A novel 3D-printed locking cage for anterior atlantoaxial fixation and fusion: case report and in vitro biomechanical evaluation

verfasst von: Shenglin Wang, Huijie Leng, Yinglun Tian, Nanfang Xu, Zhongjun Liu

Erschienen in: BMC Musculoskeletal Disorders | Ausgabe 1/2021

Abstract

Background

Treatment of atlantoaxial dislocation is aimed at reduction and stabilization of the atlantoaxial joint. 3D printing refers to a process where additive manufacturing is achieved under precise computer control. Literature on its utilization in anterior atlantoaxial fixation and fusion is rare. This study is the first report on a 3D-printed locking cage used in the anterior procedure for atlantoaxial dislocation.

Methods

A middle-aged male in his 40s presented with weakness and numbness of his extremities for 3 years and could only walk slowly with assistance. Imaging studies revealed severe anterior migration of C1, irreducible atlantoaxial dislocation, and severe cervical-medullary compression. A preoperative plan consisting of trans-oral soft tissue release and fixation using tailor-designed 3D-printed cages was devised. Following fluoroscopic confirmation of reduction of the atlantoaxial joints, two customized 3D-printed cages made of titanium alloy were inserted into the bilateral facet joints, which were then locked by six screws into the lateral masses of C1 and C2. The microstructure of the inserted cages was optimized for improved biomechanical stability and enhanced osseo-integration, without the need for bone grafting. In addition, a biomechanical test was performed on seven human cadaveric specimens comparing the novel implant with the conventional C1 lateral mass-C2 pedicle screw construct in three modes of motion (flexion-extension, lateral bending, axial rotation).

Results

Improvement of neurologic function in the patient was evident immediately after surgery. He was able to walk independently 1 month post-operatively. At the 12-month follow-up, coronal reconstruction of CT demonstrated properly-positioned 3D-printed cages, evidence of osseo-integration at the bone-implant interface, and no subsidence or displacement of the implant. Eighteen months out of surgery, the mJOA score improved to 15, and lateral X-ray confirmed reduction of atlanto-axial dislocation. Additionally, the new construct provided strong fixation comparable to that conferred by conventional constructs as there was no significant difference observed between the two groups in all three directions of motion.

Conclusions

The novel implant represents a new option in the treatment of irreducible atlantoaxial dislocation. It can provide strong anterior support for solid fixation and fusion with a low profile and a microstructure that obviates the need for bone grafting.
Begleitmaterial
Hinweise

Supplementary Information

The online version contains supplementary material available at https://​doi.​org/​10.​1186/​s12891-021-03987-2.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
mJOA
modified Japanese Orthopaedic Association
ROM
range of motion
IAAD
irreducible atlantoaxial dislocation
CVJ
cranio-vertebral junction
3DPLC
3D-printed locking cage
AAD
atlantoaxial dislocation
PMMA
polymethylmethacrylate
C1LM-C2PS
C1 lateral mass-C2 pedicle screw
C1–C2 TA
C1-C2 trans-articular screws
C1LM-C2TL
C1 lateral mass and C2 trans-laminar screws
C1–C2 LM
C1-C2 lateral mass screws
TARP
trans-oral atlantoaxial reduction plate

Background

In the management of irreducible atlantoaxial dislocation (IAAD), the goal of treatment consists of restoration of the normal anatomy of the cranio-vertebral junction (CVJ), complete decompression of the lower medulla and the upper spinal cord, and stabilization of the atlantoaxial joints [1, 2]. The main procedures reported in the literature include trans-oral odontoidectomy or peri-odontoid tissue release followed by posterior occipito-cervical or atlanto-axial fusion [35]. Anterior-only (with peri-odontoid soft tissue release followed by anterior fixation) represents another strategy which eliminates the need for a two-step operation [6]. This single-stage procedure used a reduction plate to achieve decompression, reduction and fusion, all through a trans-oral incision. However, the trans-oral reduction plate had a high profile and still required bone harvest from the iliac crest [6], as did the two-stage procedures. Although there had been attempts on using rapid prototyping (3D printing) to facilitate instrumentation in the upper cervical spine [7, 8], to our knowledge, its utilization in anterior atlantoaxial fixation has not been reported. We herein describe the use of a novel self-locking implant with semi-constrained screws in a patient with IAAD. In addition, the biomechanics of this new 3D-printed locking cage (3DPLC) was evaluated in seven cadaver specimens.

