Iterative reconstruction or filtered backprojection for semi-quantitative assessment of dopamine D2 receptor SPECT studies?

verfasst von: Walter Koch, Christine Suessmair, Klaus Tatsch, Gabriele Pöpperl

Erschienen in: European Journal of Nuclear Medicine and Molecular Imaging | Ausgabe 6/2011

Einloggen, um Zugang zu erhalten

Abstract

Purpose

In routine clinical practice striatal dopamine D2 receptor binding is generally assessed using data reconstructed by filtered backprojection (FBP). The aim of this study was to investigate the use of an iterative reconstruction algorithm (ordered subset expectation maximization, OSEM) and to assess whether it may provide comparable or even better results than those obtained by standard FBP.

Methods

In 56 patients with parkinsonian syndromes, single photon emission computed tomography (SPECT) scans were acquired 2 h after i.v. application of 185 MBq [123I]iodobenzamide (IBZM) using a triple-head gamma camera (Siemens MS 3). The scans were reconstructed both by FBP and OSEM (3 iterations, 8 subsets) and filtered using a Butterworth filter. After attenuation correction the studies were automatically fitted to a mean template with a corresponding 3-D volume of interest (VOI) map covering striatum (S), caudate (C), putamen (P) and several reference VOIs using BRASS software.

Results

Visual assessment of the fitted studies suggests a better separation between C and P in studies reconstructed by OSEM than FBP. Unspecific background activity appears more homogeneous after iterative reconstruction. The correlation shows a good accordance of dopamine receptor binding using FBP and OSEM (intra-class correlation coefficients S: 0.87; C: 0.88; P: 0.84). Receiver-operating characteristic (ROC) analyses show comparable diagnostic power of OSEM and FBP in the differentiation between idiopathic parkinsonian syndrome (IPS) and non-IPS.

