The effect of patient positioning aids on PET quantification in PET/MR imaging

verfasst von: Frederic Mantlik, Matthias Hofmann, Matthias K. Werner, Alexander Sauter, Jürgen Kupferschläger, Bernhard Schölkopf, Bernd J. Pichler, Thomas Beyer

Erschienen in: European Journal of Nuclear Medicine and Molecular Imaging | Ausgabe 5/2011

Einloggen, um Zugang zu erhalten

Abstract

Objectives

Clinical PET/MR requires the use of patient positioning aids to immobilize and support patients for the duration of the combined examination. Ancillary immobilization devices contribute to overall attenuation of the PET signal, but are not detected with conventional MR sequences and, hence, are ignored in standard MR-based attenuation correction (MR-AC). We report on the quantitative effect of not accounting for the attenuation of patient positioning aids in combined PET/MR imaging.

Methods

We used phantom and patient data acquired with positioning aids on a PET/CT scanner (Biograph 16, HI-REZ) to mimic PET/MR imaging conditions. Reference CT-based attenuation maps were generated from measured (original) CT transmission images (origCT-AC). We also created MR-like attenuation maps by following the same conversion procedure of the attenuation values except for the prior delineation and subtraction of the positioning aids from the CT images (modCT-AC). First, a uniform 68Ge cylinder was positioned centrally in the PET/CT scanner and fixed with a vacuum mattress (10 cm thick) and, in a repeat examination, with MR positioning foam pads. Second, 16 patient datasets were selected for subsequent processing. All patients were regionally immobilized with positioning aids: a vacuum mattress for head/neck imaging (nine patients) and a foam mattress for imaging of the lower extremities (seven patients). PET images were reconstructed following CT-based attenuation and scatter correction using the original and modified (MR-like) CT images: PETorigCT-AC and PETmodCT-AC, respectively. PET images following origCT-AC and modCT-AC were compared visually and in terms of mean differences of voxels with a standardized uptake value of at least 1.0. In addition, we report maximum activity concentration in lesions for selected patients.

Results

In the phantom study employing the vacuum mattress the average voxel activity in PETmodCT-AC was underestimated by 6.4% compared to PETorigCT-AC, with 3.4% of the PET voxels being underestimated by 10% or more. When the MR foam pads were not accounted for during AC, PETmodCT-AC was underestimated by 1.1% on average, with none of the PET voxels being underestimated by 10% or more. Evaluation of the head/neck patient data showed a decrease of 8.4% ([68Ga]DOTATOC) and 7.4% ([18F]FDG) when patient positioning aids were not accounted for during AC, while the corresponding decrease was insignificant for the lower extremities.

