Skip to main content
Erschienen in: International Orthopaedics 4/2014

01.04.2014 | Original Paper

Effect of design on the initial stability of press-fit cups in the presence of acetabular rim defects: experimental evaluation of the effect of adding circumferential fins

verfasst von: Wolfgang O. Huber, Philip C. Noble

Erschienen in: International Orthopaedics | Ausgabe 4/2014

Einloggen, um Zugang zu erhalten

Abstract

Purpose

This experimental study was undertaken to examine the fixation characteristics of a six-finned acetabular cup in both primary and revision arthroplasty in comparison with two other commonly used cup designs without fins.

Methods

All three cup designs (Ananova® [Intraplant], Plasmacup® NSC [Aesculap]; Exceed ABT™ [Biomet]) were implanted into validated models of normal and revision acetabula. The defect models were designed to simulate a dorso-cranial rim defect of 90° width and 10 mm in depth (moderate rim defect) and a dorso-cranial rim defect of 130° width and 15 mm in depth (severe rim defect). The fixation strength of the three cup designs was tested by cyclically edge-loading the implanted cups using a mechanical testing machine.

Results

The six-finned Ananova implant exhibited greater resistance to foam-cup interface motion than both the Plasmacup and Exceed ABT implants. The largest average differences were observed in the resistance to ultimate spin-out, with Ananova outperforming Exceed ABT and Plasmacup by 26 % and 17 % in the moderate and by 36 % and 38 % in the severe defect models, respectively.

