Background
Respiratory diseases are the most common cause of morbidity and mortality in children under 5 worldwide [
1]. In resource limited settings (RLS), options for respiratory support are often constrained. Many locations lack both expertise and access to invasive respiratory support and/or basic pediatric intensive care [
2]. Bubble Continuous Positive Airway Pressure (bCPAP) has been proposed as an alternative to meet the need for respiratory support for infants and small children for which several commercial, durable, low-cost devices are available (e.g., Pumani) [
3‐
6]. Studies have demonstrated that bCPAP can be effective, safe and can reduce morbidity/mortality in infant populations [
3]. However evidence is mixed regarding the efficacy of bCPAP to improve outcomes in older children with respiratory failure in RLS [
7,
8].
In addition to commercially available devices, instructions on how to create “low-tech” modified bCPAP devices with readily available resources have been published, [
9] endorsed by the WHO (Figs. 16 & 17 [
10]) and promulgated in pediatric global health training websites [
11]. In practice, these devices may be subject to large variability in construction based on the availability of specific equipment. This may variably affect the amount of actual positive pressure delivered. Oxygen source, delivered liter flow rates (liters per minute, lpm), circuit tubing, nasal or facial interface, miscellaneous connections, and the reliability of the bubble CPAP chamber are several of the patient-independent variables subject to potential local modification. The efficacy of positive pressure generation at the patient interface by variations of WHO-style devices has been incompletely studied [
12]. The aim of our study was to measure the amount of positive pressure generated by commercially available bCPAP devices, the standard WHO-style device and various improvised adaptations. We hypothesized that all would generate effective positive pressure (PP) matching the submersion depth at the patient interface.
Discussion
Evidence to support the use of bCPAP for older infants and children in RLS is limited, has variable results and is of lower quality [
7,
8]. Based on our informal observations of the use of various WHO-style bCPAP devices used outside of the neonatal age range, we were concerned that the support provided for those older patients (i) may not achieve any PP benefiting the patient and thus may be wasting oxygen supply resources or (ii) may delay transfer to a higher level of care during a trial of therapy or (iii) may potentially be providing excessive PP which may be harmful to the patient.
Our results demonstrate that for the commercial bCPAP devices using flow rates recommended by the manufacturers, both F&P and Pumani devices delivered at least the expected PP at the patient interface matching the submerged depth in the bubble-chamber, although the Pumani device had a steeper increase in resulting pressures (Figs.
2,
3 and
4). The etiology behind this performance difference is unclear although could be related to the performance of the flow pump inside the Pumani device at higher flow rates or the higher expiratory tubing resistance in the Pumani device [
13]. It is important to note that with the F&P device using the smallest size nasal prongs (size 3020), we observed bubbling in the bubble-chamber without occlusion, likely secondary to the intrinsic high resistance of this prong size [
14]. The intrinsic resistance creates the potential for a provider to misjudge the level of PP support actually being provided to the child, regardless of the actual leakage at the patient interface [
15]. Chamber bubbling was a reliable proxy for occlusion with the Pumani device, with the F&P nasal masks and with the resuscitation face mask (Figs.
3,
4 and
8). However, we strongly recommend against using the resuscitation face mask in this manner because of significant safety concerns related to possible hypercarbia, risk for aspiration or gastrointestinal perforations [
7,
16].
With the standard WHO-style bCPAP device, the cut nasal cannulae generated PP that far exceeded the set depth in the F&P bCPAP chamber (Fig.
5). The data from the pediatric cannula were uniformly lower than the data from the adult cannula, suggesting the possibility of unmeasured leakage. These high measured pressures were also observed in a recent publication comparing narrow and wide-diameter bCPAP circuits [
12]. The high PP occurred only with 100% occlusion of the nasal cannulae, but these data raise concern that in small infants, especially with higher flow rates, the pressure delivered could create discomfort, intolerance or barotrauma in an infant leading to use of resources without benefit. Testing a variation of the standard WHO-style bCPAP device with larger diameter F&P tubing (a combination observed in RLS by our group) prevented the generation of excessive PP values (Fig.
7). However, this improvised device was not able to generate appropriate pressures at lower flow rates, likely secondary to leaks in the system, although we did not formally test for this possibility. However, this was still a concerning result as some providers may think that a low flow rate could be a beneficial strategy to stretch limited oxygen resources while still providing PP. Our results demonstrate that using low flow rates for this variation provides oxygen only with little to no PP benefit. An improvised variation employing the Pumani device connected to F&P nasal masks also generated appropriate levels of PP, except at very low liter flow rates (Fig.
9), as did utilizing a low-flow oxygen source connected to the Pumani tubing and prongs (Fig.
6). Interestingly, the combination of low flow oxygen source, F&P bubble chamber and Pumani tubing/interface differed from the complete Pumani data at high flow rates (Fig.
4). The differences observed between Fig.
