Insertion of a pressure sensing arrayminimally affects hindfoot bone kinematics

verfasst von: Tassos Natsakis, Josefien Burg, Greta Dereymaeker, Ilse Jonkers, Jos Vander Sloten

Erschienen in: Journal of Foot and Ankle Research | Ausgabe 1/2015

Abstract

Background

Understanding the development of ankle osteoarthritis (OA) is of high importance and interest; however its causality is poorly understood and several links to joint loading conditions have been made. One way of quantifying joint loading conditions is by measuring the intra-articular pressure distribution during gait simulations performed by in-vitro experimental set-ups. However the effect of inserting a pressure sensing array in the ankle joint could potentially disturb the proper kinematics and therefore the loading conditions.

Methods

In this study, we performed in-vitro gait simulations in 7 cadaveric feet, before and after inserting a pressure sensing array and quantified the effect on the joints range of motion (ROM). The gait was simulated with a stance phase duration of one second using a custom build cadaveric gait simulator (CGS).

Results

The results show a limited effect in the ROM for all the joints of the hind foot, not exceeding the variability observed in specimens without a sensor. However, no consistent direction (increase/decrease) can be observed.

Conclusion

The results suggest that even though the effect of inserting a pressure sensing array is minimal, it needs to be evaluated against the demands/requirements of the application.
Hinweise

Competing interests

The authors declare that they have no competing interests.

Authors’ contributions

All authors were fully involved in the preparation of this study. TN, JB and GD prepared the specimens and performed the data collection during gait simulations. TN, IJ and JVS carried out the data analysis and drafted the manuscript. All authors read and approved the final manuscript.

Background

Joint loading conditions are believed to play an instrumental role in the development and progression of osteoarthritis (OA), especially in joints of the lower limbs that are subjected to higher loads during many activities of daily living [1]. Specifically for the ankle joint (i.e. the joint between the tibia, fibula and talus), several investigators have linked the onset of OA and appearance of osteophytes to the location of high strains and stresses in the joint [24]. These can be caused by trauma, inducing joint instability with an abnormal and unfavourable loading pattern on the cartilage [57]. Quantification of the intra-articular pressure distribution in normal or post-traumatic situations is thus of high importance.
One approach for quantifying intra-articular pressure is to compute it by means of modelling, either using Finite Element Analysis (FEA) [810] or musculoskeletal forward or inverse dynamics simulations [11, 12]. For FEA models the specific geometry of the bones and cartilage layers is documented through computed tomography (CT) and/or magnetic resonance imaging (MRI) and three dimensional (3D) volumetric meshes are constructed in a computer environment. Material properties are assigned and motion is imposed virtually, allowing for the pressure between the cartilage layers to be calculated. The motion imposed to the joint is based either on captured in-vivo kinematics, or is chosen to represent the standard gait cycle, and most often, a simple rotation in one direction (e.g. plantar/dorsiflexion) is imposed [9, 10]. However, as these models are sensitive to changes in material properties of the modelled bodies and the kinematics imposed, they can only give a qualitative description of the pressure distribution. To be used more quantitatively, they need to be validated against direct measurements of intra-articular pressure distribution.
Pressure sensitive arrays have been used during in-vitro experimentation to measure intra-articular pressure distribution. After tissue dissection, the sensing array is inserted in the joint and captures pressure either statically [13, 14] using Fuji-Film, which is a thin film that changes in colour proportional to the load that is applied on it. To measure pressure distribution dynamically, piezo-resistive (e.g. TekScan) or capacitive (e.g. Novel) sensors should be used. These sensors translate the force applied over an area into an electrical signal and can therefore capture pressure distribution over a period of time. During dynamic measurements, the pressure distribution throughout stance-phase can be measured in an in vitro-setup [5, 1520]. It is however important to verify whether the sensor insertion interferes with the joint kinematics, as differences in kinematics will also affect the joint loading conditions. Although the sensors used are minimal in thickness and no major dissections are performed, small changes in the configuration of the joint might have an important effect. However, so far, the effect of sensor insertion on hind foot kinematics has not yet been reported in literature.
In this study, we perform gait simulations before and after inserting a pressure array in the ankle joint of cadaveric specimens. A custom made cadaveric gait simulator (CGS), with documented high level of repeatability [21], is used. The simulations are performed under identical speed, muscle actuation pattern and imposed tibial kinematics. We hypothesise that the insertion of the pressure sensitive array does not affect the kinematics of individual bones during stance phase. To verify this hypothesis the kinematics of individual bones of the hindfoot are measured to quantify the effect of the sensor insertion on the range of motion (ROM) of hindfoot joints.

