Erschienen in: Journal of Cardiovascular Magnetic Resonance 1/2020

Open Access 01.12.2020 | Research

4D flow cardiovascular magnetic resonance for monitoring of aortic valve repair in bicuspid aortic valve disease

verfasst von: Alexander Lenz, Johannes Petersen, Christoph Riedel, Julius M. Weinrich, Hendrik Kooijman, Bjoern P. Schoennagel, Gerhard Adam, Yskert von Kodolitsch, Hermann Reichenspurner, Evaldas Girdauskas, Peter Bannas

Erschienen in: Journal of Cardiovascular Magnetic Resonance | Ausgabe 1/2020

Abstract

Background

Aortic valve repair has become a treatment option for adults with symptomatic bicuspid (BAV) or unicuspid (UAV) aortic valve insufficiency. Our aim was to demonstrate the feasibility of 4D flow cardiovascular magnetic resonance (CMR) to assess the impact of aortic valve repair on changes in blood flow dynamics in patients with symptomatic BAV or UAV.

Methods

Twenty patients with adult congenital heart disease (median 35 years, range 18–64; 16 male) and symptomatic aortic valve regurgitation (15 BAV, 5 UAV) were prospectively studied. All patients underwent 4D flow CMR before and after aortic valve repair. Aortic valve regurgitant fraction and systolic peak velocity were estimated. The degree of helical and vortical flow was evaluated according to a 3-point scale. Relative flow displacement and wall shear stress (WSS) were quantified at predefined levels in the thoracic aorta.

Results

All patients underwent successful aortic valve repair with a significant reduction of aortic valve regurgitation (16.7 ± 9.8% to 6.4 ± 4.4%, p < 0.001) and systolic peak velocity (2.3 ± 0.9 to 1.9 ± 0.4 m/s, p = 0.014). Both helical flow (1.6 ± 0.6 vs. 0.9 ± 0.5, p < 0.001) and vortical flow (1.2 ± 0.8 vs. 0.5 ± 0.6, p = 0.002) as well as both flow displacement (0.3 ± 0.1 vs. 0.25 ± 0.1, p = 0.031) and WSS (0.8 ± 0.2 N/m2 vs. 0.5 ± 0.2 N/m2, p < 0.001) in the ascending aorta were significantly reduced after aortic valve repair.

Conclusions

4D flow CMR allows assessment of the impact of aortic valve repair on changes in blood flow dynamics in patients with bicuspid aortic valve disease.
Hinweise

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
2D
Two dimensional
4D
Four dimensional
A
Anterior
AAo
Ascending aorta
AA
Aortic arch
BAV
Bicuspid aortic valve
bSSFP
Balanced steady state free precession
CMR
Cardiovascular magnetic resonance
DAo
Descending aorta
distAA
Distal aortic arch
ECG
Electrocardiogram
EDV
End-diastolic volume
EF
Ejection fraction
ESV
End-systolic volume
L
Left
LA
Left-anterior
LP
Left-posterior
LV
Left ventricle/left ventricular
midAAo
Mid-ascending aorta
P
Posterior
proxAA
Proximal aortic arch
R
Right
RA
Right-anterior
RP
Right-posterior
STJ
Sinotubular junction
UAV
Unicuspid aortic valve
WSS
Wall shear stress

Background

Bicuspid (BAV) and unicuspid (UAV) aortic valve malformations represent two forms of adult congenital heart disease with a prevalence of 1–2 and 0.02%, respectively [14]. BAV and UAV are associated with aortic valve dysfunction (regurgitation and/or stenosis), dissection, and proximal aortic dilatation, the so-called bicuspid aortopathy [5]. Genetic and hemodynamic factors contribute to the progression of BAV and UAV disease [6, 7] and play a role in the development of bicuspid aortopathy, including the increased risk for aortic dissection [5, 8].
Surgical treatment for regurgitation and/or stenosis, particularly aortic valve repair techniques, underwent major development during the last decades. Surgical repair is a promising alternative to prosthetic aortic valve replacement, especially in young patients [2, 911]. Aortic valve repair has several advantages compared to aortic valve replacement, including absence of the need for chronic anticoagulation, lower infection rate, and better hemodynamic performance [12]. From a technical point of view, aortic valve repair is a two-component surgery, consisting of cusp repair and aortic valve annulus stabilization (Fig. 1) [2]. However, recurrent aortic valve regurgitation is still a major issue in patients after aortic valve repair as compared to those after surgical aortic valve replacement [1214]. Therefore, the development of more durable aortic valve repair techniques remains an important clinical challenge. In this context, a comprehensive marker for accurate assessment of changes in hemodynamics after aortic valve repair is needed to evaluate surgical success.
Four-dimensional (4D)-flow cardiovascular magnetic resonance imaging (CMR) has been successfully used to visualize abnormal hemodynamic flow patterns such as helical and vortical flow [1518], wall shear stress [1923], and flow displacement (indicator of outflow asymmetry) [2427] in untreated adult congenital heart disease and after aortic valve replacement surgery [2831]. We hypothesize that 4D flow CMR might be a comprehensive tool to monitor aortic valve competence and hemodynamic changes after aortic valve repair. Therefore, the aim of this study was to demonstrate the feasibility of 4D flow CMR to assess the impact of aortic valve repair on changes in blood flow dynamics in adult congenital heart disease patients with symptomatic BAV or UAV.

Methods

Patients

This prospective study was approved by the local ethics board. Written informed consent was obtained from all patients.
Patients with adult congenital heart disease (BAV or UAV) and symptomatic, predominant aortic regurgitation who were referred for minimally invasive aortic valve repair between April 2017 and February 2019 were included in the study. Patients with contraindications for CMR or younger than 18 years were excluded.
Diagnosis of aortic regurgitation was based on transthoracic echocardiography. Echocardiography was also used to assess the quality of aortic cusp tissue and absence of severe calcifications prior to surgery. In case of intraoperative findings like severe cusp calcifications or fenestrations, aortic valve repair was not performed and patients were not included in this study. The procedure of aortic valve repair has been previously described [2].

Cardiovascular magnetic resonance imaging

All patients underwent non-contrast 4D flow CMR of the thoracic aorta on a 3 T system (Ingenia, Philips Healthcare, Best, The Netherlands) with a 32-channel body-phased array coil before and after surgery. Respiratory gated and cardiac triggered 4D flow CMR data were acquired over the entire cardiac cycle with full volumetric coverage of the thoracic aorta. Scan parameters included: velocity encoding 200 cm/s, temporal resolution 24–38 ms, acquired spatial resolution 2.5 × 2.5 × 2.5 mm3, field of view (280–330) x (280–330) x (50–66) mm3, flip angle = 8°. Parallel imaging (k-t BLAST) with an acceleration factor of 4 was used. Scan time for each acquisition was 13.9 ± 3.0 min, depending on heart rate, respiratory pattern, and efficiency of respiratory gating.
Electrocardiogram (ECG)-gated CMR imaging was performed for assessment of left ventricular (LV) volumes and function by using balanced steady-state free-precession (bSSFP) cine CMR [32]. ECG-gated bSSFP imaging of the thoracic aorta was performed for assessment of aortic diameters [33, 34].

