Erschienen in: BMC Oral Health 1/2020

Open Access 01.12.2020 | Research article

Computer-aided design of polyetheretherketone for application to removable pediatric space maintainers

verfasst von: Hui Guo, Yong Wang, Yijiao Zhao, He Liu

Erschienen in: BMC Oral Health | Ausgabe 1/2020

Abstract

Background

The premature loss of primary teeth is a common problem in pediatric dentistry, resulting in disruption of the arch integrity. Hence, space maintainers (SM) used for maintaining spaces are necessary. However, current methods of making removable space maintainers (RSM) have some limitations.

Methods

Digital models of dentition defects were obtained by using a scanning technique coupled with laser medical image reconstruction. The digital RSMs were designed using the 3Shape software. They were manufactured using two methods: polyetheretherketone (PEEK), and conventional methods (20 RSMs per group). For qualitative evaluation, the Likert five-point scale was used by 10 experts to obtain a score for 40 RSMs. The spaces between the tissue surfaces of the RSMs and the models were replaced using silicone, and the maximum and mean distances, as well as the standard deviation, were measured. A three-dimensional variation analysis was used to measure these spaces. The student’s t-test and Satterthwaite t-test were used to compare the differences in the spaces for the various materials.

Results

The PEEK RSMs were found to fit the models well. In the qualitative assessment, the mean experts’ scores for the PEEK and conventional groups were 1.80 ± 0.40 and 1.82 ± 0.40, and there was no significant difference between the two groups (p = 0.875). In the quantitative assessment, the mean spaces for the PEEK digital RSMs and the conventional RSMs were 44.32 ± 1.75 μm, and 137.36 ± 18.63 μm, respectively, and the differences were statistically significant (p < 0.001). In addition, there were significant differences in the maximum space and the standard deviation between the two groups.

Conclusion

Digitally designed and integrated RSMs were found to be superior to those produced using the conventional method. 3D variation analysis results showed that the mean distances and standard deviations of the PEEK groups were significantly smaller than those of conventional group (p < 0.01). A PEEK-manufactured RSM produced using CAD/CAM would be extremely suitable for clinical applications.
Hinweise
Send reprint requests and other post-publication matters to: He Liu, Department of Pediatric Dentistry, Peking University School of Stomatology, 22 Zhongguancun Nandajie, Haidian District, 100081, Beijing, China
Phone: +86-18811792829
E-mail: pediatric14@163.​com

Supplementary information

Supplementary information accompanies this paper at https://​doi.​org/​10.​1186/​s12903-020-01184-6.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Background

The premature loss of primary teeth is a common problem in the pediatric dentistry, resulting in the disruption of the arch integrity and adversely affecting the proper alignment of permanent successors [1]. Hence, space maintainer (SM) [2] are used for maintaining the space [3]. The removable space maintainer (RSM) is a kind of SM, designed for use with contiguous primary molar teeth loss, and its application is recommended by the profession [4]. There are several advantages to use an RSM [5], including maintaining the proximal, distal, and mesial lengths of a space, while maintaining the vertical height, thus restoring the aesthetics of the teeth [6], preventing speech disorders [7], and eliminating habits such as unilateral chewing. However, conventional RSMs incur some drawbacks [8], particularly in terms of their design and manufacture. For example, the manufacture of RSMs is complicated, because it is technically very sensitive, requires experienced technicians, and the product outcomes exhibit large individual variations. In addition, as the manufacture of RSMs features the use of curved snap rings and self-curing resin, it is difficult to ensure the precision of the snap rings within the space maintainers. During the polyreaction of the self-curing resin, shrinkage occurs [9], adversely affecting the fit between the tissue surface of the maintainer and the mucosa between the snap ring and the abutment. As a result, pediatric patients tend to adapt poorly to space maintainers. Furthermore, owing to the scarcity of artificial deciduous teeth products on the market, artificial permanent teeth products are usually modified to emulate the functionality of deciduous teeth; however, they do not accurately simulate their morphology. Hence, a precise, convenient, rapid design and manufacturing method is needed to address these problems.
Computer-aided design and manufacturing (CAD/CAM) has been used since the 1980s [10, 11]. It is now widely used in dentistry, for oral medicine, particularly in the field of prosthodontics [1214]. When the same material is used with both CAD/CAM and traditional restoration techniques, the former yields more stable and durable results [15]. Currently, digital design techniques and rapid molding are typically used to design removable partial denture (RPD) metal scaffolds [16] and to manufacture their resin models [17], or to directly manufacture the RPD metal scaffold [1823]. The present study set out to address the digital method of manufacturing an RPD.
PEEK [24] is a special engineering plastic with high rigidity and toughness [25], good flame-retardant properties, high mechanical performance [26], and high resistance to temperature, corrosion [27], and radiation. In addition, it offers a high level of biocompatibility [28] and resistance to fatigue caused by alternating stresses [29]. It is used in fixed prosthodontics, such as crown bridge repair and healing abutments, and to fabricate the scaffold and snap rings of an RPD. In addition, Ierardo et al. [30] conducted a pilot study using PEEK to fabricate an RSM with a dental CAD/CAM system, finding PEEK to be highly suitable for the fabrication of space maintainers.
Therefore, this study set out to investigate the application of CAD/CAM design to the RSMs used in pediatric dentistry and to evaluate the suitability of the technique for clinical applications.