Conclusion

Iterative reconstruction of IBZM SPECT studies for assessment of the D2 receptors is feasible in routine clinical practice. Close correlations between FBP and OSEM data suggest that iteratively reconstructed IBZM studies allow reliable quantification of dopamine receptor binding even though a gain in diagnostic power could not be demonstrated.
Literatur
1.
Zurück zum Zitat Vaamonde J, Ibañez R, García A, Poblete V. Study of the pre and post-synaptic dopaminergic system by DaTSCAN/IBZM SPECT in the differential diagnosis of parkinsonism in 75 patients. Neurologia 2004;19:292–300.PubMed Vaamonde J, Ibañez R, García A, Poblete V. Study of the pre and post-synaptic dopaminergic system by DaTSCAN/IBZM SPECT in the differential diagnosis of parkinsonism in 75 patients. Neurologia 2004;19:292–300.PubMed
2.
Zurück zum Zitat Tissingh G, Booij J, Winogrodzka A, van Royen EA, Wolters EC. IBZM- and CIT-SPECT of the dopaminergic system in parkinsonism. J Neural Transm Suppl 1997;50:31–7.PubMed Tissingh G, Booij J, Winogrodzka A, van Royen EA, Wolters EC. IBZM- and CIT-SPECT of the dopaminergic system in parkinsonism. J Neural Transm Suppl 1997;50:31–7.PubMed
3.
Zurück zum Zitat Cordes M, Hierholzer J, Schelosky L, Schrag A, Richter WS, Eichstädt H, et al. Iodine-123-iodo-lisuride SPECT in Parkinson’s disease. J Nucl Med 1996;37:22–5.PubMed Cordes M, Hierholzer J, Schelosky L, Schrag A, Richter WS, Eichstädt H, et al. Iodine-123-iodo-lisuride SPECT in Parkinson’s disease. J Nucl Med 1996;37:22–5.PubMed
4.
Zurück zum Zitat Hatton RL, Hutton BF, Angelides S, Choong KK, Larcos G. Improved tolerance to missing data in myocardial perfusion SPET using OSEM reconstruction. Eur J Nucl Med Mol Imaging 2004;31:857–61.PubMedCrossRef Hatton RL, Hutton BF, Angelides S, Choong KK, Larcos G. Improved tolerance to missing data in myocardial perfusion SPET using OSEM reconstruction. Eur J Nucl Med Mol Imaging 2004;31:857–61.PubMedCrossRef
5.
Zurück zum Zitat Starck SA, Ohlsson J, Carlsson S. An evaluation of reconstruction techniques and scatter correction in bone SPECT of the spine. Nucl Med Commun 2003;24:565–70.PubMedCrossRef Starck SA, Ohlsson J, Carlsson S. An evaluation of reconstruction techniques and scatter correction in bone SPECT of the spine. Nucl Med Commun 2003;24:565–70.PubMedCrossRef
6.
Zurück zum Zitat Kauppinen T, Koskinen MO, Alenius S, Vanninen E, Kuikka JT. Improvement of brain perfusion SPET using iterative reconstruction with scatter and non-uniform attenuation correction. Eur J Nucl Med 2000;27:1380–6.PubMedCrossRef Kauppinen T, Koskinen MO, Alenius S, Vanninen E, Kuikka JT. Improvement of brain perfusion SPET using iterative reconstruction with scatter and non-uniform attenuation correction. Eur J Nucl Med 2000;27:1380–6.PubMedCrossRef
7.
Zurück zum Zitat Koch W, Hamann C, Welsch J, Pöpperl G, Radau PE, Tatsch K. Is iterative reconstruction an alternative to filtered backprojection in routine processing of dopamine transporter SPECT studies? J Nucl Med 2005;46:1804–11.PubMed Koch W, Hamann C, Welsch J, Pöpperl G, Radau PE, Tatsch K. Is iterative reconstruction an alternative to filtered backprojection in routine processing of dopamine transporter SPECT studies? J Nucl Med 2005;46:1804–11.PubMed
8.
Zurück zum Zitat Pareto D, Cot A, Pavía J, Falcón C, Juvells I, Lomeña F, et al. Iterative reconstruction with correction of the spatially variant fan-beam collimator response in neurotransmission SPET imaging. Eur J Nucl Med Mol Imaging 2003;30:1322–9.PubMedCrossRef Pareto D, Cot A, Pavía J, Falcón C, Juvells I, Lomeña F, et al. Iterative reconstruction with correction of the spatially variant fan-beam collimator response in neurotransmission SPET imaging. Eur J Nucl Med Mol Imaging 2003;30:1322–9.PubMedCrossRef
