Objective

The aim of this study is to evaluate the efficacy of metal artefact reduction (MAR) in different dental materials with Picasso Trio cone-beam computed tomography (CBCT) scanner.

Materials and methods

Three imaging phantoms were custom-made of acrylic resin. Each phantom presented three cylinders of the same material: dental amalgam alloy, gutta-percha or aluminium-copper alloy. CBCT scans were performed on Picasso Trio unit with and without MAR, and artefact expression (standard deviation of grey values) was obtained and compared by Kruskal-Wallis and Student-Newman-Keuls (post hoc) (α = 0.05).

Results

Significant reduction of artefact expression (p < 0.05) was observed with MAR on areas around dental alloys. No significant difference (p > 0.05) was observed with or without MAR when gutta-percha was scanned.

Conclusion

MAR was effective in reducing artefacts arising from dental alloys on CBCT images.

Clinical relevance

Dental materials of high atomic number and density are widely used in dentistry and can produce artefact that compromise CBCT image. The present study demonstrated that metal artefact reduction algorithm is an effective tool to improve image quality.
Literatur
1.
Zurück zum Zitat Scarfe WC, Farman AG, Sukovic P (2006) Clinical applications of cone-beam computed tomography in dental practice. J Cant Dent Assoc 72:75–80 Scarfe WC, Farman AG, Sukovic P (2006) Clinical applications of cone-beam computed tomography in dental practice. J Cant Dent Assoc 72:75–80
2.
Zurück zum Zitat Bechara B, McMahan A, Moore WS, Noujeim M, Geha H (2012a) Contrast-to-noise ratio with different large volumes in a cone-beam computerized tomography machine: an in vitro study. Oral Surg Oral Med Oral Pathol Oral Radiol 114:658–665CrossRefPubMed Bechara B, McMahan A, Moore WS, Noujeim M, Geha H (2012a) Contrast-to-noise ratio with different large volumes in a cone-beam computerized tomography machine: an in vitro study. Oral Surg Oral Med Oral Pathol Oral Radiol 114:658–665CrossRefPubMed
3.
Zurück zum Zitat Oliveira ML, Freitas DQ, Ambrosano GM, Haiter-Neto F (2014) Influence of exposure factors on the variability of CBCT voxel values: a phantom study. Dentomaxillofac Radiol 43:20140128CrossRefPubMedPubMedCentral Oliveira ML, Freitas DQ, Ambrosano GM, Haiter-Neto F (2014) Influence of exposure factors on the variability of CBCT voxel values: a phantom study. Dentomaxillofac Radiol 43:20140128CrossRefPubMedPubMedCentral
4.
Zurück zum Zitat Schulze R, Heil U, Groβ D, Bruellmann DD, Dranischnikow E, Schwanecke U, Schoemer E (2011) Artefacts in CBCT: a review. Dentomaxillofac Radiol 40:265–273CrossRefPubMedPubMedCentral Schulze R, Heil U, Groβ D, Bruellmann DD, Dranischnikow E, Schwanecke U, Schoemer E (2011) Artefacts in CBCT: a review. Dentomaxillofac Radiol 40:265–273CrossRefPubMedPubMedCentral
5.
Zurück zum Zitat Gamba TO, Oliveira ML, Flores IL, Cruz AD, Almeida SM, Haiter-Neto F, Lopes SL (2014) Influence of cone-beam computed tomography image artifacts on the determination of dental arch measurements. Angle Orthod 84:274–278CrossRefPubMed Gamba TO, Oliveira ML, Flores IL, Cruz AD, Almeida SM, Haiter-Neto F, Lopes SL (2014) Influence of cone-beam computed tomography image artifacts on the determination of dental arch measurements. Angle Orthod 84:274–278CrossRefPubMed
6.
Zurück zum Zitat Bezerra ISQ, Neves FS, Vasconcelos TV, Ambrosano GMB, Freitas DQ (2015) Influence of the artefact reduction algorithm of Picasso Trio CBCT system on the diagnosis of vertical root fractures in teeth with metal posts. Dentomaxillofac Radiol 44:20140428CrossRefPubMedPubMedCentral Bezerra ISQ, Neves FS, Vasconcelos TV, Ambrosano GMB, Freitas DQ (2015) Influence of the artefact reduction algorithm of Picasso Trio CBCT system on the diagnosis of vertical root fractures in teeth with metal posts. Dentomaxillofac Radiol 44:20140428CrossRefPubMedPubMedCentral
7.
Zurück zum Zitat Barrett JF, Keat N (2004) Artifacts in CT: recognition and avoidance. Radiographics 24:1679–1691CrossRefPubMed Barrett JF, Keat N (2004) Artifacts in CT: recognition and avoidance. Radiographics 24:1679–1691CrossRefPubMed
8.
Zurück zum Zitat Neves FS, Freitas DQ, Campos PS, Ekestubbe A, Lofthag-Hansen S (2014) Evaluation of cone-beam computed tomography in the diagnosis of vertical root fractures: the influence of imaging modes and root canal materials. J Endod 40:1530–1536CrossRefPubMed Neves FS, Freitas DQ, Campos PS, Ekestubbe A, Lofthag-Hansen S (2014) Evaluation of cone-beam computed tomography in the diagnosis of vertical root fractures: the influence of imaging modes and root canal materials. J Endod 40:1530–1536CrossRefPubMed
9.
Zurück zum Zitat Mahnken AH, Raupach R, Wildberger JE, Jung B, Heussen N, Flohr TC, Günther RW, Schaller S (2003) New algorithm for metal artifact reduction in computed tomography: in vitro and in vivo evaluation after total hip replacement. Investig Radiol 38:769–775CrossRef Mahnken AH, Raupach R, Wildberger JE, Jung B, Heussen N, Flohr TC, Günther RW, Schaller S (2003) New algorithm for metal artifact reduction in computed tomography: in vitro and in vivo evaluation after total hip replacement. Investig Radiol 38:769–775CrossRef
10.
Zurück zum Zitat Maltz JS, Gangadharan B, Bose S, Hristov DH, Faddegon BA, Paidi A, Bani-Hashemi AR (2008) Algorithm for X-ray scatter, beam-hardening, and beam profile correction in diagnostic (kilovoltage) and treatment (megavoltage) cone beam CT. IEEE Trans Med Imaging 27:1791–1810CrossRefPubMed Maltz JS, Gangadharan B, Bose S, Hristov DH, Faddegon BA, Paidi A, Bani-Hashemi AR (2008) Algorithm for X-ray scatter, beam-hardening, and beam profile correction in diagnostic (kilovoltage) and treatment (megavoltage) cone beam CT. IEEE Trans Med Imaging 27:1791–1810CrossRefPubMed
11.
Zurück zum Zitat Bechara B, Moore WS, McMahan CA, Noujeim M (2012b) Metal artefact reduction with cone beam CT: an in vitro study. Dentomaxillofac Radiol 41:248–253CrossRefPubMedPubMedCentral Bechara B, Moore WS, McMahan CA, Noujeim M (2012b) Metal artefact reduction with cone beam CT: an in vitro study. Dentomaxillofac Radiol 41:248–253CrossRefPubMedPubMedCentral
12.
13.
Zurück zum Zitat Pick B, Pelka M, Beli R, Braga RR, Lohbaue U (2011) Tailoring of physical properties in highly filled experimental nanohybrid resin composites. Dent Mater 27:664–669CrossRefPubMed Pick B, Pelka M, Beli R, Braga RR, Lohbaue U (2011) Tailoring of physical properties in highly filled experimental nanohybrid resin composites. Dent Mater 27:664–669CrossRefPubMed
14.
Zurück zum Zitat Webber RL, Tzukert A, Ruttimann U (1989) The effects of beam hardening on digital subtraction radiograph. J Periodontal Res 24:53CrossRefPubMed Webber RL, Tzukert A, Ruttimann U (1989) The effects of beam hardening on digital subtraction radiograph. J Periodontal Res 24:53CrossRefPubMed
Metadaten
Titel
Evaluation of metal artefact reduction in cone-beam computed tomography images of different dental materials
verfasst von
Polyane Mazucatto Queiroz
Matheus Lima Oliveira
Francisco Carlos Groppo
Francisco Haiter-Neto
Deborah Queiroz Freitas
Publikationsdatum
23.05.2017
Verlag
Springer Berlin Heidelberg
Erschienen in
Clinical Oral Investigations / Ausgabe 1/2018
Print ISSN: 1432-6981
Elektronische ISSN: 1436-3771
DOI
https://doi.org/10.1007/s00784-017-2128-9

