Biomechanical effects of morphological variations of the cortical wall at the bone-cement interface

verfasst von: Chun-Lin Zhang, Guo-Qi Shen, Kun-Peng Zhu, Dong-xu Liu

Erschienen in: Journal of Orthopaedic Surgery and Research | Ausgabe 1/2016

Abstract

Background

The integrity of bone-cement interface is very important for the stabilization and long-term sustain of cemented prosthesis. Variations in the bone-cement interface morphology may affect the mechanical response of the shape-closed interlock.

Methods

Self-developed new reamer was used to process fresh pig reamed femoral canal, creating cortical grooves in the canal wall of experimental group. The biomechanical effects of varying the morphology with grooves of the bone-cement interface were investigated using finite element analysis (FEA) and validated using companion experimental data. Micro-CT scans were used to document interlock morphology.

Results

The contact area of the bone-cement interface was greater (P < 0.05) for the experimental group (5470 ± 265 mm2) when compared to the specimens of control group (5289 ± 299 mm2). The mechanical responses to tensile loading and anti-torsion showed that the specimens with grooves were stronger (P < 0.05) at the bone-cement interface than the specimens without grooves. There were positively significant correlation between the contact area and the tensile force (r 2 = 0.85) and the maximal torsion (r 2 = 0.77) at the bone-cement interface. The volume of cement of the experimental group (7688 ± 278 mm3) was greater (P < 0.05) than of the control group (5764 ± 186 mm3). There were positively significant correlations between the volume of cement and the tensile force (r 2 = 0.90) and the maximal torsion (r 2 = 0.97) at the bone-cement interface. The FEA results compared favorably to the tensile and torsion relationships determined experimentally. More cracks occurred in the cement than in the bone.

Conclusions

Converting the standard reaming process from a smooth bore cortical tube to the one with grooves permits the cement to interlock with the reamed bony wall. This would increase the strength of the bone-cement interface.
Hinweise
Chun-Lin Zhang and Guo-Qi Shen are co-first authors.

Background

Cemented prostheses are widely used because they generally remain firmly fixed during the early postoperative stage [15]. However, aseptic loosening is an important long-term complication of prosthesis and is in part responsible for the increasing number of patients requiring revision surgery. Clinical evidence suggests that failures and the incidence of loosening along the bone-cement interface are prevalent [6, 7]. Initial fixation strength of the prosthetic components is the main factor that influences the long-term service life of the implant. The integrity of bone-cement interface is very important for the stabilization and long-term sustain of artificial prosthesis [4].
Loosening is considered a complicated function of both mechanical and biological factors [8]. Although the incidence and progression of loosening in cemented joint replacements is poorly understood, it is generally accepted that trabecular bone resorption at the bone-cement interface contribute to failure of the bone-cement interface [9]. The fixation strength of cemented artificial joint is dependent on good interlocking of the bone-cement interface. Bone-cement interface interlocking strength is positively correlated with contact area and interlocking depth of interface. By increasing the bone-cement interface contact area and interlocking depth, artificial joint replacement surgery can be ensured to get better initial fixation strength, resulting in extended life of artificial joint and reduced complications [10, 11]. Variations in the bone-cement interface morphology may affect the mechanical response of the shape-closed interlock. Few studies to the morphology of bone-cement interface have been reported [1215]. In the present study, we postulated that using self-developed new reamer to create some superficial glossy grooves in the medullary canal wall permits increasing the interlock area of the bone-cement interface. This morphology change of the bone-cement interface may improve the mechanical strength and potential longevity to the bone-cement-implant component. The primary objective of this study was to quantify the contact area and strength of the bone-cement interface and to analyze morphological changes of the components associated with the morphological variations of the cortical wall.

Methods

This study has been granted an exemption from requiring ethics approval by the ethic committee of the Tenth People’s Hospital Affiliated to Tongji University.

