Three-dimensional analysis of cervical spine motion: reliability of a computer assisted magnetic tracking device compared to inclinometer

verfasst von: Ioannis D. Gelalis, Louis E. DeFrate, Kosmas S. Stafilas, Emilios E. Pakos, James D. Kang, Lars G. Gilbertson

Erschienen in: European Spine Journal | Ausgabe 2/2009

Einloggen, um Zugang zu erhalten

Abstract

We aimed to investigate the reliability and reproducibility of a magnetic tracking technique for the assessment of overall cervical spine motion (principal and coupled movements). Ten asymptomatic male volunteers with a mean age of 29.3 years (range 20–37 years) were included in the study. Flexion, extension, left and right lateral bending and left and right axial rotation were measured using a magnetic tracking device (MTD) mounted onto a custom head-piece. For rotational movements in the frontal and sagittal planes the results were compared with the measurements of two standard inclinometers. Intra-observer, inter-observer and intra-instrument reliability was assessed with the intraclass correlation coefficient method. There were no significant differences for all motion measurements between the MTD and the inclinometer. High inter-observer reliability was found in flexion, extension, axial rotation and lateral bending indicating that the testing routine is applicable for different examiners. The intra-observer variability was high in flexion and extension, whereas in lateral bending the reliability coefficients were lower and displayed a fair to good reliability for most of the measurements with the MTD. The results of the MTD were found to be highly comparable with the inclinometer results with an inter-instrument correlation coefficient ranging from 0.88 to 0.99. The MTD is a reliable, reproducible method for three-dimensional motion analysis of the cervical spine and therefore a valuable method both for the clinical assessment of various degenerative and traumatic disorders and as a supplement of different therapeutic procedures and rehabilitation.
Literatur
2.
Zurück zum Zitat AMA (1993) American Medical Association. Guides to the evaluation of permanent impairment. AMA, Chicago AMA (1993) American Medical Association. Guides to the evaluation of permanent impairment. AMA, Chicago
3.
Zurück zum Zitat Antonaci F, Ghirmai S, Bono G, Nappi G (2000) Current methods for cervical spine movement evaluation: a review. Clin Exp Rheumatol 18:S45–S52PubMed Antonaci F, Ghirmai S, Bono G, Nappi G (2000) Current methods for cervical spine movement evaluation: a review. Clin Exp Rheumatol 18:S45–S52PubMed
4.
Zurück zum Zitat Assink N, Bergman GJ, Knoester B, Winters JC, Dijkstra PU (2008) Assessment of the cervical range of motion over time, differences between results of the flock of birds and the EDI-320: a comparison between an electromagnetic tracking system and an electronic inclinometer. Man Ther 13:450–455. doi:10.1016/j.math.2007.05.012 PubMedCrossRef Assink N, Bergman GJ, Knoester B, Winters JC, Dijkstra PU (2008) Assessment of the cervical range of motion over time, differences between results of the flock of birds and the EDI-320: a comparison between an electromagnetic tracking system and an electronic inclinometer. Man Ther 13:450–455. doi:10.​1016/​j.​math.​2007.​05.​012 PubMedCrossRef
5.
Zurück zum Zitat Assink N, Bergman GJ, Knoester B, Winters JC, Dijkstra PU, Postema K (2005) Interobserver reliability of neck-mobility measurement by means of the flock-of-birds electromagnetic tracking system. J Manipulative Physiol Ther 28:408–413. doi:10.1016/j.jmpt.2005.06.009 PubMedCrossRef Assink N, Bergman GJ, Knoester B, Winters JC, Dijkstra PU, Postema K (2005) Interobserver reliability of neck-mobility measurement by means of the flock-of-birds electromagnetic tracking system. J Manipulative Physiol Ther 28:408–413. doi:10.​1016/​j.​jmpt.​2005.​06.​009 PubMedCrossRef
