Skip to main content
Erschienen in: Clinical Oral Investigations 6/2015

01.07.2015 | Original Article

Template-guided vs. non-guided drilling in site preparation of dental implants

verfasst von: Uta Scherer, Marcus Stoetzer, Martin Ruecker, Nils-Claudius Gellrich, Constantin von See

Erschienen in: Clinical Oral Investigations | Ausgabe 6/2015

Einloggen, um Zugang zu erhalten

Abstract

Objectives

Clinical success of oral implants is related to primary stability and osseointegration. These parameters are associated with delicate surgical techniques. We herein studied whether template-guided drilling has a significant influence on drillholes diameter and accuracy in an in vitro model.

Materials and methods

Fresh cadaveric porcine mandibles were used for drilling experiments of four experimental groups. Each group consisted of three operators, comparing guide templates for drilling with free-handed procedure. Operators without surgical knowledge were grouped together, contrasting highly experienced oral surgeons in other groups. A total of 180 drilling actions were performed, and diameters were recorded at multiple depth levels, with a precision measuring instrument.

Results

Template-guided drilling procedure improved accuracy on a very significant level in comparison with free-handed drilling operation (p ≤ 0.001). Inaccuracy of free-handed drilling became more significant in relation to measurement depth. High homogenic uniformity of template-guided drillholes was significantly stronger than unguided drilling operations by highly experienced oral surgeons (p ≤ 0.001).

Conclusion

Template-guided drilling procedure leads to significantly enhanced accuracy. Significant results compared to free-handed drilling actions were achieved, irrespective of the clinical experience level of the operator.