9.
Zurück zum Zitat Dickson JC, Tossici-Bolt L, Sera T, Erlandsson K, Varrone A, Tatsch K, et al. The impact of reconstruction method on the quantification of DaTSCAN images. Eur J Nucl Med Mol Imaging 2010;37:23–35.PubMedCrossRef Dickson JC, Tossici-Bolt L, Sera T, Erlandsson K, Varrone A, Tatsch K, et al. The impact of reconstruction method on the quantification of DaTSCAN images. Eur J Nucl Med Mol Imaging 2010;37:23–35.PubMedCrossRef
10.
Zurück zum Zitat Catafau AM, Bullich S, Danús M, Penengo MM, Cot A, Abanades S, et al. Test-retest variability and reliability of 123I-IBZM SPECT measurement of striatal dopamine D2 receptor availability in healthy volunteers and influence of iterative reconstruction algorithms. Synapse 2008;62:62–9.PubMedCrossRef Catafau AM, Bullich S, Danús M, Penengo MM, Cot A, Abanades S, et al. Test-retest variability and reliability of 123I-IBZM SPECT measurement of striatal dopamine D2 receptor availability in healthy volunteers and influence of iterative reconstruction algorithms. Synapse 2008;62:62–9.PubMedCrossRef
11.
Zurück zum Zitat Bullich S, Cot A, Gallego J, Gunn RN, Suáarez M, Pavía J, et al. Impact of scatter correction on D2 receptor occupancy measurements using 123I-IBZM SPECT: comparison to 11C-raclopride PET. Neuroimage 2010;50:1511–8.PubMedCrossRef Bullich S, Cot A, Gallego J, Gunn RN, Suáarez M, Pavía J, et al. Impact of scatter correction on D2 receptor occupancy measurements using 123I-IBZM SPECT: comparison to 11C-raclopride PET. Neuroimage 2010;50:1511–8.PubMedCrossRef
12.
Zurück zum Zitat Tatsch K, Asenbaum S, Bartenstein P, Catafau A, Halldin C, Pilowsky LS, et al. European Association of Nuclear Medicine procedure guidelines for brain neurotransmission SPET using (123)I-labelled dopamine D(2) transporter ligands. Eur J Nucl Med Mol Imaging 2002;29:BP30–5.PubMedCrossRef Tatsch K, Asenbaum S, Bartenstein P, Catafau A, Halldin C, Pilowsky LS, et al. European Association of Nuclear Medicine procedure guidelines for brain neurotransmission SPET using (123)I-labelled dopamine D(2) transporter ligands. Eur J Nucl Med Mol Imaging 2002;29:BP30–5.PubMedCrossRef
13.
Zurück zum Zitat Tatsch K, Asenbaum S, Bartenstein P, Catafau A, Halldin C, Pilowsky LS, et al. European Association of Nuclear Medicine procedure guidelines for brain neurotransmission SPET using (123)I-labelled dopamine D(2) receptor ligands. Eur J Nucl Med Mol Imaging 2002;29:BP23–9.PubMedCrossRef Tatsch K, Asenbaum S, Bartenstein P, Catafau A, Halldin C, Pilowsky LS, et al. European Association of Nuclear Medicine procedure guidelines for brain neurotransmission SPET using (123)I-labelled dopamine D(2) receptor ligands. Eur J Nucl Med Mol Imaging 2002;29:BP23–9.PubMedCrossRef
14.
Zurück zum Zitat Hudson H, Larkin R. Accelerated image reconstruction using ordered subsets of projection data. IEEE Trans Med Imaging 1994;13:601–9.PubMedCrossRef Hudson H, Larkin R. Accelerated image reconstruction using ordered subsets of projection data. IEEE Trans Med Imaging 1994;13:601–9.PubMedCrossRef
15.
Zurück zum Zitat Koch W, Radau PE, Hamann C, Tatsch K. Clinical testing of an optimized software solution for an automated, observer-independent evaluation of dopamine transporter SPECT studies. J Nucl Med 2005;46:1109–18.PubMed Koch W, Radau PE, Hamann C, Tatsch K. Clinical testing of an optimized software solution for an automated, observer-independent evaluation of dopamine transporter SPECT studies. J Nucl Med 2005;46:1109–18.PubMed
17.
