Large fixed-size metal-on-metal total hip arthroplasty: higher serum metal ion levels in patients with pain

verfasst von: Christiaan Smeekes, Bastiaan Ongkiehong, Bart van der Wal, Ron Wolterbeek, Jan-Ferdinand Henseler, Rob Nelissen

Erschienen in: International Orthopaedics | Ausgabe 4/2015

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Recently, concerns have arisen about metal-on-metal (MoM) total hip arthroplasty (THA). Therefore, the purpose of this cross-sectional cohort study was to describe the incidence of pain, pseudotumours, revisions and the relation between elevated metal ion levels, functional outcome and quality of life after MoM THA.

Methods

In 351 patients, 377 MoM THA with a fixed-size 38-mm head were evaluated with a mean follow-up of 30 months (range 11–58). Evaluation included pain, serum metal ions, patient-reported questionnaires (Short Form-36 [SF-36], Hip disability and Osteoarthritis Outcome Score [HOOS] and the Oxford Hip Score [OHS]) and radiological imaging. Sixteen patients did not participate in the screening.

Results

One hundred and eighteen (35 %) patients reported pain and showed significantly higher cobalt and chromium levels compared to patients without pain. Median serum cobalt levels were 4.4 μg/l (interquartile range [IQR] 6.6) and chromium levels were 3.6 μg/l (IQR 4.8). Patients with cobalt levels of ≤5 μg/l reported significantly better outcome on the SF-36 and HOOS. Fifty-seven pseudotumours were identified in 227 THAs. A revision rate of 19 % was observed.

