Spatial relation between a rigid (digital) intraoral X-ray receptor and longitudinal axes of maxillary teeth

verfasst von: Felix Roeder, Isabell von Rechenberg, Bernd d’Hoedt, Ralf Schulze

Erschienen in: Clinical Oral Investigations | Ausgabe 5/2011

Einloggen, um Zugang zu erhalten

Abstract

The purpose of this study was to quantify the existing (inevitable) angle which in intraoral radiology appears between tooth length axis and receptor caused by the anatomical situation. Especially in the upper jaw, due to its arched anatomy, a true “paralleling technique” is not achievable. The angulation necessarily causes distortion and a foreshortening of the image; hence, the foreshortened image leads to misinterpretations in diagnostics. We investigated the effects of the realistic angulation on these image deteriorating factors. Two hundred ninety-four plaster models of the upper jaw were collected, and the angles between a dummy receptor and the axes of the central incisor or the first molar were measured. For evaluation, a rigid dummy of an intraoral charge-coupled device (CCD) receptor (30 mm × 40 mm) was used. The mean angulation evaluated for central incisors was 36.7° (range 19–56°) and for first molars 42.5° (range 26–56°). This leads to a foreshortening of the tooth ranging from 5.4% to 44.1% in the image, when magnification is neglected. Large angles of up to 56°, in both incisor and molar region, result in a relevant underestimation of true tooth length up to 44%. It is important to note that this error cannot be simply corrected by means of local magnification correction. Techniques should be developed that allow for automated assessment of the effective angle to provide information for distortion correction.
Literatur
1.
Zurück zum Zitat Van Aken J (1969) Optimum conditions for intraoral roentgenograms. Oral Surg 27:475–491CrossRef Van Aken J (1969) Optimum conditions for intraoral roentgenograms. Oral Surg 27:475–491CrossRef
2.
Zurück zum Zitat Schulze R, d’Hoedt B (2001) Mathematical analysis of projection errors in “paralleling technique” with respect to implant geometry. Clin Oral Implants Res 12:364–371PubMedCrossRef Schulze R, d’Hoedt B (2001) Mathematical analysis of projection errors in “paralleling technique” with respect to implant geometry. Clin Oral Implants Res 12:364–371PubMedCrossRef
3.
Zurück zum Zitat Schulze R, d'Hoedt B (2002) A method to calculate angular disparities between object and receptor in “paralleling technique”. Dentomaxillofac Radiol 31:32–38PubMedCrossRef Schulze R, d'Hoedt B (2002) A method to calculate angular disparities between object and receptor in “paralleling technique”. Dentomaxillofac Radiol 31:32–38PubMedCrossRef
4.
Zurück zum Zitat Roeder F, Brüllmann D, d'Hoedt B, Schulze R (2010) Ex vivo radiographic tooth length measurements with the reference sphere method (RSM). Clin Oral Invest (in press) Roeder F, Brüllmann D, d'Hoedt B, Schulze R (2010) Ex vivo radiographic tooth length measurements with the reference sphere method (RSM). Clin Oral Invest (in press)
5.
Zurück zum Zitat Eickholz P, Hausmann E (2000) Accuracy of radiographic assessment of interproximal bone loss in intrabony defects using linear measurements. Eur J Oral Sci 108:70–73PubMedCrossRef Eickholz P, Hausmann E (2000) Accuracy of radiographic assessment of interproximal bone loss in intrabony defects using linear measurements. Eur J Oral Sci 108:70–73PubMedCrossRef
6.
Zurück zum Zitat Kim TS, Obst C, Zehaczek S, Geenen C (2008) Detection of bone loss with different X-ray techniques in periodontal patients. J Periodontol 79:1141–1149PubMedCrossRef Kim TS, Obst C, Zehaczek S, Geenen C (2008) Detection of bone loss with different X-ray techniques in periodontal patients. J Periodontol 79:1141–1149PubMedCrossRef
7.
Zurück zum Zitat Schropp L, Stavropoulos A, Gotfredsen E, Wenzal A (2009) Calibration of radiographs by a reference metal ball affects preoperative selection of implant size. Clin Oral Investig 13:375–381PubMedCrossRef Schropp L, Stavropoulos A, Gotfredsen E, Wenzal A (2009) Calibration of radiographs by a reference metal ball affects preoperative selection of implant size. Clin Oral Investig 13:375–381PubMedCrossRef
8.
Zurück zum Zitat Tyndall AA, Brooks SL (2000) Selection criteria for dental implant site imaging: a position paper of the American Academy of Oral and Maxillofacial Radiology. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 89:630–637PubMedCrossRef Tyndall AA, Brooks SL (2000) Selection criteria for dental implant site imaging: a position paper of the American Academy of Oral and Maxillofacial Radiology. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 89:630–637PubMedCrossRef
9.
Zurück zum Zitat BouSerhal C, Jacobs R, Quirynen M, van Steenberghe D (2002) Imaging technique selection for the preoperative planning of oral implants: a review of the literature. Clin Implant Dent Relat Res 4:156–172PubMedCrossRef BouSerhal C, Jacobs R, Quirynen M, van Steenberghe D (2002) Imaging technique selection for the preoperative planning of oral implants: a review of the literature. Clin Implant Dent Relat Res 4:156–172PubMedCrossRef
10.
