Artificial intelligence guided enhancement of digital PET: scans as fast as CT?

verfasst von: René Hosch, Manuel Weber, Miriam Sraieb, Nils Flaschel, Johannes Haubold, Moon-Sung Kim, Lale Umutlu, Jens Kleesiek, Ken Herrmann, Felix Nensa, Christoph Rischpler, Sven Koitka, Robert Seifert, David Kersting

Erschienen in: European Journal of Nuclear Medicine and Molecular Imaging | Ausgabe 13/2022

Abstract

Purpose

Both digital positron emission tomography (PET) detector technologies and artificial intelligence based image post-reconstruction methods allow to reduce the PET acquisition time while maintaining diagnostic quality. The aim of this study was to acquire ultra-low-count fluorodeoxyglucose (FDG) ExtremePET images on a digital PET/computed tomography (CT) scanner at an acquisition time comparable to a CT scan and to generate synthetic full-dose PET images using an artificial neural network.

Methods

This is a prospective, single-arm, single-center phase I/II imaging study. A total of 587 patients were included. For each patient, a standard and an ultra-low-count FDG PET/CT scan (whole-body acquisition time about 30 s) were acquired. A modified pix2pixHD deep-learning network was trained employing 387 data sets as training and 200 as test cohort. Three models (PET-only and PET/CT with or without group convolution) were compared. Detectability and quantification were evaluated.

Results

The PET/CT input model with group convolution performed best regarding lesion signal recovery and was selected for detailed evaluation. Synthetic PET images were of high visual image quality; mean absolute lesion SUVmax (maximum standardized uptake value) difference was 1.5. Patient-based sensitivity and specificity for lesion detection were 79% and 100%, respectively. Not-detected lesions were of lower tracer uptake and lesion volume. In a matched-pair comparison, patient-based (lesion-based) detection rate was 89% (78%) for PERCIST (PET response criteria in solid tumors)-measurable and 36% (22%) for non PERCIST-measurable lesions.

Conclusion

Lesion detectability and lesion quantification were promising in the context of extremely fast acquisition times. Possible application scenarios might include re-staging of late-stage cancer patients, in whom assessment of total tumor burden can be of higher relevance than detailed evaluation of small and low-uptake lesions.
Hinweise

Supplementary Information

The online version contains supplementary material available at https://​doi.​org/​10.​1007/​s00259-022-05901-x.
This article is part of the Topical Collection on Advanced Image Analyses (Radiomics and Artificial Intelligence).
Robert Seifert and David Kersting contributed equally.

Publisher's note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Introduction

Both the recently introduced “digital” positron emission tomography (PET) / computed tomography (CT) systems and deep-learning based PET image post-processing tools have the potential to decrease the scanning time duration while maintaining clinically relevant diagnostic information. If current standard scan protocols are applied, the typical time required for a whole-body fluorodeoxyglucose (FDG) PET scan lies in the range of 20–45 min [1]. For reasons of patient comfort, particularly for anguished, dyspneic, or pediatric patients and for reasons of cost effectiveness, shorter acquisition times are desirable [2]. Using a maximum table speed velocity of a modern digital PET/CT scanner, acquisition times of 20 to 30 s per whole-body scan that match acquisition time of a CT scan are technically feasible. However, one of the major challenges remains the tradeoff between reducing the acquisition time and conserving clinical information, as time reductions consistently result in a significant deterioration of image quality [1].
Digital PET/CT systems use silicon-photomultipliers (SiPM) that exhibit an increased detector sensitivity resulting in a higher spatial resolution and coincidence time resolution when compared to previous-generation photomultiplier-tube-based systems [36]. A higher signal recovery increases the detectability of small and low-count lesions [5, 710]. In particular, low-count PET images reconstructed with time-of-flight (TOF) option benefit from the improved time resolution characteristics of SiPM-based PET [11]. Consequently, the acquisition time might be reduced without loss of clinically relevant information. Several phantom and clinical studies by our group and others demonstrated that, depending on the clinical question, a reduction in acquisition time up to a factor of \({\raise0.7ex\hbox{$1$} \!\mathord{\left/ {\vphantom {1 3}}\right.\kern-\nulldelimiterspace} \!\lower0.7ex\hbox{$3$}}\) can be feasible for different radionuclides including 18F-FDG without image post-processing [1, 1216].
Over the last several years, deep-learning networks have become increasingly valuable and potent tools within the field of medical imaging for various tasks [17]. Deep-learning models are actively employed in the analysis of CT, computed radiography (CR), and magnetic resonance imaging (MRI) scans for purposes including, but not limited to, diagnostics, tumor segmentations, and image post-reconstruction under low dose or undersampling conditions [1821]. One field within the medical imaging domain that has recently sparked interest is medical image-to-image (I2I) translations. In this process, source domain images are transformed into synthetic images in a way that they adopt characteristics of target domain images [17, 22].
In recent years, a particular deep learning regime has shown to produce appealing results in I2I tasks, namely the Generative Adversarial Network (GAN) [23]. GANs are used for a variety of medical imaging I2I applications, such as contrast enhancement within CT scans, motion correction in MRI scans, PET to CT translations, and the de-noising of PET scans [17, 24]. To the group of GANs which have been proven to yield reliable results belong conditional GANs (cGANs) and especially the pix2pix [25] and its successor the pix2pixHD [26].
Several recent studies used artificial intelligence-based methods to enhance PET images acquired at low count rates (i.e., low acquisition times or low administered activities) [2735]. However, to our best knowledge, none of these studies used a combination of data acquisition on a digital PET/CT system and GAN-based image post-processing for whole-body PET data. Moreover, many approaches were restricted to brain PET images [3134], and in studies using whole-body data, the number of included patients was low [29, 35]. Additionally, acquisition times were longer than the maximum scan velocity of recent PET/CT scanners by a factor of at least 10 [27, 28]. Hence, the aim of this prospective phase I/II imaging study was to investigate the feasibility of acquiring maximum-speed ultra-short FDG ExtremePET images (scan time durations of about 30 s, > 33-fold reduced scan duration compared to clinical routine) on a digital PET/CT scanner and implementing a pix2pixHD network to recreate whole-body PET images (AI-ExtremePET) which are comparable to full acquisition time PET (FullTime-PET) scans. Several metrics including quantitative comparisons of PET signal recovery and manual image reading are evaluated to compare the physical image quality and the conservation of clinical information and lesion detectability among AI-ExtemePET and ground truth FullTime-PET images. We hypothesize that most of the diagnostically relevant information can be recovered from ExtremePET scans, whereas the quantitative evaluation of small and low count lesions might be challenging.

Materials and methods

Dataset

Patient cohort/ethics statement

This is a prospective, single-arm, single-center phase I/II imaging study. All patients who were referred for a clinical FDG PET/CT scan to the Department of Nuclear Medicine at the University Hospital of Essen, Germany, and who were scheduled for examination with a silicon-photomultiplier based PET/CT system between January 2020 and June 2020 were offered study participation in the order they appeared in clinical routine. Only patients < 18 years of age were excluded. A total of 587 patients were included in this prospective study (260 female patients and 327 male patients, mean age of 60.9 ± 13.2 and mean weight of 79.0 ± 18.7). The study was performed in accordance with the Declaration of Helsinki and approved by the local ethics committee (Ethics committee, University Duisburg-Essen, Faculty of Medicine, Ethics protocol number 20–9226-BO). Written informed consent was requested prior to enrollment.