Methods

7 freshly frozen cadaveric specimens were amputated mid-tibially and were used to perform gait simulations using a CGS that was previously validated against in vivo kinematics measured using intra-cortical pins [21]. The CGS is able to simulate stance phase on cadaveric feet specimens by imposing tibial kinematics in two translations (anterior-posterior and distal-proximal directions) and one rotation (sagittal plane) and by activating 9 muscles (Peroneal muscles, Extensor hallucis, Extensor digitorum, Tibialis anterior, Tibialis posterior, Flexor hallucis, Flexor digitorum, Gastrocnemius, Soleus) grouped in 6 groups [22] over the duration of stance phase (1 sec). The CGS is using a specimen specific model for the input tibial kinematics [23] and is operating in an inertial control loop [24], that allows performing physiologic simulations without a pre-defined trajectory for the vertical ground reaction force (vGRF) or the kinematics in the vertical direction. Intra-cortical pins (diameter: 4 mm, length: 50 mm, ICOS, New Deal, France) were inserted in 5 bones of each specimen (Tibia, Talus, Calcaneus, Naviculair, Cuboid). On top of each pin, a cluster of four active markers was mounted and its motion was captured by a Krypton Optoelectronic Motion Capture System (Krypton K 600, Metris, Belgium). The position of the markers was used to determine the 3D kinematics of the bones during gait simulations.
For each specimen, two sets of measurements were performed: 1) gait simulations with the specimen intact and 2) gait simulations after inserting a pressure sensitive array in the ankle joint. For each set, 15 repetitions were performed. For the second set of measurements, the intra-articular pressure distribution in the ankle joint was measured using a Tekscan #5033 sensor (Tekscan Inc, Boston, MA). The sensor holds an array of 32×46 individual sensels in an area of 38.4×26.7 mm and has a thickness of 0.1 mm. The sensor was selected given its minimal thickness and overall dimensions, to minimize interference with normal joint function. An anterior vertical incision through the skin, inferior extensor retinaculum and joint capsule was made to access the ankle joint. Furthermore, a posterior incision was performed to give access to the posterior side of the joint. To position the sensor in the articulation, it was first positioned between two rigid plastic surfaces of the same width (Fig. 1a). The three pieces were then pushed through the joint until the sensing area was positioned in the centre of the joint and was fixated on the posterior side of the tibia using a metallic screw (Fig. 1b). The two plastic sheets where then removed. No tendons or ligaments were harmed during these incisions and their line of action was not disturbed. The two incisions were finally stitched, leaving a small opening for the side of the sensor that was connected to the reader. The operation was performed by an experienced foot surgeon.

Data analysis

For determining the bone kinematics throughout stance phase, coordinate frames based on anatomical landmarks were constructed for each bone. The 3D bone motion was calculated by the projections of the coordinate frame of the distal bone on the planes of the proximal bone, over the duration of stance phase for each bone combination.
The ROM for each direction and bone combination was calculated as the difference between the minimal and maximal rotation over stance phase. An estimated difference in the ROM was calculated and a non-parametric test for statistical differences (Wilcoxon rank sum test) between the pre and post-sensor insertion sets was performed for each foot individually, as well as for the grouped results of all specimens. The non-parametric test was chosen as it does not assume normality of the measurements. Therefore, individual specimen response was differentiated from group results, as high variability between specimens in cadaveric testing could obscure significant individual effects. Finally, all calculations were performed for different periods of stance phase in order to detect the part of stance phase that the differences were introduced. The stance phase intervals that were analysed were from 0 to 20 %, from 21 to 80 % and from 81 to 100 %, representing the initial double support (IDS), single support (SS), and terminal double support (TDS) phases of the gait cycle. For each of these analyses, the level of significance was set to 95 % (p =.05). To determine whether the differences observed were relevant, the variability of the ROM before inserting the sensor was calculated based on all the measurements for each bone combination and plane.

Results

To demonstrate the repeatability of the imposed muscle actuation during the simulations prior and after the sensor insertion, a regression analysis of the applied forces on the tendons of the cadaver pre- and post-insertion was performed (Fig. 2). Very high repeatability (R2>0.97) is reported for all muscle actuations, except the Flexor Hallucis (R2=0.803). Also high repeatability (R2>0.9) is reported in the measured ground reaction forces in all three directions and for the tibial rotation that was imposed.

