Cross-camera comparison of SPECT measurements of a 3-D anthropomorphic basal ganglia phantom

verfasst von: Walter Koch, Perry E Radau, Wolfgang Münzing, Klaus Tatsch

Erschienen in: European Journal of Nuclear Medicine and Molecular Imaging | Ausgabe 4/2006

Einloggen, um Zugang zu erhalten

Abstract

Purpose

SPECT examinations of neurotransmitter systems in the brain have to be comparable between centres to generate a comprehensive data pool, e.g. for multicentre studies. Equipment-specific effects on quantitative evaluations and corresponding methods for compensation, however, have been insufficiently examined. Previous studies have shown that quantitative results may vary significantly according to the imaging equipment used, thereby affecting clinical interpretation of the data. The aim of this study was to determine correction factors for common camera/collimator combinations based on standardised measurements of an anthropomorphic 3D basal ganglia phantom to compensate for the effects of different SPECT camera/collimator equipment. The latter may serve as a model for human studies of the dopaminergic system.

Methods

The striatum and background chambers of a commercially available phantom (RSD Alderson) were filled with various 123I concentrations encompassing specific striatum/background ratios from 0.6 to 16.1. This setup was imaged with the following four camera/collimator combinations: Siemens Multispect 3 fitted with LEHR and 123I parallel-hole collimators, Siemens ECAM with LEHR parallel-hole collimators and Philips Prism 3000 fitted with LEHR fanbeam collimators, using standardised protocols for acquisition and reconstruction. All scans were automatically co-registered to a SPECT template of the phantom and quantified using a 3D volume of interest (VOI) map based on a CT scan of the phantom. All striatal/background ratios calculated by SPECT were compared with the true ratios calculated from the measurements in a well counter. Regression analyses were performed and recovery correction factors between measured and true ratios determined.

Results

The relation between true and measured ratios could be sufficiently described by a linear regression for each camera/collimator combination without relevant improvement when using second-order polynomial regression models. The recovery correction factors and standard errors were 2.04±0.04 for the Philips Prism 3000, 2.67±0.03 for the Siemens Multispect 3/LEHR parallel-hole collimators, 2.15±0.03 for the Siemens Multispect 3/123I collimators and 2.81±0.03 for the Siemens ECAM. Percentage recovery ranged from 36% to 49%.

