Effect of a novel interspinous implant on lumbar spinal range of motion

verfasst von: Robert Gunzburg, Marek Szpalski, Stuart A. Callary, Christopher J. Colloca, Victor Kosmopoulos, Deed Harrison, Robert J. Moore

Erschienen in: European Spine Journal | Ausgabe 5/2009

Einloggen, um Zugang zu erhalten

Abstract

Interspinous devices have been introduced to provide a minimally invasive surgical alternative for patients with lumbar spinal stenosis or foraminal stenosis. Little is known however, of the effect of interspinous devices on intersegmental range of motion (ROM). The aim of this in vivo study was to investigate the effect of a novel minimally invasive interspinous implant, InSwing®, on sagittal plane ROM of the lumbar spine using an ovine model. Ten adolescent Merino lambs underwent a destabilization procedure at the L1–L2 level simulating a stenotic degenerative spondylolisthesis (as described in our earlier work; Spine 15:571–576, 1990). All animals were placed in a side-lying posture and lateral radiographs were taken in full flexion and extension of the trunk in a standardized manner. Radiographs were repeated following the insertion of an 8-mm InSwing® interspinous device at L1–L2, and again with the implant secured by means of a tension band tightened to 1 N/m around the L1 and L2 spinous processes. ROM was assessed in each of the three conditions and compared using Cobb’s method. A paired t-test compared ROM for each of the experimental conditions (P < 0.05). After instrumentation with the InSwing® interspinous implant, the mean total sagittal ROM (from full extension to full flexion) was reduced by 16% from 6.3° to 5.3 ± 2.7°. The addition of the tension band resulted in a 43% reduction in total sagittal ROM to 3.6 ± 1.9° which approached significance. When looking at flexion only, the addition of the interspinous implant without the tension band did not significantly reduce lumbar flexion, however, a statistically significant 15% reduction in lumbar flexion was observed with the addition of the tension band (P = 0.01). To our knowledge, this is the first in vivo study radiographically showing the advantage of using an interspinous device to stabilize the spine in flexion. These results are important findings particularly for patients with clinical symptoms related to instable degenerative spondylolisthesis.
Literatur
6.
Zurück zum Zitat Floman Y, Millgram MA, Smorgick Y, Rand N, Ashkenazi E (2007) Failure of the Wallis interspinous implant to lower the incidence of recurrent lumbar disc herniations in patients undergoing primary disc excision. J Spinal Disord Tech 20:337–341. doi:10.1097/BSD.0b013e318030a81d PubMedCrossRef Floman Y, Millgram MA, Smorgick Y, Rand N, Ashkenazi E (2007) Failure of the Wallis interspinous implant to lower the incidence of recurrent lumbar disc herniations in patients undergoing primary disc excision. J Spinal Disord Tech 20:337–341. doi:10.​1097/​BSD.​0b013e318030a81d​ PubMedCrossRef
8.
9.
Zurück zum Zitat Graf H (1992) Lumbar instability: surgical treatment without fusion. Rachis 412:123–137 Graf H (1992) Lumbar instability: surgical treatment without fusion. Rachis 412:123–137
10.
Zurück zum Zitat Grevitt MP, Gardner AD, Spilsbury J, Shackleford IM, Baskerville R, Pursell LM, Hassaan A, Mulholland RC (1995) The Graf stabilisation system: early results in 50 patients. Eur Spine J 4:169–175. doi:10.1007/BF00298241 PubMedCrossRef Grevitt MP, Gardner AD, Spilsbury J, Shackleford IM, Baskerville R, Pursell LM, Hassaan A, Mulholland RC (1995) The Graf stabilisation system: early results in 50 patients. Eur Spine J 4:169–175. doi:10.​1007/​BF00298241 PubMedCrossRef