Clinical relevance

Template-guided drilling procedures lead to a more predictable clinical diameter. It shows that any set of instruments has to be carefully chosen to match the specific implant system. The current in vitro study is implicating an improvement of implant bed preparation but needs to be confirmed in clinical studies.
Literatur
1.
Zurück zum Zitat Shalabi MM, Gortemaker A, Van’t Hof MA, Jansen JA, Creugers NH (2006) Implant surface roughness and bone healing: a systematic review. J Dent Res 85(6):496–500PubMedCrossRef Shalabi MM, Gortemaker A, Van’t Hof MA, Jansen JA, Creugers NH (2006) Implant surface roughness and bone healing: a systematic review. J Dent Res 85(6):496–500PubMedCrossRef
2.
Zurück zum Zitat Albrektsson T, Branemark PI, Hansson HA, Lindstrom J (1981) Osseointegrated titanium implants. Requirements for ensuring a long-lasting, direct bone-to-implant anchorage in man. Acta OrthopScand 52(2):155–170CrossRef Albrektsson T, Branemark PI, Hansson HA, Lindstrom J (1981) Osseointegrated titanium implants. Requirements for ensuring a long-lasting, direct bone-to-implant anchorage in man. Acta OrthopScand 52(2):155–170CrossRef
3.
Zurück zum Zitat Atsumi M, Park SH, Wang HL (2007) Methods used to assess implant stability: current status. Int J Oral Maxillofac Implants 22(5):743–754PubMed Atsumi M, Park SH, Wang HL (2007) Methods used to assess implant stability: current status. Int J Oral Maxillofac Implants 22(5):743–754PubMed
4.
Zurück zum Zitat Akca K, Chang TL, Tekdemir I, Fanuscu MI (2006) Biomechanical aspects of initial intraosseous stability and implant design: a quantitative micro-morphometric analysis. Clin Oral Implants Res 17(4):465–472PubMedCrossRef Akca K, Chang TL, Tekdemir I, Fanuscu MI (2006) Biomechanical aspects of initial intraosseous stability and implant design: a quantitative micro-morphometric analysis. Clin Oral Implants Res 17(4):465–472PubMedCrossRef
5.
Zurück zum Zitat Akkocaoglu M, Uysal S, Tekdemir I, Akca K, Cehreli MC (2005) Implant design and intraosseous stability of immediately placed implants: a human cadaver study. Clin Oral Implants Res 16(2):202–209PubMedCrossRef Akkocaoglu M, Uysal S, Tekdemir I, Akca K, Cehreli MC (2005) Implant design and intraosseous stability of immediately placed implants: a human cadaver study. Clin Oral Implants Res 16(2):202–209PubMedCrossRef
6.
Zurück zum Zitat Brunski JB (1992) Biomechanical factors affecting the bone-dental implant interface. Clin Mater 10(3):153–201PubMedCrossRef Brunski JB (1992) Biomechanical factors affecting the bone-dental implant interface. Clin Mater 10(3):153–201PubMedCrossRef
7.
Zurück zum Zitat Haider R, Watzek G, Plenk H (1993) Effects of drill cooling and bone structure on IMZ implant fixation. Int J Oral Maxillofac Implants 8(1):83–91PubMed Haider R, Watzek G, Plenk H (1993) Effects of drill cooling and bone structure on IMZ implant fixation. Int J Oral Maxillofac Implants 8(1):83–91PubMed
8.
Zurück zum Zitat Weiss CM (1986) Tissue integration of dental endosseous implants: description and comparative analysis of the fibro-osseous integration and osseous integration systems. J Oral Implantol 12(2):169–214PubMed Weiss CM (1986) Tissue integration of dental endosseous implants: description and comparative analysis of the fibro-osseous integration and osseous integration systems. J Oral Implantol 12(2):169–214PubMed
9.
Zurück zum Zitat Queiroz TP, Souza FA, Okamoto R, Margonar R, Pereira-Filho VA, Garcia Junior IR et al (2008) Evaluation of immediate bone-cell viability and of drill wear after implant osteotomies: immunohistochemistry and scanning electron microscopy analysis. J Oral MaxillofacSurg 66(6):1233–1240CrossRef Queiroz TP, Souza FA, Okamoto R, Margonar R, Pereira-Filho VA, Garcia Junior IR et al (2008) Evaluation of immediate bone-cell viability and of drill wear after implant osteotomies: immunohistochemistry and scanning electron microscopy analysis. J Oral MaxillofacSurg 66(6):1233–1240CrossRef
10.
Zurück zum Zitat Brisman DL (1996) The effect of speed, pressure, and time on bone temperature during the drilling of implant sites. Int J Oral Maxillofac Implants 11(1):35–37PubMed Brisman DL (1996) The effect of speed, pressure, and time on bone temperature during the drilling of implant sites. Int J Oral Maxillofac Implants 11(1):35–37PubMed