Methods

Clinical report

A 44-year-old man presented with weakness and numbness of both upper and lower extremities for 3 years and had no history of cervical trauma. He could ambulate using a walker and had deficits of motor function in the right hand. His mJOA (modified Japanese Orthopaedic Association, supplementary file) [9] score was 11. Sensory dysfunction of the upper extremities was detected upon neurological examination and symmetric hyperreflexia was present in the lower extremities. Positive Hoffmann’s signs were elicited bilaterally. His gait was wide-based, spastic, and unsteady. The patient also had limited cervical flexion and extension. Imaging studies revealed AAD with severe anterior translation of C1 on X-ray (Fig. 1a) and significant cervical-medullary cord compression with high signal intensity on T2 MRI. He was diagnosed of IAAD, and as it was not reduced with a 1/5 body weight cranial traction, trans-oral peri-odontoid release was required to allow for anatomic reduction [5]. Additionally, in order to obviate an additional posterior procedure to provide fixation and fusion, with informed consent from the patient, we decided to use a novel implant, the 3DPLC, to perform an anterior-only surgery. The 3DPLC was manufactured using titanium alloy powder based on a pre-designed computer model customized to the specific anatomy of the C1-C2 facet joint of this particular patient (Fig. 1b,c, d, and e). The model was built from a thin-cut CT scan of the upper cervical spine using Materialise Interactive Medical Image Control System (MIMICS). During the procedure, complete reduction was first achieved following peri-odontoid soft tissue release, and the two tailor-fabricated 3DPLCs were implanted into the C1-C2 facet joints. Position of the cages were confirmed by C-arm image intensification, and three screws were used (one on C1 lateral mass and two on C2 vertebral body) to lock each cage (Fig. 1f, g and h). Use of this novel implant was approved by the local hospital ethics committee.

Biomechanical evaluation

Specimen preparation

Seven fresh frozen human cadaveric cervical spine specimens (kept at − 20 degrees in sealed bags) from subjects with no history of trauma or deformities involving the cervical spine were used (Fig. 2a). No specimen was osteoporotic as confirmed by densitometry. Soft tissue structures, including the ligaments, joint capsules, intervertebral discs were all intact on visual inspection. Before the test, specimens were thawed overnight at room temperature. Neutral alignment of the cervical spine was carefully maintained during preparation when both ends of the specimen were embedded with polymethylmethacrylate (PMMA) and mounted onto a spine motion simulator (MTS Corp, Eden Prairie, MN USA).

Testing protocol

Pilot markers were attached anteriorly to the C1 right lateral mass and the C2 vertebral body to track their respective segmental range of motion (ROM), which was measured and analyzed using NDI Optotrak Certus, a 3D movement measuring system (Norther Digital Inc., Ontario, Canada). Classic posterior C1 lateral mass-C2 pedicle screw (C1LM-C2PS) instrumentation [10] was first performed. Each specimen was pre-conditioned for three loading cycles with moment variation between 2 Nm and − 2 Nm to minimize viscoelasticity effects. (The moment of 2 Nm was chosen because based on the literature [9, 11], the most commonly used values in cervical spine studies range from 1.5 Nm to 2.5 Nm.) Then, starting from a neutral position, each specimen was tested in three different directions of motion in the order of extension-flexion (i.e. anterior-posterior translation), lateral bending, and axial rotation, while keeping the displacement under control between angles corresponding to the moments between 2 NM and -2NM as determined in the previous viscoelasticity mitigation step (Fig. 2b, c). The posterior implants were then removed and two personalized 3DPLCs were inserted anteriorly between the C1-C2 facet joints (Fig. 2d, e, f, g) in each specimen. A second round of biomechanical evaluation was then performed. Segmental movement between C1 and C2 was measured for each direction of motion with either type of internal fixation and data were analyzed using paired sample T-test (SPSS, 17.0 statistics). The precision of the NDI system for rigid body displacement measurement is 0.01 mm.

