Electrical impedance to detect accidental nerve puncture during ultrasound-guided peripheral nerve blocks

verfasst von: Patrick Bardou, MD, Jean-Christophe Merle, MD, Jean-Baptiste Woillard, PharmD, Nathalie Nathan-Denizot, PhD, Pierre Beaulieu, PhD

Erschienen in: Canadian Journal of Anesthesia/Journal canadien d'anesthésie | Ausgabe 3/2013

Einloggen, um Zugang zu erhalten

Abstract

Purpose

The objective of this study was to assess if an increase in electrical impedance was associated with intraneural (sub-epineural) needle tip placement.

Methods

Two electrical impedance measurements were carried out in each of 140 peripheral nerve blocks. The first measurement was performed at a distance of 0.5-1 cm from the nerve trunk (reference value), and the second measurement was performed close to the nerve, either immediately before local anesthetic injection if no nerve puncture was suspected, or immediately before repositioning the needle if nerve puncture was suspected. Nerve puncture was suspected if any one of the following indications was present: pain or paresthesia; motor responses with a minimal stimulating current < 0.4 mA; needle tip observed inside the nerve using ultrasound; nerve swelling after injection of local anesthetic. Electrical impedance variations were compared between the no puncture and the suspected puncture groups.

Results

Nerve puncture was suspected in 21 cases. The median variation [quartiles] of electrical impedance was +6.6% [−20; 36%] in the suspected puncture group (n = 21) and −10.0% [−28; 0%] in the no puncture group (n = 119) (P = 0.02). Absolute values of electrical impedance close to the nerve were greater in the suspected puncture group (15.5 kΩ [12.0; 18.0 kΩ]) vs the no puncture group (12.0 kΩ [8.9; 15.1% kΩ]) (P = 0.013). A receiver operating characteristic (ROC) curve was constructed, and the optimal cut-off for impedance was +4.3%.