11.
Zurück zum Zitat Tuijthof GJ, Fruhwirt C, Kment C (2013) Influence of tool geometry on drilling performance of cortical and trabecular bone. Med EngPhys 35(8):1165–1172CrossRef Tuijthof GJ, Fruhwirt C, Kment C (2013) Influence of tool geometry on drilling performance of cortical and trabecular bone. Med EngPhys 35(8):1165–1172CrossRef
12.
Zurück zum Zitat Ercoli C, Funkenbusch PD, Lee HJ, Moss ME, Graser GN (2004) The influence of drill wear on cutting efficiency and heat production during osteotomy preparation for dental implants: a study of drill durability. Int J Oral Maxillofac Implants 19(3):335–349PubMed Ercoli C, Funkenbusch PD, Lee HJ, Moss ME, Graser GN (2004) The influence of drill wear on cutting efficiency and heat production during osteotomy preparation for dental implants: a study of drill durability. Int J Oral Maxillofac Implants 19(3):335–349PubMed
13.
Zurück zum Zitat Karaca F, Aksakal B, Kom M (2011) Influence of orthopaedic drilling parameters on temperature and histopathology of bovine tibia: an in vitro study. Med EngPhys 33(10):1221–1227CrossRef Karaca F, Aksakal B, Kom M (2011) Influence of orthopaedic drilling parameters on temperature and histopathology of bovine tibia: an in vitro study. Med EngPhys 33(10):1221–1227CrossRef
14.
Zurück zum Zitat Wu SW, Lee CC, Fu PY, Lin SC (2012) The effects of flute shape and thread profile on the insertion torque and primary stability of dental implants. Med Eng Phys 34(7):797–805PubMedCrossRef Wu SW, Lee CC, Fu PY, Lin SC (2012) The effects of flute shape and thread profile on the insertion torque and primary stability of dental implants. Med Eng Phys 34(7):797–805PubMedCrossRef
15.
Zurück zum Zitat Yerby S, Scott CC, Evans NJ, Messing KL, Carter DR (2001) Effect of cutting flute design on cortical bone screw insertion torque and pullout strength. J Orthop Trauma 15(3):216–221PubMedCrossRef Yerby S, Scott CC, Evans NJ, Messing KL, Carter DR (2001) Effect of cutting flute design on cortical bone screw insertion torque and pullout strength. J Orthop Trauma 15(3):216–221PubMedCrossRef
16.
Zurück zum Zitat Natali C, Ingle P, Dowell J (1996) Orthopaedic bone drills—can they be improved? Temperature changes near the drilling face. J Bone Joint Surg (Br) 78(3):357–362 Natali C, Ingle P, Dowell J (1996) Orthopaedic bone drills—can they be improved? Temperature changes near the drilling face. J Bone Joint Surg (Br) 78(3):357–362
17.
Zurück zum Zitat Markovic A, Calvo-Guirado JL, Lazic Z, Gomez-Moreno G, Calasan D, Guardia J et al (2013) Evaluation of primary stability of self-tapping and non-self-tapping dental implants. A 12-week clinical study. Clin Implant Dent Relat Res 15(3):341–349PubMedCrossRef Markovic A, Calvo-Guirado JL, Lazic Z, Gomez-Moreno G, Calasan D, Guardia J et al (2013) Evaluation of primary stability of self-tapping and non-self-tapping dental implants. A 12-week clinical study. Clin Implant Dent Relat Res 15(3):341–349PubMedCrossRef
18.
Zurück zum Zitat Cordioli G, Majzoub Z (1997) Heat generation during implant site preparation: an in vitro study. Int J Oral Maxillofac Implants 12(2):186–193PubMed Cordioli G, Majzoub Z (1997) Heat generation during implant site preparation: an in vitro study. Int J Oral Maxillofac Implants 12(2):186–193PubMed
19.
Zurück zum Zitat Sutter F, Krekeler G, Schwammberger AE, Sutter FJ (1992) A traumatic surgical technique and implant bed preparation. Quintessence Int 23(12):811–816PubMed Sutter F, Krekeler G, Schwammberger AE, Sutter FJ (1992) A traumatic surgical technique and implant bed preparation. Quintessence Int 23(12):811–816PubMed
20.
Zurück zum Zitat MacAvelia T, Salahi M, Olsen M, Crookshank M, Schemitsch EH, Ghasempoor A et al (2012) Biomechanical measurements of surgical drilling force and torque in human versus artificial femurs. J Biomech Eng 134(12):124503PubMedCrossRef MacAvelia T, Salahi M, Olsen M, Crookshank M, Schemitsch EH, Ghasempoor A et al (2012) Biomechanical measurements of surgical drilling force and torque in human versus artificial femurs. J Biomech Eng 134(12):124503PubMedCrossRef
21.
Zurück zum Zitat Macavelia T, Ghasempoor A, Janabi-Sharifi F (2012) Force and torque modelling of drilling simulation for orthopaedic surgery. Comput Methods Biomech Biomed Engin Macavelia T, Ghasempoor A, Janabi-Sharifi F (2012) Force and torque modelling of drilling simulation for orthopaedic surgery. Comput Methods Biomech Biomed Engin
