Proper pedicle screw placement is an integral part of spine fusion requiring expertly trained spine surgeons. Advances in medical imaging guidance have improved accuracy. There is high interest in the emerging field of robot-assisted spine surgery; however, safety and accuracy studies are needed. This study describes the pedicle screw placement of the first 100 cases in which navigated robotic assistance was used in a private practice clinical setting. A single-surgeon, single-site retrospective Institutional Review Board-exempt review of the first 100 navigated robot-assisted spine surgery cases was performed. An orthopaedic surgeon evaluated screw placement using plain film radiographs. In addition, pedicle screw malposition, reposition, and return to operating room (OR) rates were collected. Results demonstrated a high level (99%) of successful surgeon assessed pedicle screw placement in minimally invasive navigated robot-assisted spine surgery, with no malpositions requiring return to the OR.
Hinweise
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Introduction
Pedicle screw constructs are widely used for posterior fixation in spinal surgery because of their biomechanical superiority and significant correction. However, safe pedicle screw placement is paramount to achieving successful spine surgery [1]. Specialty training is required to avoid the catastrophic neurovascular complications of misplaced screws, which occur in about 4.2% of patients [2]. Nevertheless, pedicle screws are widely used in both young and adult patients, with numerous articles documenting a favorable risk-to-benefit ratio for spinal treatment [3].
Various techniques have been used to guide and confirm pedicle screw placement [4]. The use of anatomic landmarks, plain film radiography, fluoroscopic imaging (standard or image guidance), and computed tomography (CT) image guidance are examples of these techniques [5, 6]. The procedure, benefits, and limitations of each method, as well as comparisons between different approaches, have been widely published [7, 8]. Advances in medical imaging have improved the accuracy of pedicle screw placement from fluoroscopic-guided to computer-aided navigation [8]. The most recent advancement is the use of a navigated robotic-assisted spine surgery system designed to increase the accuracy of pedicle screw placement compared to freehand placement. Clinical outcome studies are required to determine pedicle screw placement accuracy when minimally invasive navigated robotic-assisted spine surgery was performed on the first 100 patients at a single institution.
Anzeige
Methods
An Institutional Review Board-exempt retrospective chart review was conducted from October 2017 to September 2018 on the first 100 navigated robotic-assisted spine surgeries. The demographic, intraoperative, and perioperative data of 100 patients who underwent lumbosacral pedicle screw placement with minimally invasive navigated robotic guidance using preoperative or intraoperative CT were analyzed. Pedicle screw malposition and reposition rates based on the surgeon’s intraoperative radiographic observations were collected. Secondary outcome measures included patient demographics, robot workflow, length of surgery, length of hospital stay, and intraoperative blood loss.
Navigated robot-assisted pedicle screw positioning system
The robotic positioning system (Excelsius GPS®; Globus Medical, Inc. Audubon, PA, USA) (Fig. 1) uses either preoperative CT, intraoperative CT, or fluoroscopy, along with a patient reference base and positioning camera to guide pedicle screw placement in real time.
×
Preoperative CT workflow
A computed tomography (CT) scan of the spinal levels in the operative field was taken prior to the patient entering the operating room (OR) and screw placement planning was completed. The CT data set was transferred into the robotic positioning system, and then, registration was completed for vertebral levels.
Intraoperative CT workflow
In intraoperative CT mode, the image coordinate system was obtained from a portable intraoperative CT (e.g., O-arm, Medtronic SNT, Louisville, CO, USA) or standard CT scan was taken at the time of surgery, with the patient already in position on the OR table. Spinal levels were identified and a CT scan was taken. Pedicle screw trajectories were planned and saved.
Anzeige
Surgical technique
A surgeon-controlled foot pedal activated and positioned the robot arm to the planned pedicle trajectory. Stab incisions were made on the skin using a scalpel. Pedicle screws were inserted under neuromonitoring using navigated instruments guided by the robotic arm. This sequence was repeated until all pedicle screws had been placed. Rods were then placed in a standard fashion. Locking caps were set once the rods were in their proper position. Intraoperative fluoroscopy images were taken to verify the screw and rod position. Pedicle screw placement was qualitatively assessed using postoperative X-rays. Following screw and rod placement, lumbar interbody fusion was performed using 1 of 3 approaches: lateral, anterior, and posterior. The endplates were prepared and the interbody spacer was manually inserted. In the lateral approach, pedicle screws were placed, while patient remained in the lateral decubitus position. In ALIF, patient was repositioned from supine to prone position after interbody placement from the anterior approach. Screws and rods were then placed for posterior supplemental fixation.