Zurück zum Zitat Williams CB, Joyce AP, Roberts S (2006) A comparison between in vivo radiographic working length determination and measurement after extraction. J Endod 32:624–627PubMedCrossRef Williams CB, Joyce AP, Roberts S (2006) A comparison between in vivo radiographic working length determination and measurement after extraction. J Endod 32:624–627PubMedCrossRef
11.
Zurück zum Zitat Behneke N, Tetsch P (1985) Diagnostik und Planung von Implantaten im zahnlosen Unterkiefer. Fortsch Zahnärztl Implantol 1:266–271 Behneke N, Tetsch P (1985) Diagnostik und Planung von Implantaten im zahnlosen Unterkiefer. Fortsch Zahnärztl Implantol 1:266–271
12.
Zurück zum Zitat Benn DK (1990) A review of the reliability of radiographic measurements in estimating alveolar bone changes. J Clin Periodontol 17:14–21PubMedCrossRef Benn DK (1990) A review of the reliability of radiographic measurements in estimating alveolar bone changes. J Clin Periodontol 17:14–21PubMedCrossRef
13.
Zurück zum Zitat Dubrez B, Jacot-Dsecombes S, Cimasoni G (1995) Reliability of a paralleling instrument for dental radiographs. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 80:358–364PubMedCrossRef Dubrez B, Jacot-Dsecombes S, Cimasoni G (1995) Reliability of a paralleling instrument for dental radiographs. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 80:358–364PubMedCrossRef
14.
Zurück zum Zitat Zappa U, Simona C, Graf H, van Aken J (1991) In vivo determination of radiographic projection errors produced by a novel filmholder and an x-ray beam manipulator. J Periodontol 62:674–683PubMed Zappa U, Simona C, Graf H, van Aken J (1991) In vivo determination of radiographic projection errors produced by a novel filmholder and an x-ray beam manipulator. J Periodontol 62:674–683PubMed
15.
Zurück zum Zitat Dixon CA, Hildebolt CF (2005) An overview of radiographic film holders. Dentomaxillofac Radiol 34:67–73PubMedCrossRef Dixon CA, Hildebolt CF (2005) An overview of radiographic film holders. Dentomaxillofac Radiol 34:67–73PubMedCrossRef
16.
Zurück zum Zitat Wu JC, Huang JN, Zhao SF, Xu XJ, Zhang JC, Xia B, Dong YF (2005) Use of a simple intraoral instrument to standardize film alignment and improve image reproducibility. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 100:99–104PubMedCrossRef Wu JC, Huang JN, Zhao SF, Xu XJ, Zhang JC, Xia B, Dong YF (2005) Use of a simple intraoral instrument to standardize film alignment and improve image reproducibility. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 100:99–104PubMedCrossRef
17.
Zurück zum Zitat Schulze R, Bruellmann DD, Roeder F, d'Hoedt B (2004) Determination of projection geometry from quantitative assessment of the distortion of spherical references in single-view projection radiography. Med Phys 31:2849–2854PubMedCrossRef Schulze R, Bruellmann DD, Roeder F, d'Hoedt B (2004) Determination of projection geometry from quantitative assessment of the distortion of spherical references in single-view projection radiography. Med Phys 31:2849–2854PubMedCrossRef
18.
Zurück zum Zitat Schulze RKW (2010) Pose determination of a cylindrical (dental) implant in three-dimensions from a single two-dimensional radiograph. Dentomaxillofac Radiol 39:33–41PubMedCrossRef Schulze RKW (2010) Pose determination of a cylindrical (dental) implant in three-dimensions from a single two-dimensional radiograph. Dentomaxillofac Radiol 39:33–41PubMedCrossRef
19.
Zurück zum Zitat Schulze R, Roeder F (2009) Device, bite element and system for intraoral radiography. Patent application: DE102 00 901 7905.4, applicant: Johannes Gutenberg-University of Mainz Schulze R, Roeder F (2009) Device, bite element and system for intraoral radiography. Patent application: DE102 00 901 7905.4, applicant: Johannes Gutenberg-University of Mainz
20.
Zurück zum Zitat Mühlreiter E, de Jonge TE (1928) Anatomie des menschlichen Gebisses, 5th edn. Felix Verlag, Leipzig, pp 25–43 Mühlreiter E, de Jonge TE (1928) Anatomie des menschlichen Gebisses, 5th edn. Felix Verlag, Leipzig, pp 25–43
21.
Zurück zum Zitat Jenkins SM, Dummer PM, Newcombe RG (1995) Radiographic amelocemental junction and alveolar crest: effect of x-ray beam angulation. J Oral Rehabil 22:679–684PubMedCrossRef Jenkins SM, Dummer PM, Newcombe RG (1995) Radiographic amelocemental junction and alveolar crest: effect of x-ray beam angulation. J Oral Rehabil 22:679–684PubMedCrossRef
22.
Zurück zum Zitat Santos-Pinto L, Cordeiro Rde C, Zuanon AC, Basso MD, Gonçalves MA (2007) Primary tooth length determination in direct digital radiography: an in vivo study. Pediatr Dent 29:470–474PubMed Santos-Pinto L, Cordeiro Rde C, Zuanon AC, Basso MD, Gonçalves MA (2007) Primary tooth length determination in direct digital radiography: an in vivo study. Pediatr Dent 29:470–474PubMed
23.
Zurück zum Zitat Krithika AC, Kandaswamy D, Velmurugan N, Krishna VG (2008) Non-metallic grid for radiographic measurement. Aust Endod J 34:36–38PubMedCrossRef Krithika AC, Kandaswamy D, Velmurugan N, Krishna VG (2008) Non-metallic grid for radiographic measurement. Aust Endod J 34:36–38PubMedCrossRef
Metadaten
Titel
Spatial relation between a rigid (digital) intraoral X-ray receptor and longitudinal axes of maxillary teeth
verfasst von
Felix Roeder
Isabell von Rechenberg
Bernd d’Hoedt
Ralf Schulze
Publikationsdatum
01.10.2011
Verlag
Springer-Verlag
Erschienen in
Clinical Oral Investigations / Ausgabe 5/2011
Print ISSN: 1432-6981
Elektronische ISSN: 1436-3771
DOI
https://doi.org/10.1007/s00784-010-0430-x