12.
13.
Zurück zum Zitat Kettler A, Drumm J, Heuer F, Haeussler K, Mack C, Claes L, Wilke HJ (2008) Can a modified interspinous spacer prevent instability in axial rotation and lateral bending? A biomechanical in vitro study resulting in a new idea. Clin Biomech (Bristol, Avon) 23:242–247. doi:10.1016/j.clinbiomech.2007.09.004 CrossRef Kettler A, Drumm J, Heuer F, Haeussler K, Mack C, Claes L, Wilke HJ (2008) Can a modified interspinous spacer prevent instability in axial rotation and lateral bending? A biomechanical in vitro study resulting in a new idea. Clin Biomech (Bristol, Avon) 23:242–247. doi:10.​1016/​j.​clinbiomech.​2007.​09.​004 CrossRef
15.
Zurück zum Zitat Kim KA, McDonald M, Pik JH, Khoueir P, Wang MY (2007) Dynamic intraspinous spacer technology for posterior stabilization: case-control study on the safety, sagittal angulation, and pain outcome at 1-year follow-up evaluation. Neurosurg Focus 22:E7–E9PubMed Kim KA, McDonald M, Pik JH, Khoueir P, Wang MY (2007) Dynamic intraspinous spacer technology for posterior stabilization: case-control study on the safety, sagittal angulation, and pain outcome at 1-year follow-up evaluation. Neurosurg Focus 22:E7–E9PubMed
16.
Zurück zum Zitat Kong DS, Kim ES, Eoh W (2007) One-year outcome evaluation after interspinous implantation for degenerative spinal stenosis with segmental instability. J Korean Med Sci 22:330–335PubMed Kong DS, Kim ES, Eoh W (2007) One-year outcome evaluation after interspinous implantation for degenerative spinal stenosis with segmental instability. J Korean Med Sci 22:330–335PubMed
17.
Zurück zum Zitat Laudet CG, Elberg JF, Robine D (1993) Comportement biomécanique d’un ressort inter-apophysaire vertébral postérieur analyse expérimentale du comportement discal en compression et en flexion/extension. Rachis 5:3–7 Laudet CG, Elberg JF, Robine D (1993) Comportement biomécanique d’un ressort inter-apophysaire vertébral postérieur analyse expérimentale du comportement discal en compression et en flexion/extension. Rachis 5:3–7
18.
Zurück zum Zitat Leahy JC, Mathias KJ, Heaton A, Shepherd DE, Hukins DW, Deans WF, Brian MW, Wardlaw D (2000) Design of spinous process hooks for flexible fixation of the lumbar spine. Proc Inst Mech Eng [H] 214:479–487. doi:10.1243/0954411001535507 Leahy JC, Mathias KJ, Heaton A, Shepherd DE, Hukins DW, Deans WF, Brian MW, Wardlaw D (2000) Design of spinous process hooks for flexible fixation of the lumbar spine. Proc Inst Mech Eng [H] 214:479–487. doi:10.​1243/​0954411001535507​
19.
Zurück zum Zitat Levin DA, Hale JJ, Bendo JA (2007) Adjacent segment degeneration following spinal fusion for degenerative disc disease. Bull NYU Hosp Jt Dis 65:29–36PubMed Levin DA, Hale JJ, Bendo JA (2007) Adjacent segment degeneration following spinal fusion for degenerative disc disease. Bull NYU Hosp Jt Dis 65:29–36PubMed
21.
Zurück zum Zitat Mimura M, Panjabi MM, Oxland TR, Crisco JJ, Yamamoto I, Vasavada A (1994) Disc degeneration affects the multidirectional flexibility of the lumbar spine. Spine 19:1371–1380PubMed Mimura M, Panjabi MM, Oxland TR, Crisco JJ, Yamamoto I, Vasavada A (1994) Disc degeneration affects the multidirectional flexibility of the lumbar spine. Spine 19:1371–1380PubMed
26.