Statistical analysis
Statistical analysis was performed using SPSS Statistics Version 25 software (IBM, Armonk, NY, USA). Data were presented as mean ± standard deviation. The level of statistical significance was set to p < 0.05 for all statistical analysis.
Results
In the first 100 robotic cases, the average age was 63 ± 8 years and 48% (48/100) were female. The average body mass index was 30 kg/m2 (range 17–44 kg/m2). Twenty-five percent of patients were either current or former smokers. Forty-two percent of patients were retired (Table 1). Of the 100 cases, 55 were lateral lumbar interbody fusion (LLIF), 16 were anterior lumbar interbody fusion (ALIF), and 29 were posterior lumbar interbody fusion (PLIF). The three most common numbers of vertebral levels with pedicle screws inserted were 2-level (36%), 3-level (39%), and 4-level (20%). Intraoperative CT was used in 73 cases, while preoperative CT was used in the remaining 27 cases (Table 2). The most common level with screws inserted was either L4 (30%) or L5 (30%) (Fig. 2). Among the 100 cases, the majority of diagnoses for surgery was degenerative spondylolisthesis with neurogenic claudication (45%) and degenerative spondylolisthesis (17%). A total of 582 pedicle screws were placed. Of the 582 screws, 20 were placed without the robot due to surgeon discretion, leaving 562 pedicle screws inserted by navigated robotic guidance. Of the 562 screws, only 7 had to be repositioned manually due to surgeon discretion to reach a screw placement success rate of 99% (Fig. 3). There were no returns to the OR reported for screw-related complications.
Table 1
Baseline characteristics
Parameter
Overall
Number of patients
100
Gender
Female, n (%)
48 (48%)
Male, n (%)
52 (52%)
Age, mean (SD, range)
63 (8) (26–82)
BMI, mean (SD, range)
30 (6) (18–44)
Smoker, n (%)
Current
8 (8%)
Former
17 (17%)
Never
75 (75%)
Work status, n (%)
Retired
42 (42%)
Full time
41 (41%)
Part time
7 (7%)
Unemployed
4 (4%)
Disabled
3 (3%)
Unknown
3 (3%)
Table 2
Procedure characteristics
Parameter
Overall
Type of surgery, n (%)
LLIF
55 (55%)
PLIF
29 (29%)
ALIF
16 (16%)
Number of levels with screws inserted, n (%)
1
1 (1%)
2
36 (36%)
3
39 (39%)
4
20 (20%)
5
3 (3%)
6
1 (1%)
Workflow, n (%)
Preoperative CT
27 (27%)
Intraoperative CT
73 (73%)
×
×
Discussion
Navigated robotic-assisted spine procedures are in the early development [9]. In this study, a 99% screw placement success rate was reported for pedicle screw placement using minimally invasive navigated robotic-assisted spine surgery. In contrast, the pedicle screw accuracy rate reported by Kosmopoulos and Schizas [10] in a review of 130 studies with 37,337 pedicle screws was 95% with navigation and 90% without navigation [11]. According to Tang et al. [12], pedicle screw placement is variable even with new technologies; however, when compared to freehand screw placement, computer-navigated screws had substantially less risk of cortical damage [13]. Some inaccuracies may be attributed to a learning curve and adapting to a new workflow such as three-dimensional (3D) navigation. The 99% pedicle screw placement success rate using navigated robotic guidance recorded in the current study of the first 100 cases seems to indicate an extremely short learning curve.
Technological advances including navigation have improved the safety and accuracy of pedicle screw fixation. In a meta-analysis by Mason et al. [14], data were gathered from over 30 studies analyzing 9000 pedicle screws and found that the traditional fluoroscopy had an accuracy of 63.1%, two-dimensional navigation had 84.3% accuracy, and 3D navigation was most accurate at 95.5%. Gelalis et al. [15] performed similar analyses and concluded that navigation provides pedicle screw placement with higher accuracy. Jin et al. [16] reported a malposition rate of 9.8% in a series of 1145 screws placed with an intraoperative 2D/3D imaging navigation system.
