Delta Ceramic-on-Alumina Ceramic Articulation in Primary THA: Prospective, Randomized FDA-IDE Study and Retrieval Analysis

verfasst von: Adolph V. Lombardi Jr., MD, Keith R. Berend, MD, Brian E. Seng, DO, Ian C. Clarke, PhD, Joanne B. Adams, BFA

Erschienen in: Clinical Orthopaedics and Related Research® | Ausgabe 2/2010

Einloggen, um Zugang zu erhalten

Abstract

Wear and osteolysis continue to be major reasons for revision surgery in THA. Ceramic-on-ceramic bearings eliminate polyethylene wear debris. The newest generation of these bearings incorporate nanosized, yttria-stabilized tetragonal zirconia particles producing an alumina matrix composite. We asked whether this new material would perform as well as a conventional bearing in terms of functional hip scores, radiographic migration and osteolysis, complications and survival. As part of a US FDA investigational device exemption study (G000075), we conducted an initial prospective safety study of 21 alumina matrix composite femoral heads articulating on alumina liners followed by a prospective, randomized study with 44 more of these articulations and 45 zirconia femoral heads on polyethylene liners. The minimum followup for all patients was 26 months (mean, 73 months; range, 26–108 months). Harris hip scores and radiographic findings were similar in the two groups as was survivorship (trial 95% versus control 93%). There were three reoperations in the trial group and three in the control group. A fractured head retrieval showed a 33% monoclinic transformation with an increase in surface roughness from 3 to 5 nm at the main wear zone. While our numbers were insufficient to compare device-related complications, the trial device performed as well as the control device in terms of reoperation, and clinical and radiographic outcome. The alumina matrix composite femoral head on an alumina liner provided high survivorship.
Level of Evidence: Level II, therapeutic study. See Guidelines for Authors for a complete description of levels of evidence.
Literatur
1.
Zurück zum Zitat Akagi M, Nonaka T, Nishisaka F, Mori S, Fukuda K, Hamanishi C. Late dissociation of an alumina-on-alumina bearing modular acetabular component. J Arthroplasty. 2004;19:647–651.CrossRefPubMed Akagi M, Nonaka T, Nishisaka F, Mori S, Fukuda K, Hamanishi C. Late dissociation of an alumina-on-alumina bearing modular acetabular component. J Arthroplasty. 2004;19:647–651.CrossRefPubMed
2.
Zurück zum Zitat Allain J, Roudot-Thoraval F, Delecrin J, Anract P, Migaud H, Goutallier D. Revision total hip arthroplasty performed after fracture of a ceramic femoral head. J Bone Joint Surg Am. 2003;85:825–830.PubMed Allain J, Roudot-Thoraval F, Delecrin J, Anract P, Migaud H, Goutallier D. Revision total hip arthroplasty performed after fracture of a ceramic femoral head. J Bone Joint Surg Am. 2003;85:825–830.PubMed
3.
Zurück zum Zitat Bierbaum BE, Mairus J, Kuesis D, Morrison JC, Ward D. Ceramic-on-ceramic bearings in total hip arthroplasty. Clin Orthop Relat Res. 2002;405:158–163.CrossRefPubMed Bierbaum BE, Mairus J, Kuesis D, Morrison JC, Ward D. Ceramic-on-ceramic bearings in total hip arthroplasty. Clin Orthop Relat Res. 2002;405:158–163.CrossRefPubMed
4.