4D flow CMR data analysis

4D flow data were corrected for Maxwell terms, eddy currents and phase aliasing in accordance with current consensus recommendations [35]. All data sets were automatically reconstructed to 24 time frames per cardiac cycle and used to render three-dimensional phase-contrast CMR angiograms in a 3D visualization software (GTFlow, GyroTools LLC, Zurich, Switzerland).
One radiologist with 4 years of experience in the assessment of 4D flow data manually placed analysis planes at six defined anatomic landmarks in the thoracic aorta at the level of the aortic valve, the sinotubular junction, the mid-ascending aorta, the aortic arch proximal to the brachiocephalic trunc, the aortic arch distal to the left subclavian artery and the proximal descending aorta [36].
The same radiologist quantified peak velocity as well as forward and backward flow volumes at the aortic valve level and the regurgitant fraction (%) was calculated [37, 38].
Helical and vortical blood flow patterns in the ascending aorta (AAo), the aortic arch (AA), and the descending aorta (DAo) were semiquantitatively evaluated according to a 3-point scale: 0 (none), 1 (< 360°), and 2 (> 360°). A helical flow pattern was defined as a regional spiral movement along the blood flow direction and a vortical flow pattern as a regional circular movement deviating from the physiological flow-direction by > 90° [39, 40].
Flow displacement, a marker to quantify outflow asymmetry, was automatically quantified by exporting defined analyses planes into MATLAB (The MathWorks, Natick, Massachusetts, USA). Flow displacement was defined as the distance from the vessel centroid to the velocity-weighted centroid of the upper 15% of peak systolic forward flow velocity normalized to the vessel diameter, similar to the strategy reported by Sigovan et al. and Mahadevia et al [24, 26]
WSS, a time-resolved three-dimensional force, was quantified in GTFlow (Gyrotools). Magnitudinal WSS, which is the resulting net vector along the entire vascular wall, was derived from each analysis plane at peak systole [23, 41]. Values for peak systolic WSS were averaged over the five cardiac time frames centered on peak systole to reduce measurement noise [39]. Averaged circumferential WSS was assessed for each plane as well as segmental WSS at 8 standardized local anatomic orientations of the vessel wall: anterior (A), left-anterior (LA), left (L), left-posterior (LP), posterior (P), right-posterior (RP), right (R), and right-anterior (RA) [21, 26, 40].

Statistical analysis

The D’Agostino-Pearson test was used to evaluate whether parameters were normally distributed. Data before and after surgery were compared by a two-sided paired t-test if normally distributed and by Wilcoxon matched-pairs signed-rank test if non-normally distributed. Patients were further divided into subgroups: group 1: BAV type 1 L/R; group 2: UAV. All p values < 0.05 were considered statistically significant. All continuous data are presented as mean ± standard deviation. To account for skewed data, median and interquartile ranges were calculated when appropriate. Statistical analyses were performed using GraphPad Prism 7 (GraphPad Software, San Diego, California, USA).

Results

Study sample

4D flow CMR before and after aortic valve repair was successfully completed in 20 patients with adult congenital heart disease and symptomatic aortic valve regurgitation. Three other patients had to be excluded from the main analyses. In one patient, the dilated aortic root was only partially covered by the 3D volume of the 4D flow sequence. Two patients required aortic valve replacement instead of repair and therefore did not undergo the second CMR examination.
Median age of the 20 included patients at the time of surgery was 35 years (IQR 29–47). Median time between CMR examinations was 9 days (IQR 6–38). There were no relevant medications changed between CMR imaging examinations.
Fifteen patients (75%) had a BAV Sievers Type 1 phenotype with L/R cusp fusion (Fig. 2) and five patients (25%) presented with a UAV phenotype (Fig. 3) (Table 1). Seventeen patients (85%) underwent isolated aortic valve repair, while three patients (15%) needed additional aortic root remodeling [42]. After surgery, there was a significant reduction of mean aortic diameters at the level of the anulus (2.6 ± 0.4 vs. 2.3 ± 0.2 cm, p < 0.001), which is explained by suture annuloplasty aiming to reduce the basal ring diameter. The average diameter at the level of the bulbus aortae (3.8 ± 0.6 vs. 3.5 ± 0.4 cm, p = 0.03) and the mid-ascending aorta (3.3 ± 0.6 vs. 3.1 ± 0.4 cm, p = 0.03) was also significantly reduced, which is explained by the three patients with aortic root remodeling with Dacron prostheses. After surgery there was a significant reduction in stroke volume (120 ± 34 vs. 90 ± 28 ml, p < 0.001). There was no significant difference in LVEF after surgery (57.8 ± 7.8% vs. 58.7 ± 7.2%, p = 0.252). Patient characteristics of all patients and subgroup analyses of BAV and UAV patients including aortic diameters and cardiac function metrics before and after surgery are detailed in Table 1.
Table 1
Patient characteristics and aortic diameters before and after aortic valve repair
Characteristic
All patients n = 20
BAV (type L/R) n = 15
UAV n = 5
Age at time of operation (y) median
35 (IQR 29-47)
42 (IQR 30-47)
30 (IQR 24-45)
Gender, male
16
13
3
Days between imaging median
9 (IQR 6-38)
7 (IQR 6-42)
13 (IQR 6-66)
Procedures
 Isolated aortic valve repair
17
13
4
 Aortic root replacement (Yacoub)
3
2
1
 
pre OP
post OP
Pvalue
pre OP
post OP
Pvalue
pre OP
post OP
Pvalue
Heart rate (bpm)
77.8 ± 9.4
88.0 ± 14.4
0.021
79.0 ± 9.4
87.3 ± 12.1
ns
74.0 ± 9.3
90.2 ± 21.4
ns
LV parameters
 EDV (ml)
208.6 ± 71.9
149.1 ± 48.0
<0.001
223.1 ± 61.4
153.5 ± 36.9
<0.001
170.8 ± 81.9
157.0 ± 70.7
ns
 ESV (ml)
86.3 ± 45.8
64.7 ± 28.2
<0.001
100.0 ± 39.6
68.1 ± 24.9
<0.001
56.8 ± 39.3
52.5 ± 39.5
ns
 SV (ml)
119.6 ± 34.0
89.9 ± 27.9
<0.001
124.1 ± 32.9
88.4 ± 26.3
<0.001
98.3 ± 38.4
93.0 ± 36.4
ns
 EF (%)
57.8 ± 7.8
58.7 ± 7.2
ns
56.3 ± 7.7
57.5 ± 7.2
ns
63.3 ± 6.6
61.4 ± 6.2
ns
Dimensions (mm)
 Anulus
26.3 ± 3.7
23.3 ± 2.4
<0.001
28.9 ± 3.6
23.4 ± 2.0
<0.001
24.8 ± 3.8
23.0 ± 3.4
0.009
 Bulbus aortae
37.6 ± 5.5
35.2 ± 4.1
0.031
38.1 ± 6.0
35.0 ± 4.2
ns
36.0 ± 4.2
35.8 ± 4.1
ns
 Mid-ascending aorta
33.1 ± 6.4
30.6 ± 4.1
0.033
33.4 ± 5.6
31.5 ± 4.2
0.023
32.2 ± 9.4
27.8 ± 2.4
ns
 Proximal aortic arch
26.9 ± 4.0
26.2 ± 3.5
ns
27.5 ± 4.3
26.5 ± 3.7
ns
25.2 ± 2.8
25.0 ± 2.7
ns
 Distal aortic arch
22.2 ± 2.7
22.1 ± 2.6
ns
22.8 ± 2.4
22.7 ± 2.4
ns
20.4 ± 2.8
20.4 ± 2.9
ns
 Descending aorta
23.5 ± 4.0
23.2 ± 3.9
ns
24.0 ± 3.4
23.6 ± 3.4
ns
22.0 ± 5.6
22.0 ± 5.6
ns