Conclusion

A > 4.3% increase in electrical impedance may indicate accidental nerve puncture during peripheral nerve block.
Literatur
1.
Zurück zum Zitat Barrington MJ, Watts SA, Gledhill SR, et al. Preliminary results of the Australasian Regional Anaesthesia Collaboration: a prospective audit of more than 7000 peripheral nerve and plexus blocks for neurologic and other complications. Reg Anesth Pain Med 2009; 34: 534-41.PubMedCrossRef Barrington MJ, Watts SA, Gledhill SR, et al. Preliminary results of the Australasian Regional Anaesthesia Collaboration: a prospective audit of more than 7000 peripheral nerve and plexus blocks for neurologic and other complications. Reg Anesth Pain Med 2009; 34: 534-41.PubMedCrossRef
2.
Zurück zum Zitat Auroy Y, Benhamou D, Bargues L, et al. Major complications of regional anesthesia in France: The SOS Regional Anesthesia Hotline Service. Anesthesiology 2002; 97: 1274-80.PubMedCrossRef Auroy Y, Benhamou D, Bargues L, et al. Major complications of regional anesthesia in France: The SOS Regional Anesthesia Hotline Service. Anesthesiology 2002; 97: 1274-80.PubMedCrossRef
3.
Zurück zum Zitat Hogan QH. Pathophysiology of peripheral nerve injury during regional anesthesia. Reg Anesth Pain Med 2008; 33: 435-41.PubMed Hogan QH. Pathophysiology of peripheral nerve injury during regional anesthesia. Reg Anesth Pain Med 2008; 33: 435-41.PubMed
4.
Zurück zum Zitat Bigeleisen PE. Nerve puncture and apparent intraneural injection during ultrasound-guided axillary block does not invariably result in neurologic injury. Anesthesiology 2006; 105: 779-83.PubMedCrossRef Bigeleisen PE. Nerve puncture and apparent intraneural injection during ultrasound-guided axillary block does not invariably result in neurologic injury. Anesthesiology 2006; 105: 779-83.PubMedCrossRef
5.
Zurück zum Zitat Neal JM, Bernards CM, Hadzic A, et al. ASRA practice advisory on neurologic complications in regional anesthesia and pain medicine. Reg Anesth Pain Med 2008; 33: 404-15.PubMed Neal JM, Bernards CM, Hadzic A, et al. ASRA practice advisory on neurologic complications in regional anesthesia and pain medicine. Reg Anesth Pain Med 2008; 33: 404-15.PubMed
6.
Zurück zum Zitat Tsai TP, Vuckovic I, Dilberovic F, et al. Intensity of the stimulating current may not be a reliable indicator of intraneural needle placement. Reg Anesth and Pain Med 2008; 33: 207-10. Tsai TP, Vuckovic I, Dilberovic F, et al. Intensity of the stimulating current may not be a reliable indicator of intraneural needle placement. Reg Anesth and Pain Med 2008; 33: 207-10.
7.
Zurück zum Zitat Choyce A, Chan VW, Middleton WJ, Knight PR, Peng P, McCartney CJ. What is the relationship between paresthesia and nerve stimulation for axillary brachial plexus block? Reg Anesth Pain Med 2001; 26: 100-4.PubMed Choyce A, Chan VW, Middleton WJ, Knight PR, Peng P, McCartney CJ. What is the relationship between paresthesia and nerve stimulation for axillary brachial plexus block? Reg Anesth Pain Med 2001; 26: 100-4.PubMed
8.
Zurück zum Zitat Urmey WF, Stanton J. Inability to consistently elicit a motor response following sensory paresthesia during interscalene block administration. Anesthesiology 2002; 96: 552-4.PubMedCrossRef Urmey WF, Stanton J. Inability to consistently elicit a motor response following sensory paresthesia during interscalene block administration. Anesthesiology 2002; 96: 552-4.PubMedCrossRef
9.
Zurück zum Zitat Cohen JM, Gray AT. Functional deficits after intraneural injection during interscalene block. Reg Anesth Pain Med 2010; 35: 397-9.PubMedCrossRef Cohen JM, Gray AT. Functional deficits after intraneural injection during interscalene block. Reg Anesth Pain Med 2010; 35: 397-9.PubMedCrossRef
10.
Zurück zum Zitat Reiss W, Kurapati S, Shariat A, Hadzic A. Nerve injury complicating ultrasound/electrostimulation-guided supraclavicular brachial plexus block. Reg Anesth Pain Med 2010; 35: 400-1.PubMedCrossRef Reiss W, Kurapati S, Shariat A, Hadzic A. Nerve injury complicating ultrasound/electrostimulation-guided supraclavicular brachial plexus block. Reg Anesth Pain Med 2010; 35: 400-1.PubMedCrossRef
11.
Zurück zum Zitat Tsui BC, Pillay JJ, Chu KT, Dillane D. Electrical impedance to distinguish intraneural from extraneural needle placement in porcine nerves during direct exposure and ultrasound guidance. Anesthesiology 2008; 109: 479-83.PubMedCrossRef Tsui BC, Pillay JJ, Chu KT, Dillane D. Electrical impedance to distinguish intraneural from extraneural needle placement in porcine nerves during direct exposure and ultrasound guidance. Anesthesiology 2008; 109: 479-83.PubMedCrossRef
12.
Zurück zum Zitat Pethig R. Dielectric properties of body tissues. Clin Phys Physiol Meas 1987; 8(Suppl A): 5-12.PubMedCrossRef Pethig R. Dielectric properties of body tissues. Clin Phys Physiol Meas 1987; 8(Suppl A): 5-12.PubMedCrossRef
13.
Zurück zum Zitat Cory PC. Increased impedance on nerve stimulator display may actually reflect a decrease in total system impedance. Anesthesiology 2009; 110: 1192-4.PubMedCrossRef Cory PC. Increased impedance on nerve stimulator display may actually reflect a decrease in total system impedance. Anesthesiology 2009; 110: 1192-4.PubMedCrossRef
14.
Zurück zum Zitat Thomas SP, Arandia HY, Leal JJ, Chilcoat R. Tissue electrical impedance as an aid to identification of the epidural space. Reg Anesth Pain Med 1987; 12: 169-71. Thomas SP, Arandia HY, Leal JJ, Chilcoat R. Tissue electrical impedance as an aid to identification of the epidural space. Reg Anesth Pain Med 1987; 12: 169-71.
15.
Zurück zum Zitat Chin J, Tsui BC. No change in impedance upon intravascular injection of D5W. Can J Anesth 2010; 57: 559-64.PubMedCrossRef Chin J, Tsui BC. No change in impedance upon intravascular injection of D5W. Can J Anesth 2010; 57: 559-64.PubMedCrossRef
16.
Zurück zum Zitat Sites BD, Spence BC, Gallagher JD, Wiley CW, Bertrand ML, Blike GT. Characterizing novice behavior associated with learning ultrasound-guided peripheral regional anesthesia. Reg Anesth Pain Med 2007; 32: 107-15.PubMed Sites BD, Spence BC, Gallagher JD, Wiley CW, Bertrand ML, Blike GT. Characterizing novice behavior associated with learning ultrasound-guided peripheral regional anesthesia. Reg Anesth Pain Med 2007; 32: 107-15.PubMed
17.
Zurück zum Zitat Byrne K, Tsui BC. Practical concepts in nerve stimulation: impedance and other recent advances. Int Anaesthesiol Clin 2011; 49: 81-90.CrossRef Byrne K, Tsui BC. Practical concepts in nerve stimulation: impedance and other recent advances. Int Anaesthesiol Clin 2011; 49: 81-90.CrossRef
18.
Zurück zum Zitat Dillane D, Tsui BC. From basic concepts to emerging technologies in regional anesthesia. Curr Opin Anaesthesiol 2010; 23: 643-9.PubMedCrossRef Dillane D, Tsui BC. From basic concepts to emerging technologies in regional anesthesia. Curr Opin Anaesthesiol 2010; 23: 643-9.PubMedCrossRef
19.
Zurück zum Zitat Sauter AR, Dodgson MS, Kalvoy H, Grimnes S, Stubhaug A, Klaastad O. Current threshold for nerve stimulation depends on electrical impedance of the tissue: a study of ultrasound-guided electrical nerve stimulation of the median nerve. Anesth Analg 2009; 108: 1338-43.PubMedCrossRef Sauter AR, Dodgson MS, Kalvoy H, Grimnes S, Stubhaug A, Klaastad O. Current threshold for nerve stimulation depends on electrical impedance of the tissue: a study of ultrasound-guided electrical nerve stimulation of the median nerve. Anesth Analg 2009; 108: 1338-43.PubMedCrossRef
20.
Zurück zum Zitat Gielen FL, Wallinga-de Jonge W, Boon KL. Electrical conductivity of skeletal muscle tissue: experimental results from different muscles in vivo. Med Biol Eng Comput 1984; 22: 569-77.PubMedCrossRef Gielen FL, Wallinga-de Jonge W, Boon KL. Electrical conductivity of skeletal muscle tissue: experimental results from different muscles in vivo. Med Biol Eng Comput 1984; 22: 569-77.PubMedCrossRef
21.
Zurück zum Zitat Tsai JZ, Will JA, Hubbard-Van Stelle S, et al. Error analysis of tissue resistivity measurement. IEEE Trans Biomed Eng 2002; 49: 484-94.PubMedCrossRef Tsai JZ, Will JA, Hubbard-Van Stelle S, et al. Error analysis of tissue resistivity measurement. IEEE Trans Biomed Eng 2002; 49: 484-94.PubMedCrossRef
22.
Zurück zum Zitat Bigeleisen PE, Moayeri N, Groen GJ. Extraneural versus intraneural stimulation thresholds during ultrasound-guided supraclavicular block. Anesthesiology 2009; 110: 1235-43.PubMedCrossRef Bigeleisen PE, Moayeri N, Groen GJ. Extraneural versus intraneural stimulation thresholds during ultrasound-guided supraclavicular block. Anesthesiology 2009; 110: 1235-43.PubMedCrossRef
Metadaten
Titel
Electrical impedance to detect accidental nerve puncture during ultrasound-guided peripheral nerve blocks
verfasst von
Patrick Bardou, MD
Jean-Christophe Merle, MD
Jean-Baptiste Woillard, PharmD
Nathalie Nathan-Denizot, PhD
Pierre Beaulieu, PhD
Publikationsdatum
01.03.2013
Verlag
Springer-Verlag
Erschienen in
Canadian Journal of Anesthesia/Journal canadien d'anesthésie / Ausgabe 3/2013
Print ISSN: 0832-610X
Elektronische ISSN: 1496-8975
DOI
https://doi.org/10.1007/s12630-012-9845-y

