The MITRA-PRO registry revealed residual mitral regurgitation (MR) to be an important predictor of survival following transcatheter edge-to-edge repair (TEER). Intraprocedural MR assessment using 3D-Vena Contracta Area (VCA) might be a feasible tool to guide mitral TEER procedures. The study aimed to assess the impact of residual MR assessed by 3D-VCA on 1-year mortality.
Methods
823 patients with residual MR quantification using 3D-VCA in the MITRA-PRO registry, were included in this study. 1-year mortality, NYHA classification and major adverse events were assessed 1-year after mitral TEER.
Results
Patients with trace residual MR after mitral TEER were allocated to the 3D-VCA < 0.1 cm2 group (27.8%), while a 3D-VCA ≥ 0.1 < 0.3 cm2 (55.4%) was considered as mild and a 3D-VCA ≥ 0.3 cm2 (16.8%) as relevant residual MR. One-year mortality was significantly lower in patients with non-relevant residual MR (3D-VCA < 0.1 cm2: 10.5%; ≥ 0.1 < 0.3 cm2: 16.0%; ≥ 0.3: 24.8%, p = 0.003). An increasing 3D-VCA post mitral TEER was associated with a higher 1-year mortality. At a 3D-VCA of 0.07 cm2 mortality increased significantly (1-year mortality 3D-VCA post mitral TEER ≥ 0.07 cm2: 16.5% vs. < 0.07 cm2: 7.8%; p = 0.005) indicating a 3D-VCA of 0.07 cm2 to be a cut-off value for survival in daily practice.
Conclusions
Residual MR assessed by 3D-VCA after TEER is associated with 1-year mortality. Therefore, 3D-VCA is a valuable echocardiographic tool for intraprocedural MR assessment during mitral TEER and achieving a lower 3D-VCA improve patient survival. (German Clinical Trials Register: DRKS00012288).
The recently published results of the MITRA-PRO registry confirmed that residual mitral regurgitation (MR) is associated with 1-year mortality and rehospitalization in patients with functional, degenerative or mixed MR undergoing mitral transcatheter edge-to-edge repair (TEER) [1]. Assessment of MR severity via echocardiographic eyeballing is often used to determine residual regurgitation, while quantitative echocardiographic parameters like 3D-VCA are not routinely used during TEER. However, 3D-VCA is an emerging parameter for residual MR quantification following TEER [2‐4]. One might speculate, that including 3D-VCA in clinical routine of intraprocedural residual MR assessment might be beneficial for prognosis assessment and verification of procedural results. 3D-VCA correlates with hemodynamic parameters and is reduced after mitral TEER [3]. In a small cohort of patients, a 3D-VCA threshold of 0.27 cm2 was recently identified to indicate at least moderate MR with good diagnostic accuracy and a negative predictive value of 92% [3]. Also, a smaller 3D-VCA was associated with an improvement of NYHA functional class at 30-day follow-up [3]. Therefore, measurement of 3D-VCA during intraprocedural transesophageal echocardiography (TOE) might be a feasible tool for interventional cardiologists to guide mitral TEER procedures. It might provide reliable quantification of residual MR following TEER using a single parameter. However, no validated grading of 3D-VCA exists and current guidelines do not recommend 3D-VCA for residual MR graduation following TEER [5, 6]. Furthermore, applicable studies are most often retrospective, single-center and include only small sample sizes. Validated cut-off values for residual MR grading following TEER are still missing and no correlation of these cut-off values to mortality exists. We aimed to evaluate, whether intraprocedural 3D-VCA assessment of residual MR alone is able to predict prognosis following mitral TEER.