Results

Clinical results

Improvement of his neurological function was evident immediately after surgery. A Philadelphia collar was used for 3 months for outdoor activities. He was able to walk independently 1 month post-operatively. At the 12-month follow-up, CT scan (axial, sagittal, coronal) demonstrated properly-positioned 3DPLCs, evidence of osseo-integration at the bone-implant interface, and no subsidence or displacement of the 3DPLCs (Fig. 1i, j, k, and l). Eighteen months out of surgery, the mJOA score improved to 15, and lateral X-ray confirmed reduction of atlanto-axial dislocation. Even though partial osseous collapse of C1 lateral masses was suspected, there was no implant subsidence or loosening of the screws (Fig. 1m).

Biomechanical results

Both the posterior and the anterior constructs provided strong fixation. In all three directions of motion, the amount of segmental movement was limited and there was no significant difference observed between the two types of fixation (Fig. 3, Table 1).
Table 1
Comparison between 3DPLC and the conventionl posterior construct in three modes of motion
Motion mode
ROM (degree)
Anterior Fixation (Mean ± SEM)
Posterior Fixation
(Mean ± SEM)
P value
Flexion & Extension
0.024 ± 0.028
−0.004 ± 0.005
0.33
Lateral Bending
−0.013 ± 0.006
0.017 ± 0.015
0.09
Axial Rotation
−0.007 ± 0.003
−0.173 ± 0.167
0.34

Discussion

We previously described the first utilization of 3D-printed implants in spine surgery in 2016 for anterior C2 reconstruction following spondylectomy due to Ewing sarcoma in an adolescent [8]. Phan et al. [12] reported a posterior 3D-printed implant for atlanto-axial arthrodesis in a 65-year female, however, she was followed up for only 2 months. Additional report on the use of 3D-printed constructs in the upper cervical spine was rare in the literature.
The purpose of the current study was to illustrate the first anterior customized implant (3DPLC) in the upper cervical spine for an IAAD patient with mid-term follow-up. Surgical treatment of IAAD consisted of two steps: release of the peri-odontoid soft tissue to allow anatomic reduction (typically trans-oral from the anterior) and strong fixation (typically from the posterior) to allow early fusion. 3DPLC was an anterior implant that could provide solid fixation (comparable to the strongest posterior construct in all forms of ROM) along with a cage that facilitated osseo-integration. Since trans-oral release often needed to be performed in patients with IAAD, any posterior instrumentation would require a second surgical procedure with associated morbidity, which could have been avoided if an anterior instrumentation and fusion procedure were to be performed following trans-oral release. Use of our novel implant following the trans-oral release procedure could help avoid both a second-stage posterior surgery and the need for iliac bone harvesting. Additionally, it could be used as a salvage technique in cases where abnormal anatomy or previous surgical procedures precluded the use of posterior instrumentation and anterior instrumentation became the only viable option.
Existing posterior constructs for atlantoaxial fixation included C1LM–C2PS, C1-C2 trans-articular screws (C1–C2 TA), C1 lateral mass and C2 trans-laminar screws (C1LM-C2TL), and C1-C2 lateral mass screws (C1–C2 LM). The biomechanical properties of these constructs had been assessed in previous studies [1315]. A meta-analysis [13] of fifteen articles found that C1-C2 LM is less stable against axial rotation, C1LM–C2TL is less stable against lateral bending, while C1LM–C2PS represents the strongest form of atlantoaxial fixation. On the other hand, the primary form of anterior fixation in the atlantoaxial region in the literature was the anterior trans-oral atlantoaxial reduction plate (TARP). The TARP construct offered comparable stability to the C1LM-C2PS constructs in all directions of segmental movement except for flexion [16]. In comparison, results from the current study showed that 3DPLC could provide stability in all directions of movement equivalent to the conventional C1LM–C2PS construct (Fig. 3). Additionally, fixation of the new 3DPLC was easier than the TARP construct and its 3D-printed porous micro-structure could facilitate osseo-integration, such that graft harvest from the iliac crest, which could be fraught with complications (e.g. pain, wound dehiscence, etc.), became obsolete with this new implant. Lastly, the 3DPLC had a lower profile when compared to TARP, especially in the midline where the incision was made. Therefore, the 3DPLC might better facilitate wound healing in the posterior pharyngeal wall.
There were several limitations to this study. First, this was a case report with mid-term follow-up; second, the biomechanical evaluation did not include a fatigue test; and third, it currently took 2-3 weeks to manufacture the customized implant, prohibiting its use in emergency cases. However, developing anterior implants for atlantoaxial instrumentation and fusion was a project worthy of continuing efforts and could be built upon results from our current study.

