Consequences of reaming with flat and convex reamers for bone volume and surface area of the glenoid; a basic science study

verfasst von: Anne Karelse, Steven Leuridan, Alexander Van Tongel, Philippe Debeer, Jos Van Der Sloten, Kathleen Denis, Lieven F. De Wilde

Erschienen in: Journal of Orthopaedic Surgery and Research | Ausgabe 1/2015

Abstract

Background

The effect of reaming on bone volume and surface area of the glenoid is not precisely known. We hypothesize that (1) convex reamers create a larger surface area than flat reamers, (2) flat reamers cause less bone loss than convex reamers, and (3) the amount of bone loss increases with the amount of version correction.

Methods

Reaming procedures with different types of reamers are performed on similar-sized uniconcave and biconcave glenoids created from Sawbones foam blocks. The loss of bone volume, the size of the remaining surface area, and the reaming depth are measured and evaluated.

Results

Reaming with convex reamers results in a significantly larger surface area than with flat reamers for both uniconcave and biconcave glenoids (p = 0.013 and p = 0.001). Convex reamers cause more bone loss than flat reamers, but the difference is only significant for uniconcave glenoids (p = 0.007).

Conclusions

In biconcave glenoids, convex reamers remove a similar amount of bone as flat reamers, but offer a larger surface area while maximizing the correction of the retroversion. In pathological uniconcave glenoids, convex reamers are preferred because of the conforming shape.
Hinweise

Competing interests

Each author certifies that he or she has no financial or non-financial associations that might pose a conflict of interest in connection with the submitted article.
No duplicate publications or similar reports have been published and no patents are withheld.

Authors’ contributions

This basic research was performed at the University Hospital of Ghent and at the Department of Mechanical Engineering of the Catholic University of Leuven, Belgium. AK is the first author. AK, AVT, and LDW are surgeons that performed the reaming procedure and reviewed the article. SL produced and measured all the samples and wrote the material and methods section. KD and JV are mechanical engineers who developed the bone samples and the protokol for measurement and analysis of the reamed samples. All authors read and approved the final manuscript.

Background

Glenohumeral osteoarthritis is often associated with glenoid bone deformation and deficiency due to chondral and bone erosion. The erosion is concentric in approximately 60 % and eccentric in over 30 % according to Walch et al. [1] (Fig. 1). In total shoulder arthroplasty, the increased retroversion and erosion of the glenoid are associated with a higher rate of loosening of the glenoid component [2, 3]. Optimal positioning of the glenoid prosthesis seems to be essential to achieve good long-term results. To obtain this, the surgeon should aim to correct the retroversion, while minimizing glenoid bone loss and creating a maximal and congruent contact surface area to support the prosthesis [2, 4, 5]. In glenoids with concentric erosion, type A according to Walch, this brings few difficulties. With limited reaming, a congruent surface with a maximum contact area of the supporting bone offers optimal initial stability to the implant. In contrast, in type B glenoids with eccentric erosion, this cause more problems [6, 7]. It has been suggested to correct the retroversion to as close to the native version as possible (to within 5°); however, the exact amount of correction has not been clearly defined. Eccentric downreaming can correct less severe retroversion, but the amount of reaming is limited by the glenoid bone volume and by the medialization of the joint line.
It is not known how much bone is exactly removed by reaming or how this reaming affects the glenoid supporting area with respect to the pathology of the glenoid. The purpose of this study is to quantify bone loss and contact surface area of uniconcave and biconcave glenoids after reaming with different types of reamers. We hypothesize that (1) convex reamers create a larger surface area than flat reamers, (2) flat reamers cause less bone loss than convex reamers, and (3) the amount of bone loss increases with the amount of version correction.