Methods
Study population
The prospective, multicenter MITRA-PRO (MITRAclip PROgnosis) registry includes 24 centers in Germany (German Clinical Trials Register: DRKS00012288). All participating study centers were recently published [1]. The registry was designed to evaluate the association of residual intraprocedural MR following TEER on 1-year mortality and rehospitalization for heart failure. The MITRA-PRO registry was performed according to good clinical practice and in compliance with the Helsinki declaration. An individual written consent was obtained from every patient. The study was approved by the ethical committee of the University of Witten-Herdecke (approval number: 96/2016) and the local ethic committees of the 24 centers, if appropriate. The MITRA-PRO registry database contains 1546 patients as previously published [1]. 3D-VCA was assessed in 823 patients prior and post mitral TEER. It was part of the MitraScore calculation, depending on the investigator's decision [1]. Clinical follow-ups were done investigator independent by a central site (ZKS Witten/Herdecke University), while 12-month transthoracic echocardiography was performed by the individual study centers. Patients with a trace residual MR after mitral TEER were allocated to the 3D-VCA < 0.1 cm2 group (0.07 cm2 defined the cut-off for increased mortality in a ROC analysis and was rounded to 0.1 cm2), while a 3D-VCA ≥ 0.1 < 0.3 cm2 was considered as mild and a 3D-VCA ≥ 0.3 cm2 as a relevant residual MR (0.27 cm2 rounded to 0.3 cm2 as previously published [3]). 2D-TOE analysis by an independent corelab was performed in 292 patients, while 47 patients had 3D-VCA raw data pre and post mitral TEER available. Figure 1 shows the study flow chart.
Mitral TEER was performed using either MitraClip XT, NT, NTR or XTR (Abbott Structural Heart, Santa Clara, California) as previously published [7]. The procedure was guided by TOE. The study protocol for MR quantification included a detailed approach for MR assessment. This involved the investigator's visual estimation, pulmonary vein flow, color flow jet area, jet density, jet contour, and vena contracta. Proximal isovelocity surface area (PISA), effective regurgitant orifice area (EROA), and regurgitant volume were used only when appropriate and were excluded from MR quantification after mitral TEER.
3D vena contracta area
According to the study protocol all mandatory MR TOE measurements were performed at baseline after guide insertion to the left atrium and 10 min after deployment of the final TEER device (before retraction of the guiding catheter into the right atrium). In 823 patients 3D-VCA was available prior and post mitral TEER. 3D-VCA evaluation was performed from the investigators in a real-world setting at the corresponding study site. 3D color Doppler imaging of the mitral valve was performed from mid-esophageal TOE views. Six ECG triggered sequential 3D–scans were mandatory for a 3D–color full volume dataset. The software used for 3D-VCA analysis was depended on the study center. Supplementary Table 1 shows a list of the study centers including the number of 3D-VCA patients, the echocardiography system used and the software for 3D-VCA analysis. All recorded 3D–color full volume datasets were checked by the investigators for stitching artefacts, blooming effects or “dropouts” within the dataset. In case of artefacts patients were excluded from 3D-VCA measurements. The 3D frame rate was recommended to be around 15–25 Hz, as recommended at the timepoint of study initiation [8]. To avoid stitching artefacts in patients with atrial fibrillation a lower number of subvolumes was used or a 3D color Doppler loop with high volume rate was acquired. Furthermore, in patients with atrial fibrillation during the procedure 3D-VCA was calculated as mean of a minimum of 3–5 measurements. 3D-VCA was defined as the cross-sectional area of the narrowest portion of the proximal regurgitant jet through the closed mitral valve in systole [9‐11]. 3D-volumes were manually cropped by the investigators to provide a direct view perpendicular to the jet direction. For 3D-VCA quantification, the 3D dataset was rotated to bisect the long axis of the jet in each of the two orthogonal planes. The third plane (short axis) was then scanned from proximal to distal in order to identify the narrowest portion of the jet immediately distal to the regurgitant orifice. Afterwards manual planimetry of the color Doppler signal was performed. Dark or very light color signals, which are of very low or high velocities were excluded of the MR jet (which is typically 400–600 cm/s). In case of multiple residual MR jets, 3D-VCA of each jet was determined separately and added up (Fig. 2).
Fig. 2
Intraprocedural 3D-VCA assessment post mitral TEER and long-term survival. (A) Summation method for 3D-VCA assessment post mitral transcatheter edge-to-edge repair (TEER). In case of more than 1 residual regurgitant jet, each jet was assessed singularly, its orthogonal planes identified, and the total VCA is calculated as the sum of all individual 3D-VCAs. MR mitral regurgitation, MC MitraClip. (B) Kaplan–Meier curve for 1-year mortality in patients with 3D-VCA < 0.1 cm2 (green line), ≥ 0.1 < 0.3 cm2 (blue line) or ≥ 0.3 cm2 (red line). 3D-VCA assessed post TEER
In 47 patients 3D-VCA pre and post mitral TEER was measured by the investigator during the index procedure as well as an independent core-lab. The core lab was blinded to the clinical patient data and the intraprocedural results. Both, investigator and core lab were blinded to each other and an independent statistician performed interobserver variability analysis.
