Radiation exposure during transmission measurements: comparison between CT- and germanium-based techniques with a current PET scanner

verfasst von: Tung-Hsin Wu, Yung-Hui Huang, Jason J. S. Lee, Shih-Yuan Wang, Su-Cheng Wang, Cheng-Tau Su, Liang-Kung Chen, Tieh-Chi Chu

Erschienen in: European Journal of Nuclear Medicine and Molecular Imaging | Ausgabe 1/2004

Einloggen, um Zugang zu erhalten

Abstract

In positron emission tomographic (PET) scanning, transmission measurements for attenuation correction are commonly performed by using external germanium-68 rod sources. Recently, combined PET and computed tomographic (CT) scanners have been developed in which the CT data can be used for both anatomical-metabolic image formation and attenuation correction of the PET data. The purpose of this study was to evaluate the difference between germanium- and CT-based transmission scanning in terms of their radiation doses by using the same measurement technique and to compare the doses that patients receive during brain, cardiac and whole-body scans. Measurement of absorbed doses to organs was conducted by using a Rando Alderson phantom with thermoluminescent dosimeters. Effective doses were calculated according to the guidelines in the International Commission on Radiation Protection Publication Number 60. Compared with radionuclide doses used in routine 2-[fluorine-18]-fluoro-2-deoxy-d-glucose PET imaging, doses absorbed during germanium-based transmission scans were almost negligible. On the other hand, absorbed doses from CT-based transmission scans were significantly higher, particularly with a whole-body scanning protocol. Effective doses were 8.81 mSv in the high-speed mode and 18.97 mSv in the high-quality mode for whole-body CT-based transmission scans. These measurements revealed that the doses received by a patient during CT-based transmission scanning are more than those received in a typical PET examination. Therefore, the radiation doses represent a limitation to the generalised use of CT-based transmission measurements with current PET/CT scanner systems.
Literatur
1.
2.
Zurück zum Zitat Lonneux M, Borbath I, Bol A, et al. Attenuation correction in whole-body FDG oncological studies: the role of statistical reconstruction. Eur J Nucl Med 1999; 26:591–598.PubMed Lonneux M, Borbath I, Bol A, et al. Attenuation correction in whole-body FDG oncological studies: the role of statistical reconstruction. Eur J Nucl Med 1999; 26:591–598.PubMed
3.
Zurück zum Zitat Bengel FM, Ziegler SI, Avril N, Weber W, Laubenbacher C, Schwaiger M. Whole body positron emission tomography in clinical oncology: comparison between attenuation corrected and uncorrected images. Eur J Nucl Med 1997; 24:1091–1098.PubMed Bengel FM, Ziegler SI, Avril N, Weber W, Laubenbacher C, Schwaiger M. Whole body positron emission tomography in clinical oncology: comparison between attenuation corrected and uncorrected images. Eur J Nucl Med 1997; 24:1091–1098.PubMed
4.
Zurück zum Zitat Visvikis D, Costa DC, Croasdale I, Lonn AHR, Bomanji J, Gacinovic S, Ell PJ. CT-based attenuation correction in the calculation of semi-quantitative indices of [18F]FDG uptake in PET. Eur J Nucl Med Mol Imaging 2003; 30:344–353.PubMed Visvikis D, Costa DC, Croasdale I, Lonn AHR, Bomanji J, Gacinovic S, Ell PJ. CT-based attenuation correction in the calculation of semi-quantitative indices of [18F]FDG uptake in PET. Eur J Nucl Med Mol Imaging 2003; 30:344–353.PubMed
