Skip to main content
Erschienen in: Annals of Nuclear Medicine 7/2014

01.08.2014 | Original Article

A randomized study of the effect of carbonated water prior to myocardial SPECT

verfasst von: Ilse A. C. Vermeltfoort, Arjan B. van Dijk, Jeroen A. F. de Jong, Gerrit J. J. Teule, Marjon Gevers, Bas Verhoeven, Esther Raaijmakers, Paul Knaapen, Pieter G. H. M. Raijmakers

Erschienen in: Annals of Nuclear Medicine | Ausgabe 7/2014

Einloggen, um Zugang zu erhalten

Abstract

Objective

In myocardial perfusion single-photon emission computed tomography (SPECT), abdominal activity often interferes with the evaluation of perfusion in the inferior wall, especially after pharmacological stress. In this randomized study, we examined the effect of carbonated water intake versus still water intake on the quality of images obtained during myocardial perfusion images (MPI) studies.

Methods

A total of 467 MIBI studies were randomized into a carbonated water group and a water group. The presence of intestinal activity adjacent to the inferior wall was evaluated by two observers. Furthermore, a semi-quantitative analysis was performed in the adenosine subgroup, using a count ratio of the inferior myocardial wall and adjacent abdominal activity.

Results

The need for repeated SPECT in the adenosine studies was 5.3 % in the carbonated water group versus 19.4 % in the still water group (p = 0.019). The inferior wall-to-abdomen count ratio was significantly higher in the carbonated water group compared to the still water group (2.11 ± 1.00 vs. 1.72 ± 0.73, p < 0.001). The effect of carbonated water during rest and after exercise was not significant.

Conclusions

This randomized study showed that carbonated water significantly reduced the interference of extra-cardiac activity in adenosine SPECT MPI.
Literatur
1.
Zurück zum Zitat Matsumoto N, Sato Y, Suzuki Y, Yoda S, Kunimasa T, Kato M, et al. Usefulness of rapid low-dose/high-dose 1-day 99 mTc-sestamibi ECG-gated myocardial perfusion single-photon emission computed tomography. Circ J. 2006;70:1585–9.PubMedCrossRef Matsumoto N, Sato Y, Suzuki Y, Yoda S, Kunimasa T, Kato M, et al. Usefulness of rapid low-dose/high-dose 1-day 99 mTc-sestamibi ECG-gated myocardial perfusion single-photon emission computed tomography. Circ J. 2006;70:1585–9.PubMedCrossRef
2.
Zurück zum Zitat Hofman M, McKay J, Nandurkar D. Efficacy of milk versus water to reduce interfering infra-cardiac activity in 99 mTc-sestamibi myocardial perfusion scintigraphy. Nucl Med Commun. 2006;27:837–42.PubMedCrossRef Hofman M, McKay J, Nandurkar D. Efficacy of milk versus water to reduce interfering infra-cardiac activity in 99 mTc-sestamibi myocardial perfusion scintigraphy. Nucl Med Commun. 2006;27:837–42.PubMedCrossRef
3.
Zurück zum Zitat Iqbal SM, Khalil ME, Lone BA, Gorski R, Blum S, Heller EN. Simple techniques to reduce bowel activity in cardiac SPECT imaging. Nucl Med Commun. 2004;25:355–9.PubMedCrossRef Iqbal SM, Khalil ME, Lone BA, Gorski R, Blum S, Heller EN. Simple techniques to reduce bowel activity in cardiac SPECT imaging. Nucl Med Commun. 2004;25:355–9.PubMedCrossRef
4.
Zurück zum Zitat Cherng SC, Chen YH, Lee MS, Yang SP, Huang WS, Cheng CY. Acceleration of hepatobiliary excretion by lemon juice on 99 mTc-tetrofosmin cardiac SPECT. Nucl Med Commun. 2006;27:859–64.PubMedCrossRef Cherng SC, Chen YH, Lee MS, Yang SP, Huang WS, Cheng CY. Acceleration of hepatobiliary excretion by lemon juice on 99 mTc-tetrofosmin cardiac SPECT. Nucl Med Commun. 2006;27:859–64.PubMedCrossRef
5.
Zurück zum Zitat Hara M, Monzen H, Futai R, Inagaki K, Shimoyama H, Morikawa M, et al. Reduction of infracardiac intestinal activity by a small amount of carbonated water in technetium-99m tetrofosmin myocardial perfusion scintigraphy with adenosine stress. J Nucl Cardiol. 2008;15:241–5.PubMedCrossRef Hara M, Monzen H, Futai R, Inagaki K, Shimoyama H, Morikawa M, et al. Reduction of infracardiac intestinal activity by a small amount of carbonated water in technetium-99m tetrofosmin myocardial perfusion scintigraphy with adenosine stress. J Nucl Cardiol. 2008;15:241–5.PubMedCrossRef
6.
Zurück zum Zitat Boz A, Karayalcin B. Which is better for inferior wall evaluation: a full or empty stomach? J Nucl Med. 1996;37:1916–7.PubMed Boz A, Karayalcin B. Which is better for inferior wall evaluation: a full or empty stomach? J Nucl Med. 1996;37:1916–7.PubMed
7.
Zurück zum Zitat Malhotra G, Upadhye TS, Nabar A, Asopa RV, Nayak UN, Rajan MG. Can carbonated lime drink intake prior to myocardial perfusion imaging with Tc-99m MIBI reduce the extracardiac activity that degrades the image quality and leads to fallacies in interpretation? Clin Nucl Med. 2010;35:160–4.PubMedCrossRef Malhotra G, Upadhye TS, Nabar A, Asopa RV, Nayak UN, Rajan MG. Can carbonated lime drink intake prior to myocardial perfusion imaging with Tc-99m MIBI reduce the extracardiac activity that degrades the image quality and leads to fallacies in interpretation? Clin Nucl Med. 2010;35:160–4.PubMedCrossRef
8.
Zurück zum Zitat Edlund A, Sollevi A, Linde B. Haemodynamic and metabolic effects of infused adenosine in man. Clin Sci (Lond). 1990;79:131–8.PubMed Edlund A, Sollevi A, Linde B. Haemodynamic and metabolic effects of infused adenosine in man. Clin Sci (Lond). 1990;79:131–8.PubMed
Metadaten
Titel
A randomized study of the effect of carbonated water prior to myocardial SPECT
verfasst von
Ilse A. C. Vermeltfoort
Arjan B. van Dijk
Jeroen A. F. de Jong
Gerrit J. J. Teule
Marjon Gevers
Bas Verhoeven
Esther Raaijmakers
Paul Knaapen
Pieter G. H. M. Raijmakers
Publikationsdatum
01.08.2014
Verlag
Springer Japan
Erschienen in
Annals of Nuclear Medicine / Ausgabe 7/2014
Print ISSN: 0914-7187
Elektronische ISSN: 1864-6433
DOI
https://doi.org/10.1007/s12149-014-0863-4

Weitere Artikel der Ausgabe 7/2014

Annals of Nuclear Medicine 7/2014 Zur Ausgabe