Intraprocedural hemodynamics
All investigators minimized MR by performing TEER without exceeding the threshold of 5 mmHg mean pressure gradient across the mitral valve determined by continuous wave Doppler. All hemodynamic MR measurements were performed at baseline after guide insertion and 10 min after deployment of the last TEER device. Peri-procedural fluid administration was monitored by left ventricular end-diastolic pressure (LVEDP) or central venous pressure (CVP) aiming for an LVEDP greater than 10 mmHg and a central venous pressure between 10 and 15 mmHg. Systolic arterial blood pressure was adjusted to the patient´s preprocedural level by norepinephrine and/or fluid substitution. Left atrial pressure (LAP) was assessed prior and post mitral TEER.
Anzeige
Statistical analysis
Categorical parameters are described using absolute numbers and percentages and were compared using the Pearson chi-square test, continuous variables were presented as mean and standard deviation or median and quartiles and were compared using the Kruskal–Wallis test. A logistic regression model was used to identify predictors of 3D-VCA ≥ 0.3 cm2 after TEER at baseline. In addition to age and gender, the following baseline characteristics with p < 0.1 from the univariate analysis (3D-VCA ≥ 0.3 cm2 vs. < 0.3 cm2) were included in the model: sPAP per 20 mmHg, permanent atrial fibrillation, cardiac resynchronization therapy (CRT), mixed disease MR and non-ischemic cardiomyopathy. The cumulative incidence of mortality in the 3D-VCA groups was estimated using the Product-Limit method and were visualized as Kaplan–Meier curves with patients at risk and compared using log-rank test. Distribution of NYHA classes at baseline, discharge and 12-months follow-up according to the 3D-VCA groups was represented as bar chart. The impact of residual MR by 3D-VCA on the 1-year mortality was tested in an age- and sex adjusted logistic regression analyses and corresponding OR with 95%-CI were presented. The group 3D-VCA < 0.1 cm2 was used as reference. Interobserver variability of 3D-VCA (investigators versus core lab) was visualized using a Bland-Altmann-Plot. The plot provides a visual representation of the difference between two measurements on the y-axis and the average of the two measurements on the x-axis. The linear association between the 3D-VCA measurements (investigator versus core lab) was made with the Pearson correlation coefficient. The diagnostic values of echocardiographic parameters for MR quantification were analyzed by receiver operating characteristic (ROC) curve. All analyses were performed using SAS release 9.4 (Statistical Analysis Software, Cary, NC).
Results
The MITRA-PRO registry enrolled 1491 consecutive individuals with primary, secondary, or mixed MR who underwent mitral TEER and intraprocedural echocardiographic assessment of residual MR. In a real-world scenario, in 823 patients 3D-VCA before and after mitral TEER were part of the MitraScore calculations (Fig. 1). Following a ROC analysis at a 3D-VCA of 0.07 cm2, mortality significantly increased (1-year mortality for 3D-VCA post-mitral TEER ≥ 0.07 cm2: 16.5% vs. < 0.07 cm2: 7.8%; p = 0.005), highlighting a 3D-VCA value of rounded 0.1 cm2 as a relevant cutoff for survival assessment. Furthermore, a 3D-VCA of 0.27 cm2 was previously described as moderate residual MR (3). Therefore, we allocated the patients to 3 groups: 229 (27.8%) with a 3D-VCA of < 0.1 cm2, 456 (55.4%) with a 3D-VCA ≥ 0.1 < 0.3 cm2, and 138 (16.8%) with a 3D-VCA of ≥ 0.3 cm2. Group assignment was based on the 3D-VCA measurement post-mitral TEER.
Baseline and procedural characteristics
The baseline characteristics of the 3D-VCA groups are shown in Table 1 and were similar, except a higher systolic pulmonary artery pressure (sPAP) in the 3D-VCA ≥ 0.3 cm2 group. Furthermore, a greater number of TEER devices were used, and the procedure time was significantly longer in patients with a 3D-VCA ≥ 0.3 cm2 post-mitral TEER (supplementary Table 2). Both CVP and LAP were elevated before and after mitral TEER in the 3D-VCA ≥ 0.3 cm2 group compared to the 3D-VCA < 0.1 cm2 or 3D-VCA ≥ 0.1 < 0.3 cm2 groups (supplementary Table 2).
