Skip to main content
Erschienen in: Archives of Orthopaedic and Trauma Surgery 1/2015

01.01.2015 | Trauma Surgery

Distal femoral fractures in the elderly: biomechanical analysis of a polyaxial angle-stable locking plate versus a retrograde intramedullary nail in a human cadaveric bone model

verfasst von: Christopher Bliemel, Benjamin Buecking, Thorben Mueller, Christina Wack, Christos Koutras, Tabea Beck, Steffen Ruchholtz, Ralph Zettl

Erschienen in: Archives of Orthopaedic and Trauma Surgery | Ausgabe 1/2015

Einloggen, um Zugang zu erhalten

Abstract

Introduction

Compromised bone quality and the need for early mobilization still lead to high rates of implant failure in geriatric patients with distal femoral fractures. With the newest generation of polyaxial locking plates and the proven retrograde femoral nails today two minimally invasive surgical procedures have been established. Indications for both procedures overlap. This study attempts to define the strength and failure mode of both surgical procedures.

Materials and methods

A standardized fracture model was established to simulate an unstable AO/OTA 33-A3 fracture. Eight pairs of human cadaver femora (mean age 79 years, range 63–100 years) with compromised bone quality were used. Osteosyntheses with eight retrograde femoral nails and eight locking plates were randomly performed. A materials testing machine (Instron 5566) was used to perform cyclic stress tests according to a standardized loading protocol, up to a maximum load of 5,000 N.

Results

All specimens survived loading of at least 2,500 N. Three nail and one plate construct survived a maximum load of 5,000 N. The mean compressive force leading to failure was 4,400 N (CI 4,122–4,678 N) for nail osteosynthesis and 4,429 N (CI 3,653–5,204 N) for plate osteosynthesis (p = 0.943). Proximal cutting out of the osteosynthesis was the most common reason for interruption in the nail and plate osteosyntheses. Significant differences between the retrograde femoral nail and plate osteosyntheses were seen under testing conditions for plastic deformation and stiffness of the constructs (p = 0.002 and p = 0.001, respectively).

