C2 prosthesis: anterior upper cervical fixation device to reconstruct the second cervical vertebra

verfasst von: Dezső Jeszenszky, Tamás Fülöp Fekete, Robert Melcher, Jürgen Harms

Erschienen in: European Spine Journal | Ausgabe 10/2007

Abstract

Destruction of the second cervical vertebra leads to a highly unstable situation. Reconstruction is difficult because the axis plays a central role in rotatory movements and has a unique function in redistributing axial loads. The axis transfers the axial load of the two lateral masses of the atlas to three surfaces on the third cervical vertebra: the two articular facets and the vertebral body. As reconstruction is difficult and the instability in this region is life threatening, pathological processes are often treated less radically compared to other areas of the cervical spine. However, this more moderate approach may result in worse outcomes and prognoses. This paper presents the development of a new implant (C2 prosthesis) and two illustrative cases describing the implementation of this new implant. The C2 prosthesis provides anterior support and therefore allows a more radical surgical approach.

Introduction

Destruction of the body of the second cervical vertebra may develop due to primary or secondary bone tumours or infections. The treatment of these pathological processes often requires surgical resection, which may further increase instability [14]. Current treatment strategy consists of tumour removal, decompression and posterior fixation [24, 8, 9, 13, 15]. If the vertebral body of the axis—including the superior facet joints—is not present either by the destruction of a pathological process or by surgical removal, posterior fixation is widely extended. The usual extent of posterior fixation is from the occiput down to the lower subaxial spine and external immobilization is also applied [3, 6, 9, 13]. This fixation has to exceed the biomechanically compromised segments because the missing anterior support has to be compensated for. Some surgeons also perform anterior reconstructions with bone grafts and screws or bone grafts, plates and screws. However, such anterior reconstruction alone is insufficient, the posterior occipito-cervical fixation and external immobilisation cannot be avoided. These combined anterior and posterior approaches are limited to isolated case reports or small series [1, 8, 10, 11]. The low numbers of anterior reconstructions might be due to the difficulty in achieving an adequate connection between the C1 lateral masses located laterally and the C3 vertebral body in the midline.
To meet the special anatomical and biomechanical requirements of reconstructing the axis, the first and senior authors designed a C2 prosthesis by integrating the Harms anterior transoral plate and titanium mesh cage. An implant, that restores stability following total removal of the second cervical vertebra, without the need for fixation of the atlanto-occipital joints, has, to the authors’ knowledge, not been described before.

Materials and methods

During surgical interventions of C2 resection the implants used to reconstruct the axis were serially modified. As the first step of development a transoral T-shaped plate and a Harms cage was assembled prior to each surgery up to 1997 (Fig. 1a–d, g). Later a shorter and wider titanium cage to provide support for the C1 lateral masses was added (Fig. 1e, h). The prosthesis was custom-made for each patient. The size of the implant was based on measurements of preoperative radiological data. However, later it turned out that two different sizes (large and small) were sufficient for all patients. Therefore the design was finally changed to accommodate for both sizes by intraoperatively shortening the larger size implant when necessary (Fig. 1f, i).

Illustrative case report 1

A 16 year-old patient had a history of neck pain radiating to the occipital region. MRI scans showed a tumour mass at the level of C1 and C2. Suboccipital craniectomy, C1 and C2 laminectomy, and a partial tumour resection were performed through a posterior transdural approach in another hospital (Fig. 2). The histology revealed chordoma. The patient had long tract signs and lower limb ataxia at the postoperative period, which spontaneously resolved in a few weeks. The patient was referred to our hospital for complete tumour removal before proton beam therapy.
We removed the C2 vertebral body through a high cervical left anterolateral approach. Following total tumour removal it was evident that the atlas was not involved, the lateral masses and anterior arch were intact. We then implanted the C2 prosthesis. The C1 lateral mass screws were inserted through two stab incisions on the pharynx in a transoral approach. We turned the patient and performed a posterior C1–C3 fixation using polyaxial screws and rods.
The patient was mobilised without any orthosis and the recovery period was uneventful. Postoperative radiographs showed well-situated implants (Fig. 3). There was no evidence of residual tumour on the postoperative MRI.