FDG-PET/CT imaging

PET/CT data were acquired using a Biograph Vision 600 PET/CT System (Siemens Healthineers, Erlangen, Germany). PET/CT scans started with a CT scan in full-dose or low-dose technique according to the clinical routine protocol. Subsequently, reduced acquisition time PET data were first acquired in continuous bed motion mode using a table speed velocity of 50 mm/s (ExtremePET); the applied velocity was the fastest possible on the 26.3-cm field-of-view Biograph Vision PET/CT system. Normal acquisition time data were then acquired using a table speed velocity of 1.5 mm/s with an emphasis of the abdominal region by a reduced table speed velocity of 0.8 mm/s (FullTime-PET); in lung cancer patients instead of the abdominal region the thoracic region was emphasized. The mean ± SD applied activity was 327.8 ± 76.7 MBq of 18F-FDG. The mean ± SD interval between tracer application and start of PET scan was 74.6 ± 17.4 min. PET data were reconstructed using three-dimensional Poisson ordered-subsets expectation maximization with time-of-flight option (4 iterations 5 subsets, matrix size 220, voxel size 3.3 × 3.3 × 3.0 mm3, Gaussian filter 4 mm). Image datasets comprised a median number of 289 slices. Image examples are presented in Fig. 1.

Data preprocessing

Images extracted from two different methods, specifically PET and CT scans, were used as potential network inputs. Correspondingly, two different input settings were defined, namely PET and PET/CT.
For each method, 2.5-dimensional (2.5d) images were created using the axial slices. This is achieved by selecting a particular slice from the image volume, and then selecting the slices immediately prior and successive to the original slice. The first and last slices of each volume were ignored for the selection of central slices and only used to complete the 3-channel representation of the second and the penultimate slice, as, due to technical reasons in PET image acquisition, they contain mostly image noise and no clinically relevant information. Stacking results in a three-channel image; the generated 2.5d image output is a single image corresponding to the central slice that contains additional ​​information of the adjacent slides.
For the PET input setting, only 2.5d PET images were used as input for the pix2pixHD. For the PET/CT input setting, PET 2.5d and CT 2.5d images were combined on a channel axis to generate a six-channel image (Fig. 2). The six-channel image was then used as input. The intention behind the combined use of PET and CT scans is to investigate the hypothesis that CT images support the reconstruction of synthetic AI-ExtremePET images by providing more detailed anatomical information than the short acquisition time ExtremePET images.
All PET scans were transformed from Bq/ml to SUV units using the SUV Body Weight and normalized to a range of (0, 1) using a constant SUV maximum value of 50 to represent typical clinical conditions. Of note, information for lesions with SUV values > 50 could be lost. However, typical lesion SUV values in FDG PET imaging are much lower than 50 [36]. All CT scans were first resampled using the corresponding PET scan in order to match the resolution and the dimensions of the PET scans. The CT scans were then normalized to a range of (0, 1) using (− 1000, 3000) as the Hounsfield unit minimum and maximum. The resulting input samples used a resolution of 220 × 220 pixels and were then padded to a resolution of 224 × 224 pixels. Finally, the normalized slices were transformed from (0, 1) to (− 1, 1).

Network architecture

General Adversarial Networks consist of two parts, a generator and a discriminator, in competition with each other. Within an image-to-image translation setting, the generator takes an input image from a source domain and generates a corresponding image with characteristics of the target domain. The discriminator then has to distinguish between real images and the images produced by the generator. The joint training process involves optimizing both the ability of the generator to produce images as similar as possible to the target domain as well as the ability of the discriminator to assign correct labels to sample images [25, 26].
For this study, we implemented a modified TensorFlow version of the pix2pixHD. The pix2pixHD is an extension of the pix2pix architecture, but, unlike the original pix2pix, uses two generators that operate with different resolutions in order to aggregate both local and global features of the image. This pair of generators constitutes a “coarse-to-fine” generator. The discriminator of the pix2pixHD is a multi-scale discriminator that itself employs three separate discriminators, each operating on a different image scale [25, 26].
The model was trained using a GAN and a feature-matching loss function in combination, as proposed by Wang et al. [26]. Additionally, we implemented an average pooling operation within the first downsample layers. For the first convolution in the generators’ encoders, we also added the option of a group convolution. This allows us to train models that process PET and CT images separately in the first convolution layer using 2 groups with 3 channels each. All models use an initial convolution filter size of 7 × 7 and an average pooling operation in the encoders. A batch size of four images was used during training, and the output channel was set to grayscale. A learning rate of 0.002 was used, and all models were trained with 100 epochs, with a learning rate decay starting after 50 epochs. A complete list of the used model settings is listed within the Supplemental Material (Table 3).
A total of three model configurations were used and trained: a model that uses only ExtremePET images (M1), an ExtremePET/CT model that uses both ExtremePET and CT images without group convolution (M2), and an ExtremePET/CT model that uses both ExtremePET and CT images with group convolution (M3). All model configurations were trained using a fivefold cross-validation (CV), and the final prediction for each model was aggregated using the mean overall CV model predictions.
All models used an image input size of 224 × 224. A total of 387 studies were used as training, and 200 studies were used as a test cohort.

Evaluation methods

For this study, the evaluation process was performed in multiple parts. Firstly, all trained models were evaluated on the 200 test studies using the well-established I2I metrics Structural Similarity Index Measure (SSIM), Peak Signal to Noise Ratio (PSNR), and SUV-based Mean Absolute Error (MAE) [3739]. For calculation of the scores, the voxel space outside the patient was excluded (masked with 0) using patient specific body masks. With this approach, we want to ensure that the score calculation is focused on the important region of interest, the body of the patient [40].
Secondly, 50 studies were randomly selected from the 200 test studies and manually examined by an experienced nuclear medicine physician. First, the human reader segmented all clinically relevant lesions visible within the FullTime-PETscan (excluding physiological tracer uptake). In addition, all detected lesions were separated into the following anatomical categories: bones, liver, lung, lymph nodes, and other. Determination of anatomical labels was performed with the assistance of a ​​software research prototype implementing a neural network (MICIIS, formerly MI Whole Body Analysis Suite, MIWBAS, Siemens Healthineers, Knoxville, TN, United States) [41]. The lesion segmentations were then used to evaluate and compare the SUV metrics SUVmean, SUVmax, and SUVpeak for all models.
The synthetic AI-ExtremePET images of the best performing model (with respect to lesion SUV recovery) were then evaluated by the human reader to identify and segment all clinically relevant lesions (excluding physiological tracer uptake). All evaluations of synthetic AI-ExtremePET images were performed on a separate occasion from the evaluations of the original FullTime-PET scans to avoid bias due to prior knowledge. This leads to a total set of 50 Fulltime-PET segmentation masks and corresponding 50 AI-ExtremePET segmentation masks.
Lastly, the prepared segmentations were used to evaluate the lesion reconstruction quality of the best performing model. First, on the patient level, the detection of any lesion (dichotomous variable for each study) was compared among AI-ExtremePET and FullTimePET images. Next, to evaluate detection at the correct anatomical position, the given segmentation masks of the original and synthetic PETs were compared based on the Intersection over Union (IoU). In this analysis, a lesion is considered detected if the IoU is > 0. The IoU threshold is set to the proposed level because the compared masks are based on different PET images (original and synthetic). With this setting, we want to ensure that the evaluation is focused on the model’s ability to reconstruct lesions at the correct position without demanding perfect voxel matching. If more than one synthetic lesion candidate exists for an original lesion, the candidate with the highest IoU is selected. Detection at the correct anatomical position was evaluated on a patient, organ, and lesion level. Patient-based detection rate was defined as described above. On the organ level, the detection of any lesion in the specific organ (dichotomous variable for each organ and each study) was compared among AI-ExtremePET and FullTimePET images. Lesion-based detection rate includes all detected lesions in the FullTimePET images. For these evaluations, additionally, all lesions were separated into two groups: PERCIST-measurable and non PERCIST-measurable. The PERCIST group contains all lesions which satisfy the PERCIST criteria (SUVpeak > 1.5 × mean(SUVliver) + 2 × std(SUVliver)) [42, 43]. Detection rates at the correct anatomical position were then separately calculated for PERCIST-measurable and non PERCIST-measurable lesions.