Conclusions

The six-finned Ananova cup was significantly more resistant to edge loading both in the normal acetabulum and in acetabula with moderate to severe dorso-cranial rim defects than cup designs without fins, indicating that it may cover a wider range of clinical indications than conventional press-fit cups and provide clinicians with the confidence that, in primary and simple revision arthroplasty, adequate fixation strength can be obtained.
Literatur
1.
Zurück zum Zitat Widmer KH, Zurfluh B, Morscher EW (2002) Load transfer and fixation mode of press-fit acetabular sockets. J Arthroplasty 17(7):926–935PubMedCrossRef Widmer KH, Zurfluh B, Morscher EW (2002) Load transfer and fixation mode of press-fit acetabular sockets. J Arthroplasty 17(7):926–935PubMedCrossRef
2.
Zurück zum Zitat Elke R, Berli B, Wagner A, Morscher EW (2003) Acetabular revision in total hip replacement with a press-fit cup. J Bone Joint Surg Br 85(8):1114–1119PubMedCrossRef Elke R, Berli B, Wagner A, Morscher EW (2003) Acetabular revision in total hip replacement with a press-fit cup. J Bone Joint Surg Br 85(8):1114–1119PubMedCrossRef
3.
Zurück zum Zitat Pulido L, Rachala SR, Cabanela ME (2011) Cementless acetabular revision: past, present, and future. Revision total hip arthroplasty: the acetabular side using cementless implants. Int Orthop 35(2):289–298PubMedCentralPubMedCrossRef Pulido L, Rachala SR, Cabanela ME (2011) Cementless acetabular revision: past, present, and future. Revision total hip arthroplasty: the acetabular side using cementless implants. Int Orthop 35(2):289–298PubMedCentralPubMedCrossRef
4.
Zurück zum Zitat Bellini CM, Galbusera F, Ceroni RG, Raimondi MT (2007) Loss in mechanical contact of cementless acetabular prostheses due to post-operative weight bearing: a biomechanical model. Med Eng Phys 29(2):175–181PubMedCrossRef Bellini CM, Galbusera F, Ceroni RG, Raimondi MT (2007) Loss in mechanical contact of cementless acetabular prostheses due to post-operative weight bearing: a biomechanical model. Med Eng Phys 29(2):175–181PubMedCrossRef
5.
Zurück zum Zitat Pilliar RM, Lee JM, Maniatopoulos C (1986) Observations on the effect of movement on bone ingrowth into porous-surfaced implants. Clin Orthop Relat Res 208:108–113PubMed Pilliar RM, Lee JM, Maniatopoulos C (1986) Observations on the effect of movement on bone ingrowth into porous-surfaced implants. Clin Orthop Relat Res 208:108–113PubMed
6.
Zurück zum Zitat Curtis MJ, Jinnah RH, Wilson VD, Hungerford DS (1992) The initial stability of uncemented acetabular components. J Bone Joint Surg Br 74(3):372–376PubMed Curtis MJ, Jinnah RH, Wilson VD, Hungerford DS (1992) The initial stability of uncemented acetabular components. J Bone Joint Surg Br 74(3):372–376PubMed
7.
Zurück zum Zitat Adler E, Stuchin SA, Kummer FJ (1992) Stability of press-fit acetabular cups. J Arthroplasty 7(3):295–301PubMedCrossRef Adler E, Stuchin SA, Kummer FJ (1992) Stability of press-fit acetabular cups. J Arthroplasty 7(3):295–301PubMedCrossRef
8.
Zurück zum Zitat Kwong LM, O’Connor DO, Sedlacek RC, Krushell RJ, Maloney WJ, Harris WH (1994) A quantitative in vitro assessment of fit and screw fixation on the stability of a cementless hemispherical acetabular component. J Arthroplasty 9(2):163–170PubMedCrossRef Kwong LM, O’Connor DO, Sedlacek RC, Krushell RJ, Maloney WJ, Harris WH (1994) A quantitative in vitro assessment of fit and screw fixation on the stability of a cementless hemispherical acetabular component. J Arthroplasty 9(2):163–170PubMedCrossRef
9.
Zurück zum Zitat Roth A, Winzer T, Sander K, Anders JO, Venbrocks RA (2006) Press fit fixation of cementless cups: how much stability do we need indeed? Arch Orthop Trauma Surg 126(2):77–81PubMedCrossRef Roth A, Winzer T, Sander K, Anders JO, Venbrocks RA (2006) Press fit fixation of cementless cups: how much stability do we need indeed? Arch Orthop Trauma Surg 126(2):77–81PubMedCrossRef
10.
Zurück zum Zitat Larsson S, Elloy M, Hansson LI (1987) Fixation of trochanteric hip fractures. A cadaver study of static and dynamic loading. Acta Orthop Scand 58(4):365–368PubMedCrossRef Larsson S, Elloy M, Hansson LI (1987) Fixation of trochanteric hip fractures. A cadaver study of static and dynamic loading. Acta Orthop Scand 58(4):365–368PubMedCrossRef
11.
Zurück zum Zitat MacKenzie JR, Callaghan JJ, Pedersen DR, Brown TD (1994) Areas of contact and extent of gaps with implantation of oversized acetabular components in total hip arthroplasty. Clin Orthop Relat Res 298:127–136PubMed MacKenzie JR, Callaghan JJ, Pedersen DR, Brown TD (1994) Areas of contact and extent of gaps with implantation of oversized acetabular components in total hip arthroplasty. Clin Orthop Relat Res 298:127–136PubMed
12.
Zurück zum Zitat Morscher EW (1992) Current status of acetabular fixation in primary total hip arthroplasty. Clin Orthop Relat Res 274:172–193PubMed Morscher EW (1992) Current status of acetabular fixation in primary total hip arthroplasty. Clin Orthop Relat Res 274:172–193PubMed
13.
Zurück zum Zitat Schmalzried TP, Wessinger SJ, Hill GE, Harris WH (1994) The Harris-Galante porous acetabular component press-fit without screw fixation. Five-year radiographic analysis of primary cases. J Arthroplasty 9(3):235–242PubMedCrossRef Schmalzried TP, Wessinger SJ, Hill GE, Harris WH (1994) The Harris-Galante porous acetabular component press-fit without screw fixation. Five-year radiographic analysis of primary cases. J Arthroplasty 9(3):235–242PubMedCrossRef
14.
Zurück zum Zitat Won CH, Hearn TC, Tile M (1995) Micromotion of cementless hemispherical acetabular components. Does press-fit need adjunctive screw fixation? J Bone Joint Surg Br 77(3):484–489PubMed Won CH, Hearn TC, Tile M (1995) Micromotion of cementless hemispherical acetabular components. Does press-fit need adjunctive screw fixation? J Bone Joint Surg Br 77(3):484–489PubMed
15.
Zurück zum Zitat Zilkens C, Djalali S, Bittersohl B et al (2011) Migration pattern of cementless press fit cups in the presence of stabilizing screws in total hip arthroplasty. Eur J Med Res 16(3):127–132PubMedCentralPubMedCrossRef Zilkens C, Djalali S, Bittersohl B et al (2011) Migration pattern of cementless press fit cups in the presence of stabilizing screws in total hip arthroplasty. Eur J Med Res 16(3):127–132PubMedCentralPubMedCrossRef
16.
