Erschienen in: Cardiology and Therapy 1/2017

Open Access 02.01.2017 | Case Report

Late Paravalvular Aortic Regurgitation: Migration of the Valve or Late Recoil?

verfasst von: Ashkan Karimi, Negiin Pourafshar, Ki E. Park, Calvin Y. Choi, Kiran Mogali, Wade W. Stinson, Eddie W. Manning, Anthony A. Bavry

Erschienen in: Cardiology and Therapy | Ausgabe 1/2017

download
DOWNLOAD
print
DRUCKEN
insite
SUCHEN

Abstract

A 79-year-old man underwent trans-catheter aortic valve replacement for symptomatic severe aortic stenosis with a 26-mm Edwards SAPIEN XT valve. Immediately after valve deployment there was moderate amount of paravalvular leak. Post-dilation was performed with an additional 2 cc of volume, and the paravalvular leak was reduced to trace. Nine months later, trans-thoracic echocardiography revealed moderate to severe paravalvular leak and possible aortic migration of the valve. The patient was brought back for the treatment of the paravalvular leak which was suspected to be due to valve migration. However, fluoroscopy and trans-esophageal echocardiography showed good valve position. Measurement of late valve recoil in the Coplanar view using cine-angiographic analysis software showed that the lower third of the valve had the greatest late recoil (−1.74 mm, 6.55%), which presumably accounted for the progression of the paravalvular leak. Valve-in-valve trans-catheter aortic valve replacement was performed with a 26-mm SAPIEN 3 valve and the paravalvular leak was reduced to trace. This case displays late recoil as a likely mechanism for development of paravalvular leak after SAPIEN XT valve implantation. Our case illustrates that late recoil needs to be systematically evaluated in future studies, especially when trans-catheter aortic valve replacement is being expanded to lower risk and younger patients for whom the longevity and long-term performance of these valves is of critical importance.
Hinweise

Enhanced content

To view enhanced content for this article go to http://​www.​medengine.​com/​Redeem/​0237F0602E31F4A7​.

Introduction

Paravalvular leak (PVL) has been described as the Achilles’ heel of trans-catheter aortic valve replacement (TAVR). Moderate or greater degree of PVL has been clearly associated with increased mortality, and some studies have even shown an association between mild PVL and all-cause mortality [13]. What happens to PVL during the follow-up period is a matter of debate. Early experience with TAVR raised the concern that progression of PVL could lead to volume overload of the left ventricle resulting in heart failure symptoms or hemolysis and consequent blood transfusion necessitating re-intervention. In the first PARTNER trial, which used the first-generation SAPIEN valve (Edwards Lifesciences Inc., Irvine, CA), PVL had worsened by ≥1 grade at 2 years of follow-up in 22.4% of patients [1, 2]. Like other vascular scaffolds, such as coronary stents, the supporting stent scaffold of the TAVR valves is subject to acute and late recoil, and late recoil has been postulated as a potential mechanism of progression of PVL; however, this mechanism has not yet been described in the literature.