Methods

Bone models

Seventy-two glenoid bone models were created from Sawbones foam blocks: 36 with a uniconcave shape and 36 with a biconcave shape, hereby mimicking type A and type B2 glenoids according to the Walch classification [1]. The Sawbones solid rigid polyurethane foam (Sawbones, Malmo, Sweden) has material properties similar to the subchondral glenoid bone [8]. The dimensions of an original female biconcave glenoid (82-year-old woman with glenohumeral osteoarthritis) were used to create the B2 glenoid models. The dimensions of the B2 glenoid were obtained from a CT scan of the glenoid. The radius of the inferior circle of the native glenoid was 15 mm [9, 10]. The retroversion measured according to Friedman et al. [11] was 12°. From this CT scan, an Standard Tesselation Language (STL) surface was extracted using Mimics (Materialise, Haasrode, Belgium), which was used to prepare the computer-aided design (CAD) drawings and generate computer-aided manufacturing (CAM) commands for the milling process in NX 7.5 (Siemens PLM, Plano, TX, USA). In this way, the STL surface of the patient was replicated onto the Sawbones blocks. The A model glenoids were ovoid in shape [12, 13] and were not CT-based, but were chosen comparable in size to the B models, measuring 30 mm in width, 39 mm in length, and with a depth of 5 mm. The version is neutral, 0°. Again, starting from a STL surface of the CAD drawing, CAM commands were generated for the milling process in NX 7.5 (Siemens PLM, Plano, TX, USA). Both type A and B2 glenoids were milled from the polyurethane blocks using a three-axis milling machine (Haas, Oxnard, CA, USA) (Fig. 2a, b). Automation of the milling process ensured that all fabricated glenoid blocks were exact within manufacturing tolerance (10 μm).

Methodology

A setup is prepared with the bone blocks positioned vertically at surgical working height. The reaming procedure we used was validated in 2013 [14]. Three surgeons and three authors perform the reaming representing an experienced, intermediate, and inexperienced surgeon with respectively over 50, over 20, and under 20 total shoulder arthroplasties performed per year. Each surgeon individually defines the preferred center of the glenoid for the reaming procedure using a flat semicircular guide (Zimmer, Warsaw, IN). The surgeons individually determine the direction of reaming to obtain the intended correction of the version. For the A glenoids, the aim is to keep the version neutral. The B2 glenoids have a retroversion of 12°, and the aim is to correct this as close to neutral as possible. For the A glenoids, reaming is performed until the reamer is over its entire surface in contact with the glenoid bone, creating a smooth bone bed. For the B2 glenoids, reaming is performed similarly taking into account a correction to a neutral version (Fig. 3). Four different reamers with the same radius are used: a convex reamer guided by a K-wire (Global AP, diameter 33 mm, Depuy, Warsaw, Indiana), a convex reamer guided by a nipple (Global Advantage, diameter 33 mm, nipple 6 mm, Depuy, Warsaw, Indiana), a flat reamer guided by a K-wire (custom made, diameter 30 mm), and a flat reamer guided by a nipple (diameter 30 mm, nipple 6 mm, Zimmer, Warsaw, Indiana). All reamers are used with the companies’ instruments with a set arm length (18 cm). Each surgeon reams three A and three B2 glenoids with the four different reamers. This results in 24 reaming procedures per surgeon, 72 all together.

Parameters

All bone blocks are scanned using 3D laser coordinate measuring machine (CMM) (MC16, Coord3, Turin, Italy) before and after reaming. The 3D CMM uses a laser to scan the surface of the blocks resulting in a dense point cloud of points lying on that surface. These resulting point clouds are processed using GOM Inspect (Braunschweig, Germany), and STLs are built in 3-matic (Materialise, Haasrode, Belgium) based on these point clouds. These STLs are used in the further analyses. All bone models are aligned in the software to the same identical coordinate system (“world coordinate system”), ensuring comparability between the parameters of different blocks. The parameters extracted for all reamed bone blocks are the loss of bone volume, the size of the remaining surface area, and the reaming depth. The bone volume removed is calculated based on the STLs of the respective block taken before and after reaming, similar to the procedure described by Youngpravat et al. [15]. The three direction angles are calculated with respect to the local anatomical X-, Y-, and Z-axes (“anatomical coordinate system”) of the A and B2 glenoids defined according to Verstraeten et al. [16] (Fig. 4). Defining the angles with respect to this, local anatomical coordinate system allows for a uniform and clinically relevant interpretation of the angles for A and B2 blocks. Repeatability of the parameter extraction procedure is verified for all parameters with ten repetitions per parameter and results in a mean standard deviation of 0.23 % on the parameter values.