Conclusions

In conclusion, 35 % of the patients experienced pain after MoM THA. These patients showed significantly higher serum metal ion levels. The patient-reported questionnaires indicated significantly better outcome in patients with cobalt levels ≤5 μg/l.
Literatur
2.
Zurück zum Zitat Takenaga RK, Callaghan JJ, Bedard NA, Liu SS, Gao Y (2013) Which functional assessments predict long-term wear after total hip arthroplasty. Clin Orthop Relat Res 8:2586–2594. doi:10.1007/s11999-013-2968-9 CrossRef Takenaga RK, Callaghan JJ, Bedard NA, Liu SS, Gao Y (2013) Which functional assessments predict long-term wear after total hip arthroplasty. Clin Orthop Relat Res 8:2586–2594. doi:10.​1007/​s11999-013-2968-9 CrossRef
3.
4.
Zurück zum Zitat Goldsmith AA, Dowson D, Isaac GH, Lancaster JG (2000) A comparative joint simulator study of the wear of metal-on-metal and alternative material combinations in hip replacements. Proc Inst Mech Eng H 1:39–47CrossRef Goldsmith AA, Dowson D, Isaac GH, Lancaster JG (2000) A comparative joint simulator study of the wear of metal-on-metal and alternative material combinations in hip replacements. Proc Inst Mech Eng H 1:39–47CrossRef
5.
Zurück zum Zitat Zijlstra WP, van den Akken-Scheek I, Zee MJ, van Raay JJ (2011) No clinical difference between large metal-on-metal total hip arthroplasty and 28-mm-head total hip arthroplasty. Int Orthop 12:1771–1776. doi:10.1007/s00264-011-1233-7 CrossRef Zijlstra WP, van den Akken-Scheek I, Zee MJ, van Raay JJ (2011) No clinical difference between large metal-on-metal total hip arthroplasty and 28-mm-head total hip arthroplasty. Int Orthop 12:1771–1776. doi:10.​1007/​s00264-011-1233-7 CrossRef
6.
Zurück zum Zitat Back DL, Dalziel R, Young D, Shimmin A (2005) Early results of primary Birmingham hip resurfacings. An independent prospective study of the first 230 hips. J Bone Joint Surg Br 3:324–329CrossRef Back DL, Dalziel R, Young D, Shimmin A (2005) Early results of primary Birmingham hip resurfacings. An independent prospective study of the first 230 hips. J Bone Joint Surg Br 3:324–329CrossRef
7.
Zurück zum Zitat Innmann MM, Gotterbarm T, Kretzer JP, Merle C, Ewerbeck V, Weiss S et al (2014) Minimum ten-year results of a 28-mm metal-on-metal bearing in cementless total hip arthroplasty in patients fifty years of age and younger. Int Orthop 5:929–934. doi:10.1007/s00264-013-2228-3 CrossRef Innmann MM, Gotterbarm T, Kretzer JP, Merle C, Ewerbeck V, Weiss S et al (2014) Minimum ten-year results of a 28-mm metal-on-metal bearing in cementless total hip arthroplasty in patients fifty years of age and younger. Int Orthop 5:929–934. doi:10.​1007/​s00264-013-2228-3 CrossRef
8.
Zurück zum Zitat Bosker BH, Ettema HB, Boomsma MF, Kollen BJ, Maas M, Verheyen CC (2012) High incidence of pseudotumour formation after large-diameter metal-on-metal total hip replacement: a prospective cohort study. J Bone Joint Surg Br 6:755–761. doi:10.1302/0301-620X.94B6.28373 CrossRef Bosker BH, Ettema HB, Boomsma MF, Kollen BJ, Maas M, Verheyen CC (2012) High incidence of pseudotumour formation after large-diameter metal-on-metal total hip replacement: a prospective cohort study. J Bone Joint Surg Br 6:755–761. doi:10.​1302/​0301-620X.​94B6.​28373 CrossRef
9.
Zurück zum Zitat Prentice JR, Clark MJ, Hoggard N, Morton AC, Tooth C, Paley MN et al (2013) Metal-on-metal hip prostheses and systemic health: a cross-sectional association study 8 years after implantation. PLoS One 6:e66186. doi:10.1371/journal.pone.0066186 CrossRef Prentice JR, Clark MJ, Hoggard N, Morton AC, Tooth C, Paley MN et al (2013) Metal-on-metal hip prostheses and systemic health: a cross-sectional association study 8 years after implantation. PLoS One 6:e66186. doi:10.​1371/​journal.​pone.​0066186 CrossRef
10.
11.
Zurück zum Zitat Smith AJ, Dieppe P, Vernon K, Porter M, Blom AW (2012) Failure rates of stemmed metal-on-metal hip replacements: analysis of data from the National Joint Registry of England and Wales. Lancet 9822:1199–1204. doi:10.1016/S0140-6736(12)60353-5 CrossRef Smith AJ, Dieppe P, Vernon K, Porter M, Blom AW (2012) Failure rates of stemmed metal-on-metal hip replacements: analysis of data from the National Joint Registry of England and Wales. Lancet 9822:1199–1204. doi:10.​1016/​S0140-6736(12)60353-5 CrossRef