Zurück zum Zitat Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;1:307–10.PubMedCrossRef Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;1:307–10.PubMedCrossRef
18.
Zurück zum Zitat Seibyl JP, Marek K, Sheff K, Baldwin RM, Zoghbi S, Zea-Ponce Y, et al. Test/retest reproducibility of iodine-123-betaCIT SPECT brain measurement of dopamine transporters in Parkinson’s patients. J Nucl Med 1997;38:1453–9.PubMed Seibyl JP, Marek K, Sheff K, Baldwin RM, Zoghbi S, Zea-Ponce Y, et al. Test/retest reproducibility of iodine-123-betaCIT SPECT brain measurement of dopamine transporters in Parkinson’s patients. J Nucl Med 1997;38:1453–9.PubMed
19.
Zurück zum Zitat Gilbert P. Iterative methods for the three-dimensional reconstruction of an object from projections. J Theor Biol 1972;36:105–17.PubMedCrossRef Gilbert P. Iterative methods for the three-dimensional reconstruction of an object from projections. J Theor Biol 1972;36:105–17.PubMedCrossRef
20.
Zurück zum Zitat Gordon R, Bender R, Herman GT. Algebraic reconstruction techniques (ART) for three-dimensional electron microscopy and x-ray photography. J Theor Biol 1970;29:471–81.PubMedCrossRef Gordon R, Bender R, Herman GT. Algebraic reconstruction techniques (ART) for three-dimensional electron microscopy and x-ray photography. J Theor Biol 1970;29:471–81.PubMedCrossRef
21.
Zurück zum Zitat Shepp L, Vardi Y. Maximum likelihood reconstruction for emission tomography. IEEE Trans Med Imaging 1982;1:113–22.PubMedCrossRef Shepp L, Vardi Y. Maximum likelihood reconstruction for emission tomography. IEEE Trans Med Imaging 1982;1:113–22.PubMedCrossRef
22.
Zurück zum Zitat Nuyts J, Dupont P, Stroobants S, Maes A, Mortelmans L, Suetens P. Evaluation of maximum-likelihood based attenuation correction in positron emission tomography. IEEE Trans Nucl Sci 1999;46:1136–41.CrossRef Nuyts J, Dupont P, Stroobants S, Maes A, Mortelmans L, Suetens P. Evaluation of maximum-likelihood based attenuation correction in positron emission tomography. IEEE Trans Nucl Sci 1999;46:1136–41.CrossRef
23.
Zurück zum Zitat Lange K, Carson R. EM reconstruction algorithms for emission and transmission tomography. J Comput Assist Tomogr 1984;8:306–16.PubMed Lange K, Carson R. EM reconstruction algorithms for emission and transmission tomography. J Comput Assist Tomogr 1984;8:306–16.PubMed
24.
Zurück zum Zitat Kauppinen T, Yang J, Kilpeläinen H, Kuikka JT. Quantitation of neuroreceptors: a need for better SPECT imaging. Nuklearmedizin 2001;40:102–6.PubMed Kauppinen T, Yang J, Kilpeläinen H, Kuikka JT. Quantitation of neuroreceptors: a need for better SPECT imaging. Nuklearmedizin 2001;40:102–6.PubMed
25.
Zurück zum Zitat Pretorius PH, King MA, Pan TS, de Vries DJ, Glick SJ, Byrne CL. Reducing the influence of the partial volume effect on SPECT activity quantitation with 3D modelling of spatial resolution in iterative reconstruction. Phys Med Biol 1998;43:407–20.PubMedCrossRef Pretorius PH, King MA, Pan TS, de Vries DJ, Glick SJ, Byrne CL. Reducing the influence of the partial volume effect on SPECT activity quantitation with 3D modelling of spatial resolution in iterative reconstruction. Phys Med Biol 1998;43:407–20.PubMedCrossRef
26.
Zurück zum Zitat Beekman FJ, Kamphuis C, Frey EC. Scatter compensation methods in 3D iterative SPECT reconstruction: a simulation study. Phys Med Biol 1997;42:1619–32.PubMedCrossRef Beekman FJ, Kamphuis C, Frey EC. Scatter compensation methods in 3D iterative SPECT reconstruction: a simulation study. Phys Med Biol 1997;42:1619–32.PubMedCrossRef
27.
Zurück zum Zitat Hutton BF, Hudson HM, Beekman FJ. A clinical perspective of accelerated statistical reconstruction. Eur J Nucl Med 1997;24:797–808.PubMed Hutton BF, Hudson HM, Beekman FJ. A clinical perspective of accelerated statistical reconstruction. Eur J Nucl Med 1997;24:797–808.PubMed
