Background
Absorbed dose is an important parameter in characterizing the effect of radiation therapy for the efficacy of tumor eradication and protection from unacceptable damage to normal organs [
1]. For historical reasons, in terms of dose, D
w has been assumed for reporting the dose to various media. However, human body is not only composed of water. Many tissues in the body have different densities than water, especially the bones and lung. For radiation therapy the dose absorbed to water cannot accurately represent the actual dose absorbed in different tissues. In practice, traditional treatment planning system (TPS) typically takes the effect of different tissue densities with attenuation and scatter into considerations but reports the dose at each location as the dose to water. Monte Carlo (MC) algorithm is the most accurate algorithm for dose calculation in that it simulates the transport properties of various particles in various media in the region of interest and scores the dose contribution locally to the medium with its assigned chemical composition as well as density. The resulting dose distributions may be different from those calculated by traditional dose calculation algorithms, especially for tissues of heterogeneity [
2‐
4]. In recent years, MC has been increasingly adopted in clinical application [
5‐
7]. There are a number of reasons for using D
w for reporting of MC calculated doses. Two major ones are that it has been used in decades of clinical studies for outcome correlation with the dose, and that the calibration protocols are all referenced to water. A technical issue related to dose calibration is that an MC based TPS could model the chemical composition of various biological tissues by approximation as a function of Computed Tomography (CT) numbers based on data of the human body (reference International Commission on Radiation Units & Measurements reports 44 and 46). Such an approximation may not perform well for non-biological materials like in a quality assurance (QA) phantom. MC based dose calculations typically report absorbed dose to media (D
m). Therefore there is a need to convert between D
m and D
w, and, as Siebers JV et al. [
8] argued, MC is capable of doing the conversion. Siebers et al. presented a method to calculate the difference between D
m and D
w by applying the Bragg-Gray cavity theory, and their results showed a difference exceeding 10% in cortical bones.
Currently there is no consensus regarding whether D
m or D
w should be used for an MC based TPS [
9,
10]. When it comes to clinical application, the difference between D
w and D
m will affect interpretation of dose distribution and perhaps the value of prescription dose, leading to differences in plan evaluation, dose reporting, and dose verification. In this work, D
m and D
w were both calculated using Monaco TPS for 10 nasopharyngeal cancer (NPC) cases, 10 lung cancer cases and 10 bone target cases, in order to investigate the issue in two common clinical sites in which differences of dose distributions may be highlighted. Dose Volume Histogram (DVH) was used to analyze dose parameters in the target and organ at risk (OAR), and three dimensional dose difference distributions between D
m and D
w were calculated. Gamma passing rates (measurement results vs D
m/D
w plans) were calculated at different QA criteria to evaluate the dose accuracy.
Methods
Dm plan originally created for treatment
Ten NPC cases in stage T3 or T4, 10 lung cancer cases and 10 bone target cases (7 cases of lumbar vertebra metastasis, 3 cases of thoracic vertebra metastasis) treated at Sun Yat-sen University Cancer Center were retrospectively chosen in this study. The gross tumor volumes (GTVs) and clinical tumor volume (CTV) were contoured by experienced radiation oncologists according to definitions in the ICRU 50 and ICRU 62 reports [
11,
12], and the planning target volume (PTV) were generated following a set of physician prescribed margins that were consistent with departmental protocols specific to the disease sites. Monaco TPS (Version 5.0, Elekta) was used to create the treatment plans for step-and-shoot IMRT with an Elekta Synergy linac, and MC calculated D
m was chosen for dose reporting. Nine equally spaced fields were used for NPC cases. The prescription of NPC cases and Lung cancer cases were 70 Gy (32 or 33 fractions, 5 days/week) and 65 Gy (26 fractions, 5 days/week) respectively. The main planning objectives for NPC are PTV V
100% > 98% and PTV V
110% < 10% (V
x%, is the percentage volume of reign of interest (ROI) that receives at least x% prescription dose), spinal cord D
2% < 45Gy, brain stem D
2% < 54Gy, parotid gland D
50% < 30Gy, optical nerve D
2% < 54Gy, and the dose to lens as low as possible. For lung IMRT cases 5–7 fields were used. The planning objectives are PTV V
100% > 95% and PTV V
110% < 2%, spinal cord D
2% < 45Gy, normal lung V
20 Gy < 35% (V
D Gy, is the percentage volume of ROI that receives at least absorbed dose D) and normal lung mean dose <19Gy, heart V
30 Gy < 40%, and the maximum esophagus dose <65Gy. For bone target cases, 5–7 fields were used. The prescription of bone target cases was 25 Gy (5Gy/fractions, 5 days/week). The main planning objectives are for PTV, V
100% > 95% and V
110% < 10%, for spinal cord D
max < 26 Gy, for lung V
10Gy < 15%, and the maximum esophagus dose < 26 Gy.