Quantification and visualization of blood flow

4D flow CMR-derived blood flow quantification revealed that aortic valve repair resulted in a significant reduction of forward flow, backward flow, and net flow volume at the aortic valve level (all p < 0.001) (Table 2). The improved aortic root geometry also resulted in a significantly reduced regurgitant fraction (17 ± 10% vs. 6 ± 4%, p < 0.001) (Fig. 2b and e, Fig. 3b and e) and systolic peak velocity (2.2 ± 0.9 vs. 1.9 ± 0.4 m/s, p = 0.014) (Table 2).
Table 2
Comparison of 4D flow CMR-derived flow volumes and flow patterns before and after aortic valve repair
Measure
pre OP
post OP
p value
Forward flow (ml)
120 ± 44
80 ± 24
< 0.001
Backward flow (ml)
21 ± 17
5 ± 3
< 0.001
Net flow (ml)
99 ± 36
75 ± 23
< 0.001
Peak systolic velocity (cm/s)
225 ± 85
192 ± 40
0.014
Regurgitant fraction (%)
17 ± 10
6 ± 4
< 0.001
Helix grade
 Ascending aorta
1.6 ± 0.6
0.9 ± 0.5
< 0.001
 Aortic arch
0.5 ± 0.7
0.3 ± 0.5
ns
 Descending aorta
0.3 ± 0.6
0.1 ± 0.4
ns
Vortex grade
 Ascending aorta
1.2 ± 0.8
0.5 ± 0.6
0.002
 Aortic arch
0.2 ± 0.4
0.0 ± 0.0
ns
 Descending aorta
0.3 ± 0.6
0.2 ± 0.4
ns
Values represent mean ± SD. p < 0.05 indicates a statistically significant difference. Significant values are in bold
4D flow CMR allowed visualization of hemodynamic flow patterns such as helical and vortical flow in the thoracic aorta before (Fig. 2a and Fig. 3a) and after aortic valve repair (Fig. 2d and Fig. 3d).
All patients showed common global right-handed helical flow formation in the AAo before and after surgery. Pathologic, secondary local helical flow formations in the AAo were reduced after surgery in 13 of 20 patients (65%) (1.6 ± 0.6 vs. 0.9 ± 0.5, p < 0.001). Helical flow in the aortic arch and DAo was less pronounced, however, these differences where not statistically significant (Table 2).
Vortical flow patterns in the AAo were observed in 15 of 20 patients (75%) before surgery and in 9 of 20 patients (45%) after surgery, resulting in a significant decrease of the average vortical flow (1.2 ± 0.8 vs. 0.5 ± 0.6, p = 0.002). Although vortical flow was less pronounced in the aortic arch and DAo after surgery, these differences were not statistically significant (Table 2).

Flow displacement

4D flow CMR revealed highly eccentric outflow jet patterns directed towards the right-anterior wall of the mid-ascending AAo before surgery (Fig. 2c and Fig. 3c) in all patients and a more centrally located flow profile after aortic valve repair (Fig. 2f and Fig. 3f) in twelve patients (60%) (Fig. 4). Flow displacement was significantly reduced after surgery in the mid-ascending AAo (0.3 ± 0.1 vs. 0.25 ± 0.1, p= 0.031) (Fig. 4). There was no significant change of flow displacement at the level of the sinotubular junction (0.19 ± 0.1 vs. 0.18 ± 0.1, p = 0.759) and the proximal aortic arch (0.3 ± 0.1 vs. 0.29 ± 0.1, p = 0.524).
Subgroup analyses of BAV and UAV patients are detailed in Table 3.
Table 3
Flow displacement subgroups
Group
pre OP
post OP
P value
BAV
 Sinutubular junction
0.20 ± 0.1
0.20 ± 0.1
ns
 Mid-ascending aorta
0.28 ± 0.1
0.26 ± 0.1
ns
 Proximal aortic arch
0.29 ± 0.0
0.27 ± 0.1
ns
UAV
 Sinutubular junction
0.17 ± 0.1
0.10 ± 0.1
ns
 Mid-ascending aorta
0.38 ± 0.1
0.24 ± 0.1
0.026
 Proximal aortic arch
0.19 ± 0.1
0.32 ± 0.4
ns
Values represent mean ± SD. P < 0.05 indicates a statistically significant difference. Significant values are in bold

Circumferential aortic wall shear stress

Circumferential WSS was significantly reduced after aortic valve repair in the mid-ascending AAo (0.8 ± 0.2 vs. 0.5 ± 0.2 N/m2, p < 0.001), proximal aortic arch (0.8 ± 0.4 vs. 0.5 ± 0.2 N/m2, p = 0.002), and distal aortic arch (0.6 ± 0.3 vs. 0.5 ± 0.2 N/m2, P < 0.001) for all patients (Fig. 5a). There was no significant change at the level of the sinotubular junction (0.5 ± 0.2 vs. 0.4 ± 0.2 N/m2, p = 0.233) and the DAo (0.6 ± 0.3 vs. 0.5 ± 0.2 N/m2, p = 0.077). In the mid-ascending AAo, where circumferential WSS was highest, 4D flow CMR revealed decreased WSS postoperatively in 18 of 20 patients (90%) and an increased WSS in two patients (10%).
Subgroup analyses of BAV and UAV patients are detailed in Table 4.
Table 4
Circumferential aortic wall shear stress subgroups
Group
pre OP
post OP
P value
BAV
 Sinutubular junction (N/m2)
0.5 ± 0.2
0.4 ± 0.1
ns
 Mid-ascending aorta (N/m2)
0.8 ± 0.2
0.6 ± 0.2
0.009
 Proximal aortic arch (N/m2)
0.8 ± 0.3
0.6 ± 0.2
0.006
 Distal aortic arch (N/m2)
0.6 ± 0.3
0.5 ± 0.2
0.003
 Descending aorta (N/m2)
0.6 ± 0.3
0.5 ± 0.2
ns
UAV
 Sinutubular junction (N/m2)
0.4 ± 0.2
0.3 ± 0.1
ns
 Mid-ascending aorta (N/m2)
0.9 ± 0.3
0.4 ± 0.1
0.01
 Proximal aortic arch (N/m2)
0.9 ± 0.5
0.4 ± 0.1
ns
 Distal aortic arch (N/m2)
0.6 ± 0.2
0.5 ± 0.2
0.037
 Descending aorta (N/m2)
0.6 ± 0.3
0.5 ± 0.2
ns
Values represent mean ± SD. P < 0.05 indicates a statistically significant difference. Significant values are in bold

