Diagnostic equivalency of fast T2 and FLAIR sequences for pediatric brain MRI: a pilot study

verfasst von: Camilo Jaimes, Edward Yang, Pauline Connaughton, Caroline D. Robson, Richard L. Robertson

Erschienen in: Pediatric Radiology | Ausgabe 4/2020

Einloggen, um Zugang zu erhalten

Abstract

Background

Faster and motion robust magnetic resonance imaging (MRI) sequences are desirable in pediatric brain MRI as they can help reduce the need for monitored anesthesia care, which is a costly and limited resource that carries medical risks.

Objective

To evaluate the diagnostic equivalency of commercially available accelerated motion robust MR sequences relative to standard sequences.

Materials and methods

This was an institutional review board-approved prospective study. Subjects underwent a clinical brain MRI using conventional multiplanar images at 3 Tesla followed by fast axial T2 and FLAIR (fluid-attenuated inversion recovery) sequences optimized for an approximately 50% reduction in acquisition time. Conventional and fast images from each subject were reviewed by two blinded pediatric neuroradiologists. The readers evaluated the presence of 12 findings. Intra-observer agreement was estimated for fast versus conventional sequences. For each set of sequences, interobserver agreement calculations and chi-square tests were used to evaluate differences between fast and conventional acquisitions. An independent third reader reviewed the intra-observer discrepancies and adjudicated them as being more conspicuous on fast sequence, conventional sequence or the equivalent. The readers also were asked to rate motion artifacts with a previously validated score.

Results

Images from 77 children (mean age: 11.3 years) were analyzed. Intra-observer agreement (fast versus conventional) ranged between 89.2% and 92.3%. Interobserver agreement ranged between 86.1% and 88.4%. Interobserver agreement was significantly higher for conventional FLAIR relative to fast FLAIR for small (<5 mm) foci of T2 in the white matter. Otherwise, interobserver agreement was not different between the fast and conventional sequences. For awake subjects, fast sequences had significantly fewer artifacts (P<0.05).

