Factors for good near and distance visual outcomes of multifocal intraocular lens with inferior segmental near add

verfasst von: Jung Wan Kim, Youngsub Eom, Hyun Woo Chung, Jong Suk Song, Ji Won Jeong, Seh Kwang Park, Hyo Myung Kim

Erschienen in: Graefe's Archive for Clinical and Experimental Ophthalmology | Ausgabe 8/2020

Einloggen, um Zugang zu erhalten

Abstract

Purpose

This study was conducted to investigate factors related to postoperative good near and distance visual outcomes in the Lentis Mplus LS-313 MF20 (Oculentis Gmbh, Berlin, Germany) intraocular lens (IOL)-implanted eye.

Methods

A retrospective cross-sectional study analyzed a total of 198 eyes of 198 patients. Patients with 20/20-or-more uncorrected distance visual acuity (UDVA, 5 m) and J2-or-more uncorrected near visual acuity (UNVA, 40 cm) were distributed into the good visual outcome (GVO) group (n = 96), and the remaining patients were distributed into the moderate visual outcome (MVO) group (n = 102). Differences between the two groups were compared for patient age, preoperative measurements (kappa distance, kappa angle, and pupil size), and postoperative measurements (residual sphere, cylinder, and spherical equivalent) to explore factors related to good visual outcomes after IOL implantation.

Results

The average age of the GVO group (56.2 years) was significantly lower than that of the MVO group (58.6 years), and the average kappa distance of the former was significantly smaller than that of the latter. However, there were no significant differences between the two groups in kappa angle and pupil size. Notably, multivariate binary regression analysis revealed that multiple factors including age, residual cylinder, and spherical equivalent were associated with good visual outcomes. Based on the receiver operating characteristic curve, cutoff values regarded as good visual outcome following cataract surgery were calculated to be − 0.38 CD in residual cylinder and − 0.32 D in residual spherical equivalent.