Statistics

For statistical analysis, we used the two-sided nonparametric Mann–Whitney U [44] test using the python package scipy [45].

Results

Synthetic PET images for all three models were of high visual image quality and showed significant improvements in SSIM, PSNR, and MAE compared to the ultra-short ExtemePET images (using the FullTime PET as ground truth, details in Supplemental Material, Fig. 6). To select a model for detailed evaluation, first, a comparison of the lesion SUV reconstruction quality was performed. For this approach, the lesion masks (containing all lesions that were detected by a human reader in the FullTime-PET) are used to compare SUVmean, SUVmax, and SUVpeak differences (between FullTime-PET and AI-ExtremePET images) for each model. Of 50 manually evaluated cases, 33 contained lesions (66%). Within those 33 cases, 298 lesions were detected by the human reader. The organ-based SUV level comparison is presented in Table 1.
Table 1
Differences of the SUVmean, SUVmax, and SUVpeak values between the FullTime-PET and the AI-ExtremePET images for each model based on the original lesion segmentation mask. The SUVmean is calculated based on the mean SUV values present within the lesion mask. The SUVmax is calculated using the maximum SUV value for each lesion mask. The SUVpeak is calculated using a 1-cm3 sphere around the maximum voxel within a lesion mask
Body part
M1
M2
M3
 
SUVmean
SUVmax
SUVpeak
SUVmean
SUVmax
SUVpeak
SUVmean
SUVmax
SUVpeak
All
1.01 ± 1.43
1.87 ± 2.28
0.76 ± 1.60
2.11 ± 1.65
3.45 ± 2.60
1.94 ± 1.66
0.91 ± 1.54
1.50 ± 2.46
0.57 ± 1.74
Bones
1.02 ± 0.79
2.15 ± 1.86
0.72 ± 0.96
2.84 ± 1.19
4.84 ± 2.21
2.57 ± 1.39
1.05 ± 1.05
1.90 ± 2.13
0.63 ± 1.26
Liver
1.26 ± 0.50
2.62 ± 1.54
1.41 ± 0.76
2.09 ± 0.87
3.43 ± 1.95
2.31 ± 1.19
1.07 ± 0.62
2.09 ± 1.98
1.01 ± 0.98
Lung
1.16 ± 1.04
1.89 ± 1.63
0.82 ± 0.88
2.08 ± 1.01
3.05 ± 1.67
1.81 ± 1.09
0.87 ± 0.99
1.04 ± 1.94
0.39 ± 1.19
Lymph nodes
1.21 ± 0.86
1.93 ± 1.53
0.81 ± 0.87
2.01 ± 1.14
3.11 ± 1.69
1.83 ± 1.11
1.03 ± 1.00
1.38 ± 1.73
0.54 ± 0.89
Other
2.51 ± 3.73
3.78 ± 5.15
2.33 ± 4.48
3.06 ± 3.05
4.63 ± 3.80
2.94 ± 3.09
2.34 ± 3.84
3.45 ± 5.36
2.09 ± 4.69
The results show that model M3 performs best for lesion evaluation. We therefore used the M3 AI-ExtremePET images for further evaluation. Figure 3 depicts reconstructed slices from model M3, including a difference map (FullTime-PET versus AI-ExtemePET) for multiple cases.
Patient-based sensitivity and specificity for lesion detection were 79% and 100%, respectively. An exemplary patient image showing a correctly detected and a missed lesion in the AI-ExtremePET images is presented in Fig. 4.
Next, we separately calculated patient-, organ-, and lesion-based detection rates in a matched pair approach. The patient-based detection rate was 79% for all lesions. Lesions that were not detected in the AI-ExtremePET images were of smaller volume (median volume: 1.0 ml versus 2.7 ml, p = 0.06) and tracer uptake in the ExtremePET than correctly detected lesions (median SUVpeak: 3.1 versus 4.9, p = 0.05; median SUVmean: 2.7 versus 4.6, p < 0.001). We, therefore, split the data according to the clinically established PERCIST [43] criteria. From 298 detected lesions, 229 were PERCIST-measurable. The patient-based detection rate was 89% regarding only PERCIST-measurable lesions, and 36% regarding only non PERCIST-measurable lesions. The lesion-based detection rate was 65% for all lesions, 78% for PERCIST-measurable lesions, and 22% for non PERCIST-measurable lesions. Detailed organ-based detection statistics (indicating detection rates, mean SUV levels, and lesion volumes) are presented in Table 2.
Table 2
Detailed lesion-based detection characteristics. SUV values were estimated in the ExtremePET images and based on the original lesion masks
Body part
Lesion cohort
Detected
n
Volume (ml)
SUVmax
SUVmean
SUVpeak
Detection rate
All
PERCIST
True
178
8.56 ± 21.13
12.87 ± 7.22
5.78 ± 2.98
6.62 ± 4.49
0.78
False
51
1.69 ± 2.64
9.75 ± 5.23
4.06 ± 1.35
4.15 ± 1.33
Other
True
15
5.69 ± 9.72
7.62 ± 2.86
2.87 ± 0.98
2.90 ± 0.85
0.22
False
54
1.87 ± 2.55
5.82 ± 2.73
2.07 ± 0.65
2.35 ± 0.92
Bones
PERCIST
True
119
3.60 ± 3.27
12.03 ± 5.22
5.65 ± 2.33
6.07 ± 2.94
0.84
False
23
2.15 ± 3.37
8.07 ± 2.23
4.29 ± 1.11
3.79 ± 1.01
Other
True
5
0.85 ± 0.65
5.89 ± 2.74
3.19 ± 1.35
2.46 ± 0.70
0.42
False
7
1.10 ± 1.36
5.36 ± 0.93
2.13 ± 0.35
2.67 ± 1.09
Liver
PERCIST
True
5
32.74 ± 33.29
12.93 ± 2.47
4.94 ± 0.74
6.20 ± 0.94
0.83
False
1
1.63
7.87
4.01
4.27
Other
True
-
-
-
-
-
0.0
False
3
1.61 ± 2.79
8.73 ± 2.52
2.74 ± 0.29
3.48 ± 0.44
Lung
PERCIST
True
21
5.84 ± 11.18
12.34 ± 8.41
5.25 ± 3.57
6.01 ± 5.33
0.88
False
3
1.47 ± 1.51
6.97 ± 2.18
3.51 ± 1.44
3.50 ± 0.94
Other
True
3
2.35 ± 1.61
7.58 ± 3.33
2.21 ± 1.12
2.30 ± 0.87
0.23
False
10
1.16 ± 1.26
5.06 ± 3.79
1.70 ± 0.81
1.58 ± 0.67
Lymph nodes
PERCIST
True
27
11.81 ± 17.64
13.87 ± 7.36
5.96 ± 3.78
7.40 ± 4.88
0.63
False
16
1.54 ± 2.08
10.83 ± 6.20
4.10 ± 1.68
4.41 ± 1.72
Other
True
6
12.19 ± 13.34
8.81 ± 2.71
2.79 ± 0.46
3.56 ± 0.62
0.18
False
27
2.68 ± 3.17
6.04 ± 2.42
2.20 ± 0.54
2.47 ± 0.77
Other
PERCIST
True
6
81.77 ± 69.68
26.97 ± 19.27
10.10 ± 6.02
16.54 ± 11.88
0.43
False
8
0.75 ± 1.50
13.70 ± 7.89
3.54 ± 1.35
4.91 ± 1.22
Other
True
1
0.88
9.25
3.79
2.89
0.12
False
7
0.64 ± 0.98
5.28 ± 3.13
1.76 ± 0.86
2.17 ± 1.01
Moreover, lesion volume differences between AI-Extreme-PET and FullTime-PET images were lower for PERCIST-measurable in comparison to non PERCIST-measurable lesions, whereas the IoU was higher (Supplemental Material, Table 4). This, additionally, indicates an improved reproduction of PERCIST-measurable lesions.
PERCIST-measurable lesions that were not detected in the AI-ExtremePET images were of significantly smaller volume (median volume: 1.0 ml versus 2.7 ml, p < 0.0001) and tracer uptake (median SUVpeak: 4.0 versus 5.1, p < 0.0001; median SUVmean: 3.8 versus 4.9, p < 0.0001) in the ExtremePET than correctly detected lesions (Fig. 5). For non PERCIST-measurable lesions, not detected lesions were of significantly smaller tracer uptake (median SUVpeak: 2.4 versus 3.0, p = 0.05; median SUVmean: 2.2 versus 3.0, p < 0.001) but showed no significant difference in lesion volume (median volume: 1.1 ml versus 2.0 ml, p = 0.06).

