Implant fracture under dynamic fatigue loading: influence of embedded angle and depth of implant

verfasst von: Hiroaki Suzuki, Yoshiaki Hata, Fumihiko Watanabe

Erschienen in: Odontology | Ausgabe 3/2016

Einloggen, um Zugang zu erhalten

Abstract

The purpose of this study was to investigate the relationship between implant fracture under cyclic-fatigue loading at different embedding angles, embedding depths, and loading forces. Twenty-four cylinder-type implants 3.3 mm in diameter and 10 mm in length were used. Test specimens were 30 mm3 resin blocks with one surfaces inclined at angles of either 5°, 10°, 15° and 20° and embedded vertically with implants at depths of either 5 or 10 mm to the these surfaces. A straight abutment was connected to the implant and cut to 5 mm in length, and a hemispherical crown 5 mm in diameter and 7 mm in length was cast with a 12 % gold–silver–palladium alloy and cemented onto the abutment. Each specimen was mounted onto a fatigue loading device to apply repeated vertical loads of 294, 392, and 490 N to the coronal edge of the crown 60 times per min until reaching 100,000 cycles. For each respective specimen, we recorded the combined conditions of embedding and loading forces and the number of loading cycles until fracture, and then observed the fracture sites microscopically. The number of loading cycles until implant fracture tended to decrease in proportion to increased loading forces and embedded angles, and decreased embedded depths. Implant fracture was observed at angles of inclination over 10°. For specimens with an implant embedded at a depth of 5 mm, almost all fractures occurred at the center of the implant body; however, for those embedded at a depth of 10 mm, fractures occurred at the interface between the implant body and the abutment. These results demonstrate that implant fracture is associated with the loading axis, the amount of loading, and the embedded depth of the implant.
Literatur
1.
Zurück zum Zitat Jemt T. Failure and complications in 391 consecutive inserted fixed prostheses supported by Brånemark implants in edentulous jaws: a study of treatment from the time of prosthesis placement to the first annual checkup. Int J Oral Maxillofac Implants. 1991;6:270–6.PubMed Jemt T. Failure and complications in 391 consecutive inserted fixed prostheses supported by Brånemark implants in edentulous jaws: a study of treatment from the time of prosthesis placement to the first annual checkup. Int J Oral Maxillofac Implants. 1991;6:270–6.PubMed
2.
Zurück zum Zitat Lekholm U, Van Steenberghe D, Herrmann I, et al. Osseointegrated implants in the treatment of partially edentulous jaws: a prospective 5-year multicenter study. Int J Oral Maxillofac Implants. 1994;9:627–35. Lekholm U, Van Steenberghe D, Herrmann I, et al. Osseointegrated implants in the treatment of partially edentulous jaws: a prospective 5-year multicenter study. Int J Oral Maxillofac Implants. 1994;9:627–35.
3.
Zurück zum Zitat Zarb GA, Schmitt A. The longitudinal clinical effectiveness of osseointegrated implants: the Toronto study. Part III: problems and complications encountered. J Prosthet Dent. 1990;64:185–94.CrossRefPubMed Zarb GA, Schmitt A. The longitudinal clinical effectiveness of osseointegrated implants: the Toronto study. Part III: problems and complications encountered. J Prosthet Dent. 1990;64:185–94.CrossRefPubMed
4.
Zurück zum Zitat Goodacre CJ, Bernar G, Rungcharassaeng K, Kan JYK. Clinical complications with implants and implant prostheses. J Prosthet Dent. 2003;90:121–32.CrossRefPubMed Goodacre CJ, Bernar G, Rungcharassaeng K, Kan JYK. Clinical complications with implants and implant prostheses. J Prosthet Dent. 2003;90:121–32.CrossRefPubMed
5.
Zurück zum Zitat Green NT, Machtei EE, Horwitz J, Peled M. Fracture of dental implants: literature review and report of a case. Implant Dent. 2002;11:137–43.CrossRef Green NT, Machtei EE, Horwitz J, Peled M. Fracture of dental implants: literature review and report of a case. Implant Dent. 2002;11:137–43.CrossRef
6.
Zurück zum Zitat Balshi T. An analysis and management of fractured implants: a clinical report. Int J Oral Maxillofac Implants. 1996;11:660–6.PubMed Balshi T. An analysis and management of fractured implants: a clinical report. Int J Oral Maxillofac Implants. 1996;11:660–6.PubMed
7.
Zurück zum Zitat Schwarz MS. Mechanical complications of dental implants. Clin Oral Implants Res. 2000;11(Suppl 1):156–8.CrossRefPubMed Schwarz MS. Mechanical complications of dental implants. Clin Oral Implants Res. 2000;11(Suppl 1):156–8.CrossRefPubMed
8.
Zurück zum Zitat Eckert SE, Meraw SJ, Cal E, Ow RK. Analysis of incidence and associated factors with fractured implants: a retrospective study. Int J Oral Maxillofac Implants. 2000;15:662–7.PubMed Eckert SE, Meraw SJ, Cal E, Ow RK. Analysis of incidence and associated factors with fractured implants: a retrospective study. Int J Oral Maxillofac Implants. 2000;15:662–7.PubMed