Weitere Artikel der Ausgabe 5/2011

Clinical Oral Investigations 5/2011 Zur Ausgabe

Parodontalbehandlung verbessert Prognose bei Katheterablation

19.04.2024 Vorhofflimmern Nachrichten

Werden Personen mit Vorhofflimmern in der Blanking-Periode nach einer Katheterablation gegen eine bestehende Parodontitis behandelt, verbessert dies die Erfolgsaussichten. Dafür sprechen die Resultate einer prospektiven Untersuchung.

Invasive Zahnbehandlung: Wann eine Antibiotikaprophylaxe vor infektiöser Endokarditis schützt

11.04.2024 Endokarditis Nachrichten

Bei welchen Personen eine Antibiotikaprophylaxe zur Prävention einer infektiösen Endokarditis nach invasiven zahnärztlichen Eingriffen sinnvoll ist, wird diskutiert. Neue Daten stehen im Einklang mit den europäischen Leitlinienempfehlungen.

Zell-Organisatoren unter Druck: Mechanismen des embryonalen Zahnwachstums aufgedeckt

08.04.2024 Zahnmedizin Nachrichten

Der Aufbau von Geweben und Organen während der Embryonalentwicklung wird von den Zellen bemerkenswert choreografiert. Für diesen Prozess braucht es spezielle sogenannte „Organisatoren“. In einer aktuellen Veröffentlichung im Fachjournal Nature Cell Biology berichten Forschende durch welchen Vorgang diese Organisatoren im Gewebe entstehen und wie sie dann die Bildung von Zähnen orchestrieren.

Die Oralprophylaxe & Kinderzahnheilkunde umbenannt

11.03.2024 Kinderzahnmedizin Nachrichten

Infolge der Umbenennung der Deutschen Gesellschaft für Kinderzahnheilkunde in Deutsche Gesellschaft für Kinderzahnmedizin (DGKiZ) wird deren Mitgliederzeitschrift Oralprophylaxe & Kinderzahnheilkunde in Oralprophylaxe & Kinderzahnmedizin umbenannt. Aus diesem Grunde trägt die erste Ausgabe in 2024 erstmalig den neuen Titel.

Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Zahnmedizin und bleiben Sie gut informiert – ganz bequem per eMail.