01.01.2014 | Original Article

Assessment of damping capacity as an index of root proximity in self-drilling orthodontic mini-implants

verfasst von: Mitsuru Motoyoshi, Yasuki Uchida, Miho Matsuoka, Mizuki Inaba, Hiroaki Iwai, Yoko Karasawa, Noriyoshi Shimizu

Erschienen in: Clinical Oral Investigations | Ausgabe 1/2014

Einloggen, um Zugang zu erhalten

Abstract

Objectives

This study aims to investigate orthodontic mini-implant root proximity, placement torque, and damping capacity and to determine whether placement torque and damping capacity (Periotest value (PTV)) are useful indices for the estimation of mini-implant root proximity.

Materials and methods

The root proximity of 143 orthodontic mini-implants (1.6 mm diameter, 8 mm screw thread length) was evaluated in 79 patients (24 males, 55 females; mean age, 22.5 ± 8 years) using cone-beam computed tomography. The placement torque and PTV of each implant were determined using a torque tester and the Periotest, respectively. Variability in these values according to root proximity was evaluated.

Results

PTVs of mini-implants with multiple (two or more) points of contact between the root and implant were significantly larger than those of mini-implants with no root contact in the self-drilling group. Placement torque did not differ significantly according to root proximity. In the self-drilling group, the odds ratio for root contact was 20.82 (P = 0.000) for a PTV >6.

Conclusions

Placement torque could not be used to estimate root proximity. The PTV was significantly affected by root proximity in the self-drilling group.