Zurück zum Zitat Blackwelder WC. “Proving the null hypothesis” in clinical trials. Control Clin Trials. 1982;3:345–353.CrossRefPubMed Blackwelder WC. “Proving the null hypothesis” in clinical trials. Control Clin Trials. 1982;3:345–353.CrossRefPubMed
5.
Zurück zum Zitat Clarke IC, Manaka M, Green DD, Williams P, Pezzotti G, Kim YH, Ries M, Sugano N, Sedel L, Delauney C, Nissan BB, Donaldson T, Gustafson GA. Current status of zirconia used in total hip implants. J Bone Joint Surg Am. 2003;85:73–84.PubMed Clarke IC, Manaka M, Green DD, Williams P, Pezzotti G, Kim YH, Ries M, Sugano N, Sedel L, Delauney C, Nissan BB, Donaldson T, Gustafson GA. Current status of zirconia used in total hip implants. J Bone Joint Surg Am. 2003;85:73–84.PubMed
6.
Zurück zum Zitat D’Antonio J, Capello W, Manley M, Naughton M, Sutton K. Alumina ceramic bearings for total hip arthroplasty: five-year results of a prospective randomized study. Clin Orthop Relat Res. 2005;436:164–171.CrossRefPubMed D’Antonio J, Capello W, Manley M, Naughton M, Sutton K. Alumina ceramic bearings for total hip arthroplasty: five-year results of a prospective randomized study. Clin Orthop Relat Res. 2005;436:164–171.CrossRefPubMed
7.
Zurück zum Zitat Dearnley PA. A review of metallic, ceramic and surface-treated metals used for bearing surfaces in human joint replacement. Proc Inst Mech Eng H. 1999;213:107–135.CrossRefPubMed Dearnley PA. A review of metallic, ceramic and surface-treated metals used for bearing surfaces in human joint replacement. Proc Inst Mech Eng H. 1999;213:107–135.CrossRefPubMed
8.
Zurück zum Zitat De Aza AH, Chevalier J, Fantozzi G, Schehl M, Torrecillas R. Crack growth resistance of alumina, zirconia and zirconia toughened alumina ceramics for joint prostheses. Biomaterials. 2002;23:937–945.CrossRefPubMed De Aza AH, Chevalier J, Fantozzi G, Schehl M, Torrecillas R. Crack growth resistance of alumina, zirconia and zirconia toughened alumina ceramics for joint prostheses. Biomaterials. 2002;23:937–945.CrossRefPubMed
9.
Zurück zum Zitat DeLee JG, Charnley J. Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res. 1976;121:20–32.PubMed DeLee JG, Charnley J. Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res. 1976;121:20–32.PubMed
10.
Zurück zum Zitat Frndak PA, Mallory TH, Lombardi AV Jr. Translateral surgical approach to the hip: the abductor muscle “split.” Clin Orthop Relat Res. 1993;295:135–141.PubMed Frndak PA, Mallory TH, Lombardi AV Jr. Translateral surgical approach to the hip: the abductor muscle “split.” Clin Orthop Relat Res. 1993;295:135–141.PubMed
11.
Zurück zum Zitat Gruen TA, McNeice GM, Amstutz HC. “Modes of failure” of cemented stem-type femoral components: a radiographic analysis of loosening. Clin Orthop Relat Res. 1979;141:17–27.PubMed Gruen TA, McNeice GM, Amstutz HC. “Modes of failure” of cemented stem-type femoral components: a radiographic analysis of loosening. Clin Orthop Relat Res. 1979;141:17–27.PubMed
12.
Zurück zum Zitat Hamadouche M, Boutin P, Daussange J, Bolander ME, Sedel L. Alumina-on-alumina total hip arthroplasty: a minimum 18.5 year follow-up study. J Bone Joint Surg Am. 2002;84:69–77.PubMed Hamadouche M, Boutin P, Daussange J, Bolander ME, Sedel L. Alumina-on-alumina total hip arthroplasty: a minimum 18.5 year follow-up study. J Bone Joint Surg Am. 2002;84:69–77.PubMed