Statistics

Statistical analyses are performed using IBM SPSS Statistics, version 21 (SPSS Inc., Chicago, IL, USA). Hypothesis testing between two groups is performed using a t test if both groups to be compared were normally distributed according to a Shapiro-Wilk test or using a Mann-Whittney U test if one of the groups failed to pass the normality test. When more than two levels per factor are compared, an ANOVA analysis is carried out if the normality assumption is satisfied or a Kruskal-Wallis test if this assumption is not fulfilled. Regression analyses are carried out to assess the relationship between continuous parameters (e.g., direction angles) and relevant outcomes (e.g., bone loss). Significance is assessed at the 5 % level.

Results

A glenoids

Convex reamers cause significantly more bone loss than flat reamers (p = 0.007) (Table 1, Fig. 5). Reaming with convex reamers results in a significantly (p = 0.013) larger surface area than flat reamers (Table 2, Fig. 6) and a significantly greater average depth of reaming (p < 0.001). We find no significant difference in bone loss (p = 0.174), surface area (p = 0.521), and depth (p = 0.278) between reaming with a K-wire or a nipple-guided reamer, for both flat and convex reamers. The regression between bone loss and the three direction angles is not significant (p = 0.4). Hence, none of the three direction angles show a significant relation to the bone loss: X direction angle (p = 0.566), Y direction angle (p = 0.108), and Z direction angle (p = 0.568). The regression between the surface area and the three direction angles is not significant (p = 0.058): X direction angle (p = 0.083), Y direction (p = 0.070), and Z direction (p = 0.219). The regression of bone loss to the depth of reaming shows a significant relation (p < 0.001, R 2 = 0.469). Every millimeter of additional reaming depth accounts for an extra 215 mm3 of bone loss for the given A glenoid samples.
Table 1
Bone loss in A and B2 glenoids for convex and flat reamers
Reamer
Glenoid
No.
Bone loss (mm3)
Mean
SD
Convex
A
18
1686
352
Flat
A
18
1381
343
Convex
B2
18
1779
447
Flat
B2
18
1807
473
Table 2
Surface area in A and B2 glenoids for convex and flat reamers
Reamer
Glenoid
No.
Surface area (mm2)
Mean
SD
Convex
A
18
864
14
Flat
A
18
744
24
Convex
B2
18
794
38
Flat
B2
18
730
42

B2 glenoids

There is no significant difference in bone loss between flat and convex reamers (p = 0.855). Reaming with convex reamers results in a significantly larger surface than with flat reamers (p = 0.001). The average depth of reaming is significantly greater with convex reamers than with flat reamers (p < 0.001). There is no significant difference in bone loss (p = 0.174) and depth (p = 0.449) between reaming with a K-wire or a nipple-guided reamer, both for flat and convex reamers. A significant difference however exists in the reaming area (p < 0.001) between reaming with a K-wire or a nipple-guided reamer for flat reamers. No significant difference in reaming area is recorded for K-wire versus nipple-guided wires for convex reamers (p = 0.529). The regression between bone loss and the X direction angle is significant (p = 0.002, R 2 = 0.249). The regression coefficient shows that every angle degree of correction along the X-axis results in an additional bone loss of 56 mm3 for the B2 bone samples used. The Y direction angle (p = 0.943) and the Z direction angle (p = 0.288) show no significant relation to the bone loss. There is a significant difference in reaming angle between surgeons 1 and 2 for the X direction angle (p = 0.14) and between surgeons 1 and surgeon 3 (p < 0.001). No significant difference can be found between surgeons 2 and 3 (p = 0.296). Surgeon 3 corrects to an average X direction angle of 83.9° (±2.257), surgeon 2 corrects to an average X direction angle of 85.67° (±3.33), and surgeon 1 corrects to an average X direction angle of 89.58° (±3.83°). The regression between the surface area and the three direction angles is not significant (p = 0.817): X direction angle (p = 0.459), Y direction (p = 0.792), and Z direction (p = 0.856). The regression of bone loss to the depth of reaming shows a significant relation (p = 0.001, R 2 = 0.290). Every millimeter of additional reaming depth accounts for an extra 235 mm3 of bone loss for the given B2 glenoid samples.

