Real ear measurement (REM) and auditory performances with open, tulip and double closed dome in patients using hearing aids

verfasst von: Francesco Gazia, Bruno Galletti, Daniele Portelli, Giuseppe Alberti, Francesco Freni, Rocco Bruno, Francesco Galletti

Erschienen in: European Archives of Oto-Rhino-Laryngology | Ausgabe 5/2020

Einloggen, um Zugang zu erhalten

Abstract

Objective

The aim of the study is to evaluate the importance of acoustic modifications generated by different commercially available ear-tips, focused on domes of receiver in the canal hearing aids using Real Ear Measurement (REM).

Methods

We enrolled 110 people selecting 200 ears bearers of hearing aids. In every patient, we performed REM and audiological tests with three different dome types: Open, Tulip and Double Closed (DC). Data about real-ear occluded gain (REOG), Pure Tone Average (PTA), Word Recognition Score (WRS) with aids switched on in Free Field, Ear and Auditory Comfort were collected and analyzed.

Results

REOG gain was statistically significant different between the three types of dome, with a DC that always closes the external auditory canal (EAC) (p < 0.001). There was no statistically significant difference between the PTA (p = 0.11). Regarding the WRS there were statistically significant differences between Open and DC dome (p < 0.001) and between Tulip and DC dome (p < 0.001), with worse discrimination when using DC. Both auditory and ear comfort are worse in the DC than in the other two domes (p < 0.001). From measured REOG gain values, in 135 cases Tulip dome does not occlude the EAC, with a statistically significant difference compared to DC (p < 0.001; Odd Ratio 0.0012; 95% CI 0.001–0.0196).