Materials and methods

Construction of digital model for dentition defects

A standard negative model was used (Nissin™, China) to produce super-hard mixed dentition plaster (Type I dental plaster, Pegasus™, China) which was scanned using a three-dimensional (3D) model scanner (D800, 3Shape A/S, Denmark) to produce digital models (Fig. 1a). A partial-defect dentition model was constructed by removing the first and second deciduous molars from both sides of the standard model’s mandible. This standard model includes all the primary teeth and the four first molars. Then, a clear digital edentulism model was obtained by scanning the partial model (Fig. 1b). To ensure that the digital arrangement requirements for artificial teeth were satisfied by this model, the upper jaw model was also scanned and its occlusal relationship with the mandible model was evaluated.

Digital design of RSMs

Dental CAD software (Dental System 2017, 3Shape A/S, Denmark) and reverse engineering software (Geomagic Studio 2014, Geomagic Inc., USA) were jointly used to design all the components of the RSM.
As the 3Shape software database does not include deciduous teeth and given that the anatomical differences between deciduous and permanent molars are considerable, artificial digitized models of standard teeth crowns were constructed for this experiment. The complete standard digital jaw model data were then imported into the Geomagic Studio 2014 software and the “constructing sample boundary” function was used to define the boundary of each tooth. The processed data were then exported in stereolithography (STL) format. The anatomical and implant database in the anatomy element was selected from the control panel of the 3Shape software, and the new database was created by clicking “add” under the database menu and named “deciduous teeth database.” The model in the scan database was selected and the constructed teeth data were then imported into the database. The “Scan It Library” function was then used for tooth editing (Fig. 2). Bottom cutting, dental crown changing, dental crown shrinking, debris removal, and boundary and deformation point addition were all performed. Then, the models in the teeth database were edited. The teeth sequence and positions were arranged according to professional requirements and “smile editing” was performed. Hence, artificial digitized models of standard teeth crowns were constructed.
The RPD+ temporary tooth design module in 3Shape was selected. As RSMs are transitional prosthodontics, the selection of materials for RSMs should be different from that of RPD for adults. This is because during the growth and development phase, snap rings at the buccal side of the teeth are used minimally to avoid affecting the development of the width of the dental arch. Therefore, the effects of the maintainer base on the jaw and dental arch should be considered in the design, and the base should either be very short or have no vestibular base, and should instead have a long lingual base. Under conditions in which the mandible is fastened, snap rings would not be required, and fixation would be solely dependent on the lingual base.
The sites of missing teeth were labelled on the mandibular edentulism digital model, after which the digitized models of standard teeth crown were imported into the software. The “virtual articulator” function was used to adjust the anatomical module of the artificial teeth to ensure the functionality of the occlusion. The “major connector” design was chosen for the maintainer base and its thickness was set to the conventional value of 1.5 mm. Since the connector to the partial denture is relatively thin and represents the location where stress is most concentrated, its thickness was increased to 2.5 mm. The lingual side of the base was extended to the distal and middle parts of the first molar to form a base for the RSM. Finally, the finished set of RSM data were exported in STL format (Fig. 3a–h).
A 5-axis numerical-control milling machine (Organical Multi, R + K GmbH, Germany) was used to shape 20 PEEK material samples (Fig. 4a) into RSMs. After removing the support structure, the maintainer was polished and fitted onto a super-hard plaster model. Accordingly, a total of 40 super-hard plasters were fabricated. These were divided into two groups of 20.
The mandibular edentulism model described in Section 2.1 was used to fabricate the conventional RSM. The base was fabricated using Type 2 Class I denture base polymer powder (self-curing denture powder, 1R, biomimetic color, NISSIN™, Japan) and Heraeus three-layer synthetic resin teeth (Heraeus GmbH, Germany). After the maintainer was fabricated, it was fitted to a different super-hard plaster model (Fig. 4b).

Suitability of different morphological components of RSMs

To qualitatively assess each RSM, the RSM was fitted to the mandibular plaster model, and then a preliminary evaluation of the model’s feasibility was performed through observations using a compression method. The assessment criteria for RPDs were used as a reference for the observations, based on the following principles proposed by Frank et al. [31], confirming that: [1] all occlusal rests were completely in place [2]; the rigid components of the denture contacted the corresponding abutment; and [3] no visible space greater than 1 mm existed between the main connector and the model. In this experiment, since no occlusal rests were involved, the second and third of Frank’s criteria were used for the assessment. In the compression method, a cement filler was used to compress the occlusal support vertically. The suitability was considered to be good if the denture did not significantly warp. Ten experts from the Department of Pediatrics of Peking University Hospital of Stomatology assessed the suitability of the 40 RSMs, assigning a score between 1 and 5 (Likert scale), where 1 was the most satisfactory and 5 was the least satisfactory.
For the quantitative assessment of RSMs, a silicone impression material (Variotime Light Flow, Heraeus GmbH, Germany) was injected onto the tissue surface of the RSM [32]. The RSM was then placed onto the super-hard plaster model, and a vertical force of 20 N was applied for 10 min until the silicone had completely solidified. Excess silicone was eliminated to prevent the removal of the RSM, thereby affecting the analysis results. The RSM was then removed, allowing the silicone gel to remain on the super-hard plaster model.
The space between the denture and the model was then measured using digital analysis [33], and a dental scanner (Smart Optics, 880 GmbH, Germany) was used to obtain the corresponding digital silicone film model data. Sets of film model and raw digital model data were imported to the Geomagic software, and the data from the raw digital model were fixed as a reference. The digital silicone film model was set as the measurement subject. Three identical position points within the two models were selected for registration. The silicone film region in the digital silicone film model was selected as the measurement area and the “3D variation analysis” function was used to calculate the mean differences in the selected area (i.e., the differences in the mean thickness of the silicone film) between the two models. At the same time, a variation chromatogram was generated to display the thickness of the silicone film in various regions (Fig. 5 a, b). This analysis function was able to simultaneously calculate the maximum distance (i.e., the maximum thickness of the silicone film) for the two digital models; and these data were recorded as the “maximum space,” which was then used to express the maximum space between the denture tissue surface and the model. The “standard deviation” was recorded to express the uniformity of the space between the denture tissue surface and the model. This method was repeated for each of the forty model samples.

