Skip to main content
Erschienen in: Clinical Oral Investigations 9/2018

17.02.2018 | Original Article

Cyclic fatigue resistance of R-Pilot, WaveOne Gold Glider, and ProGlider glide path instruments

verfasst von: Cangül Keskin, Uğur İnan, Murat Demiral, Ali Keleş

Erschienen in: Clinical Oral Investigations | Ausgabe 9/2018

Einloggen, um Zugang zu erhalten

Abstract

Objective

The aim of the present study was to compare the cyclic fatigue resistance of R-Pilot (VDW; Munich, Germany) with ProGlider (Denstply Sirona; Ballaigues, Switzerland) and WaveOne Gold Glider (Denstply Sirona; Ballaigues, Switzerland) glide path instruments.

Materials and methods

R-Pilot, ProGlider, and WaveOne Gold Glider instruments were collected (n = 15) and tested in a dynamic cyclic fatigue test device, which has an artificial canal with 60° angle of curvature and a 5-mm radius of curvature. All instruments were operated until fracture occurred, and both time to fracture (TF) and the lengths of the fractured fragments were recorded. Mean and standard deviations of TF and fragment length were calculated for each reciprocating system. TF data and fractured fragment length data were subjected to one-way ANOVA and post-hoc Tukey tests (P < 0.05). Also a Weibull analysis was performed on TF data.

Results

The cyclic fatigue resistance values of the WaveOne Gold Glider and R-Pilot were significantly higher than those of the ProGlider (P < 0.05), with no significant difference between them (P > 0.05). Weibull analysis revealed that WaveOne Gold Glider showed the highest predicted TF value for 99% survival rate, which was followed by R-Pilot and ProGlider. Regarding the length of the fractured tips, there were no significant differences among the instruments (P > 0.05).

Conclusions

The reciprocating WaveOne Gold Glider and R-Pilot instruments had significantly higher cyclic fatigue resistance than rotary ProGlider instruments.