Conclusion

Based on our results, no statements regarding the superiority of either of the devices can be made. Even though the load to failure values for both osteosyntheses were much higher than the loads experienced during normal walking; however, because only axial loading was applied, it remains unclear whether both osteosyntheses meet the estimated requirements for postoperative full weight-bearing for an average heavy patient with a distal femoral fracture.
Literatur
1.
Zurück zum Zitat Aldrian S, Schuster R, Haas N, Erhart J, Strickner M, Blutsch B, Wernhart S, Leitgeb J, Platzer P (2013) Fixation of supracondylar femoral fractures following total knee arthroplasty: is there any difference comparing angular stable plate fixation versus rigid interlocking nail fixation? Arch Orthop Trauma Surg 133:921–927PubMedCrossRef Aldrian S, Schuster R, Haas N, Erhart J, Strickner M, Blutsch B, Wernhart S, Leitgeb J, Platzer P (2013) Fixation of supracondylar femoral fractures following total knee arthroplasty: is there any difference comparing angular stable plate fixation versus rigid interlocking nail fixation? Arch Orthop Trauma Surg 133:921–927PubMedCrossRef
2.
Zurück zum Zitat Bergmann G, Deuretzbacher G, Heller M, Graichen F, Rohlmann A, Strauss J, Duda GN (2001) Hip contact forces and gait patterns from routine activities. J Biomech 34:859–871PubMedCrossRef Bergmann G, Deuretzbacher G, Heller M, Graichen F, Rohlmann A, Strauss J, Duda GN (2001) Hip contact forces and gait patterns from routine activities. J Biomech 34:859–871PubMedCrossRef
3.
Zurück zum Zitat Bottlang M, Doornink J, Fitzpatrick DC, Madey SM (2009) Far cortical locking can reduce stiffness of locked plating constructs while retaining construct strength. J Bone Joint Surg Am 91:1985–1994PubMedCentralPubMedCrossRef Bottlang M, Doornink J, Fitzpatrick DC, Madey SM (2009) Far cortical locking can reduce stiffness of locked plating constructs while retaining construct strength. J Bone Joint Surg Am 91:1985–1994PubMedCentralPubMedCrossRef
4.
Zurück zum Zitat Cristofolini L, Viceconti M, Cappello A, Toni A (1996) Mechanical validation of whole bone composite femur models. J Biomech 29:525–535PubMedCrossRef Cristofolini L, Viceconti M, Cappello A, Toni A (1996) Mechanical validation of whole bone composite femur models. J Biomech 29:525–535PubMedCrossRef
5.
Zurück zum Zitat Duan X, Li T, Mohammed AQ, Xiang Z (2011) Reamed intramedullary nailing versus unreamed intramedullary nailing for shaft fracture of femur: a systematic literature review. Arch Orthop Trauma Surg 131:1445–1452PubMedCrossRef Duan X, Li T, Mohammed AQ, Xiang Z (2011) Reamed intramedullary nailing versus unreamed intramedullary nailing for shaft fracture of femur: a systematic literature review. Arch Orthop Trauma Surg 131:1445–1452PubMedCrossRef
6.
Zurück zum Zitat Döbele S, Horn C, Eichhorn S, Buchholtz A, Lenich A, Burgkart R, Nüssler AK, Lucke M, Andermatt D, Koch R, Stöckle U (2010) The dynamic locking screw (DLS) can increase interfragmentary motion on the near cortex of locked plating constructs by reducing the axial stiffness. Langenbecks Arch Surg 395:421–428PubMedCrossRef Döbele S, Horn C, Eichhorn S, Buchholtz A, Lenich A, Burgkart R, Nüssler AK, Lucke M, Andermatt D, Koch R, Stöckle U (2010) The dynamic locking screw (DLS) can increase interfragmentary motion on the near cortex of locked plating constructs by reducing the axial stiffness. Langenbecks Arch Surg 395:421–428PubMedCrossRef
7.