5.
Zurück zum Zitat Kluetz PG, Meltzer CC, Villemagne VL, et al. Combined PET/CT imaging in oncology: impact on patient management. Clin Positron Imaging 2000; 3:223–230.PubMed Kluetz PG, Meltzer CC, Villemagne VL, et al. Combined PET/CT imaging in oncology: impact on patient management. Clin Positron Imaging 2000; 3:223–230.PubMed
6.
Zurück zum Zitat Charron M, Beyer T, Bohnen NN, et al. Image analysis in patients with cancer studied with a combined PET and CT scanner. Clin Nucl Med 2000; 25:905–910.PubMed Charron M, Beyer T, Bohnen NN, et al. Image analysis in patients with cancer studied with a combined PET and CT scanner. Clin Nucl Med 2000; 25:905–910.PubMed
7.
Zurück zum Zitat Townsend DW. A combined PET/CT scanner: the choices. J Nucl Med 2001; 42:533–534. Townsend DW. A combined PET/CT scanner: the choices. J Nucl Med 2001; 42:533–534.
8.
Zurück zum Zitat Israel O, Mor M, Gaitini D, et al. Combined functional and structural evaluation of cancer patients with a hybrid camera-based PET/CT system using18F-FDG. J Nucl Med 2002; 43:1129–1136.PubMed Israel O, Mor M, Gaitini D, et al. Combined functional and structural evaluation of cancer patients with a hybrid camera-based PET/CT system using18F-FDG. J Nucl Med 2002; 43:1129–1136.PubMed
9.
Zurück zum Zitat Kamel EM, Goerres GW, Burger C, von Schulthess GK, Steinert HC. Recurrent laryngeal nerve palsy in patients with lung cancer: detection with PET-CT image fusion—report of six cases. Radiology 2002; 224:153–156.PubMed Kamel EM, Goerres GW, Burger C, von Schulthess GK, Steinert HC. Recurrent laryngeal nerve palsy in patients with lung cancer: detection with PET-CT image fusion—report of six cases. Radiology 2002; 224:153–156.PubMed
10.
Zurück zum Zitat Kamel E, Hany TF, Burger C, et al. CT vs68Ge attenuation correction in a combined PET/CT system: evaluation of the effect of lowering the CT tube current. Eur J Nucl Med Mol Imaging 2002; 29:346–350.PubMed Kamel E, Hany TF, Burger C, et al. CT vs68Ge attenuation correction in a combined PET/CT system: evaluation of the effect of lowering the CT tube current. Eur J Nucl Med Mol Imaging 2002; 29:346–350.PubMed
11.
Zurück zum Zitat Nakamoto Y, Osman M, Cohade C, Marshall LT, Links JM, Kohlmyer S, Wahl RL. PET/CT: comparison of quantitative tracer uptake between germanium and CT transmission attenuation-corrected images. J Nucl Med 2002; 43:1137–1143. Nakamoto Y, Osman M, Cohade C, Marshall LT, Links JM, Kohlmyer S, Wahl RL. PET/CT: comparison of quantitative tracer uptake between germanium and CT transmission attenuation-corrected images. J Nucl Med 2002; 43:1137–1143.
12.
Zurück zum Zitat Beyer T, Townsend DW, Brun T, et al. A combined PET/CT scanner for clinical oncology. J Nucl Med 2000; 41:1369–1379.PubMed Beyer T, Townsend DW, Brun T, et al. A combined PET/CT scanner for clinical oncology. J Nucl Med 2000; 41:1369–1379.PubMed
13.
Zurück zum Zitat Hany TF, Steinert HC, Goerres GW, Buck A, von Schulthess GK. PET diagnostic accuracy: improvement with in-line PET-CT system: initial results. Radiology 2002; 225:575–581.PubMed Hany TF, Steinert HC, Goerres GW, Buck A, von Schulthess GK. PET diagnostic accuracy: improvement with in-line PET-CT system: initial results. Radiology 2002; 225:575–581.PubMed