Specimen and model preparation

After removing the surrounding soft tissues, the muscles and the femoral condyles of the fresh pig’s femurs which were purchased from the supermarket, 40 samples were cut. The length of the model femur bone was 60 mm each. The 64-slice spiral computed tomography (CT) was used to scan the femur samples. Forty samples were randomly assigned to the experimental group and the control group, with each group 20 samples. The bone marrow and most of cancellous bone of all samples were removed. In control group, the traditional reamer (Link Company, Germany) was used to ream the samples. In experimental group, the traditional reamer was used to ream the samples firstly, then the self-developed reamer which had two cutting teeth (Fig. 1) was used to create two cortical grooves in the canal wall. The two grooves respectively located at 20 and 40 mm of the proximal femoral shaft, and both were 2 mm wide and 2 mm deep. This treatment of the inner wall of the medullary cavity was not performed in the control group which was reamed only by the traditional reamer. After reaming, pulsed lavage was used to clean bone particles inside the medullary canal.
The CoCrMo prosthesis was inserted to the femur after cement importing (Fig. 2). The stem of CoCrMo prosthesis was cylindrical with 12 mm diameter, 50 mm length, and with a distal smooth cambered surface. A 200-N constant pressure was applied to the fixed rod of the prosthesis until the cement fully solidified in 10 min. The thickness of cement was 3 mm due to the diameter of the femoral canal and the properties of the prosthesis. All operations were done at temperature of 21 ± 1 °C and humidity of 55 ± 5 %.

Detection of micro-computed tomography

All specimens were scanned with a micro-computed tomography (micro-CT) scanner (phoenix-v|tome|x m, General Electric Company, USA) to document cement and bone morphology before biomechanical testing. Models were vertical inverted to place on fixed jaw of instrument. Each scanning was at clockwise rotation with 0.19° and voltage of 240 kV, current of 200 μA, and resolution of 45 μm. The average scanned time was 90 ± 5 min. We used the DATOSX software (General Electric Company, USA) after scanning for image acquisition and reconstructed visualization. Three-dimensional morphology measurements were made using the VGstudio software (Volume Graphics, Germany), which was based on a single threshold of CT gray value of the bone, cement, and artificial prosthesis, including the total volume (TV), the volume of cement (CV), the bone volume (BV), the pore volume (PV), and the contact area of the bone-cement interface (CA). The porosity was the ratio of the pore volume to the total volume of the specimen (PV/TV). The bone volume fraction was the ratio of the bone volume to the total volume (BV/TV). The metal prosthesis may have great image effects on cement and bone during the image reconstruction due to the prosthesis was a cobalt-chromium-molybdenum material. Therefore, we must extract the images of the metal prosthesis from models, and then the bone and cement for analysis and calculation, to avoid the effects caused by metal prosthesis.

Biomechanical testing

After micro-CT detection, all 40 specimens were randomly divided into experimental group and control group, including 10 specimens of tensile experimental group, 10 specimens of tensile control group, 10 specimens of torsion experimental group, and 10 specimens of torsion control group. The biomechanical tests of all specimens were performed by the USA pull torsional biomechanical testing machine (dynamic displacement ±50 mm; dynamic axial load ±14.2 kN; static torsion angle ±40°; static torque 200 Nm).
The test specimens were fixed vertically on the biomechanical testing machine. Before testing, the samples were adapted under the loading from −100 to 100 N and the frequency of 2 Hz. The initial load of the tensile testing was performed vertically upward to 100 N and 2 mm/min of displacement to implant fixation rod. In the torsion group, a similarly initial load that perpendicular to the longitudinal axis of the model was performed, 5°/min of rotational rate. When the specimens showed cortical bone fractures, separation of the bone-cement interface or cement-prosthesis interface, and fracture of prosthesis, which were considered model failure. The maximal load was the loading of model failure.

Finite element modeling of the bone-cement-prosthesis composite

FEA models were created using micro-CT scans of specimens. The FEA meshes of the two models with and without grooves were created by meshing using ANSYS14.0 program. The initial material properties of the models were considered to be linear elastic and isotropic. Young’s modulus [16] and Poisson’s ratio of the femur, cement, and prosthesis were set to 16.7, 3.0, 200 Gpa, and 0.3, respectively. The bone properties were based upon micro-CT gray scale values, which were converted to equivalent HA-densities using a calibration phantom. The assumption of a linear relationship between the HA-density and the Young’s modulus resulted in Young’s moduli ranging from 0.1 to 20,000 MPa for the bone (n = 0.3). In the tension test, the load was added to the top of the prosthesis pointing downward vertically to simulate the biomechanical experimental load by an axial strength of 2.4 kN. Similarly, in the torsion test, a torque of 40 Nm was added to the top of the prosthesis vertically.

Statistics

The data were analyzed descriptively using the arithmetic mean, standard deviation (SD), minimum value, and maximum value. The skewness and kurtosis coefficient were used to analyze the distribution of data. Paired t tests were used to determine if the data was normally distributed, otherwise, the rank test was used. P < 0.05 was considered statistically significant. Linear regression analysis was used to determine correlations between average contact area and tensile strength, maximal torsion, and porosity. The results of biomechanical experimental tests and FEA were compared. All data analyses were performed using SPSS 18.0 (SPSS, Chicago, Illinois).

