A case report of traumatic osteoarthritis associated with LARS artificial ligament use in anterior cruciate ligament reconstruction

verfasst von: Yuanliang Du, Haifeng Dai, Zhihui Wang, Di Wu, Changjiang Shi, Tianjie Xiao, Zhihuai Li

Erschienen in: BMC Musculoskeletal Disorders | Ausgabe 1/2020

Abstract

Background

A ligament advanced reinforcement system (LARS) artificial ligament has been proposed for use in anterior cruciate ligament (ACL) reconstruction, and many reports have shown its success in ACL reconstruction. However, there are great concerns about the potential risk of complications, which might prevent its extensive use. Late failure may occur due to serious complications.

Case presentation

We report a rare case of serious osteoarthritis that occurred 2 years postoperatively in a 51-year-old man who underwent reconstruction with an LARS artificial ligament. In X-rays, the tibial tunnel was placed too posteriorly. MRI showed that the tibial tunnel was enlarged, and there was a large effusion in the knee joint. The LARS device was rough and worn. Histologically, a large number of fibroblasts and a few multinucleated giant cells infiltrated the graft fibres.

Conclusion

Our findings remind surgeons that an LARS device should be with great caution in ACL reconstruction.
Hinweise

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
ACL
Anterior cruciate ligament
LARS
Ligament advanced reinforcement system

Background

Anterior cruciate ligament (ACL) injury is the most common reason for knee instability, and injuries to the ligament may lead to osteoarthritis and other degenerative joint disease [1, 2]. For patients eager to achieve orthobiosis, the most effective treatments for ACL injury are reconstruction of the ligament and postoperative physiotherapeutic procedures [3]. Reconstruction of the ligament could be conducive to knee stability and, in turn, reduce the risk of secondary injuries [4, 5]. A variety of grafts are available for use in ACL reconstruction surgery, such as allografts and synthetic ligaments. The use of allograft autografts are the first option in ACL-reconstruction in recent years with the attraction of absent donor site morbidity [6]. The ligament advanced reinforcement system (LARS), a synthetic, non-absorbable augmentation device made of polyethylene terephthalate (PET), has been also widely used in many countries [7]. With the aim of removing potential machining residues and oils, the LARS is intensively cleaned to further encourage soft tissue in-growth and reduce the risk of reactive synovitis [8]. In a recent study, the medium-term clinical outcomes of ACL reconstruction using an LARS artificial ligament were assessed, and among all patients enrolled in the study, the failure rate was 4.4%, with an overall complication rate of 2.2% [9]. Although these preliminary clinical investigations have shown encouraging results, there is concern that late failure may occur due to serious complications. We present a case report of osteoarthritis after LARS artificial ligament implantation that required revision at 2 years and a histologic analysis of the retrieved artificial ligament. This case highlights the seriousness of potential hazards when using an LARS artificial ligament, and further research may focus on its long-term outcomes after ACL reconstruction.

Case presentation

A 51-year-old worker sustained serious knee synovitis and osteoarthritis after ACL reconstruction with an LARS ligament 2 years previously. Knee arthroscopy was performed, and the ruptured artificial ligament was retrieved for histologic study. Subsequently, the patient underwent prosthetic replacement of the knee joint. Originally, the patient underwent ACL reconstruction with an LARS artificial ligament in the right knee for an ACL injury in 2011 in our department. After surgery, the postoperative physiotherapeutic procedure was followed. Postoperative recovery consisted of non-weightbearing on crutches in a hinged brace with a progressive increase in range of motion. The brace was removed 8 weeks after surgery, and full weightbearing was allowed to commence at 12 weeks. At 9 months, the patient reported continual aching in the knee joint after a long walk. No extra treatment was started except for more rest, and physical therapy was resumed. However, the patient had sustained, repeated pain and swelling in the right knee since then. The patient could not walk and suffered from continual aching, which could not be relieved after rest, 1 year later. During the initial physical examination, the patella tap test, anterior drawer test, and Lachman test were all positive. The right knee joint was obviously swollen before surgery (Fig. 1). X-rays demonstrated that the tibial tunnel was placed slightly posteriorly (Fig. 2). Magnetic resonance imaging showed that the tibial tunnel was enlarged on the T1 sequence, which led to loss of stability and chronic inflammation (Fig. 3a). On the T2-weighted images, there was high signal intensity on both the femoral and tibial sides, and there was an effusion in the joint (Fig. 3b). Internal fixation removal of the LARS artificial ligament and prosthetic replacement of the knee joint were recommended.