Discussion

In this study, we demonstrated the use of a pix2pixHD network to generate synthetic full acquisition time PET images from 33-fold reduced acquisition time ExtremePET images that were exclusively acquired on a “digital” silicon-photomultiplier based PET/CT scanner. Only few previous approaches have been described that use GANs for enhancement of low-count whole-body PET images [27, 28, 46]. Some previously published approaches were restricted to brain PET imaging [17, 3134, 47], which is an easier process due to the limited anatomical variance compared to whole-body images [35]. In the following, we intend to evaluate the results of this study in the context of whole-body imaging.
Some approaches for whole-body low-count imaging enhancement do not implement GANs [28, 48, 49]. Compared to the few previous works that use GANs on whole-body PET images, in our study the acquisition time was shorter by about one magnitude, and the number of included patients was significantly larger. For example, Lei et al. use a CycleGAN for enhancement of 25 whole-body eightfold reduced PET data sets acquired on a conventional PET scanner [46]. Sanaat et al. implemented a ResNET and a CycleGAN and included 85 conventional sevenfold reduced PET data sets [27]. No previous study focused on the benefits of a digital PET scanner for low-count PET imaging, and no study that used a digital scanner implemented a GAN. Kaplan and Zhu used an estimator and an adversarial discriminator network on tenfold reduced digital PET data but included only 2 patients [49]. Most recently, Chaudhari et al. applied a commercially available encoder-decoder U-Net based solution on 50 fourfold reduced PET data sets that were only in parts acquired on a digital scanner [28]. The improved detector sensitivity, time and spatial resolution, and noise characteristics [3, 5, 6, 11] of SiPM-based PET provide the basis for acquiring ultra-low-count PET data with a whole-body scan time comparable to a CT scan.
In this proof-of-concept approach, we used the maximum possible PET scan velocity of a latest-generation digital PET/CT scanner. A visual evaluation revealed the technical feasibility to generate PET images of high visual image quality from noisy ultra-short acquisition time data (example in Fig. 5). A preserved high visual image quality is the prerequisite for further evaluation in terms of quantification and detectability, and a first relevant finding since the acquisition time in this study was in the range of a standard CT scan and, thus, significantly shorter than in previously published studies. Next, we compared SSIM, PSNR, and MAE, commonly applied criteria for assessment of quantification performance [27], among three different models using PET data only, PET/CT data without group convolution, or PET/CT data with group convolution in comparison to the ground truth full acquisition time images. The results were excellent for all three models and significantly improved compared to the ExtremePET images without post-reconstruction (Fig. 3). Next, we analyzed SUV differences for all lesions that were detected and segmented by a human reader in the ground truth FullTime PET images. This analysis showed best performance for the combined input model (M3) with group convolution (Table 1). We therefore selected this model for detailed evaluation. In most previous works, only PET data were used as input [2729, 48, 49]. However, our results are in line with a PET/MRI study that describes benefits by simultaneous input of PET and MRI images for enhancement of ultra-low-dose PET images in children [35].
The detailed lesion quantification analysis for all lesions that were detected in the ground truth images revealed a mean ± SD absolute SUVmax difference of 1.5 ± 2.5 and a mean ± SD absolute SUVmean difference of 0.9 ± 1.6. These are appealing results in the context of the extremely short acquisition time. For previous approaches, using longer acquisition times, lower SUV differences were described. Chaudhari et al. report a lesion mean SUV difference of approximately zero (95%-confidence limit of 1.8 for the SUVmax of the lesions) [28]. Sanaat et al. indicate a mean SUVmax difference of − 0.01 [27]. However, slight SUV differences, as in our study, do not influence patient management in most clinical circumstances, as clinical image reading in most cases does not depend on exact lesion quantification but on detectability of lesions with increased tracer uptake. The reason for the reduced SUV reproduction in our account approach is most probably by a factor of ≥ 10 reduced acquisition time in our approach and not a structural problem of the applied GAN. With regard to the ultra-low-count PET data we used, deviations in quantification are expectable, and exact quantification was not an aim of this study.
More than lesion quantification, lesion detectability is decisive to assess the possible value of ultra-short acquisition time PET images in a clinical context, as missed lesions can be of big influence on patient management. The patient-based analysis revealed a sensitivity of 79% and specificity of 100%; no patients were classified as false-positive. This is a relevant finding, as misclassification of patients as false-positive could have consequences to further patient/therapy management. Chaudari et al. reached a patient-level sensitivity of 94% and specificity of 98%, Sanaat et al. do not report a patient-based analysis but a lesion-based sensitivity of 97% [27]. The detailed analysis showed that missed lesions were of lower tracer uptake and lesion volume in the input ExtremePET images. To evaluate the clinical usability, we, therefore, performed additional separate analyses for PERCIST-measurable and non PERCIST-measurable lesions in a matched-pair approach and calculated patient-based, organ-based and lesion-based detection rates. These were largely increased for PERCIST-measurable lesions (for example, 89% versus 36% on the patient level and 78% versus 22% on the lesion level, details in Table 2). For PERCIST-measurable lesions not detected lesions were of significantly smaller tracer uptake (SUVmean) and volume, whereas for non PERCIST-measurable lesions only the difference in tracer uptake was significant (Fig. 5).
As the reproduction of lesions with low tracer uptake and small lesion volume remains challenging, the clinical applicability of AI-ExtremePET images is limited. As detectability was restricted for small lesions with low tracer uptake, primary staging and investigation of patients in the early phase of disease will probably be problematic. The analysis of PERCIST-measurable lesions showed an improved detectability for larger lesions of higher tracer uptake. Moreover, for PERCIST-measurable lesions the volume reproduction was improved (Supplemental Material, Table 4). Therefore, a possible setting for a clinical use might be follow-up of metastatic cancer patients in whom an evaluation of total tumor burden (with, potentially, high tracer uptake) is of larger clinical relevance than detection of single lesions. In this context, typically pain-stricken patients with high tumor burden could benefit from a short acquisition time. Reduced motion artifacts can be an additional advantage of short emission time PET scans [28].
To assess whether the AI-ExtremePET technique is suitable for follow-up staging, future studies could investigate reproducibility of quantitative measures and metabolic tumor volumes. Typically, test–retest SUV deviation in FDG PET scans is about ± 20% [50]. A detailed investigation would require two separate PET scans in short temporal distance and is, therefore, not possible using the data set of this study. Moreover, future studies could evaluate patients with a follow-up scan to investigate whether AI-ExtremePET images can be used for oncological response assessment.
For clinical applications in which a characterization of single lesions is decisive (e.g., initial staging or detection of primary tumor), accuracy of lesion quantification and detectability must be improved. Future studies might focus on probing different acquisition times to investigate the optimal tradeoff between scan duration and conservation of clinical information. Since previous studies, which show better results, use substantially lower reductions in acquisition time [27, 28], an investigation of the acquisition time range between their approach and ours could be promising. For example, an extension of the ExtremePET acquisition time by a factor of 2–5 could be investigated. However, an acquisition time optimization was beyond the protocol of this prospective study in which the fastest possible acquisition time of a current-generation digital PET/CT scanner was evaluated. Moreover, future approaches might use data acquired on total-body PET/CT scanners that exhibit improved counting statistics and might allow for even shorter acquisition times than standard digital PET/CT scanners while maintaining acceptable image quality [50]. However, until now, total-body PET scanners are expensive and have not widespreadly been introduced. As reductions in acquisition time below the scan duration that was applied in this study might not be more beneficial in daily practice, deep-learning based image enhancement might in future be used for total-body PET data to reduce the administered activity. A reduced acquisition time can be used as a surrogate for a reduced administered tracer activity, as these values, in a first approach, correlate linearly [16, 51].
In our study, ExtremePET data were used that were acquired in a separate ultra-short scan, whereas most previous studies use simulated short acquisition time data that are created by undersampling of original list-mode data sets [29, 35, 48, 49]. Sanaat et al. also use separately acquired low-dose PET data [27]. Chaudhari et al. describe a multicenter study that uses undersampled data from one center and separately acquired data sets from two centers [28]. The usage of real data is beneficial, since larger image noise in short acquisition time data sets might not be fully reproduced by undersampling [27, 51]. On the other hand, the separate PET data acquisition contributes to a further main limitation of the study. Possible deviations in spatial pairing between FullTime-PET and ExtremePET images due to patient motion might impair the reconstruction quality of the applied Pix2PixHD network. However, the ExtremePET scan was started directly after the FullTime-PET scan using the same scan area and patient position; this ensures a good co-registration between both image data sets. Moreover, most clinically relevant lesions are located in the trunk for which motion is low. Therefore, the benefits of separate PET acquisitions justify the limitations of spatial deviations.
Besides the Pix2PixHD that was used in this study, other neural network architectures could be used for I2I translations of PET images. For example, a cycleGAN setting could be used, which would tackle the problem of the spatial alignment of the different image types since the CycleGAN is used for unpaired image to image tasks [52]. Another future possibility might be the application of a transformer-based CycleGAN. These networks may be less susceptible to variations in spatial pairing and could therefore be promising for PET images, which are of lower resolution and higher image noise than CT or MRI images. A comparison of different networks could be the subject of future studies.
The study is affected by further limitations. Only PET/CT data from a single center and a single PET/CT scanner were included in the evaluation. For generalizability, a multicenter study using digital PET/CT scanners from different vendors should be performed. However, generalizability was not an aim of this proof-of-concept study, as first further steps of improvement and validation are necessary. In addition, the trained network only used stacked 2D axial slices within a 2.5D approach which could be updated within further studies to a 3D approach. An attention-weighted loss function could be used to emphasize the most significant body parts [32, 33]. Furthermore, the study was only performed for 18F-FDG PET/CT imaging. Future studies could cover different PET tracers. Previous approaches were described for brain PET imaging using 18F-Florbetaben [34], whereas for DOTATATE/DOTATOC and PSMA PET, to our best knowledge, no deep learning models to enhance low count images have been described.