6.
Zurück zum Zitat Bell KM, Frazier EC, Shively CM, Hartman RA, Ulibarri JC, Lee JY, Kang JD, Donaldson WF 3rd (2008) Assessing range of motion to evaluate the adverse effects of ill-fitting cervical orthoses. Spine J Bell KM, Frazier EC, Shively CM, Hartman RA, Ulibarri JC, Lee JY, Kang JD, Donaldson WF 3rd (2008) Assessing range of motion to evaluate the adverse effects of ill-fitting cervical orthoses. Spine J
9.
Zurück zum Zitat Buck CA, Dameron FB, Dow MJ, Skowlund HV (1959) Study of normal range of motion in the neck utilizing a bubble goniometer. Arch Phys Med Rehabil 40:390–392PubMed Buck CA, Dameron FB, Dow MJ, Skowlund HV (1959) Study of normal range of motion in the neck utilizing a bubble goniometer. Arch Phys Med Rehabil 40:390–392PubMed
10.
Zurück zum Zitat Capuano-Pucci D, Rheault W, Aukai J, Bracke M, Day R, Pastrick M (1991) Intratester and intertester reliability of the cervical range of motion device. Arch Phys Med Rehabil 72:338–340PubMed Capuano-Pucci D, Rheault W, Aukai J, Bracke M, Day R, Pastrick M (1991) Intratester and intertester reliability of the cervical range of motion device. Arch Phys Med Rehabil 72:338–340PubMed
13.
14.
Zurück zum Zitat Dvorak J, Panjabi M, Gerber M, Wichmann W (1987) CT-functional diagnostics of the rotatory instability of upper cervical spine. 1. An experimental study on cadavers. Spine 12:197–205PubMedCrossRef Dvorak J, Panjabi M, Gerber M, Wichmann W (1987) CT-functional diagnostics of the rotatory instability of upper cervical spine. 1. An experimental study on cadavers. Spine 12:197–205PubMedCrossRef
15.
Zurück zum Zitat Fielding JW (1957) Cineroentgenography of the normal cervical spine. J Bone Joint Surg Am 39-A:1280–1288PubMed Fielding JW (1957) Cineroentgenography of the normal cervical spine. J Bone Joint Surg Am 39-A:1280–1288PubMed
16.
Zurück zum Zitat Fisher SV, Bowar JF, Awad EA, Gullickson G Jr (1977) Cervical orthoses effect on cervical spine motion: roentgenographic and goniometric method of study. Arch Phys Med Rehabil 58:109–115PubMed Fisher SV, Bowar JF, Awad EA, Gullickson G Jr (1977) Cervical orthoses effect on cervical spine motion: roentgenographic and goniometric method of study. Arch Phys Med Rehabil 58:109–115PubMed
17.
Zurück zum Zitat Hof AL, Koerhuis CL, Winters JC (2001) ‘Coupled motions’ in cervical spine rotation can be misleading Comment on V. Feipel, B. Rondelet, J.-P. Le Pallec and M. Rooze. Normal global motion of the cervical spine: an electrogoniometric study. Clin. Biomechanics 1999; 14: 462–470. Clin Biomech (Bristol, Avon) 16:455–458. doi:10.1016/S0268-0033(01)00012-2 Hof AL, Koerhuis CL, Winters JC (2001) ‘Coupled motions’ in cervical spine rotation can be misleading Comment on V. Feipel, B. Rondelet, J.-P. Le Pallec and M. Rooze. Normal global motion of the cervical spine: an electrogoniometric study. Clin. Biomechanics 1999; 14: 462–470. Clin Biomech (Bristol, Avon) 16:455–458. doi:10.​1016/​S0268-0033(01)00012-2
18.
Zurück zum Zitat Jones MD, Bard G (1966) Cineradiography of the unstable cervical spine. Am J Phys Med 45:291–295PubMed Jones MD, Bard G (1966) Cineradiography of the unstable cervical spine. Am J Phys Med 45:291–295PubMed
20.
Zurück zum Zitat Kadir N, Grayson MF, Goldberg AA, Swain MC (1981) A new goniometer. Rheumatol Rehabil 20:219–226PubMedCrossRef Kadir N, Grayson MF, Goldberg AA, Swain MC (1981) A new goniometer. Rheumatol Rehabil 20:219–226PubMedCrossRef
21.