22.
Zurück zum Zitat Hobkirk JA, Rusiniak K (1977) Investigation of variable factors in drilling bone. J Oral Surg 35(12):968–973PubMed Hobkirk JA, Rusiniak K (1977) Investigation of variable factors in drilling bone. J Oral Surg 35(12):968–973PubMed
23.
Zurück zum Zitat Tehemar SH (1999) Factors affecting heat generation during implant site preparation: a review of biologic observations and future considerations. Int J Oral Maxillofac Implants 14(1):127–136PubMed Tehemar SH (1999) Factors affecting heat generation during implant site preparation: a review of biologic observations and future considerations. Int J Oral Maxillofac Implants 14(1):127–136PubMed
24.
Zurück zum Zitat Vercruyssen M, Jacobs R, Van Assche N, van Steenberghe D (2008) The use of CT scan based planning for oral rehabilitation by means of implants and its transfer to the surgical field: a critical review on accuracy. J Oral Rehabil 35(6):454–474PubMedCrossRef Vercruyssen M, Jacobs R, Van Assche N, van Steenberghe D (2008) The use of CT scan based planning for oral rehabilitation by means of implants and its transfer to the surgical field: a critical review on accuracy. J Oral Rehabil 35(6):454–474PubMedCrossRef
25.
Zurück zum Zitat Van Assche N, Vercruyssen M, Coucke W, Teughels W, Jacobs R, Quirynen M (2012) Accuracy of computer-aided implant placement. Clin Oral Implants Res 23(Suppl 6):112–123PubMedCrossRef Van Assche N, Vercruyssen M, Coucke W, Teughels W, Jacobs R, Quirynen M (2012) Accuracy of computer-aided implant placement. Clin Oral Implants Res 23(Suppl 6):112–123PubMedCrossRef
26.
Zurück zum Zitat Koop R, Vercruyssen M, Vermeulen K, Quirynen M (2013) Tolerance within the sleeve inserts of different surgical guides for guided implant surgery. Clin Oral Implants Res 24(6):630–634PubMedCrossRef Koop R, Vercruyssen M, Vermeulen K, Quirynen M (2013) Tolerance within the sleeve inserts of different surgical guides for guided implant surgery. Clin Oral Implants Res 24(6):630–634PubMedCrossRef
27.
Zurück zum Zitat Schneider D, Marquardt P, Zwahlen M, Jung RE (2009) A systematic review on the accuracy and the clinical outcome of computer-guided template-based implant dentistry. Clin Oral Implants Res 20(Suppl 4):73–86PubMedCrossRef Schneider D, Marquardt P, Zwahlen M, Jung RE (2009) A systematic review on the accuracy and the clinical outcome of computer-guided template-based implant dentistry. Clin Oral Implants Res 20(Suppl 4):73–86PubMedCrossRef
28.
Zurück zum Zitat Huang HM, Chiu CL, Yeh CY, Lee SY (2003) Factors influencing the resonance frequency of dental implants. J Oral MaxillofacSurg 61(10):1184–1188CrossRef Huang HM, Chiu CL, Yeh CY, Lee SY (2003) Factors influencing the resonance frequency of dental implants. J Oral MaxillofacSurg 61(10):1184–1188CrossRef
29.
Zurück zum Zitat Chong L, Khocht A, Suzuki JB, Gaughan J (2009) Effect of implant design on initial stability of tapered implants. J Oral Implantol 35(3):130–135PubMedCrossRef Chong L, Khocht A, Suzuki JB, Gaughan J (2009) Effect of implant design on initial stability of tapered implants. J Oral Implantol 35(3):130–135PubMedCrossRef
30.
Zurück zum Zitat Dos Santos MV, Elias CN, Cavalcanti Lima JH (2011) The effects of superficial roughness and design on the primary stability of dental implants. Clin Implant Dent Relat Res 13(3):215–223PubMedCrossRef Dos Santos MV, Elias CN, Cavalcanti Lima JH (2011) The effects of superficial roughness and design on the primary stability of dental implants. Clin Implant Dent Relat Res 13(3):215–223PubMedCrossRef
31.
Zurück zum Zitat Dhore CR, Snel SJ, Jacques SV, Naert IE, Walboomers XF, Jansen JA (2008) In vitro osteogenic potential of bone debris resulting from placement of titanium screw-type implants. Clin Oral Implants Res 19(6):606–611PubMedCrossRef Dhore CR, Snel SJ, Jacques SV, Naert IE, Walboomers XF, Jansen JA (2008) In vitro osteogenic potential of bone debris resulting from placement of titanium screw-type implants. Clin Oral Implants Res 19(6):606–611PubMedCrossRef
32.