Conclusion

A combination of digital PET/CT and artificial intelligence-based image post-reconstruction allows the generation of high quality images from PET data that were acquired as fast as CT scans. Detectability (79% on a patient level) and lesion quantification revealed promising results; lesion tracer uptake and volume were lower for not-detected lesions. In the current form, the number of missed lesions still prevents a broad clinical use, but the approach could be applied in late-stage cancer patients to monitor total tumor burden. Future studies investigating ultra-fast PET imaging are warranted.

Declarations

Ethics approval

This study was conducted in compliance with the guidelines of the Institutional Review Board of the University Hospital Essen (approval number 20–9226-BO). The data were completely anonymized before being included in the study.
Informed consent was obtained from participants included in the study.
Informed consent was obtained from participants included in the study.

Competing interests

Lale Umutlu is a Speaker/Advisory Board Member for Bayer Healthcare and Siemens Healthcare and received research grants from Siemens Healthcare outside of the submitted work. Ken Herrmann reports personal fees from Bayer SIRTEX Adacap Curium Endocyte IPSEN Siemens Healthineers GE Healthcare Amgen Novartis and ymabs personal fees and other from Sofie Biosciences non-financial support from ABX grants and personal fees from BTG outside the submitted work. Christoph Rischpler reports a research grant from Pfizer, consultancy for Adacap and Pfizer, speaker honoraria from Adacap, Alnylam, BTG, Curium, GE Healthcare, Pfizer, and Siemens Healthineers. David Kersting is supported by the Clinician Scientist Program of the Universitätsmedizin Essen Clinician Scientist Academy (UMEA)/German Research Foundation (DFG, Deutsche Forschungsgemeinschaft) and received research funding from Pfizer outside of the submitted work. Robert Seifert is supported by the Junior Clinician Scientist Program of the Universitätsmedizin Essen Clinician Scientist Academy (UMEA)/German Research Foundation (DFG, Deutsche Forschungsgemeinschaft). Manuel Weber reports fees from Boston Scientific, Terumo, Eli Lilly, and Advanced Accelerator Applications.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher's note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