Zurück zum Zitat Udomkiat P, Dorr LD, Wan Z (2002) Cementless hemispheric porous-coated sockets implanted with press-fit technique without screws: average ten-year follow-up. J Bone Joint Surg Am 84-A(7):1195–1200PubMed Udomkiat P, Dorr LD, Wan Z (2002) Cementless hemispheric porous-coated sockets implanted with press-fit technique without screws: average ten-year follow-up. J Bone Joint Surg Am 84-A(7):1195–1200PubMed
17.
Zurück zum Zitat Templeton JE, Callaghan JJ, Goetz DD, Sullivan PM, Johnston RC (2001) Revision of a cemented acetabular component to a cementless acetabular component. A ten to fourteen-year follow-up study. J Bone Joint Surg Am 83-A(11):1706–1711PubMed Templeton JE, Callaghan JJ, Goetz DD, Sullivan PM, Johnston RC (2001) Revision of a cemented acetabular component to a cementless acetabular component. A ten to fourteen-year follow-up study. J Bone Joint Surg Am 83-A(11):1706–1711PubMed
18.
Zurück zum Zitat Paprosky WG, Perona PG, Lawrence JM (1994) Acetabular defect classification and surgical reconstruction in revision arthroplasty. A 6-year follow-up evaluation. J Arthroplasty 9(1):33–44PubMedCrossRef Paprosky WG, Perona PG, Lawrence JM (1994) Acetabular defect classification and surgical reconstruction in revision arthroplasty. A 6-year follow-up evaluation. J Arthroplasty 9(1):33–44PubMedCrossRef
19.
Zurück zum Zitat Keating EM, Ritter MA, Faris PM (1990) Structures at risk from medially placed acetabular screws. J Bone Joint Surg Am 72(4):509–511PubMed Keating EM, Ritter MA, Faris PM (1990) Structures at risk from medially placed acetabular screws. J Bone Joint Surg Am 72(4):509–511PubMed
20.
Zurück zum Zitat Wasielewski RC, Galat DD, Sheridan KC, Rubash HE (2005) Acetabular anatomy and transacetabular screw fixation at the high hip center. Clin Orthop Relat Res 438:171–176PubMedCrossRef Wasielewski RC, Galat DD, Sheridan KC, Rubash HE (2005) Acetabular anatomy and transacetabular screw fixation at the high hip center. Clin Orthop Relat Res 438:171–176PubMedCrossRef
21.
Zurück zum Zitat Kirkpatrick JS, Callaghan JJ, Vandemark RM, Goldner RD (1990) The relationship of the intrapelvic vasculature to the acetabulum. Implications in screw-fixation acetabular components. Clin Orthop Relat Res 258:183–190PubMed Kirkpatrick JS, Callaghan JJ, Vandemark RM, Goldner RD (1990) The relationship of the intrapelvic vasculature to the acetabulum. Implications in screw-fixation acetabular components. Clin Orthop Relat Res 258:183–190PubMed
22.
Zurück zum Zitat Cipriano CA, Issack PS, Beksac B, Della Valle AG, Sculco TP, Salvati EA (2008) Metallosis after metal-on-polyethylene total hip arthroplasty. Am J Orthop (Belle Mead NJ) 37(2):E18–E25 Cipriano CA, Issack PS, Beksac B, Della Valle AG, Sculco TP, Salvati EA (2008) Metallosis after metal-on-polyethylene total hip arthroplasty. Am J Orthop (Belle Mead NJ) 37(2):E18–E25
23.
Zurück zum Zitat Jasty M, Bragdon C, Jiranek W, Chandler H, Maloney W, Harris WH (1994) Etiology of osteolysis around porous-coated cementless total hip arthroplasties. Clin Orthop Relat Res 308:111–126PubMed Jasty M, Bragdon C, Jiranek W, Chandler H, Maloney W, Harris WH (1994) Etiology of osteolysis around porous-coated cementless total hip arthroplasties. Clin Orthop Relat Res 308:111–126PubMed
24.
Zurück zum Zitat Wilson-MacDonald J, Morscher E, Masar Z (1990) Cementless uncoated polyethylene acetabular components in total hip replacement. Review of five- to 10-year results. J Bone Joint Surg Br 72(3):423–430PubMed Wilson-MacDonald J, Morscher E, Masar Z (1990) Cementless uncoated polyethylene acetabular components in total hip replacement. Review of five- to 10-year results. J Bone Joint Surg Br 72(3):423–430PubMed
25.
Zurück zum Zitat Maloney WJ, Peters P, Engh CA, Chandler H (1993) Severe osteolysis of the pelvic in association with acetabular replacement without cement. J Bone Joint Surg Am 75(11):1627–1635PubMed Maloney WJ, Peters P, Engh CA, Chandler H (1993) Severe osteolysis of the pelvic in association with acetabular replacement without cement. J Bone Joint Surg Am 75(11):1627–1635PubMed
26.
Zurück zum Zitat Berman AT, Avolio A Jr, DelGallo W (1994) Acetabular osteolysis in total hip arthroplasty: prevention and treatment. Orthopedics 17(10):963–965PubMed Berman AT, Avolio A Jr, DelGallo W (1994) Acetabular osteolysis in total hip arthroplasty: prevention and treatment. Orthopedics 17(10):963–965PubMed
27.
Zurück zum Zitat Galat DD, Petrucci JA, Wasielewski RC (2004) Radiographic evaluation of screw position in revision total hip arthroplasty. Clin Orthop Relat Res 419:124–129PubMedCrossRef Galat DD, Petrucci JA, Wasielewski RC (2004) Radiographic evaluation of screw position in revision total hip arthroplasty. Clin Orthop Relat Res 419:124–129PubMedCrossRef
28.
Zurück zum Zitat Jamieson ML, Russell RD, Incavo SJ, Noble PC (2011) Does an enhanced surface finish improve acetabular fixation in revision total hip arthroplasty? J Arthroplasty 26(4):644–648PubMedCrossRef Jamieson ML, Russell RD, Incavo SJ, Noble PC (2011) Does an enhanced surface finish improve acetabular fixation in revision total hip arthroplasty? J Arthroplasty 26(4):644–648PubMedCrossRef
29.
Zurück zum Zitat Ries MD, Salehi A, Shea J (1999) Photoelastic analysis of stresses produced by different acetabular cups. Clin Orthop Relat Res 369:165–174PubMedCrossRef Ries MD, Salehi A, Shea J (1999) Photoelastic analysis of stresses produced by different acetabular cups. Clin Orthop Relat Res 369:165–174PubMedCrossRef
30.
Zurück zum Zitat Baleani M, Fognani R, Toni A (2001) Initial stability of a cementless acetabular cup design: experimental investigation on the effect of adding fins to the rim of the cup. Artif Organs 25(8):664–669PubMedCrossRef Baleani M, Fognani R, Toni A (2001) Initial stability of a cementless acetabular cup design: experimental investigation on the effect of adding fins to the rim of the cup. Artif Organs 25(8):664–669PubMedCrossRef
Metadaten
Titel
Effect of design on the initial stability of press-fit cups in the presence of acetabular rim defects: experimental evaluation of the effect of adding circumferential fins
verfasst von
Wolfgang O. Huber
Philip C. Noble
Publikationsdatum
01.04.2014
Verlag
Springer Berlin Heidelberg
Erschienen in
International Orthopaedics / Ausgabe 4/2014
Print ISSN: 0341-2695
Elektronische ISSN: 1432-5195
DOI
https://doi.org/10.1007/s00264-013-2187-8

