Accuracy of blood glucose measurements using the NOVA StatStrip® glucometer during cardiac surgery: a prospective observational study

verfasst von: Yosuke Nakadate, MD, PhD, Hiroaki Sato, MD, PhD, Patricia Roque, MD, Tamaki Sato, MD, Takashi Matsukawa, MD, PhD, Linda Wykes, PhD, Akiko Kawakami, MD, Thomas Schricker, MD, PhD

Erschienen in: Canadian Journal of Anesthesia/Journal canadien d'anesthésie | Ausgabe 8/2019

Einloggen, um Zugang zu erhalten

Abstract

Purpose

The Nova StatStrip® Glucose Hospital Meter System (Nova Biomedical, Waltham, MA, USA) is United States Food and Drug Administration approved for point-of-care use in critically ill patients, but its use during cardiac surgery has not been evaluated. In this study, we compare glucose values obtained during cardiac surgery by StatStrip® with values obtained by a blood gas analyzer.

Methods

Blood glucose concentrations were analyzed in 121 patients by the StatStrip point- of-care test (POCT) glucose monitor and the GEM® Premier™ 3000 blood gas analyzer (Instrumentation Laboratory Company, Bedford MA, USA). Arterial blood samples were taken at baseline (before surgery), before cardiopulmonary bypass (CPB), during early and late CPB, and 30 min after CPB. Accuracy of the StatStrip glucometer was analyzed using the Clinical and Laboratory Standards Institute (CLSI) POCT12-A3 criteria (criterion 1; 95% of samples should be ± 0.66 mMol·L−1 of reference glucose values < 5.5 mMol·L−1 and ± 12.5% for reference glucose values > 5.5 mMol·L−1, criterion 2; 98% of samples should be ± 0.83 mMol·L−1 of reference glucose values < 4.1 mMol·L−1 or 20% of the reference glucose for values > 4.1 mMol·L−1).

Results

The accuracy of StatStrip glucose measurements at baseline (99%, 100%) and before CPB (95%, 98%), but not during (early: 84%, 97%; late: 83%, 96%) and after (92%, 100%) CPB, satisfied the CLSI POCT12-A3 criteria.

Conclusion

Arterial blood glucose measurement by StatStrip was accurate before CPB, but lacked accuracy during and after CPB. Glucose values should be interpreted with caution when intensive glucose control protocols are being used during cardiac surgery.

