Skip to main content
Erschienen in: Archives of Orthopaedic and Trauma Surgery 10/2008

01.10.2008 | Orthopaedic Surgery

Femoral fit in ABG-II hip stems, influence on clinical outcome and bone remodeling: a radiographic study

verfasst von: B. C. H. van der Wal, B. J. de Kramer, B. Grimm, W. Vencken, I. C. Heyligers, A. J. Tonino

Erschienen in: Archives of Orthopaedic and Trauma Surgery | Ausgabe 10/2008

Einloggen, um Zugang zu erhalten

Abstract

Introduction

In a consecutive series of 64 patients with an ABG-II stem, we analyzed whether the tightness of the fit of the prosthesis affected bone remodeling and if there was a relation between clinical and radiological results.

Materials and methods

Radiographic analysis of bone remodeling in different Gruen zones was done. Bone density changes were graded as present or absent in the AP and lateral radiographs as compared with the previous sets of radiographs. Bone remodeling was compared to literature values of the ABG-I stem. Three stem levels were defined. The proximal level was set at the upper border of the lesser trochanter, the mid-stem level at halfway the stem and the distal level 1 cm above the tip of the prosthesis. Femoral fit was defined as tight when the ratio of the fit was ≥0.8 and as non-tight if the fit was <0.8. The incidence of thigh pain was scored using the Merle d’Aubigne (MdA) hip score.

Results

Proximal bone resorption in Gruen zone 1 was 26.6% and in zone 7, 34.4% compared to 48 and 45% for the ABG-I stem after 5 years. No correlation was found between femoral fit and radiological changes. Proximal and distal fit was significantly lower for patients with thigh pain than without thigh pain. Patients with a non-tight proximal fit produced significantly more varus (17/30 = 56.7%) than patients with a tight proximal fit (2/34 = 5.9%; P < 0.01).

