: Children with heart murmurs: can ventricular septal defect be diagnosed reliably without an echocardiogram?. J Am Coll Cardiol. 1997, 30: 243-246. 10.1016/S0735-1097(97)00101-0.CrossRef Danfor DA, Martin AB, Fletcher SE, Gumbiner CH, Cheatham JP, Hofschire PJ, Kugler JD: Children with heart murmurs: can ventricular septal defect be diagnosed reliably without an echocardiogram?. J Am Coll Cardiol. 1997, 30: 243-246. 10.1016/S0735-1097(97)00101-0.CrossRef
37.
Zurück zum Zitat Danford DA, Salaymeh KJ, Martin AB, Fletcher SE, Gumbiner CH: Pulmonary stenosis: defect-spesific diagnostic accurancy of heart murmurs in children. J Pediatr. 1999, 134: 76-81. 10.1016/S0022-3476(99)70375-6.CrossRefPubMed Danford DA, Salaymeh KJ, Martin AB, Fletcher SE, Gumbiner CH: Pulmonary stenosis: defect-spesific diagnostic accurancy of heart murmurs in children. J Pediatr. 1999, 134: 76-81. 10.1016/S0022-3476(99)70375-6.CrossRefPubMed
38.
Zurück zum Zitat Muta H, Akagi T, Egami K, Furui J, Sugahara Y, Ishii M, Matsuishi T: Incidence and clinical features of asymptomatic atrial septal defect in scool children diagnosed by heart disease screening. Circ J. 2003, 67: 112-115. 10.1253/circj.67.112.CrossRefPubMed Muta H, Akagi T, Egami K, Furui J, Sugahara Y, Ishii M, Matsuishi T: Incidence and clinical features of asymptomatic atrial septal defect in scool children diagnosed by heart disease screening. Circ J. 2003, 67: 112-115. 10.1253/circj.67.112.CrossRefPubMed
39.
Zurück zum Zitat Van Oort A, Hopman J, De Boo T, Van Der Werf T, Rohmer J, Daniels O: The vibratory innocent heart murmur in schoolchildren: A case-control doppler echocardiographic study. Pediatric Cardiol. 1994, 15: 275-281. 10.1007/BF00798120.CrossRef Van Oort A, Hopman J, De Boo T, Van Der Werf T, Rohmer J, Daniels O: The vibratory innocent heart murmur in schoolchildren: A case-control doppler echocardiographic study. Pediatric Cardiol. 1994, 15: 275-281. 10.1007/BF00798120.CrossRef
40.
41.
Zurück zum Zitat Pelech AN: The physiolgy of cardiac auscultation. Pediatr Clin North Am. 2004, 51: 1515-1535. 10.1016/j.pcl.2004.08.004.CrossRefPubMed Pelech AN: The physiolgy of cardiac auscultation. Pediatr Clin North Am. 2004, 51: 1515-1535. 10.1016/j.pcl.2004.08.004.CrossRefPubMed
42.
Zurück zum Zitat Biancaniello T: Innocent murmurs. Circulation. 2005, 111: 20-22. 10.1161/01.CIR.0000153388.41229.CB.CrossRef Biancaniello T: Innocent murmurs. Circulation. 2005, 111: 20-22. 10.1161/01.CIR.0000153388.41229.CB.CrossRef
43.
Zurück zum Zitat Advani N, Menahem S, Wilkinson JL: The diagnosis of innocent murmur in childhood. Cardiol Young. 2000, 10: 340-342.CrossRefPubMed Advani N, Menahem S, Wilkinson JL: The diagnosis of innocent murmur in childhood. Cardiol Young. 2000, 10: 340-342.CrossRefPubMed
44.
Zurück zum Zitat Donnerstein RL, Thomssen VS: Hemodynamic and anatomic factors affecting the frequency content of Still's innocent murmur. Am J Cardiol. 1994, 74: 508-510. 10.1016/0002-9149(94)90917-2.CrossRefPubMed Donnerstein RL, Thomssen VS: Hemodynamic and anatomic factors affecting the frequency content of Still's innocent murmur. Am J Cardiol. 1994, 74: 508-510. 10.1016/0002-9149(94)90917-2.CrossRefPubMed
45.
Zurück zum Zitat Noponen AL, Lukkarinen S, Sikiö K, Angerla A, Sepponen R: How to recognize the innocent vibratory murmur. Comput Cardiol. 2000, 27: 561-564. Noponen AL, Lukkarinen S, Sikiö K, Angerla A, Sepponen R: How to recognize the innocent vibratory murmur. Comput Cardiol. 2000, 27: 561-564.
46.
Zurück zum Zitat Lukkarinen S, Noponen AL: Heart auscultation: audio samples and phonocardiograms. In: Evidence-Based Medicine Guidelines [online]. 2004, Helsinki, Duodecim Medical Publications Ltd, WWW-document (password required):, [http://www.ebm-guidelines.com/] Lukkarinen S, Noponen AL: Heart auscultation: audio samples and phonocardiograms. In: Evidence-Based Medicine Guidelines [online]. 2004, Helsinki, Duodecim Medical Publications Ltd, WWW-document (password required):, [http://​www.​ebm-guidelines.​com/​]
47.
Zurück zum Zitat Lukkarinen S, Noponen AL, Angerla A, Sikiö K, Sepponen R: Experiences in videoconference system in heart murmur screening. Comput Cardiol. 1998, 25: 53-56. Lukkarinen S, Noponen AL, Angerla A, Sikiö K, Sepponen R: Experiences in videoconference system in heart murmur screening. Comput Cardiol. 1998, 25: 53-56.
48.
Zurück zum Zitat Dahl LB, Hasvold P, Arild E, Hasvold T: Heart murmurs recorded by a sensor based electronic stethoscope and e-mailed for remote assessment. Arch Dis Child. 2002, 87: 297-300. 10.1136/adc.87.4.297.CrossRefPubMedPubMedCentral Dahl LB, Hasvold P, Arild E, Hasvold T: Heart murmurs recorded by a sensor based electronic stethoscope and e-mailed for remote assessment. Arch Dis Child. 2002, 87: 297-300. 10.1136/adc.87.4.297.CrossRefPubMedPubMedCentral
49.
Zurück zum Zitat Belmont JM, Mattioli LF: Accurancy of analog telephonic stethoscopy for pediatric telecardiology. Pediatrics. 2003, 112: 780-786. 10.1542/peds.112.4.780.CrossRefPubMed Belmont JM, Mattioli LF: Accurancy of analog telephonic stethoscopy for pediatric telecardiology. Pediatrics. 2003, 112: 780-786. 10.1542/peds.112.4.780.CrossRefPubMed
Metadaten
Titel
Phono-spectrographic analysis of heart murmur in children
verfasst von
Anna-Leena Noponen
Sakari Lukkarinen
Anna Angerla
Raimo Sepponen
Publikationsdatum
01.12.2007
Verlag
BioMed Central
Erschienen in
BMC Pediatrics / Ausgabe 1/2007
Elektronische ISSN: 1471-2431
DOI
https://doi.org/10.1186/1471-2431-7-23