Differences in ROM during stance phase

The resulting kinematics of the six bone combinations for the rotations around three anatomical planes, for the measurements before and after the sensor insertion, are presented in Fig. 3. The summary of the differences in ROM for all specimens and for each bone combination and direction is presented in Fig. 4. The greatest differences were detected for the talo-navicular in transverse and coronal planes with a 3.2 and 1.6 ° decreased ROM (Table 1). Furthermore, the talo-calcaneal joint had a 1.2 ° decreased ROM in the coronal plane. The tibio-talar joint exhibited a less than 1 ° difference in all three plane. None of these differences was statistically significant.
Table 1
Changes in ROM after the sensor insertion for all specimens and bone combinations
Plane
 
tib-tal
tal-calc
tal-nav
calc-cub
nav-cub
SAGITTAL
Specimen 1
-1.3
-0.5
-0.8
0.4
-0.1
 
Specimen 2
-4.2
-0.7
0.7
0.1
-0.4
 
Specimen 3
3.8
-9.5 *
-10.6 *
-0.3
0
 
Specimen 4
0
-7.8 *
-0.7
-14.5 *
-0.9
 
Specimen 5
6.7 *
2.1
5.7 *
1.2
-1.5 *
 
Specimen 6
-7.9
-2.4
4.3
0.1
-0.1
 
Specimen 7
-2.3
3.2
4.9
1.1
-0.2
 
All Specimens
-0.9
-0.7
1
-0.1
-0.3
 
StD
6.7
6.4
4.9
5.6
1.1
CORONAL
Specimen 1
0.1
-0.4
-0.6
0.3
-0.5
 
Specimen 2
0.4
-0.7
-2.3
2
-1
 
Specimen 3
-14.1 *
-10.1 *
2
1
1.4
 
Specimen 4
0.9
-0.4
-1.9
-12.3 *
-1.4
 
Specimen 5
3.7
1.3
-2.8
0.4
0.9
 
Specimen 6
8.5 *
-2
-4.9
3.2
3.8 *
 
Specimen 7
-1.2
-1.5
-3.4
1.7
-1.9 *
 
All specimens
-0.9
-1.2
-1.6
0.1
-0.6
 
StD
4.4
4.1
5.4
4.4
1.7
TRANSVERSE
Specimen 1
-0.2
0.6
0
0.2
-0.8
 
Specimen 2
2.5
1.8
-3.5
-0.3
0.7
 
Specimen 3
-16.4 *
-15.9 *
-34.3 *
0.9
-1.4
 
Specimen 4
-0.3
-0.8
-4.2
-8.3 *
0.8
 
Specimen 5
3.5
1.4
-4.7
3.8 *
-0.1
 
Specimen 6
1.6
3.7
-1.5
0.2
-0.4
 
Specimen 7
-2
-1.9
7.7
2.8 *
2 *
 
All specimens
0.6
-0.1
-3.2
-0.6
-0.2
 
StD
6.3
5
10.8
2.6
1.9
Statistical significant differences within the standard deviation rate are reported with a star (*, p<0.05)

Changes in ROM during different phases of stance

A similar response as when analysing the ROM for stance phase duration, is present for the ROM during the three different phases of stance (Fig. 5). For the IDS phase, the greatest differences appear for the talo-navicular joint in the transverse plane (1.7 ° decrease, Table 2), the tibio-talar joint in the coronal plane (1.6 ° decrease, Table 3) and the talo-calcaneal joint in the sagittal plane (1.3 ° decrease, Table 4). The differences in the SS phase are further limited to the talo-naviculair joint in the transverse plane (2.4 ° decrease) and the tibio-talar joint in the sagittal plane (1.4 ° increase). Finally, for the TDS phase the talo-navicular joint ROM decreased in the transverse plane (1.4 °) and the tibio-talar joint ROM decreased in the sagittal plane (1 °). None of these differences were statistically significant.
Table 2
Differences of ROM pre- and post- sensor insertion for all bone combinations for the transverse direction. The results are broken down for the three phases of stance, corresponding to the initial double support phase (0 to 20 %), single stance (21 to 79 %) and final double support phase (80 to 100 %)
Phase
 
