An in vitro assessment of light intensity provided during direct laryngeal visualization by videolaryngoscopes with Macintosh geometry blades

verfasst von: Madeleine Harlow, BSc, George Kovacs, MD, FRCPC, Paul Brousseau, BEd, RRT (A), J. Adam Law, MD, FRCPC

Erschienen in: Canadian Journal of Anesthesia/Journal canadien d'anesthésie | Ausgabe 12/2021

Einloggen, um Zugang zu erhalten

Abstract

Background

Adequate illumination of the larynx is needed during laryngoscopy to facilitate tracheal intubation. The International Organization for Standardization (ISO) has established a minimum light intensity for direct laryngoscopy (DL) of over 500 lux for at least ten minutes, but no such standard exists for Macintosh geometry videolaryngoscope (Mac-VL) blades, which allow for both direct or indirect (videoscopic) viewing of the larynx. Using in situ bench and in vitro testing in a human cadaver, we determined illumination and luminance values delivered by various Mac-VLs and compared these with published minimum lighting benchmarks as well as a reference direct laryngoscope.

Methods

We tested six Mac-VLs (i-view™, McGRATH™ MAC, GlideScope® Spectrum™ [single-use] DVM S4, GlideScope® Titanium [reusable] Mac T4, C-MAC® S [single-use] Macintosh #4, C-MAC® [reusable] Macintosh #4) together with one direct laryngoscope (Heine LED). Each laryngoscope was assessed with three measurements, as follows: part 1: illuminance (lux) was measured in situ using a purpose-designed benchtop light intensity measurement apparatus; part 2: luminance (light reflected back to the eye) was measured (in candela m-2 [cd·m−2]) during videolaryngoscopy (VL) and DL in a human cadaver using a spot meter pointed at the interarytenoid notch; part 3: illuminance (lux) was measured during VL and DL in a human cadaver using a light meter surgically implanted just proximal to the vocal cords.

Results

Illuminance and luminance varied significantly among the Mac-VLs. Mean (standard devitation) illuminance among the six tested Mac-VLs ranged from 117 (11) to 2,626 (42) lux in the measurement apparatus and from 228 (11) to 2,900 (374) lux by the surgically implanted light meter in the cadaver. All values were less than the reference Heine direct laryngoscope and some fell below the published ISO standard of 500 lux for DL. Luminance testing by spot meter had a similarly wide range, varying from 3.78 (0.60) to 49.1 (10.4) cd·m−2, with some Mac-VLs delivering less luminance than the reference Heine direct laryngoscope.