Clinical relevance

This study reported that novel reciprocating glide path instruments exhibited higher cyclic fatigue resistance than rotating glide path instrument.
Literatur
1.
Zurück zum Zitat Patiño PV, Biedma BM, Liébana CR et al (2005) The influence of a manual glide path on the separation rate of NiTi rotary instruments. J Endod 31:114–116CrossRef Patiño PV, Biedma BM, Liébana CR et al (2005) The influence of a manual glide path on the separation rate of NiTi rotary instruments. J Endod 31:114–116CrossRef
2.
Zurück zum Zitat Berutti E, Negro AR, Lendini M, Pasqualini D (2004) Influence of manual preflaring and torque on the failure rate of ProTaper rotary instruments. J Endod 30:228–230CrossRef Berutti E, Negro AR, Lendini M, Pasqualini D (2004) Influence of manual preflaring and torque on the failure rate of ProTaper rotary instruments. J Endod 30:228–230CrossRef
3.
Zurück zum Zitat Berutti E, Cantatore G, Castellucci A, Chiandussi G, Pera F, Migliaretti G, Pasqualini D (2009) Use of nickel-titanium rotary PathFile to create the glide path: comparison with manual preflaring in simulated root canals. J Endod 35:408–412CrossRef Berutti E, Cantatore G, Castellucci A, Chiandussi G, Pera F, Migliaretti G, Pasqualini D (2009) Use of nickel-titanium rotary PathFile to create the glide path: comparison with manual preflaring in simulated root canals. J Endod 35:408–412CrossRef
4.
Zurück zum Zitat Nakagawa R, Alves J, Buono V, Bahia M (2014) Flexibility and torsional behaviour of rotary nickel-titanium PathFile, RaCe ISO 10, Scout RaCe and stainless steel K-File hand instruments. Int Endod J 47:290–297CrossRef Nakagawa R, Alves J, Buono V, Bahia M (2014) Flexibility and torsional behaviour of rotary nickel-titanium PathFile, RaCe ISO 10, Scout RaCe and stainless steel K-File hand instruments. Int Endod J 47:290–297CrossRef
5.
Zurück zum Zitat Berutti E, Angelini E, Rigolone M, Migliaretti G, Pasqualini D (2006) Influence of sodium hypochlorite on fracture properties and corrosion of ProTaper Rotary instruments. Int Endod J 39:693–699CrossRef Berutti E, Angelini E, Rigolone M, Migliaretti G, Pasqualini D (2006) Influence of sodium hypochlorite on fracture properties and corrosion of ProTaper Rotary instruments. Int Endod J 39:693–699CrossRef
6.
Zurück zum Zitat Walsch H (2004) The hybrid concept of nickel–titanium rotary instrumentation. Dent Clin N Am 48:183–202CrossRef Walsch H (2004) The hybrid concept of nickel–titanium rotary instrumentation. Dent Clin N Am 48:183–202CrossRef
7.
Zurück zum Zitat Gianluca G (2005) The K3 rotary nickel titanium instrument system. Endod Topics 10:179–182CrossRef Gianluca G (2005) The K3 rotary nickel titanium instrument system. Endod Topics 10:179–182CrossRef
8.
Zurück zum Zitat Paleker F, van der Vyver PJ (2017) Glide path enlargement of mandibular molar canals by using K-files, the ProGlider file, and G-files: a comparative study of the preparation times. J Endod 43:609–612CrossRef Paleker F, van der Vyver PJ (2017) Glide path enlargement of mandibular molar canals by using K-files, the ProGlider file, and G-files: a comparative study of the preparation times. J Endod 43:609–612CrossRef
9.
Zurück zum Zitat Walia H, Brantley WA, Gerstein H (1988) An initial investigation of the bending and torsional properties of Nitinol root canal files. J Endod 14:346–351CrossRef Walia H, Brantley WA, Gerstein H (1988) An initial investigation of the bending and torsional properties of Nitinol root canal files. J Endod 14:346–351CrossRef
10.
Zurück zum Zitat Shen Y, Cheung GS, Bian Z, Peng B (2006) Comparison of defects in ProFile and ProTaper systems after clinical use. J Endod 32:61–65CrossRef Shen Y, Cheung GS, Bian Z, Peng B (2006) Comparison of defects in ProFile and ProTaper systems after clinical use. J Endod 32:61–65CrossRef
11.
Zurück zum Zitat Cheung GS (2007) Instrument fracture: mechanisms, removal of fragments, and clinical outcomes. Endod Topics 16:1–26CrossRef Cheung GS (2007) Instrument fracture: mechanisms, removal of fragments, and clinical outcomes. Endod Topics 16:1–26CrossRef
12.
Zurück zum Zitat Kitchens GG, Liewehr FR, Moon PC (2007) The effect of operational speed on the fracture of nickel-titanium rotary instruments. J Endod 33:52–54CrossRef Kitchens GG, Liewehr FR, Moon PC (2007) The effect of operational speed on the fracture of nickel-titanium rotary instruments. J Endod 33:52–54CrossRef