Illustrative case report 2

A 51 year-old patient suffered from progressive neck pain. Radiological investigation revealed a left-sided tumour partially destroying the C2 and C3 vertebral bodies. Posterior decompression, C2, C3, C4 laminectomy and tumour resection was performed in another institution. The result of the histological analysis was chordoma.
The patient was referred to our hospital to complete the removal of the tumour before a proton beam irradiation.
We performed a posterior fusion from C1 to C5, using polyaxial lateral mass screws and rods, and restored a lordotic curve. We decided to include the C4–5 segment in the stabilisation, because of significant instability from previous surgery. At that stage we turned the patient and performed total tumour removal, an anterior C2, C3 vertebral body resection and a left-sided vertebral artery resection. We also performed a C4–5 discectomy and fusion. A custom-made C2 prosthesis was used to reconstruct the anterior aspect of the spine. The length of the cylindrical cage was long enough to replace both the C2 and C3 vertebral bodies. The right third of the C2 vertebral body seemed to be intact and preservation was possible. The wing of the C2 prosthesis on the right side was removed accordingly (Fig. 4). The anterior plate of the prosthesis was made especially long to incorporate the C5 vertebra into the fusion. The third approach in the same session was transoral. Two screws to the lateral mass of C1 were inserted through two stab incisions on the posterior wall of the pharynx to stabilise the cranial end of the T-shaped plate of the C2 prosthesis.
The patient was extubated on the third postoperative day and mobilised without any external immobilisation device. Unfortunately, there was a haematoma formation at the anterior surgical site so the anterior wound on the tenth postoperative day was revised and the haematoma was evacuated. There was no implant failure and there was no neurological complication. The postoperative radiographs showed a good alignment of the cervical spine (Fig. 5).
Later the patient received proton beam therapy. At 2-year follow-up the patient is still tumour-free. The patient is fully reintegrated into normal daily activities.

Results: description of C2 prosthesis

The prosthesis consists of a T-shaped plate, a long cylinder mesh cage, and a short elliptical cylinder mesh cage (Fig. 1f, i). The width of the elliptical cylinder cage is 25 mm, the depth is 14 mm and the height is 10 mm. The diameter of the cylinder cage is 14 mm. The elliptical cylinder cage is situated at the cranial end of the main cylinder cage. The horizontal part of the T-shaped plate is just above the main cylinder cage and the caudal part of the plate extends below the main cylinder cage. All three components are built together as a robust construct.
The elliptical cylinder cage provides axial support to the inferior articular facets of the atlas. The load of the elliptical cylinder cage is transferred to the caudally adjacent vertebra via the main cylinder cage (Fig. 8).
There is one hole on each side at the horizontal part of the plate for screws to the lateral mass of C1. There are four holes in two rows at the caudal end of the plate for two screws to the next adjacent subaxial vertebral body. The screw holes in the plate are shaped in a compressive fashion, i.e. when tightening the screws the prosthesis is compressed between the lateral mass of the atlas and the caudally adjacent subaxial vertebra. The holes are designed for bicortical screws with a diameter of 3.5 or 4 mm.
A combined anterolateral and transoral surgical approach was used in both cases described here, however, a transoral approach described by Schmelzle and Harms [12] is generally used.

Transition of the large C2 prosthesis to the small size

There are four holes at the caudal end of the plate, as mentioned above. By cutting off the distal end of the plate with the two holes and, accordingly, cutting off the caudal end of the mesh cage, the implant can be used where the smaller size is needed. Furthermore, the prosthesis can be individually modified intraoperatively just like the Harms mesh cage. The shape of implant can be modified according to the type and extent of resection.
The length of the main cylinder cage can depend on the extent of resection. In individual cases an implant can be made to reconstruct a longer segment of the cervical spine with an appropriately long main cylinder cage and plate. This allows for the resection of additional cervical vertebral bodies. The extent of resection this implant still provides a safe support for is still unknown.

Discussion

Anatomical and biomechanical considerations

The craniovertebral junction is the most mobile portion of the spinal column accounting for about half of all movements in the cervical spine. The second cervical vertebra is a transition between the atlas and the more typical third cervical vertebra. The role of axis in the rotatory movement of the atlantoaxial complex is well known, as its name implies. However, the unique arrangement of its articular facets, along with its special role in the redistribution of axial load, is rarely discussed.
Unlike any other vertebra, the superior articular facets of the axis originate from the vertebral body and extend posteriorly onto the adjoining parts of the pedicles. The inferior articular facets are situated at the junction of the pedicles and laminae. Thus the superior articular facets are anteriorly situated with respect to the inferior articular facets. The consequences are twofold.
First, the superior facets do not form an articular pillar with the inferior facets. As C1 lacks a vertebral body, the axial load of atlas is transferred through the C1–C2 facet joints on both sides and can be regarded as a two-column (left and right) system. The joints between the axis and the third cervical vertebra are the C2–C3 facet joints and the C2–C3 intervertebral disc. Thus, the C2–C3 joints represent a three-column system. The unique role of the body of axis is to redistribute the two-column axial load of the atlas into a three-column system of the subaxial spine (Fig. 6).
Second, the removal of the body of the axis includes removal of the superior articular facets, thereby separating all the connections of the atlas to the pedicle of the axis and thus to the inferior articular facets of C2. This is practically a iatrogenic lesion in the pars interarticaluris of the axis. Vertebral body resection at C2 level through a single anterior approach results in complete disconnection of the spine (Fig. 7)! This is equivalent to a 360° surgical decompression/resection at any other lower spinal level. Therefore following removal of the body of the axis, both anterior and posterior structures have to be reconstructed.