Conclusion

Femoral fit in ABG-II does not predict certain radiological changes, but less proximal bone resorption confirms the design changes from ABG-I to ABG-II. A non-tight proximal fit is correlated with varus position of the stem. Thigh pain is correlated with a poor fit and fill of the femoral stem.
Literatur
1.
Zurück zum Zitat Ang KC, Das De S, Goh JC, Low SL, Bose K (1997) Periprosthetic bone remodelling after cementless total hip replacement. A prospective comparison of two different implant designs. J Bone Joint Surg Br 79:675–679PubMedCrossRef Ang KC, Das De S, Goh JC, Low SL, Bose K (1997) Periprosthetic bone remodelling after cementless total hip replacement. A prospective comparison of two different implant designs. J Bone Joint Surg Br 79:675–679PubMedCrossRef
2.
Zurück zum Zitat Archibeck MJ, Berger RA, Jacobs JJ, Quigley LR, Gitelis S, Rosenberg AG, Galante JO (2001) Second-generation cementless total hip arthroplasty. Eight to eleven-year results. J Bone Joint Surg Am 83-A:1666–1673PubMed Archibeck MJ, Berger RA, Jacobs JJ, Quigley LR, Gitelis S, Rosenberg AG, Galante JO (2001) Second-generation cementless total hip arthroplasty. Eight to eleven-year results. J Bone Joint Surg Am 83-A:1666–1673PubMed
3.
Zurück zum Zitat Bourne RB, Rorabeck CH, Ghazal ME, Lee MH (1994) Pain in the thigh following total hip replacement with a porous-coated anatomic prosthesis for osteoarthrosis. A five-year follow-up study. J Bone Joint Surg Am 76:1464–1470PubMed Bourne RB, Rorabeck CH, Ghazal ME, Lee MH (1994) Pain in the thigh following total hip replacement with a porous-coated anatomic prosthesis for osteoarthrosis. A five-year follow-up study. J Bone Joint Surg Am 76:1464–1470PubMed
4.
Zurück zum Zitat Bourne RB, Rorabeck CH, Patterson JJ, Guerin J (2001) Tapered titanium cementless total hip replacements: a 10- to 13-year follow-up study. Clin Orthop Relat Res:112–120 Bourne RB, Rorabeck CH, Patterson JJ, Guerin J (2001) Tapered titanium cementless total hip replacements: a 10- to 13-year follow-up study. Clin Orthop Relat Res:112–120
5.
Zurück zum Zitat Brooker AF, Bowerman JW, Robinson RA, Riley LH Jr (1973) Ectopic ossification following total hip replacement. Incidence and a method of classification. J Bone Joint Surg Am 55:1629–1632PubMed Brooker AF, Bowerman JW, Robinson RA, Riley LH Jr (1973) Ectopic ossification following total hip replacement. Incidence and a method of classification. J Bone Joint Surg Am 55:1629–1632PubMed
6.
Zurück zum Zitat Burkart BC, Bourne RB, Rorabeck CH, Kirk PG (1993) Thigh pain in cementless total hip arthroplasty. A comparison of two systems at 2 years’ follow-up. Orthop Clin North Am 24:645–653PubMed Burkart BC, Bourne RB, Rorabeck CH, Kirk PG (1993) Thigh pain in cementless total hip arthroplasty. A comparison of two systems at 2 years’ follow-up. Orthop Clin North Am 24:645–653PubMed
7.
Zurück zum Zitat Danesh-Clough T, Bourne RB, Rorabeck CH, McCalden R (2007) The mid-term results of a dual offset uncemented stem for total hip arthroplasty. J Arthroplasty 22:195–203PubMedCrossRef Danesh-Clough T, Bourne RB, Rorabeck CH, McCalden R (2007) The mid-term results of a dual offset uncemented stem for total hip arthroplasty. J Arthroplasty 22:195–203PubMedCrossRef
8.
Zurück zum Zitat D’Antonio JA, Capello WN, Jaffe WL (1992) Hydroxylapatite-coated hip implants. Multicenter three-year clinical and roentgenographic results. Clin Orthop Relat Res:102–15 D’Antonio JA, Capello WN, Jaffe WL (1992) Hydroxylapatite-coated hip implants. Multicenter three-year clinical and roentgenographic results. Clin Orthop Relat Res:102–15
9.
Zurück zum Zitat Dorr LD, Faugere MC, Mackel AM, Gruen TA, Bognar B, Malluche HH (1993) Structural and cellular assessment of bone quality of proximal femur. Bone 14:231–42PubMedCrossRef Dorr LD, Faugere MC, Mackel AM, Gruen TA, Bognar B, Malluche HH (1993) Structural and cellular assessment of bone quality of proximal femur. Bone 14:231–42PubMedCrossRef