Conclusions

The novel implant herein described represented a new option in the treatment of irreducible atlantoaxial dislocation that could provide strong anterior support for solid fixation and fusion with a low profile and a microstructure that obviated the need for bone grafting.

Acknowledgements

Not Applicable.
This study was approved by the Ethics Committee of Peking University Third Hospital and written consent to participate was obtained from the patient.
Not Applicable.

Competing interests

None.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Anhänge

Supplementary Information

Literatur
1.
Zurück zum Zitat Wang S, et al. Novel surgical classification and treatment strategy for atlantoaxial dislocations. Spine (Phila Pa 1976). 2013;38(21):E1348–56.CrossRef Wang S, et al. Novel surgical classification and treatment strategy for atlantoaxial dislocations. Spine (Phila Pa 1976). 2013;38(21):E1348–56.CrossRef
2.
Zurück zum Zitat Wu X, et al. Surgical strategies for the treatment of os odontoideum with atlantoaxial dislocation. J Neurosurg Spine. 2018;28(2):131–9.CrossRef Wu X, et al. Surgical strategies for the treatment of os odontoideum with atlantoaxial dislocation. J Neurosurg Spine. 2018;28(2):131–9.CrossRef
3.
Zurück zum Zitat Guan J, et al. Effectiveness of posterior reduction and fixation in atlantoaxial dislocation: a retrospective cohort study of 135 patients with a treatment algorithm proposal. Eur Spine J. 2019. Guan J, et al. Effectiveness of posterior reduction and fixation in atlantoaxial dislocation: a retrospective cohort study of 135 patients with a treatment algorithm proposal. Eur Spine J. 2019.
4.
Zurück zum Zitat Wang S, et al. Atlantoaxial rotatory fixed dislocation: report on a series of 32 pediatric cases. Spine (Phila Pa 1976). 2016;41(12):E725–32.CrossRef Wang S, et al. Atlantoaxial rotatory fixed dislocation: report on a series of 32 pediatric cases. Spine (Phila Pa 1976). 2016;41(12):E725–32.CrossRef
5.
Zurück zum Zitat Wang C, et al. Open reduction of irreducible atlantoaxial dislocation by transoral anterior atlantoaxial release and posterior internal fixation. Spine (Phila Pa 1976). 2006;31(11):E306–13.CrossRef Wang C, et al. Open reduction of irreducible atlantoaxial dislocation by transoral anterior atlantoaxial release and posterior internal fixation. Spine (Phila Pa 1976). 2006;31(11):E306–13.CrossRef
6.
Zurück zum Zitat Yin QS, et al. Transoral atlantoaxial reduction plate fixation for irreducible atlantoaxial dislocation. Chin J Traumatol. 2006;9(1):14–20.PubMed Yin QS, et al. Transoral atlantoaxial reduction plate fixation for irreducible atlantoaxial dislocation. Chin J Traumatol. 2006;9(1):14–20.PubMed
7.
Zurück zum Zitat Cai H, et al. 3D Printing in Spine Surgery. Adv Exp Med Biol. 2018;1093:345–59.CrossRef Cai H, et al. 3D Printing in Spine Surgery. Adv Exp Med Biol. 2018;1093:345–59.CrossRef
8.
Zurück zum Zitat Xu N, et al. Reconstruction of the upper cervical spine using a personalized 3D-printed vertebral body in an adolescent with Ewing sarcoma. Spine (Phila Pa 1976). 2016;41(1):E50–4.CrossRef Xu N, et al. Reconstruction of the upper cervical spine using a personalized 3D-printed vertebral body in an adolescent with Ewing sarcoma. Spine (Phila Pa 1976). 2016;41(1):E50–4.CrossRef
9.
Zurück zum Zitat Tetreault L, et al. The modified Japanese Orthopaedic association scale: establishing criteria for mild, moderate and severe impairment in patients with degenerative cervical myelopathy. Eur Spine J. 2017;26(1):78–84.CrossRef Tetreault L, et al. The modified Japanese Orthopaedic association scale: establishing criteria for mild, moderate and severe impairment in patients with degenerative cervical myelopathy. Eur Spine J. 2017;26(1):78–84.CrossRef