Segmental aortic wall shear stress

Segmental WSS analyses along the aortic circumference are illustrated in Fig. 5b. The highest segmental WSS was observed in the mid-ascending AAo and the proximal aortic arch before aortic valve repair.
In the mid-ascending AAo, consistent with the regions of eccentric flow, segmental WSS before aortic valve repair was most pronounced at the anterior, right-anterior, and right aortic wall. After surgery, WSS was significantly reduced in the A, RA, R, P, L, and LA segments.
In the proximal aortic arch, segmental WSS magnitude and asymmetry was most pronounced before aortic valve repair at the anterior and left-anterior aortic wall. After aortic valve repair, WSS was significantly reduced in the A, R, RP, P, LP, L, and LA segments.
At the level of the distal aortic arch, aortic valve repair significantly reduced segmental WSS at the A, RA, and R segments. At the level of the DAo, aortic valve repair significantly reduced segmental WSS at the LA and A segments. At the level of the sinotubular junction, segmental WSS was lowest and without significant changes after aortic valve repair.

Discussion

This study demonstrated the feasibility of 4D flow CMR to evaluate the hemodynamic changes of aortic valve repair in patients with BAV and UAV. 4D flow CMR-derived angiograms and the ability to quantify abnormal flow parameters in the thoracic aorta allowed for both qualitative and quantitative assessment of the hemodynamic changes after surgical reconstruction of aortic root geometry. Our study revealed that the extent of secondary flow patterns, such as vortices and helices, flow eccentricity, as well as global and regional WSS were all significantly reduced in the ascending aorta after aortic valve repair.
Previous studies demonstrated that 4D flow CMR allows visualization of pronounced helical and vortical flow formations in the AAo in untreated patients with adult congenital heart disease [18, 40]. Such secondary flow patterns are caused by the eccentric outflow jet due to the asymmetry of the aortic valve. Bissel et al. [15] and Mahadevia et al. [26] showed that patients with BAV and L/R cusp fusion had highly eccentric outflow jets directed towards the right-anterior wall of the AAo before aortic valve repair. In our patients, the observed flow eccentricity resulted in localized elevated WSS in the mid-ascending AAo and the proximal aortic arch before aortic valve repair, which is in line with results of previous 4D flow CMR studies [21, 22].
Our study focused on changes in systolic outflow after aortic valve repair, as quantified by the extent of systolic peak velocity, aortic regurgitation, secondary flow patterns, flow displacement, and WSS. 4D flow CMR revealed a significantly decreased aortic valve regurgitation as well as reduced flow displacement and regional WSS after aortic valve repair. These 4D flow CMR-derived findings indicate an improved competence of the aortic valve and more symmetrical valve anatomy, thus success of the surgical procedure to normalize flow hemodynamics.
In our study, the recreation of the optimal aortic root geometry, including reduction of basal ring diameter and restoration of effective cusp height had two important hemodynamic effects as demonstrated by 4D flow CMR: First, reduced regurgitation, which consecutively reduces the stroke volume, and secondly, a more centralized outflow due the improved symmetry of the aortic valve. Corresponding to this postoperative hemodynamic improvement, a significant reduction of WSS was observed in the mid-ascending AAo, as well as in the proximal and distal aortic arch. A significant reduction of flow displacement was observed only in the mid-ascending AAo. This discrepancy may be explained by the fact that WSS is not only caused by local eccentric flow jets, but also by the stroke volume, which was significantly reduced after surgical aortic valve repair, thus reducing WSS.
Patients with highest flow displacement and highest WSS in the mid-ascending AAo before aortic valve repair showed the greatest changes in blood flow hemodynamics and were among the patients with lowest flow displacement and lowest WSS after aortic valve repair. This may be explained by the fact that patients with the highest flow displacement had the most asymmetric aortic valve geometry and the most severe aortic regurgitation. These patients had the greatest hemodynamic benefit from symmetrical aortic valve rearrangement correction of symptomatic aortic regurgitation.
Two patients showed an increase in circumferential WSS in the mid-ascending aorta after isolated aortic valve repair with either just a relatively small improvement or even a small increase in flow displacement. The first patient developed severe recurrent aortic regurgitation 6 weeks after surgery and underwent a redo aortic valve repair 2 months after the initial surgery. A tearing of the fused cusp at the commissural level was found intraoperatively. The second patient had an uneventful postoperative course with an excellent echocardiographic result and no residual aortic regurgitation.
Our study has potentially important clinical implications. Adult congenital heart disease, namely BAV and UAV, is associated with aortopathy and an increased risk for aortic dissection, even after aortic valve repair [5, 8]. It remains unclear, how a modified, more physiologic flow after bicuspid aortic valve repair affects the long-term outcome of aortopathy. In this context, 4D flow CMR is a comprehensive tool to monitor not only aortic valve competence but also hemodynamic changes in the thoracic aorta. Particularly 4D flow CMR-derived WSS and flow eccentricity may provide insights into the mechanisms involved in aortopathy formation.
Several studies have successfully investigated 4D flow CMR for the evaluation of hemodynamic changes after aortic valve replacement surgery [28, 43]. However, none of these studies has assessed the impact of the modified transvalvular flow patterns on the progression of aortopathy in prospective longitudinal studies. The ability of 4D flow CMR to illustrate non-physiologic blood flow patterns and measure hemodynamic parameters beyond standard metrics may allow to guide physicians towards an individualized clinical decision approach regarding aortic root replacement in patients with increased risk for progression of aortopathy after aortic valve repair or replacement.
However, we are well aware that our results represent only the first step toward long term studies with outcome metrics. Future prospective and longitudinal studies are needed to assess whether post-operative 4D flow CMR can predict the progression of aortopathy and the risk of late aortic complications as well as patient outcomes.
Our study has several limitations. First, a limited study sample size prevented us from further post-hoc subanalysis and comparison of different BAV morphologies (Sievers type 0, type 1, and type 2) and UAV morphology as well as comparison of patients with and without aortic root replacement. Future 4D flow CMR studies with a larger sample size are needed to compare the surgical impact on blood flow dynamics between patients with these different valvular morphotypes.
Second, a limited spatial resolution of our 4D flow CMR may result in underestimation of WSS [44, 45]. However, our main focus was the comparison of 4D flow CMR-derived parameters before and after aortic valve repair and all scans were performed on the same CMR unit with identical scan parameters. Therefore, relative differences in WSS before and after aortic valve repair remain reliable results.
The third limitation is the manual positioning of 2D analysis planes. We tried to minimize this potential bias by adhering to defined anatomical landmarks for positioning of the analysis planes. Still, 2D analyses may result in limited coverage and distorted quantification of complex WSS distribution when compared to more advanced 3D WSS quantification techniques [20, 27, 39]. However, we do not have access to these advanced and dedicated analyses techniques. Instead, we aimed to validate the clinical utility of currently available 4D flow CMR analyses techniques in the setting of adult congenital heart disease. Therefore, we focused on analyses methods that are established in our department and widely accepted in the 4D flow CMR community, and therefore also more likely available to other clinical sites.