Clinical relevance

A threshold of PTV >6 could be applied clinically for the estimation of self-drilling mini-implant root proximity.
Literatur
1.
Zurück zum Zitat Kanomi R (1997) Mini-implant for orthodontic anchorage. J Clin Orthod 31:763–767PubMed Kanomi R (1997) Mini-implant for orthodontic anchorage. J Clin Orthod 31:763–767PubMed
2.
Zurück zum Zitat Miyawaki S, Koyama I, Inoue M, Mishima K, Sugawara T, Yamamoto T (2003) Factors associated with the stability of titanium screws placed in the posterior region for orthodontic anchorage. Am J Orthod Dentofac Orthop 124:373–378CrossRef Miyawaki S, Koyama I, Inoue M, Mishima K, Sugawara T, Yamamoto T (2003) Factors associated with the stability of titanium screws placed in the posterior region for orthodontic anchorage. Am J Orthod Dentofac Orthop 124:373–378CrossRef
3.
Zurück zum Zitat Sugawara J, Daimaruya T, Umemori M, Nagasaka H, Takahashi I, Kawamura H, Mitani H (2004) Distal movement of mandibular molars in adult patients with the skeletal anchorage system. Am J Orthod Dentofac Orthop 125:130–138CrossRef Sugawara J, Daimaruya T, Umemori M, Nagasaka H, Takahashi I, Kawamura H, Mitani H (2004) Distal movement of mandibular molars in adult patients with the skeletal anchorage system. Am J Orthod Dentofac Orthop 125:130–138CrossRef
4.
Zurück zum Zitat Herman RJ, Currier GF, Miyake A (2006) Mini-implant anchorage for maxillary canine retraction: a pilot study. Am J Orthod Dentofac Orthop 130:228–235CrossRef Herman RJ, Currier GF, Miyake A (2006) Mini-implant anchorage for maxillary canine retraction: a pilot study. Am J Orthod Dentofac Orthop 130:228–235CrossRef
5.
Zurück zum Zitat Kim TW, Kim H, Lee SJ (2006) Correction of deep overbite and gummy smile by using a mini-implant with a segmented wire in a growing class II division 2 patient. Am J Orthod Dentofac Orthop 130:676–685CrossRef Kim TW, Kim H, Lee SJ (2006) Correction of deep overbite and gummy smile by using a mini-implant with a segmented wire in a growing class II division 2 patient. Am J Orthod Dentofac Orthop 130:676–685CrossRef
6.
Zurück zum Zitat Liou EJ, Pai BC, Lin JC (2004) Do miniscrews remain stationary under orthodontic forces? Am J Orthod Dentofac Orthop 126:42–47CrossRef Liou EJ, Pai BC, Lin JC (2004) Do miniscrews remain stationary under orthodontic forces? Am J Orthod Dentofac Orthop 126:42–47CrossRef
7.
Zurück zum Zitat Kuroda S, Yamada K, Deguchi T, Hashimoto T, Kyung HM, Takano-Yamamoto T (2007) Root proximity is a major factor for screw failure in orthodontic anchorage. Am J Orthod Dentofac Orthop 131:S68–S73CrossRef Kuroda S, Yamada K, Deguchi T, Hashimoto T, Kyung HM, Takano-Yamamoto T (2007) Root proximity is a major factor for screw failure in orthodontic anchorage. Am J Orthod Dentofac Orthop 131:S68–S73CrossRef
8.
Zurück zum Zitat Min KI, Kim SC, Kang KH, Cho JH, Lee EH, Chang NY, Chae JM (2012) Root proximity and cortical bone thickness effects on the success rate of orthodontic micro-implants using cone beam computed tomography. Angle Orthod 82:1014–1021PubMedCrossRef Min KI, Kim SC, Kang KH, Cho JH, Lee EH, Chang NY, Chae JM (2012) Root proximity and cortical bone thickness effects on the success rate of orthodontic micro-implants using cone beam computed tomography. Angle Orthod 82:1014–1021PubMedCrossRef
9.
Zurück zum Zitat Kim H, Kim TW (2011) Histologic evaluation of root-surface healing after root contact or approximation during placement of mini-implants. Am J Orthod Dentofac Orthop 139:752–760CrossRef Kim H, Kim TW (2011) Histologic evaluation of root-surface healing after root contact or approximation during placement of mini-implants. Am J Orthod Dentofac Orthop 139:752–760CrossRef
10.
Zurück zum Zitat Estelita CB, Janson G, Chiqueto K, Freitas MR, Henriques JFC, Pinzan A (2006) A three-dimensional radiographic surgical guide for mini-implant placement. J Clin Orthod 40:548–554 Estelita CB, Janson G, Chiqueto K, Freitas MR, Henriques JFC, Pinzan A (2006) A three-dimensional radiographic surgical guide for mini-implant placement. J Clin Orthod 40:548–554
11.
Zurück zum Zitat Choi HJ, Kim TW, Kim HW (2007) A precise wire guide for positioning interradicular miniscrews. J Clin Orthod 41:258–261PubMed Choi HJ, Kim TW, Kim HW (2007) A precise wire guide for positioning interradicular miniscrews. J Clin Orthod 41:258–261PubMed
12.
Zurück zum Zitat Kim SH, Choi YS, Hwang EH, Chung KR, Kook YA, Nelson G (2007) Surgical positioning of orthodontic mini-implants with guides fabricated on models replicated with cone-beam computed tomography. Am J Orthod Dentofac Orthop 131:S82–S89CrossRef Kim SH, Choi YS, Hwang EH, Chung KR, Kook YA, Nelson G (2007) Surgical positioning of orthodontic mini-implants with guides fabricated on models replicated with cone-beam computed tomography. Am J Orthod Dentofac Orthop 131:S82–S89CrossRef
13.
Zurück zum Zitat Kim SH, Kang SM, Choi YS, Kook YA, Chung KR, John C, Huang JC (2010) Cone-beam computed tomography evaluation of mini-implants after placement: is root proximity a major risk factor for failure? Am J Orthod Dentofac Orthop 138:264–276CrossRef Kim SH, Kang SM, Choi YS, Kook YA, Chung KR, John C, Huang JC (2010) Cone-beam computed tomography evaluation of mini-implants after placement: is root proximity a major risk factor for failure? Am J Orthod Dentofac Orthop 138:264–276CrossRef
14.