Discussion

Glenoid component failure remains the most important indication for revision surgery of total shoulder arthroplasty [1719]. Biomechanical studies have shown that placement of a glenoid component in more than 10° of retroversion causes eccentric loading of the prosthesis, and this can lead to instability, rocking horse phenomenon, and early loosening [7, 2024]. Correction of the version helps to restore the glenohumeral relationship and rebalances the force couple of the rotator cuff. Downreaming of the anterior glenoid is an accepted method to correct the retroversion, but limited by the volume of the glenoid vault. Excessive reaming can result in loss of glenoid bone stock and medialization of the joint line jeopardizing solid fixation and with the risk of peg perforation [2527]. If the retroversion is less than 15° to 20°, downreaming of the anterior glenoid is advised. However, there are no explicit guidelines regarding the amount of version that can be safely corrected by eccentric reaming without compromising the glenoid bone stock [28]. We do know that there are limits and clinical experience proved reverse shoulder arthroplasty to be a viable surgical option to solve both the problem of severe glenoid erosion in patients with a biconcave glenoid without rotator cuff insufficiency [29].
The amount of bone resected by the different types of reamers (nipple or K-wire guided, flat and convex) is unknown. To our knowledge, this is the first study investigating the effect of reaming with different reamers on bone volume and surface area in two different-shaped glenoids. This study shows that convex reamers cause more bone loss than flat reamers in uniconcave type A glenoids. This is partly due to the deeper reaming range as a result of the convexity. Corrective reaming of biconcave type B2 glenoids with convex reamers tends to cause slightly more bone loss than with flat reamers, but the difference is not significant. In A glenoids, the reaming angle is as close to neutral as possible, so this does not interfere with bone loss. The depth of reaming does have a significant effect on bone loss, and every millimeter of additional reaming depth accounts for an extra 215 mm3 of bone loss. In B2 glenoids, the angle of correction along the X-axis (representing the version angle correction) is an important factor in determining the bone loss; every additional degree of correction along the X-axis results in an extra 56 mm3 of bone loss. Similarly, the depth of reaming has an important effect on bone loss; every millimeter of additional reaming depth accounts for an extra 235 mm3 of bone loss. Obviously, it is the degree of retroversion and biconcavity, and the intended correction, which dictates the loss of bone volume after reaming in biconcave glenoids. If a surgeon decides to correct more by reaming, this has a direct effect on the amount of bone loss. There is a significant difference between surgeons in the correction of version in the B2 glenoids in this study. This is probably due to the surgeons’ intention and experience to correct as close as possible to the native version [15]. In recent publications, Iannotti et al. [30] and Karelse et al. [14] came to a similar conclusion that in biconcave glenoids, correction of version by reaming is not reproducible. Until now, corrective reaming is performed by “carpenters eye,” helped by the experience and the natural 3D orientation of the surgeon. Several studies show that the accuracy of the position of the glenoid prosthesis in the transverse plane can be improved by intraoperative navigation and patient-specific instrumentation; this is particularly so in severely retroverted glenoids [3135]. Convex reamers create a larger surface area than flat reamers in both A and B2 glenoids, and this is not affected by the correction angle. This finding differs from the results from Youngpravat et al. [15], where smaller version corrections increase the surface area. A larger surface area obviously increases the contact area for a glenoid component. In biconcave glenoids, the convex reamers are at slight disadvantage to flat reamers concerning bone loss, but they win back in a larger surface area of the glenoid after reaming. For uniconcave type A glenoids, which are considered non-pathological glenoids, reaming with convex reamers causes more bone loss than with flat reamers. The difference in surface area between the reamers is small given the fact that reaming depth must be minimal in these non-eroded glenoids. If however glenoids are centrally eroded to type A1 and A2 glenoids according to Walch, and excessive medialization of the joint line should be avoided, minimal reaming with a more conforming reamer is the objective. A convex reamer with a radius of curvature mimicking the radius of the native articular surface can maximally preserve the surface area and existing bone stock in centrally eroded glenoids [36], whereas flat reamers would reduce both surface area and bone stock. Another explanation for the reduced bone loss after flat reaming can be that the radius of flat reamers is chosen accordingly to the largest radius of the glenoid, thereby reaming mainly the circumferential bone and not reaching the centrally eroded part.
The surface area of B2 glenoids is larger after reaming over K-wires than nipple guided using flat reamers. The difference may partly be explained by the difference in diameter of the K-wire and the nipple, 2 and 6 mm, respectively.
We are aware of the limitations of this study. We performed reaming procedures on foam blocks in a surgical setup but without the intraoperative conditions that can be of great influence to a procedure. We created only two types of morphology while we are aware of the large variation of the concavity of the glenoid. Nevertheless, we believe that this study offers valuable information that can be of help in future decisions on reaming strategy and possibly influence the choice and development of flat or curved backed glenoid prostheses for certain pathological glenoids [3639].