Dw calculation
The MC algorithm in the Monaco TPS used for this study, called XVMC, calculates dose based upon mass density. A technical issue of dose calculation with MC in treatment planning is how to obtain the density and chemical composition data for the patient model from the CT. An approximation is made by assigning a voxel to certain type of tissue in the human body based on its Hounsfield unit (HU) in a certain range, and the mass density and composition data can be looked up in the International Commission on Radiation Units & Measurements Reports No. 46 [
13]. XVMC algorithm converts CT numbers to ED numbers using the user-defined CT-to-ED calibration table and takes with a fit function that maps continuously the electron density to mass density for matching a tissue with approximating cross section and attenuation coefficient data [
14].
The conversion to D
w can be calculated based on the distribution of D
m plan according to the Bragg-Gray cavity theory:
$$ {\mathrm{D}}_{\mathrm{w}}={\mathrm{D}}_{\mathrm{m}}\ {s}_{w, med} $$
(1)
where
sw,med is the mean unconstrained mass stop power ratio of water to media of primary electron spectrum, and D
w is understood as the dose to the voxel replacement of water embedded to the actual media. Theoretically mass stop power ratio can be calculated by the following formula [
8]:
$$ {s}_{w, med}={\int}_0^{E_{max}}{\left({\Phi}_E\right)}_m{\left(S/\rho \right)}_w dE/{\int}_0^{E_{max}}{\left({\Phi}_E\right)}_m{\left(S/\rho \right)}_{med} dE $$
(2)
where (
S/
ρ)
w and (
S/
ρ)
med are the unconstrained mass stop power of water and media, respectively. (Φ
E)
m is the primary electron fluence in the medium and
Emax is the maximum energy in the (Φ
E)
m distribution. The stopping power ratio in Moncao was pre-calculated by approximation for tissue-like media.
The conversion from Dm to Dw in Monaco with a clinically accepted plan involved a simple recalculation with exactly the same set of plan parameters (all the geometric parameters and monitor units (MU)) retained. The stopping power ratios dependent of mass density were applied voxel by voxel. The matrix of dose calculation grid was 0.3 cm × 0.3 cm × 0.3 cm, and the Monte Carlo statistical uncertainty was set at 3% per control point.
Dm and Dw dose verification
All the plans were measured with MapCHECK2 (Sun Nuclear, Florida, USA) to verify the dose distribution. MacpCHECK2 was mounted in a water-equivalent phantom (MapPHAN) with a 5 cm equivalent depth from the surface to the detectors. The TPS planed dose was calculated on the real phantom CT images without overriding the density. The measured dose distributions of composite fields were compared with the corresponding planned dose distributions (Dm or Dw), and the local dose normalization gamma (γ) passing rates were calculated at the setting dose difference (DD) and distance to agreement (DTA). In order to eliminate dose in the out-of-field region where a large relative dose difference can be calculated and hence skew theγ result, a lower dose threshold (10%) was set and below the threshold theγ result was ignored. Using 3%&3 mm, 2%&2 mm and 1%&1 mm tolerances, the gamma passing rates were calculated to find how the pass rates change with reduction of dose difference and DTA limits.
Data analysis
According to the ICRU 83 report, the volume-dose is recommended to describe the dose information in the ROIs, as D
x% to note the dose that X% of volume of ROI receives [
15]. For example, D
98% means 98% of volume received the dose at specified value such as 65Gy. These DVH parameters were used for statistical analysis of D
w and D
m dose distributions. The bin width of the DVHs was 1 cGy, and the resolution for DVH sampling was 0.1 cm. The difference between the D
w and D
m was calculated by:
$$ \mathrm{Diff}\ \left(\%\right)=\left({\left({\mathrm{D}}_{\mathrm{x}\%}\right)}_{\mathrm{w}}-{\left({\mathrm{D}}_{\mathrm{x}\%}\right)}_{\mathrm{m}}\right)/{\left({\mathrm{D}}_{\mathrm{x}\%}\right)}_{\mathrm{w}}\times 100 $$
(3)
The plan subtraction method was used to evaluate the spatial dose difference distribution of Dw and Dm.
