Skip to main content
Erschienen in: Clinical Orthopaedics and Related Research® 12/2014

01.12.2014 | Clinical Research

Does Fluoroscopy Improve Acetabular Component Placement in Total Hip Arthroplasty?

verfasst von: Brandon S. Beamer, MD, Jordan H. Morgan, BS, Christopher Barr, BS, Michael J. Weaver, MD, Mark S. Vrahas, MD

Erschienen in: Clinical Orthopaedics and Related Research® | Ausgabe 12/2014

Einloggen, um Zugang zu erhalten

Abstract

Background

The success of THA largely depends on correct placement of the individual components. Traditionally, these have been placed freehand using anatomic landmarks, but studies have shown poor accuracy with this method.

Questions/purposes

Specifically, we asked (1) does using fluoroscopy lead to more accurate and greater likelihood of cup placement with the Lewinnek safe zone than does freehand cup placement; (2) is there a learning curve associated with the use of fluoroscopy for cup placement; (3) does the use of fluoroscopy increase operative time; and (4) is there a difference in leg length discrepancy between freehand and fluoroscopic techniques?

Methods

This series consisted of 109 consecutive patients undergoing primary THA, conversion of a previous hip surgery to THA, and revision THA during a 24-month period. No patients were excluded from analysis during this time. The first 52 patients had cups placed freehand, and then the next 57 patients had acetabular components placed using fluoroscopy; the analysis began with the first patient treated using fluoroscopy, to include our initial experience with the technique. The abduction, version, and limb length discrepancy were measured on 6-week postoperative pelvic radiographs obtained with the patient in the supine position. Operative time, sex, age, BMI, diagnosis, operative side, and femoral head size were recorded as possible confounders.

Results

Cups inserted freehand were placed in the ideal range of abduction (30°–45°) and anteversion (5°–25°) 44% of the time. With fluoroscopy, placement in the Lewinnek safe zone for both measures significantly increased to 65%. The odds of placing the cup in the Lewinnek safe zone for abduction and version were 2.3 times greater with the use of fluoroscopy (95% CI, 1.2–5.0; p = 0.03). Patients undergoing primary THAs (32 freehand, 35 C-arm) had cup placement in the safe zone for abduction and version 44% of the time freehand and 57% of the time with fluoroscopy, which failed to reach statistical significance. There was no difference in operative time, patient age, sex, operative side, diagnosis, limb length discrepancy, or femoral head size between the two groups.

Conclusions

The use of fluoroscopy to directly observe pelvic position and acetabular component placement increased the success of placement in the Lewinnek safe zone in this cohort of patients having complex and primary THAs. This is a simple, low-cost, and quick method for increasing successful acetabular component alignment. The study population included a large proportion of patients having complex THAs, and further validation of this technique in patients undergoing straightforward, primary THAs needs to be done to understand if similar gains in accuracy for component placement can be expected in that group.