28.
Zurück zum Zitat Sjörgreen K, Ljungberg M, Strand SE. Parameters influencing volume and activity quantitation in SPECT. Acta Oncol 1996;35:323–30.PubMedCrossRef Sjörgreen K, Ljungberg M, Strand SE. Parameters influencing volume and activity quantitation in SPECT. Acta Oncol 1996;35:323–30.PubMedCrossRef
29.
Zurück zum Zitat Van Laere KJ, Warwick J, Versijpt J, Goethals I, Audenaert K, Van Heerden B, et al. Analysis of clinical brain SPECT data based on anatomic standardization and reference to normal data: an ROC-based comparison of visual, semiquantitative, and voxel-based methods. J Nucl Med 2002;43:458–69.PubMed Van Laere KJ, Warwick J, Versijpt J, Goethals I, Audenaert K, Van Heerden B, et al. Analysis of clinical brain SPECT data based on anatomic standardization and reference to normal data: an ROC-based comparison of visual, semiquantitative, and voxel-based methods. J Nucl Med 2002;43:458–69.PubMed
30.
Zurück zum Zitat Radau PE, Linke R, Slomka PJ, Tatsch K. Optimization of automated quantification of 123I-IBZM uptake in the striatum applied to parkinsonism. J Nucl Med 2000;41:220–7.PubMed Radau PE, Linke R, Slomka PJ, Tatsch K. Optimization of automated quantification of 123I-IBZM uptake in the striatum applied to parkinsonism. J Nucl Med 2000;41:220–7.PubMed
31.
Zurück zum Zitat Radau PE, Slomka PJ, Julin P, Svensson L, Wahlund LO. Evaluation of linear registration algorithms for brain SPECT and the errors due to hypoperfusion lesions. Med Phys 2001;28:1660–8.PubMedCrossRef Radau PE, Slomka PJ, Julin P, Svensson L, Wahlund LO. Evaluation of linear registration algorithms for brain SPECT and the errors due to hypoperfusion lesions. Med Phys 2001;28:1660–8.PubMedCrossRef
32.
Zurück zum Zitat Slomka PJ, Radau P, Hurwitz GA, Dey D. Automated three-dimensional quantification of myocardial perfusion and brain SPECT. Comput Med Imaging Graph 2001;25:153–64.PubMedCrossRef Slomka PJ, Radau P, Hurwitz GA, Dey D. Automated three-dimensional quantification of myocardial perfusion and brain SPECT. Comput Med Imaging Graph 2001;25:153–64.PubMedCrossRef
33.
Zurück zum Zitat Tatsch K. Can SPET imaging of dopamine uptake sites replace PET imaging in Parkinson’s disease? For. Eur J Nucl Med Mol Imaging 2002;29:711–4.PubMedCrossRef Tatsch K. Can SPET imaging of dopamine uptake sites replace PET imaging in Parkinson’s disease? For. Eur J Nucl Med Mol Imaging 2002;29:711–4.PubMedCrossRef
34.
Zurück zum Zitat Van Laere K, Koole M, Kauppinen T, Monsieurs M, Bouwens L, Dierck R. Nonuniform transmission in brain SPECT using 201Tl, 153Gd, and 99mTc static line sources: anthropomorphic dosimetry studies and influence on brain quantification. J Nucl Med 2000;41:2051–62.PubMed Van Laere K, Koole M, Kauppinen T, Monsieurs M, Bouwens L, Dierck R. Nonuniform transmission in brain SPECT using 201Tl, 153Gd, and 99mTc static line sources: anthropomorphic dosimetry studies and influence on brain quantification. J Nucl Med 2000;41:2051–62.PubMed
Metadaten
Titel
Iterative reconstruction or filtered backprojection for semi-quantitative assessment of dopamine D2 receptor SPECT studies?
verfasst von
Walter Koch
Christine Suessmair
Klaus Tatsch
Gabriele Pöpperl
Publikationsdatum
01.06.2011
Verlag
Springer-Verlag
Erschienen in
European Journal of Nuclear Medicine and Molecular Imaging / Ausgabe 6/2011
Print ISSN: 1619-7070
Elektronische ISSN: 1619-7089
DOI
https://doi.org/10.1007/s00259-011-1737-9

Weitere Artikel der Ausgabe 6/2011

European Journal of Nuclear Medicine and Molecular Imaging 6/2011 Zur Ausgabe