Conclusions

Our results indicate that illuminance and luminance provided by Mac-VLs used for direct laryngeal viewing varies substantially between devices, with some falling below standards previously suggested as the minimum required for DL. While this may have no implications for the quality of image visible on a device’s video monitor, the clinician should be aware that when Mac-VLs are used for direct viewing of the larynx, lighting may not be optimal. This might adversely affect ease or success of tracheal intubation.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat International Organization for Standardization. Anaesthetic and respiratory equipment - laryngoscopes for tracheal intubation 7376:2020(E). Geneva, Switzerland: ISO; 2020. International Organization for Standardization. Anaesthetic and respiratory equipment - laryngoscopes for tracheal intubation 7376:2020(E). Geneva, Switzerland: ISO; 2020.
3.
Zurück zum Zitat Skilton RW, Parry D, Arthurs GJ, Hiles P. A study of the brightness of laryngoscope light. Anaesthesia 1996; 51: 667-72.CrossRef Skilton RW, Parry D, Arthurs GJ, Hiles P. A study of the brightness of laryngoscope light. Anaesthesia 1996; 51: 667-72.CrossRef
4.
Zurück zum Zitat Baker PA, Raos AS, Thompson JM, Jacobs RJ. Visual acuity during direct laryngoscopy at different illuminance levels. Anesth Analg 2013; 116: 343-50.CrossRef Baker PA, Raos AS, Thompson JM, Jacobs RJ. Visual acuity during direct laryngoscopy at different illuminance levels. Anesth Analg 2013; 116: 343-50.CrossRef
5.
Zurück zum Zitat Cheung KW, Kovacs G, Law JA, Brousseau P, Hill W. Illumination of bulb-on-blade laryngoscopes in the out-of-hospital setting. Acad Emerg Med 2007; 14: 496-9.CrossRef Cheung KW, Kovacs G, Law JA, Brousseau P, Hill W. Illumination of bulb-on-blade laryngoscopes in the out-of-hospital setting. Acad Emerg Med 2007; 14: 496-9.CrossRef
6.
Zurück zum Zitat Arora G, Brousseau PA, Milne AD. Laryngoscope brightness determined by visual inspection and during mannequin laryngoscopy. Can J Anesth 2019; 66: 123-4.CrossRef Arora G, Brousseau PA, Milne AD. Laryngoscope brightness determined by visual inspection and during mannequin laryngoscopy. Can J Anesth 2019; 66: 123-4.CrossRef
7.
Zurück zum Zitat Dodd KW, Prekker ME, Robinson AE, Buckley R, Reardon RF, Driver BE. Video screen viewing and first intubation attempt success with standard geometry video laryngoscope use. Am J Emerg Med 2019; 37: 1336-9.PubMed Dodd KW, Prekker ME, Robinson AE, Buckley R, Reardon RF, Driver BE. Video screen viewing and first intubation attempt success with standard geometry video laryngoscope use. Am J Emerg Med 2019; 37: 1336-9.PubMed
8.
Zurück zum Zitat Kovacs G, Levitan R, Sandeski R. Clinical cadavers as a simulation resource for procedural learning. AEM Educ Train 2018; 2: 239-47.CrossRef Kovacs G, Levitan R, Sandeski R. Clinical cadavers as a simulation resource for procedural learning. AEM Educ Train 2018; 2: 239-47.CrossRef
9.
Zurück zum Zitat Ludbrook J. Multiple inferences using confidence intervals. Clin Exp Pharmacol Physiol 2000; 27: 212-5.CrossRef Ludbrook J. Multiple inferences using confidence intervals. Clin Exp Pharmacol Physiol 2000; 27: 212-5.CrossRef
10.
Zurück zum Zitat International Organization for Standardization. Anaesthetic and respiratory equipment - Laryngoscopes for tracheal intubation. Second edition 2009-08-15. Geneva, Switzerland: ISO; 2009. International Organization for Standardization. Anaesthetic and respiratory equipment - Laryngoscopes for tracheal intubation. Second edition 2009-08-15. Geneva, Switzerland: ISO; 2009.
11.
Zurück zum Zitat Baker PA, McQuoid S, Thompson JM, Jacobs RJ. An audit of laryngoscopes and application of a new ISO standard. Paediatr Anaesth 2011; 21: 428-34.CrossRef Baker PA, McQuoid S, Thompson JM, Jacobs RJ. An audit of laryngoscopes and application of a new ISO standard. Paediatr Anaesth 2011; 21: 428-34.CrossRef
12.
Zurück zum Zitat Pieters BM, van Zundert AA. Videolaryngoscopes differ substantially in illumination of the oral cavity: a manikin study. Indian J Anaesth 2016; 60: 325-9.CrossRef Pieters BM, van Zundert AA. Videolaryngoscopes differ substantially in illumination of the oral cavity: a manikin study. Indian J Anaesth 2016; 60: 325-9.CrossRef
13.
Zurück zum Zitat Milne AD, Brousseau PA, Brousseau CA. Effects of laryngoscope handle light source on the light intensity from disposable laryngoscope blades. Anaesthesia 2014; 69: 1331-6.CrossRef Milne AD, Brousseau PA, Brousseau CA. Effects of laryngoscope handle light source on the light intensity from disposable laryngoscope blades. Anaesthesia 2014; 69: 1331-6.CrossRef
14.
Zurück zum Zitat Crosby E, Cleland M. An assessment of the luminance and light field characteristics of used direct laryngoscopes. Can J Anesth 1999; 46: 792-6.CrossRef Crosby E, Cleland M. An assessment of the luminance and light field characteristics of used direct laryngoscopes. Can J Anesth 1999; 46: 792-6.CrossRef
15.
Zurück zum Zitat Cierniak M, Timler D, Wieczorek A, Sekalski P, Borkowska N, Gaszynski T. The comparison of the technical parameters in endotracheal intubation devices: the Cmac, the Vividtrac, the McGrath Mac and the Kingvision. J Clin Monit Comput 2016; 30: 379-87.CrossRef Cierniak M, Timler D, Wieczorek A, Sekalski P, Borkowska N, Gaszynski T. The comparison of the technical parameters in endotracheal intubation devices: the Cmac, the Vividtrac, the McGrath Mac and the Kingvision. J Clin Monit Comput 2016; 30: 379-87.CrossRef
17.
Zurück zum Zitat Galvez JA, Acquah S, Ahumada L, et al. Hypoxemia, bradycardia, and multiple laryngoscopy attempts during anesthetic induction in infants: a single-center, retrospective study. Anesthesiology 2019; 131: 830-9.CrossRef Galvez JA, Acquah S, Ahumada L, et al. Hypoxemia, bradycardia, and multiple laryngoscopy attempts during anesthetic induction in infants: a single-center, retrospective study. Anesthesiology 2019; 131: 830-9.CrossRef
18.