Zurück zum Zitat Vodicka P, Smetana K Jr, Dvorankova B, Emerick T, Xu YZ, Ourednik J et al (2005) The miniature pig as an animal model in biomedical research. Ann N Y Acad Sci 1049:161–171PubMedCrossRef Vodicka P, Smetana K Jr, Dvorankova B, Emerick T, Xu YZ, Ourednik J et al (2005) The miniature pig as an animal model in biomedical research. Ann N Y Acad Sci 1049:161–171PubMedCrossRef
33.
Zurück zum Zitat Swindle MM, Makin A, Herron AJ, ClubbFJ J, Frazier KS (2012) Swine as models in biomedical research and toxicology testing. Vet Pathol 49(2):344–356PubMedCrossRef Swindle MM, Makin A, Herron AJ, ClubbFJ J, Frazier KS (2012) Swine as models in biomedical research and toxicology testing. Vet Pathol 49(2):344–356PubMedCrossRef
34.
Zurück zum Zitat Pearce AI, Richards RG, Milz S, Schneider E, Pearce SG (2007) Animal models for implant biomaterial research in bone: a review. Eur Cell Mater 13:1–10PubMed Pearce AI, Richards RG, Milz S, Schneider E, Pearce SG (2007) Animal models for implant biomaterial research in bone: a review. Eur Cell Mater 13:1–10PubMed
35.
Zurück zum Zitat Saha S, Pal S, Albright JA (1982) Surgical drilling: design and performance of an improved drill. J Biomech Eng 104(3):245–252PubMedCrossRef Saha S, Pal S, Albright JA (1982) Surgical drilling: design and performance of an improved drill. J Biomech Eng 104(3):245–252PubMedCrossRef
37.
Zurück zum Zitat Han S, Rho J, Medige J, Ziv I (1996) Ultrasound velocity and broadband attenuation over a wide range of bone mineral density. OsteoporosInt 6(4):291–296CrossRef Han S, Rho J, Medige J, Ziv I (1996) Ultrasound velocity and broadband attenuation over a wide range of bone mineral density. OsteoporosInt 6(4):291–296CrossRef
38.
Zurück zum Zitat Ansell RH, Scales JT (1968) A study of some factors which affect the strength of screws and their insertion and holding power in bone. J Biomech 1(4):279–302PubMedCrossRef Ansell RH, Scales JT (1968) A study of some factors which affect the strength of screws and their insertion and holding power in bone. J Biomech 1(4):279–302PubMedCrossRef
39.
Zurück zum Zitat Siu WS, Qin L, Cheung WH, Leung KS (2004) A study of trabecular bones in ovariectomized goats with micro-computed tomography and peripheral quantitative computed tomography. Bone 35(1):21–26PubMedCrossRef Siu WS, Qin L, Cheung WH, Leung KS (2004) A study of trabecular bones in ovariectomized goats with micro-computed tomography and peripheral quantitative computed tomography. Bone 35(1):21–26PubMedCrossRef
40.
Zurück zum Zitat Shuaib I, Hillery M (1995) Forces generated in guide-wires when drilling human bone. ProcInst MechEng H 209(3):157–162CrossRef Shuaib I, Hillery M (1995) Forces generated in guide-wires when drilling human bone. ProcInst MechEng H 209(3):157–162CrossRef
41.
Zurück zum Zitat Majumdar S (2003) Current technologies in the evaluation of bone architecture. CurrOsteoporos Rep 1(3):105–109 Majumdar S (2003) Current technologies in the evaluation of bone architecture. CurrOsteoporos Rep 1(3):105–109
42.
Zurück zum Zitat Bonfield W, Li CH (1968) The temperature dependence of the deformation of bone. J Biomech 1(4):323–329PubMedCrossRef Bonfield W, Li CH (1968) The temperature dependence of the deformation of bone. J Biomech 1(4):323–329PubMedCrossRef
43.
Zurück zum Zitat Martinez H, Davarpanah M, Missika P, Celletti R, Lazzara R (2001) Optimal implant stabilization in low density bone. Clin Oral Implants Res 12(5):423–432PubMedCrossRef Martinez H, Davarpanah M, Missika P, Celletti R, Lazzara R (2001) Optimal implant stabilization in low density bone. Clin Oral Implants Res 12(5):423–432PubMedCrossRef
44.
Zurück zum Zitat O’Sullivan D, Sennerby L, Meredith N (2004) Influence of implant taper on the primary and secondary stability of osseointegrated titanium implants. Clin Oral Implants Res 15(4):474–480PubMedCrossRef O’Sullivan D, Sennerby L, Meredith N (2004) Influence of implant taper on the primary and secondary stability of osseointegrated titanium implants. Clin Oral Implants Res 15(4):474–480PubMedCrossRef
45.
Zurück zum Zitat Scarano A, Degidi M, Iezzi G, Petrone G, Piattelli A (2006) Correlation between implant stability quotient and bone-implant contact: a retrospective histological and histomorphometrical study of seven titanium implants retrieved from humans. Clin Implant Dent Relat Res 8(4):218–222PubMedCrossRef Scarano A, Degidi M, Iezzi G, Petrone G, Piattelli A (2006) Correlation between implant stability quotient and bone-implant contact: a retrospective histological and histomorphometrical study of seven titanium implants retrieved from humans. Clin Implant Dent Relat Res 8(4):218–222PubMedCrossRef