Paired t-tests were performed using the SPSS software (Version 19, SPSS, Inc., USA) to determine the statistical significance of the difference between Dw and Dm, with a p-value < 0.05 as the threshold for consideration as statistically significant.
Discussions
With the application of MC algorithm for dose calculation in radiation therapy, whether the dose should be calculated to medium or to water has been an unsettled debate [
9,
10,
16]. The arguments that support D
w include that beam data was measured in water, that the beam output was calibrated in water, and that most clinical experience were based on dose to water, etc. However, the compelling argument to support the use of D
m is that it represents the true dose at each location of specific medium. It is the unique advantage of Monte Carlo in that D
m can be calculated directly, but D
m to D
w using stopping power ratios may involve an uncertainty [
17]. In reality, different TPS use different dose calculation algorithms to produce D
w, from direct calculation to applying conversion factors. According to the AAPM TG 105 report [
18], when the element components are considered in dose calculation, both D
m and D
w should be available for evaluation. When comes to a specific clinical situation, the difference between D
m and D
w should be known. N Dogan et al. [
19] showed that converting D
m to D
w in EGS4 MC-calculated IMRT treatment plans introduces a systematic error in target and critical structure DVHs, and this systematic error may reach up to 5.8% for H&N and 8.0% for prostate cases when the hard-bone-containing structures such as femoral heads are present.
From our work using Monaco for NPC and lung cancer, D
m was less than D
w. The mean deviation for soft tissues was within 2%. For T-M joints and mandibular, the mean deviation was greater than 5%, and in regions of unspecified normal bone the difference could reach 10%. Our results agreed nicely with the work by Siebers et al. [
8]. It is interesting to find, based on our study, that there was hardly any difference between D
w and D
m in low density regions. Although the stopping power ratio for both cortical bone and air can be above 1.10, the stopping power ratio is close to 1 for low density tissues like lung. For this reason, the issue with using D
w or D
m may have a minimal effect for majority of clinical situations.
The dose difference between D
w and D
m in bony structures may become clinically significant if the OAR is receiving doses close to its tolerance dose limit which can influence selection or rejection of a particular plan. The dose calculated by MC may need to be carefully evaluated in certain situations, e.g. bone metastasis, bone tumor, or constraining a hot spot in bone that becomes a limiting factor in plan optimization. From the Fig.
3, for PTV of the bone target cases, though the target dose coverages (the target volume (%) received the prescription dose) of D
m and converted D
w plan were similar, the mean median dose of D
w plan increased by 3.5% comparing with that of D
m plan (Table
3). That means the dose prescription for bone target could be about 3.5% higher than that of using D
w dose, and their treatment response and outcome may need further study in the future.
Previous studies [
16,
20] using EGS4/MCSIM Monte Carlo and AXB dose calculations proved that conventional model based algorithms predicted dose distributions in bone that were closer to D
m distributions than to D
w distributions. It is therefore better to use D
m for consistency with previous radiation therapy experience. Our measurements showed that at widely used reference standard, 3% dose difference and 3 mm DTA, the D
m and D
w plan gamma passing rates were very close, but when the gamma calculation standard became stricter, the D
w was closer to the result of measurement than the D
m. That’s because the MapCheck2 CT images without forcing density were used to calculate the planned dose distribution, where the MapCheck2 detectors are made of high density metallic elements and the detectors are always calibrated by D
w. The CT scanner used for acquisition of patient simulation images has the limitation of scanning high density material such as the diode and the TPS also has limitation while accepting CT images with high density material. In our practice, D
m is used for treatment planning, and physicians and physicists will be consulted in case conversion to D
w in bone may affect the decisions to choose the appropriate dose distribution for treatment.
Conversion to D
w may be necessary for dose verification in the quality assurance phantom. If a water phantom is used, the difference between D
m and D
w can be ignored. Kan MW et al. [
20] showed that for a heterogeneous phantom with high density materials contained the difference between D
m and D
w has an effect on the passing rate of QA measurement. Our results (Table
4) showed there were obvious differences between the D
m and D
w plan gamma passing rates when the QA criteria became strict. A simple method to bypass the problem is to assign a uniform density to the phantom and calculate to either D
m or D
w in a consistent manner. The choice of an appropriate density needs to be validated by an independent method such as point dose measurement.