Zurück zum Zitat Koerhuis CL, Winters JC, van der Helm FC, Hof AL (2003) Neck mobility measurement by means of the ‘flock of birds’ electromagnetic tracking system. Clin Biomech (Bristol, Avon) 18:14–18. doi:10.1016/S0268-0033(02)00146-8 CrossRef Koerhuis CL, Winters JC, van der Helm FC, Hof AL (2003) Neck mobility measurement by means of the ‘flock of birds’ electromagnetic tracking system. Clin Biomech (Bristol, Avon) 18:14–18. doi:10.​1016/​S0268-0033(02)00146-8 CrossRef
22.
Zurück zum Zitat Lysell E (1969) Motion in the cervical spine. An experimental study on autopsy specimens. Acta Orthop Scand Suppl 123:121 Lysell E (1969) Motion in the cervical spine. An experimental study on autopsy specimens. Acta Orthop Scand Suppl 123:121
23.
Zurück zum Zitat Mayer T, Brady S, Bovasso E, Pope P, Gatchel RJ (1993) Noninvasive measurement of cervical tri-planar motion in normal subjects. Spine 18:2191–2195PubMed Mayer T, Brady S, Bovasso E, Pope P, Gatchel RJ (1993) Noninvasive measurement of cervical tri-planar motion in normal subjects. Spine 18:2191–2195PubMed
24.
Zurück zum Zitat Nilsson N, Hartvigsen J, Christensen HW (1996) Normal ranges of passive cervical motion for women and men 20–60 years old. J Manipulative Physiol Ther 19:306–309PubMed Nilsson N, Hartvigsen J, Christensen HW (1996) Normal ranges of passive cervical motion for women and men 20–60 years old. J Manipulative Physiol Ther 19:306–309PubMed
27.
Zurück zum Zitat Rheault W, Albright B, Byers C, Franta M, Johnson A, Skowronek M, Dougherty J (1992) Intertester reliability of the cervical range of motion device. J Orthop Sports Phys Ther 15:147–150PubMed Rheault W, Albright B, Byers C, Franta M, Johnson A, Skowronek M, Dougherty J (1992) Intertester reliability of the cervical range of motion device. J Orthop Sports Phys Ther 15:147–150PubMed
28.
Zurück zum Zitat Tucci SM, Hicks JE, Gross EG, Campbell W, Danoff J (1986) Cervical motion assessment: a new, simple and accurate method. Arch Phys Med Rehabil 67:225–230PubMed Tucci SM, Hicks JE, Gross EG, Campbell W, Danoff J (1986) Cervical motion assessment: a new, simple and accurate method. Arch Phys Med Rehabil 67:225–230PubMed
29.
Zurück zum Zitat Youdas JW, Carey JR, Garrett TR (1991) Reliability of measurements of cervical spine range of motion–comparison of three methods. Phys Ther 71:98–104 discussion 105–106PubMed Youdas JW, Carey JR, Garrett TR (1991) Reliability of measurements of cervical spine range of motion–comparison of three methods. Phys Ther 71:98–104 discussion 105–106PubMed
30.
Zurück zum Zitat Youdas JW, Garrett TR, Suman VJ, Bogard CL, Hallman HO, Carey JR (1992) Normal range of motion of the cervical spine: an initial goniometric study. Phys Ther 72:770–780PubMed Youdas JW, Garrett TR, Suman VJ, Bogard CL, Hallman HO, Carey JR (1992) Normal range of motion of the cervical spine: an initial goniometric study. Phys Ther 72:770–780PubMed
Metadaten
Titel
Three-dimensional analysis of cervical spine motion: reliability of a computer assisted magnetic tracking device compared to inclinometer
verfasst von
Ioannis D. Gelalis
Louis E. DeFrate
Kosmas S. Stafilas
Emilios E. Pakos
James D. Kang
Lars G. Gilbertson
Publikationsdatum
01.02.2009
Verlag
Springer-Verlag
Erschienen in
European Spine Journal / Ausgabe 2/2009
Print ISSN: 0940-6719
Elektronische ISSN: 1432-0932
DOI
https://doi.org/10.1007/s00586-008-0853-0

Weitere Artikel der Ausgabe 2/2009

European Spine Journal 2/2009 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Arthroskopie kann Knieprothese nicht hinauszögern

25.04.2024 Gonarthrose Nachrichten

Ein arthroskopischer Eingriff bei Kniearthrose macht im Hinblick darauf, ob und wann ein Gelenkersatz fällig wird, offenbar keinen Unterschied.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.