Zurück zum Zitat Oh JS, Kim SG, Lim SC, Ong JL (2009) A comparative study of two noninvasive techniques to evaluate implant stability: Periotest and Osstell Mentor. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 107:513–518PubMedCrossRef Oh JS, Kim SG, Lim SC, Ong JL (2009) A comparative study of two noninvasive techniques to evaluate implant stability: Periotest and Osstell Mentor. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 107:513–518PubMedCrossRef
15.
Zurück zum Zitat Nienkemper M, Wilmes B, Panayotidis A, Pauls A, Golubovic V, Schwarz F, Drescher D (2012) Measurement of mini-implant stability using resonance frequency analysis. Angle Orthod. doi:10.2319/043012-354.1 Nienkemper M, Wilmes B, Panayotidis A, Pauls A, Golubovic V, Schwarz F, Drescher D (2012) Measurement of mini-implant stability using resonance frequency analysis. Angle Orthod. doi:10.​2319/​043012-354.​1
16.
Zurück zum Zitat Tachibana R, Motoyoshi M, Shinohara A, Shigeeda T, Shimizu N (2012) Safe placement techniques for self-drilling orthodontic mini-implants. Int J Oral Maxillofac Surg 41:1439–1444PubMedCrossRef Tachibana R, Motoyoshi M, Shinohara A, Shigeeda T, Shimizu N (2012) Safe placement techniques for self-drilling orthodontic mini-implants. Int J Oral Maxillofac Surg 41:1439–1444PubMedCrossRef
17.
Zurück zum Zitat Motoyoshi M, Hirabayashi M, Uemura M, Shimizu N (2006) Recommended placement torque when tightening an orthodontic miniimplant. Clin Oral Implants Res 17:109–114PubMedCrossRef Motoyoshi M, Hirabayashi M, Uemura M, Shimizu N (2006) Recommended placement torque when tightening an orthodontic miniimplant. Clin Oral Implants Res 17:109–114PubMedCrossRef
18.
Zurück zum Zitat Motoyoshi M, Yoshida T, Ono A, Shimizu N (2007) Effect of cortical bone thickness and implant placement torque on stability of orthodontic mini-implants. Int J Oral Maxillofac Implants 22:779–784PubMed Motoyoshi M, Yoshida T, Ono A, Shimizu N (2007) Effect of cortical bone thickness and implant placement torque on stability of orthodontic mini-implants. Int J Oral Maxillofac Implants 22:779–784PubMed
19.
Zurück zum Zitat Zix J, Hug S, Kessler-Liechti G, Mericske-Stern R (2008) Measurement of dental implant stability by resonance frequency analysis and damping capacity assessment: comparison of both techniques in a clinical trial. Int J Oral Maxillofac Implants 3:525–530 Zix J, Hug S, Kessler-Liechti G, Mericske-Stern R (2008) Measurement of dental implant stability by resonance frequency analysis and damping capacity assessment: comparison of both techniques in a clinical trial. Int J Oral Maxillofac Implants 3:525–530
20.
Zurück zum Zitat Lachmann S, Jäger B, Axmann D, Gomez-Roman G, Groten M, Weber H (2006) Resonance frequency analysis and damping capacity assessment. Part I: an in vitro study on measurement reliability and a method of comparison in the determination of primary dental implant stability. Clin Oral Implants Res 17:75–79PubMedCrossRef Lachmann S, Jäger B, Axmann D, Gomez-Roman G, Groten M, Weber H (2006) Resonance frequency analysis and damping capacity assessment. Part I: an in vitro study on measurement reliability and a method of comparison in the determination of primary dental implant stability. Clin Oral Implants Res 17:75–79PubMedCrossRef
21.
Zurück zum Zitat Lachmann S, Laval JY, Jäger B, Axmann D, Gomez-Roman G, Groten M, Weber H (2006) Resonance frequency analysis and damping capacity assessment. Part 2: peri-implant bone loss follow-up. An in vitro study with the Periotest and Osstell instruments. Clin Oral Implants Res 17:80–84PubMedCrossRef Lachmann S, Laval JY, Jäger B, Axmann D, Gomez-Roman G, Groten M, Weber H (2006) Resonance frequency analysis and damping capacity assessment. Part 2: peri-implant bone loss follow-up. An in vitro study with the Periotest and Osstell instruments. Clin Oral Implants Res 17:80–84PubMedCrossRef
22.
Zurück zum Zitat Inaba M (2009) Evaluation of primary stability of inclined orthodontic mini-implants. J Oral Sci 51:347–353PubMedCrossRef Inaba M (2009) Evaluation of primary stability of inclined orthodontic mini-implants. J Oral Sci 51:347–353PubMedCrossRef
23.
Zurück zum Zitat Omasa S, Motoyoshi M, Arai Y, Ejima K, Shimizu N (2012) Low-level laser therapy enhances the stability of orthodontic mini-implants via bone formation related to BMP-2 expression in a rat model. Photomed Laser Surg 30:255–261PubMedCrossRef Omasa S, Motoyoshi M, Arai Y, Ejima K, Shimizu N (2012) Low-level laser therapy enhances the stability of orthodontic mini-implants via bone formation related to BMP-2 expression in a rat model. Photomed Laser Surg 30:255–261PubMedCrossRef
24.
Zurück zum Zitat Uemura M, Motoyoshi M, Yano S, Sakaguchi M, Igarashi Y, Shimizu N (2012) Orthodontic mini-implant stability and the ratio of pilot hole–implant diameter. Euro J Orthod 34:52–55CrossRef Uemura M, Motoyoshi M, Yano S, Sakaguchi M, Igarashi Y, Shimizu N (2012) Orthodontic mini-implant stability and the ratio of pilot hole–implant diameter. Euro J Orthod 34:52–55CrossRef
Metadaten
Titel
Assessment of damping capacity as an index of root proximity in self-drilling orthodontic mini-implants
verfasst von
Mitsuru Motoyoshi
Yasuki Uchida
Miho Matsuoka
Mizuki Inaba
Hiroaki Iwai
Yoko Karasawa
Noriyoshi Shimizu
Publikationsdatum
01.01.2014
Verlag
Springer Berlin Heidelberg
Erschienen in
Clinical Oral Investigations / Ausgabe 1/2014
Print ISSN: 1432-6981
Elektronische ISSN: 1436-3771
DOI
https://doi.org/10.1007/s00784-013-0965-8