Peri-interventional complications and in-hospital outcomes
Peri-interventional complications were rare across all groups, with no significant differences observed, as detailed in supplementary Table 2. During the hospital stay following the index procedure, there were 3 deaths in the 3D-VCA < 0.1 cm2 group, 9 deaths in the 3D-VCA ≥ 0.1 < 0.3 cm2 group, and 4 deaths in the 3D-VCA ≥ 0.3 cm2 group, resulting in in-hospital mortality rates of 1.3%, 2.0%, and 2.9%, respectively (p = 0.56; supplementary Table 2). Additionally, the rates of in-hospital major adverse cardiovascular events (MACE) and major adverse cardiovascular and cerebrovascular events (MACCE) were similar across all groups (supplementary Table 2).
Predictors of 3D-VCA ≥ 0.3 cm2
To identify predictors of residual MR a multivariate analysis was performed. It revealed sPAP (per 20 mmHg), severe baseline MR and permanent AF as independent predictors of a 3D-VCA ≥ 0.3 cm2 (Table 2).
Table 2
Multivariable predictors of 3D VCA ≥ 0.3 cm2 after TEER at baseline
Odds ratio estimates and wald confidence intervals
1-year clinical follow-up was completed in 779 patients (94.7%) of the total cohort (3D-VCA < 0.1 cm2: 221 patients, 96.5%; 3D-VCA ≥ 0.1 < 0.3 cm2: 435 patients, 95.4%; 3D-VCA ≥ 0.3 cm2: 123 patients, 89.1%). Notably, patients with trace or mild residual MR showed significantly lower 1-year mortality rates. Specifically, the 1-year mortality rates were 10.5% in the 3D-VCA < 0.1 cm2 group, 16.0% in the 3D-VCA ≥ 0.1 < 0.3 cm2 group, and 24.8% in the 3D-VCA ≥ 0.3 cm2 group (Fig. 2). A higher 3D-VCA measurement post-mitral TEER was associated with increased 1-year mortality, as shown in Fig. 3.
Fig. 3
Association of 3D-VCA and 1-year mortality. 3D-VCA assessing residual MR following transcatheter mitral valve repair (TEER) related to 1-year mortality. Bar graphs indicate grouped 3D-VCA post TEER (< 0.1 cm2, ≥ 0.1 < 0.3 cm2 and ≥ 0.3 cm2). OR odds ratio
An analysis with the 3 predefined 3D-VCA groups revealed that group ≥ 0.1 < 0.3 cm2 is significantly different from the 3D-VCA < 0.1 group (p = 0.01) and 3D-VCA ≥ 0.3 group (p < 0.001) demonstrating increasing 1-year mortality over these groups (Fig. 3). Furthermore, re-intervention or mitral valve surgery was more frequently required in the 3D-VCA ≥ 0.3 cm2 group (3D-VCA < 0.1 cm2: 1.1%; 3D-VCA ≥ 0.1 < 0.3 cm2: 4.4%; and 3D-VCA ≥ 0.3 cm2: 8.8%; p = 0.016).
Comparison of 3D-VCA to LAP and 2D-TOE
Patients were categorized based on MR severity: MR 0/I (none, trace, mild MR), MR II (moderate MR), or MR III (severe MR), as determined by post-procedural TOE following mitral TEER. We observed a 1-year mortality rate of 15.9% for MR 0/I, 22.4% for MR II, and 25.0% for MR III, as assessed by post-procedural 2D-TOE, indicating a trend without a significant difference (log-rank p-value: 0.063). Similarly, we conducted an analysis based on LAP post-TEER, categorizing patients into LAP < 20 mmHg, LAP 20–35 mmHg, and LAP > 35 mmHg. The 1-year mortality rates were 16.2% for LAP < 20 mmHg, 16.3% for LAP 20–35 mmHg, and 25.0% for LAP > 35 mmHg, again showing a trend without a significant difference (log-rank p-value: 0.064). A c-statistic for predicting 1-year mortality was performed: 3D-VCA: 0.6 (CI: 0.52–0.63), 2D-TOE: 0.5 (CI: 0.43–0.58) and LAP:0.5 (CI: 0.43–0.56).