Statistical analyses

Data management and statistical analysis were performed using the SPSS software program (version 22.0 IBM). Shapiro-Wilk and Bartlett tests were applied to the continuous variables. A student’s t-test was applied to the continuous variables to compare the PEEK and conventional groups for which the continuous variables satisfy the normal distribution and homogeneity of variance. Cohen’s d value was calculated for these effect sizes. A Satterthwaite t-test was applied to those continuous variables which were consistent with the normal distribution but not the homogeneity of variance. Glass’s △ value was calculated for these effect sizes. For the qualitative evaluation, 10 experts used the Likert five-point scale to apply a score to 40 RSMs. Kendall’s coefficient of concordance was used to determine the consistency of scoring. The average score assigned by the experts was used as the comprehensive expert score for the models. The power of the test was achieved using Proc power in the University Edition of the Statistical Analysis System (version SAS Studio 9.4). For the qualitative assessment of the RSMs, the null hypothesis was that there is no significant difference between the scores assigned by the experts to the two groups. For the quantitative assessment of the RSMs, the null hypothesis was that there is no significant difference in the mean distances, maximum distances, and standard deviations of the two groups. All p values were two-tailed, with a significance level of 0.05.

Results

Qualitative assessment results

The two types of RSMs exhibited a good fit onto the super-hard plaster standard model—there was tight contact between the denture base and the model, and between the major connecter and its lower plaster model; no 1 mm-spaces were observed and no significant warping was present. The compression of the RSM components using the cement filler also produced no significant warping, thus making them suitable for clinical applications.
The data is normally distributed (the results of the normality are listed in attached Table 1). Table 1 shows that the consistency evaluation of the experts for the PEEK group (Kendall = 0.556) is higher than that for the conventional group (Kendall = 0.484), which means that the suitability of the RMSs in the PEEK group were easier to assess and the scores were easier to converge.
Table 1
Comparison of consistency of experts’ evaluations of the two processing methods for the PEEK and conventional groups
Group
Kendall W
χ2
p value
PEEK
0.556
105.705
< 0.001
Conventional
0.484
92.037
< 0.001
All
0.498
194.263
< 0.001
As listed in Table 2, the mean scores assigned by the experts for the PEEK and conventional groups were 1.80 ± 0.40 and 1.82 ± 0.40, respectively, while there was no significant difference between the two groups (p = 0.875). (Fig. 6).
Table 2
Comparison of experts’ comprehensive scores for the two processing methods for the PEEK and conventional groups
 
Comprehensive experts’ score
PEEK group (n = 20)
1.80 ± 0.40
Conventional group (n = 20)
1.82 ± 0.40
Power
0.053
t value
−0.158
p value
0.875
Cohen’s d
0.050

Quantitative assessment results

The data is normally distributed (the results of the normality test are listed in attached Table 2). Table 3 shows that the 3D variation analysis results revealed that the mean spaces for the PEEK digital RSMs and the conventional RSMs were 44.32 ± 1.75 μm, and 137.36 ± 18.63 μm, respectively, with the differences being statistically significant (p < 0.001). The maximum space values for the PEEK digital RSMs and the conventional RSMs were 602.71 ± 11.94 μm and 1261.88 ± 249.34 μm, respectively, while the differences were also statistically significant (p < 0.001). The standard deviation values for the PEEK digital RSMs and the conventional RSMs were 72.21 ± 5.72 μm and 361.22 ± 91.30 μm, respectively, while the differences were again statistically significant (p < 0.001).
Table 3
Comparison of markers for the two processing methods for the PEEK and conventional groups
 
PEEK group
Conventional group
Power
t value
p value
Glass’s△ (95%CI)
Maximum distance
602.71 ± 11.94
1261.88 ± 249.34
> 0.999
11.809
< 0.001
2.64 (2.19–3.10)
Mean distance
44.32 ± 1.75
137.36 ± 18.63
> 0.999
22.241
< 0.001
4.99 (4.52–5.46)
Standard deviation
72.21 ± 5.72
361.22 ± 91.30
> 0.999
14.129
< 0.001
3.17 (2.70–3.63)
As listed in Table 3, for the maximum distances, mean distances, and standard deviation, the Glass’s △ were 2.64(2.19–3.10), 4.99(4.52–5.46), and 3.17(2.70–3.63), respectively. According to the relative and effective sizes of the differences shown in Table 4, the maximum distances, mean distances, and standard deviation of the two groups were large (effect size > 1.4). The power analysis is illustrated in attached Fig. 1.
Table 4
Effect size index
Relative Size
Effect Size
% of control group below the mean of the experimental group
 
0.0
50%
Small
0.2
58%
Medium
0.5
69%
Large
0.8
79%
 
1.4
92%
As shown in Fig. 7, the maximum distances, mean distances and standard deviation of the PEEK group were significantly smaller than those of the conventional group, with the differences being statistically significant (p < 0.05).