Surgery

First, the skin and subcutaneous tissue were cut, freeing approximately 3 cm on both sides along the deep subfascial space and exposing the front of the knee joint and the quadriceps tendon. The suprapatellar capsule and joint capsule were cut along the quadriceps tendon and the medial edge of the patella to the inside of the tibial tubercle. The effusion was drained, the tibia was evaginated, and the knee joint was flexed. Villous synovitis was evident in all compartments, and the medial tibial plateau, medial femur and patellar cartilage were all worn (Fig. 4). The intercondylar notch was narrow, and the meniscus had metamorphosed. The tibial tunnel was placed too posteriorly, and the LARS device was rough, worn and incompletely enshrouded in fibrous tissue (Fig. 5). The intra-articular LARS prosthesis was excised to reveal the titanium femoral interference screw, which was removed completely by special equipment. Simultaneously, the femoral and tibial ends of the prosthesis were able to be removed. After that, the femoral condyle, meniscus and tibia were replaced by artificial joint prostheses.

Histopathological analysis

The excised LARS prosthesis was divided into twelve segments to perform histopathological analysis. On light microscopy, marked hyperplasia was observed around all parts of the graft, with many multinucleate cells surrounding and inside the graft fibres (Fig. 6a). Approaching the tibial end of the graft, there were large amounts of proliferating chondrocytes inside the lacuna, along with some visible calcification (Fig. 6b). The closer the graft was to the tibia, the more obvious the phenomenon. However, only fibroblasts and poorly multinucleated cells infiltrated into the artificial ligament in the femoral end without any chondrocytes. There was no evidence of malignancy. The findings were consistent with potential involvement of the LARS ligament, which allows new ligamentous and neurovascular tissue to regenerate along a synthetic scaffold.

Discussion and conclusion

Theoretically, LARS artificial ligaments made of terephthalic polyethylene polyester fibres are a great innovation. The aims of its design are to prevent fibre breakdown early and facilitate tensioning of the graft fibres during knee movement. In contrast to traditional ACL reconstruction techniques, the LARS surgical technique reduces trauma to the soft tissues of the knee with less surgical time [10, 11]. Although the LARS has made achievements in the short term, serious complications have been reported, such as graft rupture, foreign-body inflammation, and serious knee synovitis [1214].
In one study that evaluated the clinical outcome of ACL reconstruction with LARS with at least a 7-year follow-up, the failure rate was only 4.4%, including two cases of recurrent instability, one case of limitation of the range of motion and one artificial graft rupture; however, no synovitis occurred in the study [9]. In a multicentre study with 3 to 5 years of follow-up in 159 patients, the overall complication rate for ACL reconstruction with LARS artificial ligaments was 5.7%, and serious knee synovitis developed only in one case [15]. Although synovitis appears to be a rare complication, it is very serious and can result in ligament rupture and surgical failure [1618]. In addition, synthetic grafts in ruptured ACL were associated with a lack of subjective satisfaction in half of patients treated with LARS [19]. The most common cause of complications was closely related to imperfect graft positioning: either the tibial or femoral tunnel was too far anteriorly or both [20, 21]. Amis and Kempson examined 25 ruptured polyethylene terephthalate fibre ACL implants retrieved clinically and believed that ACL implant failure was often caused by bone impingement during knee extension after malpositioning of the tibial tunnel (often too far anteriorly) [22]. However, in our case, the tibial tunnel was placed slightly posteriorly, and the ligament was ruptured near the tibial tunnel. The tunnel positioning in this case might have led to knee joint laxity and loss of stability, causing signs of osteoarthritis. With osteoarthritis in the knee joint, in turn, the LARS ligament could not achieve adequate tensile strength to maintain joint stability, which accelerated the process of ligament rupture. Therefore, non-anatomical reconstruction is the direct reason for surgical failure.
In this case, we left sufficient intact tissue to completely enshroud the LARS ligament in the femoral stump, intending to encourage fibrovascular ingrowth around individual fibres and hence prevent the articular release of wear particles. Based on observation of the histological analysis, only a few chondrocytes grew well along with the parallel fibres of the LARS ligament. However, the ligament should be highly cleaned to remove potential machining residues and oils to further reduce the risk of reactive synovitis and osteoarthritis. While coverage may theoretically prevent wear particles from entering the joint, the native residues or tissue remains shielded from stress by the nonabsorbable LARS graft, and when it does fail, the tissue shielded from stress will also fail, resulting in laxity and exposure of the joint to wear debris.
Although it would be unwise to draw too firm a conclusion from one case, the present case confirmed that LARS ligaments promote chondrocytes proliferation in terephthalic polyethylene polyester fibres with enough native residues by histologically analysis. However, the synovitis and osteoarthritis in this case might have resulted from shielded tissue failure. How to modify the residue surface requires further research to promote the biocompatibility of the LARS artificial ligament. Additionally, in the absence of peer-reviewed long-term outcome data, further use of the LARS device should be done with great caution.