Anzeige
Interobserver variability
To estimate interobserver variability corelab 3D-VCA data of 47 patients were compared to the corresponding 3D-VCA assessed by the investigators. Pearson correlation coefficient (PCC) and Bland Altman Analysis of 3D-VCA (investigator versus corelab) demonstrated good reliability (PCC: 0.86; p < 0.001; supp. Figure 1).
12-months clinical and echocardiographic follow-up
In the MITRA-PRO cohort, mitral TEER reduced 3D-VCA significantly compared to baseline. In transthoracic echocardiography, MR grading improved at discharge compared to baseline across all groups, with significantly better results in patients with a 3D-VCA < 0.1 cm2 (Supplemental Table 3). However, at 12-month follow-up, the number of patients who underwent transthoracic echocardiography at individual study centers was low (3D-VCA < 0.1 cm2: 38; 3D-VCA ≥ 0.1 < 0.3 cm2: 83; 3D-VCA ≥ 0.3 cm2: 33), leading to no statistical difference in MR grading between the three 3D-VCA groups (supplemental Table 3).
Furthermore, NYHA classification was improved at 12-months follow-up after mitral TEER (Fig. 4A). Baseline NYHA classification was comparable in all 3D-VCA groups, while in 3D-VCA < 0.1 cm2 the number of patients with NYHA I/II at 12-months follow-up was significantly higher compared to 3D-VCA ≥ 0.3 cm2 indicating an association of residual MR and improvement of symptoms (Fig. 4A). Also, improvement in NYHA classification (at least one class) at 12-months follow-up following mitral TEER was significantly higher in 3D-VCA < 0.1 cm2 compared to 3D-VCA ≥ 0.3 cm2 (Fig. 4B), which further strengthens the hypothesis, that residual MR after mitral TEER is directly linked to symptomatic improvement.
Fig. 4
NYHA classification according to 3D-VCA groups. A Distribution of NYHA classification at baseline, discharge and 12-months follow-up according to 3D-VCA groups (< 0.1 cm2, ≥ 0.1 < 0.3 cm2 and ≥ 0.3 cm2). B NYHA improvement ≥ 1 class according to 3D-VCA groups (< 0.1 cm2, ≥ 0.1 < 0.3 cm2 and ≥ 0.3 cm2). 3D-VCA 3D-Vena Contracta Area
The MITRA-PRO registry evaluated the impact of intraprocedural residual mitral regurgitation on clinical outcomes after mitral TEER and revealed in patients with mild residual MR a significantly lower mortality and rehospitalization rate at one year after TEER [1]. No difference between degenerative (primary), functional (secondary) or mixed MR was observed [1]. However, in a real-world setting interventional cardiologists seek for a simple and straight-forward approach of evaluating residual MR following mitral TEER. Color Doppler echocardiography is able to allow rapid regurgitation assessment, but has limitations [13]. Especially in the context of multiple residual MR jets post TEER color Doppler echocardiography might lead to over- or underestimation of residual MR [14]. Furthermore, effective regurgitant orifice area calculation based on the PISA method is not recommended for flow quantification especially in case of multiple residual regurgitation jets [13]. Finally, the classical multi-parametric MR assessment by TTE is not feasible during the interventional procedure. The c-statistic of 0.5 for 2D-TOE assessed in this MITRA-PRO analysis highlights its limitations in post mitral-TEER assessment, and reinforces the additional usage of 3D techniques for residual MR quantification in this context. Therefore, 3D-VCA is an emerging parameter for residual MR evaluation following mitral TEER [2‐4]. However, no validated grading of 3D-VCA exists and today current guidelines do not recommend 3D-VCA for residual MR gradation [5, 6]. This study confirmed that intraprocedural 3D-VCA assessment by experienced interventional echocardiographers of residual MR alone with standardized hemodynamic conditions is able to predict 1-year mortality. Therefore, a 3D-VCA of < 0.07 cm2 showed the lowest 1-year mortality, while a 3D-VCA ≥ 0.3 cm2 is associated with the poorest outcome. Also, improvement of NYHA classification at 12-months follow-up was associated with a lower 3D-VCA post mitral TEER. Avenatti et al. showed a 3D-VCA threshold of 0.27 cm2 to be indicative for moderate residual MR, which is confirmed by our findings of excess mortality in patients with a 3D-VCA ≥ 0.3 cm2 [3]. However, we showed that also a 3D-VCA ≥ 0.07 cm2 is associated with increased 1-year mortality compared to less residual MR post mitral TEER.