Anhänge

Supplementary Information

Below is the link to the electronic supplementary material.
Literatur
1.
Zurück zum Zitat Boellaard R, Delgado-Bolton R, Oyen WJG, Giammarile F, Tatsch K, Eschner W, et al. FDG PET/CT: EANM procedure guidelines for tumour imaging: version 2.0. Eur J Nucl Med Mol Imaging. 2015;42:328–54.CrossRef Boellaard R, Delgado-Bolton R, Oyen WJG, Giammarile F, Tatsch K, Eschner W, et al. FDG PET/CT: EANM procedure guidelines for tumour imaging: version 2.0. Eur J Nucl Med Mol Imaging. 2015;42:328–54.CrossRef
2.
Zurück zum Zitat Lasnon C, Coudrais N, Houdu B, Nganoa C, Salomon T, Enilorac B, et al. How fast can we scan patients with modern (digital) PET/CT systems? Eur J Radiol. 2020;129:109144.CrossRef Lasnon C, Coudrais N, Houdu B, Nganoa C, Salomon T, Enilorac B, et al. How fast can we scan patients with modern (digital) PET/CT systems? Eur J Radiol. 2020;129:109144.CrossRef
3.
Zurück zum Zitat Van Sluis J, De Jong J, Schaar J, Noordzij W, Van Snick P, Dierckx R, et al. Performance characteristics of the digital biograph vision PET/CT system. J Nucl Med Off Publ Soc Nucl Med United States. 2019;60:1031–6. Van Sluis J, De Jong J, Schaar J, Noordzij W, Van Snick P, Dierckx R, et al. Performance characteristics of the digital biograph vision PET/CT system. J Nucl Med Off Publ Soc Nucl Med United States. 2019;60:1031–6.
4.
Zurück zum Zitat Surti S, Viswanath V, Daube-Witherspoon ME, Conti M, Casey ME, Karp JS. Benefit of improved performance with state-of-the art digital PET/CT for lesion detection in oncology. J Nucl Med Off Publ Soc Nucl Med United States. 2020;61:1684–90. Surti S, Viswanath V, Daube-Witherspoon ME, Conti M, Casey ME, Karp JS. Benefit of improved performance with state-of-the art digital PET/CT for lesion detection in oncology. J Nucl Med Off Publ Soc Nucl Med United States. 2020;61:1684–90.
5.
Zurück zum Zitat Koopman D, van Dalen JA, Stevens H, Slump CH, Knollema S, Jager PL. Performance of digital PET compared with high-resolution conventional PET in patients with cancer. J Nucl Med Off Publ Soc Nucl Med United States. 2020;61:1448–54. Koopman D, van Dalen JA, Stevens H, Slump CH, Knollema S, Jager PL. Performance of digital PET compared with high-resolution conventional PET in patients with cancer. J Nucl Med Off Publ Soc Nucl Med United States. 2020;61:1448–54.
6.
Zurück zum Zitat van Sluis J, Boellaard R, Somasundaram A, van Snick PH, Borra RJH, Dierckx RAJO, et al. Image quality and semiquantitative measurements on the biograph vision PET/CT system: initial experiences and comparison with the biograph mCT. J Nucl Med Off Publ Soc Nucl Med United States. 2020;61:129–35. van Sluis J, Boellaard R, Somasundaram A, van Snick PH, Borra RJH, Dierckx RAJO, et al. Image quality and semiquantitative measurements on the biograph vision PET/CT system: initial experiences and comparison with the biograph mCT. J Nucl Med Off Publ Soc Nucl Med United States. 2020;61:129–35.
7.
Zurück zum Zitat Kersting D, Jentzen W, Sraieb M, Costa PF, Conti M, Umutlu L, et al. Comparing lesion detection efficacy and image quality across different PET system generations to optimize the iodine-124 PET protocol for recurrent thyroid cancer. EJNMMI Phys. 2021;8:14.CrossRef Kersting D, Jentzen W, Sraieb M, Costa PF, Conti M, Umutlu L, et al. Comparing lesion detection efficacy and image quality across different PET system generations to optimize the iodine-124 PET protocol for recurrent thyroid cancer. EJNMMI Phys. 2021;8:14.CrossRef
8.
Zurück zum Zitat López-Mora DA, Flotats A, Fuentes-Ocampo F, Camacho V, Fernández A, Ruiz A, et al. Comparison of image quality and lesion detection between digital and analog PET/CT. Eur J Nucl Med Mol Imaging. 2019;46:1383–90.CrossRef López-Mora DA, Flotats A, Fuentes-Ocampo F, Camacho V, Fernández A, Ruiz A, et al. Comparison of image quality and lesion detection between digital and analog PET/CT. Eur J Nucl Med Mol Imaging. 2019;46:1383–90.CrossRef
9.
Zurück zum Zitat Alberts I, Prenosil G, Sachpekidis C, Weitzel T, Shi K, Rominger A, et al. Digital versus analogue PET in [(68)Ga]Ga-PSMA-11 PET/CT for recurrent prostate cancer: a matched-pair comparison. Eur J Nucl Med Mol Imaging Germany. 2020;47:614–23.CrossRef Alberts I, Prenosil G, Sachpekidis C, Weitzel T, Shi K, Rominger A, et al. Digital versus analogue PET in [(68)Ga]Ga-PSMA-11 PET/CT for recurrent prostate cancer: a matched-pair comparison. Eur J Nucl Med Mol Imaging Germany. 2020;47:614–23.CrossRef
10.
Zurück zum Zitat Kersting D, Jentzen W, Fragoso Costa P, Sraieb M, Sandach P, Umutlu L, et al. Silicon-photomultiplier-based PET/CT reduces the minimum detectable activity of iodine-124. Sci Rep. 2021;11:17477.CrossRef Kersting D, Jentzen W, Fragoso Costa P, Sraieb M, Sandach P, Umutlu L, et al. Silicon-photomultiplier-based PET/CT reduces the minimum detectable activity of iodine-124. Sci Rep. 2021;11:17477.CrossRef
11.
Zurück zum Zitat Conti M, Bendriem B. The new opportunities for high time resolution clinical TOF PET. Clin Transl Imaging. 2019;7:139–47.CrossRef Conti M, Bendriem B. The new opportunities for high time resolution clinical TOF PET. Clin Transl Imaging. 2019;7:139–47.CrossRef
12.
Zurück zum Zitat Hatami S, Frye S, McMunn A, Botkin C, Muzaffar R, Christopher K, et al. Added value of digital over analog PET/CT: more significant as image field of view and body mass index increase. J Nucl Med Technol United States. 2020;48(354):360. Hatami S, Frye S, McMunn A, Botkin C, Muzaffar R, Christopher K, et al. Added value of digital over analog PET/CT: more significant as image field of view and body mass index increase. J Nucl Med Technol United States. 2020;48(354):360.
13.
Zurück zum Zitat Weber M, Jentzen W, Hofferber R, Herrmann K, Fendler WP, Rischpler C, et al. Evaluation of (18)F-FDG PET/CT images acquired with a reduced scan time duration in lymphoma patients using the digital biograph vision. BMC Cancer. 2021;21:62.CrossRef Weber M, Jentzen W, Hofferber R, Herrmann K, Fendler WP, Rischpler C, et al. Evaluation of (18)F-FDG PET/CT images acquired with a reduced scan time duration in lymphoma patients using the digital biograph vision. BMC Cancer. 2021;21:62.CrossRef
15.
Zurück zum Zitat Weber M, Jentzen W, Hofferber R, Herrmann K, Fendler WP, Conti M, et al. Evaluation of [(68)Ga]Ga-PSMA PET/CT images acquired with a reduced scan time duration in prostate cancer patients using the digital biograph vision. EJNMMI Res. 2021;11:21.CrossRef Weber M, Jentzen W, Hofferber R, Herrmann K, Fendler WP, Conti M, et al. Evaluation of [(68)Ga]Ga-PSMA PET/CT images acquired with a reduced scan time duration in prostate cancer patients using the digital biograph vision. EJNMMI Res. 2021;11:21.CrossRef
16.
Zurück zum Zitat van Sluis J, Boellaard R, Dierckx RAJO, Stormezand GN, Glaudemans AWJM, Noordzij W. Image quality and activity optimization in oncologic (18)F-FDG PET using the digital biograph vision PET/CT system. J Nucl Med Off Publ Soc Nucl Med United States. 2020;61:764–71. van Sluis J, Boellaard R, Dierckx RAJO, Stormezand GN, Glaudemans AWJM, Noordzij W. Image quality and activity optimization in oncologic (18)F-FDG PET using the digital biograph vision PET/CT system. J Nucl Med Off Publ Soc Nucl Med United States. 2020;61:764–71.