Conclusion

Measurements of a 3D basal ganglia phantom with various imaging devices revealed linear correlations between measured and true striatal/background ratios. Based on these findings, adjustment of quantitative results between different equipment seems possible, provided that acquisition, reconstruction and evaluation are adequately standardised. The use of identical evaluation methods in phantom and patient studies (comparable shape, size and location of the VOIs) might allow transfer of the calculated correction factors from phantom to studies of the dopaminergic system in patients.
Literatur
1.
Zurück zum Zitat Bergmann H, Busemann-Sokole E, Horton PW. Quality assurance and harmonisation of nuclear medicine investigations in Europe. Eur J Nucl Med 1995;22:477–480PubMedCrossRef Bergmann H, Busemann-Sokole E, Horton PW. Quality assurance and harmonisation of nuclear medicine investigations in Europe. Eur J Nucl Med 1995;22:477–480PubMedCrossRef
2.
Zurück zum Zitat Meyer PT, Sattler B, Lincke T, Seese A, Sabri O. Investigating dopaminergic neurotransmission with 123I-FP-CIT SPECT: comparability of modern SPECT systems. J Nucl Med 2003;44:839–845PubMed Meyer PT, Sattler B, Lincke T, Seese A, Sabri O. Investigating dopaminergic neurotransmission with 123I-FP-CIT SPECT: comparability of modern SPECT systems. J Nucl Med 2003;44:839–845PubMed
3.
Zurück zum Zitat Stodilka RZ, Kemp BJ, Prato FS, Nicholson RL. Importance of bone attenuation in brain SPECT quantification. J Nucl Med 1998;39:190–197PubMed Stodilka RZ, Kemp BJ, Prato FS, Nicholson RL. Importance of bone attenuation in brain SPECT quantification. J Nucl Med 1998;39:190–197PubMed
4.
Zurück zum Zitat Tatsch K, Asenbaum S, Bartenstein P, Catafau A, Halldin C, Pilowsky LS, et al European Association of Nuclear Medicine procedure guidelines for brain neurotransmission SPET using 123I-labelled dopamine D2 transporter ligands. Eur J Nucl Med Mol Imaging 2002;29:BP30–BP35PubMedCrossRef Tatsch K, Asenbaum S, Bartenstein P, Catafau A, Halldin C, Pilowsky LS, et al European Association of Nuclear Medicine procedure guidelines for brain neurotransmission SPET using 123I-labelled dopamine D2 transporter ligands. Eur J Nucl Med Mol Imaging 2002;29:BP30–BP35PubMedCrossRef
5.
Zurück zum Zitat Radau PE, Linke R, Slomka PJ, Tatsch K. Optimization of automated quantification of 123I-IBZM uptake in the striatum applied to parkinsonism. J Nucl Med 2000;41:220–227PubMed Radau PE, Linke R, Slomka PJ, Tatsch K. Optimization of automated quantification of 123I-IBZM uptake in the striatum applied to parkinsonism. J Nucl Med 2000;41:220–227PubMed
6.
Zurück zum Zitat Radau PE, Koch W, Holtmannspoetter M, Poepperl G, Tatsch K. Automated, objective software for 3D registration and analysis of dopamine transporter SPECT studies. J Nucl Med 2003;44:P12 Radau PE, Koch W, Holtmannspoetter M, Poepperl G, Tatsch K. Automated, objective software for 3D registration and analysis of dopamine transporter SPECT studies. J Nucl Med 2003;44:P12
7.
Zurück zum Zitat Koch W, Radau PE, Hamann C, Tatsch K. Clinical testing of an optimized software solution for an automated, observer-independent evaluation of dopamine transporter SPECT studies. J Nucl Med 2005;46:1109–1118PubMed Koch W, Radau PE, Hamann C, Tatsch K. Clinical testing of an optimized software solution for an automated, observer-independent evaluation of dopamine transporter SPECT studies. J Nucl Med 2005;46:1109–1118PubMed
8.
Zurück zum Zitat Koch W, Münzing W, Geworski L, Sonnenschein W, Höffken H, Czech N, et al Zentrumübergreifende Messungen an einem Basalganglienphantom zur Vereinheitlichung der SPECT Diagnostik des dopaminergen Systems. In: Brink I, editor. Nuklearmedizin als Paradigma molekularer Bildgebung. Berlin, Germany: Blackwell; 2002. p. 20 Koch W, Münzing W, Geworski L, Sonnenschein W, Höffken H, Czech N, et al Zentrumübergreifende Messungen an einem Basalganglienphantom zur Vereinheitlichung der SPECT Diagnostik des dopaminergen Systems. In: Brink I, editor. Nuklearmedizin als Paradigma molekularer Bildgebung. Berlin, Germany: Blackwell; 2002. p. 20
9.
Zurück zum Zitat Hudson H, Larkin RS. Accelerated image reconstruction using ordered subsets of projection data. IEEE Trans Med Imaging 1994;13:594–600CrossRef Hudson H, Larkin RS. Accelerated image reconstruction using ordered subsets of projection data. IEEE Trans Med Imaging 1994;13:594–600CrossRef
10.