Zurück zum Zitat Egol KA, Kubiak EN, Fulkerson E, Kummer FJ, Koval KJ (2004) Biomechanics of locked plates and screws. J Orthop Trauma 18:488–493PubMedCrossRef Egol KA, Kubiak EN, Fulkerson E, Kummer FJ, Koval KJ (2004) Biomechanics of locked plates and screws. J Orthop Trauma 18:488–493PubMedCrossRef
8.
Zurück zum Zitat Fankhauser F, Gruber G, Schippinger G, Boldin C, Hofer HP, Grechenig W, Szyszkowitz R (2004) Minimal-invasive treatment of distal femoral fractures with the LISS (Less Invasive Stabilization System): a prospective study of 30 fractures with a follow up of 20 months. Acta Orthop Scand 75:56–60PubMedCrossRef Fankhauser F, Gruber G, Schippinger G, Boldin C, Hofer HP, Grechenig W, Szyszkowitz R (2004) Minimal-invasive treatment of distal femoral fractures with the LISS (Less Invasive Stabilization System): a prospective study of 30 fractures with a follow up of 20 months. Acta Orthop Scand 75:56–60PubMedCrossRef
9.
Zurück zum Zitat Fensky F, Nüchtern JV, Kolb JP, Huber S, Rupprecht M, Jauch SY, Sellenschloh K, Püschel K, Morlock MM, Rueger JM, Lehmann W (2013) Cement augmentation of the proximal femoral nail antirotation for the treatment of osteoporotic pertrochanteric fractures—a biomechanical cadaver study. Injury 44:802–807PubMedCrossRef Fensky F, Nüchtern JV, Kolb JP, Huber S, Rupprecht M, Jauch SY, Sellenschloh K, Püschel K, Morlock MM, Rueger JM, Lehmann W (2013) Cement augmentation of the proximal femoral nail antirotation for the treatment of osteoporotic pertrochanteric fractures—a biomechanical cadaver study. Injury 44:802–807PubMedCrossRef
10.
Zurück zum Zitat Firoozbakhsh K, Behzadi K, DeCoster TA, Moneim MS, Naraghi FF (1995) Mechanics of retrograde nail versus plate fixation for supracondylar femur fractures. J Orthop Trauma 9:152–157PubMedCrossRef Firoozbakhsh K, Behzadi K, DeCoster TA, Moneim MS, Naraghi FF (1995) Mechanics of retrograde nail versus plate fixation for supracondylar femur fractures. J Orthop Trauma 9:152–157PubMedCrossRef
11.
Zurück zum Zitat Greiwe RM, Archdeacon MT (2007) Locking plate technology: current concepts. J Knee Surg 20:50–55PubMed Greiwe RM, Archdeacon MT (2007) Locking plate technology: current concepts. J Knee Surg 20:50–55PubMed
12.
Zurück zum Zitat Gwathmey FW, Jones-Quaidoo SM, Kahler D, Hurwitz S, Cui Q (2010) Distal femoral fractures: current concepts. J Am Acad Orthop Surg 18:597–607PubMed Gwathmey FW, Jones-Quaidoo SM, Kahler D, Hurwitz S, Cui Q (2010) Distal femoral fractures: current concepts. J Am Acad Orthop Surg 18:597–607PubMed
13.
Zurück zum Zitat Hartin NL, Harris I, Hazratwala K (2006) Retrograde nailing versus fixed-angle blade plating for supracondylar femoral fractures: a randomized controlled trial. ANZ J Surg 76:290–294PubMedCrossRef Hartin NL, Harris I, Hazratwala K (2006) Retrograde nailing versus fixed-angle blade plating for supracondylar femoral fractures: a randomized controlled trial. ANZ J Surg 76:290–294PubMedCrossRef
14.
Zurück zum Zitat Heiney JP, Barnett MD, Vrabec GA, Schoenfeld AJ, Baji A, Njus GO (2009) Distal femoral fixation: a biomechanical comparison of trigen retrograde intramedullary (i.m.) nail, dynamic condylar screw (DCS), and locking compression plate (LCP) condylar plate. J Trauma 66:443–449PubMedCrossRef Heiney JP, Barnett MD, Vrabec GA, Schoenfeld AJ, Baji A, Njus GO (2009) Distal femoral fixation: a biomechanical comparison of trigen retrograde intramedullary (i.m.) nail, dynamic condylar screw (DCS), and locking compression plate (LCP) condylar plate. J Trauma 66:443–449PubMedCrossRef
15.