10.
Zurück zum Zitat Duwelius PJ, Schmidt AH, Kyle RF, Talbott V, Ellis TJ, Butler JB (2004) A prospective, modernized treatment protocol for periprosthetic femur fractures. Orthop Clin North Am 35, vi:485–492CrossRef Duwelius PJ, Schmidt AH, Kyle RF, Talbott V, Ellis TJ, Butler JB (2004) A prospective, modernized treatment protocol for periprosthetic femur fractures. Orthop Clin North Am 35, vi:485–492CrossRef
11.
Zurück zum Zitat Engh CA, Jr., McAuley JP, Sychterz CJ, Sacco ME, Engh CA, Sr (2000) The accuracy and reproducibility of radiographic assessment of stress-shielding. A postmortem analysis J Bone Joint Surg Am 82-A:1414–1420PubMed Engh CA, Jr., McAuley JP, Sychterz CJ, Sacco ME, Engh CA, Sr (2000) The accuracy and reproducibility of radiographic assessment of stress-shielding. A postmortem analysis J Bone Joint Surg Am 82-A:1414–1420PubMed
12.
Zurück zum Zitat Engh CA, Massin P, Suthers KE (1990) Roentgenographic assessment of the biologic fixation of porous-surfaced femoral components. Clin Orthop Relat Res:107–128 Engh CA, Massin P, Suthers KE (1990) Roentgenographic assessment of the biologic fixation of porous-surfaced femoral components. Clin Orthop Relat Res:107–128
13.
Zurück zum Zitat Gosens T, Sluimer JC, Kester AD, van Langelaan EJ (2005) Femoral fit predicts radiologic changes, but not clinical results, in Mallory-head total hip arthroplasties. Clin Orthop Relat Res:138–147 Gosens T, Sluimer JC, Kester AD, van Langelaan EJ (2005) Femoral fit predicts radiologic changes, but not clinical results, in Mallory-head total hip arthroplasties. Clin Orthop Relat Res:138–147
14.
Zurück zum Zitat Gruen TA, McNeice GM, Amstutz HC (1979) “Modes of failure” of cemented stem-type femoral components: a radiographic analysis of loosening. Clin Orthop Relat Res:17–27 Gruen TA, McNeice GM, Amstutz HC (1979) “Modes of failure” of cemented stem-type femoral components: a radiographic analysis of loosening. Clin Orthop Relat Res:17–27
15.
Zurück zum Zitat Hellman EJ, Capello WN, Feinberg JR (1999) Omnifit cementless total hip arthroplasty. A 10-year average followup. Clin Orthop Relat Res:164–174 Hellman EJ, Capello WN, Feinberg JR (1999) Omnifit cementless total hip arthroplasty. A 10-year average followup. Clin Orthop Relat Res:164–174
16.
Zurück zum Zitat Herrera A, Canales V, Anderson J, Garcia-Araujo C, Murcia-Mazon A, Tonino AJ (2004) Seven to 10 years followup of an anatomic hip prosthesis: an international study. Clin Orthop Relat Res (423):129–137 Herrera A, Canales V, Anderson J, Garcia-Araujo C, Murcia-Mazon A, Tonino AJ (2004) Seven to 10 years followup of an anatomic hip prosthesis: an international study. Clin Orthop Relat Res (423):129–137
17.
Zurück zum Zitat Iorio R, Healy WL, Warren PD, Appleby D (2006) Lateral trochanteric pain following primary total hip arthroplasty. J Arthroplasty 21:233–236PubMedCrossRef Iorio R, Healy WL, Warren PD, Appleby D (2006) Lateral trochanteric pain following primary total hip arthroplasty. J Arthroplasty 21:233–236PubMedCrossRef
18.
Zurück zum Zitat Kim YH, Kim VE (1993) Uncemented, porous-coated anatomic total hip replacement. Results at six years in a consecutive series. J Bone Joint Surg Br 75:6–13PubMed Kim YH, Kim VE (1993) Uncemented, porous-coated anatomic total hip replacement. Results at six years in a consecutive series. J Bone Joint Surg Br 75:6–13PubMed
19.
Zurück zum Zitat Laine HJ, Puolakka TJ, Moilanen T, Pajamaki KJ, Wirta J, Lehto MU (2000) The effects of cementless femoral stem shape and proximal surface texture on ‘fit-and-fill’ characteristics and on bone remodeling. Int Orthop 24:184–190PubMedCrossRef Laine HJ, Puolakka TJ, Moilanen T, Pajamaki KJ, Wirta J, Lehto MU (2000) The effects of cementless femoral stem shape and proximal surface texture on ‘fit-and-fill’ characteristics and on bone remodeling. Int Orthop 24:184–190PubMedCrossRef
20.