Conclusion

This study shows that the characteristics of the reamer and the experience of the surgeon influence the amount of bone removal and the remaining surface area of the glenoid. These findings account for the two morphologic types studied: A and B2 glenoids. Convex reamers are due to their conforming shape that is better indicated in pathological A glenoids, but the convexity of the reamer should be optimally adapted to the pathological curvature [38]. In B glenoids, convex reamers are preferred because they remove a similar amount of bone as flat reamers but offer a larger surface area while maximizing the correction of the retroversion.

Acknowledgements

We are grateful for the financial funding received from the BVOT (Belgische Vereniging voor Orthopedie en Traumatologie). This enabled us to finance the production of the glenoid bone models.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.

Competing interests

Each author certifies that he or she has no financial or non-financial associations that might pose a conflict of interest in connection with the submitted article.
No duplicate publications or similar reports have been published and no patents are withheld.

Authors’ contributions

This basic research was performed at the University Hospital of Ghent and at the Department of Mechanical Engineering of the Catholic University of Leuven, Belgium. AK is the first author. AK, AVT, and LDW are surgeons that performed the reaming procedure and reviewed the article. SL produced and measured all the samples and wrote the material and methods section. KD and JV are mechanical engineers who developed the bone samples and the protokol for measurement and analysis of the reamed samples. All authors read and approved the final manuscript.
Literatur
1.
Zurück zum Zitat Walch G, Badet R, Boulahia A, Khoury A. Morphologic study of the glenoid in primary glenohumeral osteoarthritis. J Arthroplasty. 1999;14:756–60.CrossRefPubMed Walch G, Badet R, Boulahia A, Khoury A. Morphologic study of the glenoid in primary glenohumeral osteoarthritis. J Arthroplasty. 1999;14:756–60.CrossRefPubMed
5.
Zurück zum Zitat Walch G, Young AA, Boileau P, Loew M, Gazielly D, Mole D. Patterns of loosening of polyethylene keeled glenoid components after shoulder arthroplasty for primary osteoarthritis: results of a multicenter study with more than five years of follow-up. J Bone Joint Surg Am. 2012;94:145–50. doi:10.2106/JBJS.J.00699.CrossRefPubMed Walch G, Young AA, Boileau P, Loew M, Gazielly D, Mole D. Patterns of loosening of polyethylene keeled glenoid components after shoulder arthroplasty for primary osteoarthritis: results of a multicenter study with more than five years of follow-up. J Bone Joint Surg Am. 2012;94:145–50. doi:10.​2106/​JBJS.​J.​00699.CrossRefPubMed
7.
Zurück zum Zitat Gillespie R, Lyons R, Lazarus M. Eccentric reaming in total shoulder arthroplasty: a cadaveric study. Orthopedics. 2009;32(1):21.CrossRefPubMed Gillespie R, Lyons R, Lazarus M. Eccentric reaming in total shoulder arthroplasty: a cadaveric study. Orthopedics. 2009;32(1):21.CrossRefPubMed
10.
Zurück zum Zitat Verstraeten TR, Deschepper E, Jacxsens M, Walravens S, De Coninck B, De Wilde LF. Operative guidelines for the reconstruction of the native glenoid plane: an anatomic three-dimensional computed tomography-scan reconstruction study. J Shoulder Elbow Surg. 2012;21:1565–72. doi:10.1016/j.jse.2011.10.030.CrossRefPubMed Verstraeten TR, Deschepper E, Jacxsens M, Walravens S, De Coninck B, De Wilde LF. Operative guidelines for the reconstruction of the native glenoid plane: an anatomic three-dimensional computed tomography-scan reconstruction study. J Shoulder Elbow Surg. 2012;21:1565–72. doi:10.​1016/​j.​jse.​2011.​10.​030.CrossRefPubMed