13.
Zurück zum Zitat Haraguchi K, Sugano N, Nishii T, Miki H, Oka K, Yoshikawa H. Phase transformation of a zirconia ceramic head after total hip arthroplasty. J Bone Joint Surg Br. 2001;83:996–1000.CrossRefPubMed Haraguchi K, Sugano N, Nishii T, Miki H, Oka K, Yoshikawa H. Phase transformation of a zirconia ceramic head after total hip arthroplasty. J Bone Joint Surg Br. 2001;83:996–1000.CrossRefPubMed
14.
Zurück zum Zitat Harris WH. Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty. An end-result study using a new method of result evaluation. J Bone Joint Surg Am. 1969;51:737–755.PubMed Harris WH. Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty. An end-result study using a new method of result evaluation. J Bone Joint Surg Am. 1969;51:737–755.PubMed
15.
Zurück zum Zitat Hasegawa M, Sudo A, Hirata H, Uchida A. Ceramic acetabular liner fracture in total hip arthroplasty with a ceramic sandwich cup. J Arthroplasty. 2003;18:658–661.CrossRefPubMed Hasegawa M, Sudo A, Hirata H, Uchida A. Ceramic acetabular liner fracture in total hip arthroplasty with a ceramic sandwich cup. J Arthroplasty. 2003;18:658–661.CrossRefPubMed
16.
Zurück zum Zitat Hasegawa M, Sudo A, Uchida A. Cobalt-chromium head wear following revision hip arthroplasty performed after ceramic fracture—a case report. Acta Orthop. 2006;77:833–835.CrossRefPubMed Hasegawa M, Sudo A, Uchida A. Cobalt-chromium head wear following revision hip arthroplasty performed after ceramic fracture—a case report. Acta Orthop. 2006;77:833–835.CrossRefPubMed
17.
Zurück zum Zitat Hasegawa M, Sudo A, Uchida A. Alumina ceramic-on-ceramic total hip replacement with a layered acetabular component. J Bone Joint Surg Br. 2006;88:877–882.CrossRefPubMed Hasegawa M, Sudo A, Uchida A. Alumina ceramic-on-ceramic total hip replacement with a layered acetabular component. J Bone Joint Surg Br. 2006;88:877–882.CrossRefPubMed
18.
Zurück zum Zitat Huo MH, Martin RP, Zatorski LE, Keggi KJ. Ceramic total hip replacements done without cement: an average 9-year follow-up study. Clin Orthop Relat Res. 1996;332:143–150.CrossRefPubMed Huo MH, Martin RP, Zatorski LE, Keggi KJ. Ceramic total hip replacements done without cement: an average 9-year follow-up study. Clin Orthop Relat Res. 1996;332:143–150.CrossRefPubMed
19.
Zurück zum Zitat Hwang SK, Oh JR, Her MS, Shim YJ, Cho TY, Kwon SM. Fracture-dissociation of ceramic liner. Orthopedics. 2008;31:804.CrossRefPubMed Hwang SK, Oh JR, Her MS, Shim YJ, Cho TY, Kwon SM. Fracture-dissociation of ceramic liner. Orthopedics. 2008;31:804.CrossRefPubMed
20.
Zurück zum Zitat Inzerillo CV, Garino JP. Alternative bearing surfaces in total hip arthroplasty. J Southern Orthop Assoc. 2003;12:106–111. Inzerillo CV, Garino JP. Alternative bearing surfaces in total hip arthroplasty. J Southern Orthop Assoc. 2003;12:106–111.
21.
Zurück zum Zitat Jazrawi LM, Kummer FJ, DiCesare PE. Alternative bearing surfaces for total joint arthroplasty. J Am Acad Orthop Surg. 1998;6:198–203.PubMed Jazrawi LM, Kummer FJ, DiCesare PE. Alternative bearing surfaces for total joint arthroplasty. J Am Acad Orthop Surg. 1998;6:198–203.PubMed