tib-tal
tal-calc
tal-nav
calc-cub
nav-cub
IDS
Specimen 1
0
0.1
-0.2
0.7
-0.6
 
Specimen 2
-0.4
0.5
-0.9
-0.3
0.4
 
Specimen 3
-16.9 *
-14.1 *
-33.1 *
-0.5
1.9 *
 
Specimen 4
-0.5
-3.3
1
-8.5 *
0.5
 
Specimen 5
1.3
-0.1
-6.6
1.3
-0.7
 
Specimen 6
-1
2.3
-2.4
-0.4
0
 
Specimen 7
-0.9
-1.7
11.8 *
3.9 *
0.9
 
All specimens
-0.2
-0.3
-1.7
-0.6
0.5
 
StD
6.4
4.4
10.1
2.8
1.2
SS
Specimen 1
-0.1
0.3
0.1
-0.2
-0.2
 
Specimen 2
-0.1
0.3
-2.3
0
0.9
 
Specimen 3
-5.8 *
-6 *
-13.7 *
1.3 *
-0.3
 
Specimen 4
0.5
2.1
-4.7
-1.6 *
0.3
 
Specimen 5
2.3
0.8
2.3
2.2 *
0.9
 
Specimen 6
0.8
0.1
-3.2
1 *
-0.1
 
Specimen 7
0.4
0.6
-1
0.3
0.3
 
All specimens
0
-0.1
-2.4
0.5
0.3
 
StD
2.6
2.8
6
0.9
1.1
TDS
Specimen 1
0
0
-0.2
0.5
-0.2
 
Specimen 2
1.3
2.3 *
-1.1
-0.4
0
 
Specimen 3
-4 *
-3.7 *
-6.4 *
-1.1 *
0.1
 
Specimen 4
0.1
1.2 *
-1.3
-1.2 *
0.6
 
Specimen 5
-0.2
0.1
-0.4
0.6
0
 
Specimen 6
0.3
1.1
-0.3
0.2
0
 
Specimen 7
-0.3
-0.2
-0.6
0.1
0.1
 
All specimens
-0.2
0.1
-1.4
-0.3
0
 
StD
1.5
1.1
2.3
0.7
0.6
Statistical significant differences within the standard deviation rate are reported with a star (*, p<0.05)
Table 3
Differences of ROM pre- and post- sensor insertion for all bone combinations for the coronal direction
Phase
 
tib-tal
tal-calc
tal-nav
calc-cub
nav-cub
IDS
Specimen 1
-0.1
-0.5
-0.4
0
-0.5
 
Specimen 2
-1.8
0.8
0.3
1.2
-0.2
 
Specimen 3
-14.1 *
-10.3 *
2.2
5.8 *
0
 
Specimen 4
0.3
-0.9
0.5
-13.9 *
0.5
 
Specimen 5
1
-1
-2.8
-1.1
-0.4
 
Specimen 6
1.3
-2.1
-7.7
1.6
2.2 *
 
Specimen 7
-2.9
-1.9
-4.6
1.9
-0.6
 
All specimens
-1.6
-1.2
-0.4
0.3
0.1
 
StD
4.5
4.1
5
4.5
1.3
SS
Specimen 1
-0.1
0.1
-0.1
0.4
0.1
 
Specimen 2
0.1
-1.3
-3 *
0.8
-0.1
 
Specimen 3
-3.4 *
-1.7
-4.7 *
1.7
0.5
 
Specimen 4
0.5
-3 *
-2
-8 *
-2.5 *
 
Specimen 5
1.9 *
2.8 *
1.1
0.4
1.9 *
 
Specimen 6
3.1 *
1
0.7
1.1
1.6 *
 
Specimen 7
-0.9
1
3.4 *
0.7
-1
 
All specimens
0.5
0
-0.3
0.3
-0.1
 
StD
1.6
2.5
2.3
3.2
1.1
TDS
Specimen 1
0.1
-0.1
-0.4
0
-0.1
 
Specimen 2
-1
-1.4
-3.8 *
0.4
-0.7 *
 
Specimen 3
-3.6 *
-4.1 *
0.7
0.3
-0.2
 
Specimen 4
0.3
-1.6 *
-0.5
-0.6
-0.9 *
 
Specimen 5
2.1 *
0.5
0.3
0.1
0.3
 
Specimen 6
-0.1
1
1.6
0.2
0.3
 
Specimen 7
0.1
-0.3
0.2
-0.1
0.1
 
All specimens
-0.2
-0.6
0.1
0.1
-0.1
 
StD
1.1
1.6
1.2
0.8
0.6
Statistical significant differences within the standard deviation rate are reported with a star (*, p<0.05)
Table 4
Differences of ROM pre- and post- sensor insertion for all bone combinations for the sagittal direction
Phase
 