Results

The morphology of bone-cement interface (micro-CT)

Compared with the control group, bone cement in the specimens of experimental group had better penetration and better interlock between cement and bone, due to annular grooves of the inner wall of the bone medullary canal. The contact area of the bone-cement interface was greater (P < 0.05) for the experimental group (5470 ± 265 mm2) when compared to the specimens of control group (5289 ± 299 mm2). The volume of cement for the experimental group (7688 ± 278 mm3) was greater (P < 0.05) when compared to the specimens of control group (5764 ± 186 mm3). However, the porosity for the experimental group (1.50 ± 0.382 %) was similar (P > 0.05) to the control group (1.59 ± 0.496 %). There was no significant difference in the bone volume fraction (BV/TV) (P > 0.05) between the experimental group (0.493 ± 0.019) and control group (0.495 ± 0.031) (Fig. 3).

Mechanical properties

Under the loading of 2400 N, there was no significant difference of the average peak stress and the total deformation, respectively, for the components with grooves and the components without grooves. It showed that the creation of grooves in the intramedullary canal did not affect the integrity of the femur and jeopardize the strength of the femur either. The mechanical responses to tensile loading showed that the specimens of experimental group (7337 ± 1825 N) had stronger force (P < 0.05) at the bone-cement interface than the control group (5564 ± 1359 N). The anti-torsion capability was greater (P < 0.05) for the experimental group (65.70 ± 4.83 Nm) when compared to the control group (60.60 ± 4.43 Nm) (Fig. 3). There were positively significant correlations between the contact area and the tensile force (r 2 = 0.85) and the maximal torsion (r 2 = 0.77) at the bone-cement interface. Furthermore, the cement volume was strongly and significantly correlated with tensile force and maximal torsion, respectively (r 2 = 0.90, P < 0.05 and r 2 = 0.97, P < 0.05, Fig. 4). The relationships between the porosity and the tensile force and the maximal torsion were observed (r 2 = 0.57, P < 0.05 and r 2 = 0.43, P < 0.05, Fig. 4). In the tensile testing, the failure of all specimens (20/20) including the experimental group and the control group occurred at the bone-cement interface. There was a macroscopic displacement between the bone and cement. In the rotary testing, the failure of 17 specimens (17/20) occurred at the bone-cement interface (Fig. 5). More cracks occurred in the cement than in the bone.
In FEA, under the tension and torsion loads, the peak stresses of the components with grooves (23.48 and 161.5 MPa, respectively) were both lower than those of the components without grooves (27.62 and 182.3 MPa, respectively) (Fig. 6). The peak stresses of the femur of the experimental group were also lower than that of the control group under both tension and torsion loads (Fig. 7). Although stress concentration inside the grooves was observed, those peak stress values (around 7.9 MPa under tension load and 30 MPa under torsion load) were lower than the yielding strength (over 150 MPa) of the cortical bone and were still within an elastic range, indicating that failure may not occur. To conclude, the grooves would increase the ability of the composite against tension and torsion loads.
The FEA results compared favorably to the biomechanical experimental tensile strength-torsion relationships of the specimens; all FEA results presented here fell within the distribution of the biomechanical experimental data (Fig. 8).