Trial registration

ClinicalTrials.gov (NCT02729064); registered 5 April, 2016.
Literatur
1.
Zurück zum Zitat Thiessen S, Vanhorebeek I, Van den Berghe G. Glycemic control and outcome related to cardiopulmonary bypass. Best Pract Res Clin Anaesthesiol 2015; 29: 177-87.CrossRefPubMed Thiessen S, Vanhorebeek I, Van den Berghe G. Glycemic control and outcome related to cardiopulmonary bypass. Best Pract Res Clin Anaesthesiol 2015; 29: 177-87.CrossRefPubMed
2.
Zurück zum Zitat Rice MJ, Pitkin AD, Coursin DB. Review article: glucose measurement in the operating room: more complicated than it seems. Anesth Analg 2010; 110: 1056-65.CrossRefPubMed Rice MJ, Pitkin AD, Coursin DB. Review article: glucose measurement in the operating room: more complicated than it seems. Anesth Analg 2010; 110: 1056-65.CrossRefPubMed
4.
Zurück zum Zitat Clinical and Laboratory Standard Institute. Point-of-Care Blood Glucose Testing in Acute and Chronic Care Facilities; Approved Guideline - Third Edition; 2013. Clinical and Laboratory Standard Institute. Point-of-Care Blood Glucose Testing in Acute and Chronic Care Facilities; Approved Guideline - Third Edition; 2013.
5.
Zurück zum Zitat Parkes JL, Slatin SL, Pardo S, Ginsberg BH. A new consensus error grid to evaluate the clinical significance of inaccuracies in the measurement of blood glucose. Diabetes Care 2000; 23: 1143-8.CrossRefPubMed Parkes JL, Slatin SL, Pardo S, Ginsberg BH. A new consensus error grid to evaluate the clinical significance of inaccuracies in the measurement of blood glucose. Diabetes Care 2000; 23: 1143-8.CrossRefPubMed
6.
Zurück zum Zitat Clinical and Laboratory Standard Institute. In vitro diagnostic test systems – Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus, 2nd edition. International Organization for Standardization; 2013. Clinical and Laboratory Standard Institute. In vitro diagnostic test systems – Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus, 2nd edition. International Organization for Standardization; 2013.
8.
Zurück zum Zitat Nashef SA, Roques F, Sharples LD, et al. EuroSCORE II. Eur J Cardiothorac Surg 2012; 41: 734-44; discussion 744-5. Nashef SA, Roques F, Sharples LD, et al. EuroSCORE II. Eur J Cardiothorac Surg 2012; 41: 734-44; discussion 744-5.
9.
Zurück zum Zitat Dobson G. Special announcement: Guidelines to the practice of anesthesia - revised edition 2018. Can J Anesth 2018; 65: 1-6. Dobson G. Special announcement: Guidelines to the practice of anesthesia - revised edition 2018. Can J Anesth 2018; 65: 1-6.
10.
Zurück zum Zitat Karon BS, Donato LJ, Larsen CM, et al. Accuracy of capillary and arterial whole blood glucose measurements using a glucose meter in patients under general anesthesia in the operating room. Anesthesiology 2017; 127: 466-74.CrossRefPubMedPubMedCentral Karon BS, Donato LJ, Larsen CM, et al. Accuracy of capillary and arterial whole blood glucose measurements using a glucose meter in patients under general anesthesia in the operating room. Anesthesiology 2017; 127: 466-74.CrossRefPubMedPubMedCentral
11.
Zurück zum Zitat Inoue S, Egi M, Kotani J, Morita K. Accuracy of blood-glucose measurements using glucose meters and arterial blood gas analyzers in critically ill adult patients: systematic review. Crit Care 2013; 17: R48.CrossRefPubMedPubMedCentral Inoue S, Egi M, Kotani J, Morita K. Accuracy of blood-glucose measurements using glucose meters and arterial blood gas analyzers in critically ill adult patients: systematic review. Crit Care 2013; 17: R48.CrossRefPubMedPubMedCentral
12.
Zurück zum Zitat Steinfelder-Visscher J, Weerwind PW, Teerenstra S, Brouwer MH. Reliability of point-of-care hematocrit, blood gas, electrolyte, lactate and glucose measurement during cardiopulmonary bypass. Perfusion 2006; 21: 33-7.CrossRefPubMed Steinfelder-Visscher J, Weerwind PW, Teerenstra S, Brouwer MH. Reliability of point-of-care hematocrit, blood gas, electrolyte, lactate and glucose measurement during cardiopulmonary bypass. Perfusion 2006; 21: 33-7.CrossRefPubMed
13.
Zurück zum Zitat Ghys T, Goedhuys W, Spincemaille K, Gorus F, Gerlo E. Plasma-equivalent glucose at the point-of-care: evaluation of Roche Accu-Chek Inform and Abbott Precision PCx glucose meters. Clin Chim Acta 2007; 386: 63-8.CrossRefPubMed Ghys T, Goedhuys W, Spincemaille K, Gorus F, Gerlo E. Plasma-equivalent glucose at the point-of-care: evaluation of Roche Accu-Chek Inform and Abbott Precision PCx glucose meters. Clin Chim Acta 2007; 386: 63-8.CrossRefPubMed
14.
Zurück zum Zitat Karon BS, Griesmann L, Scott R, et al. Evaluation of the impact of hematocrit and other interference on the accuracy of hospital-based glucose meters. Diabetes Technol Ther 2008; 10: 111-20.CrossRefPubMed Karon BS, Griesmann L, Scott R, et al. Evaluation of the impact of hematocrit and other interference on the accuracy of hospital-based glucose meters. Diabetes Technol Ther 2008; 10: 111-20.CrossRefPubMed
15.
Zurück zum Zitat Tang Z, Louie RF, Lee JH, Lee DM, Miller EE, Kost GJ. Oxygen effects on glucose meter measurements with glucose dehydrogenase- and oxidase-based test strips for point-of-care testing. Crit Care Med 2001; 29: 1062-70.CrossRefPubMed Tang Z, Louie RF, Lee JH, Lee DM, Miller EE, Kost GJ. Oxygen effects on glucose meter measurements with glucose dehydrogenase- and oxidase-based test strips for point-of-care testing. Crit Care Med 2001; 29: 1062-70.CrossRefPubMed
16.
Zurück zum Zitat Pulzi Junior SA, Assuncao MS, Mazza BF, et al. Accuracy of different methods for blood glucose measurement in critically ill patients. Sao Paulo Med J 2009; 127: 259-65.CrossRefPubMed Pulzi Junior SA, Assuncao MS, Mazza BF, et al. Accuracy of different methods for blood glucose measurement in critically ill patients. Sao Paulo Med J 2009; 127: 259-65.CrossRefPubMed
17.
Zurück zum Zitat Kanji S, Buffie J, Hutton B, et al. Reliability of point-of-care testing for glucose measurement in critically ill adults. Crit Care Med 2005; 33: 2778-85.CrossRefPubMed Kanji S, Buffie J, Hutton B, et al. Reliability of point-of-care testing for glucose measurement in critically ill adults. Crit Care Med 2005; 33: 2778-85.CrossRefPubMed
18.
Zurück zum Zitat Mraovic B, Schwenk ES, Epstein RH. Intraoperative accuracy of a point-of-care glucose meter compared with simultaneous central laboratory measurements. J Diabetes Sci Technol 2012; 6: 541-6.CrossRefPubMedPubMedCentral Mraovic B, Schwenk ES, Epstein RH. Intraoperative accuracy of a point-of-care glucose meter compared with simultaneous central laboratory measurements. J Diabetes Sci Technol 2012; 6: 541-6.CrossRefPubMedPubMedCentral
19.
Zurück zum Zitat Karon BS, Blanshan CT, Deobald GR, Wockenfus AM. Retrospective evaluation of the accuracy of Roche AccuChek Inform and Nova StatStrip glucose meters when used on critically ill patients. Diabetes Technol Ther 2014; 16: 828-32.CrossRefPubMed Karon BS, Blanshan CT, Deobald GR, Wockenfus AM. Retrospective evaluation of the accuracy of Roche AccuChek Inform and Nova StatStrip glucose meters when used on critically ill patients. Diabetes Technol Ther 2014; 16: 828-32.CrossRefPubMed
20.
Zurück zum Zitat Mitsios JV, Ashby LA, Haverstick DM, Bruns DE, Scott MG. Analytic evaluation of a new glucose meter system in 15 different critical care settings. J Diabetes Sci Technol 2013; 7: 1282-7.CrossRefPubMedPubMedCentral Mitsios JV, Ashby LA, Haverstick DM, Bruns DE, Scott MG. Analytic evaluation of a new glucose meter system in 15 different critical care settings. J Diabetes Sci Technol 2013; 7: 1282-7.CrossRefPubMedPubMedCentral
21.
Zurück zum Zitat Murray DJ, Brosnahan WJ, Pennell B, Kapalanski D, Weiler JM, Olson J. Heparin detection by the activated coagulation time: a comparison of the sensitivity of coagulation tests and heparin assays. J Cardiothorac Vasc Anesth 1997; 11: 24-8.CrossRefPubMed Murray DJ, Brosnahan WJ, Pennell B, Kapalanski D, Weiler JM, Olson J. Heparin detection by the activated coagulation time: a comparison of the sensitivity of coagulation tests and heparin assays. J Cardiothorac Vasc Anesth 1997; 11: 24-8.CrossRefPubMed
22.
Zurück zum Zitat U.S. Food and Drug Administration. Blood Glucose Monitoring Test Systems for Prescription Point-of-Care Use - Guidance for Industry and Food and Drug Administration Staff, October 2016; Issued January 2014. Available from URL: https://www.fda.gov/downloads/ucm380325.pdf (accessed February 2019). U.S. Food and Drug Administration. Blood Glucose Monitoring Test Systems for Prescription Point-of-Care Use - Guidance for Industry and Food and Drug Administration Staff, October 2016; Issued January 2014. Available from URL: https://​www.​fda.​gov/​downloads/​ucm380325.​pdf (accessed February 2019).
23.
Zurück zum Zitat Mazer CD, Whitlock RP, Fergusson DA, et al. Six-month outcomes after restrictive or liberal transfusion for cardiac surgery. N Engl J Med 2018; 379: 1224-33.CrossRefPubMed Mazer CD, Whitlock RP, Fergusson DA, et al. Six-month outcomes after restrictive or liberal transfusion for cardiac surgery. N Engl J Med 2018; 379: 1224-33.CrossRefPubMed
Metadaten
Titel
Accuracy of blood glucose measurements using the NOVA StatStrip® glucometer during cardiac surgery: a prospective observational study
verfasst von
Yosuke Nakadate, MD, PhD
Hiroaki Sato, MD, PhD
Patricia Roque, MD
Tamaki Sato, MD
Takashi Matsukawa, MD, PhD
Linda Wykes, PhD
Akiko Kawakami, MD
Thomas Schricker, MD, PhD
Publikationsdatum
20.03.2019
Verlag
Springer International Publishing
Erschienen in
Canadian Journal of Anesthesia/Journal canadien d'anesthésie / Ausgabe 8/2019
Print ISSN: 0832-610X
Elektronische ISSN: 1496-8975
DOI
https://doi.org/10.1007/s12630-019-01350-7