Conclusions

4D flow CMR allows assessment of the impact of aortic valve repair on changes in blood flow dynamics in patients with bicuspid aortic valve disease. Further prospective studies are required to determine whether 4D flow CMR can play an important clinical role, such as predicting progression of aortopathy and long-term risk of late aortic complications as well as patient outcomes.

Acknowledgements

Not applicable.
This study was approved by the local ethics board University Medical Center Hamburg-Eppendorf. Written informed consent was obtained from all patients.
All participants in this study gave written consent to participate and to publish.

Competing interests

No financial support for this study was provided by industry. One coauthor (HK) is an employee of Koninklijke Philips NV but had no control of the data.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
Zurück zum Zitat Sievers H-H, Schmidtke C. A classification system for the bicuspid aortic valve from 304 surgical specimens. J Thorac Cardiovasc Surg. 2007;133:1226–33.PubMed Sievers H-H, Schmidtke C. A classification system for the bicuspid aortic valve from 304 surgical specimens. J Thorac Cardiovasc Surg. 2007;133:1226–33.PubMed
2.
Zurück zum Zitat Girdauskas E, Petersen J, Sachweh J, Kozlik-Feldmann R, Sinning C, Rickers C, et al. Aortic valve repair in adult congenital heart disease. Cardiovasc Diagn Ther. 2018;8:789–98.PubMedPubMedCentral Girdauskas E, Petersen J, Sachweh J, Kozlik-Feldmann R, Sinning C, Rickers C, et al. Aortic valve repair in adult congenital heart disease. Cardiovasc Diagn Ther. 2018;8:789–98.PubMedPubMedCentral
3.
Zurück zum Zitat Wu MH, Lu CW, Chen HC, Kao FY, Huang SK. Adult congenital heart disease in a Nationwide population 2000–2014: epidemiological trends, arrhythmia, and standardized mortality ratio. J Am Heart Assoc. 2018;7:1233–10. Wu MH, Lu CW, Chen HC, Kao FY, Huang SK. Adult congenital heart disease in a Nationwide population 2000–2014: epidemiological trends, arrhythmia, and standardized mortality ratio. J Am Heart Assoc. 2018;7:1233–10.
4.
Zurück zum Zitat Weinrich JM, Lenz A, Girdauskas E, Adam G, von Kodolitsch Y, Bannas P. Current and emerging imaging techniques in patients with genetic aortic syndromes. Rofo. 2019;192:50–8. Weinrich JM, Lenz A, Girdauskas E, Adam G, von Kodolitsch Y, Bannas P. Current and emerging imaging techniques in patients with genetic aortic syndromes. Rofo. 2019;192:50–8.
5.
Zurück zum Zitat Girdauskas E, Borger MA. Bicuspid aortic valve and associated aortopathy: an update. Semin Thorac Cardiovasc Surg. 2013;25:310–6.PubMed Girdauskas E, Borger MA. Bicuspid aortic valve and associated aortopathy: an update. Semin Thorac Cardiovasc Surg. 2013;25:310–6.PubMed
6.
Zurück zum Zitat Girdauskas E, Rouman M, Disha K, Fey B, Dubslaff G, Theis B, et al. Functional aortic root parameters and expression of aortopathy in bicuspid versus tricuspid aortic valve stenosis. J Am Coll Cardiol. 2016;67:1786–96.PubMed Girdauskas E, Rouman M, Disha K, Fey B, Dubslaff G, Theis B, et al. Functional aortic root parameters and expression of aortopathy in bicuspid versus tricuspid aortic valve stenosis. J Am Coll Cardiol. 2016;67:1786–96.PubMed
7.
Zurück zum Zitat Grewal N, Girdauskas E, DeRuiter M, Goumans M-J, Poelmann RE, Klautz RJM, et al. The role of hemodynamics in bicuspid aortopathy: a histopathologic study. Cardiovasc Pathol. 2019;41:1–9. Grewal N, Girdauskas E, DeRuiter M, Goumans M-J, Poelmann RE, Klautz RJM, et al. The role of hemodynamics in bicuspid aortopathy: a histopathologic study. Cardiovasc Pathol. 2019;41:1–9.
8.
Zurück zum Zitat Girdauskas E, Rouman M, Disha K, Espinoza A, Misfeld M, Borger MA, et al. Aortic dissection after previous aortic valve replacement for bicuspid aortic valve disease. J Am Coll Cardiol. 2015;66:1409–11.PubMed Girdauskas E, Rouman M, Disha K, Espinoza A, Misfeld M, Borger MA, et al. Aortic dissection after previous aortic valve replacement for bicuspid aortic valve disease. J Am Coll Cardiol. 2015;66:1409–11.PubMed
9.
Zurück zum Zitat Boodhwani M, De Kerchove L, Glineur D, Rubay J, Vanoverschelde JL, Noirhomme P, et al. Repair of regurgitant bicuspid aortic valves: a systematic approach. J Thorac Cardiovasc Surg. 2010;140:276–84 e1.PubMed Boodhwani M, De Kerchove L, Glineur D, Rubay J, Vanoverschelde JL, Noirhomme P, et al. Repair of regurgitant bicuspid aortic valves: a systematic approach. J Thorac Cardiovasc Surg. 2010;140:276–84 e1.PubMed
10.
Zurück zum Zitat Vohra HA, Whistance RN, De Kerchove L, Punjabi P, EL Khoury G. Valve-preserving surgery on the bicuspid aortic valve. Eur J Cardiothorac Surg. 2013;43:888–98.PubMed Vohra HA, Whistance RN, De Kerchove L, Punjabi P, EL Khoury G. Valve-preserving surgery on the bicuspid aortic valve. Eur J Cardiothorac Surg. 2013;43:888–98.PubMed
11.
Zurück zum Zitat Borger MA, Fedak PWM, Stehens EH, Gleason TG, Girdauskas E, Ikonomidis JS, et al. The American Association for Thoracic Surgery consensus guidelines on bicuspid aortic valve-related aortopathy. J Thorac Cardiovasc Surg. 2018;156:e41–74.PubMedPubMedCentral Borger MA, Fedak PWM, Stehens EH, Gleason TG, Girdauskas E, Ikonomidis JS, et al. The American Association for Thoracic Surgery consensus guidelines on bicuspid aortic valve-related aortopathy. J Thorac Cardiovasc Surg. 2018;156:e41–74.PubMedPubMedCentral
12.
Zurück zum Zitat Aicher D, Fries R, Rodionycheva S, Schmidt K, Langer F, Schäfers H-J. Aortic valve repair leads to a low incidence of valve-related complications. Eur J Cardiothorac Surg. 2010;37:127–32.PubMed Aicher D, Fries R, Rodionycheva S, Schmidt K, Langer F, Schäfers H-J. Aortic valve repair leads to a low incidence of valve-related complications. Eur J Cardiothorac Surg. 2010;37:127–32.PubMed
13.