Zurück zum Zitat Malchau H, Karrholm J, Wang YX, Herberts P (1995) Accuracy of migration analysis in hip arthroplasty. Digitized and conventional radiography, compared to radiostereometry in 51 patients. Acta Orthop Scand 66:418–424PubMedCrossRef Malchau H, Karrholm J, Wang YX, Herberts P (1995) Accuracy of migration analysis in hip arthroplasty. Digitized and conventional radiography, compared to radiostereometry in 51 patients. Acta Orthop Scand 66:418–424PubMedCrossRef
21.
Zurück zum Zitat Meding JB, Keating EM, Ritter MA, Faris PM, Berend ME (2004) Minimum ten-year follow-up of a straight-stemmed, plasma-sprayed, titanium-alloy, uncemented femoral component in primary total hip arthroplasty. J Bone Joint Surg Am 86-A:92–97PubMed Meding JB, Keating EM, Ritter MA, Faris PM, Berend ME (2004) Minimum ten-year follow-up of a straight-stemmed, plasma-sprayed, titanium-alloy, uncemented femoral component in primary total hip arthroplasty. J Bone Joint Surg Am 86-A:92–97PubMed
22.
Zurück zum Zitat Mulliken BD, Nayak N, Bourne RB, Rorabeck CH, Bullas R (1996) Early radiographic results comparing cemented and cementless total hip arthroplasty. J Arthroplasty 11:24–33PubMedCrossRef Mulliken BD, Nayak N, Bourne RB, Rorabeck CH, Bullas R (1996) Early radiographic results comparing cemented and cementless total hip arthroplasty. J Arthroplasty 11:24–33PubMedCrossRef
23.
Zurück zum Zitat Noble PC, Alexander JW, Lindahl LJ, Yew DT, Granberry WM, Tullos HS (1988) The anatomic basis of femoral component design. Clin Orthop Relat Res (244):148–165 Noble PC, Alexander JW, Lindahl LJ, Yew DT, Granberry WM, Tullos HS (1988) The anatomic basis of femoral component design. Clin Orthop Relat Res (244):148–165
24.
Zurück zum Zitat Oosterbos CJ, Rahmy AI, Tonino AJ (2001) Hydroxyapatite coated hip prosthesis followed up for 5 years. Int Orthop 25:17–21PubMedCrossRef Oosterbos CJ, Rahmy AI, Tonino AJ (2001) Hydroxyapatite coated hip prosthesis followed up for 5 years. Int Orthop 25:17–21PubMedCrossRef
25.
Zurück zum Zitat Oosterbos CJ, Rahmy AI, Tonino AJ, Witpeerd W (2004) High survival rate of hydroxyapatite-coated hip prostheses: 100 consecutive hips followed for 10 years. Acta Orthop Scand 75:127–133PubMedCrossRef Oosterbos CJ, Rahmy AI, Tonino AJ, Witpeerd W (2004) High survival rate of hydroxyapatite-coated hip prostheses: 100 consecutive hips followed for 10 years. Acta Orthop Scand 75:127–133PubMedCrossRef
26.
Zurück zum Zitat Park MS, Choi BW, Kim SJ, Park JH (2003) Plasma spray-coated Ti femoral component for cementless total hip arthroplasty. J Arthroplasty 18:626–30PubMedCrossRef Park MS, Choi BW, Kim SJ, Park JH (2003) Plasma spray-coated Ti femoral component for cementless total hip arthroplasty. J Arthroplasty 18:626–30PubMedCrossRef
27.
Zurück zum Zitat Rahmy AI, Gosens T, Blake GM, Tonino A, Fogelman I (2004) Periprosthetic bone remodelling of two types of uncemented femoral implant with proximal hydroxyapatite coating: a 3-year follow-up study addressing the influence of prosthesis design and preoperative bone density on periprosthetic bone loss. Osteoporos Int 15:281–289PubMedCrossRef Rahmy AI, Gosens T, Blake GM, Tonino A, Fogelman I (2004) Periprosthetic bone remodelling of two types of uncemented femoral implant with proximal hydroxyapatite coating: a 3-year follow-up study addressing the influence of prosthesis design and preoperative bone density on periprosthetic bone loss. Osteoporos Int 15:281–289PubMedCrossRef
28.
Zurück zum Zitat Rossi P, Sibelli P, Fumero S, Crua E (1995) Short-term results of hydroxyapatite-coated primary total hip arthroplasty. Clin Orthop Relat Res (310):98–102 Rossi P, Sibelli P, Fumero S, Crua E (1995) Short-term results of hydroxyapatite-coated primary total hip arthroplasty. Clin Orthop Relat Res (310):98–102
29.
Zurück zum Zitat Singh M, Riggs BL, Beabout JW, Jowsey J (1973) Femoral trabecular pattern index for evaluation of spinal osteoporosis. A detailed methodologic description. Mayo Clin Proc 48:184–189PubMed Singh M, Riggs BL, Beabout JW, Jowsey J (1973) Femoral trabecular pattern index for evaluation of spinal osteoporosis. A detailed methodologic description. Mayo Clin Proc 48:184–189PubMed