Weitere Artikel der Ausgabe 1/2007

BMC Pediatrics 1/2007 Zur Ausgabe

ADHS-Medikation erhöht das kardiovaskuläre Risiko

16.05.2024 Herzinsuffizienz Nachrichten

Erwachsene, die Medikamente gegen das Aufmerksamkeitsdefizit-Hyperaktivitätssyndrom einnehmen, laufen offenbar erhöhte Gefahr, an Herzschwäche zu erkranken oder einen Schlaganfall zu erleiden. Es scheint eine Dosis-Wirkungs-Beziehung zu bestehen.

Erstmanifestation eines Diabetes-Typ-1 bei Kindern: Ein Notfall!

16.05.2024 DDG-Jahrestagung 2024 Kongressbericht

Manifestiert sich ein Typ-1-Diabetes bei Kindern, ist das ein Notfall – ebenso wie eine diabetische Ketoazidose. Die Grundsäulen der Therapie bestehen aus Rehydratation, Insulin und Kaliumgabe. Insulin ist das Medikament der Wahl zur Behandlung der Ketoazidose.

Frühe Hypertonie erhöht späteres kardiovaskuläres Risiko

Wie wichtig es ist, pädiatrische Patienten auf Bluthochdruck zu screenen, zeigt eine kanadische Studie: Hypertone Druckwerte in Kindheit und Jugend steigern das Risiko für spätere kardiovaskuläre Komplikationen.

Betalaktam-Allergie: praxisnahes Vorgehen beim Delabeling

16.05.2024 Pädiatrische Allergologie Nachrichten

Die große Mehrheit der vermeintlichen Penicillinallergien sind keine. Da das „Etikett“ Betalaktam-Allergie oft schon in der Kindheit erworben wird, kann ein frühzeitiges Delabeling lebenslange Vorteile bringen. Ein Team von Pädiaterinnen und Pädiatern aus Kanada stellt vor, wie sie dabei vorgehen.

Update Pädiatrie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.