13.
Zurück zum Zitat Bui TB, Mitchell JC, Baumgartner JC (2008) Effect of electropolishing ProFile nickel–titanium rotary instruments on cyclic fatigue resistance, torsional resistance, and cutting efficiency. J Endod 34:190–193CrossRef Bui TB, Mitchell JC, Baumgartner JC (2008) Effect of electropolishing ProFile nickel–titanium rotary instruments on cyclic fatigue resistance, torsional resistance, and cutting efficiency. J Endod 34:190–193CrossRef
14.
Zurück zum Zitat Ounsi HF, Al-Shalan T, Salameh Z et al (2008) Quantitative and qualitative elemental analysis of different nickel–titanium rotary instruments by using scanning electron microscopy and energy dispersive spectroscopy. J Endod 34:53–55CrossRef Ounsi HF, Al-Shalan T, Salameh Z et al (2008) Quantitative and qualitative elemental analysis of different nickel–titanium rotary instruments by using scanning electron microscopy and energy dispersive spectroscopy. J Endod 34:53–55CrossRef
15.
Zurück zum Zitat Elnaghy AM, Elsaka SE (2015) Evaluation of the mechanical behaviour of PathFile and ProGlider pathfinding nickel-titanium rotary instruments. Int Endod J 48:894–901CrossRef Elnaghy AM, Elsaka SE (2015) Evaluation of the mechanical behaviour of PathFile and ProGlider pathfinding nickel-titanium rotary instruments. Int Endod J 48:894–901CrossRef
16.
Zurück zum Zitat Yared G (2008) Canal preparation using only one Ni-Ti rotary instrument: preliminary observations. Int Endod J 41:339–344CrossRef Yared G (2008) Canal preparation using only one Ni-Ti rotary instrument: preliminary observations. Int Endod J 41:339–344CrossRef
17.
Zurück zum Zitat Plotino G, Ahmed HMA, Grande NM, Cohen S, Bukiet F (2015) Current assessment of reciprocation in endodontic preparation: a comprehensive review—part II: properties and effectiveness. J Endod 41:1939–1950CrossRef Plotino G, Ahmed HMA, Grande NM, Cohen S, Bukiet F (2015) Current assessment of reciprocation in endodontic preparation: a comprehensive review—part II: properties and effectiveness. J Endod 41:1939–1950CrossRef
20.
Zurück zum Zitat Pruett JP, Clement DJ, Carnes DL (1997) Cyclic fatigue testing of nickel-titanium endodontic instruments. J Endod 23:77–85CrossRef Pruett JP, Clement DJ, Carnes DL (1997) Cyclic fatigue testing of nickel-titanium endodontic instruments. J Endod 23:77–85CrossRef
21.
Zurück zum Zitat Weibull W (1951) A statistical distribution function of wide applicability. J Appl Mech 128:293–297 Weibull W (1951) A statistical distribution function of wide applicability. J Appl Mech 128:293–297
22.
Zurück zum Zitat Nguyen HH, Fong H, Paranjpe A, Flake NM, Johnson JD, Peters OA (2014) Evaluation of the resistance to cyclic fatigue between ProTaper Next, ProTaper Universal, and Vortex Blue rotary instruments. J Endod 40:1190–1994CrossRef Nguyen HH, Fong H, Paranjpe A, Flake NM, Johnson JD, Peters OA (2014) Evaluation of the resistance to cyclic fatigue between ProTaper Next, ProTaper Universal, and Vortex Blue rotary instruments. J Endod 40:1190–1994CrossRef
23.
Zurück zum Zitat Castelló-Escrivá R, Alegre-Domingo T, Faus-Matoses V, Román-Richon S, Faus-Llácer VJ (2012) In vitro comparison of cyclic fatigue resistance of ProTaper, WaveOne, and Twisted Files. J Endod 38:1521–1524CrossRef Castelló-Escrivá R, Alegre-Domingo T, Faus-Matoses V, Román-Richon S, Faus-Llácer VJ (2012) In vitro comparison of cyclic fatigue resistance of ProTaper, WaveOne, and Twisted Files. J Endod 38:1521–1524CrossRef
24.
Zurück zum Zitat De-Deus G, Moreira E, Lopes H, Elias C (2010) Extended cyclic fatigue life of F2 ProTaper instruments used in reciprocating movement. Int Endod J 43:1063–1068CrossRef De-Deus G, Moreira E, Lopes H, Elias C (2010) Extended cyclic fatigue life of F2 ProTaper instruments used in reciprocating movement. Int Endod J 43:1063–1068CrossRef
25.
Zurück zum Zitat Lopes HP, Elias CN, Vieira MV et al (2013) Fatigue life of Reciproc and Mtwo instruments subjected to static and dynamic tests. J Endod 39:693–696CrossRef Lopes HP, Elias CN, Vieira MV et al (2013) Fatigue life of Reciproc and Mtwo instruments subjected to static and dynamic tests. J Endod 39:693–696CrossRef
26.