Weitere Artikel der Ausgabe 1/2014

Clinical Oral Investigations 1/2014 Zur Ausgabe

Endlich: Zi zeigt, mit welchen PVS Praxen zufrieden sind

IT für Ärzte Nachrichten

Darauf haben viele Praxen gewartet: Das Zi hat eine Liste von Praxisverwaltungssystemen veröffentlicht, die von Nutzern positiv bewertet werden. Eine gute Grundlage für wechselwillige Ärzte und Psychotherapeuten.

Parodontalbehandlung verbessert Prognose bei Katheterablation

19.04.2024 Vorhofflimmern Nachrichten

Werden Personen mit Vorhofflimmern in der Blanking-Periode nach einer Katheterablation gegen eine bestehende Parodontitis behandelt, verbessert dies die Erfolgsaussichten. Dafür sprechen die Resultate einer prospektiven Untersuchung.

Invasive Zahnbehandlung: Wann eine Antibiotikaprophylaxe vor infektiöser Endokarditis schützt

11.04.2024 Endokarditis Nachrichten

Bei welchen Personen eine Antibiotikaprophylaxe zur Prävention einer infektiösen Endokarditis nach invasiven zahnärztlichen Eingriffen sinnvoll ist, wird diskutiert. Neue Daten stehen im Einklang mit den europäischen Leitlinienempfehlungen.

Zell-Organisatoren unter Druck: Mechanismen des embryonalen Zahnwachstums aufgedeckt

08.04.2024 Zahnmedizin Nachrichten

Der Aufbau von Geweben und Organen während der Embryonalentwicklung wird von den Zellen bemerkenswert choreografiert. Für diesen Prozess braucht es spezielle sogenannte „Organisatoren“. In einer aktuellen Veröffentlichung im Fachjournal Nature Cell Biology berichten Forschende durch welchen Vorgang diese Organisatoren im Gewebe entstehen und wie sie dann die Bildung von Zähnen orchestrieren.

Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Zahnmedizin und bleiben Sie gut informiert – ganz bequem per eMail.