Discussion

It is important to obtain a sufficient level of initial bone-cement interlock because some loss of interlock can be expected owing to trabecular resorption with long-term in vivo service. With 10 or more years in service, the loss of interdigitation depth was approximately 75 % [9]. Greater initial interlock between the bone and cement results in more stable constructs with less micro-motion with in vivo service [17, 18]. The quantity of interlock required for stable clinical fixation is not known, but interfaces that have micro-motion in the range of 150 μm [19] are known to result in fibrous tissue formation. The location of failure changed from the metal-cement interface for short-term implantation to the bone-cement interface for longer-term implantation. A shift in the failure mode from the implant-cement interface to the bone-cement interface with time in vivo was observed [20].
Few efforts to the morphology of the bone-cement interface in order to increase the interdigitation have been reported. Arola et al. [12] produced a variety of surface textures within the bone and demonstrated that the stability and initial displacement of the implant were dependent on the surface morphology of medullary canal. The contact area can be enlarged by reducing interfacial gaps, which can be obtained by reducing polymerization shrinkage and air and fluid inclusions [14, 15]. It might be beneficial to brush the cortical bone before cement insertion to increase the bone-cement contact area. More recent investigators have modeled bone-cement interfaces with part of a whole bone structure that included geometries of individual trabeculae determined by micro-CT [10, 21, 22]. We used self-developed reamer to create cortical grooves of the femoral inner wall and form cement circular rings in the canal wall. This increased the interlock contact area between the bone and cement. Theoretically, during infusion of bone cement, more cement will enter the inner wall of the medullary canal. The strength of the bone-cement interface can be improved through the increase of contact area and interdigitation. We investigated the difference in biomechanical behavior of the bone-cement interface in response to tension and torsion loading as a result of changing the morphology of inner wall of the bone medullary canal with grooves. The contact area of the bone-cement interface was greater for the group with cortical grooves when compared to the group without cortical grooves. The tensile strength and anti-torsion of the bone-cement interface were linearly dependent on the average contact area between the bone and cement for both tension and torsion loading conditions. There were very strong correlations between the contact area and model failure for both tensile force and maximal torsion. It suggests that achieving greater contact area of the bone-cement interface is beneficial to optimizing the interfacial strength.
Many techniques aim to improve the cement penetration between the bone and cement and to establish a durable interface. The technologies include pulsed lavage canal, place distal plug, retrograde perfusion of cement, biological type of cement, and oscillation of bone cement [2325]. Use of pulsatile lavage and pressurization of cement has been shown to reduce the initial occurrence of radiolucent lines at the bone-cement interface [26] and increase the amount of cement penetration into the trabecular bone bed [27]. An ideal cement penetration of 3–4 mm has been proposed based on the concept that a minimum of 2 mm is needed to interlock with transverse trabeculae [28]. The penetration depth of bone cement depends on several factors, including the viscosity of cement [29], bone preparation technique [8], pressurized infusion of cement [30], and bone quality and shape, which increases the mechanical properties of bone-cement interface. Mann et al. [31] evaluated mechanical properties of the bone-cement interface under tension and the effects of bone density and bone-cement interdigitation. Increasing the volume of cement in the trabecular bone was recommended for maximizing the interface strength and reducing the incidence of interface failure. Larger cement volumes and contact between cement and cortex are both strongly negatively correlated with high trabecular bone stresses surrounding the cement and may provide a stronger support of cement mantle [32]. A large average interdigitation and contact area in trabecular bone can be achieved by preparing the bone with pulsatile lavage to allow for cement infiltration, but will be more difficult to achieve in the cortical bone. In the present study, the volume of cement was significantly greater for the specimens with grooves when compared to the specimens without grooves. The cement volume was strongly and significantly correlated with tensile force and maximal torsion, respectively. The amount of cement penetration was increased through improving the contact area which created a better initial fixing strength.
In addition, there was a very strong correlation between the contact area and model failure for both tensile force and maximal torsion. The displacement and destruction of the models occurred most at the bone-cement interface. However, the cement-prosthesis interface did not change significantly. The majority of cracks occur in the cement than the bone which is consistent with what was found experimentally [3, 10]. Porosity has been shown to affect the fatigue life of bone cements, and vacuum mixing is widely used to reduce porosity in the clinical setting [33]. In the present study, there was strongly negative correlation between the porosity and the tensile force and maximal torsion. Reduced porosity may increase the static mechanical strength of bone cement and improve the fatigue life of the cement.
As a validation, the FEA results were compared with biomechanical experimental tests. The FEA simulations compared satisfactorily with biomechanical experimental results. The high positive correlation between contact area and interface strength found for the models performed has also been noted in experimental studies of bone-cement specimens. This is consistent with the previous study [3, 10]. Combining the FEA findings with biomechanical experimental tests suggests that achieving a greater contact area between the bone and cement is essential for increasing the interfacial strength. It is believed that the increased contact area between bone and cement could enhance the initial fixation of artificial prosthesis. Our study supports the concept of obtaining sufficient initial interlock [34].
An obvious limitation of this study was that loosening of the interface was examined in an environment comprised of mechanical factors only. Complications associated with the development of interface debris and development of fibrous tissue were not considered. These factors would most likely elicit a synergistic response that potentially accelerates the loosening process with respect to results of the present study. Over the long term, bone resorption may occur at the interface, which considerably weakens the interface [35]. The study only considered time-zero biomechanics and did not consider any changes in vivo such as bone adaptation or resorption. Additional experimental work is needed to delineate the role of micromechanical behavior and gross structural response under fatigue loading. Another recognized limitation of the study was the limited number of specimens prepared and examined. Nevertheless, this study has examined mechanical properties of the bone-cement interface and distinguished that the bone surface morphology contributes to increasing the interlock of the bone-cement interface.