Zurück zum Zitat Hente R, Füchtmeier B, Schlegel U, Ernstberger A, Perren SM (2004) The influence of cyclic compression and distraction on the healing of experimental tibial fractures. J Orthop Res 22:709–715PubMedCrossRef Hente R, Füchtmeier B, Schlegel U, Ernstberger A, Perren SM (2004) The influence of cyclic compression and distraction on the healing of experimental tibial fractures. J Orthop Res 22:709–715PubMedCrossRef
16.
Zurück zum Zitat Hwang JH, Oh JK, Oh CW, Yoon YC, Choi HW (2012) Mismatch of anatomically pre-shaped locking plate on Asian femurs could lead to malalignment in the minimally invasive plating of distal femoral fractures: a cadaveric study. Arch Orthop Trauma Surg 132:51–56PubMedCrossRef Hwang JH, Oh JK, Oh CW, Yoon YC, Choi HW (2012) Mismatch of anatomically pre-shaped locking plate on Asian femurs could lead to malalignment in the minimally invasive plating of distal femoral fractures: a cadaveric study. Arch Orthop Trauma Surg 132:51–56PubMedCrossRef
17.
Zurück zum Zitat Hüfner T, Citak M, Suero EM, Miller B, Kendoff D, Krettek C (2011) Femoral malrotation after unreamed intramedullary nailing: an evaluation of influencing operative factors. J Orthop Trauma 25:224–227PubMedCrossRef Hüfner T, Citak M, Suero EM, Miller B, Kendoff D, Krettek C (2011) Femoral malrotation after unreamed intramedullary nailing: an evaluation of influencing operative factors. J Orthop Trauma 25:224–227PubMedCrossRef
18.
Zurück zum Zitat Kenwright J, Goodship AE (1989) Controlled mechanical stimulation in the treatment of tibial fractures. Clin Orthop Relat Res 241:36–47PubMed Kenwright J, Goodship AE (1989) Controlled mechanical stimulation in the treatment of tibial fractures. Clin Orthop Relat Res 241:36–47PubMed
19.
Zurück zum Zitat Kreb DL, Blokhuis TJ, van Wessem KJ, Bemelman M, Lansink KW, Leenen LP (2013) Intramedullary nailing without interlocking screws for femoral and tibial shaft fractures. Arch Orthop Trauma Surg 133:1109–1113PubMedCrossRef Kreb DL, Blokhuis TJ, van Wessem KJ, Bemelman M, Lansink KW, Leenen LP (2013) Intramedullary nailing without interlocking screws for femoral and tibial shaft fractures. Arch Orthop Trauma Surg 133:1109–1113PubMedCrossRef
20.
Zurück zum Zitat Kregor PJ, Stannard JA, Zlowodzki M, Cole PA (2004) Treatment of distal femur fractures using the less invasive stabilization system: surgical experience and early clinical results in 103 fractures. J Orthop Trauma 18:509–520PubMedCrossRef Kregor PJ, Stannard JA, Zlowodzki M, Cole PA (2004) Treatment of distal femur fractures using the less invasive stabilization system: surgical experience and early clinical results in 103 fractures. J Orthop Trauma 18:509–520PubMedCrossRef
21.
Zurück zum Zitat Leggon RE, Feldmann DD (2001) Retrograde femoral nailing: a focus on the knee. Am J Knee Surg 14:109–118PubMed Leggon RE, Feldmann DD (2001) Retrograde femoral nailing: a focus on the knee. Am J Knee Surg 14:109–118PubMed
22.
Zurück zum Zitat Link BC, Babst R (2012) Current concepts in fractures of the distal femur. Acta Chir Orthop Traumatol Cech 79:11–20PubMed Link BC, Babst R (2012) Current concepts in fractures of the distal femur. Acta Chir Orthop Traumatol Cech 79:11–20PubMed
23.
Zurück zum Zitat Meyer RW, Plaxton NA, Postak PD, Gilmore A, Froimson MI, Greenwald AS (2000) Mechanical comparison of a distal femoral side plate and a retrograde intramedullary nail. J Orthop Trauma 14:398–404PubMedCrossRef Meyer RW, Plaxton NA, Postak PD, Gilmore A, Froimson MI, Greenwald AS (2000) Mechanical comparison of a distal femoral side plate and a retrograde intramedullary nail. J Orthop Trauma 14:398–404PubMedCrossRef
24.