Weitere Artikel der Ausgabe 1/2018

Clinical Oral Investigations 1/2018 Zur Ausgabe

Parodontalbehandlung verbessert Prognose bei Katheterablation

19.04.2024 Vorhofflimmern Nachrichten

Werden Personen mit Vorhofflimmern in der Blanking-Periode nach einer Katheterablation gegen eine bestehende Parodontitis behandelt, verbessert dies die Erfolgsaussichten. Dafür sprechen die Resultate einer prospektiven Untersuchung.

Invasive Zahnbehandlung: Wann eine Antibiotikaprophylaxe vor infektiöser Endokarditis schützt

11.04.2024 Endokarditis Nachrichten

Bei welchen Personen eine Antibiotikaprophylaxe zur Prävention einer infektiösen Endokarditis nach invasiven zahnärztlichen Eingriffen sinnvoll ist, wird diskutiert. Neue Daten stehen im Einklang mit den europäischen Leitlinienempfehlungen.

Zell-Organisatoren unter Druck: Mechanismen des embryonalen Zahnwachstums aufgedeckt

08.04.2024 Zahnmedizin Nachrichten

Der Aufbau von Geweben und Organen während der Embryonalentwicklung wird von den Zellen bemerkenswert choreografiert. Für diesen Prozess braucht es spezielle sogenannte „Organisatoren“. In einer aktuellen Veröffentlichung im Fachjournal Nature Cell Biology berichten Forschende durch welchen Vorgang diese Organisatoren im Gewebe entstehen und wie sie dann die Bildung von Zähnen orchestrieren.

Die Oralprophylaxe & Kinderzahnheilkunde umbenannt

11.03.2024 Kinderzahnmedizin Nachrichten

Infolge der Umbenennung der Deutschen Gesellschaft für Kinderzahnheilkunde in Deutsche Gesellschaft für Kinderzahnmedizin (DGKiZ) wird deren Mitgliederzeitschrift Oralprophylaxe & Kinderzahnheilkunde in Oralprophylaxe & Kinderzahnmedizin umbenannt. Aus diesem Grunde trägt die erste Ausgabe in 2024 erstmalig den neuen Titel.

Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Zahnmedizin und bleiben Sie gut informiert – ganz bequem per eMail.