Conclusion

Depending on the size and density of the positioning aids used, a regionally variable underestimation of PET activity following AC is observed when positioning aids are not accounted for. This underestimation may become relevant in combined PET/MR imaging of patients with neuropsychiatric indications, but appears to be of no clinical relevance in imaging the extremities.
Literatur
3.
Zurück zum Zitat Ratib O, Becker M, Vallee JP, Loubeyre P, Wissmeyer M, Willi J-P, et al. Whole body PET-MRI scanner: first experience in oncology. J Nucl Med. 2010;51 Suppl 2:165. Ratib O, Becker M, Vallee JP, Loubeyre P, Wissmeyer M, Willi J-P, et al. Whole body PET-MRI scanner: first experience in oncology. J Nucl Med. 2010;51 Suppl 2:165.
6.
Zurück zum Zitat Beyer T, Tellmann L, Nickel I, Pietrzyk U. On the use of positioning aids to reduce misregistration in the head and neck in whole-body PET/CT studies. J Nucl Med. 2005;46(4):596–602.PubMed Beyer T, Tellmann L, Nickel I, Pietrzyk U. On the use of positioning aids to reduce misregistration in the head and neck in whole-body PET/CT studies. J Nucl Med. 2005;46(4):596–602.PubMed
9.
Zurück zum Zitat Martinez-Möller A, Souvatzoglou M, Delso G, Bundschuh RA, Chefd'hotel C, Ziegler SI, et al. Tissue classification as a potential approach for attenuation correction in whole-body PET/MRI: evaluation with PET/CT data. J Nucl Med. 2009;50(4):520–6. doi:10.2967/jnumed.108.054726.PubMedCrossRef Martinez-Möller A, Souvatzoglou M, Delso G, Bundschuh RA, Chefd'hotel C, Ziegler SI, et al. Tissue classification as a potential approach for attenuation correction in whole-body PET/MRI: evaluation with PET/CT data. J Nucl Med. 2009;50(4):520–6. doi:10.​2967/​jnumed.​108.​054726.PubMedCrossRef
10.
11.
Zurück zum Zitat Mantlik F, Hofmann M, Kupferschläger J, Werner M, Pichler B, Beyer T. The effect of positioning aids on PET quantification following MR-based attenuation correction (AC) in PET/MR imaging. J Nucl Med. 2010;51 Suppl 2:1418. Mantlik F, Hofmann M, Kupferschläger J, Werner M, Pichler B, Beyer T. The effect of positioning aids on PET quantification following MR-based attenuation correction (AC) in PET/MR imaging. J Nucl Med. 2010;51 Suppl 2:1418.
14.
Zurück zum Zitat Catana C, van der Kouwe A, Benner T, Michel CJ, Hamm M, Fenchel M, et al. Toward implementing an MRI-based PET attenuation-correction method for neurologic studies on the MR-PET brain prototype. J Nucl Med. 2010;51(9):1431–8. doi:10.2967/jnumed.109.069112.PubMedCrossRef Catana C, van der Kouwe A, Benner T, Michel CJ, Hamm M, Fenchel M, et al. Toward implementing an MRI-based PET attenuation-correction method for neurologic studies on the MR-PET brain prototype. J Nucl Med. 2010;51(9):1431–8. doi:10.​2967/​jnumed.​109.​069112.PubMedCrossRef
15.
Zurück zum Zitat Nawaz A, Torigian DA, Siegelman ES, Basu S, Chryssikos T, Alavi A. Diagnostic performance of FDG-PET, MRI, and plain film radiography (PFR) for the diagnosis of osteomyelitis in the diabetic foot. Mol Imaging Biol. 2010;12(3):335–42. doi:10.1007/s11307-009-0268-2.PubMedCrossRef Nawaz A, Torigian DA, Siegelman ES, Basu S, Chryssikos T, Alavi A. Diagnostic performance of FDG-PET, MRI, and plain film radiography (PFR) for the diagnosis of osteomyelitis in the diabetic foot. Mol Imaging Biol. 2010;12(3):335–42. doi:10.​1007/​s11307-009-0268-2.PubMedCrossRef
16.
Zurück zum Zitat Sauter A, Horger M, Boss A, Kolb A, Mantlik F, Kanz L, et al. Simultaneous PET/MRI for the evaluation of hemato-oncological diseases with lower extremity manifestations. J Nucl Med. 2010;51 Suppl 2:1001. Sauter A, Horger M, Boss A, Kolb A, Mantlik F, Kanz L, et al. Simultaneous PET/MRI for the evaluation of hemato-oncological diseases with lower extremity manifestations. J Nucl Med. 2010;51 Suppl 2:1001.
17.
Zurück zum Zitat Ladebeck R. Method for determining attenuation values of an object. US Patent Appl. No. 12/458368, Pub. No. US 2010/0007346 A1. Published Jan 14, 2010. Ladebeck R. Method for determining attenuation values of an object. US Patent Appl. No. 12/458368, Pub. No. US 2010/0007346 A1. Published Jan 14, 2010.
Metadaten
Titel
The effect of patient positioning aids on PET quantification in PET/MR imaging
verfasst von
Frederic Mantlik
Matthias Hofmann
Matthias K. Werner
Alexander Sauter
Jürgen Kupferschläger
Bernhard Schölkopf
Bernd J. Pichler
Thomas Beyer
Publikationsdatum
01.05.2011
Verlag
Springer-Verlag
Erschienen in
European Journal of Nuclear Medicine and Molecular Imaging / Ausgabe 5/2011
Print ISSN: 1619-7070
Elektronische ISSN: 1619-7089
DOI
https://doi.org/10.1007/s00259-010-1721-9

Weitere Artikel der Ausgabe 5/2011

European Journal of Nuclear Medicine and Molecular Imaging 5/2011 Zur Ausgabe