Zurück zum Zitat Miclau T, Holmes W, Martin RE, Krettek C, Schandelmaier P (1998) Plate osteosynthesis of the distal femur: surgical techniques and results. J South Orthop Assoc 7:161–170PubMed Miclau T, Holmes W, Martin RE, Krettek C, Schandelmaier P (1998) Plate osteosynthesis of the distal femur: surgical techniques and results. J South Orthop Assoc 7:161–170PubMed
25.
Zurück zum Zitat Moed BR, Watson JT (1995) Retrograde intramedullary nailing, without reaming, of fractures of the femoral shaft in multiply injured patients. J Bone Joint Surg Am 77:1520–1527PubMed Moed BR, Watson JT (1995) Retrograde intramedullary nailing, without reaming, of fractures of the femoral shaft in multiply injured patients. J Bone Joint Surg Am 77:1520–1527PubMed
26.
Zurück zum Zitat Neubauer T, Krawany M, Leitner L, Karlbauer A, Wagner M, Plecko M (2012) Retrograde femoral nailing in elderly patients: outcome and functional results. Orthopedics 35:e855–e861PubMedCrossRef Neubauer T, Krawany M, Leitner L, Karlbauer A, Wagner M, Plecko M (2012) Retrograde femoral nailing in elderly patients: outcome and functional results. Orthopedics 35:e855–e861PubMedCrossRef
27.
Zurück zum Zitat Ng AC, Drake MT, Clarke BL, Sems SA, Atkinson EJ, Achenbach SJ, Melton LJ (2012) Trends in subtrochanteric, diaphyseal, and distal femur fractures, 1984–2007. Osteoporos Int 23:1721–1726PubMedCentralPubMedCrossRef Ng AC, Drake MT, Clarke BL, Sems SA, Atkinson EJ, Achenbach SJ, Melton LJ (2012) Trends in subtrochanteric, diaphyseal, and distal femur fractures, 1984–2007. Osteoporos Int 23:1721–1726PubMedCentralPubMedCrossRef
28.
Zurück zum Zitat Niedhart C, Braun K, Graf Stenbock-Fermor N, Bours F, Schneider P, Zilkens KW, Niethard FU (2003) The value of peripheral quantitative computed tomography (pQCT) in the diagnosis of osteoporosis. Z Orthop Ihre Grenzgeb 141:135–142PubMedCrossRef Niedhart C, Braun K, Graf Stenbock-Fermor N, Bours F, Schneider P, Zilkens KW, Niethard FU (2003) The value of peripheral quantitative computed tomography (pQCT) in the diagnosis of osteoporosis. Z Orthop Ihre Grenzgeb 141:135–142PubMedCrossRef
29.
Zurück zum Zitat Perren SM (2002) Evolution of the internal fixation of long bone fractures. The scientific basis of biological internal fixation: choosing a new balance between stability and biology. J Bone Joint Surg Br 84:1093–1110PubMedCrossRef Perren SM (2002) Evolution of the internal fixation of long bone fractures. The scientific basis of biological internal fixation: choosing a new balance between stability and biology. J Bone Joint Surg Br 84:1093–1110PubMedCrossRef
30.
Zurück zum Zitat Ricci WM, Bellabarba C, Evanoff B, Herscovici D, DiPasquale T, Sanders R (2001) Retrograde versus antegrade nailing of femoral shaft fractures. J Orthop Trauma 15:161–169PubMedCrossRef Ricci WM, Bellabarba C, Evanoff B, Herscovici D, DiPasquale T, Sanders R (2001) Retrograde versus antegrade nailing of femoral shaft fractures. J Orthop Trauma 15:161–169PubMedCrossRef
31.
Zurück zum Zitat Ricci WM, Streubel PN, Morshed S, Collinge C, Nork SE, Gardner MJ (2013) Risk factors for failure of locked plate fixation of distal femur fractures: an analysis of 335 cases. J Orthop Trauma 28:83–89CrossRef Ricci WM, Streubel PN, Morshed S, Collinge C, Nork SE, Gardner MJ (2013) Risk factors for failure of locked plate fixation of distal femur fractures: an analysis of 335 cases. J Orthop Trauma 28:83–89CrossRef
32.
Zurück zum Zitat Ruchholtz S, El-Zayat B, Kreslo D, Bücking B, Lewan U, Krüger A, Zettl R (2013) Less invasive polyaxial locking plate fixation in periprosthetic and peri-implant fractures of the femur—a prospective study of 41 patients. Injury 44:239–248PubMedCrossRef Ruchholtz S, El-Zayat B, Kreslo D, Bücking B, Lewan U, Krüger A, Zettl R (2013) Less invasive polyaxial locking plate fixation in periprosthetic and peri-implant fractures of the femur—a prospective study of 41 patients. Injury 44:239–248PubMedCrossRef