22.
Zurück zum Zitat Kaplan EL, Meier P. Nonparametric estimation from incomplete observations. J Am Statistical Assoc. 1958;53:457–481.CrossRef Kaplan EL, Meier P. Nonparametric estimation from incomplete observations. J Am Statistical Assoc. 1958;53:457–481.CrossRef
23.
Zurück zum Zitat Keurentjes JC, Kuipers RM, Wever DJ, Schreurs BW. High incidence of squeaking in THAs with alumina ceramic-on-ceramic bearings. Clin Orthop Relat Res. 2008;466:1438–1443.CrossRefPubMed Keurentjes JC, Kuipers RM, Wever DJ, Schreurs BW. High incidence of squeaking in THAs with alumina ceramic-on-ceramic bearings. Clin Orthop Relat Res. 2008;466:1438–1443.CrossRefPubMed
24.
Zurück zum Zitat Kraay MJ, Thomas RD, Rimnac CM, Fitzgerald SJ, Goldberg VM. Zirconia versus Co-Cr femoral heads in total hip arthroplasty. Clin Orthop Relat Res. 2006;453:86–90.CrossRefPubMed Kraay MJ, Thomas RD, Rimnac CM, Fitzgerald SJ, Goldberg VM. Zirconia versus Co-Cr femoral heads in total hip arthroplasty. Clin Orthop Relat Res. 2006;453:86–90.CrossRefPubMed
25.
Zurück zum Zitat Lusty PJ, Tai CC, Sew-Hoy RP, Walter WL, Walter WK, Zicat BA. Third-generation alumina-on-alumina ceramic bearings in cementless total hip arthroplasty. J Bone Joint Surg Am. 2007;89:2676–2683.CrossRefPubMed Lusty PJ, Tai CC, Sew-Hoy RP, Walter WL, Walter WK, Zicat BA. Third-generation alumina-on-alumina ceramic bearings in cementless total hip arthroplasty. J Bone Joint Surg Am. 2007;89:2676–2683.CrossRefPubMed
26.
Zurück zum Zitat Masonis JL, Bourne RB, Ries MD, McCalden RW, Salehi A, Kelman DC. Zirconia femoral head fracture: a clinical and retrieval analysis. J Arthroplasty. 2004;19:898–905.CrossRefPubMed Masonis JL, Bourne RB, Ries MD, McCalden RW, Salehi A, Kelman DC. Zirconia femoral head fracture: a clinical and retrieval analysis. J Arthroplasty. 2004;19:898–905.CrossRefPubMed
27.
Zurück zum Zitat Masson B. Emergence of the alumina matrix composite in total hip arthroplasty. Int Orthop. 2009;33:359–363.CrossRefPubMed Masson B. Emergence of the alumina matrix composite in total hip arthroplasty. Int Orthop. 2009;33:359–363.CrossRefPubMed
28.
Zurück zum Zitat Mehmood S, Riyaz JH, Pandit H. Review of ceramic-on-ceramic total hip arthroplasty. J Surg Ortho Adv. 2008;17:45–50. Mehmood S, Riyaz JH, Pandit H. Review of ceramic-on-ceramic total hip arthroplasty. J Surg Ortho Adv. 2008;17:45–50.
29.
Zurück zum Zitat Morita Y, Nakata K, Kim YH, Sekino T, Niihara K, Ikeuchi K. Wear properties of alumina/zirconia composite ceramics for joint prostheses measured with an end-face apparatus. Biomed Mater Eng. 2004;14:263–270.PubMed Morita Y, Nakata K, Kim YH, Sekino T, Niihara K, Ikeuchi K. Wear properties of alumina/zirconia composite ceramics for joint prostheses measured with an end-face apparatus. Biomed Mater Eng. 2004;14:263–270.PubMed
30.
Zurück zum Zitat Morlock M, Nassutt R, Janssen R, Willmann G, Honl M. Mismatched wear couple zirconium oxide and aluminum oxide in total hip arthroplasty. J Arthroplasty. 2001;16:1071–1074.CrossRefPubMed Morlock M, Nassutt R, Janssen R, Willmann G, Honl M. Mismatched wear couple zirconium oxide and aluminum oxide in total hip arthroplasty. J Arthroplasty. 2001;16:1071–1074.CrossRefPubMed