Discussion

An RSM offers multiple advantages; however, to date, it has presented problems in terms of its manufacture. Given that the denture base is composed of self-curing resin, shrinkage will occur as part of the curing process. The frictional resistance between the resin and the gypsum model partially inhibits the shrinkage. Thus, when the RSM cools to room temperature, latent stress will exist inside the base. Over the long term, this stress will be released slowly, resulting in the deformation of the base, the formation of micro cracks or cracks in and on the surface of the base resin, and even eventually lead to the fracture of the denture [34]. In this study, a new CAD/CAM method was used to help overcome these problems.
Currently, the application of CAD/CAM techniques to RPD manufacture focuses primarily on the design and manufacture of the scaffolds. Meanwhile, other denture components are still manufactured using the conventional processes. There are several related studies, such as that by Pooya et al. [35], who evaluated the overall accuracy and fit of conventional versus CAD/CAM RPD frameworks based on the results of a STL data analysis, and further evaluated the accuracy and fit of each component of the RPD framework. In a case report, Virard et al. [36] proposed an innovative procedure for producing an immediate removable denture, based on the use of three components: an intra-oral scanner, CAD with two different software used sequentially, and CAM with a 5-axis machine. The prosthodontics department at the Peking University Hospital of Stomatology [37] has reported on the case of a removable partial denture that was completely manufactured using CAD/CAM. Digital design has also been used in orthodontic dentistry, such as orthodontic space maintainers [38]. In this study, an update of the CAD/CAM software allowed us to use a new version of the 3Shape software, such that the frameworks and artificial teeth could be designed in the same design module, thus realizing a fully integrated design.
To the best of the authors’ knowledge, few studies have reported on the use of CAD/CAM to design and manufacture RSMs in pediatric dentistry. The use of this method could overcome the problems associated with traditional manufacturing. Such problems include the cost and the poor cooperation of children, as well as the size of intraoral scanners, in that they are too big to be applied to primary dentitions. In addition, there have been no reports detailing the use of CAD/CAM to design and manufacture RSMs. Studies into the area of digital technology being applied to space maintainers have also been rarely reported. Ji et al. [39] designed SM “band with loop” devices using CAD software, and these were manufactured by milling of PEEK blocks. Ierardo et al. (2017) fabricated orthodontic SMs in PEEK polymer through a digital workflow. However, they did not design or use artificial teeth. Soni [38] used digital methods to make a BruxZir zirconia SM for a six-year-old female patient with a chronic intra-radicular abscess in the upper right first primary molar which was treated by extraction. In this study, the technical route of using CAD/CAM to design and manufacture RSMs was explored, and the digitally prepared RSMs fitted the model well, which provided an important basis for clinical application.
In this study, PEEK was used to manufacture digital RSMs. This material has been widely used for the fabrication of clinical denture bases due to its good physiochemical, mechanical, and biological performance, and given that it may be easily cut and molded through the application of CAD/CAM technology. For children, it is safer and easier to make, and possibly superior to conventional methods which use artificial teeth and self-curing resin.
The present study found that, when using a CAD/CAM system, the maximum and mean distances and standard deviation for the PEEK group were significantly smaller than those for the conventional group, which indicates that the use of PEEK is promising for clinical applications. To qualitatively assess the compatibility of digitally designed RSMs, visual observation and a compression technique commonly used in clinical treatment and research were employed, and the results showed a good fit for the two types of RSMs. For quantitative assessment, reports by Stern et al. [40] and Dunham et al. [41] were referenced. They measured the space between the denture and the model to assess the former’s compatibility. According to the mean scores assigned by the experts, there was no significant difference between the two groups. This result may because the differences in the spaces were difficult to distinguish with the naked eye. Hence, to measure the space between the denture and the model, a 3D variation analysis commonly used to assess the suitability of fixed prostheses was employed [35]. The results showed that the maximum and mean distances and standard deviation for the PEEK group were significantly smaller than those for the conventional group, and these differences were statistically significant (p < 0.05). Therefore, the compatibility of the digital RSMs fabricated in this study is better than that of traditional types, largely because the conventional fabrication of RSMs is overly complex in terms of model construction; difficulties include the polymerization shrinkage of the self-curing resin, and the grinding and polishing of the RSM, which are prone to errors. Ye et al. [42] found that the space under a cast metal large connecter was 131.1 ± 87.1 μm (7.5–353.0 μm), while that under a PEEK large connecter was 52.8 ± 44.6 μm (0.5–177.4 μm). In this study, the mean space of the PEEK and conventional RMSs were 44.32 ± 1.75 μm and 137.36 ± 18.63 μm, respectively, and their values were all smaller than the above research results. The differences between the results of this study and the above research results may be caused by the progress of fabrication technology, the development of the associated equipment, the different structures of the RPD and RSM, and the different measurement areas.
In this study, digital design and fabrication simplified the manufacturing process, reduced deformation and errors, and improved the suitability of RSMs. Apart from the abovementioned three case reports, there have been no further clinical studies about the utilization of PEEK polymers in pediatric dentistry.
Currently, the fabrication cost of PEEK maintainers is high, and the color options are limited. When used in the frontal teeth region, the color of the base and the artificial teeth may affect the aesthetics. The color of PEEK maintainers can be improved by changing the performance of the material or by dyeing it. Future studies will focus on improving the color of the PEEK maintainers, the mechanical performance of various other RSMs, and seek out additional materials similar to PEEK. Moreover, research into the wear resistance of PEEK artificial teeth is also underway.

Conclusion

In conclusion, the feasibility of CAD/CAM was examined to develop new PEEK-integrated RSMs and validated CAD/CAM, which provides a foundation for future clinical applications. This research has produced an original and creative method for combining digitalization technology and the clinical requirements of pediatric dentistry, thus opening up many new possibilities. The results of 3D variation analysis showed that the mean spaces for the PEEK digital RSMs and the conventional RSMs were 44.32 ± 1.75 μm, and 137.36 ± 18.63 μm, respectively, and the differences were statistically significant (p < 0.001). Design improvements and the search for new manufacturing materials require further research.