Zurück zum Zitat Stinson HR, Srinivasan V, Topjian AA, et al. Failure of invasive airway placement on the first attempt is associated with progression to cardiac arrest in pediatric acute respiratory compromise. Pediatr Crit Care Med 2018; 19: 9-16.CrossRef Stinson HR, Srinivasan V, Topjian AA, et al. Failure of invasive airway placement on the first attempt is associated with progression to cardiac arrest in pediatric acute respiratory compromise. Pediatr Crit Care Med 2018; 19: 9-16.CrossRef
19.
Zurück zum Zitat Engelhardt T, Virag K, Veyckemans F, Habre W; APRICOT Group of the European Society of Anaesthesiology Clinical Trial Network. Airway management in paediatric anaesthesia in Europe-insights from APRICOT (Anaesthesia Practice In Children Observational Trial): a prospective multicentre observational study in 261 hospitals in Europe. Br J Anaesth 2018; 121: 66-75. Engelhardt T, Virag K, Veyckemans F, Habre W; APRICOT Group of the European Society of Anaesthesiology Clinical Trial Network. Airway management in paediatric anaesthesia in Europe-insights from APRICOT (Anaesthesia Practice In Children Observational Trial): a prospective multicentre observational study in 261 hospitals in Europe. Br J Anaesth 2018; 121: 66-75.
20.
Zurück zum Zitat Fiadjoe JE, Nishisaki A, Jagannathan N, et al. Airway management complications in children with difficult tracheal intubation from the Pediatric Difficult Intubation (PeDI) registry: a prospective cohort analysis. Lancet Respir Med 2016; 4: 37-48.CrossRef Fiadjoe JE, Nishisaki A, Jagannathan N, et al. Airway management complications in children with difficult tracheal intubation from the Pediatric Difficult Intubation (PeDI) registry: a prospective cohort analysis. Lancet Respir Med 2016; 4: 37-48.CrossRef
21.
Zurück zum Zitat Bodily JB, Webb HR, Weiss SJ, Braude DA. Incidence and duration of continuously measured oxygen desaturation during emergency department intubation. Ann Emerg Med 2016; 67: 389-95.CrossRef Bodily JB, Webb HR, Weiss SJ, Braude DA. Incidence and duration of continuously measured oxygen desaturation during emergency department intubation. Ann Emerg Med 2016; 67: 389-95.CrossRef
22.
Zurück zum Zitat Kerslake D, Oglesby AJ, Di Rollo N, et al. Tracheal intubation in an urban emergency department in Scotland: a prospective, observational study of 3738 intubations. Resuscitation 2015; 89: 20-4.CrossRef Kerslake D, Oglesby AJ, Di Rollo N, et al. Tracheal intubation in an urban emergency department in Scotland: a prospective, observational study of 3738 intubations. Resuscitation 2015; 89: 20-4.CrossRef
23.
Zurück zum Zitat Goto T, Watase H, Morita H, et al. Repeated attempts at tracheal intubation by a single intubator associated with decreased success rates in emergency departments: an analysis of a multicentre prospective observational study. Emerg Med J 2015; 32: 781-6.CrossRef Goto T, Watase H, Morita H, et al. Repeated attempts at tracheal intubation by a single intubator associated with decreased success rates in emergency departments: an analysis of a multicentre prospective observational study. Emerg Med J 2015; 32: 781-6.CrossRef
24.
Zurück zum Zitat Sakles JC, Chiu S, Mosier J, Walker C, Stolz U. The importance of first pass success when performing orotracheal intubation in the emergency department. Acad Emerg Med 2013; 20: 71-8.CrossRef Sakles JC, Chiu S, Mosier J, Walker C, Stolz U. The importance of first pass success when performing orotracheal intubation in the emergency department. Acad Emerg Med 2013; 20: 71-8.CrossRef
25.
Zurück zum Zitat Hasegawa K, Shigemitsu K, Hagiwara Y, et al. Association between repeated intubation attempts and adverse events in emergency departments: an analysis of a multicenter prospective observational study. Ann Emerg Med 2012; 60: 749-54.e2.CrossRef Hasegawa K, Shigemitsu K, Hagiwara Y, et al. Association between repeated intubation attempts and adverse events in emergency departments: an analysis of a multicenter prospective observational study. Ann Emerg Med 2012; 60: 749-54.e2.CrossRef
26.
Zurück zum Zitat Martin LD, Mhyre JM, Shanks AM, Tremper KK, Kheterpal S. 3,423 emergency tracheal intubations at a university hospital: airway outcomes and complications. Anesthesiology 2011; 114: 42-8.CrossRef Martin LD, Mhyre JM, Shanks AM, Tremper KK, Kheterpal S. 3,423 emergency tracheal intubations at a university hospital: airway outcomes and complications. Anesthesiology 2011; 114: 42-8.CrossRef
27.
Zurück zum Zitat Moore S, Dwyer D, Arendts G. Laryngoscope illumination grade does not influence time to successful manikin intubation. Emerg Med Australas 2009; 21: 131-5.PubMed Moore S, Dwyer D, Arendts G. Laryngoscope illumination grade does not influence time to successful manikin intubation. Emerg Med Australas 2009; 21: 131-5.PubMed
28.
Zurück zum Zitat Cormack RS, Lehane J. Difficult tracheal intubation in obstetrics. Anaesthesia 1984; 39: 1105-11.CrossRef Cormack RS, Lehane J. Difficult tracheal intubation in obstetrics. Anaesthesia 1984; 39: 1105-11.CrossRef
29.
Zurück zum Zitat Cook TM. A new practical classification of laryngeal view. Anaesthesia 2000; 55: 274-9.CrossRef Cook TM. A new practical classification of laryngeal view. Anaesthesia 2000; 55: 274-9.CrossRef
30.
Zurück zum Zitat Tousignant G, Tessler MJ. Light intensity and area of illumination provided by various laryngoscope blades. Can J Anaesth 1994; 41: 865-9.CrossRef Tousignant G, Tessler MJ. Light intensity and area of illumination provided by various laryngoscope blades. Can J Anaesth 1994; 41: 865-9.CrossRef
31.
Zurück zum Zitat Goodwin N, Wilkes AR, Hall JE. Flexibility and light emission of disposable paediatric Miller 1 laryngoscope blades. Anaesthesia 2006; 61: 792-9.CrossRef Goodwin N, Wilkes AR, Hall JE. Flexibility and light emission of disposable paediatric Miller 1 laryngoscope blades. Anaesthesia 2006; 61: 792-9.CrossRef
Metadaten
Titel
An in vitro assessment of light intensity provided during direct laryngeal visualization by videolaryngoscopes with Macintosh geometry blades
verfasst von
Madeleine Harlow, BSc
George Kovacs, MD, FRCPC
Paul Brousseau, BEd, RRT (A)
J. Adam Law, MD, FRCPC
Publikationsdatum
08.09.2021
Verlag
Springer International Publishing
Erschienen in
Canadian Journal of Anesthesia/Journal canadien d'anesthésie / Ausgabe 12/2021
Print ISSN: 0832-610X
Elektronische ISSN: 1496-8975
DOI
https://doi.org/10.1007/s12630-021-02099-8