17.
Zurück zum Zitat Armanious K, Jiang C, Fischer M, Küstner T, Hepp T, Nikolaou K, et al. MedGAN: Medical image translation using GANs. Comput Med Imaging Graph. 2019;79:101684.CrossRef Armanious K, Jiang C, Fischer M, Küstner T, Hepp T, Nikolaou K, et al. MedGAN: Medical image translation using GANs. Comput Med Imaging Graph. 2019;79:101684.CrossRef
18.
Zurück zum Zitat Cao H, Wang Y, Chen J, Jiang D, Zhang X, Tian Q, et al. Swin-Unet: Unet-like pure transformer for medical image segmentation. ArXiv210505537 Cs Eess [Internet]. 2021 [cited 2021 Jun 15]; Available from: http://arxiv.org/abs/2105.05537 Cao H, Wang Y, Chen J, Jiang D, Zhang X, Tian Q, et al. Swin-Unet: Unet-like pure transformer for medical image segmentation. ArXiv210505537 Cs Eess [Internet]. 2021 [cited 2021 Jun 15]; Available from: http://​arxiv.​org/​abs/​2105.​05537
19.
Zurück zum Zitat Gong E, Pauly JM, Wintermark M, Zaharchuk G. Deep learning enables reduced gadolinium dose for contrast-enhanced brain MRI. J Magn Reson Imaging JMRI. 2018;48:330–40.CrossRef Gong E, Pauly JM, Wintermark M, Zaharchuk G. Deep learning enables reduced gadolinium dose for contrast-enhanced brain MRI. J Magn Reson Imaging JMRI. 2018;48:330–40.CrossRef
20.
Zurück zum Zitat Haubold J, Hosch R, Umutlu L, Wetter A, Haubold P, Radbruch A, et al. Contrast agent dose reduction in computed tomography with deep learning using a conditional generative adversarial network. Eur Radiol. 2021;31:6087–95.CrossRef Haubold J, Hosch R, Umutlu L, Wetter A, Haubold P, Radbruch A, et al. Contrast agent dose reduction in computed tomography with deep learning using a conditional generative adversarial network. Eur Radiol. 2021;31:6087–95.CrossRef
21.
Zurück zum Zitat Zhao J, Li D, Kassam Z, Howey J, Chong J, Chen B, et al. Tripartite-GAN: Synthesizing liver contrast-enhanced MRI to improve tumor detection. Med Image Anal. 2020;63:101667.CrossRef Zhao J, Li D, Kassam Z, Howey J, Chong J, Chen B, et al. Tripartite-GAN: Synthesizing liver contrast-enhanced MRI to improve tumor detection. Med Image Anal. 2020;63:101667.CrossRef
22.
Zurück zum Zitat Kaji S, Kida S. Overview of image-to-image translation by use of deep neural networks: denoising, super-resolution, modality conversion, and reconstruction in medical imaging. Radiol Phys Technol. 2019;12:235–48.CrossRef Kaji S, Kida S. Overview of image-to-image translation by use of deep neural networks: denoising, super-resolution, modality conversion, and reconstruction in medical imaging. Radiol Phys Technol. 2019;12:235–48.CrossRef
24.
Zurück zum Zitat Seibold C, Fink MA, Goos C, Kauczor H-U, Schlemmer H-P, Stiefelhagen R, et al. Prediction of low-kev monochromatic images from polyenergetic CT scans for improved automatic detection of pulmonary embolism. 2021 IEEE 18th Int Symp Biomed Imaging ISBI. 2021. p. 1017–20. Seibold C, Fink MA, Goos C, Kauczor H-U, Schlemmer H-P, Stiefelhagen R, et al. Prediction of low-kev monochromatic images from polyenergetic CT scans for improved automatic detection of pulmonary embolism. 2021 IEEE 18th Int Symp Biomed Imaging ISBI. 2021. p. 1017–20.
25.
Zurück zum Zitat Isola P, Zhu J-Y, Zhou T, Efros AA. Image-to-image translation with conditional adversarial networks. IEEE Conf Comput Vis Pattern Recognit CVPR. 2017. p. 5967–76. Isola P, Zhu J-Y, Zhou T, Efros AA. Image-to-image translation with conditional adversarial networks. IEEE Conf Comput Vis Pattern Recognit CVPR. 2017. p. 5967–76.
26.
Zurück zum Zitat Wang T-C, Liu M-Y, Zhu J-Y, Tao A, Kautz J, Catanzaro B. High-resolution image synthesis and semantic manipulation with conditional gans. Proc IEEE Conf Comput Vis Pattern Recognit. 2018. p. 8798–807. Wang T-C, Liu M-Y, Zhu J-Y, Tao A, Kautz J, Catanzaro B. High-resolution image synthesis and semantic manipulation with conditional gans. Proc IEEE Conf Comput Vis Pattern Recognit. 2018. p. 8798–807.
27.
Zurück zum Zitat Sanaat A, Shiri I, Arabi H, Mainta I, Nkoulou R, Zaidi H. Deep learning-assisted ultra-fast/low-dose whole-body PET/CT imaging. Eur J Nucl Med Mol Imaging. 2021;48:2405–15.CrossRef Sanaat A, Shiri I, Arabi H, Mainta I, Nkoulou R, Zaidi H. Deep learning-assisted ultra-fast/low-dose whole-body PET/CT imaging. Eur J Nucl Med Mol Imaging. 2021;48:2405–15.CrossRef
28.
Zurück zum Zitat Chaudhari AS, Mittra E, Davidzon GA, Gulaka P, Gandhi H, Brown A, et al. Low-count whole-body PET with deep learning in a multicenter and externally validated study. NPJ Digit Med England. 2021;4:127.CrossRef Chaudhari AS, Mittra E, Davidzon GA, Gulaka P, Gandhi H, Brown A, et al. Low-count whole-body PET with deep learning in a multicenter and externally validated study. NPJ Digit Med England. 2021;4:127.CrossRef
29.
Zurück zum Zitat Lu W, Onofrey JA, Lu Y, Shi L, Ma T, Liu Y, et al. An investigation of quantitative accuracy for deep learning based denoising in oncological PET. Phys Med Biol England. 2019;64:165019.CrossRef Lu W, Onofrey JA, Lu Y, Shi L, Ma T, Liu Y, et al. An investigation of quantitative accuracy for deep learning based denoising in oncological PET. Phys Med Biol England. 2019;64:165019.CrossRef
30.
Zurück zum Zitat Gong K, Guan J, Kim K, Zhang X, Yang J, Seo Y, et al. Iterative PET image reconstruction using convolutional neural network representation. IEEE Trans Med Imaging. 2019;38:675–85.CrossRef Gong K, Guan J, Kim K, Zhang X, Yang J, Seo Y, et al. Iterative PET image reconstruction using convolutional neural network representation. IEEE Trans Med Imaging. 2019;38:675–85.CrossRef
31.
Zurück zum Zitat Kang J, Gao Y, Shi F, Lalush DS, Lin W, Shen D. Prediction of standard-dose brain PET image by using MRI and low-dose brain [18F]FDG PET images. Med Phys. 2015;42:5301–9.CrossRef Kang J, Gao Y, Shi F, Lalush DS, Lin W, Shen D. Prediction of standard-dose brain PET image by using MRI and low-dose brain [18F]FDG PET images. Med Phys. 2015;42:5301–9.CrossRef
32.
Zurück zum Zitat Wang Y, Yu B, Wang L, Zu C, Lalush DS, Lin W, et al. 3D conditional generative adversarial networks for high-quality PET image estimation at low dose. Neuroimage. 2018;174:550–62.CrossRef Wang Y, Yu B, Wang L, Zu C, Lalush DS, Lin W, et al. 3D conditional generative adversarial networks for high-quality PET image estimation at low dose. Neuroimage. 2018;174:550–62.CrossRef
33.
Zurück zum Zitat Sanaat A, Arabi H, Mainta I, Garibotto V, Zaidi H. Projection space implementation of deep learning-guided low-dose brain PET imaging improves performance over implementation in image space. J Nucl Med Off Publ Soc Nucl Med. 2020;61:1388–96. Sanaat A, Arabi H, Mainta I, Garibotto V, Zaidi H. Projection space implementation of deep learning-guided low-dose brain PET imaging improves performance over implementation in image space. J Nucl Med Off Publ Soc Nucl Med. 2020;61:1388–96.
34.