Zurück zum Zitat Arnaoutakis GJ, Sultan I, Siki M, Bavaria JE. Bicuspid aortic valve repair: systematic review on long-term outcomes. Ann Cardiothorac Surg. 2019;8:302–12.PubMedPubMedCentral Arnaoutakis GJ, Sultan I, Siki M, Bavaria JE. Bicuspid aortic valve repair: systematic review on long-term outcomes. Ann Cardiothorac Surg. 2019;8:302–12.PubMedPubMedCentral
14.
Zurück zum Zitat Svensson LG, Kindi Al AH, Vivacqua A, Pettersson GB, Gillinov AM, Mihaljevic T, et al. Long-term durability of bicuspid aortic valve repair. Ann Thorac Surg. 2014;97:1539–47 discussion 1548.PubMed Svensson LG, Kindi Al AH, Vivacqua A, Pettersson GB, Gillinov AM, Mihaljevic T, et al. Long-term durability of bicuspid aortic valve repair. Ann Thorac Surg. 2014;97:1539–47 discussion 1548.PubMed
15.
Zurück zum Zitat Bissell MM, Hess AT, Biasiolli L, Glaze SJ, Loudon M, Pitcher A, et al. Aortic dilation in bicuspid aortic valve disease. Circ Cardiovasc Imaging. 2013;6:499–507.PubMed Bissell MM, Hess AT, Biasiolli L, Glaze SJ, Loudon M, Pitcher A, et al. Aortic dilation in bicuspid aortic valve disease. Circ Cardiovasc Imaging. 2013;6:499–507.PubMed
16.
Zurück zum Zitat Schnell S, Smith DA, Barker AJ, Entezari P, Honarmand AR, Carr ML, et al. Altered aortic shape in bicuspid aortic valve relatives influences blood flow patterns. Eur Heart J Cardiovasc Imaging. 2016;17:1239–47.PubMedPubMedCentral Schnell S, Smith DA, Barker AJ, Entezari P, Honarmand AR, Carr ML, et al. Altered aortic shape in bicuspid aortic valve relatives influences blood flow patterns. Eur Heart J Cardiovasc Imaging. 2016;17:1239–47.PubMedPubMedCentral
17.
Zurück zum Zitat Bannas P, Lenz A, Petersen J, Sinn M, Adam G, Reichenspurner H, et al. Normalization of transvalvular flow patterns after bicuspid aortic valve repair: insights from 4D flow cardiovascular magnetic resonance imaging. Ann Thorac Surg. 2018;106(6):e319–20.PubMed Bannas P, Lenz A, Petersen J, Sinn M, Adam G, Reichenspurner H, et al. Normalization of transvalvular flow patterns after bicuspid aortic valve repair: insights from 4D flow cardiovascular magnetic resonance imaging. Ann Thorac Surg. 2018;106(6):e319–20.PubMed
18.
Zurück zum Zitat Hope MD, Hope TA, Meadows AK, Ordovas KG, Urbania TH, Alley MT, et al. Bicuspid aortic valve: four-dimensional MR evaluation of ascending aortic systolic flow patterns. Radiology. 2010;255:53–61.PubMed Hope MD, Hope TA, Meadows AK, Ordovas KG, Urbania TH, Alley MT, et al. Bicuspid aortic valve: four-dimensional MR evaluation of ascending aortic systolic flow patterns. Radiology. 2010;255:53–61.PubMed
19.
Zurück zum Zitat Malek AM, Jackman R, Rosenberg RD, Izumo S. Endothelial expression of thrombomodulin is reversibly regulated by fluid shear stress. Circ Res. 1994;74:852–60.PubMed Malek AM, Jackman R, Rosenberg RD, Izumo S. Endothelial expression of thrombomodulin is reversibly regulated by fluid shear stress. Circ Res. 1994;74:852–60.PubMed
20.
Zurück zum Zitat van Ooij P, Garcia J, Potters WV, Malaisrie SC, Collins JD, Carr JC, et al. Age-related changes in aortic 3D blood flow velocities and wall shear stress: implications for the identification of altered hemodynamics in patients with aortic valve disease. J Magn Reson Imaging. 2015;43:1239–49.PubMedPubMedCentral van Ooij P, Garcia J, Potters WV, Malaisrie SC, Collins JD, Carr JC, et al. Age-related changes in aortic 3D blood flow velocities and wall shear stress: implications for the identification of altered hemodynamics in patients with aortic valve disease. J Magn Reson Imaging. 2015;43:1239–49.PubMedPubMedCentral
21.
Zurück zum Zitat Barker AJ, Markl M, Bürk J, Lorenz R, Bock J, Bauer S, et al. Bicuspid aortic valve is associated with altered wall shear stress in the ascending aorta. Circ Cardiovasc Imaging. 2012;5:457–66.PubMed Barker AJ, Markl M, Bürk J, Lorenz R, Bock J, Bauer S, et al. Bicuspid aortic valve is associated with altered wall shear stress in the ascending aorta. Circ Cardiovasc Imaging. 2012;5:457–66.PubMed
22.
Zurück zum Zitat Farag ES, van Ooij P, Planken RN, Dukker KCP, de Heer F, Bouma BJ, et al. Aortic valve stenosis and aortic diameters determine the extent of increased wall shear stress in bicuspid aortic valve disease. J Magn Reson Imaging. 2018;48:522–30.PubMedPubMedCentral Farag ES, van Ooij P, Planken RN, Dukker KCP, de Heer F, Bouma BJ, et al. Aortic valve stenosis and aortic diameters determine the extent of increased wall shear stress in bicuspid aortic valve disease. J Magn Reson Imaging. 2018;48:522–30.PubMedPubMedCentral
23.
Zurück zum Zitat Meierhofer C, Schneider EP, Lyko C, Hutter A, Martinoff S, Markl M, et al. Wall shear stress and flow patterns in the ascending aorta in patients with bicuspid aortic valves differ significantly from tricuspid aortic valves: a prospective study. Eur Heart J Cardiovasc Imaging. 2013;14:797–804.PubMed Meierhofer C, Schneider EP, Lyko C, Hutter A, Martinoff S, Markl M, et al. Wall shear stress and flow patterns in the ascending aorta in patients with bicuspid aortic valves differ significantly from tricuspid aortic valves: a prospective study. Eur Heart J Cardiovasc Imaging. 2013;14:797–804.PubMed
24.
Zurück zum Zitat Sigovan M, Hope MD, Dyverfeldt P, Saloner D. Comparison of four-dimensional flow parameters for quantification of flow eccentricity in the ascending aorta. J Magn Reson Imaging. 2011;34:1226–30.PubMed Sigovan M, Hope MD, Dyverfeldt P, Saloner D. Comparison of four-dimensional flow parameters for quantification of flow eccentricity in the ascending aorta. J Magn Reson Imaging. 2011;34:1226–30.PubMed
25.
Zurück zum Zitat Burris NS, Sigovan M, Knauer HA, Tseng EE, Saloner D, Hope MD. Systolic flow displacement correlates with future ascending aortic growth in patients with bicuspid aortic valves undergoing magnetic resonance surveillance. Invest Radiol. 2014;49:635–9.PubMed Burris NS, Sigovan M, Knauer HA, Tseng EE, Saloner D, Hope MD. Systolic flow displacement correlates with future ascending aortic growth in patients with bicuspid aortic valves undergoing magnetic resonance surveillance. Invest Radiol. 2014;49:635–9.PubMed