Conclusions

The majority of cracks occurred in the cement than the bone. Changing the surface morphology of the bone medullary canal from a smooth bore cortical tube to one with grooves would make a better interlocking between the bone and cement, which may potentially increase the initial stability and biomechanical strength of the bone-cement interface, and prolongs the survival life of prosthesis.

Acknowledgements

The study was supported by the Shanghai Pujiang Talent Project Grant 13PJD023 and Shanghai Jiaotong University Grant YG2012MS49. The authors wish to thank vice professor Hua Zikai for contributions made to the FEA support used in the present study.

Authors’ contributions

C-LZ designed and organized the experiments. G-QS performed the majority of the experiments and drafted the manuscript. C-LZ and G-QS are co-first authors. K-PZ contributed to technical assistance. D-L contributed to the FEA support. The manuscript was written through contributions of all authors. All authors read and approved the final manuscript.

Competing interests

The authors declare that they have no competing interests.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.
Literatur
1.
Zurück zum Zitat Graham J, Ries M, Pruitt L. Effect of bone porosity on the mechanical integrity of the bone-cement interface. J Bone Joint Surg Am. 2003;85(10):1901–8.PubMed Graham J, Ries M, Pruitt L. Effect of bone porosity on the mechanical integrity of the bone-cement interface. J Bone Joint Surg Am. 2003;85(10):1901–8.PubMed
2.
Zurück zum Zitat Amirfeyz R, Bannister G. The effect of bone porosity on the shear strength of the bone–cement interface. Int Orthop. 2009;33(3):843–6.CrossRefPubMed Amirfeyz R, Bannister G. The effect of bone porosity on the shear strength of the bone–cement interface. Int Orthop. 2009;33(3):843–6.CrossRefPubMed
4.
Zurück zum Zitat Jasty M, Maloney WJ, Bragdon CR. Histomorphological studies of the long-term skeletal responses to well fixed cemented femoral components. J Bone Joint Surg Am. 1990;72(8):1220–9.PubMed Jasty M, Maloney WJ, Bragdon CR. Histomorphological studies of the long-term skeletal responses to well fixed cemented femoral components. J Bone Joint Surg Am. 1990;72(8):1220–9.PubMed
5.
Zurück zum Zitat Janssen D, Mann KA, Verdonschot N. Micro-mechanical modeling of the cement–bone interface: the effect of friction, morphology and material properties on the micromechanical response. J Biomech. 2008;41(15):3158–63.CrossRefPubMedPubMedCentral Janssen D, Mann KA, Verdonschot N. Micro-mechanical modeling of the cement–bone interface: the effect of friction, morphology and material properties on the micromechanical response. J Biomech. 2008;41(15):3158–63.CrossRefPubMedPubMedCentral
6.
Zurück zum Zitat Gardiner RC, Hozack WJ. Failure of the cement–bone interface. A consequence of strengthening the cement–prosthesis interface? J Bone Joint Surg Br. 1994;76B:49–52. Gardiner RC, Hozack WJ. Failure of the cement–bone interface. A consequence of strengthening the cement–prosthesis interface? J Bone Joint Surg Br. 1994;76B:49–52.
7.
Zurück zum Zitat Mohler CG, Callaghan JJ, Collis DK, Johnston RC. Early loosening of the femoral component at the cement–prosthesis interface after total hip replacement. J Bone Joint Surg Am. 1995;77A:1315–22. Mohler CG, Callaghan JJ, Collis DK, Johnston RC. Early loosening of the femoral component at the cement–prosthesis interface after total hip replacement. J Bone Joint Surg Am. 1995;77A:1315–22.
8.
Zurück zum Zitat Berry DJ. Cemented femoral stems: what matters most1 1No benefits or funds were received in support of this study. J Arthroplasty. 2004;19(4):83–4.CrossRefPubMed Berry DJ. Cemented femoral stems: what matters most1 1No benefits or funds were received in support of this study. J Arthroplasty. 2004;19(4):83–4.CrossRefPubMed
