14.06.2016 | Original Article

Comparison of laser fiber passage in ureteroscopic maximum deflection and their influence on deflection and irrigation: Do we really need the ball tip concept?

verfasst von: Mohammed Baghdadi, Esteban Emiliani, Michele Talso, Pol Servián, Aaron Barreiro, Andrea Orosa, Silvia Proietti, Olivier Traxer

Erschienen in: World Journal of Urology | Ausgabe 2/2017

Einloggen, um Zugang zu erhalten

Abstract

Purpose

To examine laser fiber passage capabilities through flexible ureterorenoscopes (fURSs) and to measure deflections and flow characteristics.

Methods

For this in vitro study, eight fURSs were examined (Olympus® URF-P6, URF-P6, URF-V, URF-V2; Storz® Xc and Flex-X2; Richard Wolf® Cobra Vision; and Lithovue). Four laser fibers standard 200- and 273-μm (uncleaved and cleaved), sheath-coated and ball-tip fibers were attempted to pass through each fURS while deflected at 120°, 180°, maximum deflection, and maximum deflection with reduced 9-mm radius. Measurements included maximal (up/down) deflections and irrigation flow rates achieved with each fiber.

Results

Wolf Cobra Vision demonstrated minimal loss of deflections with mean differences of −2°/0° (p > 0.05) when loaded with the 200-μm fiber. The 273-μm fiber provoked utmost deflections that decline when loaded in Olympus URF-P5: mean differences of −52°/−35° (p < 0.001 for upward deflection). Of overall deflections, sheath-coated fiber induced least insult (p > 0.05), while standard 273-μm fiber incited maximum degradation (p < 0.00001). With few exceptions, sheath-coated and ball-tip fibers passed through all maximally deflected scopes. Uncleaved 200- and 273-μm fibers failed to pass through most maximally deflected fURS. However, cleaving their ends allowed 200- and 273-μm fiber to pass through all angles of deflections expect in the Olympus URF-P5 and Olympus URF-P5 and URF-V, respectively. The irrigation through all fURSs was significantly impaired (p < 0.00001).

Conclusions

fURS deflection was least affected by sheath-coated fibers and most affected by the 273-μm fiber. Uncleaved 200- and 273-μm fibers showed least passage capabilities; while removing the ends, the fibers greatly facilitated their passage capabilities as much as the other fibers tested.
Literatur
3.
Zurück zum Zitat Al-Qahtani SM, Gil-deiz-de-Medina S, Traxer O (2012) Predictors of clinical outcomes of flexible ureterorenoscopy with Holmium laser for renal stone greater than 2 cm. Adv Urol 2012:543537. doi:10.1155/2012/543537 CrossRefPubMed Al-Qahtani SM, Gil-deiz-de-Medina S, Traxer O (2012) Predictors of clinical outcomes of flexible ureterorenoscopy with Holmium laser for renal stone greater than 2 cm. Adv Urol 2012:543537. doi:10.​1155/​2012/​543537 CrossRefPubMed
6.
Zurück zum Zitat Wiesenthal JD, Ghiculete D, DA Honey RJ et al (2011) A comparison of treatment modalities for renal calculi between 100 and 300 mm2: are shockwave lithotripsy, ureteroscopy, and percutaneous nephrolithotomy equivalent? J Endourol 25:481–485. doi:10.1089/end.2010.0208 CrossRefPubMed Wiesenthal JD, Ghiculete D, DA Honey RJ et al (2011) A comparison of treatment modalities for renal calculi between 100 and 300 mm2: are shockwave lithotripsy, ureteroscopy, and percutaneous nephrolithotomy equivalent? J Endourol 25:481–485. doi:10.​1089/​end.​2010.​0208 CrossRefPubMed
8.
Zurück zum Zitat Sooriakumaran P, Kaba R, Andrews HO et al (2005) Evaluation of the mechanisms of damage to flexible ureteroscopes and suggestions for ureteroscope preservation. Asian J Androl 7:433–438CrossRefPubMed Sooriakumaran P, Kaba R, Andrews HO et al (2005) Evaluation of the mechanisms of damage to flexible ureteroscopes and suggestions for ureteroscope preservation. Asian J Androl 7:433–438CrossRefPubMed
9.
12.
Zurück zum Zitat Clayman RV (2003) Flexible ureterorenoscopy for the treatment of lower pole calyx stones: influence of different lithotripsy probes and stone extraction tools on scope reflection and irrigation flow. J Urol 170:686–687CrossRef Clayman RV (2003) Flexible ureterorenoscopy for the treatment of lower pole calyx stones: influence of different lithotripsy probes and stone extraction tools on scope reflection and irrigation flow. J Urol 170:686–687CrossRef
16.
Zurück zum Zitat Knudsen B, De S, Monga M (2014) Abstract pd37-05 Ball tipped holmium:YAG optical fiber: pulse energy settings increase degradation during clinical use. J Urol 191:947CrossRef Knudsen B, De S, Monga M (2014) Abstract pd37-05 Ball tipped holmium:YAG optical fiber: pulse energy settings increase degradation during clinical use. J Urol 191:947CrossRef
Metadaten
Titel
Comparison of laser fiber passage in ureteroscopic maximum deflection and their influence on deflection and irrigation: Do we really need the ball tip concept?
verfasst von
Mohammed Baghdadi
Esteban Emiliani
Michele Talso
Pol Servián
Aaron Barreiro
Andrea Orosa
Silvia Proietti
Olivier Traxer
Publikationsdatum
14.06.2016
Verlag
Springer Berlin Heidelberg
Erschienen in
World Journal of Urology / Ausgabe 2/2017
Print ISSN: 0724-4983
Elektronische ISSN: 1433-8726
DOI
https://doi.org/10.1007/s00345-016-1873-8

Weitere Artikel der Ausgabe 2/2017

World Journal of Urology 2/2017 Zur Ausgabe

Update Urologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.