Level of Evidence

Level III, therapeutic study. See the Instructions for Authors for a complete description of levels of evidence.
Literatur
1.
Zurück zum Zitat Biedermann R, Tonin A, Krismer M, Rachbauer F, Eibl G, Stockl B. Reducing the risk of dislocation after total hip arthroplasty: the effect of orientation of the acetabular component. J Bone Joint Surg Br. 2005;87:762–769.PubMedCrossRef Biedermann R, Tonin A, Krismer M, Rachbauer F, Eibl G, Stockl B. Reducing the risk of dislocation after total hip arthroplasty: the effect of orientation of the acetabular component. J Bone Joint Surg Br. 2005;87:762–769.PubMedCrossRef
2.
Zurück zum Zitat Bragdon CR, Martell JM, Greene ME, Estok DM 2nd, Thanner J, Karrholm J, Harris WH, Malchau H. Comparison of femoral head penetration using RSA and the Martell method. Clin Orthop Relat Res. 2006;448:52–57.PubMedCrossRef Bragdon CR, Martell JM, Greene ME, Estok DM 2nd, Thanner J, Karrholm J, Harris WH, Malchau H. Comparison of femoral head penetration using RSA and the Martell method. Clin Orthop Relat Res. 2006;448:52–57.PubMedCrossRef
3.
Zurück zum Zitat Callanan MC, Jarrett B, Bragdon CR, Zurakowski D, Rubash HE, Freiberg AA, Malchau H. The John Charnley Award: risk factors for cup malpositioning: quality improvement through a joint registry at a tertiary hospital. Clin Orthop Relat Res. 2011;469:319–329.PubMedCentralPubMedCrossRef Callanan MC, Jarrett B, Bragdon CR, Zurakowski D, Rubash HE, Freiberg AA, Malchau H. The John Charnley Award: risk factors for cup malpositioning: quality improvement through a joint registry at a tertiary hospital. Clin Orthop Relat Res. 2011;469:319–329.PubMedCentralPubMedCrossRef
4.
Zurück zum Zitat Digioia AM 3rd, Jaramaz B, Plakseychuk AY, Moody JE Jr, Nikou C, Labarca RS, Levison TJ, Picard F. Comparison of a mechanical acetabular alignment guide with computer placement of the socket. J Arthroplasty. 2002;17:359–364.PubMedCrossRef Digioia AM 3rd, Jaramaz B, Plakseychuk AY, Moody JE Jr, Nikou C, Labarca RS, Levison TJ, Picard F. Comparison of a mechanical acetabular alignment guide with computer placement of the socket. J Arthroplasty. 2002;17:359–364.PubMedCrossRef
5.
Zurück zum Zitat Hosmer DW, Lemeshow S. Applied Logistic Regression. New York, NY: Wiley; 2000.CrossRef Hosmer DW, Lemeshow S. Applied Logistic Regression. New York, NY: Wiley; 2000.CrossRef
6.
Zurück zum Zitat Hui AJ, McCalden RW, Martell JM, MacDonald SJ, Bourne RB, Rorabeck CH. Validation of two and three-dimensional radiographic techniques for measuring polyethylene wear after total hip arthroplasty. J Bone Joint Surg Am. 2003;85:505–511.PubMed Hui AJ, McCalden RW, Martell JM, MacDonald SJ, Bourne RB, Rorabeck CH. Validation of two and three-dimensional radiographic techniques for measuring polyethylene wear after total hip arthroplasty. J Bone Joint Surg Am. 2003;85:505–511.PubMed
7.
Zurück zum Zitat Katz JN, Losina E, Barrett J, Phillips CB, Mahomed NN, Lew RA, Guadagnoli E, Harris WH, Poss R, Baron JA. Association between hospital and surgeon procedure volume and outcomes of total hip replacement in the United States medicare population. J Bone Joint Surg Am. 2001;83:1622–1629.PubMedCrossRef Katz JN, Losina E, Barrett J, Phillips CB, Mahomed NN, Lew RA, Guadagnoli E, Harris WH, Poss R, Baron JA. Association between hospital and surgeon procedure volume and outcomes of total hip replacement in the United States medicare population. J Bone Joint Surg Am. 2001;83:1622–1629.PubMedCrossRef
8.
Zurück zum Zitat Katz MH. Multivariable Analysis : A Practical Guide for Clinicians and Public Health Researchers. New York, NY: Cambridge University Press; 2011.CrossRef Katz MH. Multivariable Analysis : A Practical Guide for Clinicians and Public Health Researchers. New York, NY: Cambridge University Press; 2011.CrossRef
9.
Zurück zum Zitat Kennedy JG, Rogers WB, Soffe KE, Sullivan RJ, Griffen DG, Sheehan LJ. Effect of acetabular component orientation on recurrent dislocation, pelvic osteolysis, polyethylene wear, and component migration. J Arthroplasty. 1998;13:530–534.PubMedCrossRef Kennedy JG, Rogers WB, Soffe KE, Sullivan RJ, Griffen DG, Sheehan LJ. Effect of acetabular component orientation on recurrent dislocation, pelvic osteolysis, polyethylene wear, and component migration. J Arthroplasty. 1998;13:530–534.PubMedCrossRef
10.
Zurück zum Zitat Kjellberg M, Al-Amiry B, Englund E, Sjoden GO, Sayed-Noor AS. Measurement of leg length discrepancy after total hip arthroplasty: the reliability of a plain radiographic method compared to CT-scanogram. Skeletal Radiol. 2012;41:187–191.PubMedCrossRef Kjellberg M, Al-Amiry B, Englund E, Sjoden GO, Sayed-Noor AS. Measurement of leg length discrepancy after total hip arthroplasty: the reliability of a plain radiographic method compared to CT-scanogram. Skeletal Radiol. 2012;41:187–191.PubMedCrossRef
11.
Zurück zum Zitat Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR. Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am. 1978;60:217–220.PubMed Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR. Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am. 1978;60:217–220.PubMed
12.