4 and Fig.
6 and the small standard deviation measurements could be due to unmeasured leakage or differences in the actual air flow delivered by the Pumani air pump versus a low flow wall oxygen source.
F&P nasal prongs are designed to “fill the nares completely without stretching the skin” (F&P bCPAP brochure). This is to create a seal without causing tissue pressure injuries. In our testing, PP was only consistently achieved when there was 100% occlusion of the distal end of the apparatus (prongs or mask). If the apparatus was not occluded (as would be the case in older children), the PP delivered was necessarily less than the submersion depth. This observation has particular relevance when considering the implications of attempting to provide PP to older children in clinical contexts such as septic shock. When treating septic shock, PP (non-invasive or invasive) is often administered with a goal to not only reduce work of breathing, but also to reduce left ventricular afterload (assuming adequate preload), thereby reducing the overall metabolic consumption of oxygen which can help to optimize oxygen delivery [
17]. In shock states, failure to achieve these goals can have detrimental, potentially fatal, effects. This physiology underscores the risk of using improvised low-tech bCPAP devices that are used outside of previously tested age-ranges.
Either soft cloth or gel-plastic nasal masks are the patient interface typically used to deliver CPAP or BIPAP to older children in resource-replete settings. Testing the Pumani device with a typical commercially available CPAP/BIPAP mask designed for older children (SleepWeaver™, Circadiance, Export, PA) did not generate any measurable PP (Table
1, data not shown) due to the inherent controlled leakage of the mask, which the low liter bias flow rate of Pumani was not able to overcome. Typical commercially available CPAP/BIPAP masks designed for older children have vented interfaces and require high liter bias flow rates of > 40 lpm to function appropriately and prevent CO
2 rebreathing [
18]. Importantly, in a RLS, flow rates greater than 10 lpm are practically hard to generate. Oxygen cylinders deplete quickly if used at > 10 lpm, raising stability of supply chain issues if the liter flow rate is raised > 10 lpm. Typical electric oxygen concentrators only generate
< 10 lpm. Therefore, it may not be possible to provide CPAP/BIPAP to older children whose faces/noses are too large to fit a non-vented nasal mask (e.g. an F&P device) in the absence of reliable, built-in wall-piped-high-flow oxygen or air (rare in a RLS) or without a device employing a self-contained flow pump.
There are several limitations to this in vitro study. The devices and improvisations were tested in static conditions, in the absence of any type of lung simulator and were not attached to any type of simulated pharynx, preventing the estimation of pharyngeal pressures and any step down in pressure transmission [
19]. With static testing, the choice of interface should not significantly affect the generated pressure unless the outlet resistance is markedly higher. This phenomenon was only observed with the smallest F&P prong set similar to prior observations, [
14] where the authors observed a large increase in resistance for the F&P 3020 prongs independent of flow rate. We did test the F&P device with fully occluded and fully-unoccluded conditions, but this does not appropriately simulate the effect of having the mouth open or closed. We only tested flows
< 10 lpm and did not measure air flow leakage for any device. Reduction of leakage is crucial for PP generation [
15].
Although there have been some attempts to create higher flow devices with Venturi valves that can service older patients (e.g. ECPAP), [
20] we are hopeful that the emergence of high flow nasal cannula devices which have built-in flow pumps and which can generate flows up to 60 lpm without a high-pressure wall air source (e.g. F&P Airvo™ 2 system), as well as the general robustness of the Pumani model, point to the possible future development of new innovative devices that could safely, successfully and reliably generate non-invasive positive pressure without high-pressure wall-piped air/oxygen in a RLS for older children.
Conclusion
In well-resourced settings, a trial of non-invasive respiratory support to avoid endotracheal intubation and mechanical ventilation is generally accepted as safe and effective therapy for infants and children with moderate to severe respiratory illnesses. All of these interventions require high-level technical knowledge and robust supply chains which may be scarce in RLS [
21]. Bubble CPAP has been proposed as an alternative in resource-limited settings to meet the need for respiratory support for infants [
6]. However, there is an urgent need to develop reliable non-invasive positive pressure capacity for children above the infant age range.
All of the WHO-style bCPAP devices that we tested were able to generate PP that matched, or markedly exceeded, the submersion depth set in the associated bubble-chamber. PP was only measurable with 100% seal of the distal end of the bCPAP patient interfaces (commercial or improvised) but perfect seal is anatomically unlikely in older children. Therefore, the use of WHO-style bCPAP devices in patients outside the infant age range may be limited. Our data do support the effectiveness of bCPAP to generate appropriate pressures in newborns and young infants. However, based on our results, we recommend against using WHO-style bCPAP devices for non-infant patients with respiratory failure due to the low likelihood of any PP generation and instead recommend using standard oxygen therapy with nasal cannulae or face-masks, as well as early consideration of transfer to a higher level of care.