Case Report

A 79-year-old man with a history of hypertension, diabetes, prior coronary bypass, atrial fibrillation, carotid artery stenosis, colon cancer, and frailty who had been turned down for surgical AVR at his referring hospital underwent TAVR for symptomatic severe aortic stenosis with a 26-mm SAPIEN XT valve (Edwards Lifesciences Inc.). Computed tomography prior to the procedure showed minimal amount of aortic annular calcification—the calcium volume was 1670 mm3 and mainly deposited in the valve leaflets (Fig. 1a). The aortic valve annular area measured 459 mm2, which corresponded to 15.7% oversizing. Immediately after the SAPIEN XT valve deployment there was moderate amount of PVL (Enhanced Content Video 1). Acute valve recoil was measured at three levels in the Coplanar view by subtracting valve diameters immediately after balloon deflation from the valve diameters at maximum balloon inflation. Measurements were performed with the image processing software from the Philips Xcelera package (Philips Medical Systems, Veenpluis, The Netherlands). Using this software, calibration was performed on a straight segment of a contrast-filled pigtail catheter several centimeters from its tip using the known diameter of the catheter based on its French size. After calibration, the valve scaffold was measured from outer edge to outer edge in the coplanar view at three levels: high, mid, and low level. The numbers were reported as absolute values and as the percentage of the valve diameter at maximum balloon inflation. The valve had −0.45 mm (1.68%) recoil at the higher level, −1.01 mm (3.77%) recoil at the mid-level, and −0.66 mm (2.49%) recoil at the lower level (Fig. 1b). Post-dilation was performed with 2 cc of extra volume, and the PVL was reduced to trace (Enchanced Content Video 2). Acute valve recoil was re-measured and the valve had −1.1 mm (4%) recoil at the higher level, −1.02 mm (3.75%) recoil at the mid-level, and −0.14 mm (0.52%) recoil at the lower level (Fig. 1c). Nine months later, a trans-thoracic echocardiography performed for atypical chest pain showed moderate to severe PVL and possible aortic migration of the valve (Enhanced Content Video 3). The patient was brought back for the treatment of PVL which was suspected to be due to valve migration. Trans-esophageal echocardiography confirmed severe PVL and showed three separate regurgitation jets in the mid-esophageal short-axis view (lower left inset of Video 4 in Enhanced Content). However, fluoroscopy and trans-esophageal echocardiography showed good valve position and no evidence of valve migration (Fig. 2, Video 4). To discover the mechanism for PVL progression we decided to measure the amount of late recoil. To this end, we obtained three angiographic cine images in the coplanar view, in the same projection that the initial TAVR had been performed. Valve diameters were measured at three levels from these films and averaged (Fig. 1d). Late recoil was calculated by subtracting the average numbers at each level from valve diameters after balloon deflation at the post-dilation phase of the initial TAVR (middle table in Fig. 1c). Late recoil was measured to be −0.91 mm (3.45%) at the higher level, −1.44 mm (5.5%) at the mid-level, and −1.74 mm (6.55%) at the lower level of the valve (Fig. 1d). Therefore, the lower level of the valve showed the greatest amount of late recoil, which in our opinion was the mechanism for progression of the PVL. The summation of acute and late recoil at 12 months was (4.00 + 3.45 = 7.45%) at the higher level, (3.75 + 5.50 = 9.25%) at the mid-level, and (0.52 + 6.55 = 7.07%) at the lower level of the valve. To treat the PVL, valve-in-valve TAVR was performed with a 26-mm SAPIEN 3 valve (Edwards Lifesciences Inc.) and the PVL was reduced to trace (Enhanced Content Video 5).
Informed consent was obtained from the patient for receiving treatment.