30.
Zurück zum Zitat Theis JC, Beadel G (2003) Changes in proximal femoral bone mineral density around a hydroxyapatite-coated hip joint arthroplasty. J Orthop Surg (Hong Kong) 11:48–52 Theis JC, Beadel G (2003) Changes in proximal femoral bone mineral density around a hydroxyapatite-coated hip joint arthroplasty. J Orthop Surg (Hong Kong) 11:48–52
31.
Zurück zum Zitat Tonino AJ, Rahmy AI (2000) The hydroxyapatite-ABG hip system: 5- to 7-year results from an international multicentre study. The International ABG Study Group. J Arthroplasty 15:274–282PubMedCrossRef Tonino AJ, Rahmy AI (2000) The hydroxyapatite-ABG hip system: 5- to 7-year results from an international multicentre study. The International ABG Study Group. J Arthroplasty 15:274–282PubMedCrossRef
32.
Zurück zum Zitat Tonino AJ, Therin M, Doyle C (1999) Hydroxyapatite-coated femoral stems. Histology and histomorphometry around five components retrieved at postmortem. J Bone Joint Surg Br 81:148–154PubMedCrossRef Tonino AJ, Therin M, Doyle C (1999) Hydroxyapatite-coated femoral stems. Histology and histomorphometry around five components retrieved at postmortem. J Bone Joint Surg Br 81:148–154PubMedCrossRef
33.
Zurück zum Zitat van der Wal BC, Rahmy AIA, Grimm B, Blake GM, Heyligers IC, Tonino AJ (2006) The influence of implant design on periprosthetic bone remodelling of two types of uncemented HA-coated hip stems. A two-year follow-up study using DEXA. Hip Int 1:8–17 van der Wal BC, Rahmy AIA, Grimm B, Blake GM, Heyligers IC, Tonino AJ (2006) The influence of implant design on periprosthetic bone remodelling of two types of uncemented HA-coated hip stems. A two-year follow-up study using DEXA. Hip Int 1:8–17
34.
Zurück zum Zitat Venesmaa PK, Kroger HP, Miettinen HJ, Jurvelin JS, Suomalainen OT, Alhava EM (2001) Monitoring of periprosthetic BMD after uncemented total hip arthroplasty with dual-energy X-ray absorptiometry–a 3-year follow-up study. J Bone Miner Res 16:1056–1061PubMedCrossRef Venesmaa PK, Kroger HP, Miettinen HJ, Jurvelin JS, Suomalainen OT, Alhava EM (2001) Monitoring of periprosthetic BMD after uncemented total hip arthroplasty with dual-energy X-ray absorptiometry–a 3-year follow-up study. J Bone Miner Res 16:1056–1061PubMedCrossRef
35.
Zurück zum Zitat Whiteside LA (1989) The effect of stem fit on bone hypertrophy and pain relief in cementless total hip arthroplasty. Clin Orthop Relat Res (247):138–147 Whiteside LA (1989) The effect of stem fit on bone hypertrophy and pain relief in cementless total hip arthroplasty. Clin Orthop Relat Res (247):138–147
36.
Zurück zum Zitat Wu CC, Au MK, Wu SS, Lin LC (1999) Risk factors for postoperative femoral fracture in cementless hip arthroplasty. J Formos Med Assoc 98:190–194PubMed Wu CC, Au MK, Wu SS, Lin LC (1999) Risk factors for postoperative femoral fracture in cementless hip arthroplasty. J Formos Med Assoc 98:190–194PubMed
Metadaten
Titel
Femoral fit in ABG-II hip stems, influence on clinical outcome and bone remodeling: a radiographic study
verfasst von
B. C. H. van der Wal
B. J. de Kramer
B. Grimm
W. Vencken
I. C. Heyligers
A. J. Tonino
Publikationsdatum
01.10.2008
Verlag
Springer-Verlag
Erschienen in
Archives of Orthopaedic and Trauma Surgery / Ausgabe 10/2008
Print ISSN: 0936-8051
Elektronische ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-007-0537-y

Weitere Artikel der Ausgabe 10/2008

Archives of Orthopaedic and Trauma Surgery 10/2008 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Arthroskopie kann Knieprothese nicht hinauszögern

25.04.2024 Gonarthrose Nachrichten

Ein arthroskopischer Eingriff bei Kniearthrose macht im Hinblick darauf, ob und wann ein Gelenkersatz fällig wird, offenbar keinen Unterschied.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.