12.
Zurück zum Zitat Meding JB, Meding LK, Keating EM, Berend ME (2012) Low incidence of groin pain and early failure with large metal articulation total hip arthroplasty. Clin Orthop Relat Res 2:388–394. doi:10.1007/s11999-011-2069-6 CrossRef Meding JB, Meding LK, Keating EM, Berend ME (2012) Low incidence of groin pain and early failure with large metal articulation total hip arthroplasty. Clin Orthop Relat Res 2:388–394. doi:10.​1007/​s11999-011-2069-6 CrossRef
13.
Zurück zum Zitat Keurentjes JC, Blane D, Bartley M, Keurentjes JJ, Fiocco M, Nelissen RG (2013) Socio-economic position has no effect on improvement in health-related quality of life and patient satisfaction in total hip and knee replacement: a cohort study. PLoS One 3:e56785. doi:10.1371/journal.pone.0056785 CrossRef Keurentjes JC, Blane D, Bartley M, Keurentjes JJ, Fiocco M, Nelissen RG (2013) Socio-economic position has no effect on improvement in health-related quality of life and patient satisfaction in total hip and knee replacement: a cohort study. PLoS One 3:e56785. doi:10.​1371/​journal.​pone.​0056785 CrossRef
14.
Zurück zum Zitat de Groot IB, Reijman M, Terwee CB, Bierma-Zeinstra S, Favejee MM, Roos E et al (2009) Validation of the Dutch version of the Hip disability and Osteoarthritis Outcome Score. Osteoarthritis Cartilage 1:132. doi:10.1016/j.joca.2008.05.014 CrossRef de Groot IB, Reijman M, Terwee CB, Bierma-Zeinstra S, Favejee MM, Roos E et al (2009) Validation of the Dutch version of the Hip disability and Osteoarthritis Outcome Score. Osteoarthritis Cartilage 1:132. doi:10.​1016/​j.​joca.​2008.​05.​014 CrossRef
15.
Zurück zum Zitat Ware JE Jr, Keller SD, Hatoum HT, Kong SX (1999) The SF-36 Arthritis-Specific Health Index (ASHI): I. Development and cross-validation of scoring algorithms. Med Care 5(Suppl):MS40–MS50 Ware JE Jr, Keller SD, Hatoum HT, Kong SX (1999) The SF-36 Arthritis-Specific Health Index (ASHI): I. Development and cross-validation of scoring algorithms. Med Care 5(Suppl):MS40–MS50
16.
Zurück zum Zitat Gosens T, Hoefnagels NH, de Vet RC, Dhert WJ, van Langelaan EJ, Bulstra SK et al (2005) The “Oxford Heup Score”: the translation and validation of a questionnaire into Dutch to evaluate the results of total hip arthroplasty. Acta Orthop 2:204–211CrossRef Gosens T, Hoefnagels NH, de Vet RC, Dhert WJ, van Langelaan EJ, Bulstra SK et al (2005) The “Oxford Heup Score”: the translation and validation of a questionnaire into Dutch to evaluate the results of total hip arthroplasty. Acta Orthop 2:204–211CrossRef
17.
Zurück zum Zitat Hart AJ, Sabah SA, Bandi AS, Maggiore P, Tarassoli P, Sampson B et al (2011) Sensitivity and specificity of blood cobalt and chromium metal ions for predicting failure of metal-on-metal hip replacement. J Bone Joint Surg Br 10:1308–1313. doi:10.1302/0301-620X.93B10.26249 CrossRef Hart AJ, Sabah SA, Bandi AS, Maggiore P, Tarassoli P, Sampson B et al (2011) Sensitivity and specificity of blood cobalt and chromium metal ions for predicting failure of metal-on-metal hip replacement. J Bone Joint Surg Br 10:1308–1313. doi:10.​1302/​0301-620X.​93B10.​26249 CrossRef
18.
Zurück zum Zitat Anderson H, Toms AP, Cahir JG, Goodwin RW, Wimhurst J, Nolan JF (2011) Grading the severity of soft tissue changes associated with metal-on-metal hip replacements: reliability of an MR grading system. Skeletal Radiol 3:303–307. doi:10.1007/s00256-010-1000-7 CrossRef Anderson H, Toms AP, Cahir JG, Goodwin RW, Wimhurst J, Nolan JF (2011) Grading the severity of soft tissue changes associated with metal-on-metal hip replacements: reliability of an MR grading system. Skeletal Radiol 3:303–307. doi:10.​1007/​s00256-010-1000-7 CrossRef
19.
Zurück zum Zitat Hauptfleisch J, Pandit H, Grammatopoulos G, Gill HS, Murray DW, Ostlere S (2012) A MRI classification of periprosthetic soft tissue masses (pseudotumours) associated with metal-on-metal resurfacing hip arthroplasty. Skeletal Radiol 2:149–155. doi:10.1007/s00256-011-1329-6 CrossRef Hauptfleisch J, Pandit H, Grammatopoulos G, Gill HS, Murray DW, Ostlere S (2012) A MRI classification of periprosthetic soft tissue masses (pseudotumours) associated with metal-on-metal resurfacing hip arthroplasty. Skeletal Radiol 2:149–155. doi:10.​1007/​s00256-011-1329-6 CrossRef