Zurück zum Zitat Kauppinen T, Yang J, Kilpelainen H, Kuikka JT. Quantitation of neuroreceptors: a need for better SPECT imaging. Nuklearmedizin 2001;40:102–106PubMed Kauppinen T, Yang J, Kilpelainen H, Kuikka JT. Quantitation of neuroreceptors: a need for better SPECT imaging. Nuklearmedizin 2001;40:102–106PubMed
11.
Zurück zum Zitat Booij J, Hemelaar TG, Speelman JD, de Bruin K, Janssen AG, van Royen EA. One-day protocol for imaging of the nigrostriatal dopaminergic pathway in Parkinson’s disease by [123I]FPCIT SPECT. J Nucl Med 1999;40:753–761PubMed Booij J, Hemelaar TG, Speelman JD, de Bruin K, Janssen AG, van Royen EA. One-day protocol for imaging of the nigrostriatal dopaminergic pathway in Parkinson’s disease by [123I]FPCIT SPECT. J Nucl Med 1999;40:753–761PubMed
12.
Zurück zum Zitat Pirker W, Asenbaum S, Hauk M, Kandlhofer S, Tauscher J, Willeit M, et al Imaging serotonin and dopamine transporters with 123I-beta-CIT SPECT: binding kinetics and effects of normal aging. J Nucl Med 2000;41:36–44PubMed Pirker W, Asenbaum S, Hauk M, Kandlhofer S, Tauscher J, Willeit M, et al Imaging serotonin and dopamine transporters with 123I-beta-CIT SPECT: binding kinetics and effects of normal aging. J Nucl Med 2000;41:36–44PubMed
13.
Zurück zum Zitat Brucke T, Kornhuber J, Angelberger P, Asenbaum S, Frassine H, Podreka I, et al SPECT imaging of dopamine and serotonin transporters with [123I]beta-CIT. Binding kinetics in the human brain. Imaging serotonin and dopamine transporters with 123I-beta-CIT SPECT: binding kinetics and effects of normal aging. J Neural Transm Gen Sect 1993;94:137–146PubMedCrossRef Brucke T, Kornhuber J, Angelberger P, Asenbaum S, Frassine H, Podreka I, et al SPECT imaging of dopamine and serotonin transporters with [123I]beta-CIT. Binding kinetics in the human brain. Imaging serotonin and dopamine transporters with 123I-beta-CIT SPECT: binding kinetics and effects of normal aging. J Neural Transm Gen Sect 1993;94:137–146PubMedCrossRef
14.
Zurück zum Zitat Geworski L, Knoop BO, de Cabrejas ML, Knapp WH, Munz DL. Recovery correction for quantitation in emission tomography: a feasibility study. Eur J Nucl Med 2000;27:161–169PubMedCrossRef Geworski L, Knoop BO, de Cabrejas ML, Knapp WH, Munz DL. Recovery correction for quantitation in emission tomography: a feasibility study. Eur J Nucl Med 2000;27:161–169PubMedCrossRef
15.
Zurück zum Zitat Tatsch K, Asenbaum S, Bartenstein P, Catafau A, Halldin C, Pilowsky LS, et al European Association of Nuclear medicine procedure guidelines for brain neurotransmission SPET using 123I-labelled dopamine D2 receptor ligands. Eur J Nucl Med Mol Imaging 2002;29:BP23–BP29PubMedCrossRef Tatsch K, Asenbaum S, Bartenstein P, Catafau A, Halldin C, Pilowsky LS, et al European Association of Nuclear medicine procedure guidelines for brain neurotransmission SPET using 123I-labelled dopamine D2 receptor ligands. Eur J Nucl Med Mol Imaging 2002;29:BP23–BP29PubMedCrossRef
16.
Zurück zum Zitat Hamann C, Koch W, Radau PE, Tatsch K. Iterative reconstruction or filtered backprojection for quantitative assessment of dopamine d2 receptor studies? J Nucl Med 2003;44:114P Hamann C, Koch W, Radau PE, Tatsch K. Iterative reconstruction or filtered backprojection for quantitative assessment of dopamine d2 receptor studies? J Nucl Med 2003;44:114P
17.
Zurück zum Zitat Hashimoto J, Kubo A, Ogawa K, Amano T, Fukuuchi Y, Motomura N, et al Scatter and attenuation correction in technetium-99m brain SPECT. J Nucl Med 1997;38:157–162PubMed Hashimoto J, Kubo A, Ogawa K, Amano T, Fukuuchi Y, Motomura N, et al Scatter and attenuation correction in technetium-99m brain SPECT. J Nucl Med 1997;38:157–162PubMed
18.
Zurück zum Zitat Hashimoto J, Sasaki T, Ogawa K, Kubo A, Motomura N, Ichihara T, et al Effects of scatter and attenuation correction on quantitative analysis of beta-CIT brain SPET. Nucl Med Commun 1999;20:159–165PubMedCrossRef Hashimoto J, Sasaki T, Ogawa K, Kubo A, Motomura N, Ichihara T, et al Effects of scatter and attenuation correction on quantitative analysis of beta-CIT brain SPET. Nucl Med Commun 1999;20:159–165PubMedCrossRef
19.
Zurück zum Zitat Soret M, Koulibaly PM, Darcourt J, Hapdey S, Buvat I. Quantitative accuracy of dopaminergic neurotransmission imaging with 123I SPECT. J Nucl Med 2003;44:1184–1193PubMed Soret M, Koulibaly PM, Darcourt J, Hapdey S, Buvat I. Quantitative accuracy of dopaminergic neurotransmission imaging with 123I SPECT. J Nucl Med 2003;44:1184–1193PubMed