tib-tal
tal-calc
tal-nav
calc-cub
nav-cub
IDS
Specimen 1
-0.5
-0.9
-0.7
0.4
0.1
 
Specimen 2
-7
0.3
0.1
0
-0.1
 
Specimen 3
3.9
-5.5 *
-2.9
0.6
-1.4 *
 
Specimen 4
-1.1
-12.9 *
-0.9
-15.2 *
-1.3 *
 
Specimen 5
2.1
0.6
0.6
0.3
-0.7
 
Specimen 6
-18.5
-3.6
0.8
0.5
-0.4
 
Specimen 7
-2.5
2.7
1.7
1.6
0.1
 
All specimens
-1.2
-1.3
0.4
0.3
-0.6
 
StD
7.2
5.1
3.6
6
0.9
SS
Specimen 1
-0.2
0.3
-0.2
-0.1
-0.1
 
Specimen 2
-0.8
-0.8
0.3
-0.1
-0.3
 
Specimen 3
1.5
-3.4
-5.6 *
0.3
1 *
 
Specimen 4
-0.5
-8.6 *
-1.7
-6.5 *
0.4
 
Specimen 5
4.1 *
1.9
3.6 *
0.4
-0.5 *
 
Specimen 6
3.3
0.7
1.8
0.2
0
 
Specimen 7
0.1
-0.7
1.4
0.2
-0.3
 
All specimens
1.4
-0.3
0.3
-0.2
0
 
StD
3.9
6.3
2.7
2.5
0.4
TDS
Specimen 1
-0.9
-0.1
0
-0.2
-0.1
 
Specimen 2
0.4
-0.7
0.7
0.2
-0.4
 
Specimen 3
-4 *
-2.1 *
-1.3 *
0.1
0.1
 
Specimen 4
-0.3
2.9 *
-0.1
-0.9 *
0
 
Specimen 5
1.3
-0.1
0.9
0.5 *
-0.3
 
Specimen 6
-0.9
1
0.3
0.2
0
 
Specimen 7
-0.8
-0.4
-0.3
-0.1
0.1
 
All specimens
-1
-0.1
0.2
-0.1
-0.1
 
StD
2.8
1.1
1.1
0.4
0.5
Statistical significant differences that were observed are reported with a star (*, p<.05)

Discussion

In this study we investigate the effect of the insertion of a pressure sensing array in the ankle joint, by quantifying the difference in ROM in cadaveric specimens during simulated gait. Several gait cycles were performed using freshly frozen cadaveric specimens, and the cycles were performed prior and after the insertion of a pressure sensing array. The ROM of different joints was calculated and compared statistically between the two situations. Finally, the input forces applied on the specimens during the measurements were compared between the two sets of measurements in order to showcase the similarity of the boundary conditions. The comparison of the forces suggests that the conditions after inserting the sensor were identical to those before, as high values of correlation appear (R2>0.97 for all muscles, except for flexor hallucis R2=0.8 and R2>0.92 for all directions of the ground reaction force). Therefore the differences found in ROM can be assumed to be affected only by the sensor insertion. The lowest value of correlation for the flexor hallucis muscle relates to the fact that this muscle has the least actuation during stance phase (max F=8.5 N) and thus the noise-to-signal ratio of the measured force is higher. However, the combination of a lower contribution of the muscle in the gait and the still relatively high R2 value does not contaminate the repeatability of the measurements.
The results of the differences in ROM demonstrate a median difference in ROM after the sensor insertion less than 2.5 °, for all bone combinations and directions. The largest differences occurred in the tibio-talar and talo-navicular joints in sagittal plane rotation, and in talo-navicular joint in transverse plane. However, these differences were within the variability observed in the specimens without the sensor and were not detected to be statistically significant. The individual specimens demonstrated only limited significant differences, with the exception of specimen 3, that demonstrated significant differences in many joints and directions, both for the whole stance phase duration but also for the specific phases of stance.
Even though larger changes in ROM were expected in the ankle joint, where the incision was made, the largest differences appear in the talo-naviculair joint. This finding can be explained by the fact that this joint is less supported, especially during the initial phase of stance. This seems to be confirmed by the larger differences in ROM at the beginning than in the middle and end of stance phase. Such results agree with previous findings by Okita et al. [25], who observed higher compliance of the midtarsal joints during foot flat and push-off phases, compared to the loading response phase.
This study is the first one to quantify the effect of a sensing array insertion in the ankle joint. Several researches have reported findings on intra-articular pressure in the ankle joint [9, 15] or validated finite element models with it [8]. However, even in the absence of significant and relevant differences in the ROM after sensor insertion, it is important to notice that, there is some variability between different specimens, with some specimens presenting increase and others decrease in ROM. This variability shows that even with a repeatable experimental protocol, it is difficult to predict the effect of inserting a sensor in the ankle joint. Therefore attention is needed when interpreting intra-articular pressure measurements obtained from in-vitro simulations, as such measurements might not necessarily reflect to the pre-sensor insertion situation. An example of this can be seen for specimen 3, where much higher difference in ROM is observed, probably due to slightly different incision during the insertion of the sensor. When such a difference is detected, the results of the intra-articular pressure distribution cannot be considered reliable, and the measurements from such specimens should be discarded. Finally, when using the experimental pressure measurements for model validation, a margin of error of 2.5 ° on the associated kinematics should be taken into account, and a sensitivity analysis should be performed to judge if such an error margin would induce relevant differences for the envisaged application.