20.
Zurück zum Zitat Lavigne M, Laffosse JM, Ganapathi M, Girard J, Vendittoli P (2011) Residual groin pain at a minimum of two years after metal-on-metal THA with a twenty-eight-millimeter femoral head, THA with a large-diameter femoral head, and hip resurfacing. J Bone Joint Surg Am 93:93–98. doi:10.2106/JBJS.J.01711 CrossRefPubMed Lavigne M, Laffosse JM, Ganapathi M, Girard J, Vendittoli P (2011) Residual groin pain at a minimum of two years after metal-on-metal THA with a twenty-eight-millimeter femoral head, THA with a large-diameter femoral head, and hip resurfacing. J Bone Joint Surg Am 93:93–98. doi:10.​2106/​JBJS.​J.​01711 CrossRefPubMed
23.
Zurück zum Zitat Nilsdotter AK, Lohmander LS, Klassbo M, Roos EM (2003) Hip disability and osteoarthritis outcome score (HOOS)—validity and responsiveness in total hip replacement. BMC Musculoskelet Disord:10. doi:10.1186/1471-2474-4-10 Nilsdotter AK, Lohmander LS, Klassbo M, Roos EM (2003) Hip disability and osteoarthritis outcome score (HOOS)—validity and responsiveness in total hip replacement. BMC Musculoskelet Disord:10. doi:10.​1186/​1471-2474-4-10
24.
Zurück zum Zitat van der Weegen W, Smolders JM, Sijbesma T, Hoekstra HJ, Brakel K, van Susante JL (2012) High incidence of pseudotumours after hip resurfacing even in low risk patients; results from an intensified MRI screening protocol. Hip Int:0. doi:10.5301/hipint.5000004 van der Weegen W, Smolders JM, Sijbesma T, Hoekstra HJ, Brakel K, van Susante JL (2012) High incidence of pseudotumours after hip resurfacing even in low risk patients; results from an intensified MRI screening protocol. Hip Int:0. doi:10.​5301/​hipint.​5000004
25.
Zurück zum Zitat Keegan GM, Learmonth ID, Case CP (2008) A systematic comparison of the actual, potential, and theoretical health effects of cobalt and chromium exposures from industry and surgical implants. Crit Rev Toxicol 8:645–674. doi:10.1080/10408440701845534 CrossRef Keegan GM, Learmonth ID, Case CP (2008) A systematic comparison of the actual, potential, and theoretical health effects of cobalt and chromium exposures from industry and surgical implants. Crit Rev Toxicol 8:645–674. doi:10.​1080/​1040844070184553​4 CrossRef
26.
Zurück zum Zitat Zeng Y, Feng W, Li J, Lu L, Ma C, Zeng J et al (2014) A prospective study concerning the relationship between metal allergy and post-operative pain following total hip and knee arthroplasty. Int Orthop. doi:10.1007/s00264-014-2367-1 Zeng Y, Feng W, Li J, Lu L, Ma C, Zeng J et al (2014) A prospective study concerning the relationship between metal allergy and post-operative pain following total hip and knee arthroplasty. Int Orthop. doi:10.​1007/​s00264-014-2367-1
29.
Zurück zum Zitat Wimmer MA, Fischer A, Buscher R, Pourzal R, Sprecher C, Hauert R et al (2010) Wear mechanisms in metal-on-metal bearings: the importance of tribochemical reaction layers. J Orthop Res 4:436–443. doi:10.1002/jor.21020 Wimmer MA, Fischer A, Buscher R, Pourzal R, Sprecher C, Hauert R et al (2010) Wear mechanisms in metal-on-metal bearings: the importance of tribochemical reaction layers. J Orthop Res 4:436–443. doi:10.​1002/​jor.​21020
31.
Metadaten
Titel
Large fixed-size metal-on-metal total hip arthroplasty: higher serum metal ion levels in patients with pain
verfasst von
Christiaan Smeekes
Bastiaan Ongkiehong
Bart van der Wal
Ron Wolterbeek
Jan-Ferdinand Henseler
Rob Nelissen
Publikationsdatum
01.04.2015
Verlag
Springer Berlin Heidelberg
Erschienen in
International Orthopaedics / Ausgabe 4/2015
Print ISSN: 0341-2695
Elektronische ISSN: 1432-5195
DOI
https://doi.org/10.1007/s00264-014-2605-6

Weitere Artikel der Ausgabe 4/2015

International Orthopaedics 4/2015 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Arthroskopie kann Knieprothese nicht hinauszögern

25.04.2024 Gonarthrose Nachrichten

Ein arthroskopischer Eingriff bei Kniearthrose macht im Hinblick darauf, ob und wann ein Gelenkersatz fällig wird, offenbar keinen Unterschied.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.