Zurück zum Zitat Gao Y, Gutmann JL, Wilkinson K, Maxwell R, Ammon D (2012) Evaluation of the impact of raw materials on the fatigue and mechanical properties of ProFile Vortex rotary instruments. J Endod 38:398–401CrossRef Gao Y, Gutmann JL, Wilkinson K, Maxwell R, Ammon D (2012) Evaluation of the impact of raw materials on the fatigue and mechanical properties of ProFile Vortex rotary instruments. J Endod 38:398–401CrossRef
27.
Zurück zum Zitat Frick CP, Ortega AM, Tyber J, Maksound AEM, Maier HJ, Liu Y, Gall K (2005) Thermal processing of polycrystalline NiTi shape memory alloys. Mater Sci Eng A 405:34–49CrossRef Frick CP, Ortega AM, Tyber J, Maksound AEM, Maier HJ, Liu Y, Gall K (2005) Thermal processing of polycrystalline NiTi shape memory alloys. Mater Sci Eng A 405:34–49CrossRef
28.
Zurück zum Zitat Versluis A, Kim HC, Lee W, Kim BM, Lee CJ (2012) Flexural stiffness and stresses in nickel-titanium rotary files for various pitch and cross-sectional geometries. J Endod 38:1399–1403CrossRef Versluis A, Kim HC, Lee W, Kim BM, Lee CJ (2012) Flexural stiffness and stresses in nickel-titanium rotary files for various pitch and cross-sectional geometries. J Endod 38:1399–1403CrossRef
29.
Zurück zum Zitat He R, Ni J (2010) Design improvement and failure reduction of endodontic files through finite element analysis: application to V-Taper file designs. J Endod 36:1552–1557CrossRef He R, Ni J (2010) Design improvement and failure reduction of endodontic files through finite element analysis: application to V-Taper file designs. J Endod 36:1552–1557CrossRef
30.
Zurück zum Zitat Adıgüzel M, Capar ID (2017) Comparison of cyclic fatigue resistance of WaveOne and WaveOne Gold Small, Primary, and Large instruments. J Endod 43:623–627CrossRef Adıgüzel M, Capar ID (2017) Comparison of cyclic fatigue resistance of WaveOne and WaveOne Gold Small, Primary, and Large instruments. J Endod 43:623–627CrossRef
31.
Zurück zum Zitat Keskin C, Inan U, Demiral M, Keleş A (2017) Cyclic fatigue resistance of Reciproc Blue, Reciproc and WaveOne Gold reciprocating instruments. J Endod 43:1360–1363CrossRef Keskin C, Inan U, Demiral M, Keleş A (2017) Cyclic fatigue resistance of Reciproc Blue, Reciproc and WaveOne Gold reciprocating instruments. J Endod 43:1360–1363CrossRef
32.
Zurück zum Zitat Hieawy A, Haapasalo M, Zhou H, Wang ZJ, Shen Y (2015) Phase transformation behavior and resistance to bending and cyclic fatigue of ProTaper Gold and ProTaper Universal instruments. J Endod 41:1134–1138CrossRef Hieawy A, Haapasalo M, Zhou H, Wang ZJ, Shen Y (2015) Phase transformation behavior and resistance to bending and cyclic fatigue of ProTaper Gold and ProTaper Universal instruments. J Endod 41:1134–1138CrossRef
34.
Zurück zum Zitat Capar ID, Kaval ME, Ertas H, Sen BH (2015) Comparison of the cyclic fatigue resistance of 5 different rotary pathfinding instruments made of conventional nickel-titanium wire, M-wire, and controlled memory wire. J Endod 41:535–538CrossRef Capar ID, Kaval ME, Ertas H, Sen BH (2015) Comparison of the cyclic fatigue resistance of 5 different rotary pathfinding instruments made of conventional nickel-titanium wire, M-wire, and controlled memory wire. J Endod 41:535–538CrossRef
35.
Zurück zum Zitat Tokita D, Ebihara A, Miyara K, Okiji T (2017) Dynamic torsional and cyclic fracture behavior of ProFile rotary instruments at continuous or reciprocating rotation as visualized with high-speed digital video imaging. J Endod 43:1337–1342CrossRef Tokita D, Ebihara A, Miyara K, Okiji T (2017) Dynamic torsional and cyclic fracture behavior of ProFile rotary instruments at continuous or reciprocating rotation as visualized with high-speed digital video imaging. J Endod 43:1337–1342CrossRef
36.
Zurück zum Zitat Klocke A, Kahl-Nieke B (2005) Influence of force location in orthodontic shear bond strength testing. Dent Mater 21:391–396CrossRef Klocke A, Kahl-Nieke B (2005) Influence of force location in orthodontic shear bond strength testing. Dent Mater 21:391–396CrossRef
Metadaten
Titel
Cyclic fatigue resistance of R-Pilot, WaveOne Gold Glider, and ProGlider glide path instruments
verfasst von
Cangül Keskin
Uğur İnan
Murat Demiral
Ali Keleş
Publikationsdatum
17.02.2018
Verlag
Springer Berlin Heidelberg
Erschienen in
Clinical Oral Investigations / Ausgabe 9/2018
Print ISSN: 1432-6981
Elektronische ISSN: 1436-3771
DOI
https://doi.org/10.1007/s00784-018-2391-4