9.
Zurück zum Zitat Miller MA, Goodheart JR, Izant TH, Rimnac CM, Cleary RJ, Mann KA. Loss of cement-bone interlock in retrieved tibial components from total knee arthroplasties. Clin Orthop Relat Res. 2014;472(1):304–13.CrossRefPubMed Miller MA, Goodheart JR, Izant TH, Rimnac CM, Cleary RJ, Mann KA. Loss of cement-bone interlock in retrieved tibial components from total knee arthroplasties. Clin Orthop Relat Res. 2014;472(1):304–13.CrossRefPubMed
10.
Zurück zum Zitat Waanders D, Janssen D, Mann KA. The mechanical effects of different levels of cement penetration at the cement–bone interface. J Biomech. 2010;43(6):1167–75.CrossRefPubMedPubMedCentral Waanders D, Janssen D, Mann KA. The mechanical effects of different levels of cement penetration at the cement–bone interface. J Biomech. 2010;43(6):1167–75.CrossRefPubMedPubMedCentral
11.
Zurück zum Zitat Miller MA, Eberhardt AW, Cleary RJ. Micromechanics of postmortem-retrieved cement–bone interfaces. J Orthop Res. 2010;28(2):170–7.PubMedPubMedCentral Miller MA, Eberhardt AW, Cleary RJ. Micromechanics of postmortem-retrieved cement–bone interfaces. J Orthop Res. 2010;28(2):170–7.PubMedPubMedCentral
12.
Zurück zum Zitat Arola D, Stoffel KA, Yang DT. Fatigue of the cement/bone interface: the surface texture of bone and loosening. J Biomed Mater Res B Appl Biomater. 2006;76(2):287–97.CrossRefPubMed Arola D, Stoffel KA, Yang DT. Fatigue of the cement/bone interface: the surface texture of bone and loosening. J Biomed Mater Res B Appl Biomater. 2006;76(2):287–97.CrossRefPubMed
13.
Zurück zum Zitat Tong J, Wong KY, Lupton C. Determination of interfacial fracture toughness of bone–cement interface using sandwich Brazilian disks. Eng Fract Mech. 2007;74(12):1904–16.CrossRefPubMedPubMedCentral Tong J, Wong KY, Lupton C. Determination of interfacial fracture toughness of bone–cement interface using sandwich Brazilian disks. Eng Fract Mech. 2007;74(12):1904–16.CrossRefPubMedPubMedCentral
14.
Zurück zum Zitat Wang JS, Franzen H, Lidgren L. Interface gap after implantation of a cemented femoral stem in pigs. Acta Orthop Scand. 1999;70:234–9.CrossRefPubMed Wang JS, Franzen H, Lidgren L. Interface gap after implantation of a cemented femoral stem in pigs. Acta Orthop Scand. 1999;70:234–9.CrossRefPubMed
15.
Zurück zum Zitat Race A, Miller MA, Clarke MT, Mann KA. Cement-implant interface gaps explain the poor results of CMW3 for femoral stem fixation: a cadaver study of migration, fatigue and mantle morphology. Acta Orthop. 2005;76:679–87.CrossRefPubMedPubMedCentral Race A, Miller MA, Clarke MT, Mann KA. Cement-implant interface gaps explain the poor results of CMW3 for femoral stem fixation: a cadaver study of migration, fatigue and mantle morphology. Acta Orthop. 2005;76:679–87.CrossRefPubMedPubMedCentral
16.
Zurück zum Zitat Lotz JC, Gerhart TN, Hayes WC. Mechanical properties of metaphyseal bone in the proximal femur. J Biomech. 1991;24:317–29.CrossRefPubMed Lotz JC, Gerhart TN, Hayes WC. Mechanical properties of metaphyseal bone in the proximal femur. J Biomech. 1991;24:317–29.CrossRefPubMed
17.
Zurück zum Zitat Reading AD, McCaskie AW, Barnes MR. A comparison of 2 modern femoral cementing techniques: analysis by cement–bone interface pressure measurements, computerized image analysis, and static mechanical testing. J Arthroplasty. 2000;15(4):479–87.CrossRefPubMed Reading AD, McCaskie AW, Barnes MR. A comparison of 2 modern femoral cementing techniques: analysis by cement–bone interface pressure measurements, computerized image analysis, and static mechanical testing. J Arthroplasty. 2000;15(4):479–87.CrossRefPubMed
18.