10.
Zurück zum Zitat Wang S, et al. Radiographic evaluation of the technique for C1 lateral mass and C2 pedicle screw fixation in three hundred nineteen cases. Spine (Phila Pa 1976). 2011;36(1):3–8.CrossRef Wang S, et al. Radiographic evaluation of the technique for C1 lateral mass and C2 pedicle screw fixation in three hundred nineteen cases. Spine (Phila Pa 1976). 2011;36(1):3–8.CrossRef
11.
Zurück zum Zitat Yang J, et al. In vivo study of a self-stabilizing artificial vertebral body fabricated by electron beam melting. Spine (Phila Pa 1976). 2014;39(8):E486–92.CrossRef Yang J, et al. In vivo study of a self-stabilizing artificial vertebral body fabricated by electron beam melting. Spine (Phila Pa 1976). 2014;39(8):E486–92.CrossRef
12.
Zurück zum Zitat Phan K, et al. Application of a 3D custom printed patient specific spinal implant for C1/2 arthrodesis. J Spine Surg. 2016;2(4):314–8.CrossRef Phan K, et al. Application of a 3D custom printed patient specific spinal implant for C1/2 arthrodesis. J Spine Surg. 2016;2(4):314–8.CrossRef
13.
Zurück zum Zitat Du JY, et al. Biomechanical analysis of screw constructs for atlantoaxial fixation in cadavers: a systematic review and meta-analysis. J Neurosurg Spine. 2015;22(2):151–61.CrossRef Du JY, et al. Biomechanical analysis of screw constructs for atlantoaxial fixation in cadavers: a systematic review and meta-analysis. J Neurosurg Spine. 2015;22(2):151–61.CrossRef
14.
Zurück zum Zitat Chun DH, et al. Biomechanical comparison of four different Atlantoaxial posterior fixation constructs in adults: a finite element study. Spine (Phila Pa 1976). 2018;43(15):E891–7.CrossRef Chun DH, et al. Biomechanical comparison of four different Atlantoaxial posterior fixation constructs in adults: a finite element study. Spine (Phila Pa 1976). 2018;43(15):E891–7.CrossRef
15.
Zurück zum Zitat Liu S, et al. Biomechanical evaluation of C1 lateral mass and C2 translaminar bicortical screws in atlantoaxial fixation: an in vitro human cadaveric study. Spine J. 2018;18(4):674–81.CrossRef Liu S, et al. Biomechanical evaluation of C1 lateral mass and C2 translaminar bicortical screws in atlantoaxial fixation: an in vitro human cadaveric study. Spine J. 2018;18(4):674–81.CrossRef
16.
Zurück zum Zitat Zhang B, et al. Biomechanical comparison of modified TARP technique versus modified Goel technique for the treatment of basilar invagination: a finite element analysis. Spine (Phila Pa 1976). 2016;41(8):E459–66.CrossRef Zhang B, et al. Biomechanical comparison of modified TARP technique versus modified Goel technique for the treatment of basilar invagination: a finite element analysis. Spine (Phila Pa 1976). 2016;41(8):E459–66.CrossRef
Metadaten
Titel
A novel 3D-printed locking cage for anterior atlantoaxial fixation and fusion: case report and in vitro biomechanical evaluation
verfasst von
Shenglin Wang
Huijie Leng
Yinglun Tian
Nanfang Xu
Zhongjun Liu
Publikationsdatum
01.12.2021
Verlag
BioMed Central
Erschienen in
BMC Musculoskeletal Disorders / Ausgabe 1/2021
Elektronische ISSN: 1471-2474
DOI
https://doi.org/10.1186/s12891-021-03987-2

Weitere Artikel der Ausgabe 1/2021

BMC Musculoskeletal Disorders 1/2021 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Arthroskopie kann Knieprothese nicht hinauszögern

25.04.2024 Gonarthrose Nachrichten

Ein arthroskopischer Eingriff bei Kniearthrose macht im Hinblick darauf, ob und wann ein Gelenkersatz fällig wird, offenbar keinen Unterschied.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.