26.
Zurück zum Zitat Mahadevia R, Barker AJ, Schnell S, Entezari P, Kansal P, Fedak PWM, et al. Bicuspid aortic cusp fusion morphology alters aortic three-dimensional outflow patterns, wall shear stress, and expression of aortopathy. Circulation. 2014;129:673–82.PubMed Mahadevia R, Barker AJ, Schnell S, Entezari P, Kansal P, Fedak PWM, et al. Bicuspid aortic cusp fusion morphology alters aortic three-dimensional outflow patterns, wall shear stress, and expression of aortopathy. Circulation. 2014;129:673–82.PubMed
27.
Zurück zum Zitat Guzzardi DG, Barker AJ, van Ooij P, Malaisrie SC, Puthumana JJ, Belke DD, et al. Valve-related hemodynamics mediate human bicuspid aortopathy: insights from wall shear stress mapping. J Am Coll Cardiol. 2015;66:892–900.PubMedPubMedCentral Guzzardi DG, Barker AJ, van Ooij P, Malaisrie SC, Puthumana JJ, Belke DD, et al. Valve-related hemodynamics mediate human bicuspid aortopathy: insights from wall shear stress mapping. J Am Coll Cardiol. 2015;66:892–900.PubMedPubMedCentral
28.
Zurück zum Zitat Bissell MM, Loudon M, Hess AT, Stoll V, Orchard E, Neubauer S, et al. Differential flow improvements after valve replacements in bicuspid aortic valve disease: a cardiovascular magnetic resonance assessment. J Cardiovasc Magn Reson. 2018;20:1–10. Bissell MM, Loudon M, Hess AT, Stoll V, Orchard E, Neubauer S, et al. Differential flow improvements after valve replacements in bicuspid aortic valve disease: a cardiovascular magnetic resonance assessment. J Cardiovasc Magn Reson. 2018;20:1–10.
29.
Zurück zum Zitat Galea N, Piatti F, Sturla F, Weinsaft JW, Lau C, Chirichilli I, et al. Novel insights by 4D flow imaging on aortic flow physiology after valve-sparing root replacement with or without neosinuses. Interact Cardiovasc Thorac Surg. 2018;26:957–64.PubMed Galea N, Piatti F, Sturla F, Weinsaft JW, Lau C, Chirichilli I, et al. Novel insights by 4D flow imaging on aortic flow physiology after valve-sparing root replacement with or without neosinuses. Interact Cardiovasc Thorac Surg. 2018;26:957–64.PubMed
30.
Zurück zum Zitat Semaan E, Markl M, Chris Malaisrie S, Barker A, Allen B, McCarthy P, et al. Haemodynamic outcome at four-dimensional flow magnetic resonance imaging following valve-sparing aortic root replacement with tricuspid and bicuspid valve morphology. Eur J Cardiothorac Surg. 2014;45:818–25.PubMed Semaan E, Markl M, Chris Malaisrie S, Barker A, Allen B, McCarthy P, et al. Haemodynamic outcome at four-dimensional flow magnetic resonance imaging following valve-sparing aortic root replacement with tricuspid and bicuspid valve morphology. Eur J Cardiothorac Surg. 2014;45:818–25.PubMed
31.
Zurück zum Zitat Stephens EH, Hope TA, Kari FA, Kvitting JP, Liang DH, Herfkens RJ, et al. Greater asymmetric wall shear stress in Sievers' type 1/LR compared with 0/LA bicuspid aortic valves after valve-sparing aortic root replacement. J Thorac Cardiovasc Surg. 2015;150:59–68.PubMed Stephens EH, Hope TA, Kari FA, Kvitting JP, Liang DH, Herfkens RJ, et al. Greater asymmetric wall shear stress in Sievers' type 1/LR compared with 0/LA bicuspid aortic valves after valve-sparing aortic root replacement. J Thorac Cardiovasc Surg. 2015;150:59–68.PubMed
32.
Zurück zum Zitat Hundley WG, Bluemke DA, Finn JP, Flamm SD, Fogel MA, Friedrich MG, et al. ACCF/ACR/AHA/NASCI/SCMR 2010 expert consensus document on cardiovascular magnetic resonance. J Am Coll Cardiol. 2010;55:2614–62.PubMed Hundley WG, Bluemke DA, Finn JP, Flamm SD, Fogel MA, Friedrich MG, et al. ACCF/ACR/AHA/NASCI/SCMR 2010 expert consensus document on cardiovascular magnetic resonance. J Am Coll Cardiol. 2010;55:2614–62.PubMed
33.
Zurück zum Zitat Veldhoen S, Behzadi C, Derlin T, Rybczinsky M, von Kodolitsch Y, Sheikhzadeh S, et al. Exact monitoring of aortic diameters in Marfan patients without gadolinium contrast: intraindividual comparison of 2D SSFP imaging with 3D CE-MRA and echocardiography. Eur Radiol. 2014;25:872–82.PubMed Veldhoen S, Behzadi C, Derlin T, Rybczinsky M, von Kodolitsch Y, Sheikhzadeh S, et al. Exact monitoring of aortic diameters in Marfan patients without gadolinium contrast: intraindividual comparison of 2D SSFP imaging with 3D CE-MRA and echocardiography. Eur Radiol. 2014;25:872–82.PubMed
34.
Zurück zum Zitat Veldhoen S, Behzadi C, Lenz A, Henes FO, Rybczynski M, von Kodolitsch Y, et al. Non-contrast MR angiography at 1.5 tesla for aortic monitoring in Marfan patients after aortic root surgery. J Cardiovasc Magn Reson. 2017;19:1–10. Veldhoen S, Behzadi C, Lenz A, Henes FO, Rybczynski M, von Kodolitsch Y, et al. Non-contrast MR angiography at 1.5 tesla for aortic monitoring in Marfan patients after aortic root surgery. J Cardiovasc Magn Reson. 2017;19:1–10.
35.
Zurück zum Zitat Dyverfeldt P, Bissell M, Barker AJ, Bolger AF, Carlhäll C-J, Ebbers T, et al. 4D flow cardiovascular magnetic resonance consensus statement. J Cardiovasc Magn Reson. 2015;17:72.PubMedPubMedCentral Dyverfeldt P, Bissell M, Barker AJ, Bolger AF, Carlhäll C-J, Ebbers T, et al. 4D flow cardiovascular magnetic resonance consensus statement. J Cardiovasc Magn Reson. 2015;17:72.PubMedPubMedCentral
36.
Zurück zum Zitat Hiratzka LF, Bakris GL, Beckman JA, Bersin RM, Carr VF, Casey DE Jr, et al. 2010 ACCF/AHA/AATS/ACR/ASA/SCA/SCAI/SIR/STS/SVM guidelines for the diagnosis and management of patients with thoracic aortic disease. Circulation. 2010;121:818–104. Hiratzka LF, Bakris GL, Beckman JA, Bersin RM, Carr VF, Casey DE Jr, et al. 2010 ACCF/AHA/AATS/ACR/ASA/SCA/SCAI/SIR/STS/SVM guidelines for the diagnosis and management of patients with thoracic aortic disease. Circulation. 2010;121:818–104.