14.
Zurück zum Zitat Cao Z, Tsui BM. Performance characteristics of transmission imaging using a uniform sheet source with parallel hole collimaton. Med Phys 1992; 19:1205–1212.CrossRefPubMed Cao Z, Tsui BM. Performance characteristics of transmission imaging using a uniform sheet source with parallel hole collimaton. Med Phys 1992; 19:1205–1212.CrossRefPubMed
15.
Zurück zum Zitat Prvulovich EM, Lonn AHR, Bomanji JB, Jarrit PH, Ell PJ. Effect of attenuation correction on myocardial thallium-201 distribution in patients with a low likelihood of coronary artery disease. Eur J Nucl Med 1997; 24:266–275.CrossRefPubMed Prvulovich EM, Lonn AHR, Bomanji JB, Jarrit PH, Ell PJ. Effect of attenuation correction on myocardial thallium-201 distribution in patients with a low likelihood of coronary artery disease. Eur J Nucl Med 1997; 24:266–275.CrossRefPubMed
16.
Zurück zum Zitat Hashimoto J, Kubo A, Ogawa K, Amano T, Fukuuchi Y, Motomura N, Ichiara T. Scatter and attenuation correction in technetium-99m brain SPET. J Nucl Med 1997; 38:157–162.PubMed Hashimoto J, Kubo A, Ogawa K, Amano T, Fukuuchi Y, Motomura N, Ichiara T. Scatter and attenuation correction in technetium-99m brain SPET. J Nucl Med 1997; 38:157–162.PubMed
17.
Zurück zum Zitat Almeida P, Bendriem B, de Dreuille O, Peltier A, Perrot C, Brulon V. Dosimetry of transmission measurements in nuclear medicine: a study using anthropomorphic phantoms and thermoluminescent dosimeters. Eur J Nucl Med 1998; 25:1435–1441.CrossRefPubMed Almeida P, Bendriem B, de Dreuille O, Peltier A, Perrot C, Brulon V. Dosimetry of transmission measurements in nuclear medicine: a study using anthropomorphic phantoms and thermoluminescent dosimeters. Eur J Nucl Med 1998; 25:1435–1441.CrossRefPubMed
18.
Zurück zum Zitat Burger C, Goerres G, Schoenes S, Buck A, Lonn AHR, von Schulthess GK. PET attenuation coefficients from CT images: experimental evaluation of the transformation of CT into PET 511 keV attenuation coefficients. Eur J Nucl Med Mol Imaging 2002; 29:922–927.CrossRefPubMed Burger C, Goerres G, Schoenes S, Buck A, Lonn AHR, von Schulthess GK. PET attenuation coefficients from CT images: experimental evaluation of the transformation of CT into PET 511 keV attenuation coefficients. Eur J Nucl Med Mol Imaging 2002; 29:922–927.CrossRefPubMed
19.
Zurück zum Zitat Deloar HM, Fujiwara T, Shidahara M, Nakamura T, Yamadera A, Itoh M. Internal absorbed dose estimation by a TLD method for18F-FDG and comparison with dose estimates from whole body PET. Phys Med Biol 1999; 44:595–606.CrossRefPubMed Deloar HM, Fujiwara T, Shidahara M, Nakamura T, Yamadera A, Itoh M. Internal absorbed dose estimation by a TLD method for18F-FDG and comparison with dose estimates from whole body PET. Phys Med Biol 1999; 44:595–606.CrossRefPubMed
20.
Zurück zum Zitat Jurik AG, Jensen LC, Hansen J. Radiation dose by spiral CT and conventional tomography of the sternoclavicular joints and the manubrium sterni. Skeletal Radiol 1996; 25:467–470.CrossRefPubMed Jurik AG, Jensen LC, Hansen J. Radiation dose by spiral CT and conventional tomography of the sternoclavicular joints and the manubrium sterni. Skeletal Radiol 1996; 25:467–470.CrossRefPubMed
21.