20.
Zurück zum Zitat Yang J, Kuikka JT, Vanninen E, Kauppinen T, Lansimies E, Patomaki L. Evaluation of scatter correction using a single isotope for simultaneous emission and transmission data. Phantom and clinical patient studies. Nuklearmedizin 1999;38:49–55PubMed Yang J, Kuikka JT, Vanninen E, Kauppinen T, Lansimies E, Patomaki L. Evaluation of scatter correction using a single isotope for simultaneous emission and transmission data. Phantom and clinical patient studies. Nuklearmedizin 1999;38:49–55PubMed
21.
Zurück zum Zitat Kauppinen T, Koskinen MO, Alenius S, Vanninen E, Kuikka JT. Improvement of brain perfusion SPET using iterative reconstruction with scatter and non-uniform attenuation correction. Eur J Nucl Med 2000;27:1380–1386PubMedCrossRef Kauppinen T, Koskinen MO, Alenius S, Vanninen E, Kuikka JT. Improvement of brain perfusion SPET using iterative reconstruction with scatter and non-uniform attenuation correction. Eur J Nucl Med 2000;27:1380–1386PubMedCrossRef
22.
Zurück zum Zitat Ichihara T, Ogawa K, Motomura N, Kubo A, Hashimoto S. Compton scatter compensation using the triple-energy window method for single- and dual-isotope SPECT. J Nucl Med 1993;34:2216–2221PubMed Ichihara T, Ogawa K, Motomura N, Kubo A, Hashimoto S. Compton scatter compensation using the triple-energy window method for single- and dual-isotope SPECT. J Nucl Med 1993;34:2216–2221PubMed
23.
Zurück zum Zitat El Fakhri G, Buvat I, Benali H, Todd-Pokropek A, Di Paola R. Relative impact of scatter, collimator response, attenuation, and finite spatial resolution corrections in cardiac SPECT. J Nucl Med 2000;41:1400–1408PubMed El Fakhri G, Buvat I, Benali H, Todd-Pokropek A, Di Paola R. Relative impact of scatter, collimator response, attenuation, and finite spatial resolution corrections in cardiac SPECT. J Nucl Med 2000;41:1400–1408PubMed
24.
Zurück zum Zitat Yanch JC, Flower MA, Webb S. Improved quantification of radionuclide uptake using deconvolution and windowed subtraction techniques for scatter compensation in single photon emission computed tomography. Med Phys 1990;17:1011–1022PubMedCrossRef Yanch JC, Flower MA, Webb S. Improved quantification of radionuclide uptake using deconvolution and windowed subtraction techniques for scatter compensation in single photon emission computed tomography. Med Phys 1990;17:1011–1022PubMedCrossRef
25.
Zurück zum Zitat Jaszczak RJ, Greer KL, Floyd CE Jr, Harris CC, Coleman RE. Improved SPECT quantification using compensation for scattered photons. J Nucl Med 1984;25:893–900PubMed Jaszczak RJ, Greer KL, Floyd CE Jr, Harris CC, Coleman RE. Improved SPECT quantification using compensation for scattered photons. J Nucl Med 1984;25:893–900PubMed
26.
Zurück zum Zitat El Fakhri G, Moore SC, Maksud P, Aurengo A, Kijewski MF. Absolute activity quantitation in simultaneous 123I/99mTc brain SPECT. J Nucl Med 2001;42:300–308PubMed El Fakhri G, Moore SC, Maksud P, Aurengo A, Kijewski MF. Absolute activity quantitation in simultaneous 123I/99mTc brain SPECT. J Nucl Med 2001;42:300–308PubMed
27.
Zurück zum Zitat Linke R, Gostomzyk J, Hahn K, Tatsch K. [123I]IPT binding to the presynaptic dopamine transporter: variation of intra- and interobserver data evaluation in parkinsonian patients and controls. Eur J Nucl Med 2000;27:1809–1812PubMedCrossRef Linke R, Gostomzyk J, Hahn K, Tatsch K. [123I]IPT binding to the presynaptic dopamine transporter: variation of intra- and interobserver data evaluation in parkinsonian patients and controls. Eur J Nucl Med 2000;27:1809–1812PubMedCrossRef
28.
Zurück zum Zitat Hamann C, Koch W, Radau PE, Tatsch K. Automated evaluation of dopamine D2 receptor SPECT studies: Benefit in clinical routine? J Nucl Med 2003;44:P261–P262 Hamann C, Koch W, Radau PE, Tatsch K. Automated evaluation of dopamine D2 receptor SPECT studies: Benefit in clinical routine? J Nucl Med 2003;44:P261–P262
Metadaten
Titel
Cross-camera comparison of SPECT measurements of a 3-D anthropomorphic basal ganglia phantom
verfasst von
Walter Koch
Perry E Radau
Wolfgang Münzing
Klaus Tatsch
Publikationsdatum
01.04.2006
Verlag
Springer-Verlag
Erschienen in
European Journal of Nuclear Medicine and Molecular Imaging / Ausgabe 4/2006
Print ISSN: 1619-7070
Elektronische ISSN: 1619-7089
DOI
https://doi.org/10.1007/s00259-005-0036-8

Weitere Artikel der Ausgabe 4/2006

European Journal of Nuclear Medicine and Molecular Imaging 4/2006 Zur Ausgabe