37.
Zurück zum Zitat Cawley PJ, Hamilton-Craig C, Owens DS, Krieger EV, Strugnell WE, Mitsumori L, et al. Prospective comparison of valve regurgitation quantitation by cardiac magnetic resonance imaging and transthoracic echocardiography. Circ Cardiovasc Imaging. 2013;6:48–57.PubMed Cawley PJ, Hamilton-Craig C, Owens DS, Krieger EV, Strugnell WE, Mitsumori L, et al. Prospective comparison of valve regurgitation quantitation by cardiac magnetic resonance imaging and transthoracic echocardiography. Circ Cardiovasc Imaging. 2013;6:48–57.PubMed
38.
Zurück zum Zitat Feneis JF, Kyubwa E, Atianzar K, Cheng JY, Alley MT, Vasanawala SS, et al. 4D flow MRI quantification of mitral and tricuspid regurgitation: reproducibility and consistency relative to conventional MRI. J Magn Reson Imaging. 2018;48:1147–58.PubMed Feneis JF, Kyubwa E, Atianzar K, Cheng JY, Alley MT, Vasanawala SS, et al. 4D flow MRI quantification of mitral and tricuspid regurgitation: reproducibility and consistency relative to conventional MRI. J Magn Reson Imaging. 2018;48:1147–58.PubMed
39.
Zurück zum Zitat Geiger J, Hirtler D, Gottfried K, Rahman O, Bollache E, Barker AJ, et al. Longitudinal evaluation of aortic hemodynamics in Marfan syndrome: new insights from a 4D flow cardiovascular magnetic resonance multi-year follow-up study. J Cardiovasc Magn Reson. 2017;19:33.PubMedPubMedCentral Geiger J, Hirtler D, Gottfried K, Rahman O, Bollache E, Barker AJ, et al. Longitudinal evaluation of aortic hemodynamics in Marfan syndrome: new insights from a 4D flow cardiovascular magnetic resonance multi-year follow-up study. J Cardiovasc Magn Reson. 2017;19:33.PubMedPubMedCentral
40.
Zurück zum Zitat von Knobelsdorff-Brenkenhoff F, Karunaharamoorthy A, Trauzeddel RF, Barker AJ, Blaszczyk E, Markl M, et al. Evaluation of aortic blood flow and wall shear stress in aortic stenosis and its association with left ventricular remodeling. Circ Cardiovasc Imaging. 2016;9:e004038. von Knobelsdorff-Brenkenhoff F, Karunaharamoorthy A, Trauzeddel RF, Barker AJ, Blaszczyk E, Markl M, et al. Evaluation of aortic blood flow and wall shear stress in aortic stenosis and its association with left ventricular remodeling. Circ Cardiovasc Imaging. 2016;9:e004038.
41.
Zurück zum Zitat Stalder AF, Russe MF, Frydrychowicz A, Bock J, Hennig J, Markl M. Quantitative 2D and 3D phase contrast MRI: optimized analysis of blood flow and vessel wall parameters. Magn Reson Med. 2008;60:1218–31.PubMed Stalder AF, Russe MF, Frydrychowicz A, Bock J, Hennig J, Markl M. Quantitative 2D and 3D phase contrast MRI: optimized analysis of blood flow and vessel wall parameters. Magn Reson Med. 2008;60:1218–31.PubMed
42.
Zurück zum Zitat Sarsam MA, Yacoub M. Remodeling of the aortic valve anulus. J Thorac Cardiovasc Surg. 1993;105:435–8.PubMed Sarsam MA, Yacoub M. Remodeling of the aortic valve anulus. J Thorac Cardiovasc Surg. 1993;105:435–8.PubMed
43.
Zurück zum Zitat Oechtering TH, Sieren M, Schubert K, Schaller T, Scharfschwerdt M, Panagiotopoulos A, et al. In vitro 4D flow MRI evaluation of aortic valve replacements reveals disturbed flow distal to biological but not to mechanical valves. J Card Surg. 2019;34(12):1452-7. Oechtering TH, Sieren M, Schubert K, Schaller T, Scharfschwerdt M, Panagiotopoulos A, et al. In vitro 4D flow MRI evaluation of aortic valve replacements reveals disturbed flow distal to biological but not to mechanical valves. J Card Surg. 2019;34(12):1452-7.
44.
Zurück zum Zitat Boussel L, Rayz V, McCulloch C, Martin A, Acevedo-Bolton G, Lawton M, et al. Aneurysm growth occurs at region of low wall shear stress: patient-specific correlation of hemodynamics and growth in a longitudinal study. Stroke. 2008;39(11):2997–3002.PubMedPubMedCentral Boussel L, Rayz V, McCulloch C, Martin A, Acevedo-Bolton G, Lawton M, et al. Aneurysm growth occurs at region of low wall shear stress: patient-specific correlation of hemodynamics and growth in a longitudinal study. Stroke. 2008;39(11):2997–3002.PubMedPubMedCentral
45.
Zurück zum Zitat Wood NB, Weston SJ, Kilner PJ, Gosman AD, Firmin DN. Combined MR imaging and CFD simulation of flow in the human descending aorta. J Magn Reson Imaging. 2001;13(5):699–713.PubMed Wood NB, Weston SJ, Kilner PJ, Gosman AD, Firmin DN. Combined MR imaging and CFD simulation of flow in the human descending aorta. J Magn Reson Imaging. 2001;13(5):699–713.PubMed
Metadaten
Titel
4D flow cardiovascular magnetic resonance for monitoring of aortic valve repair in bicuspid aortic valve disease
verfasst von
Alexander Lenz
Johannes Petersen
Christoph Riedel
Julius M. Weinrich
Hendrik Kooijman
Bjoern P. Schoennagel
Gerhard Adam
Yskert von Kodolitsch
Hermann Reichenspurner
Evaldas Girdauskas
Peter Bannas
Publikationsdatum
01.12.2020
Verlag
BioMed Central
Erschienen in
Journal of Cardiovascular Magnetic Resonance / Ausgabe 1/2020
Elektronische ISSN: 1532-429X
DOI
https://doi.org/10.1186/s12968-020-00608-0

Weitere Artikel der Ausgabe 1/2020

Journal of Cardiovascular Magnetic Resonance 1/2020 Zur Ausgabe

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Ein Drittel der jungen Ärztinnen und Ärzte erwägt abzuwandern

07.05.2024 Klinik aktuell Nachrichten

Extreme Arbeitsverdichtung und kaum Supervision: Dr. Andrea Martini, Sprecherin des Bündnisses Junge Ärztinnen und Ärzte (BJÄ) über den Frust des ärztlichen Nachwuchses und die Vorteile des Rucksack-Modells.

Endlich: Zi zeigt, mit welchen PVS Praxen zufrieden sind

IT für Ärzte Nachrichten

Darauf haben viele Praxen gewartet: Das Zi hat eine Liste von Praxisverwaltungssystemen veröffentlicht, die von Nutzern positiv bewertet werden. Eine gute Grundlage für wechselwillige Ärztinnen und Psychotherapeuten.

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.