31.
Zurück zum Zitat Nevelos J, Ingham E, Doyle C, Streicher R, Nevelos A, Walter W, Fisher J, Eng D. Microseparation of the centers of alumina-alumina artificial hip joints during simulator testing produces clinically relevant wear rates and patterns. J Arthroplasty. 2000;15:793–795.CrossRefPubMed Nevelos J, Ingham E, Doyle C, Streicher R, Nevelos A, Walter W, Fisher J, Eng D. Microseparation of the centers of alumina-alumina artificial hip joints during simulator testing produces clinically relevant wear rates and patterns. J Arthroplasty. 2000;15:793–795.CrossRefPubMed
32.
Zurück zum Zitat Nizard R, Pourreyron D, Raould A, Hannouche D, Sedel L. Alumina-on-alumina hip arthroplasty in patients younger than 30 years old. Clin Orthop Relat Res. 2008;466:317–323.CrossRefPubMed Nizard R, Pourreyron D, Raould A, Hannouche D, Sedel L. Alumina-on-alumina hip arthroplasty in patients younger than 30 years old. Clin Orthop Relat Res. 2008;466:317–323.CrossRefPubMed
33.
Zurück zum Zitat Pulliam IT, Trousdale RT. Fracture of a ceramic femoral head after a revision operation. J Bone Joint Surg Am. 1997;79:118–121.PubMed Pulliam IT, Trousdale RT. Fracture of a ceramic femoral head after a revision operation. J Bone Joint Surg Am. 1997;79:118–121.PubMed
34.
Zurück zum Zitat Ravasi F, Sansone V. Five-year follow-up with a ceramic sandwich cup in total hip replacement. Arch Orthop Trauma Surg. 2002;122:350–353.PubMed Ravasi F, Sansone V. Five-year follow-up with a ceramic sandwich cup in total hip replacement. Arch Orthop Trauma Surg. 2002;122:350–353.PubMed
35.
Zurück zum Zitat Restrepo C, Parvizi J, Kurtz SM, Sharkey PF, Hozack WJ, Rothman RH. The noisy ceramic hip: is component malposition the cause? J Arthroplasty. 2008;23:643–649.CrossRefPubMed Restrepo C, Parvizi J, Kurtz SM, Sharkey PF, Hozack WJ, Rothman RH. The noisy ceramic hip: is component malposition the cause? J Arthroplasty. 2008;23:643–649.CrossRefPubMed
36.
Zurück zum Zitat Rhoads DP, Baker KC, Israel R, Greene PW. Fracture of an alumina femoral head used in ceramic-on-ceramic total hip arthroplasty. J Arthroplasty. 2008;23:1239.e25–e30. Rhoads DP, Baker KC, Israel R, Greene PW. Fracture of an alumina femoral head used in ceramic-on-ceramic total hip arthroplasty. J Arthroplasty. 2008;23:1239.e25–e30.
37.
Zurück zum Zitat Rodríguez JA, Gonzalez DelaValle A, McCook N. Squeaking in total hip replacement: a cause for concern. Orthopedics. 2008;31:874, 877–878. Rodríguez JA, Gonzalez DelaValle A, McCook N. Squeaking in total hip replacement: a cause for concern. Orthopedics. 2008;31:874, 877–878.
38.
Zurück zum Zitat Simon JA, Dayan AJ, Enrique E, Stuchin SA, Di Cesare PE. Catastrophic failure of the acetabular component in a ceramic-polyethylene bearing total hip arthroplasty. J Arthroplasty. 1998;13:108–113.CrossRefPubMed Simon JA, Dayan AJ, Enrique E, Stuchin SA, Di Cesare PE. Catastrophic failure of the acetabular component in a ceramic-polyethylene bearing total hip arthroplasty. J Arthroplasty. 1998;13:108–113.CrossRefPubMed
39.