Zurück zum Zitat ICRP Publication Number 60. Recommendations of the International Commission on Radiation Protection. Oxford: Pergamon Press, 1990. ICRP Publication Number 60. Recommendations of the International Commission on Radiation Protection. Oxford: Pergamon Press, 1990.
22.
Zurück zum Zitat Deloar HM, Fujiwara T, Shidahara M, et al. Estimation of absorbed dose for 2-[F-18]fluoro-2-deoxy-d-glucose using whole-body positron emission tomography and magnetic resonance imaging. Eur J Nucl Med 1998; 25:565–574. Deloar HM, Fujiwara T, Shidahara M, et al. Estimation of absorbed dose for 2-[F-18]fluoro-2-deoxy-d-glucose using whole-body positron emission tomography and magnetic resonance imaging. Eur J Nucl Med 1998; 25:565–574.
23.
Zurück zum Zitat Dowd MT, Chen CT, Wendel MJ, et al. Radiation dose to the bladder wall from 2-[18F]-Fluoro-2-deoxy-d-glucose in adult humans. J Nucl Med 1991; 32:707–712.PubMed Dowd MT, Chen CT, Wendel MJ, et al. Radiation dose to the bladder wall from 2-[18F]-Fluoro-2-deoxy-d-glucose in adult humans. J Nucl Med 1991; 32:707–712.PubMed
24.
Zurück zum Zitat Thomas SR, Stabin MG, Chen CT, Samaratunga RC. A dynamic urinary bladder model for radiation dose calculations. MIRD pamphlet No 14, revised. J Nucl Med 1999; 40:102S–123S.PubMed Thomas SR, Stabin MG, Chen CT, Samaratunga RC. A dynamic urinary bladder model for radiation dose calculations. MIRD pamphlet No 14, revised. J Nucl Med 1999; 40:102S–123S.PubMed
25.
Zurück zum Zitat Wu TH, Liu RS, Dong SL, Chung YW, Chou KL, Lee JS. Dynamic evaluation of absorbed dose to the bladder wall with a balloon-bladder phantom during a study using [18F]fluorodeoxyglucose positron emission imaging. Nucl Med Commun 2002; 23:749–755.CrossRefPubMed Wu TH, Liu RS, Dong SL, Chung YW, Chou KL, Lee JS. Dynamic evaluation of absorbed dose to the bladder wall with a balloon-bladder phantom during a study using [18F]fluorodeoxyglucose positron emission imaging. Nucl Med Commun 2002; 23:749–755.CrossRefPubMed
26.
Zurück zum Zitat Visvikis D, Cheze-LeRest C, Costa DC, Bomanji J, Gacinovic S, Ell PJ. Influence of OSEM and segmented attenuation correction in the calculation of standardised uptake values for [18F]FDG PET. Eur J Nucl Med 2001; 28:1326–1335.PubMed Visvikis D, Cheze-LeRest C, Costa DC, Bomanji J, Gacinovic S, Ell PJ. Influence of OSEM and segmented attenuation correction in the calculation of standardised uptake values for [18F]FDG PET. Eur J Nucl Med 2001; 28:1326–1335.PubMed
27.
Zurück zum Zitat Xu EZ, Mullani NA, Gould LK, Anderson WL. A segmented attenuation correction for PET. J Nucl Med 1991; 32:161–165.PubMed Xu EZ, Mullani NA, Gould LK, Anderson WL. A segmented attenuation correction for PET. J Nucl Med 1991; 32:161–165.PubMed
Metadaten
Titel
Radiation exposure during transmission measurements: comparison between CT- and germanium-based techniques with a current PET scanner
verfasst von
Tung-Hsin Wu
Yung-Hui Huang
Jason J. S. Lee
Shih-Yuan Wang
Su-Cheng Wang
Cheng-Tau Su
Liang-Kung Chen
Tieh-Chi Chu
Publikationsdatum
01.01.2004
Verlag
Springer-Verlag
Erschienen in
European Journal of Nuclear Medicine and Molecular Imaging / Ausgabe 1/2004
Print ISSN: 1619-7070
Elektronische ISSN: 1619-7089
DOI
https://doi.org/10.1007/s00259-003-1327-6

Weitere Artikel der Ausgabe 1/2004

European Journal of Nuclear Medicine and Molecular Imaging 1/2004 Zur Ausgabe