01.04.2010 | Original Article
99mTc-DTPA dynamic SPECT and CT volumetry for measuring split renal function in live kidney donors
Erschienen in: Annals of Nuclear Medicine | Ausgabe 3/2010
Einloggen, um Zugang zu erhaltenAbstract
Objectives
Split renal function (SRF) estimated from the posterior view of 99mTc-diethylenetriaminepentaacetic acid planar scintigraphy (DTPA/P) is not sufficiently accurate even after correction for kidney depth by computed tomography (CT). To obtain more accurate SRF using 99mTc-DTPA, dynamic SPECT method was carried out for the initial 5 min after bolus injection of 99mTc-DTPA (DTPA/SPECT). Also SRF was evaluated from the renal volume measured by CT. We compared the results with 99mTc-dimercaptosuccinic acid SPECT (DMSA/SPECT).
Methods
In 60 consecutive live kidney donors, 30 DTPA/P, 30 DTPA/SPECT, 60 99mTc-DMSA/SPECT, and 60 CT studies were performed. In the DTPA/P studies, SRF was calculated from the posterior image recorded during 2–3-min postinjection with attenuation correction for kidney depth measured by CT. In the DTPA/SPECT studies, SPECT images were acquired continuously for 5 min with a dual-headed gamma camera. In 99mTc-DMSA scintigraphy, DMSA/SPECT images were acquired 3-h postinjection. The SRF on both SPECT studies was calculated from the total counts of each kidney. In the DTPA/SPECT study, SRF was evaluated on the three images summed for 1 min: 1–2 min (DTPA/SPECT1–2), 2–3 min (DTPA/SPECT2–3), and 1–3 min (DTPA/SPECT1–3). In the CT examination, to assess the global renal volume, the axial images in the excretory phase were chosen. Renal contours were identified on each image, and the areas (cm2) of these regions were summed and multiplied by the slice thickness (10 mm) to yield global renal volume (ml). Right renal function from DTPA/P, DTPA/SPECT, and CT were compared with that from DMSA/SPECT as a reference.
Results
Correlation coefficients of the right renal function between DMSA/SPECT and DTPA/P, DTPA/SPECT2–3, and CT were 0.663, 0.849 and 0.907, respectively (P < 0.0001). The differences between DMSA/SPECT and DTPA/P, DTPA/SPECT2–3 and CT were 2.42 ± 3.878, 0.867 ± 1.672, and −0.421 ± 1.077% (mean ± SD), respectively.
Conclusion
SRF derived from DTPA/SPECT showed a better correlation with DMSA/SPECT and significantly fewer errors (P < 0.05) than DTPA/P. A significant strong correlation was observed between SRF from DMSA/SPECT and CT, indicating the utility of CT renal volumetry for evaluating SRF.