33.
Zurück zum Zitat Sasaki D, Hatori M, Kotajima S, Kokubun S (2005) Fatigue fracture of the distal femur arising in the elderly. Arch Orthop Trauma Surg 125:422–425PubMedCrossRef Sasaki D, Hatori M, Kotajima S, Kokubun S (2005) Fatigue fracture of the distal femur arising in the elderly. Arch Orthop Trauma Surg 125:422–425PubMedCrossRef
34.
Zurück zum Zitat Schütz M, Müller M, Regazzoni P, Höntzsch D, Krettek C, Van der Werken C, Haas N (2005) Use of the less invasive stabilization system (LISS) in patients with distal femoral (AO33) fractures: a prospective multicenter study. Arch Orthop Trauma Surg 125:102–108PubMedCrossRef Schütz M, Müller M, Regazzoni P, Höntzsch D, Krettek C, Van der Werken C, Haas N (2005) Use of the less invasive stabilization system (LISS) in patients with distal femoral (AO33) fractures: a prospective multicenter study. Arch Orthop Trauma Surg 125:102–108PubMedCrossRef
35.
Zurück zum Zitat Sievänen H, Koskue V, Rauhio A, Kannus P, Heinonen A, Vuori I (1998) Peripheral quantitative computed tomography in human long bones: evaluation of in vitro and in vivo precision. J Bone Miner Res 13:871–882PubMedCrossRef Sievänen H, Koskue V, Rauhio A, Kannus P, Heinonen A, Vuori I (1998) Peripheral quantitative computed tomography in human long bones: evaluation of in vitro and in vivo precision. J Bone Miner Res 13:871–882PubMedCrossRef
36.
Zurück zum Zitat Singh SK, El-Gendy KA, Chikkamuniyappa C, Houshian S (2006) The retrograde nail for distal femoral fractures in the elderly: high failure rate of the condyle screw and nut. Injury 37:1004–1010PubMedCrossRef Singh SK, El-Gendy KA, Chikkamuniyappa C, Houshian S (2006) The retrograde nail for distal femoral fractures in the elderly: high failure rate of the condyle screw and nut. Injury 37:1004–1010PubMedCrossRef
37.
Zurück zum Zitat Soveid M, Serati AR, Masoompoor M (2005) Incidence of hip fracture in Shiraz, Iran. Osteoporos Int 16:1412–1416PubMedCrossRef Soveid M, Serati AR, Masoompoor M (2005) Incidence of hip fracture in Shiraz, Iran. Osteoporos Int 16:1412–1416PubMedCrossRef
38.
Zurück zum Zitat Topp T, Müller T, Huss S, Kann PH, Weihe E, Ruchholtz S, Zettl RP (2012) Embalmed and fresh frozen human bones in orthopedic cadaveric studies: which bone is authentic and feasible? Acta Orthop 83:543–547PubMedCentralPubMedCrossRef Topp T, Müller T, Huss S, Kann PH, Weihe E, Ruchholtz S, Zettl RP (2012) Embalmed and fresh frozen human bones in orthopedic cadaveric studies: which bone is authentic and feasible? Acta Orthop 83:543–547PubMedCentralPubMedCrossRef
39.
Zurück zum Zitat Wang Z, Bhattacharyya T (2011) Trends in incidence of subtrochanteric fragility fractures and bisphosphonate use among the US elderly, 1996–2007. J Bone Miner Res 26:553–560PubMedCentralPubMedCrossRef Wang Z, Bhattacharyya T (2011) Trends in incidence of subtrochanteric fragility fractures and bisphosphonate use among the US elderly, 1996–2007. J Bone Miner Res 26:553–560PubMedCentralPubMedCrossRef
40.
Zurück zum Zitat Weiss RJ, Montgomery SM, Al Dabbagh Z, Jansson KA (2009) National data of 6409 Swedish inpatients with femoral shaft fractures: stable incidence between 1998 and 2004. Injury 40:304–308PubMedCrossRef Weiss RJ, Montgomery SM, Al Dabbagh Z, Jansson KA (2009) National data of 6409 Swedish inpatients with femoral shaft fractures: stable incidence between 1998 and 2004. Injury 40:304–308PubMedCrossRef