11.
Zurück zum Zitat Friedman RJ, Hawthorne KB, Genez BM. The use of computerized tomography in the measurement of glenoid version. J Bone Joint Surg Am. 1992;74A:1032–37. Friedman RJ, Hawthorne KB, Genez BM. The use of computerized tomography in the measurement of glenoid version. J Bone Joint Surg Am. 1992;74A:1032–37.
13.
Zurück zum Zitat Madsen DA. Geometric dimensioning and tolerancing: basic fundamentals: 392 Goodheart-Willcox Co.; 1977. (ISBN No. 9780870062315). Madsen DA. Geometric dimensioning and tolerancing: basic fundamentals: 392 Goodheart-Willcox Co.; 1977. (ISBN No. 9780870062315).
16.
Zurück zum Zitat Verstraeten TR, Deschepper E, Jacxsens M, Walravens S, De Coninck B, Pouliart N, et al. Determination of a reference system for the three-dimensional study of the glenohumeral relationship. Skeletal Radiol. 2013;42:1061–71. doi:10.1007/s00256-013-1572-0.CrossRefPubMed Verstraeten TR, Deschepper E, Jacxsens M, Walravens S, De Coninck B, Pouliart N, et al. Determination of a reference system for the three-dimensional study of the glenohumeral relationship. Skeletal Radiol. 2013;42:1061–71. doi:10.​1007/​s00256-013-1572-0.CrossRefPubMed
17.
Zurück zum Zitat Favard L, Katz D, Colmar M, Benkalfate T, Thomazeau H, Emily S. Total shoulder arthroplasty—arthroplasty for glenohumeral arthropathies: results and complications after a minimum follow-up of 8 years according to the type of arthroplasty and etiology. Orthop Traumatol Surg Res. 2012;98:S41–7. doi:10.1016/j.otsr.2012.04.003.CrossRefPubMed Favard L, Katz D, Colmar M, Benkalfate T, Thomazeau H, Emily S. Total shoulder arthroplasty—arthroplasty for glenohumeral arthropathies: results and complications after a minimum follow-up of 8 years according to the type of arthroplasty and etiology. Orthop Traumatol Surg Res. 2012;98:S41–7. doi:10.​1016/​j.​otsr.​2012.​04.​003.CrossRefPubMed
19.
Zurück zum Zitat Norris TR, Iannotti JP. Functional outcome after shoulder arthroplasty for primary osteoarthritis: a multicenter study. J Shoulder Elbow Surg. 2002;11:130–5.CrossRefPubMed Norris TR, Iannotti JP. Functional outcome after shoulder arthroplasty for primary osteoarthritis: a multicenter study. J Shoulder Elbow Surg. 2002;11:130–5.CrossRefPubMed
21.
Zurück zum Zitat Collins D, Tencer A, Sidles J, Matsen F. Edge displacement and deformation of glenoid components in response to eccentric loading—the effect of preparation of the glenoid bone. J Bone Joint Surg Am. 1992;74A:501–07. Collins D, Tencer A, Sidles J, Matsen F. Edge displacement and deformation of glenoid components in response to eccentric loading—the effect of preparation of the glenoid bone. J Bone Joint Surg Am. 1992;74A:501–07.
24.
27.
Zurück zum Zitat Nowak DD, Bahu MJ, Gardner TR, Dyrszka MD, Levine WN, Bigliani LU, et al. Simulation of surgical glenoid resurfacing using three-dimensional computed tomography of the arthritic glenohumeral joint: the amount of glenoid retroversion that can be corrected. J Shoulder Elbow Surg. 2009;18:680–88. doi:10.1016/j.jse.2009.03.019.CrossRefPubMed Nowak DD, Bahu MJ, Gardner TR, Dyrszka MD, Levine WN, Bigliani LU, et al. Simulation of surgical glenoid resurfacing using three-dimensional computed tomography of the arthritic glenohumeral joint: the amount of glenoid retroversion that can be corrected. J Shoulder Elbow Surg. 2009;18:680–88. doi:10.​1016/​j.​jse.​2009.​03.​019.CrossRefPubMed
29.