Discussion

The incidence of PVL has been reduced significantly as a result of more accurate valve sizing based on computed tomography measurements, availability of the 29-mm valve in addition to the 23-mm valve, and the 26-mm valve as a third size option, improved operators’ skills, and innovative valve designs. The upshot of studies that have reported degree of PVL at different follow-up points is that the incidence of moderate or greater PVL decreases during the follow-up period in most patients [4]. The reduction in incidence of moderate or greater PVL in the follow-up could be due to increased mortality of patients who had or developed moderate to severe PVL or it could be related to remodeling of the aortic valve annulus. Echocardiographic analysis of patients participating in the PARTNER trial, which used the first-generation SAPIEN valve, showed that at 2 years of follow-up PVL remained the same in 46.2% of patients, improved in 31.5%, and worsened by ≥1 grade in 22.4% [1, 2]. Studies have not specifically looked at the progression of PVL with the newer generation SAPIEN XT or SAPEIN 3 valves over time, but PVL remains an important complication of TAVR. In Cohort A of the PARTNER II trial, which used the second-generation SAPIEN XT valve in intermediate-risk patients, moderate to severe PVL was reported in 3.7% of patients at 30 days of follow-up and was associated with a hazard ratio of 2.85 for all-cause mortality in the follow-up period compared to patients with no PVL or trace PVL [5]. In Cohort B of the PARTNER II trial, which compared SAPIEN XT to the first-generation SAPIEN valve in inoperable patients, the reported incidence of moderate to severe PVL with the SAPIEN XT valve was markedly higher than that reported in other studies—24.2% at 30 days of follow-up and 27.5% at 1 year [6]. The higher rate of PVL in that study was speculated to be due to the ambiguity and difficulty in quantifying PVL and the tendency of the echocardiography core laboratory to overestimate its amount. It has however been postulated that acute and late recoil are at least partially responsible for the development and progression of PVL.
To date, only two studies have systemically assessed acute recoil of the Edwards balloon-expandable valves. Both studies only included the first-generation SAPIEN valve and the second-generation SAPIEN XT valve, and both reported higher degree of acute recoil with the SAPIEN XT valve [7, 8].
The stent scaffold of the SAPIEN XT valve is made of cobalt–chromium (CoCr) alloy as opposed to stainless steel which is used in the first-generation SAPIEN valve, but the stent struts of the SAPIEN XT valve are thinner than those of the older generation SAPIEN valve. For the same strut thickness, CoCr alloy is stiffer and more resistant to recoil than stainless steel. This quality is exploited to reduce the thickness of the stent struts and the crimped valve profile to allow delivery of the valve through a smaller sheath. Thinner stent struts, however, reduce the stiffness of the valve scaffold, making it more prone to recoil. Additionally, the SAPIEN XT stent scaffold possesses three rungs of struts instead of the four rungs present in the older generation SAPIEN valve, which also makes it more susceptible to recoil. To our knowledge, late recoil has not been previously investigated or measurably reported nor has acute recoil of the SAPIEN 3 valve been studied. The SAPIEN 3 stent scaffold is made of CoCr alloy and consists of five rungs of struts with a wider strut angle. These characteristics provide higher radial strength to keep circularity compared to the SAPIEN XT valve (internal data provided by Edward Lifesciences Inc.) [9, 10] and likely make the SAPIEN 3 valve less prone to recoil compared to the SAPIEN XT valve; however, the degree of acute or late recoil of the SAPIEN 3 valve has not yet been studied in vivo. Nevertheless, we postulate recoil would have less effect on PVL of the SAPIEN 3 valve, given the presence of the outer skirt, and therefore we chose to treat our patient with the SAPIEN 3 valve. In a large observational study of slightly over 1000 patients, the SAPIEN 3 valve was associated with 3.4% incidence of moderate to severe PVL at the 30-day follow-up [11] and only 2% at 1 year [12].
It is also worthwhile mentioning that although we initially suspected aortic migration of the valve as the mechanism for progression of the PVL, late aortic migration of TAVR valves is an extremely rare phenomenon, and late migration of the TAVR valve is most commonly retrograde towards the left ventricle as opposed to towards the aorta. Interestingly, a computational fluid dynamics model showed that the antegrade force exerted on the TAVR valve during ejection is tenfold weaker than the retrograde force applied to it in diastole, which would explain why late migration tends to occur towards the left ventricular outflow tract [13].

Conclusion

Acute and late recoil are mechanisms for development and progression of PVL. Studies have shown an average 3.75–5.18% of acute recoil associated with the SAPIEN XT valve depending on the valve size or the study [7, 8]. This case report shows late recoil as the likely mechanism for progression of PVL during follow-up after a 26-mm SAPIEN XT implantation and in summation shows 7.07–9.25% of acute plus late recoil at different valve levels at 1 year of follow-up. The effect of valve recoil on PVL has not been previously studied. Future studies need to systematically evaluate and report acute and late recoil and how they affect PVL.

Acknowledgements

No funding or sponsorship was received for this study or publication of this article.
All named authors meet the International Committee of Medical Journal Editors (ICMJE) criteria for authorship for this manuscript, take responsibility for the integrity of the work as a whole, and have given final approval for the version to be published.