Conclusions

Patients with implanted Lentis Mplus LS-313 MF20 IOL who are young, had small kappa distance, and had low levels of postoperative residual cylinder and spherical equivalent tended to exhibit good vision at both near and distance. These results imply that surgeons should minimize postoperative residual cylinder and spherical equivalent to provide good visual outcomes to patients with implanted Lentis Mplus LS-313 MF20 IOL.
Literatur
1.
Zurück zum Zitat Alio JL, Plaza-Puche AB, Fernandez-Buenaga R, Pikkel J, Maldonado M (2017) Multifocal intraocular lenses: an overview. Surv Ophthalmol 62:611–634CrossRef Alio JL, Plaza-Puche AB, Fernandez-Buenaga R, Pikkel J, Maldonado M (2017) Multifocal intraocular lenses: an overview. Surv Ophthalmol 62:611–634CrossRef
2.
Zurück zum Zitat Chung JK, Lee HK, Kim MK, Kim HK, Kim SW, Kim EC, Kim HS (2019) Cataract surgery practices in the Republic of Korea: a survey of the Korean Society of Cataract and Refractive Surgery 2018. Korean J Ophthalmol 33:451–457CrossRef Chung JK, Lee HK, Kim MK, Kim HK, Kim SW, Kim EC, Kim HS (2019) Cataract surgery practices in the Republic of Korea: a survey of the Korean Society of Cataract and Refractive Surgery 2018. Korean J Ophthalmol 33:451–457CrossRef
3.
Zurück zum Zitat Kim MJ, Zheleznyak L, Macrae S, Tchah H, Yoon G (2011) Objective evaluation of through-focus optical performance of presbyopia-correcting intraocular lenses using an optical bench system. J Cataract Refract Surg 37:1305–1312CrossRef Kim MJ, Zheleznyak L, Macrae S, Tchah H, Yoon G (2011) Objective evaluation of through-focus optical performance of presbyopia-correcting intraocular lenses using an optical bench system. J Cataract Refract Surg 37:1305–1312CrossRef
4.
Zurück zum Zitat Montes-Mico R, Lopez-Gil N, Perez-Vives C, Bonaque S, Ferrer-Blasco T (2012) In vitro optical performance of nonrotational symmetric and refractive-diffractive aspheric multifocal intraocular lenses: impact of tilt and decentration. J Cataract Refract Surg 38:1657–1663CrossRef Montes-Mico R, Lopez-Gil N, Perez-Vives C, Bonaque S, Ferrer-Blasco T (2012) In vitro optical performance of nonrotational symmetric and refractive-diffractive aspheric multifocal intraocular lenses: impact of tilt and decentration. J Cataract Refract Surg 38:1657–1663CrossRef
5.
Zurück zum Zitat Karhanova M, Maresova K, Pluhacek F, Mlcak P, Vlacil O, Sin M (2013) The importance of angle kappa for centration of multifocal intraocular lenses. Cesk Slov Oftalmol 69:64–68PubMed Karhanova M, Maresova K, Pluhacek F, Mlcak P, Vlacil O, Sin M (2013) The importance of angle kappa for centration of multifocal intraocular lenses. Cesk Slov Oftalmol 69:64–68PubMed
6.
Zurück zum Zitat Moore JE, McNeely RN, Pazo EE, Moore TC (2017) Rotationally asymmetric multifocal intraocular lenses: preoperative considerations and postoperative outcomes. Curr Opin Ophthalmol 28:9–15CrossRef Moore JE, McNeely RN, Pazo EE, Moore TC (2017) Rotationally asymmetric multifocal intraocular lenses: preoperative considerations and postoperative outcomes. Curr Opin Ophthalmol 28:9–15CrossRef
7.
Zurück zum Zitat Alio JL, Plaza-Puche AB, Javaloy J, Ayala MJ, Moreno LJ, Pinero DP (2012) Comparison of a new refractive multifocal intraocular lens with an inferior segmental near add and a diffractive multifocal intraocular lens. Ophthalmology 119:555–563CrossRef Alio JL, Plaza-Puche AB, Javaloy J, Ayala MJ, Moreno LJ, Pinero DP (2012) Comparison of a new refractive multifocal intraocular lens with an inferior segmental near add and a diffractive multifocal intraocular lens. Ophthalmology 119:555–563CrossRef
8.
Zurück zum Zitat Alio JL, Plaza-Puche AB, Pinero DP (2012) Rotationally asymmetric multifocal IOL implantation with and without capsular tension ring: refractive and visual outcomes and intraocular optical performance. J Refract Surg 28:253–258CrossRef Alio JL, Plaza-Puche AB, Pinero DP (2012) Rotationally asymmetric multifocal IOL implantation with and without capsular tension ring: refractive and visual outcomes and intraocular optical performance. J Refract Surg 28:253–258CrossRef