41.
Zurück zum Zitat Wähnert D, Hoffmeier K, Fröber R, Hofmann GO, Mückley T (2011) Distal femur fractures of the elderly—different treatment options in a biomechanical comparison. Injury 42:655–659PubMedCrossRef Wähnert D, Hoffmeier K, Fröber R, Hofmann GO, Mückley T (2011) Distal femur fractures of the elderly—different treatment options in a biomechanical comparison. Injury 42:655–659PubMedCrossRef
42.
Zurück zum Zitat Wähnert D, Hofmann-Fliri L, Götzen M, Kösters C, Windolf M, Raschke MJ (2013) Feasibility study on the potential of a spiral blade in osteoporotic distal femur fracture fixation. Arch Orthop Trauma Surg 133:1675–1679PubMedCrossRef Wähnert D, Hofmann-Fliri L, Götzen M, Kösters C, Windolf M, Raschke MJ (2013) Feasibility study on the potential of a spiral blade in osteoporotic distal femur fracture fixation. Arch Orthop Trauma Surg 133:1675–1679PubMedCrossRef
43.
Zurück zum Zitat Zlowodzki M, Bhandari M, Marek DJ, Cole PA, Kregor PJ (2006) Operative treatment of acute distal femur fractures: systematic review of 2 comparative studies and 45 case series (1989 to 2005). J Orthop Trauma 20:366–371PubMedCrossRef Zlowodzki M, Bhandari M, Marek DJ, Cole PA, Kregor PJ (2006) Operative treatment of acute distal femur fractures: systematic review of 2 comparative studies and 45 case series (1989 to 2005). J Orthop Trauma 20:366–371PubMedCrossRef
44.
Zurück zum Zitat Zlowodzki M, Williamson S, Cole PA, Zardiackas LD, Kregor PJ (2004) Biomechanical evaluation of the less invasive stabilization system, angled blade plate, and retrograde intramedullary nail for the internal fixation of distal femur fractures. J Orthop Trauma 18:494–502PubMedCrossRef Zlowodzki M, Williamson S, Cole PA, Zardiackas LD, Kregor PJ (2004) Biomechanical evaluation of the less invasive stabilization system, angled blade plate, and retrograde intramedullary nail for the internal fixation of distal femur fractures. J Orthop Trauma 18:494–502PubMedCrossRef
45.
Zurück zum Zitat Zlowodzki M, Williamson S, Zardiackas LD, Kregor PJ (2006) Biomechanical evaluation of the less invasive stabilization system and the 95-degree angled blade plate for the internal fixation of distal femur fractures in human cadaveric bones with high bone mineral density. J Trauma 60:836–840PubMedCrossRef Zlowodzki M, Williamson S, Zardiackas LD, Kregor PJ (2006) Biomechanical evaluation of the less invasive stabilization system and the 95-degree angled blade plate for the internal fixation of distal femur fractures in human cadaveric bones with high bone mineral density. J Trauma 60:836–840PubMedCrossRef
Metadaten
Titel
Distal femoral fractures in the elderly: biomechanical analysis of a polyaxial angle-stable locking plate versus a retrograde intramedullary nail in a human cadaveric bone model
verfasst von
Christopher Bliemel
Benjamin Buecking
Thorben Mueller
Christina Wack
Christos Koutras
Tabea Beck
Steffen Ruchholtz
Ralph Zettl
Publikationsdatum
01.01.2015
Verlag
Springer Berlin Heidelberg
Erschienen in
Archives of Orthopaedic and Trauma Surgery / Ausgabe 1/2015
Print ISSN: 0936-8051
Elektronische ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-014-2111-8

Weitere Artikel der Ausgabe 1/2015

Archives of Orthopaedic and Trauma Surgery 1/2015 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Arthroskopie kann Knieprothese nicht hinauszögern

25.04.2024 Gonarthrose Nachrichten

Ein arthroskopischer Eingriff bei Kniearthrose macht im Hinblick darauf, ob und wann ein Gelenkersatz fällig wird, offenbar keinen Unterschied.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.