Clinical implications

Anterior reconstruction following axis removal is difficult with the usual strut grafts. The caudal end of the graft material can be secured to the subaxial vertebral body just like anywhere else in the subaxial spine. However, to establish a secure connection at the cranial end of the conventional implant to the atlas is more difficult. The cranial end of a conventional cage or the bone graft can be secured either by an anterior plate or direct screws to the tiny anterior arch of the atlas [11]. However, neither of these methods provide safe fixation. The anterior support is not reliable, because the bone graft or the cage with a relatively small diameter can only support the thin anterior arch of the atlas and not the more important lateral masses. In addition to this the screw purchase in the arch of the atlas is also very poor.
As a result the benefit of adding such an anterior support is at least questionable. This is supported by the authors’ experience and by the small number of reported cases where the fixation failed and/or extensive posterior fixation and external immobilisation was used [11, 15].
Another difficult area following removal of the axis is the counteraction of active rotational forces as muscles acting to rotate the head still remain intact.
In contrast, the C2 prosthesis provides a reliable anterior support and rotational stability. The implant presented in this paper was designed to replace the above-mentioned roles of the axis in axial load redistribution.

Application of the C2 prosthesis

The C2 prosthesis provides anterior support to the atlas by the wide elliptical cylinder cage (wings of the prosthesis), which transfers the axial loads of the C1 lateral masses to the vertebral body of the subaxial vertebra in the midline (Fig. 8). In addition, the prosthesis also provides rotational stability as it is secured to the lateral masses of the atlas by screws through the wide upper portion of the T-shaped plate.
The corrugated ends of the elliptical cylinder cage are designed to embed themselves at C1 lateral masses and at the most cranially available subaxial vertebra to prevent any translational or rotational movement. The plate prevents migration of the implant toward the spinal canal.
As mentioned earlier, an additional posterior fixation is unavoidable since the removal of the body of the axis completely disconnects the spine. This is also supported by a biomechanical study, where additional posterior stabilisation improved stability of transoral atlantoaxial plating [7]. The authors’ method of choice for a posterior fixation is polyaxial screws and rods. However, the extent of the posterior fixation can be reduced in comparison to previously described methods. In contrast to formerly known methods of anterior support, a stable connection is provided between the atlas and the anterior construct. There are four screws in the atlas with the anterior and posterior lateral mass screws. Anterior support is provided by the wings of the implant supporting the lateral masses. As a consequence, an occipitocervical fixation is unnecessary; the motion of C0–C1 segment can be preserved. It also allows a more moderate approach to the rest of the subaxial spine, as the posterior fixation can caudally end at the first available unaffected subaxial vertebra.
The combined anterior C2 prosthesis and posterior screw–rod construct achieves immediate stability. The safe reconstruction allows a more radical approach to pathologies affecting the axis. Thus the generally accepted concepts in spine surgery can be applied in cases of destruction of the axis vertebra. This might result in improved clinical outcome, longer survival, and higher rate of curative surgical interventions.
The decreased extent of fixation preserves unaffected motion segments. Furthermore, the use of an external orthosis is not necessary. All these improve the quality of life.
To the authors’ knowledge, there has been no previously published description of total axis removal without immobilizing the C0–1 segment to date.
Initial results with the C2 prosthesis proved to be excellent and without hardware failure [5].

Conclusions

The new prosthesis was constructed to replace the second cervical vertebra. This is the first implant which allows complete removal of C2 while preserving C0–C1 motion. The combined anterior C2 prosthesis and posterior screw–rod construct achieves immediate stability, obviating the need for an external orthosis and allowing for early mobilisation.
The new implant allows a more radical surgical approach to spinal pathologies involving the axis vertebra. The implant can also be applied in a clinical setting, where the pathological process exceeds the confines of the axis in the caudal direction and destroys subaxial vertebrae. The C2 prosthesis allows the application of concepts generally accepted in other areas of spine surgery.
The benefit of this new fixation method is currently being investigated in a clinical study with 10-year follow-up at the time of writing.