9.
Zurück zum Zitat Eom Y, Kim DW, Ryu D, Kim JH, Yang SK, Song JS, Kim SW, Kim HM (2017) Ring-shaped dysphotopsia associated with posterior chamber phakic implantable collamer lenses with a central hole. Acta Ophthalmol 95:e170–e178CrossRef Eom Y, Kim DW, Ryu D, Kim JH, Yang SK, Song JS, Kim SW, Kim HM (2017) Ring-shaped dysphotopsia associated with posterior chamber phakic implantable collamer lenses with a central hole. Acta Ophthalmol 95:e170–e178CrossRef
10.
Zurück zum Zitat Kim JS, Jung JW, Lee JM, Seo KY, Kim EK, Kim TI (2015) Clinical outcomes following implantation of diffractive multifocal intraocular lenses with varying add powers. Am J Ophthalmol 160(702–709):e701 Kim JS, Jung JW, Lee JM, Seo KY, Kim EK, Kim TI (2015) Clinical outcomes following implantation of diffractive multifocal intraocular lenses with varying add powers. Am J Ophthalmol 160(702–709):e701
11.
Zurück zum Zitat Walkow L, Klemen UM (2001) Patient satisfaction after implantation of diffractive designed multifocal intraocular lenses in dependence on objective parameters. Graefes Arch Clin Exp Ophthalmol 239:683–687CrossRef Walkow L, Klemen UM (2001) Patient satisfaction after implantation of diffractive designed multifocal intraocular lenses in dependence on objective parameters. Graefes Arch Clin Exp Ophthalmol 239:683–687CrossRef
12.
Zurück zum Zitat Woodward MA, Randleman JB, Stulting RD (2009) Dissatisfaction after multifocal intraocular lens implantation. J Cataract Refract Surg 35:992–997CrossRef Woodward MA, Randleman JB, Stulting RD (2009) Dissatisfaction after multifocal intraocular lens implantation. J Cataract Refract Surg 35:992–997CrossRef
13.
Zurück zum Zitat Prakash G, Prakash DR, Agarwal A, Kumar DA, Agarwal A, Jacob S (2011) Predictive factor and kappa angle analysis for visual satisfactions in patients with multifocal IOL implantation. Eye (Lond) 25:1187–1193CrossRef Prakash G, Prakash DR, Agarwal A, Kumar DA, Agarwal A, Jacob S (2011) Predictive factor and kappa angle analysis for visual satisfactions in patients with multifocal IOL implantation. Eye (Lond) 25:1187–1193CrossRef
14.
Zurück zum Zitat Tchah H, Nam K, Yoo A (2017) Predictive factors for photic phenomena after refractive, rotationally asymmetric, multifocal intraocular lens implantation. Int J Ophthalmol 10:241–245PubMedPubMedCentral Tchah H, Nam K, Yoo A (2017) Predictive factors for photic phenomena after refractive, rotationally asymmetric, multifocal intraocular lens implantation. Int J Ophthalmol 10:241–245PubMedPubMedCentral
15.
Zurück zum Zitat Song IS, Yoon SY, Kim JY, Kim MJ, Tchah H (2016) Influence of near-segment positioning in a rotationally asymmetric multifocal intraocular Lens. J Refract Surg 32:238–243CrossRef Song IS, Yoon SY, Kim JY, Kim MJ, Tchah H (2016) Influence of near-segment positioning in a rotationally asymmetric multifocal intraocular Lens. J Refract Surg 32:238–243CrossRef
16.
Zurück zum Zitat Yoshino M, Bissen-Miyajima H, Minami K (2013) Assessment of whether visual outcomes with diffractive multifocal intraocular lenses vary with patient age. J Cataract Refract Surg 39:1502–1506CrossRef Yoshino M, Bissen-Miyajima H, Minami K (2013) Assessment of whether visual outcomes with diffractive multifocal intraocular lenses vary with patient age. J Cataract Refract Surg 39:1502–1506CrossRef
17.
Zurück zum Zitat Plaza-Puche AB, Alio JL, Sala E, Mojzis P (2016) Impact of low mesopic contrast sensitivity outcomes in different types of modern multifocal intraocular lenses. Eur J Ophthalmol 26:612–617CrossRef Plaza-Puche AB, Alio JL, Sala E, Mojzis P (2016) Impact of low mesopic contrast sensitivity outcomes in different types of modern multifocal intraocular lenses. Eur J Ophthalmol 26:612–617CrossRef
18.