Supplementary information

Supplementary information accompanies this paper at https://​doi.​org/​10.​1186/​s12903-020-01184-6.

Acknowledgements

We thank Wenjing Li, Ph. D candidate of Department of Periodontology, Peking University School of Stomatology for her helpful review and revision of this manuscript.
There is no need for ethics approval and consent. The manuscript does not report on or involve any animals, humans, human data, human tissue, or plants.
The manuscript contains no individual person’s data in any form (including individual details, images or videos).

Competing interests

The authors declare that they have no competing interests.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
Zurück zum Zitat Park K, Jung DW, Kim JY. Three-dimensional space changes after premature loss of a maxillary primary first molar. Int J Paediatr Dent. 2009;19:383–9.PubMed Park K, Jung DW, Kim JY. Three-dimensional space changes after premature loss of a maxillary primary first molar. Int J Paediatr Dent. 2009;19:383–9.PubMed
2.
Zurück zum Zitat Qudeimat MA, Sasa IS. Clinical success and longevity of band and loop compared to crown and loop space maintainers. Eur Arch Paediatr Dent. 2015;16:391–6.PubMed Qudeimat MA, Sasa IS. Clinical success and longevity of band and loop compared to crown and loop space maintainers. Eur Arch Paediatr Dent. 2015;16:391–6.PubMed
3.
Zurück zum Zitat Ahmad AJ, Parekh S, Ashley PF. Methods of space maintenance for premature loss of a primary molar: a review. Eur Arch Paediatr Dent. 2018;9:311–20. Ahmad AJ, Parekh S, Ashley PF. Methods of space maintenance for premature loss of a primary molar: a review. Eur Arch Paediatr Dent. 2018;9:311–20.
4.
Zurück zum Zitat Law CS. Management of premature primary tooth loss in the child patient. J Calif Dent Assoc. 2013;41:612–8.PubMed Law CS. Management of premature primary tooth loss in the child patient. J Calif Dent Assoc. 2013;41:612–8.PubMed
5.
Zurück zum Zitat Goenka P, Sarawgi A, Marwah N, Gumber P, Dutta S. Simple fixed functional space maintainer. Int J Clin Pediatr Dent. 2014;7:225–8.PubMed Goenka P, Sarawgi A, Marwah N, Gumber P, Dutta S. Simple fixed functional space maintainer. Int J Clin Pediatr Dent. 2014;7:225–8.PubMed
6.
Zurück zum Zitat Mohammad Z, Cheruku SR, Penmetcha S, Bagalkotkar A, Kumari S. A novel approach to regain anterior space using modified 2 by 3 fixed appliance: a report of two cases. J Clin Diagn Res. 2015;9:ZD23–5.PubMedPubMedCentral Mohammad Z, Cheruku SR, Penmetcha S, Bagalkotkar A, Kumari S. A novel approach to regain anterior space using modified 2 by 3 fixed appliance: a report of two cases. J Clin Diagn Res. 2015;9:ZD23–5.PubMedPubMedCentral
7.
Zurück zum Zitat Kalia G, Tandon S, Bhupali NR, Rathore A, Mathur R, Rathore K. Speech evaluation in children with missing anterior teeth and after prosthetic rehabilitation with fixed functional space maintainer. J Indian Soc Pedod Prev Dent. 2018;36:391–5.PubMed Kalia G, Tandon S, Bhupali NR, Rathore A, Mathur R, Rathore K. Speech evaluation in children with missing anterior teeth and after prosthetic rehabilitation with fixed functional space maintainer. J Indian Soc Pedod Prev Dent. 2018;36:391–5.PubMed
8.
Zurück zum Zitat Srivastava N, Grover J, Panthri P. Space maintenance with an innovative “tube and loop” space maintainer (Nikhil appliance)[J]. Int J Clin Pediatr Dent. 2016;9(1):86.PubMedPubMedCentral Srivastava N, Grover J, Panthri P. Space maintenance with an innovative “tube and loop” space maintainer (Nikhil appliance)[J]. Int J Clin Pediatr Dent. 2016;9(1):86.PubMedPubMedCentral
9.
Zurück zum Zitat Kawara M, Komiyama O, Kimoto S, Kobayashi N, Kobayashi K, Nemoto K. Distortion behavior of heat-activated acrylic denture-base resin in conventional and long, low-temperature processing methods. J Dent Res. 1998;77:1446–53.PubMed Kawara M, Komiyama O, Kimoto S, Kobayashi N, Kobayashi K, Nemoto K. Distortion behavior of heat-activated acrylic denture-base resin in conventional and long, low-temperature processing methods. J Dent Res. 1998;77:1446–53.PubMed
10.
Zurück zum Zitat Kanazawa M, Inokoshi M, Minakuchi S, Ohbayashi N. Trial of a CAD/CAM system for fabricating complete dentures. Dent Mater J. 2011;30:93–6.PubMed Kanazawa M, Inokoshi M, Minakuchi S, Ohbayashi N. Trial of a CAD/CAM system for fabricating complete dentures. Dent Mater J. 2011;30:93–6.PubMed
11.
Zurück zum Zitat Duret F, Blouin JL, Duret B. CAD/CAM in dentistry. J Am Dent Assoc. 1988;117(1939):715–20.PubMed Duret F, Blouin JL, Duret B. CAD/CAM in dentistry. J Am Dent Assoc. 1988;117(1939):715–20.PubMed
12.
Zurück zum Zitat Moldovan O, Luthardt RG, Corcodel N, Rudolph H. Three-dimensional fit of cad/cam-made zirconia copings. Dent Mater. 2011;27:1273–8.PubMed Moldovan O, Luthardt RG, Corcodel N, Rudolph H. Three-dimensional fit of cad/cam-made zirconia copings. Dent Mater. 2011;27:1273–8.PubMed
13.
Zurück zum Zitat Beuer F, Schweiger J, Edelhoff D. Digital dentistry: an overview of recent developments for CAD/CAM generated restorations. Br Dent J. 2008;204:505–11.PubMed Beuer F, Schweiger J, Edelhoff D. Digital dentistry: an overview of recent developments for CAD/CAM generated restorations. Br Dent J. 2008;204:505–11.PubMed
14.
Zurück zum Zitat Quaas S, Rudolph H, Luthardt RG. Direct mechanical data acquisition of dental impressions for the manufacturing of CAD/CAM restorations. J Dent. 2007;35:903–8.PubMed Quaas S, Rudolph H, Luthardt RG. Direct mechanical data acquisition of dental impressions for the manufacturing of CAD/CAM restorations. J Dent. 2007;35:903–8.PubMed