46.
Zurück zum Zitat Wong M, Eulenberger J, Schenk R, Hunziker E (1995) Effect of surface topology on the osseointegration of implant materials in trabecular bone. J Biomed Mater Res 29(12):1567–1575PubMedCrossRef Wong M, Eulenberger J, Schenk R, Hunziker E (1995) Effect of surface topology on the osseointegration of implant materials in trabecular bone. J Biomed Mater Res 29(12):1567–1575PubMedCrossRef
47.
Zurück zum Zitat Li J (1997) Bone-implant interface and remaining tissues on the implant surface after push-out test: an SEM observation. Biomed Mater Eng 7(6):379–385PubMed Li J (1997) Bone-implant interface and remaining tissues on the implant surface after push-out test: an SEM observation. Biomed Mater Eng 7(6):379–385PubMed
48.
Zurück zum Zitat Sakoh J, Wahlmann U, Stender E, Nat R, Al-Nawas B, Wagner W (2006) Primary stability of a conical implant and a hybrid, cylindric screw-type implant in vitro. Int J Oral Maxillofac Implants 21(4):560–566PubMed Sakoh J, Wahlmann U, Stender E, Nat R, Al-Nawas B, Wagner W (2006) Primary stability of a conical implant and a hybrid, cylindric screw-type implant in vitro. Int J Oral Maxillofac Implants 21(4):560–566PubMed
49.
Zurück zum Zitat Kay JF, Gilman L, May TC (1991) The tri-spade drill for endosseous dental implant installation. J Oral Implantol 17(4):424–428PubMed Kay JF, Gilman L, May TC (1991) The tri-spade drill for endosseous dental implant installation. J Oral Implantol 17(4):424–428PubMed
50.
Zurück zum Zitat Toyoshima T, Wagner W, Klein MO, Stender E, Wieland M, Al-Nawas B (2011) Primary stability of a hybrid self-tapping implant compared to a cylindrical non-self-tapping implant with respect to drilling protocols in an ex vivo model. Clin Implant Dent Relat Res 13(1):71–78PubMedCrossRef Toyoshima T, Wagner W, Klein MO, Stender E, Wieland M, Al-Nawas B (2011) Primary stability of a hybrid self-tapping implant compared to a cylindrical non-self-tapping implant with respect to drilling protocols in an ex vivo model. Clin Implant Dent Relat Res 13(1):71–78PubMedCrossRef
51.
Zurück zum Zitat Beaman FD, Bancroft LW, Peterson JJ, Kransdorf MJ, Menke DM, DeOrio JK (2006) Imaging characteristics of bone graft materials. Radiographics 26(2):373–388PubMedCrossRef Beaman FD, Bancroft LW, Peterson JJ, Kransdorf MJ, Menke DM, DeOrio JK (2006) Imaging characteristics of bone graft materials. Radiographics 26(2):373–388PubMedCrossRef
52.
Zurück zum Zitat Hultin M, Davidson T, Gynther G, Helgesson G, Jemt T, Lekholm U et al (2012) Oral rehabilitation of tooth loss: a systematic review of quantitative studies of OHRQoL. Int J Prosthodont 25(6):543–552PubMed Hultin M, Davidson T, Gynther G, Helgesson G, Jemt T, Lekholm U et al (2012) Oral rehabilitation of tooth loss: a systematic review of quantitative studies of OHRQoL. Int J Prosthodont 25(6):543–552PubMed
53.
Zurück zum Zitat Amorim MA, Takayama L, Jorgetti V, Pereira RM (2006) Comparative study of axial and femoral bone mineral density and parameters of mandibular bone quality in patients receiving dental implants. OsteoporosInt 17(10):1494–1500CrossRef Amorim MA, Takayama L, Jorgetti V, Pereira RM (2006) Comparative study of axial and femoral bone mineral density and parameters of mandibular bone quality in patients receiving dental implants. OsteoporosInt 17(10):1494–1500CrossRef
54.
Zurück zum Zitat Esposito M, Hirsch JM, Lekholm U, Thomsen P (1998) Biological factors contributing to failures of osseointegrated oral implants. (II). Etiopathogenesis. Eur J Oral Sci 106(3):721–764PubMedCrossRef Esposito M, Hirsch JM, Lekholm U, Thomsen P (1998) Biological factors contributing to failures of osseointegrated oral implants. (II). Etiopathogenesis. Eur J Oral Sci 106(3):721–764PubMedCrossRef
Metadaten
Titel
Template-guided vs. non-guided drilling in site preparation of dental implants
verfasst von
Uta Scherer
Marcus Stoetzer
Martin Ruecker
Nils-Claudius Gellrich
Constantin von See
Publikationsdatum
01.07.2015
Verlag
Springer Berlin Heidelberg
Erschienen in
Clinical Oral Investigations / Ausgabe 6/2015
Print ISSN: 1432-6981
Elektronische ISSN: 1436-3771
DOI
https://doi.org/10.1007/s00784-014-1346-7