Conclusion

Our study confirms the necessity to perform REM to evaluate if the prescription target is achieved, especially when tulip domes are used, because they may not occlude the ear canal, causing in some cases the reduction of the vocal discrimination.
Literatur
1.
Zurück zum Zitat Ciodaro F, Freni F, Mannella VK, Gazia F, Maceri A, Bruno R et al (2019) Use of 3D volume rendering based on high-resolution computed tomography temporal bone in patients with cochlear implants. Am J Case Rep 20:184–188CrossRef Ciodaro F, Freni F, Mannella VK, Gazia F, Maceri A, Bruno R et al (2019) Use of 3D volume rendering based on high-resolution computed tomography temporal bone in patients with cochlear implants. Am J Case Rep 20:184–188CrossRef
2.
Zurück zum Zitat Galletti B, Gazia F, Freni F, Nicita RA, Bruno R, Galletti F (2019) chronic otitis media associated with cholesteatoma in a case of the Say-Barber-Biesecker-Young-Simpson Variant of Ohdo Syndrome. Am J Case Rep 20:175–178CrossRef Galletti B, Gazia F, Freni F, Nicita RA, Bruno R, Galletti F (2019) chronic otitis media associated with cholesteatoma in a case of the Say-Barber-Biesecker-Young-Simpson Variant of Ohdo Syndrome. Am J Case Rep 20:175–178CrossRef
3.
Zurück zum Zitat Amorini M, Romeo P, Bruno R, Galletti F, Di Bella C, Longo P et al (2015) Prevalence of Deafness-Associated Connexin-26 (GJB2) and Connexin-30 (GJB6) Pathogenic Alleles in a Large Patient Cohort from Eastern Sicily. Ann Hum Genet 79(5):341–349CrossRef Amorini M, Romeo P, Bruno R, Galletti F, Di Bella C, Longo P et al (2015) Prevalence of Deafness-Associated Connexin-26 (GJB2) and Connexin-30 (GJB6) Pathogenic Alleles in a Large Patient Cohort from Eastern Sicily. Ann Hum Genet 79(5):341–349CrossRef
4.
Zurück zum Zitat Kiessling J (2001) Hearing aid fitting procedures–state-of-the-art and current issues. Scand Audiol Suppl 52:57–59CrossRef Kiessling J (2001) Hearing aid fitting procedures–state-of-the-art and current issues. Scand Audiol Suppl 52:57–59CrossRef
5.
Zurück zum Zitat Levitt H (1992) Adaptive procedures for hearing aid prescription and other audiologic applications. J Am Acad Audiol 3(2):119–131PubMed Levitt H (1992) Adaptive procedures for hearing aid prescription and other audiologic applications. J Am Acad Audiol 3(2):119–131PubMed
6.
Zurück zum Zitat Winkler A, Latzel M, Holube I (2016) Open versus closed hearing-aid fittings: a literature review of both fitting approaches. Trends Hear 15:20 Winkler A, Latzel M, Holube I (2016) Open versus closed hearing-aid fittings: a literature review of both fitting approaches. Trends Hear 15:20
7.
Zurück zum Zitat Robinson SR, Nguyen CT, Allen JB (2013) Characterizing the ear canal acoustic impedance and reflectance by pole-zero fitting. Hear Res 301:168–182CrossRef Robinson SR, Nguyen CT, Allen JB (2013) Characterizing the ear canal acoustic impedance and reflectance by pole-zero fitting. Hear Res 301:168–182CrossRef
8.
Zurück zum Zitat Jespersen CT, Møller KN (2013) Reliability of real ear insertion gain in behind-the-ear hearing aids with different coupling systems to the ear canal. Int J Audiol 52(3):169–176CrossRef Jespersen CT, Møller KN (2013) Reliability of real ear insertion gain in behind-the-ear hearing aids with different coupling systems to the ear canal. Int J Audiol 52(3):169–176CrossRef
9.
Zurück zum Zitat Denys S, Latzel M, Francart T, Wouters J (2019) A preliminary investigation into hearing aid fitting based on automated real-ear measurements integrated in the fitting software: test-retest reliability, matching accuracy and perceptual outcomes. Int J Audiol 58(3):132–140CrossRef Denys S, Latzel M, Francart T, Wouters J (2019) A preliminary investigation into hearing aid fitting based on automated real-ear measurements integrated in the fitting software: test-retest reliability, matching accuracy and perceptual outcomes. Int J Audiol 58(3):132–140CrossRef
10.
Zurück zum Zitat Scollie S, Bagatto M, Moodie S, Crukley J (2011) Accuracy and reliability of a real-ear-to-coupler difference measurement procedure implemented within a behind-the-ear hearing aid. J Am Acad Audiol 22(9):612–622CrossRef Scollie S, Bagatto M, Moodie S, Crukley J (2011) Accuracy and reliability of a real-ear-to-coupler difference measurement procedure implemented within a behind-the-ear hearing aid. J Am Acad Audiol 22(9):612–622CrossRef
11.
Zurück zum Zitat Vaisberg JM, Folkeard P, Pumford J, Narten P, Scollie S (2018) Evaluation of the repeatability and accuracy of the wideband real-ear-to-coupler difference. J Am Acad Audiol 29(6):520–532CrossRef Vaisberg JM, Folkeard P, Pumford J, Narten P, Scollie S (2018) Evaluation of the repeatability and accuracy of the wideband real-ear-to-coupler difference. J Am Acad Audiol 29(6):520–532CrossRef
12.
Zurück zum Zitat Munro KJ, Toal S (2005) Measuring the real-ear to coupler difference transfer function with an insert earphone and a hearing instrument: are they the same? Ear Hear 26(1):27–34CrossRef Munro KJ, Toal S (2005) Measuring the real-ear to coupler difference transfer function with an insert earphone and a hearing instrument: are they the same? Ear Hear 26(1):27–34CrossRef
13.
Zurück zum Zitat Moodie S, Pietrobon J, Rall E, Lindley G, Eiten L, Gordey D et al (2016) Using the real-ear-to-coupler difference within the American Academy of Audiology Pediatric Amplification Guideline: Protocols for Applying and Predicting Earmold RECDs. J Am Acad Audiol 27(3):264–275CrossRef Moodie S, Pietrobon J, Rall E, Lindley G, Eiten L, Gordey D et al (2016) Using the real-ear-to-coupler difference within the American Academy of Audiology Pediatric Amplification Guideline: Protocols for Applying and Predicting Earmold RECDs. J Am Acad Audiol 27(3):264–275CrossRef
14.
Zurück zum Zitat Gustafson S, Pittman A, Fanning R (2013) Effects of tubing length and coupling method on hearing threshold and real-ear to coupler difference measures. Am J Audiol 22(1):190–199CrossRef Gustafson S, Pittman A, Fanning R (2013) Effects of tubing length and coupling method on hearing threshold and real-ear to coupler difference measures. Am J Audiol 22(1):190–199CrossRef
15.
Zurück zum Zitat Bagatto M, Moodie S, Scollie S, Seewald R, Moodie S, Pumford J et al (2005) Clinical protocols for hearing instrument fitting in the Desired Sensation Level method. Trends Amplif 9(4):199–226CrossRef Bagatto M, Moodie S, Scollie S, Seewald R, Moodie S, Pumford J et al (2005) Clinical protocols for hearing instrument fitting in the Desired Sensation Level method. Trends Amplif 9(4):199–226CrossRef
16.
Zurück zum Zitat Munro KJ, Millward KE (2006) The influence of RECD transducer when deriving real-ear sound pressure level. Ear Hear 27(4):409–423CrossRef Munro KJ, Millward KE (2006) The influence of RECD transducer when deriving real-ear sound pressure level. Ear Hear 27(4):409–423CrossRef
17.
Zurück zum Zitat Gazia F, Abita P, Alberti G, Loteta S, Longo P, Caminiti F et al (2019) NICU infants & SNHL: experience of a western Sicily tertiary care centre. Acta Medica Mediterranea 35(2):1001–1007 Gazia F, Abita P, Alberti G, Loteta S, Longo P, Caminiti F et al (2019) NICU infants & SNHL: experience of a western Sicily tertiary care centre. Acta Medica Mediterranea 35(2):1001–1007
18.
Zurück zum Zitat Munro KJ, Salisbury VA (2002) Is the real-ear to coupler difference independent of the measurement earphone? Int J Audiol 41(7):408–413CrossRef Munro KJ, Salisbury VA (2002) Is the real-ear to coupler difference independent of the measurement earphone? Int J Audiol 41(7):408–413CrossRef
19.
Zurück zum Zitat Galletti F, Freni F, Gazia F, Galletti B (2019) Endomeatal approach in cochlear implant surgery in a patient with small mastoid cavity and procident lateral sinus. BMJ Case Rep 12(6) Galletti F, Freni F, Gazia F, Galletti B (2019) Endomeatal approach in cochlear implant surgery in a patient with small mastoid cavity and procident lateral sinus. BMJ Case Rep 12(6)
20.
Zurück zum Zitat Bagatto MP, Scollie SD, Seewald RC, Moodie KS, Hoover BM (2002) Real-ear-to-coupler difference predictions as a function of age for two coupling procedures. J Am Acad Audiol 13(8):407–415PubMed Bagatto MP, Scollie SD, Seewald RC, Moodie KS, Hoover BM (2002) Real-ear-to-coupler difference predictions as a function of age for two coupling procedures. J Am Acad Audiol 13(8):407–415PubMed
21.
Zurück zum Zitat Stuart A, Butler AK, Jones SM, Jones TA (2013) Real-ear output measures of ear level fluency devices. Int J Audiol 52(6):413–418CrossRef Stuart A, Butler AK, Jones SM, Jones TA (2013) Real-ear output measures of ear level fluency devices. Int J Audiol 52(6):413–418CrossRef
Metadaten
Titel
Real ear measurement (REM) and auditory performances with open, tulip and double closed dome in patients using hearing aids
verfasst von
Francesco Gazia
Bruno Galletti
Daniele Portelli
Giuseppe Alberti
Francesco Freni
Rocco Bruno
Francesco Galletti
Publikationsdatum
01.02.2020
Verlag
Springer Berlin Heidelberg
Erschienen in
European Archives of Oto-Rhino-Laryngology / Ausgabe 5/2020
Print ISSN: 0937-4477
Elektronische ISSN: 1434-4726
DOI
https://doi.org/10.1007/s00405-020-05822-1