15.
Zurück zum Zitat Stawarczyk B, Ender A, Trottmann A, Özcan M, Fischer J, Hämmerle CHF. Load-bearing capacity of CAD/CAM milled polymeric three-unit fixed dental prostheses: effect of aging regimens. Clin Oral Investig. 2012;16:1669–77.PubMed Stawarczyk B, Ender A, Trottmann A, Özcan M, Fischer J, Hämmerle CHF. Load-bearing capacity of CAD/CAM milled polymeric three-unit fixed dental prostheses: effect of aging regimens. Clin Oral Investig. 2012;16:1669–77.PubMed
16.
Zurück zum Zitat Bibb R, Eggbeer D, Williams R. Rapid manufacture of removable partial denture frameworks. Rapid Prototyp J. 2006;12:95–9. Bibb R, Eggbeer D, Williams R. Rapid manufacture of removable partial denture frameworks. Rapid Prototyp J. 2006;12:95–9.
17.
Zurück zum Zitat Wu L, Lv PJ, Wang Y, Ai H. Hanjun. Establishment of 3D digital observability of dentition defect model. Chin J Stomatol. 2009;44:117–21. Wu L, Lv PJ, Wang Y, Ai H. Hanjun. Establishment of 3D digital observability of dentition defect model. Chin J Stomatol. 2009;44:117–21.
18.
Zurück zum Zitat Wu L, Lv PJ, Wang Y, Ai H, Dong D. Computer-aided design and fabrication of removable partial denture stent framework. Chin J Stomatol. 2006;41:432–5. Wu L, Lv PJ, Wang Y, Ai H, Dong D. Computer-aided design and fabrication of removable partial denture stent framework. Chin J Stomatol. 2006;41:432–5.
19.
Zurück zum Zitat Bibb RJ, Eggbeer D, Williams RJ, Woodward A. Trial fitting of a removable partial denture framework made using computer-aided design and rapid prototyping techniques. Proc Inst Mech Eng H. 2006;220:793–7.PubMed Bibb RJ, Eggbeer D, Williams RJ, Woodward A. Trial fitting of a removable partial denture framework made using computer-aided design and rapid prototyping techniques. Proc Inst Mech Eng H. 2006;220:793–7.PubMed
20.
Zurück zum Zitat Abduo J, Lyons K, Bennamoun M. Trends in computer-aided manufacturing in prosthodontics: a review of the available streams. Int J Dent. 2014;783948. Abduo J, Lyons K, Bennamoun M. Trends in computer-aided manufacturing in prosthodontics: a review of the available streams. Int J Dent. 2014;783948.
21.
Zurück zum Zitat Williams RJ, Bibb R, Eggbeer D, Collis J. Use of CAD/CAM technology to fabricate a removable partial denture framework. J Prosthet Dent. 2006;96:96–9.PubMed Williams RJ, Bibb R, Eggbeer D, Collis J. Use of CAD/CAM technology to fabricate a removable partial denture framework. J Prosthet Dent. 2006;96:96–9.PubMed
22.
Zurück zum Zitat Zhu SY, Cai YH, Dai N, Yuanjun Y. Preliminary clinical application of CAD/CAM casting technology for removable partial denture casting stent. Stomatology. 2009;03:126–9. Zhu SY, Cai YH, Dai N, Yuanjun Y. Preliminary clinical application of CAD/CAM casting technology for removable partial denture casting stent. Stomatology. 2009;03:126–9.
23.
Zurück zum Zitat Eggbeer D, Bibb R, Williams R. The computer-aided design and rapid prototyping fabrication of removable partial denture frameworks. Proc Inst Mech Eng H J Eng Med. 2005;219:195–202. Eggbeer D, Bibb R, Williams R. The computer-aided design and rapid prototyping fabrication of removable partial denture frameworks. Proc Inst Mech Eng H J Eng Med. 2005;219:195–202.
24.
Zurück zum Zitat Bezzon OL, Mattos MG, Ribero RF. Surveying removable partial dentures: the importance of guiding planes and path of insertion for stability. J Prosthet Dent. 1997;78:412–8.PubMed Bezzon OL, Mattos MG, Ribero RF. Surveying removable partial dentures: the importance of guiding planes and path of insertion for stability. J Prosthet Dent. 1997;78:412–8.PubMed
25.
Zurück zum Zitat Fan JP, Tsui CP, Tang CY, Chow CL. Influence of interphase layer on the overall elasto-plastic behaviors of HA/PEEK biocomposite. Biomaterials. 2004;25(23):5363–73.PubMed Fan JP, Tsui CP, Tang CY, Chow CL. Influence of interphase layer on the overall elasto-plastic behaviors of HA/PEEK biocomposite. Biomaterials. 2004;25(23):5363–73.PubMed
26.
Zurück zum Zitat Schmidlin PR, Stawarczyk B, Wieland M, Attin T, Hämmerle CHF, Fischer J. Effect of different surface pre-treatments and luting materials on shear bond strength to peek [J]. Dent Mater. 2010;26:553–9.PubMed Schmidlin PR, Stawarczyk B, Wieland M, Attin T, Hämmerle CHF, Fischer J. Effect of different surface pre-treatments and luting materials on shear bond strength to peek [J]. Dent Mater. 2010;26:553–9.PubMed
27.
Zurück zum Zitat Uhrenbacher J, Schmidlin PR, Keul C, Eichberger M, Roos M, Gernet W, Stawarczyk B. The effect of surface modification on the retention strength of polyetheretherketone crowns adhesively bonded to dentin abutments. J Prosthet Dent. 2014;112:1489–97.PubMed Uhrenbacher J, Schmidlin PR, Keul C, Eichberger M, Roos M, Gernet W, Stawarczyk B. The effect of surface modification on the retention strength of polyetheretherketone crowns adhesively bonded to dentin abutments. J Prosthet Dent. 2014;112:1489–97.PubMed
28.
29.
Zurück zum Zitat Schwitalla AD, Abou-Emara M, Spintig T, Lackmann J, Müller WD. Finite element analysis of the biomechanical effects of peek dental implants on the peri-implant bone. J Biomech. 2015;48:1–7.PubMed Schwitalla AD, Abou-Emara M, Spintig T, Lackmann J, Müller WD. Finite element analysis of the biomechanical effects of peek dental implants on the peri-implant bone. J Biomech. 2015;48:1–7.PubMed
30.
Zurück zum Zitat Ierardo G, Luzzi V, Lesti M, Vozza I, Brugnoletti O, Polimeni A, Bossù M. PEEK polymer in orthodontics: a pilot study on children. J Clin Exp Dent. 2017;9:e1271–5.PubMedPubMedCentral Ierardo G, Luzzi V, Lesti M, Vozza I, Brugnoletti O, Polimeni A, Bossù M. PEEK polymer in orthodontics: a pilot study on children. J Clin Exp Dent. 2017;9:e1271–5.PubMedPubMedCentral