Weitere Artikel der Ausgabe 6/2015

Clinical Oral Investigations 6/2015 Zur Ausgabe

Parodontalbehandlung verbessert Prognose bei Katheterablation

19.04.2024 Vorhofflimmern Nachrichten

Werden Personen mit Vorhofflimmern in der Blanking-Periode nach einer Katheterablation gegen eine bestehende Parodontitis behandelt, verbessert dies die Erfolgsaussichten. Dafür sprechen die Resultate einer prospektiven Untersuchung.

Invasive Zahnbehandlung: Wann eine Antibiotikaprophylaxe vor infektiöser Endokarditis schützt

11.04.2024 Endokarditis Nachrichten

Bei welchen Personen eine Antibiotikaprophylaxe zur Prävention einer infektiösen Endokarditis nach invasiven zahnärztlichen Eingriffen sinnvoll ist, wird diskutiert. Neue Daten stehen im Einklang mit den europäischen Leitlinienempfehlungen.

Zell-Organisatoren unter Druck: Mechanismen des embryonalen Zahnwachstums aufgedeckt

08.04.2024 Zahnmedizin Nachrichten

Der Aufbau von Geweben und Organen während der Embryonalentwicklung wird von den Zellen bemerkenswert choreografiert. Für diesen Prozess braucht es spezielle sogenannte „Organisatoren“. In einer aktuellen Veröffentlichung im Fachjournal Nature Cell Biology berichten Forschende durch welchen Vorgang diese Organisatoren im Gewebe entstehen und wie sie dann die Bildung von Zähnen orchestrieren.

Die Oralprophylaxe & Kinderzahnheilkunde umbenannt

11.03.2024 Kinderzahnmedizin Nachrichten

Infolge der Umbenennung der Deutschen Gesellschaft für Kinderzahnheilkunde in Deutsche Gesellschaft für Kinderzahnmedizin (DGKiZ) wird deren Mitgliederzeitschrift Oralprophylaxe & Kinderzahnheilkunde in Oralprophylaxe & Kinderzahnmedizin umbenannt. Aus diesem Grunde trägt die erste Ausgabe in 2024 erstmalig den neuen Titel.

Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Zahnmedizin und bleiben Sie gut informiert – ganz bequem per eMail.