Weitere Artikel der Ausgabe 5/2020

European Archives of Oto-Rhino-Laryngology 5/2020 Zur Ausgabe

Erhebliches Risiko für Kehlkopfkrebs bei mäßiger Dysplasie

29.05.2024 Larynxkarzinom Nachrichten

Fast ein Viertel der Personen mit mäßig dysplastischen Stimmlippenläsionen entwickelt einen Kehlkopftumor. Solche Personen benötigen daher eine besonders enge ärztliche Überwachung.

Hörschwäche erhöht Demenzrisiko unabhängig von Beta-Amyloid

29.05.2024 Hörstörungen Nachrichten

Hört jemand im Alter schlecht, nimmt das Hirn- und Hippocampusvolumen besonders schnell ab, was auch mit einem beschleunigten kognitiven Abbau einhergeht. Und diese Prozesse scheinen sich unabhängig von der Amyloidablagerung zu ereignen.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Betalaktam-Allergie: praxisnahes Vorgehen beim Delabeling

16.05.2024 Pädiatrische Allergologie Nachrichten

Die große Mehrheit der vermeintlichen Penicillinallergien sind keine. Da das „Etikett“ Betalaktam-Allergie oft schon in der Kindheit erworben wird, kann ein frühzeitiges Delabeling lebenslange Vorteile bringen. Ein Team von Pädiaterinnen und Pädiatern aus Kanada stellt vor, wie sie dabei vorgehen.

Update HNO

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert – ganz bequem per eMail.