Zurück zum Zitat Stewart TD, Tipper JL, Streicher RM, Ingham E, Fisher J. Long-term wear of HIPed alumina on alumina bearings for THR under microseparation conditions. J Mater Sci Mater Med. 2001;12:1053–1056.CrossRefPubMed Stewart TD, Tipper JL, Streicher RM, Ingham E, Fisher J. Long-term wear of HIPed alumina on alumina bearings for THR under microseparation conditions. J Mater Sci Mater Med. 2001;12:1053–1056.CrossRefPubMed
40.
Zurück zum Zitat Streicher RM, Semlitsch M, Schoen R, Weber H, Rieker C. Metal-on-metal articulation for artificial hip joints: laboratory study and clinical results. Proc Inst Mech Eng. 1996;210:223–232. Streicher RM, Semlitsch M, Schoen R, Weber H, Rieker C. Metal-on-metal articulation for artificial hip joints: laboratory study and clinical results. Proc Inst Mech Eng. 1996;210:223–232.
41.
Zurück zum Zitat Sychterz CJ, Engh CA, Swope SW, McNulty DE, Engh CA. Analysis of prosthetic femoral heads retrieved at autopsy. Clin Orthop Relat Res. 1999;358:223–234.CrossRefPubMed Sychterz CJ, Engh CA, Swope SW, McNulty DE, Engh CA. Analysis of prosthetic femoral heads retrieved at autopsy. Clin Orthop Relat Res. 1999;358:223–234.CrossRefPubMed
42.
Zurück zum Zitat Tateiwa T, Clarke IC, Williams PA, Garino J, Manaka M, Shishido T, Yamamoto K, Imakiire A. Ceramic total hip arthroplasty in the United States: safety and risk issues revisited. Am J Orthop. 2008;37:E26–E31.PubMed Tateiwa T, Clarke IC, Williams PA, Garino J, Manaka M, Shishido T, Yamamoto K, Imakiire A. Ceramic total hip arthroplasty in the United States: safety and risk issues revisited. Am J Orthop. 2008;37:E26–E31.PubMed
43.
Zurück zum Zitat Taylor S, Manley MT, Sutton K. The role of stripe wear in causing acoustic emissions from alumina ceramic-on-ceramic bearings. J Arthroplasty. 2007;22(suppl 3):47–51.CrossRefPubMed Taylor S, Manley MT, Sutton K. The role of stripe wear in causing acoustic emissions from alumina ceramic-on-ceramic bearings. J Arthroplasty. 2007;22(suppl 3):47–51.CrossRefPubMed
44.
Zurück zum Zitat Tradonsky S, Postak PD, Froimson AI, Greenwald AS. A comparison of the disassociation strength of modular acetabular components. Clin Orthop Relat Res. 1993;296:154–160.PubMed Tradonsky S, Postak PD, Froimson AI, Greenwald AS. A comparison of the disassociation strength of modular acetabular components. Clin Orthop Relat Res. 1993;296:154–160.PubMed
45.
Zurück zum Zitat Urban JA, Garvin KL, Boese CK, Bryson L, Pedersen DR, Callaghan JJ, Miller RK. Ceramic-on-polyethylene bearing surfaces in total hip arthroplasty: seventeen to twenty-one year results. J Bone Joint Surg Am. 2001;83:1688–1694.PubMed Urban JA, Garvin KL, Boese CK, Bryson L, Pedersen DR, Callaghan JJ, Miller RK. Ceramic-on-polyethylene bearing surfaces in total hip arthroplasty: seventeen to twenty-one year results. J Bone Joint Surg Am. 2001;83:1688–1694.PubMed
46.
Zurück zum Zitat Walter WL, Insley GM, Walter WK, Tuke MA. Edge loading in third generation alumina ceramic-on-ceramic bearings: stripe wear. J Arthroplasty. 2004;19:402–413.CrossRefPubMed Walter WL, Insley GM, Walter WK, Tuke MA. Edge loading in third generation alumina ceramic-on-ceramic bearings: stripe wear. J Arthroplasty. 2004;19:402–413.CrossRefPubMed
47.