Weitere Artikel der Ausgabe 12/2021

Canadian Journal of Anesthesia/Journal canadien d'anesthésie 12/2021 Zur Ausgabe

Acknowledgement

Guest Reviewers

Mehr Frauen im OP – weniger postoperative Komplikationen

21.05.2024 Allgemeine Chirurgie Nachrichten

Ein Frauenanteil von mindestens einem Drittel im ärztlichen Op.-Team war in einer großen retrospektiven Studie aus Kanada mit einer signifikanten Reduktion der postoperativen Morbidität assoziiert.

Delir bei kritisch Kranken – Antipsychotika versus Placebo

16.05.2024 Delir Nachrichten

Um die Langzeitfolgen eines Delirs bei kritisch Kranken zu mildern, wird vielerorts auf eine Akuttherapie mit Antipsychotika gesetzt. Eine US-amerikanische Forschungsgruppe äußert jetzt erhebliche Vorbehalte gegen dieses Vorgehen. Denn es gibt neue Daten zum Langzeiteffekt von Haloperidol bzw. Ziprasidon versus Placebo.

Eingreifen von Umstehenden rettet vor Erstickungstod

15.05.2024 Fremdkörperaspiration Nachrichten

Wer sich an einem Essensrest verschluckt und um Luft ringt, benötigt vor allem rasche Hilfe. Dass Umstehende nur in jedem zweiten Erstickungsnotfall bereit waren, diese zu leisten, ist das ernüchternde Ergebnis einer Beobachtungsstudie aus Japan. Doch es gibt auch eine gute Nachricht.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.