Conclusion

This study investigated the influence of inserting a pressure sensing array in the ankle joint on the kinematics of the hindfoot bones, during simulated roll-offs. The influence was determined by the difference in ROM of five joints during pre- and post-insertion simulations. The limited differences in ROM indicate that the kinematics are not affected significantly by such a procedure and therefore the measurements obtained can be considered representative of the pre-insertion conditions.

Acknowledgements

This work was funded by the Chair Berghmans- Dereymaeker, the Research Foundation Flanders and the Agency for Innovation by Science and Technology in Flanders (IWT).
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://​creativecommons.​org/​licenses/​by/​4.​0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.

Competing interests

The authors declare that they have no competing interests.

Authors’ contributions

All authors were fully involved in the preparation of this study. TN, JB and GD prepared the specimens and performed the data collection during gait simulations. TN, IJ and JVS carried out the data analysis and drafted the manuscript. All authors read and approved the final manuscript.
Literatur
2.
Zurück zum Zitat Buckwalter JA, Mankin HJ, Grodzinsky AJ. Articular cartilage and Osteoarthritis. Instr Course Lect -American Acad Orthop Surg. 2005; 54:465–80. Buckwalter JA, Mankin HJ, Grodzinsky AJ. Articular cartilage and Osteoarthritis. Instr Course Lect -American Acad Orthop Surg. 2005; 54:465–80.
3.
Zurück zum Zitat Anderson DD, Van Hofwegen C, Marsh JL, Brown TD. Is elevated contact stress predictive of post-traumatic osteoarthritis for imprecisely reduced tibial plafond fractures?J Orthop Res. 2011; 29(1):33–9.CrossRefPubMedPubMedCentral Anderson DD, Van Hofwegen C, Marsh JL, Brown TD. Is elevated contact stress predictive of post-traumatic osteoarthritis for imprecisely reduced tibial plafond fractures?J Orthop Res. 2011; 29(1):33–9.CrossRefPubMedPubMedCentral
4.
Zurück zum Zitat Buckwalter JA, Anderson DD, Brown TD, Tochigi Y, Martin Ja. The roles of mechanical stresses in the pathogenesis of osteoarthritis: implications for treatment of joint injuries. Cartilage. 2013; 4(4):286–94.CrossRefPubMedPubMedCentral Buckwalter JA, Anderson DD, Brown TD, Tochigi Y, Martin Ja. The roles of mechanical stresses in the pathogenesis of osteoarthritis: implications for treatment of joint injuries. Cartilage. 2013; 4(4):286–94.CrossRefPubMedPubMedCentral
5.
Zurück zum Zitat Krause FG, Schmid T. Ankle arthrodesis versus total ankle replacement: how do I decide?Foot Ankle Clin. 2012; 17(4):529–43.CrossRefPubMed Krause FG, Schmid T. Ankle arthrodesis versus total ankle replacement: how do I decide?Foot Ankle Clin. 2012; 17(4):529–43.CrossRefPubMed
6.
Zurück zum Zitat Valderrabano V, Horisberger M, Russell I, Dougall H, Hintermann B. Etiology of ankle osteoarthritis. Clin Orthop Relat Res. 2009; 467(7):1800–6.CrossRefPubMed Valderrabano V, Horisberger M, Russell I, Dougall H, Hintermann B. Etiology of ankle osteoarthritis. Clin Orthop Relat Res. 2009; 467(7):1800–6.CrossRefPubMed
7.
Zurück zum Zitat Horisberger M, Valderrabano V. Ankle Osteoarthritis - a review of the current state of knowledge. Eur Musculoskelet Rev. 2011; 6(2):114–8. Horisberger M, Valderrabano V. Ankle Osteoarthritis - a review of the current state of knowledge. Eur Musculoskelet Rev. 2011; 6(2):114–8.
8.
Zurück zum Zitat Anderson DD, Goldsworthy JK, Li W, James Rudert M, Tochigi Y, Brown TD. Physical validation of a patient-specific contact finite element model of the ankle. J Biomech. 2007; 40(8):1662–9.CrossRefPubMedPubMedCentral Anderson DD, Goldsworthy JK, Li W, James Rudert M, Tochigi Y, Brown TD. Physical validation of a patient-specific contact finite element model of the ankle. J Biomech. 2007; 40(8):1662–9.CrossRefPubMedPubMedCentral
9.