Conclusion

Conventional T2 and FLAIR sequences can be optimized to shorten acquisition while maintaining diagnostic equivalency. These faster sequences were also less susceptible to motion artifacts.
Literatur
1.
Zurück zum Zitat Robertson RL, Silk S, Ecklund K et al (2018) Imaging optimization in children. J Am Coll Radiol 15:440–443CrossRef Robertson RL, Silk S, Ecklund K et al (2018) Imaging optimization in children. J Am Coll Radiol 15:440–443CrossRef
2.
Zurück zum Zitat Jaimes C, Gee MS (2016) Strategies to minimize sedation in pediatric body magnetic resonance imaging. Pediatr Radiol 46:916–927CrossRef Jaimes C, Gee MS (2016) Strategies to minimize sedation in pediatric body magnetic resonance imaging. Pediatr Radiol 46:916–927CrossRef
3.
Zurück zum Zitat Vanderby SA, Babyn PS, Carter MW et al (2010) Effect of anesthesia and sedation on pediatric MR imaging patient flow. Radiology 256:229–237CrossRef Vanderby SA, Babyn PS, Carter MW et al (2010) Effect of anesthesia and sedation on pediatric MR imaging patient flow. Radiology 256:229–237CrossRef
4.
Zurück zum Zitat Jaimes C, Murcia DJ, Miguel K et al (2018) Identification of quality improvement areas in pediatric MRI from analysis of patient safety reports. Pediatr Radiol 48:66–73CrossRef Jaimes C, Murcia DJ, Miguel K et al (2018) Identification of quality improvement areas in pediatric MRI from analysis of patient safety reports. Pediatr Radiol 48:66–73CrossRef
5.
Zurück zum Zitat Afacan O, Erem B, Roby DP et al (2016) Evaluation of motion and its effect on brain magnetic resonance image quality in children. Pediatr Radiol 46:1728–1735CrossRef Afacan O, Erem B, Roby DP et al (2016) Evaluation of motion and its effect on brain magnetic resonance image quality in children. Pediatr Radiol 46:1728–1735CrossRef
6.
Zurück zum Zitat Jaimes C, Kirsch JE, Gee MS (2018) Fast, free-breathing and motion-minimized techniques for pediatric body magnetic resonance imaging. Pediatr Radiol 48:1197–1208CrossRef Jaimes C, Kirsch JE, Gee MS (2018) Fast, free-breathing and motion-minimized techniques for pediatric body magnetic resonance imaging. Pediatr Radiol 48:1197–1208CrossRef
7.
Zurück zum Zitat Bilgic B, Gagoski BA, Cauley SF et al (2015) Wave-CAIPI for highly accelerated 3D imaging. Magn Reson Med 73:2152–2162CrossRef Bilgic B, Gagoski BA, Cauley SF et al (2015) Wave-CAIPI for highly accelerated 3D imaging. Magn Reson Med 73:2152–2162CrossRef
8.
Zurück zum Zitat Prakkamakul S, Witzel T, Huang S et al (2016) Ultrafast brain MRI: clinical deployment and comparison to conventional brain MRI at 3T. J Neuroimaging 26:503–510CrossRef Prakkamakul S, Witzel T, Huang S et al (2016) Ultrafast brain MRI: clinical deployment and comparison to conventional brain MRI at 3T. J Neuroimaging 26:503–510CrossRef
9.
Zurück zum Zitat Fagundes J, Longo MG, Huang SY et al (2017) Diagnostic performance of a 10-minute gadolinium-enhanced brain MRI protocol compared with the standard clinical protocol for detection of intracranial enhancing lesions. AJNR Am J Neuroradiol 38:1689–1694CrossRef Fagundes J, Longo MG, Huang SY et al (2017) Diagnostic performance of a 10-minute gadolinium-enhanced brain MRI protocol compared with the standard clinical protocol for detection of intracranial enhancing lesions. AJNR Am J Neuroradiol 38:1689–1694CrossRef
10.
Zurück zum Zitat Trofimova A, Kadom N (2019) Added value from abbreviated brain MRI in children with headache. AJR Am J Roentgenol 12:1–6 Trofimova A, Kadom N (2019) Added value from abbreviated brain MRI in children with headache. AJR Am J Roentgenol 12:1–6
11.
Zurück zum Zitat Saleh A, Wenserski F, Cohnen M et al (2004) Exclusion of brain lesions: is MR contrast medium required after a negative fluid-attenuated inversion recovery sequence? Br J Radiol 77:183–188CrossRef Saleh A, Wenserski F, Cohnen M et al (2004) Exclusion of brain lesions: is MR contrast medium required after a negative fluid-attenuated inversion recovery sequence? Br J Radiol 77:183–188CrossRef
12.
Zurück zum Zitat Uffman JC, Tumin D, Raman V et al (2017) MRI utilization and the associated use of sedation and anesthesia in a pediatric ACO. J Am Coll Radiol 14:924–930CrossRef Uffman JC, Tumin D, Raman V et al (2017) MRI utilization and the associated use of sedation and anesthesia in a pediatric ACO. J Am Coll Radiol 14:924–930CrossRef
13.
Zurück zum Zitat Strauss S, Gavish E, Gottlieb P, Katsnelson L (2007) Interobserver and intraobserver variability in the sonographic assessment of fatty liver. AJR Am J Roentgenol 189:W320–W323CrossRef Strauss S, Gavish E, Gottlieb P, Katsnelson L (2007) Interobserver and intraobserver variability in the sonographic assessment of fatty liver. AJR Am J Roentgenol 189:W320–W323CrossRef
14.
Zurück zum Zitat Redondo A, Comas M, Macia F et al (2012) Inter- and intraradiologist variability in the BI-RADS assessment and breast density categories for screening mammograms. Br J Radiol 85:1465–1470CrossRef Redondo A, Comas M, Macia F et al (2012) Inter- and intraradiologist variability in the BI-RADS assessment and breast density categories for screening mammograms. Br J Radiol 85:1465–1470CrossRef
15.
Zurück zum Zitat Schnetzke M, Schuler S, Hoffend J et al (2017) Interobserver and intraobserver agreement of ligamentous injuries on conventional MRI after simple elbow dislocation. BMC Musculoskelet Disord 18:85CrossRef Schnetzke M, Schuler S, Hoffend J et al (2017) Interobserver and intraobserver agreement of ligamentous injuries on conventional MRI after simple elbow dislocation. BMC Musculoskelet Disord 18:85CrossRef
16.
Zurück zum Zitat Bayram E, Topcu Y, Karaoglu P et al (2013) Incidental white matter lesions in children presenting with headache. Headache 53:970–976CrossRef Bayram E, Topcu Y, Karaoglu P et al (2013) Incidental white matter lesions in children presenting with headache. Headache 53:970–976CrossRef
17.
Zurück zum Zitat Schmidt RL, Factor RE (2013) Understanding sources of bias in diagnostic accuracy studies. Arch Pathol Lab Med 137:558–565CrossRef Schmidt RL, Factor RE (2013) Understanding sources of bias in diagnostic accuracy studies. Arch Pathol Lab Med 137:558–565CrossRef
Metadaten
Titel
Diagnostic equivalency of fast T2 and FLAIR sequences for pediatric brain MRI: a pilot study
verfasst von
Camilo Jaimes
Edward Yang
Pauline Connaughton
Caroline D. Robson
Richard L. Robertson
Publikationsdatum
20.12.2019
Verlag
Springer Berlin Heidelberg
Erschienen in
Pediatric Radiology / Ausgabe 4/2020
Print ISSN: 0301-0449
Elektronische ISSN: 1432-1998
DOI
https://doi.org/10.1007/s00247-019-04584-1

Weitere Artikel der Ausgabe 4/2020

Pediatric Radiology 4/2020 Zur Ausgabe

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.