Disclosures

Anthony Bavry is a contractor for the American College of Cardiology and has received travel support from Edwards.
Ashkan Karimi, Negiin Pourafsha, Ki Park, Calvin Choi, Kiran Mogali, Wade Stinson and Eddie Manning do not have any financial disclosure relevant to the content of this case report.

Compliance with Ethics Guidelines

Informed consent was obtained from the patient for receiving treatment.

Open Access

This article is distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 International License (http://​creativecommons.​org/​licenses/​by-nc/​4.​0/​), which permits any noncommercial use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (https://​creativecommons.​org/​licenses/​by/​4.​0), which permits use, duplication, adaptation, distribution, and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
Literatur
1.
Zurück zum Zitat Genereux P, Head SJ, Hahn R, et al. Paravalvular leak after transcatheter aortic valve replacement: the new Achilles’ heel? A comprehensive review of the literature. J Am Coll Cardiol. 2013;61(11):1125–36.CrossRefPubMed Genereux P, Head SJ, Hahn R, et al. Paravalvular leak after transcatheter aortic valve replacement: the new Achilles’ heel? A comprehensive review of the literature. J Am Coll Cardiol. 2013;61(11):1125–36.CrossRefPubMed
2.
Zurück zum Zitat Kodali SK, Williams MR, Smith CR, et al. Two-year outcomes after transcatheter or surgical aortic-valve replacement. N Engl J Med. 2012;366(18):1686–95.CrossRefPubMed Kodali SK, Williams MR, Smith CR, et al. Two-year outcomes after transcatheter or surgical aortic-valve replacement. N Engl J Med. 2012;366(18):1686–95.CrossRefPubMed
3.
Zurück zum Zitat Mack MJ, Leon MB, Smith CR, et al. 5-year outcomes of transcatheter aortic valve replacement or surgical aortic valve replacement for high surgical risk patients with aortic stenosis (PARTNER 1): a randomised controlled trial. Lancet. 2015;385(9986):2477–84.CrossRefPubMed Mack MJ, Leon MB, Smith CR, et al. 5-year outcomes of transcatheter aortic valve replacement or surgical aortic valve replacement for high surgical risk patients with aortic stenosis (PARTNER 1): a randomised controlled trial. Lancet. 2015;385(9986):2477–84.CrossRefPubMed
4.
Zurück zum Zitat Lerakis S, Hayek SS, Douglas PS. Paravalvular aortic leak after transcatheter aortic valve replacement: current knowledge. Circulation. 2013;127(3):397–407.CrossRefPubMed Lerakis S, Hayek SS, Douglas PS. Paravalvular aortic leak after transcatheter aortic valve replacement: current knowledge. Circulation. 2013;127(3):397–407.CrossRefPubMed
5.
Zurück zum Zitat Leon MB, Smith CR, Mack MJ, et al. Transcatheter or surgical aortic-valve replacement in intermediate-risk patients. N Engl J Med. 2016;374(17):1609–20.CrossRefPubMed Leon MB, Smith CR, Mack MJ, et al. Transcatheter or surgical aortic-valve replacement in intermediate-risk patients. N Engl J Med. 2016;374(17):1609–20.CrossRefPubMed
6.
Zurück zum Zitat Webb JG, Doshi D, Mack MJ, et al. A randomized evaluation of the SAPIEN XT transcatheter heart valve system in patients with aortic stenosis who are not candidates for surgery. JACC Cardiovasc Interv. 2015;8(14):1797–806.CrossRefPubMed Webb JG, Doshi D, Mack MJ, et al. A randomized evaluation of the SAPIEN XT transcatheter heart valve system in patients with aortic stenosis who are not candidates for surgery. JACC Cardiovasc Interv. 2015;8(14):1797–806.CrossRefPubMed
7.