Zurück zum Zitat Resnick DK, Choudhri TF, Dailey AT, Groff MW, Khoo L, Matz PG, Mummaneni P, Watters WCIII, Wang J, Walters BC, Hadley MN (2005) Guidelines for the performance of fusion procedures for degenerative disease of the lumbar spine. Part 9: fusion in patients with stenosis and spondylolisthesis. J Neurosurg Spine 2:679–685PubMedCrossRef Resnick DK, Choudhri TF, Dailey AT, Groff MW, Khoo L, Matz PG, Mummaneni P, Watters WCIII, Wang J, Walters BC, Hadley MN (2005) Guidelines for the performance of fusion procedures for degenerative disease of the lumbar spine. Part 9: fusion in patients with stenosis and spondylolisthesis. J Neurosurg Spine 2:679–685PubMedCrossRef
28.
Zurück zum Zitat Schmoelz W, Huber JF, Nydegger T, Dipl I, Claes L, Wilke HJ (2003) Dynamic stabilization of the lumbar spine and its effects on adjacent segments: an in vitro experiment. J Spinal Disord Tech 16:418–423PubMed Schmoelz W, Huber JF, Nydegger T, Dipl I, Claes L, Wilke HJ (2003) Dynamic stabilization of the lumbar spine and its effects on adjacent segments: an in vitro experiment. J Spinal Disord Tech 16:418–423PubMed
29.
Zurück zum Zitat Senegas J (2002) Mechanical supplementation by non-rigid fixation in degenerative intervertebral lumbar segments: the Wallis system. Eur Spine J 11(Suppl 2):S164–S169PubMed Senegas J (2002) Mechanical supplementation by non-rigid fixation in degenerative intervertebral lumbar segments: the Wallis system. Eur Spine J 11(Suppl 2):S164–S169PubMed
31.
Zurück zum Zitat Singer KP, Edmondston SJ, Day RE, Breidahl WH (1994) Computer-assisted curvature assessment and Cobb angle determination of the thoracic kyphosis. An in vivo and in vitro comparison. Spine 19:1381–1384PubMedCrossRef Singer KP, Edmondston SJ, Day RE, Breidahl WH (1994) Computer-assisted curvature assessment and Cobb angle determination of the thoracic kyphosis. An in vivo and in vitro comparison. Spine 19:1381–1384PubMedCrossRef
32.
Zurück zum Zitat Sterna J, Chopek L, Ciupik A, Dobkiewicz J, Pienazek J, Radek M, Szpalski M (2006) Evaluation of implantation procedure and research of influence of Inter-S dynamic tensioning stabilization system. In: XIII Meeting of the Neuroorthopedic Section of the Polish Society of Neurosurgery, Zakopane, April 27–29, 2006 Sterna J, Chopek L, Ciupik A, Dobkiewicz J, Pienazek J, Radek M, Szpalski M (2006) Evaluation of implantation procedure and research of influence of Inter-S dynamic tensioning stabilization system. In: XIII Meeting of the Neuroorthopedic Section of the Polish Society of Neurosurgery, Zakopane, April 27–29, 2006
33.
Zurück zum Zitat Stoll TM, Dubois G, Schwarzenbach O (2002) The dynamic neutralization system for the spine: a multi-center study of a novel non-fusion system. Eur Spine J 11(Suppl 2):S170–S178PubMed Stoll TM, Dubois G, Schwarzenbach O (2002) The dynamic neutralization system for the spine: a multi-center study of a novel non-fusion system. Eur Spine J 11(Suppl 2):S170–S178PubMed
35.
36.
Zurück zum Zitat Tsai KJ, Murakami H, Lowery GL, Hutton WC (2006) A biomechanical evaluation of an interspinous device (Coflex) used to stabilize the lumbar spine. J Surg Orthop Adv 15:167–172PubMed Tsai KJ, Murakami H, Lowery GL, Hutton WC (2006) A biomechanical evaluation of an interspinous device (Coflex) used to stabilize the lumbar spine. J Surg Orthop Adv 15:167–172PubMed
37.