Weitere Artikel der Ausgabe 4/2014

International Orthopaedics 4/2014 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Aquatherapie bei Fibromyalgie wirksamer als Trockenübungen

03.05.2024 Fibromyalgiesyndrom Nachrichten

Bewegungs-, Dehnungs- und Entspannungsübungen im Wasser lindern die Beschwerden von Patientinnen mit Fibromyalgie besser als das Üben auf trockenem Land. Das geht aus einer spanisch-brasilianischen Vergleichsstudie hervor.

Endlich: Zi zeigt, mit welchen PVS Praxen zufrieden sind

IT für Ärzte Nachrichten

Darauf haben viele Praxen gewartet: Das Zi hat eine Liste von Praxisverwaltungssystemen veröffentlicht, die von Nutzern positiv bewertet werden. Eine gute Grundlage für wechselwillige Ärzte und Psychotherapeuten.

Proximale Humerusfraktur: Auch 100-Jährige operieren?

01.05.2024 DCK 2024 Kongressbericht

Mit dem demographischen Wandel versorgt auch die Chirurgie immer mehr betagte Menschen. Von Entwicklungen wie Fast-Track können auch ältere Menschen profitieren und bei proximaler Humerusfraktur können selbst manche 100-Jährige noch sicher operiert werden.

Sind Frauen die fähigeren Ärzte?

30.04.2024 Gendermedizin Nachrichten

Patienten, die von Ärztinnen behandelt werden, dürfen offenbar auf bessere Therapieergebnisse hoffen als Patienten von Ärzten. Besonders scheint das auf weibliche Kranke zuzutreffen, wie eine Studie zeigt.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.