Acknowledgments

The authors would like to thank Biedermann-Motech GmbH for construction and development of the C2 prosthesis; Wilfried Matthis for all the support in biomechanics and the preparations of the 3D illustrations; Andreas Lütscher for the photographs; and Charles McCammon for helping to prepare the manuscript.
Open Access This is an open access article distributed under the terms of the Creative Commons Attribution Noncommercial License ( https://​creativecommons.​org/​licenses/​by-nc/​2.​0 ), which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Orthopädie & Unfallchirurgie

Kombi-Abonnement

Mit e.Med Orthopädie & Unfallchirurgie erhalten Sie Zugang zu CME-Fortbildungen der Fachgebiete, den Premium-Inhalten der dazugehörigen Fachzeitschriften, inklusive einer gedruckten Zeitschrift Ihrer Wahl.

Literatur
1.
Zurück zum Zitat Bailey CS, Fisher CG, Boyd MC, Dvorak MF (2006) En bloc marginal excision of a multilevel cervical chordoma. Case report. J Neurosurg Spine 4:409–414PubMed Bailey CS, Fisher CG, Boyd MC, Dvorak MF (2006) En bloc marginal excision of a multilevel cervical chordoma. Case report. J Neurosurg Spine 4:409–414PubMed
2.
Zurück zum Zitat Dickman CA, Locantro J, Fessler RG (1992) The influence of transoral odontoid resection on stability of the craniovertebral junction. J Neurosurg 77:525–530PubMed Dickman CA, Locantro J, Fessler RG (1992) The influence of transoral odontoid resection on stability of the craniovertebral junction. J Neurosurg 77:525–530PubMed
3.
Zurück zum Zitat Fourney DR, York JE, Cohen ZR, Suki D, Rhines LD, Gokaslan ZL (2003) Management of atlantoaxial metastases with posterior occipitocervical stabilization. J Neurosurg 98:165–170PubMed Fourney DR, York JE, Cohen ZR, Suki D, Rhines LD, Gokaslan ZL (2003) Management of atlantoaxial metastases with posterior occipitocervical stabilization. J Neurosurg 98:165–170PubMed
4.
Zurück zum Zitat Hertlein H, Mittlmeier T, Schurmann M, Lob G (1994) Posterior stabilization of C2 metastases by combination of atlantoaxial screw fixation and hook plate. Eur Spine J 3:52–55PubMedCrossRef Hertlein H, Mittlmeier T, Schurmann M, Lob G (1994) Posterior stabilization of C2 metastases by combination of atlantoaxial screw fixation and hook plate. Eur Spine J 3:52–55PubMedCrossRef
5.
Zurück zum Zitat Jeszenszky D, Harms J, Hadasch R, Preiss S (1999) C2 prosthesis allowing optimal stabilisation after C2 resection following destructive lesions. Eur Spine J 8:S40, P38 Jeszenszky D, Harms J, Hadasch R, Preiss S (1999) C2 prosthesis allowing optimal stabilisation after C2 resection following destructive lesions. Eur Spine J 8:S40, P38
6.
Zurück zum Zitat Jonsson B, Jonsson H Jr., Karlstrom G, Sjostrom L (1994) Surgery of cervical spine metastases: a retrospective study. Eur Spine J 3:76–83PubMedCrossRef Jonsson B, Jonsson H Jr., Karlstrom G, Sjostrom L (1994) Surgery of cervical spine metastases: a retrospective study. Eur Spine J 3:76–83PubMedCrossRef
7.
Zurück zum Zitat Kandziora F, Kerschbaumer F, Starker M, Mittlmeier T (2000) Biomechanical assessment of transoral plate fixation for atlantoaxial instability. Spine 25:1555–1561PubMedCrossRef Kandziora F, Kerschbaumer F, Starker M, Mittlmeier T (2000) Biomechanical assessment of transoral plate fixation for atlantoaxial instability. Spine 25:1555–1561PubMedCrossRef
8.
Zurück zum Zitat Levin DA, Hensinger RN, Graziano GP (2006) Aneurysmal bone cyst of the second cervical vertebrae causing multilevel upper cervical instability. J Spinal Disord Tech 19:73–75PubMedCrossRef Levin DA, Hensinger RN, Graziano GP (2006) Aneurysmal bone cyst of the second cervical vertebrae causing multilevel upper cervical instability. J Spinal Disord Tech 19:73–75PubMedCrossRef
9.
Zurück zum Zitat Piper JG, Menezes AH (1996) Management strategies for tumors of the axis vertebra. J Neurosurg 84:543–551PubMed Piper JG, Menezes AH (1996) Management strategies for tumors of the axis vertebra. J Neurosurg 84:543–551PubMed