Zurück zum Zitat Anderson DD, Goldsworthy JK, Shivanna K, Grosland NM, Pedersen DR, Thomas TP. Intra-articular contact stress distributions at the ankle throughout stance phase-patient-specific finite element analysis as a metric of degeneration propensity. Biomech Model Mechanobiol. 2006; 5(2–3):82–9.CrossRefPubMedPubMedCentral Anderson DD, Goldsworthy JK, Shivanna K, Grosland NM, Pedersen DR, Thomas TP. Intra-articular contact stress distributions at the ankle throughout stance phase-patient-specific finite element analysis as a metric of degeneration propensity. Biomech Model Mechanobiol. 2006; 5(2–3):82–9.CrossRefPubMedPubMedCentral
10.
Zurück zum Zitat Li G, Wan L, Kozanek M. Determination of real-time in-vivo cartilage contact deformation in the ankle joint. J Biomech. 2008; 41(1):128–36.CrossRefPubMed Li G, Wan L, Kozanek M. Determination of real-time in-vivo cartilage contact deformation in the ankle joint. J Biomech. 2008; 41(1):128–36.CrossRefPubMed
11.
Zurück zum Zitat Sasimontonkul S, Bay BK, Pavol MJ. Bone contact forces on the distal tibia during the stance phase of running. J Biomech. 2007; 40(15):3503–9.CrossRefPubMed Sasimontonkul S, Bay BK, Pavol MJ. Bone contact forces on the distal tibia during the stance phase of running. J Biomech. 2007; 40(15):3503–9.CrossRefPubMed
12.
Zurück zum Zitat Moissenet F, Chèze L, Dumas R. A 3D lower limb musculoskeletal model for simultaneous estimation of musculo-tendon, joint contact, ligament and bone forces during gait. J Biomech. 2014; 47(1):50–8.CrossRefPubMed Moissenet F, Chèze L, Dumas R. A 3D lower limb musculoskeletal model for simultaneous estimation of musculo-tendon, joint contact, ligament and bone forces during gait. J Biomech. 2014; 47(1):50–8.CrossRefPubMed
13.
Zurück zum Zitat Calhoun JH, Li F, Ledbetter BR, Viegas SF. A comprehensive study of pressure distribution in the ankle joint with inversion and eversion. Foot Ankle Int. 1994; 15(3):125–33.CrossRefPubMed Calhoun JH, Li F, Ledbetter BR, Viegas SF. A comprehensive study of pressure distribution in the ankle joint with inversion and eversion. Foot Ankle Int. 1994; 15(3):125–33.CrossRefPubMed
14.
Zurück zum Zitat Matricali GA, Bartels W, Labey L, Dereymaeker GPE, Luyten FP, Vander Sloten J. High inter-specimen variability of baseline data for the tibio-talar contact area. Clin Biomech. 2009; 24(1):117–120.CrossRef Matricali GA, Bartels W, Labey L, Dereymaeker GPE, Luyten FP, Vander Sloten J. High inter-specimen variability of baseline data for the tibio-talar contact area. Clin Biomech. 2009; 24(1):117–120.CrossRef
15.
Zurück zum Zitat Suckel A, Muller O, Wachter N, Kluba T. In vitro measurement of intraarticular pressure in the ankle joint Knee Surgery. Sport Traumatol Arthrosc. 2010; 18(5):664–8.CrossRef Suckel A, Muller O, Wachter N, Kluba T. In vitro measurement of intraarticular pressure in the ankle joint Knee Surgery. Sport Traumatol Arthrosc. 2010; 18(5):664–8.CrossRef
16.
Zurück zum Zitat Anderson AE, Ellis BJ, Maas SA, Peters CL, Weiss JA. Validation of finite element predictions of cartilage contact pressure in the human hip joint. J Biomech Eng. 2008; 130(5):1–10.CrossRef Anderson AE, Ellis BJ, Maas SA, Peters CL, Weiss JA. Validation of finite element predictions of cartilage contact pressure in the human hip joint. J Biomech Eng. 2008; 130(5):1–10.CrossRef
17.
Zurück zum Zitat Potthast W, Lersch C, Segesser B, Koebke J, Brüggemann G-P. Intraarticular pressure distribution in the talocrural joint is related to lower leg muscle forces. Clin Biomech (Bristol, Avon). 2008; 23(5):632–9.CrossRef Potthast W, Lersch C, Segesser B, Koebke J, Brüggemann G-P. Intraarticular pressure distribution in the talocrural joint is related to lower leg muscle forces. Clin Biomech (Bristol, Avon). 2008; 23(5):632–9.CrossRef
18.