Zurück zum Zitat Garg A, Parashar A, Agarwal S, et al. Comparison of acute elastic recoil between the SAPIEN-XT and SAPIEN valves in transfemoral-transcatheter aortic valve replacement. Catheter Cardiovasc Interv. 2015;85(3):490–6.CrossRefPubMed Garg A, Parashar A, Agarwal S, et al. Comparison of acute elastic recoil between the SAPIEN-XT and SAPIEN valves in transfemoral-transcatheter aortic valve replacement. Catheter Cardiovasc Interv. 2015;85(3):490–6.CrossRefPubMed
8.
Zurück zum Zitat Nombela-Franco L, Ribeiro HB, Urena M, et al. Incidence, predictive factors and haemodynamic consequences of acute stent recoil following transcatheter aortic valve implantation with a balloon-expandable valve. EuroIntervention. 2014;9(12):1398–406.CrossRefPubMed Nombela-Franco L, Ribeiro HB, Urena M, et al. Incidence, predictive factors and haemodynamic consequences of acute stent recoil following transcatheter aortic valve implantation with a balloon-expandable valve. EuroIntervention. 2014;9(12):1398–406.CrossRefPubMed
10.
Zurück zum Zitat Schymik G, Schrofel H, Heimeshoff M, Luik A, Thoenes M, Mandinov L. How to adapt the implantation technique for the new SAPIEN 3 transcatheter heart valve design. J Interv Cardiol. 2015;28(1):82–9.CrossRefPubMed Schymik G, Schrofel H, Heimeshoff M, Luik A, Thoenes M, Mandinov L. How to adapt the implantation technique for the new SAPIEN 3 transcatheter heart valve design. J Interv Cardiol. 2015;28(1):82–9.CrossRefPubMed
11.
Zurück zum Zitat Kodali S, Thourani VH, White J, et al. Early clinical and echocardiographic outcomes after SAPIEN 3 transcatheter aortic valve replacement in inoperable, high-risk, and intermediate-risk patients with aortic stenosis. Eur Heart J. 2016;37(28):2252–62. Kodali S, Thourani VH, White J, et al. Early clinical and echocardiographic outcomes after SAPIEN 3 transcatheter aortic valve replacement in inoperable, high-risk, and intermediate-risk patients with aortic stenosis. Eur Heart J. 2016;37(28):2252–62.
12.
Zurück zum Zitat Thourani VH, Kodali S, Makkar RR, et al. Transcatheter aortic valve replacement versus surgical valve replacement in intermediate-risk patients: a propensity score analysis. Lancet. 2016;387(10034):2218–225. Thourani VH, Kodali S, Makkar RR, et al. Transcatheter aortic valve replacement versus surgical valve replacement in intermediate-risk patients: a propensity score analysis. Lancet. 2016;387(10034):2218–225.
13.
Zurück zum Zitat Dwyer HA, Matthews PB, Azadani A, Ge L, Guy TS, Tseng EE. Migration forces of transcatheter aortic valves in patients with noncalcific aortic insufficiency. J Thorac Cardiovasc Surg. 2009;138(5):1227–33.CrossRefPubMed Dwyer HA, Matthews PB, Azadani A, Ge L, Guy TS, Tseng EE. Migration forces of transcatheter aortic valves in patients with noncalcific aortic insufficiency. J Thorac Cardiovasc Surg. 2009;138(5):1227–33.CrossRefPubMed
Metadaten
Titel
Late Paravalvular Aortic Regurgitation: Migration of the Valve or Late Recoil?
verfasst von
Ashkan Karimi
Negiin Pourafshar
Ki E. Park
Calvin Y. Choi
Kiran Mogali
Wade W. Stinson
Eddie W. Manning
Anthony A. Bavry
Publikationsdatum
02.01.2017
Verlag
Springer Healthcare
Erschienen in
Cardiology and Therapy / Ausgabe 1/2017
Print ISSN: 2193-8261
Elektronische ISSN: 2193-6544
DOI
https://doi.org/10.1007/s40119-016-0077-z

Weitere Artikel der Ausgabe 1/2017

Cardiology and Therapy 1/2017 Zur Ausgabe

Leitlinien kompakt für die Innere Medizin

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

Update Innere Medizin

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.