Zurück zum Zitat Pinero DP, Camps VJ, Ramon ML, Mateo V, Perez-Cambrodi RJ (2015) Error induced by the estimation of the corneal power and the effective lens position with a rotationally asymmetric refractive multifocal intraocular lens. Int J Ophthalmol 8:501–507PubMedPubMedCentral Pinero DP, Camps VJ, Ramon ML, Mateo V, Perez-Cambrodi RJ (2015) Error induced by the estimation of the corneal power and the effective lens position with a rotationally asymmetric refractive multifocal intraocular lens. Int J Ophthalmol 8:501–507PubMedPubMedCentral
19.
Zurück zum Zitat Ramji H, Moore J, Moore CB, Shah S (2016) Can the accuracy of multifocal intraocular lens power calculation be improved to make patients spectacle free? Cont Lens Anterior Eye 39:160–166CrossRef Ramji H, Moore J, Moore CB, Shah S (2016) Can the accuracy of multifocal intraocular lens power calculation be improved to make patients spectacle free? Cont Lens Anterior Eye 39:160–166CrossRef
20.
Zurück zum Zitat Calvo-Sanz JA, Sanchez-Tena MA (2018) Characterization of optical performance with defocusing curve: analysis of two refractive intraocular lens models with high and medium addition. J Optom Calvo-Sanz JA, Sanchez-Tena MA (2018) Characterization of optical performance with defocusing curve: analysis of two refractive intraocular lens models with high and medium addition. J Optom
21.
Zurück zum Zitat Kohnen T, Hemkeppler E, Herzog M, Schonbrunn S, DeLorenzo N, Petermann K, Bohm M (2018) Visual outcomes after implantation of a segmental refractive multifocal intraocular lens following cataract surgery. Am J Ophthalmol 191:156–165CrossRef Kohnen T, Hemkeppler E, Herzog M, Schonbrunn S, DeLorenzo N, Petermann K, Bohm M (2018) Visual outcomes after implantation of a segmental refractive multifocal intraocular lens following cataract surgery. Am J Ophthalmol 191:156–165CrossRef
22.
Zurück zum Zitat Kohnen T, Titke C, Bohm M (2016) Trifocal intraocular Lens implantation to treat visual demands in various distances following lens removal. Am J Ophthalmol 161:71–77 e71CrossRef Kohnen T, Titke C, Bohm M (2016) Trifocal intraocular Lens implantation to treat visual demands in various distances following lens removal. Am J Ophthalmol 161:71–77 e71CrossRef
23.
Zurück zum Zitat Mastropasqua L, Toto L, Mattei PA, Vecchiarino L, Mastropasqua A, Navarra R, Di Nicola M, Nubile M (2014) Optical coherence tomography and 3-dimensional confocal structured imaging system-guided femtosecond laser capsulotomy versus manual continuous curvilinear capsulorhexis. J Cataract Refract Surg 40:2035–2043CrossRef Mastropasqua L, Toto L, Mattei PA, Vecchiarino L, Mastropasqua A, Navarra R, Di Nicola M, Nubile M (2014) Optical coherence tomography and 3-dimensional confocal structured imaging system-guided femtosecond laser capsulotomy versus manual continuous curvilinear capsulorhexis. J Cataract Refract Surg 40:2035–2043CrossRef
24.
Zurück zum Zitat Lee DH, Lee HY, Lee KH, Chung KH, Joo CK (2001) Effect of a capsular tension ring on the shape of the capsular bag and opening and the intraocular lens. J Cataract Refract Surg 27:452–456CrossRef Lee DH, Lee HY, Lee KH, Chung KH, Joo CK (2001) Effect of a capsular tension ring on the shape of the capsular bag and opening and the intraocular lens. J Cataract Refract Surg 27:452–456CrossRef
25.
Zurück zum Zitat Kim JH, Kim H, Joo CK (2005) The effect of capsular tension ring on posterior capsular opacity in cataract surgery. Korean J Ophthalmol 19:23–28CrossRef Kim JH, Kim H, Joo CK (2005) The effect of capsular tension ring on posterior capsular opacity in cataract surgery. Korean J Ophthalmol 19:23–28CrossRef
Metadaten
Titel
Factors for good near and distance visual outcomes of multifocal intraocular lens with inferior segmental near add
verfasst von
Jung Wan Kim
Youngsub Eom
Hyun Woo Chung
Jong Suk Song
Ji Won Jeong
Seh Kwang Park
Hyo Myung Kim
Publikationsdatum
29.05.2020
Verlag
Springer Berlin Heidelberg
Erschienen in
Graefe's Archive for Clinical and Experimental Ophthalmology / Ausgabe 8/2020
Print ISSN: 0721-832X
Elektronische ISSN: 1435-702X
DOI
https://doi.org/10.1007/s00417-020-04761-1

Weitere Artikel der Ausgabe 8/2020

Graefe's Archive for Clinical and Experimental Ophthalmology 8/2020 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.