Acknowledgements

We thank the Pathology Department of the Affiliated Hospital of Chengde Medical University and the patient for making the time and effort for this case report.
This case report was produced in accordance with the Ethics Committee of the Affiliated Hospital of Chengde Medical University.
The authors have obtained the patient’s informed written consent for publication of this case report. All participants provided their written consent to publish their data and accompanying images.

Competing interests

The authors declare that they have no competing interests.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
4.
Zurück zum Zitat Sajovic M, Strahovnik A, Dernovsek MZ, Skaza K. Quality of life and clinical outcome comparison of semitendinosus and gracilis tendon versus patellar tendon autografts for anterior cruciate ligament reconstruction: an 11-year follow-up of a randomized controlled trial. Am J Sports Med. 2011;39:2161–9. https://doi.org/10.1177/0363546511411702.CrossRefPubMed Sajovic M, Strahovnik A, Dernovsek MZ, Skaza K. Quality of life and clinical outcome comparison of semitendinosus and gracilis tendon versus patellar tendon autografts for anterior cruciate ligament reconstruction: an 11-year follow-up of a randomized controlled trial. Am J Sports Med. 2011;39:2161–9. https://​doi.​org/​10.​1177/​0363546511411702​.CrossRefPubMed
5.
Zurück zum Zitat Lewek MD, Chmielewski TL, Risberg MA, Snyder-Mackler L. Dynamic knee stability after anterior cruciate ligament rupture. Exerc Sport Sci Rev. 2003;31:195–200.CrossRef Lewek MD, Chmielewski TL, Risberg MA, Snyder-Mackler L. Dynamic knee stability after anterior cruciate ligament rupture. Exerc Sport Sci Rev. 2003;31:195–200.CrossRef
8.
Zurück zum Zitat Lavoie P, Fletcher J, Duval N. Patient satisfaction needs as related to knee stability and objective findings after ACL reconstruction using the LARS artificial ligament. Knee. 2000;7:157–63.CrossRef Lavoie P, Fletcher J, Duval N. Patient satisfaction needs as related to knee stability and objective findings after ACL reconstruction using the LARS artificial ligament. Knee. 2000;7:157–63.CrossRef
13.
21.
Zurück zum Zitat Estrada-Cruz K, Pérez-Meave JA. Postoperative functional angular anatomy in patients with anterior cruciate ligament with autologous graft. Acta Ortop Mex. 2018;32:157–62.PubMed Estrada-Cruz K, Pérez-Meave JA. Postoperative functional angular anatomy in patients with anterior cruciate ligament with autologous graft. Acta Ortop Mex. 2018;32:157–62.PubMed
22.
Zurück zum Zitat Amis AA, Kempson SA. Failure mechanisms of polyester fiber anterior cruciate ligament implants: a human retrieval and laboratory study. J Biomed Mater Res. 1999;48:534–9.CrossRef Amis AA, Kempson SA. Failure mechanisms of polyester fiber anterior cruciate ligament implants: a human retrieval and laboratory study. J Biomed Mater Res. 1999;48:534–9.CrossRef
Metadaten
Titel
A case report of traumatic osteoarthritis associated with LARS artificial ligament use in anterior cruciate ligament reconstruction
verfasst von
Yuanliang Du
Haifeng Dai
Zhihui Wang
Di Wu
Changjiang Shi
Tianjie Xiao
Zhihuai Li
Publikationsdatum
01.12.2020
Verlag
BioMed Central
Erschienen in
BMC Musculoskeletal Disorders / Ausgabe 1/2020
Elektronische ISSN: 1471-2474
DOI
https://doi.org/10.1186/s12891-020-03764-7

Weitere Artikel der Ausgabe 1/2020

BMC Musculoskeletal Disorders 1/2020 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Proximale Humerusfraktur: Auch 100-Jährige operieren?

01.05.2024 DCK 2024 Kongressbericht

Mit dem demographischen Wandel versorgt auch die Chirurgie immer mehr betagte Menschen. Von Entwicklungen wie Fast-Track können auch ältere Menschen profitieren und bei proximaler Humerusfraktur können selbst manche 100-Jährige noch sicher operiert werden.

Sind Frauen die fähigeren Ärzte?

30.04.2024 Gendermedizin Nachrichten

Patienten, die von Ärztinnen behandelt werden, dürfen offenbar auf bessere Therapieergebnisse hoffen als Patienten von Ärzten. Besonders gilt das offenbar für weibliche Kranke, wie eine Studie zeigt.

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Arthroskopie kann Knieprothese nicht hinauszögern

25.04.2024 Gonarthrose Nachrichten

Ein arthroskopischer Eingriff bei Kniearthrose macht im Hinblick darauf, ob und wann ein Gelenkersatz fällig wird, offenbar keinen Unterschied.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.