Zurück zum Zitat Miller MA, Terbush MJ, Goodheart JR, Izant TH, Mann KA. Increased initial cement-bone interlock correlates with reduced total knee arthroplasty micro-motion following in vivo service. J Biomech. 2014;47(10):2460–6.CrossRefPubMedPubMedCentral Miller MA, Terbush MJ, Goodheart JR, Izant TH, Mann KA. Increased initial cement-bone interlock correlates with reduced total knee arthroplasty micro-motion following in vivo service. J Biomech. 2014;47(10):2460–6.CrossRefPubMedPubMedCentral
19.
Zurück zum Zitat Jasty M, Bragdon C, Burke D, O'Connor D, Lowenstein J, Harris WH. In vivo skeletal responses to porous-surfaced implants subjected to small induced motions. J Bone Joint Surg Am. 1997;79(5):707–14.PubMed Jasty M, Bragdon C, Burke D, O'Connor D, Lowenstein J, Harris WH. In vivo skeletal responses to porous-surfaced implants subjected to small induced motions. J Bone Joint Surg Am. 1997;79(5):707–14.PubMed
20.
Zurück zum Zitat Gebert de Uhlenbrock A, Puschel V, Puschel K, et al. Influence of time in-situ and implant type on fixation strength of cemented tibial trays—a post mortem retrieval analysis. Clin Biomech (Bristol, Avon). 2012;27:929–35.CrossRef Gebert de Uhlenbrock A, Puschel V, Puschel K, et al. Influence of time in-situ and implant type on fixation strength of cemented tibial trays—a post mortem retrieval analysis. Clin Biomech (Bristol, Avon). 2012;27:929–35.CrossRef
21.
Zurück zum Zitat Janssen D, Mann KA, Verdonschot N. Finite element simulation of cement-bone interface micromechanics: a comparison to experimental results. J Orthop Res. 2009;27:1312–8.CrossRefPubMedPubMedCentral Janssen D, Mann KA, Verdonschot N. Finite element simulation of cement-bone interface micromechanics: a comparison to experimental results. J Orthop Res. 2009;27:1312–8.CrossRefPubMedPubMedCentral
22.
Zurück zum Zitat Tozzi G, Zhang Q-H, Tong J. 3D real-time micromechanical compressive behaviour of bone-cement interface:experimental and finite element studies. J Biomech. 2012;45:356–63.CrossRefPubMed Tozzi G, Zhang Q-H, Tong J. 3D real-time micromechanical compressive behaviour of bone-cement interface:experimental and finite element studies. J Biomech. 2012;45:356–63.CrossRefPubMed
23.
Zurück zum Zitat Schmale GA, Lachiewicz PF, Kelley SS. Early failure of revision total hip arthroplasty with cemented precoated femoral components. J Arthroplasty. 2000;15(6):718–29.CrossRefPubMed Schmale GA, Lachiewicz PF, Kelley SS. Early failure of revision total hip arthroplasty with cemented precoated femoral components. J Arthroplasty. 2000;15(6):718–29.CrossRefPubMed
24.
Zurück zum Zitat Herberts P, Malchau H. Long-term registration has improved the quality of hip replacement: a review of the Swedish THR Register comparing 160,000 cases. Acta Orthop. 2000;71(2):111–21.CrossRef Herberts P, Malchau H. Long-term registration has improved the quality of hip replacement: a review of the Swedish THR Register comparing 160,000 cases. Acta Orthop. 2000;71(2):111–21.CrossRef
25.
Zurück zum Zitat Wang Y, Han P, Gu W. Cement oscillation increases interlock strength at the cement–bone interface. Orthopedics. 2009;32(5):325.CrossRefPubMed Wang Y, Han P, Gu W. Cement oscillation increases interlock strength at the cement–bone interface. Orthopedics. 2009;32(5):325.CrossRefPubMed
26.
Zurück zum Zitat Ritter MA, Herbst SA, Keating EM. Radiolucency at the bone-cement interface in total knee replacement: the effects of bone-surface preparation and cement technique. J Bone Joint Surg Am. 1994;76(1):60–5.PubMed Ritter MA, Herbst SA, Keating EM. Radiolucency at the bone-cement interface in total knee replacement: the effects of bone-surface preparation and cement technique. J Bone Joint Surg Am. 1994;76(1):60–5.PubMed
27.
Zurück zum Zitat Schlegel UJ, Siewe J, Delank KS. Pulsed lavage improves fixation strength of cemented tibial components. Int Orthop. 2011;35:1165–9.CrossRefPubMed Schlegel UJ, Siewe J, Delank KS. Pulsed lavage improves fixation strength of cemented tibial components. Int Orthop. 2011;35:1165–9.CrossRefPubMed