Zurück zum Zitat Verhoof OJ, Bron JL, Wapstra FH, Van Royen BJ (2008) High failure rate of the interspinous distraction device (X-Stop) for the treatment of lumbar spinal stenosis caused by degenerative spondylolisthesis. Eur Spine J 17:188–192. doi:10.1007/s00586-007-0492-x PubMedCrossRef Verhoof OJ, Bron JL, Wapstra FH, Van Royen BJ (2008) High failure rate of the interspinous distraction device (X-Stop) for the treatment of lumbar spinal stenosis caused by degenerative spondylolisthesis. Eur Spine J 17:188–192. doi:10.​1007/​s00586-007-0492-x PubMedCrossRef
38.
Zurück zum Zitat Weinstein JN, Tosteson TD, Lurie JD, Tosteson AN, Blood E, Hanscom B, Herkowitz H, Cammisa F, Albert T, Boden SD, Hilibrand A, Goldberg H, Berven S, An H (2008) Surgical versus nonsurgical therapy for lumbar spinal stenosis. N Engl J Med 358:794–810. doi:10.1056/NEJMoa0707136 PubMedCrossRef Weinstein JN, Tosteson TD, Lurie JD, Tosteson AN, Blood E, Hanscom B, Herkowitz H, Cammisa F, Albert T, Boden SD, Hilibrand A, Goldberg H, Berven S, An H (2008) Surgical versus nonsurgical therapy for lumbar spinal stenosis. N Engl J Med 358:794–810. doi:10.​1056/​NEJMoa0707136 PubMedCrossRef
40.
Zurück zum Zitat Zucherman JF, Hsu KY, Hartjen CA, Mehalic TF, Implicito DA, Martin MJ, Johnson DR, Skidmore GA, Vessa PP, Dwyer JW, Puccio ST, Cauthen JC, Ozuna RM (2005) A multicenter, prospective, randomized trial evaluating the X STOP interspinous process decompression system for the treatment of neurogenic intermittent claudication: two-year follow-up results. Spine 30:1351–1358. doi:10.1097/01.brs.0000166618.42749.d1 PubMedCrossRef Zucherman JF, Hsu KY, Hartjen CA, Mehalic TF, Implicito DA, Martin MJ, Johnson DR, Skidmore GA, Vessa PP, Dwyer JW, Puccio ST, Cauthen JC, Ozuna RM (2005) A multicenter, prospective, randomized trial evaluating the X STOP interspinous process decompression system for the treatment of neurogenic intermittent claudication: two-year follow-up results. Spine 30:1351–1358. doi:10.​1097/​01.​brs.​0000166618.​42749.​d1 PubMedCrossRef
Metadaten
Titel
Effect of a novel interspinous implant on lumbar spinal range of motion
verfasst von
Robert Gunzburg
Marek Szpalski
Stuart A. Callary
Christopher J. Colloca
Victor Kosmopoulos
Deed Harrison
Robert J. Moore
Publikationsdatum
01.05.2009
Verlag
Springer-Verlag
Erschienen in
European Spine Journal / Ausgabe 5/2009
Print ISSN: 0940-6719
Elektronische ISSN: 1432-0932
DOI
https://doi.org/10.1007/s00586-009-0890-3

Weitere Artikel der Ausgabe 5/2009

European Spine Journal 5/2009 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Proximale Humerusfraktur: Auch 100-Jährige operieren?

01.05.2024 DCK 2024 Kongressbericht

Mit dem demographischen Wandel versorgt auch die Chirurgie immer mehr betagte Menschen. Von Entwicklungen wie Fast-Track können auch ältere Menschen profitieren und bei proximaler Humerusfraktur können selbst manche 100-Jährige noch sicher operiert werden.

Sind Frauen die fähigeren Ärzte?

30.04.2024 Gendermedizin Nachrichten

Patienten, die von Ärztinnen behandelt werden, dürfen offenbar auf bessere Therapieergebnisse hoffen als Patienten von Ärzten. Besonders gilt das offenbar für weibliche Kranke, wie eine Studie zeigt.

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Arthroskopie kann Knieprothese nicht hinauszögern

25.04.2024 Gonarthrose Nachrichten

Ein arthroskopischer Eingriff bei Kniearthrose macht im Hinblick darauf, ob und wann ein Gelenkersatz fällig wird, offenbar keinen Unterschied.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.