Zurück zum Zitat Martell JM, Berdia S. Determination of polyethylene wear in total hip replacements with use of digital radiographs. J Bone Joint Surg Am. 1997;79:1635–1641.PubMed Martell JM, Berdia S. Determination of polyethylene wear in total hip replacements with use of digital radiographs. J Bone Joint Surg Am. 1997;79:1635–1641.PubMed
13.
Zurück zum Zitat Martell JM, Berkson E, Berger R, Jacobs J. Comparison of two and three-dimensional computerized polyethylene wear analysis after total hip arthroplasty. J Bone Joint Surg Am. 2003;85:1111–1117.PubMed Martell JM, Berkson E, Berger R, Jacobs J. Comparison of two and three-dimensional computerized polyethylene wear analysis after total hip arthroplasty. J Bone Joint Surg Am. 2003;85:1111–1117.PubMed
14.
Zurück zum Zitat Meermans G, Malik A, Witt J, Haddad F. Preoperative radiographic assessment of limb-length discrepancy in total hip arthroplasty. Clin Orthop Relat Res. 2011;469:1677–1682.PubMedCentralPubMedCrossRef Meermans G, Malik A, Witt J, Haddad F. Preoperative radiographic assessment of limb-length discrepancy in total hip arthroplasty. Clin Orthop Relat Res. 2011;469:1677–1682.PubMedCentralPubMedCrossRef
15.
Zurück zum Zitat Moskal JT, Capps SG. Acetabular component positioning in total hip arthroplasty: an evidence-based analysis. J Arthroplasty. 2011;26:1432–1437.PubMedCrossRef Moskal JT, Capps SG. Acetabular component positioning in total hip arthroplasty: an evidence-based analysis. J Arthroplasty. 2011;26:1432–1437.PubMedCrossRef
16.
Zurück zum Zitat Murray DW. The definition and measurement of acetabular orientation. J Bone Joint Surg Br. 1993;75:228–232.PubMed Murray DW. The definition and measurement of acetabular orientation. J Bone Joint Surg Br. 1993;75:228–232.PubMed
17.
Zurück zum Zitat Nishihara S, Sugano N, Nishii T, Ohzono K, Yoshikawa H. Measurements of pelvic flexion angle using three-dimensional computed tomography. Clin Orthop Relat Res. 2003;411:140–151.PubMedCrossRef Nishihara S, Sugano N, Nishii T, Ohzono K, Yoshikawa H. Measurements of pelvic flexion angle using three-dimensional computed tomography. Clin Orthop Relat Res. 2003;411:140–151.PubMedCrossRef
18.
Zurück zum Zitat Patil S, Bergula A, Chen PC, Colwell CW Jr, D’Lima DD. Polyethylene wear and acetabular component orientation. J Bone Joint Surg Am. 2003;85(suppl 4):56–63.PubMed Patil S, Bergula A, Chen PC, Colwell CW Jr, D’Lima DD. Polyethylene wear and acetabular component orientation. J Bone Joint Surg Am. 2003;85(suppl 4):56–63.PubMed
19.
Zurück zum Zitat Siebenrock KA, Kalbermatten DF, Ganz R. Effect of pelvic tilt on acetabular retroversion: a study of pelves from cadavers. Clin Orthop Relat Res. 2003;407:241–248.PubMedCrossRef Siebenrock KA, Kalbermatten DF, Ganz R. Effect of pelvic tilt on acetabular retroversion: a study of pelves from cadavers. Clin Orthop Relat Res. 2003;407:241–248.PubMedCrossRef
20.
Zurück zum Zitat Tsalafoutas IA, Tsapaki V, Kaliakmanis A, Pneumaticos S, Tsoronis F, Koulentianos ED, Papachristou G. Estimation of radiation doses to patients and surgeons from various fluoroscopically guided orthopaedic surgeries. Radiat Prot Dosimetry. 2008;128:112–119.PubMedCrossRef Tsalafoutas IA, Tsapaki V, Kaliakmanis A, Pneumaticos S, Tsoronis F, Koulentianos ED, Papachristou G. Estimation of radiation doses to patients and surgeons from various fluoroscopically guided orthopaedic surgeries. Radiat Prot Dosimetry. 2008;128:112–119.PubMedCrossRef
21.
Zurück zum Zitat Wera GD, Ting NT, Moric M, Paprosky WG, Sporer SM, Della Valle CJ. Classification and management of the unstable total hip arthroplasty. J Arthroplasty. 2012;27:710–715.PubMedCrossRef Wera GD, Ting NT, Moric M, Paprosky WG, Sporer SM, Della Valle CJ. Classification and management of the unstable total hip arthroplasty. J Arthroplasty. 2012;27:710–715.PubMedCrossRef
Metadaten
Titel
Does Fluoroscopy Improve Acetabular Component Placement in Total Hip Arthroplasty?
verfasst von
Brandon S. Beamer, MD
Jordan H. Morgan, BS
Christopher Barr, BS
Michael J. Weaver, MD
Mark S. Vrahas, MD
Publikationsdatum
01.12.2014
Verlag
Springer US
Erschienen in
Clinical Orthopaedics and Related Research® / Ausgabe 12/2014
Print ISSN: 0009-921X
Elektronische ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-014-3944-8

Weitere Artikel der Ausgabe 12/2014

Clinical Orthopaedics and Related Research® 12/2014 Zur Ausgabe

Symposium: Award Papers from Turkish Society of Orthopaedics and Traumatology 2013

Apical and Intermediate Anchors Without Fusion Improve Cobb Angle and Thoracic Kyphosis in Early-onset Scoliosis

Symposium: 2013 Limb Lengthening and Reconstruction Society

External Fixation for Closed Pediatric Femoral Shaft Fractures: Where Are We Now?

Symposium: ABJS Carl T. Brighton Workshop on Implant Wear and Tribocorrosion of Total Joint Replacements

Which Design and Biomaterial Factors Affect Clinical Wear Performance of Total Disc Replacements? A Systematic Review

Symposium: 2013 Limb Lengthening and Reconstruction Society

Editorial Comment: 2013 Limb Lengthening and Reconstruction Society

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Arthroskopie kann Knieprothese nicht hinauszögern

25.04.2024 Gonarthrose Nachrichten

Ein arthroskopischer Eingriff bei Kniearthrose macht im Hinblick darauf, ob und wann ein Gelenkersatz fällig wird, offenbar keinen Unterschied.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.