Zurück zum Zitat Walter WL, O’Toole GC, Walter WK, Ellis A, Zicat BA. Squeaking in ceramic-on-ceramic hips: the importance of acetabular orientation. J Arthroplasty. 2007;22:496–503.CrossRefPubMed Walter WL, O’Toole GC, Walter WK, Ellis A, Zicat BA. Squeaking in ceramic-on-ceramic hips: the importance of acetabular orientation. J Arthroplasty. 2007;22:496–503.CrossRefPubMed
48.
Zurück zum Zitat Walter WL, Waters TS, Gillies M, Donohoo S, Kurtz SM, Ranawat AS, Hozack WJ, Tuke MA. Squeaking hips. J Bone Joint Surg Am. 2008;90 (suppl 4):102–111.CrossRefPubMed Walter WL, Waters TS, Gillies M, Donohoo S, Kurtz SM, Ranawat AS, Hozack WJ, Tuke MA. Squeaking hips. J Bone Joint Surg Am. 2008;90 (suppl 4):102–111.CrossRefPubMed
49.
Zurück zum Zitat Williams VG 2nd, Whiteside LA, White SE, McCarthy DS. Fixation of ultrahigh-molecular-weight polyethylene liners to metal-backed acetabular cups. J Arthroplasty. 1997;12:25–31.CrossRefPubMed Williams VG 2nd, Whiteside LA, White SE, McCarthy DS. Fixation of ultrahigh-molecular-weight polyethylene liners to metal-backed acetabular cups. J Arthroplasty. 1997;12:25–31.CrossRefPubMed
50.
51.
Zurück zum Zitat Yamamoto K, Shishido T, Tateiwa T, Katori Y, Masaoka T, Imakiire A, Clarke IC. Failure of ceramic THR with liner dissociation—a case report. Acta Orthop Scand. 2004;75:500–502.PubMedCrossRef Yamamoto K, Shishido T, Tateiwa T, Katori Y, Masaoka T, Imakiire A, Clarke IC. Failure of ceramic THR with liner dissociation—a case report. Acta Orthop Scand. 2004;75:500–502.PubMedCrossRef
52.
Zurück zum Zitat Yang CC, Kim RH, Dennis DA. The squeaking hip: a cause for concern—Disagrees. Orthopedics. 2007;30:739–742.PubMed Yang CC, Kim RH, Dennis DA. The squeaking hip: a cause for concern—Disagrees. Orthopedics. 2007;30:739–742.PubMed
53.
Zurück zum Zitat Yoo JJ, Kim YM, Yoon KS, Koo KH, Song WS, Kim HJ. Alumina-on-alumina total hip arthroplasty: a five-year minimum follow-up study. J Bone Joint Surg Am. 2005;87:530–535.CrossRefPubMed Yoo JJ, Kim YM, Yoon KS, Koo KH, Song WS, Kim HJ. Alumina-on-alumina total hip arthroplasty: a five-year minimum follow-up study. J Bone Joint Surg Am. 2005;87:530–535.CrossRefPubMed
Metadaten
Titel
Delta Ceramic-on-Alumina Ceramic Articulation in Primary THA: Prospective, Randomized FDA-IDE Study and Retrieval Analysis
verfasst von
Adolph V. Lombardi Jr., MD
Keith R. Berend, MD
Brian E. Seng, DO
Ian C. Clarke, PhD
Joanne B. Adams, BFA
Publikationsdatum
01.02.2010
Verlag
Springer-Verlag
Erschienen in
Clinical Orthopaedics and Related Research® / Ausgabe 2/2010
Print ISSN: 0009-921X
Elektronische ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-009-1143-9

Weitere Artikel der Ausgabe 2/2010

Clinical Orthopaedics and Related Research® 2/2010 Zur Ausgabe

Symposium: Papers Presented at the Hip Society Meetings 2009

Outcomes of Isolated Acetabular Revision

Letter to the Editor

Reply to Letter to the Editor

Symposium: Papers Presented at the Hip Society Meetings 2009

The Classic: Chapter XVIII. Operative Treatment in Chronic Articular Ostitis

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Arthroskopie kann Knieprothese nicht hinauszögern

25.04.2024 Gonarthrose Nachrichten

Ein arthroskopischer Eingriff bei Kniearthrose macht im Hinblick darauf, ob und wann ein Gelenkersatz fällig wird, offenbar keinen Unterschied.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.