31.
Zurück zum Zitat Frank RP, Brudvik JS, Leroux B, Milgrom P, Hawkins N. Relationship between the standards of removable partial denture construction, clinical acceptability, and patient satisfaction. J Prosthet Dent. 2000;83:521–7.PubMed Frank RP, Brudvik JS, Leroux B, Milgrom P, Hawkins N. Relationship between the standards of removable partial denture construction, clinical acceptability, and patient satisfaction. J Prosthet Dent. 2000;83:521–7.PubMed
32.
Zurück zum Zitat Sarot JR, Contar CM, da Cruz AC, de Souza Magini R. Evaluation of the stress distribution in CFR-PEEK dental implants by the three-dimensional finite element method. J Mater Sci Mater Med. 2010;21:2079–85.PubMed Sarot JR, Contar CM, da Cruz AC, de Souza Magini R. Evaluation of the stress distribution in CFR-PEEK dental implants by the three-dimensional finite element method. J Mater Sci Mater Med. 2010;21:2079–85.PubMed
33.
Zurück zum Zitat Zoidis P, Bakiri E, Polyzois G. Using modified polyetheretherketone (PEEK) as an alternative material for endocrown restorations: a short-term clinical report [J]. J Prosthet Dent. 2017;117:335–9.PubMed Zoidis P, Bakiri E, Polyzois G. Using modified polyetheretherketone (PEEK) as an alternative material for endocrown restorations: a short-term clinical report [J]. J Prosthet Dent. 2017;117:335–9.PubMed
34.
Zurück zum Zitat Nogueira SS, Ogle RE, Davis EL. Comparison of accuracy between compression-and injection-molded complete dentures [J]. J Prosthet Dent. 1999;82(3):291–300.PubMed Nogueira SS, Ogle RE, Davis EL. Comparison of accuracy between compression-and injection-molded complete dentures [J]. J Prosthet Dent. 1999;82(3):291–300.PubMed
35.
Zurück zum Zitat Soltanzadeh P, Suprono MS, Kattadiyil MT, Goodacre C, Gregorius W. An in vitro investigation of accuracy and fit of conventional and CAD/CAM removable partial denture frameworks. J Prosthodont. 2019;28:547–55.PubMed Soltanzadeh P, Suprono MS, Kattadiyil MT, Goodacre C, Gregorius W. An in vitro investigation of accuracy and fit of conventional and CAD/CAM removable partial denture frameworks. J Prosthodont. 2019;28:547–55.PubMed
36.
Zurück zum Zitat Virard F, Venet L, Richert R, Pfeffer D, Viguié G, Bienfait A, Farges JC, Ducret M. Manufacturing of an immediate removable partial denture with an intraoral scanner and CAD-CAM technology: a case report. BMC Oral Health. 2018;18:120.PubMedPubMedCentral Virard F, Venet L, Richert R, Pfeffer D, Viguié G, Bienfait A, Farges JC, Ducret M. Manufacturing of an immediate removable partial denture with an intraoral scanner and CAD-CAM technology: a case report. BMC Oral Health. 2018;18:120.PubMedPubMedCentral
37.
Zurück zum Zitat Ye H, Ning J, Li M, Niu L, Yang J, Sun Y, Zhou Y. Preliminary clinical application of removable partial denture frameworks fabricated using computer-aided design and rapid prototyping techniques. Int J Prosthodont. 2017;30:348–53.PubMed Ye H, Ning J, Li M, Niu L, Yang J, Sun Y, Zhou Y. Preliminary clinical application of removable partial denture frameworks fabricated using computer-aided design and rapid prototyping techniques. Int J Prosthodont. 2017;30:348–53.PubMed
38.
Zurück zum Zitat Soni HK. Application of CAD-CAM for fabrication of metal-free band and loop space maintainer. J Clin Diagn Res. 2017;11:ZD14–6.PubMedPubMedCentral Soni HK. Application of CAD-CAM for fabrication of metal-free band and loop space maintainer. J Clin Diagn Res. 2017;11:ZD14–6.PubMedPubMedCentral
39.
Zurück zum Zitat Kun J, Dinggui Z, Wei L, Li D, Wang X, Jing JL, Wu G. Clinical application of digital space maintainer fabricated by polyetherketoneketone for premature loss of deciduous teeth [J/CD]. Chin J Stomatol. 2019;13:368–72. Kun J, Dinggui Z, Wei L, Li D, Wang X, Jing JL, Wu G. Clinical application of digital space maintainer fabricated by polyetherketoneketone for premature loss of deciduous teeth [J/CD]. Chin J Stomatol. 2019;13:368–72.
40.
Zurück zum Zitat Stern MA, Brudvik JS, Frank RP. Clinical evaluation of removable partial denture rest seat adaptation. J Prosthet Dent. 1985;53:658–62.PubMed Stern MA, Brudvik JS, Frank RP. Clinical evaluation of removable partial denture rest seat adaptation. J Prosthet Dent. 1985;53:658–62.PubMed
41.
Zurück zum Zitat Dunham D, Brudvik JS, Morris WJ, Plummer KD, Cameron SM. A clinical investigation of the fit of removable partial dental prosthesis clasp assemblies. J Prosthet Dent. 2006;95:323–6.PubMed Dunham D, Brudvik JS, Morris WJ, Plummer KD, Cameron SM. A clinical investigation of the fit of removable partial dental prosthesis clasp assemblies. J Prosthet Dent. 2006;95:323–6.PubMed
42.
Zurück zum Zitat Ye H, Li X, Wang G, et al. A novel computer-aided design/computer-assisted manufacture method for one-piece removable partial denture and evaluation of fit [J]. Int J Prosthodont. 2018;31(2):149–51.PubMed Ye H, Li X, Wang G, et al. A novel computer-aided design/computer-assisted manufacture method for one-piece removable partial denture and evaluation of fit [J]. Int J Prosthodont. 2018;31(2):149–51.PubMed
Metadaten
Titel
Computer-aided design of polyetheretherketone for application to removable pediatric space maintainers
verfasst von
Hui Guo
Yong Wang
Yijiao Zhao
He Liu
Publikationsdatum
01.12.2020
Verlag
BioMed Central
Erschienen in
BMC Oral Health / Ausgabe 1/2020
Elektronische ISSN: 1472-6831
DOI
https://doi.org/10.1186/s12903-020-01184-6