Zurück zum Zitat Mizuno N, Denard PJ, Raiss P, Walch G. Reverse total shoulder arthroplasty for primary glenohumeral osteoarthritis in patients with a biconcave glenoid. J Bone Joint Surg Am. 2013;95(14):1297–304. doi:10.2106/JBJS.L.00820.CrossRefPubMed Mizuno N, Denard PJ, Raiss P, Walch G. Reverse total shoulder arthroplasty for primary glenohumeral osteoarthritis in patients with a biconcave glenoid. J Bone Joint Surg Am. 2013;95(14):1297–304. doi:10.​2106/​JBJS.​L.​00820.CrossRefPubMed
32.
Zurück zum Zitat Hendel MD, Bryan JA, Barsoum WK, Rodriguez EJ, Brems JJ, Evans PJ, et al. Comparison of patient-specific instruments with standard surgical instruments in determining glenoid component position: a randomized prospective clinical trial. J Bone Joint Surg Am. 2012;94A:2167–75. doi:10.2106/Jbjs.K.01209.CrossRef Hendel MD, Bryan JA, Barsoum WK, Rodriguez EJ, Brems JJ, Evans PJ, et al. Comparison of patient-specific instruments with standard surgical instruments in determining glenoid component position: a randomized prospective clinical trial. J Bone Joint Surg Am. 2012;94A:2167–75. doi:10.​2106/​Jbjs.​K.​01209.CrossRef
33.
Zurück zum Zitat Kircher J, Wiedemann M, Magosch P, Lichtenberg S, Habermeyer P. Improved accuracy of glenoid positioning in total shoulder arthroplasty with intraoperative navigation: a prospective-randomized clinical study. J Shoulder Elbow Surg. 2009;18:515–20. doi:10.1016/j.jse.2009.03.014.CrossRefPubMed Kircher J, Wiedemann M, Magosch P, Lichtenberg S, Habermeyer P. Improved accuracy of glenoid positioning in total shoulder arthroplasty with intraoperative navigation: a prospective-randomized clinical study. J Shoulder Elbow Surg. 2009;18:515–20. doi:10.​1016/​j.​jse.​2009.​03.​014.CrossRefPubMed
34.
Zurück zum Zitat Nguyen D, Ferreira LM, Brownhill JR, King GJ, Drosdowech DS, Faber KJ, et al. Improved accuracy of computer assisted glenoid implantation in total shoulder arthroplasty: an in-vitro randomized controlled trial. J Shoulder Elbow Surg. 2009;18:907–14. doi:10.1016/j.jse.2009.02.022.CrossRefPubMed Nguyen D, Ferreira LM, Brownhill JR, King GJ, Drosdowech DS, Faber KJ, et al. Improved accuracy of computer assisted glenoid implantation in total shoulder arthroplasty: an in-vitro randomized controlled trial. J Shoulder Elbow Surg. 2009;18:907–14. doi:10.​1016/​j.​jse.​2009.​02.​022.CrossRefPubMed
35.
Zurück zum Zitat Suero EM, Citak M, Lo D, Krych AJ, Craig EV, Pearle AD. Use of a custom alignment guide to improve glenoid component position in total shoulder arthroplasty. Knee Surg Sports Traumatol Arthrosc. 2012;21(12):2860–6. doi:10.1007/s00167-012-2177-1.CrossRefPubMed Suero EM, Citak M, Lo D, Krych AJ, Craig EV, Pearle AD. Use of a custom alignment guide to improve glenoid component position in total shoulder arthroplasty. Knee Surg Sports Traumatol Arthrosc. 2012;21(12):2860–6. doi:10.​1007/​s00167-012-2177-1.CrossRefPubMed
36.
Metadaten
Titel
Consequences of reaming with flat and convex reamers for bone volume and surface area of the glenoid; a basic science study
verfasst von
Anne Karelse
Steven Leuridan
Alexander Van Tongel
Philippe Debeer
Jos Van Der Sloten
Kathleen Denis
Lieven F. De Wilde
Publikationsdatum
01.12.2015
Verlag
BioMed Central
Erschienen in
Journal of Orthopaedic Surgery and Research / Ausgabe 1/2015
Elektronische ISSN: 1749-799X
DOI
https://doi.org/10.1186/s13018-015-0312-7

Weitere Artikel der Ausgabe 1/2015

Journal of Orthopaedic Surgery and Research 1/2015 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Arthroskopie kann Knieprothese nicht hinauszögern

25.04.2024 Gonarthrose Nachrichten

Ein arthroskopischer Eingriff bei Kniearthrose macht im Hinblick darauf, ob und wann ein Gelenkersatz fällig wird, offenbar keinen Unterschied.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.