9.
Zurück zum Zitat Perriard J, Wiskott WA, Mellal A, Scherrer SS, Botsis J, Belser UC. Fatigue resistance of ITI implant-abutment connectors—a comparison of the standard cone with a novel internally keyed design. Clin Oral Implants Res. 2002;13:542–9.CrossRefPubMed Perriard J, Wiskott WA, Mellal A, Scherrer SS, Botsis J, Belser UC. Fatigue resistance of ITI implant-abutment connectors—a comparison of the standard cone with a novel internally keyed design. Clin Oral Implants Res. 2002;13:542–9.CrossRefPubMed
10.
Zurück zum Zitat Steinebrunner L, Wolfart S, Ludwig K, Kern M. Implant-abutment interface design affects fatigue and fracture strength of implants. Clin Oral Implants Res. 2008;19:1276–84.CrossRefPubMed Steinebrunner L, Wolfart S, Ludwig K, Kern M. Implant-abutment interface design affects fatigue and fracture strength of implants. Clin Oral Implants Res. 2008;19:1276–84.CrossRefPubMed
11.
Zurück zum Zitat Japan Industrial Standard. Zinc phosphate dental cement, T6602, 1993. Japan Industrial Standard. Zinc phosphate dental cement, T6602, 1993.
12.
Zurück zum Zitat ISO 14801:2007. Dentistry—Implant—Dynamic fatigue test for endosseous dental implants. ISO 14801:2007. Dentistry—Implant—Dynamic fatigue test for endosseous dental implants.
13.
Zurück zum Zitat ASTM E486-90:2004 Standard practice for presentation of constant amplitude fatigue test results for metallic materials. ASTM E486-90:2004 Standard practice for presentation of constant amplitude fatigue test results for metallic materials.
14.
Zurück zum Zitat Takamizawa T. Studies on the co-relative and individual biting forces of normal permanent teeth. J Jpn Prosthodont Soc. 1965;9:217–36.CrossRef Takamizawa T. Studies on the co-relative and individual biting forces of normal permanent teeth. J Jpn Prosthodont Soc. 1965;9:217–36.CrossRef
15.
Zurück zum Zitat Goodacre CJ, Kan JY, Rungcharassaeng K. Clinical complications of osseointegrated implants. J Prosthet Dent. 1999;81:537–52.CrossRefPubMed Goodacre CJ, Kan JY, Rungcharassaeng K. Clinical complications of osseointegrated implants. J Prosthet Dent. 1999;81:537–52.CrossRefPubMed
16.
Zurück zum Zitat Green NT, Machtei EE, Horwitz J, Peled M. Fracture of dental implants; literature review and report of a case. Implant Dent. 2002;1:137–43.CrossRef Green NT, Machtei EE, Horwitz J, Peled M. Fracture of dental implants; literature review and report of a case. Implant Dent. 2002;1:137–43.CrossRef
17.
Zurück zum Zitat Morgan MJ, James DF, Pilliar RM. Fractures of the fixture component of an osseointegrated implant. Int J Oral Maxillofac Implants. 1993;8:409–14.PubMed Morgan MJ, James DF, Pilliar RM. Fractures of the fixture component of an osseointegrated implant. Int J Oral Maxillofac Implants. 1993;8:409–14.PubMed
18.
Zurück zum Zitat Rangert B, Krogh P, Langer B, Van Roekel N. Bending overload and implant failure: a retrospective clinical analysis. Int J Oral Maxillofac Implants. 1995;10:326–34.PubMed Rangert B, Krogh P, Langer B, Van Roekel N. Bending overload and implant failure: a retrospective clinical analysis. Int J Oral Maxillofac Implants. 1995;10:326–34.PubMed
19.
Zurück zum Zitat Huang HW, Tsai CH, Chang CC, Lin CT, Lee SY. Evaluation of loading conditions on fatigue-failed implants by fracture surface analysis. Int J Oral Maxillofac Implants. 2005;20:854–9.PubMed Huang HW, Tsai CH, Chang CC, Lin CT, Lee SY. Evaluation of loading conditions on fatigue-failed implants by fracture surface analysis. Int J Oral Maxillofac Implants. 2005;20:854–9.PubMed
20.
Zurück zum Zitat Dittmer MP, Dittmer S, Borchersh L, Kohorst P, Stiesch M. Influence of the interface design on the yield force of the implant-abutment complex before and after cyclic mechanical loading. J Prosthodont Res. 2012;56:19–24.CrossRefPubMed Dittmer MP, Dittmer S, Borchersh L, Kohorst P, Stiesch M. Influence of the interface design on the yield force of the implant-abutment complex before and after cyclic mechanical loading. J Prosthodont Res. 2012;56:19–24.CrossRefPubMed
21.
Zurück zum Zitat Chan HL, Oh WS, Ong HS, Fu JH, Steigmann M, Sierraalta M, Wang HL. Impact of implantoplasty on strength of the implant-abutment complex. Int J Oral Maxillofac Implants. 2013;28:1530–5.CrossRefPubMed Chan HL, Oh WS, Ong HS, Fu JH, Steigmann M, Sierraalta M, Wang HL. Impact of implantoplasty on strength of the implant-abutment complex. Int J Oral Maxillofac Implants. 2013;28:1530–5.CrossRefPubMed
Metadaten
Titel
Implant fracture under dynamic fatigue loading: influence of embedded angle and depth of implant
verfasst von
Hiroaki Suzuki
Yoshiaki Hata
Fumihiko Watanabe
Publikationsdatum
23.12.2015
Verlag
Springer Japan
Erschienen in
Odontology / Ausgabe 3/2016
Print ISSN: 1618-1247
Elektronische ISSN: 1618-1255
DOI
https://doi.org/10.1007/s10266-015-0228-z