28.
Zurück zum Zitat Cawley DT, Kelly N, McGarry JP, Shannon FJ. Cementing techniques for the tibial component in primary total knee replacement. Bone Joint J. 2013;95-B:295–300.CrossRefPubMed Cawley DT, Kelly N, McGarry JP, Shannon FJ. Cementing techniques for the tibial component in primary total knee replacement. Bone Joint J. 2013;95-B:295–300.CrossRefPubMed
29.
Zurück zum Zitat Race A, Miller MA, Clarke MT. The effect of low-viscosity cement on mantle morphology and femoral stem micromotion: a cadaver model with simulated blood flow. Acta Orthop. 2006;77(4):607–16.CrossRefPubMed Race A, Miller MA, Clarke MT. The effect of low-viscosity cement on mantle morphology and femoral stem micromotion: a cadaver model with simulated blood flow. Acta Orthop. 2006;77(4):607–16.CrossRefPubMed
30.
Zurück zum Zitat Gozzard C, Gheduzzi S, Miles AW. An in-vitro investigation into the cement pressurization achieved during insertion of four different femoral stems. Proc Inst Mech Eng H J Eng Med. 2005;219(6):407–13.CrossRef Gozzard C, Gheduzzi S, Miles AW. An in-vitro investigation into the cement pressurization achieved during insertion of four different femoral stems. Proc Inst Mech Eng H J Eng Med. 2005;219(6):407–13.CrossRef
31.
Zurück zum Zitat Mann KA, Ayers DC, Werner FW. Tensile strength of the bone-cement interface depends on the amount of bone interdigitated with PMMA cement. J Biomech. 1997;30:339–46.CrossRefPubMed Mann KA, Ayers DC, Werner FW. Tensile strength of the bone-cement interface depends on the amount of bone interdigitated with PMMA cement. J Biomech. 1997;30:339–46.CrossRefPubMed
32.
Zurück zum Zitat Wee H, Armstrong AD, Flint WW, Kunselman AR, Lewis GS. Peri-implant stress correlates with bone and cement morphology: Micro-FE modeling of implanted cadaveric glenoids. J Orthop Res. 2015;33(11):1671–9.CrossRefPubMed Wee H, Armstrong AD, Flint WW, Kunselman AR, Lewis GS. Peri-implant stress correlates with bone and cement morphology: Micro-FE modeling of implanted cadaveric glenoids. J Orthop Res. 2015;33(11):1671–9.CrossRefPubMed
33.
Zurück zum Zitat Baleani M, Bialoblocka-Juszczyk E, Engels GE, et al. The effect of vacuum mixing and pre-heating the femoral component on the mechanical properties of the cement mantle. J Bone Joint Surg. 2010;92(3):454–60.CrossRef Baleani M, Bialoblocka-Juszczyk E, Engels GE, et al. The effect of vacuum mixing and pre-heating the femoral component on the mechanical properties of the cement mantle. J Bone Joint Surg. 2010;92(3):454–60.CrossRef
34.
Zurück zum Zitat Messick KJ, Miller MA, Damron LA. Vacuum-mixing cement does not decrease overall porosity in cemented femoral stems: an in vitro laboratory investigation. J Bone Joint Surg. 2007;89(8):1115–21.CrossRef Messick KJ, Miller MA, Damron LA. Vacuum-mixing cement does not decrease overall porosity in cemented femoral stems: an in vitro laboratory investigation. J Bone Joint Surg. 2007;89(8):1115–21.CrossRef
35.
Zurück zum Zitat Sundfeldt M, Carlsson LV, Johansson CB. Aseptic loosening, not only a question of wear: a review of different theories. Acta Orthop. 2006;77(2):177–97.CrossRefPubMed Sundfeldt M, Carlsson LV, Johansson CB. Aseptic loosening, not only a question of wear: a review of different theories. Acta Orthop. 2006;77(2):177–97.CrossRefPubMed
Metadaten
Titel
Biomechanical effects of morphological variations of the cortical wall at the bone-cement interface
verfasst von
Chun-Lin Zhang
Guo-Qi Shen
Kun-Peng Zhu
Dong-xu Liu
Publikationsdatum
01.12.2016
Verlag
BioMed Central
Erschienen in
Journal of Orthopaedic Surgery and Research / Ausgabe 1/2016
Elektronische ISSN: 1749-799X
DOI
https://doi.org/10.1186/s13018-016-0405-y

Weitere Artikel der Ausgabe 1/2016

Journal of Orthopaedic Surgery and Research 1/2016 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.