This initial study of the first 100 cases at a single institution in the clinical use of navigated, robot-assisted spine surgery demonstrated a high pedicle screw placement success rate. There were no postoperative screw malpositions requiring a return to the OR. While this is a single-surgeon, single-site retrospective study, the pedicle screw placement success rate is better than the rates reported in the literature using robot-assisted techniques.
Conclusion
Navigated robotic guidance provides successful pedicle screw placement at a rate of 99% at this single institution, with a 0% return to OR rate.
Compliance with ethical standards
Conflict of interest
Author KTH receives royalties from and is a consultant for NuVasive, and receives royalties from Titan Spine and ODC. Authors LAA, JRR, and CGL are salaried employees (with stock options) of Globus Medical, Inc. The Excelsius GPS™ robot described in this manuscript is manufactured by Globus Medical, Inc., (GMI) where authors LAA, JRR, and CGL are employees. KTH was a paid consultant to GMI at the time of writing, is a consultant with royalties to NuVasive, and receives royalties from Titan Spine and ODC.
Anzeige
Informed consent
As the research conducted for this manuscript was a retrospective study on patient data, compliance with the ethical standards of the responsible committee on human experimentation (institutional and national) and the Helsinki Declaration of 1975 were not required.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Mit der Zeitschrift Die Chirurgie erhalten Sie zusätzlich Online-Zugriff auf weitere 43 chirurgische Fachzeitschriften, CME-Fortbildungen, Webinare, Vorbereitungskursen zur Facharztprüfung und die digitale Enzyklopädie e.Medpedia.
Bis 30. April 2024 bestellen und im ersten Jahr nur 199 € zahlen!
Nach PCI besteht ein erhöhtes Blutungsrisiko, wenn die Behandelten eine verminderte linksventrikuläre Ejektionsfraktion aufweisen. Das Risiko ist umso höher, je stärker die Pumpfunktion eingeschränkt ist.
In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.
Der Einsatz von Wundprotektoren bei offenen Eingriffen am unteren Gastrointestinaltrakt schützt vor Infektionen im Op.-Gebiet – und dient darüber hinaus der besseren Sicht. Das bestätigt mit großer Robustheit eine randomisierte Studie im Fachblatt JAMA Surgery.
Der belastende Arbeitsalltag wirkt sich negativ auf die psychische Gesundheit der Angehörigen ärztlicher Berufsgruppen aus. Chirurginnen und Chirurgen bilden da keine Ausnahme, im Gegenteil.
Update Chirurgie
Bestellen Sie unseren Fach-Newsletterund bleiben Sie gut informiert.
Das Karpaltunnelsyndrom ist die häufigste Kompressionsneuropathie peripherer Nerven. Obwohl die Anamnese mit dem nächtlichen Einschlafen der Hand (Brachialgia parästhetica nocturna) sehr typisch ist, ist eine klinisch-neurologische Untersuchung und Elektroneurografie in manchen Fällen auch eine Neurosonografie erforderlich. Im Anfangsstadium sind konservative Maßnahmen (Handgelenksschiene, Ergotherapie) empfehlenswert. Bei nicht Ansprechen der konservativen Therapie oder Auftreten von neurologischen Ausfällen ist eine Dekompression des N. medianus am Karpaltunnel indiziert.
Das Webinar beschäftigt sich mit Fragen und Antworten zu Diagnostik und Klassifikation sowie Möglichkeiten des Ausschlusses von Zusatzverletzungen. Die Referenten erläutern, welche Frakturen konservativ behandelt werden können und wie. Das Webinar beantwortet die Frage nach aktuellen operativen Therapiekonzepten: Welcher Zugang, welches Osteosynthesematerial? Auf was muss bei der Nachbehandlung der distalen Radiusfraktur geachtet werden?
Inhalte des Webinars zur S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“ sind die Darstellung des Projektes und des Erstellungswegs zur S1-Leitlinie, die Erläuterung der klinischen Relevanz der Klassifikation EAES 2015, die wissenschaftliche Begründung der wichtigsten Empfehlungen und die Darstellung stadiengerechter Therapieoptionen.