Anzeige
3D-VCA assessment has been reported feasible and reproducible for both eccentric and multiple regurgitation jets, which are characteristic findings after mitral TEER [12, 14, 15]. Furthermore, in-vitro validation and cardiac MRI comparison studies showed promising results for mitral regurgitation quantification using 3D-VCA [16‐19]. Moreover, a low interobserver variability was reported for both baseline and post-TEER 3D-VCA assessment [3]. In the MITRA-PRO registry 3D-VCA was used in a real-world scenario of a large cohort of patients with primary, secondary or mixed MR. Of note, experience in this technique was a requirement for usage in the study protocol. MITRA-PRO was a nonrandomized registry and intraprocedural evaluation of MR was done at the end of mitral TEER using a web-based system, which excluded post-hoc changes of the MR assessment. Clinical follow-ups were done investigator independent by a central site excluding investigator bias. In this real-world setting 3D-VCA proofed feasibility and predicted 1-year mortality, while lowest 1-year mortality was found for 3D-VCA < 0.1 cm2. Severe MR was defined as a 3D-VCA of > 0.4 cm2 and moderate MR as 3D-VCA of > 0.3 cm2 in previous studies [3]. In MITRA-PRO a 3D-VCA ≥ 0.3 revealed the highest mortality and showed reduced clinical improvement following mitral TEER.
A recent study demonstrated the value of real-time monitoring of LAP during TEER to predict clinical outcomes. An increase in mean LAP was associated with worse clinical outcomes at short-term follow-up [20]. Therefore, we analyzed the impact of LAP assessed after deployment of the last TEER device on 1-year mortality. No significant difference between 3 predefined LAP groups and a c-statistic of 0.5 for 1-year mortality were found. Furthermore, an analysis of 1-year mortality for residual mitral regurgitation class 0/I (none, trace, mild MR), class II (moderate) and III (severe) assessed by intraprocedural 2D-TOE post TEER was performed. The predictive performance regarding 1-year mortality was similar between 2D-TOE and LAP (both with a c-statistic of 0.5). A c-statistic revealed 0.6 for 3D-VCA, which indicate that 3D-VCA has modest predictive power and should not be used in isolation for clinical prognosis assessment. Furthermore, the overlapping confidence intervals of 3D-VCA, LAP and 2D-TOE indicate no superiority of a single parameter in the assessment of long-term survival. Predicting 1-year mortality after mitral TEER is complex, involving multiple factors beyond residual MR severity, such as renal function or ventricular function. An optimal approach would involve integrating 3D-VCA with other modalities, such as comprehensive 2D echocardiographic assessments, clinical parameters, and hemodynamic data such as LAP, to provide a robust evaluation of patient outcomes post-TEER.
Avenatti et al. documented good interobserver variability for baseline 3D-VCA (ICC = 0.84) and excellent variability for post-procedural 3D-VCA (ICC = 0.93) [3]. Our findings align with these results through a Bland–Altman Analysis and Pearson's Correlation Coefficient comparing 3D-VCA values obtained by investigators and an independent core lab. However, our analysis was not designed to investigate interobserver variability and is constrained by the limited number of patients with core lab 3D-VCA assessment due to missing raw 3D-VCA data. Also, 3D-VCA is an echocardiographic parameter that can be influenced by image quality and the investigator’s experience, factors that may contribute to interobserver variability.
This analysis of the MITRA-PRO registry demonstrated the impact of residual MR assessed by 3D-VCA following TEER on 1-year mortality. Predictors of a residual MR with a 3D-VCA ≥ 0.3 cm2 were permanent atrial fibrillation, severe baseline MR and increased sPAP per 20 mmHg. A significant residual MR (3D-VCA ≥ 0.3) was associated with more severe MR grading at baseline, higher baseline sPAP, a higher proportion of mixed MR cases, the use of multiple TEER devices, and extended procedure times, all of which indicate more complex technical cases. Complex TEER cases lead to suboptimal results, as indicated by residual MR, which is directly linked to increased mortality. This critical aspect needs careful consideration in patient selection before mitral TEER.