Weitere Artikel der Ausgabe 3/2016

Odontology 3/2016 Zur Ausgabe

Parodontalbehandlung verbessert Prognose bei Katheterablation

19.04.2024 Vorhofflimmern Nachrichten

Werden Personen mit Vorhofflimmern in der Blanking-Periode nach einer Katheterablation gegen eine bestehende Parodontitis behandelt, verbessert dies die Erfolgsaussichten. Dafür sprechen die Resultate einer prospektiven Untersuchung.

Invasive Zahnbehandlung: Wann eine Antibiotikaprophylaxe vor infektiöser Endokarditis schützt

11.04.2024 Endokarditis Nachrichten

Bei welchen Personen eine Antibiotikaprophylaxe zur Prävention einer infektiösen Endokarditis nach invasiven zahnärztlichen Eingriffen sinnvoll ist, wird diskutiert. Neue Daten stehen im Einklang mit den europäischen Leitlinienempfehlungen.

Zell-Organisatoren unter Druck: Mechanismen des embryonalen Zahnwachstums aufgedeckt

08.04.2024 Zahnmedizin Nachrichten

Der Aufbau von Geweben und Organen während der Embryonalentwicklung wird von den Zellen bemerkenswert choreografiert. Für diesen Prozess braucht es spezielle sogenannte „Organisatoren“. In einer aktuellen Veröffentlichung im Fachjournal Nature Cell Biology berichten Forschende durch welchen Vorgang diese Organisatoren im Gewebe entstehen und wie sie dann die Bildung von Zähnen orchestrieren.

Die Oralprophylaxe & Kinderzahnheilkunde umbenannt

11.03.2024 Kinderzahnmedizin Nachrichten

Infolge der Umbenennung der Deutschen Gesellschaft für Kinderzahnheilkunde in Deutsche Gesellschaft für Kinderzahnmedizin (DGKiZ) wird deren Mitgliederzeitschrift Oralprophylaxe & Kinderzahnheilkunde in Oralprophylaxe & Kinderzahnmedizin umbenannt. Aus diesem Grunde trägt die erste Ausgabe in 2024 erstmalig den neuen Titel.

Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Zahnmedizin und bleiben Sie gut informiert – ganz bequem per eMail.