Zurück zum Zitat Ouyang J, Chen KT, Gong E, Pauly J, Zaharchuk G. Ultra-low-dose PET reconstruction using generative adversarial network with feature matching and task-specific perceptual loss. Med Phys. 2019;46:3555–64.CrossRef Ouyang J, Chen KT, Gong E, Pauly J, Zaharchuk G. Ultra-low-dose PET reconstruction using generative adversarial network with feature matching and task-specific perceptual loss. Med Phys. 2019;46:3555–64.CrossRef
35.
Zurück zum Zitat Wang Y-RJ, Baratto L, Hawk KE, Theruvath AJ, Pribnow A, Thakor AS, et al. Artificial intelligence enables whole-body positron emission tomography scans with minimal radiation exposure. Eur J Nucl Med Mol Imaging. 2021;48:2771–81. Wang Y-RJ, Baratto L, Hawk KE, Theruvath AJ, Pribnow A, Thakor AS, et al. Artificial intelligence enables whole-body positron emission tomography scans with minimal radiation exposure. Eur J Nucl Med Mol Imaging. 2021;48:2771–81.
36.
Zurück zum Zitat Nguyen NC, Kaushik A, Wolverson MK, Osman MM. Is there a common SUV threshold in oncological FDG PET/CT, at least for some common indications? A retrospective study. Acta Oncol Taylor & Francis. 2011;50:670–7.CrossRef Nguyen NC, Kaushik A, Wolverson MK, Osman MM. Is there a common SUV threshold in oncological FDG PET/CT, at least for some common indications? A retrospective study. Acta Oncol Taylor & Francis. 2011;50:670–7.CrossRef
37.
Zurück zum Zitat Horé A, Ziou D. Image quality metrics: PSNR vs. SSIM. 2010 20th Int Conf Pattern Recognit. 2010. 2366–9. Horé A, Ziou D. Image quality metrics: PSNR vs. SSIM. 2010 20th Int Conf Pattern Recognit. 2010. 2366–9.
40.
Zurück zum Zitat Ma L, Jia X, Sun Q, Schiele B, Tuytelaars T, Van Gool L. Pose guided person image generation. Adv Neural Inf Process Syst. 2017;30. Ma L, Jia X, Sun Q, Schiele B, Tuytelaars T, Van Gool L. Pose guided person image generation. Adv Neural Inf Process Syst. 2017;30.
41.
Zurück zum Zitat Sibille L, Seifert R, Avramovic N, Vehren T, Spottiswoode B, Zuehlsdorff S, et al. 18F-FDG PET/CT uptake classification in lymphoma and lung cancer by using deep convolutional neural networks. Radiology Radiological Society of North America. 2020;294:445–52. Sibille L, Seifert R, Avramovic N, Vehren T, Spottiswoode B, Zuehlsdorff S, et al. 18F-FDG PET/CT uptake classification in lymphoma and lung cancer by using deep convolutional neural networks. Radiology Radiological Society of North America. 2020;294:445–52.
42.
Zurück zum Zitat Weber M, Kersting D, Umutlu L, Schäfers M, Rischpler C, Fendler WP, et al. Just another “Clever Hans”? Neural networks and FDG PET-CT to predict the outcome of patients with breast cancer. Eur J Nucl Med Mol Imaging. 2021;48:3141–50.CrossRef Weber M, Kersting D, Umutlu L, Schäfers M, Rischpler C, Fendler WP, et al. Just another “Clever Hans”? Neural networks and FDG PET-CT to predict the outcome of patients with breast cancer. Eur J Nucl Med Mol Imaging. 2021;48:3141–50.CrossRef
43.
Zurück zum Zitat O JH, Lodge MA, Wahl RL. Practical PERCIST: a simplified guide to PET response criteria in solid tumors 1.0. Radiology. Radiological Society of North America; 2016;280:576–84. O JH, Lodge MA, Wahl RL. Practical PERCIST: a simplified guide to PET response criteria in solid tumors 1.0. Radiology. Radiological Society of North America; 2016;280:576–84.
45.
Zurück zum Zitat Virtanen P, Gommers R, Oliphant TE, Haberland M, Reddy T, Cournapeau D, et al. SciPy 1.0: fundamental algorithms for scientific computing in Python. Nat Methods. 2020;17:261–72.CrossRef Virtanen P, Gommers R, Oliphant TE, Haberland M, Reddy T, Cournapeau D, et al. SciPy 1.0: fundamental algorithms for scientific computing in Python. Nat Methods. 2020;17:261–72.CrossRef
46.
Zurück zum Zitat Lei Y, Dong X, Wang T, Higgins K, Liu T, Curran WJ, et al. Whole-body PET estimation from low count statistics using cycle-consistent generative adversarial networks. Phys Med Biol. 2019;64:215017.CrossRef Lei Y, Dong X, Wang T, Higgins K, Liu T, Curran WJ, et al. Whole-body PET estimation from low count statistics using cycle-consistent generative adversarial networks. Phys Med Biol. 2019;64:215017.CrossRef
47.
Zurück zum Zitat Wang Y, Zhou L, Yu B, Wang L, Zu C, Lalush DS, et al. 3D Auto-context-based locality adaptive multi-modality GANs for PET synthesis. IEEE Trans Med Imaging. 2019;38:1328–39.CrossRef Wang Y, Zhou L, Yu B, Wang L, Zu C, Lalush DS, et al. 3D Auto-context-based locality adaptive multi-modality GANs for PET synthesis. IEEE Trans Med Imaging. 2019;38:1328–39.CrossRef
48.
Zurück zum Zitat Wang X, Zhou L, Wang Y, Jiang H, Ye H. Improved low-dose positron emission tomography image reconstruction using deep learned prior. Phys Med Biol IOP Publishing. 2021;66:115001.CrossRef Wang X, Zhou L, Wang Y, Jiang H, Ye H. Improved low-dose positron emission tomography image reconstruction using deep learned prior. Phys Med Biol IOP Publishing. 2021;66:115001.CrossRef
49.
Zurück zum Zitat Kaplan S, Zhu Y-M. Full-dose PET image estimation from low-dose PET image using deep learning: a pilot study. J Digit Imaging. 2019;32:773–8.CrossRef Kaplan S, Zhu Y-M. Full-dose PET image estimation from low-dose PET image using deep learning: a pilot study. J Digit Imaging. 2019;32:773–8.CrossRef
50.
Zurück zum Zitat de Langen AJ, Vincent A, Velasquez LM, van Tinteren H, Boellaard R, Shankar LK, et al. Repeatability of 18F-FDG uptake measurements in tumors: a metaanalysis. J Nucl Med Off Publ Soc Nucl Med United States. 2012;53:701–8. de Langen AJ, Vincent A, Velasquez LM, van Tinteren H, Boellaard R, Shankar LK, et al. Repeatability of 18F-FDG uptake measurements in tumors: a metaanalysis. J Nucl Med Off Publ Soc Nucl Med United States. 2012;53:701–8.
51.
Zurück zum Zitat Schaefferkoetter J, Nai Y-H, Reilhac A, Townsend DW, Eriksson L, Conti M. Low dose positron emission tomography emulation from decimated high statistics: a clinical validation study. Med Phys John & Sons Wiley Ltd. 2019;46:2638–45. Schaefferkoetter J, Nai Y-H, Reilhac A, Townsend DW, Eriksson L, Conti M. Low dose positron emission tomography emulation from decimated high statistics: a clinical validation study. Med Phys John & Sons Wiley Ltd. 2019;46:2638–45.
52.
Zurück zum Zitat Zhu J-Y, Park T, Isola P, Efros AA. Unpaired image-to-image translation using cycle-consistent adversarial networks. IEEE Int Conf Comput Vis ICCV. 2017. p. 2242–51. Zhu J-Y, Park T, Isola P, Efros AA. Unpaired image-to-image translation using cycle-consistent adversarial networks. IEEE Int Conf Comput Vis ICCV. 2017. p. 2242–51.
Metadaten
Titel
Artificial intelligence guided enhancement of digital PET: scans as fast as CT?
verfasst von
René Hosch
Manuel Weber
Miriam Sraieb
Nils Flaschel
Johannes Haubold
Moon-Sung Kim
Lale Umutlu
Jens Kleesiek
Ken Herrmann
Felix Nensa
Christoph Rischpler
Sven Koitka
Robert Seifert
David Kersting
Publikationsdatum
29.07.2022
Verlag
Springer Berlin Heidelberg
Erschienen in
European Journal of Nuclear Medicine and Molecular Imaging / Ausgabe 13/2022
Print ISSN: 1619-7070
Elektronische ISSN: 1619-7089
DOI
https://doi.org/10.1007/s00259-022-05901-x

Weitere Artikel der Ausgabe 13/2022

European Journal of Nuclear Medicine and Molecular Imaging 13/2022 Zur Ausgabe