Anzeige
In summary, an optimal procedural result is beneficial for long-term survival after TEER. Procedural 3D-VCA alone is a valuable parameter to quantify residual MR following mitral TEER and it is associated with 1-year mortality in a setup of experienced interventional echocardiographers.
Limitations
Assessment of residual MR was done by MITRA-PRO investigators and investigator bias cannot be excluded. Also, 3D-VCA was only assessed in a limited number of patients by the MITRA-PRO corelab due to missing 3D-VCA raw data. However, our results on interobserver variability were supported by a recent publication with similar findings (3). Furthermore, in this real-world, all-comer registry guideline directed medical treatment (GDMT) was not controlled at 1-year follow-up. Therefore, changes in GDMT might have influenced the outcome results. Also, accuracy of 3D-VCA tracing can be affected by non-standardized settings of different ultrasound machine manufacturers, making generalization of cutoff values difficult. MITRA-PRO started enrolling patients in 2016 and modern rendering tools were not available, which could increase accuracy of 3D-VCA assessment. Residual MR jets follow different trajectories and identification of the true perpendicular plane for each jet is necessary. Transthoracic echocardiography at the 12-month follow-up was conducted by individual study centers. However, the number of echocardiograms performed at 12 months was low. Due to this loss to follow-up of patients, no valid conclusions can be drawn about the durability of mitral TEER in the MITRA-PRO study.
Conclusion
These results of a MITRA-PRO registry demonstrated the feasibility of 3D-VCA assessment for quantification of residual MR following mitral TEER. Furthermore, residual MR assessed by 3D-VCA is associated with postprocedural 1-year mortality in a prospective large scale multicenter trial. A 3D-VCA ≥ 0.07 cm2 is associated with increased 1-year mortality, a higher rate of reintervention and reduced symptomatic improvement.
Acknowledgements
We thank Zentrum für klinische Studien (ZKS) of University Witten/Herdecke for performing clinical follow-ups of the MITRA-PRO registry.
Declarations
Competing Interests
Peter Boekstegers: Proctoring and Center of Excellence for Abbott Vascular and Edwards Lifesciences. Jörg Hausleiter: Speakers honorary and research support by Abbott Vascular and Edwards Lifesciences. Ralph Stephan van Bardeleben: Advisory board or Speakers Bureau for Abbott Cardiovascular, BioVentrix, Boston Scientific, Cardiac Dimensions, Edwards Lifesciences, and NeoChord. The other authors have no conflicts of interest to declare.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
Intraprocedural 3D-vena contracta area predicts survival after transcatheter edge-to-edge repair: results from MITRA-PRO registry
verfasst von
Dennis Rottländer Jörg Hausleiter Thomas Schmitz Alexander Bufe Melchior Seyfarth Ralph Stephan von Bardeleben Harald Beucher Taoufik Ouarrak Steffen Schneider Peter Boekstegers the MITRA-PRO investigators
Hausärztinnen und -ärzte werden immer noch am liebsten im traditionellen weißen Kittel gesehen, so das Ergebnis eines systematischen Reviews. Dabei scheinen vor allem männliche Patienten zunehmend auch ein saloppes Outfit zu billigen – allerdings nur beim Arzt, nicht bei der Ärztin.
Finden sich im Kardio-MRT gesunder älterer Ausdauersportler Zeichen einer lokalen Myokardfibrose, steigt laut einer aktuellen Studie die Wahrscheinlichkeit ventrikulärer Arrhythmien. Unklar bleibt, ob die Fibrose Zeichen einer Kardiomyopathie ist.
Der duale GLP-1/GIP-Rezeptoragonist Tirzepatid hat in einer großen Studie bewiesen, dass sich damit bei Patienten und Patientinnen mit Typ-2-Diabetes und manifester Gefäßerkrankung kardiovaskulären Ereignissen wirksam vorbeugen lässt.
Ist es von Vorteil, gebrechliche ältere Patienten mit Vorhofflimmern von einem Vitamin-K-Antagonisten auf ein DOAK umzustellen? Eine umfangreiche Metaanalyse gibt darauf nun eine fundierte Antwort.
In diesem CME-Kurs können Sie Ihr Wissen zur EKG-Befundung anhand von zwölf Video-Tutorials auffrischen und 10 CME-Punkte sammeln.
Praxisnah, relevant und mit vielen Tipps & Tricks vom Profi.