Weitere Artikel der Ausgabe 1/2020

BMC Oral Health 1/2020 Zur Ausgabe

Parodontalbehandlung verbessert Prognose bei Katheterablation

19.04.2024 Vorhofflimmern Nachrichten

Werden Personen mit Vorhofflimmern in der Blanking-Periode nach einer Katheterablation gegen eine bestehende Parodontitis behandelt, verbessert dies die Erfolgsaussichten. Dafür sprechen die Resultate einer prospektiven Untersuchung.

Invasive Zahnbehandlung: Wann eine Antibiotikaprophylaxe vor infektiöser Endokarditis schützt

11.04.2024 Endokarditis Nachrichten

Bei welchen Personen eine Antibiotikaprophylaxe zur Prävention einer infektiösen Endokarditis nach invasiven zahnärztlichen Eingriffen sinnvoll ist, wird diskutiert. Neue Daten stehen im Einklang mit den europäischen Leitlinienempfehlungen.

Zell-Organisatoren unter Druck: Mechanismen des embryonalen Zahnwachstums aufgedeckt

08.04.2024 Zahnmedizin Nachrichten

Der Aufbau von Geweben und Organen während der Embryonalentwicklung wird von den Zellen bemerkenswert choreografiert. Für diesen Prozess braucht es spezielle sogenannte „Organisatoren“. In einer aktuellen Veröffentlichung im Fachjournal Nature Cell Biology berichten Forschende durch welchen Vorgang diese Organisatoren im Gewebe entstehen und wie sie dann die Bildung von Zähnen orchestrieren.

Die Oralprophylaxe & Kinderzahnheilkunde umbenannt

11.03.2024 Kinderzahnmedizin Nachrichten

Infolge der Umbenennung der Deutschen Gesellschaft für Kinderzahnheilkunde in Deutsche Gesellschaft für Kinderzahnmedizin (DGKiZ) wird deren Mitgliederzeitschrift Oralprophylaxe & Kinderzahnheilkunde in Oralprophylaxe & Kinderzahnmedizin umbenannt. Aus diesem Grunde trägt die erste Ausgabe in 2024 erstmalig den neuen Titel.

Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Zahnmedizin und bleiben Sie gut informiert – ganz bequem per eMail.