Weitere Artikel der Ausgabe 8/2019

Canadian Journal of Anesthesia/Journal canadien d'anesthésie 8/2019 Zur Ausgabe

Mehr Frauen im OP – weniger postoperative Komplikationen

21.05.2024 Allgemeine Chirurgie Nachrichten

Ein Frauenanteil von mindestens einem Drittel im ärztlichen Op.-Team war in einer großen retrospektiven Studie aus Kanada mit einer signifikanten Reduktion der postoperativen Morbidität assoziiert.

Delir bei kritisch Kranken – Antipsychotika versus Placebo

16.05.2024 Delir Nachrichten

Um die Langzeitfolgen eines Delirs bei kritisch Kranken zu mildern, wird vielerorts auf eine Akuttherapie mit Antipsychotika gesetzt. Eine US-amerikanische Forschungsgruppe äußert jetzt erhebliche Vorbehalte gegen dieses Vorgehen. Denn es gibt neue Daten zum Langzeiteffekt von Haloperidol bzw. Ziprasidon versus Placebo.

Eingreifen von Umstehenden rettet vor Erstickungstod

15.05.2024 Fremdkörperaspiration Nachrichten

Wer sich an einem Essensrest verschluckt und um Luft ringt, benötigt vor allem rasche Hilfe. Dass Umstehende nur in jedem zweiten Erstickungsnotfall bereit waren, diese zu leisten, ist das ernüchternde Ergebnis einer Beobachtungsstudie aus Japan. Doch es gibt auch eine gute Nachricht.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.