10.
Zurück zum Zitat Rao G, Ha CS, Chakrabarti I, Feiz-Erfan I, Mendel E, Rhines LD (2006) Multiple myeloma of the cervical spine: treatment strategies for pain and spinal instability. J Neurosurg Spine 5:140–145PubMed Rao G, Ha CS, Chakrabarti I, Feiz-Erfan I, Mendel E, Rhines LD (2006) Multiple myeloma of the cervical spine: treatment strategies for pain and spinal instability. J Neurosurg Spine 5:140–145PubMed
11.
Zurück zum Zitat Rhines LD, Fourney DR, Siadati A, Suk I, Gokaslan ZL (2005) En bloc resection of multilevel cervical chordoma with C-2 involvement. Case report and description of operative technique. J Neurosurg Spine 2:199–205PubMedCrossRef Rhines LD, Fourney DR, Siadati A, Suk I, Gokaslan ZL (2005) En bloc resection of multilevel cervical chordoma with C-2 involvement. Case report and description of operative technique. J Neurosurg Spine 2:199–205PubMedCrossRef
12.
Zurück zum Zitat Schmelzle R, Harms J (1987) Craniocervical junction-diseases, diagnostic application of imaging procedures, surgical techniques. Fortschr Kiefer Gesichtschir 32:206–208PubMed Schmelzle R, Harms J (1987) Craniocervical junction-diseases, diagnostic application of imaging procedures, surgical techniques. Fortschr Kiefer Gesichtschir 32:206–208PubMed
13.
Zurück zum Zitat Shin H, Barrenechea IJ, Lesser J, Sen C, Perin NI (2006) Occipitocervical fusion after resection of craniovertebral junction tumors. J Neurosurg Spine 4:137–144PubMed Shin H, Barrenechea IJ, Lesser J, Sen C, Perin NI (2006) Occipitocervical fusion after resection of craniovertebral junction tumors. J Neurosurg Spine 4:137–144PubMed
14.
Zurück zum Zitat Stoltze D, Harms J (1991) Cervico-occipitale Stabilisierungen unter besonderer Berücksichtigung der transoralen Technik. Georg Thieme Verlag, Stuttgart Stoltze D, Harms J (1991) Cervico-occipitale Stabilisierungen unter besonderer Berücksichtigung der transoralen Technik. Georg Thieme Verlag, Stuttgart
15.
Zurück zum Zitat Tuite GF, Veres R, Crockard HA, Sell D (1996) Pediatric transoral surgery: indications, complications, and long-term outcome. J Neurosurg 84:573–583PubMed Tuite GF, Veres R, Crockard HA, Sell D (1996) Pediatric transoral surgery: indications, complications, and long-term outcome. J Neurosurg 84:573–583PubMed
Metadaten
Titel
C2 prosthesis: anterior upper cervical fixation device to reconstruct the second cervical vertebra
verfasst von
Dezső Jeszenszky
Tamás Fülöp Fekete
Robert Melcher
Jürgen Harms
Publikationsdatum
01.10.2007
Verlag
Springer-Verlag
Erschienen in
European Spine Journal / Ausgabe 10/2007
Print ISSN: 0940-6719
Elektronische ISSN: 1432-0932
DOI
https://doi.org/10.1007/s00586-007-0435-6

Weitere Artikel der Ausgabe 10/2007

European Spine Journal 10/2007 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Proximale Humerusfraktur: Auch 100-Jährige operieren?

01.05.2024 DCK 2024 Kongressbericht

Mit dem demographischen Wandel versorgt auch die Chirurgie immer mehr betagte Menschen. Von Entwicklungen wie Fast-Track können auch ältere Menschen profitieren und bei proximaler Humerusfraktur können selbst manche 100-Jährige noch sicher operiert werden.

Sind Frauen die fähigeren Ärzte?

30.04.2024 Gendermedizin Nachrichten

Patienten, die von Ärztinnen behandelt werden, dürfen offenbar auf bessere Therapieergebnisse hoffen als Patienten von Ärzten. Besonders gilt das offenbar für weibliche Kranke, wie eine Studie zeigt.

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Arthroskopie kann Knieprothese nicht hinauszögern

25.04.2024 Gonarthrose Nachrichten

Ein arthroskopischer Eingriff bei Kniearthrose macht im Hinblick darauf, ob und wann ein Gelenkersatz fällig wird, offenbar keinen Unterschied.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.