Zurück zum Zitat Jung H-G, Parks BG, Nguyen A, Schon LC. Effect of tibiotalar joint arthrodesis on adjacent tarsal joint pressure in a cadaver model. Foot ankle Int Am Orthop Foot Ankle Soc Swiss Foot Ankle Soc. 2007; 28(1):103–8.CrossRef Jung H-G, Parks BG, Nguyen A, Schon LC. Effect of tibiotalar joint arthrodesis on adjacent tarsal joint pressure in a cadaver model. Foot ankle Int Am Orthop Foot Ankle Soc Swiss Foot Ankle Soc. 2007; 28(1):103–8.CrossRef
19.
Zurück zum Zitat Lee DG, Davis BL. Assessment of the effects of diabetes on midfoot joint pressures using a robotic gait simulator. Foot ankle Int Am Orthop Foot Ankle Soc Swiss Foot Ankle Soc. 2009; 30(8):767–72.CrossRef Lee DG, Davis BL. Assessment of the effects of diabetes on midfoot joint pressures using a robotic gait simulator. Foot ankle Int Am Orthop Foot Ankle Soc Swiss Foot Ankle Soc. 2009; 30(8):767–72.CrossRef
20.
Zurück zum Zitat Michelson JD, Checcone M, Kuhn T, Varner K. Intra-articular load distribution in the human ankle joint during motion. Foot Ankle Int. 2001; 22(3):226–33.PubMed Michelson JD, Checcone M, Kuhn T, Varner K. Intra-articular load distribution in the human ankle joint during motion. Foot Ankle Int. 2001; 22(3):226–33.PubMed
21.
Zurück zum Zitat Peeters K, Natsakis T, Burg J, Spaepen P, Jonkers I, Dereymaeker G. An in vitro approach to the evaluation of foot-ankle kinematics: performance evaluation of a custom-built gait simulator. Proc Inst Mech Eng H. 2013; 227(9):955–67.CrossRefPubMed Peeters K, Natsakis T, Burg J, Spaepen P, Jonkers I, Dereymaeker G. An in vitro approach to the evaluation of foot-ankle kinematics: performance evaluation of a custom-built gait simulator. Proc Inst Mech Eng H. 2013; 227(9):955–67.CrossRefPubMed
22.
Zurück zum Zitat Bogey R, Perry J, Gitter A. An EMG-to-force processing approach for determining ankle muscle forces during normal human gait. IEEE Trans Neural Syst Rehabil Eng. 2005; 13(3):302–10.CrossRefPubMed Bogey R, Perry J, Gitter A. An EMG-to-force processing approach for determining ankle muscle forces during normal human gait. IEEE Trans Neural Syst Rehabil Eng. 2005; 13(3):302–10.CrossRefPubMed
23.
Zurück zum Zitat Natsakis T, Peeters K, Burg F, Dereymaeker G, Vander Sloten J, Jonkers I. Specimen-specific tibial kinematics model for in vitro gait simulations. Proc Inst Mech Eng Part H J Eng Med. 2012; 227(4):454–63.CrossRef Natsakis T, Peeters K, Burg F, Dereymaeker G, Vander Sloten J, Jonkers I. Specimen-specific tibial kinematics model for in vitro gait simulations. Proc Inst Mech Eng Part H J Eng Med. 2012; 227(4):454–63.CrossRef
24.
Zurück zum Zitat Natsakis T, Burg J, Dereymaeker G, Jonkers I, Vander Sloten J. Inertial control as novel technique for in vitro gait simulations. J Biomech. 2015; 48(2):392–5.CrossRefPubMed Natsakis T, Burg J, Dereymaeker G, Jonkers I, Vander Sloten J. Inertial control as novel technique for in vitro gait simulations. J Biomech. 2015; 48(2):392–5.CrossRefPubMed
25.
Zurück zum Zitat Okita N, Meyers S, Challis JH, Sharkey N. Midtarsal joint locking: new perspectives on an old paradigm. J Orthop Res. 2014; 32(1):110–5.CrossRefPubMed Okita N, Meyers S, Challis JH, Sharkey N. Midtarsal joint locking: new perspectives on an old paradigm. J Orthop Res. 2014; 32(1):110–5.CrossRefPubMed
Metadaten
Titel
Insertion of a pressure sensing arrayminimally affects hindfoot bone kinematics
verfasst von
Tassos Natsakis
Josefien Burg
Greta Dereymaeker
Ilse Jonkers
Jos Vander Sloten
Publikationsdatum
01.12.2015
Verlag
BioMed Central
Erschienen in
Journal of Foot and Ankle Research / Ausgabe 1/2015
Elektronische ISSN: 1757-1146
DOI
https://doi.org/10.1186/s13047-015-0081-x

Weitere Artikel der Ausgabe 1/2015

Journal of Foot and Ankle Research 1/2015 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Arthroskopie kann Knieprothese nicht hinauszögern

25.04.2024 Gonarthrose Nachrichten

Ein arthroskopischer Eingriff bei Kniearthrose macht im Hinblick darauf, ob und wann ein Gelenkersatz fällig wird, offenbar keinen Unterschied.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.