Weitere Artikel der Ausgabe 9/2018

Clinical Oral Investigations 9/2018 Zur Ausgabe

Parodontalbehandlung verbessert Prognose bei Katheterablation

19.04.2024 Vorhofflimmern Nachrichten

Werden Personen mit Vorhofflimmern in der Blanking-Periode nach einer Katheterablation gegen eine bestehende Parodontitis behandelt, verbessert dies die Erfolgsaussichten. Dafür sprechen die Resultate einer prospektiven Untersuchung.

Invasive Zahnbehandlung: Wann eine Antibiotikaprophylaxe vor infektiöser Endokarditis schützt

11.04.2024 Endokarditis Nachrichten

Bei welchen Personen eine Antibiotikaprophylaxe zur Prävention einer infektiösen Endokarditis nach invasiven zahnärztlichen Eingriffen sinnvoll ist, wird diskutiert. Neue Daten stehen im Einklang mit den europäischen Leitlinienempfehlungen.

Zell-Organisatoren unter Druck: Mechanismen des embryonalen Zahnwachstums aufgedeckt

08.04.2024 Zahnmedizin Nachrichten

Der Aufbau von Geweben und Organen während der Embryonalentwicklung wird von den Zellen bemerkenswert choreografiert. Für diesen Prozess braucht es spezielle sogenannte „Organisatoren“. In einer aktuellen Veröffentlichung im Fachjournal Nature Cell Biology berichten Forschende durch welchen Vorgang diese Organisatoren im Gewebe entstehen und wie sie dann die Bildung von Zähnen orchestrieren.

Die Oralprophylaxe & Kinderzahnheilkunde umbenannt

11.03.2024 Kinderzahnmedizin Nachrichten

Infolge der Umbenennung der Deutschen Gesellschaft für Kinderzahnheilkunde in Deutsche Gesellschaft für Kinderzahnmedizin (DGKiZ) wird deren Mitgliederzeitschrift Oralprophylaxe & Kinderzahnheilkunde in Oralprophylaxe & Kinderzahnmedizin umbenannt. Aus diesem Grunde trägt die erste Ausgabe in 2024 erstmalig den neuen Titel.

Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Zahnmedizin und bleiben Sie gut informiert – ganz bequem per eMail.