Weitere Artikel der Ausgabe 3/2013

Canadian Journal of Anesthesia/Journal canadien d'anesthésie 3/2013 Zur Ausgabe

Mehr Frauen im OP – weniger postoperative Komplikationen

21.05.2024 Allgemeine Chirurgie Nachrichten

Ein Frauenanteil von mindestens einem Drittel im ärztlichen Op.-Team war in einer großen retrospektiven Studie aus Kanada mit einer signifikanten Reduktion der postoperativen Morbidität assoziiert.

Delir bei kritisch Kranken – Antipsychotika versus Placebo

16.05.2024 Delir Nachrichten

Um die Langzeitfolgen eines Delirs bei kritisch Kranken zu mildern, wird vielerorts auf eine Akuttherapie mit Antipsychotika gesetzt. Eine US-amerikanische Forschungsgruppe äußert jetzt erhebliche Vorbehalte gegen dieses Vorgehen. Denn es gibt neue Daten zum Langzeiteffekt von Haloperidol bzw. Ziprasidon versus Placebo.

Eingreifen von Umstehenden rettet vor Erstickungstod

15.05.2024 Fremdkörperaspiration Nachrichten

Wer sich an einem Essensrest verschluckt und um Luft ringt, benötigt vor allem rasche Hilfe. Dass Umstehende nur in jedem zweiten Erstickungsnotfall bereit waren, diese zu leisten, ist das ernüchternde Ergebnis einer Beobachtungsstudie aus Japan. Doch es gibt auch eine gute Nachricht.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.